AC Current Probe
Model MD301
User Manual
DESCRIPTION
The MD301 (Catalog #1201.07) is designed for use in industrial environments. The hook-shaped jaws
enable the user to “pry” into or “hook” onto cables (will accept 2 x 500kcmil) or even smaller bus bars.
The voltage output (1mV DC/A AC) enables instruments without current ranges to measure, display or log
currents on a DC voltage range. The Model MD301 is compatible with any DC voltmeter, multimeter, or
other voltage measurement instruments with an input impedance greater than 100k.
WARNING
These safety warnings are provided to ensure the safety of personnel and proper operation of the
instrument.
• Read the instruction manual completely and follow all the safety information before attempting
to use or service this instrument.
• Use caution on any circuit: Potentially high voltages and currents may be present and may
pose a shock hazard.
• Read the Safety Specifications section prior to using the current probe. Never exceed the
maximum voltage ratings given.
• Safety is the responsibility of the operator.
• ALWAYS connect the current probe to the display device before clamping the probe onto the
sample being tested.
• ALWAYS inspect the instrument, probe, probe cable, and output terminals prior to use.
Replace any defective parts immediately.
• NEVER use the current probe on electrical conductors rated above 600V in overvoltage
category III (CAT III). Use extreme caution when clamping around bare conductors or bus bars.
INTERNATIONAL ELECTRICAL SYMBOLS
This symbol signifies that the current probe is protected by double or reinforced insulation. Use
only factory specified replacement parts when servicing the instrument.
This symbol signifies CAUTION! and requests that the user refer to the user manual before
using the instrument.
This is a type A current sensor. This symbol signifies that application around and removal
from HAZARDOUS LIVE conductors is permitted.
DEFINITION OF MEASUREMENT CATEGORIES
Cat. II: For measurements performed on circuits directly connected to the electrical distribution system.
.
Examples are measurements on household appliances or portable tools.
Cat. III: For measurements performed in the building installation at the distribution level such as on
hardwired equipment in fixed installation and circuit breakers.
Cat. IV: For measurements performed at the primary electrical supply (<1000V) such as on primary
overcurrent protection devices, ripple control units, or meters
RECEIVING YOUR SHIPMENT
Upon receiving your shipment, make sure that the contents are consistent with the packing list. Notify
your distributor of any missing items. If the equipment appears to be damaged, file a claim immediately
with the carrier and notify your distributor at once, giving a detailed description of any damage.
![](/html/54/5422/54223394b84fa99fe24233100b0650e979b1ea28b3bdbf22707487d255adf9c9/bg2.png)
Ø =1.18" (30 mm)
Cable max.
2.5 x 0.19" (63 x 5 mm)
Bus Bar max.
Ø =0.98" (25 mm)
Dual Cable max.
2.6"
(66 mm)
ELECTRICAL SPECIFICATIONS
Current Range:
2 to 500A AC
Transformation Ratio:
1000:1
Output Signal:
1mV DC/A AC
Overload: 700A for 10mn
I
7.68"
(195 mm)
Black
S2
S1
Red
1.34"
(34 mm)
Accuracy*:
Primary current 25A 100A 250A 500A
Accuracy % 5% 2% 1% 1%
(*Reference conditions: 23°C ± 5°K, 20 to 75%
RH, external magnetic field < 40A/m, no DC
component, no external current carrying conductor,
test sample centered.) 10MΩ load.
Frequency Range: 48 to 1000Hz
(error: add 2% to the ref.)