General Instructions ............................................................................................................................................. 4
Contents of the box ............................................................................................................................................ 4
Front Face .......................................................................................................................................................... 6
Back Face .......................................................................................................................................................... 7
Front Face ........................................................................................................................................................ 12
Back Face ........................................................................................................................................................ 13
Pressing keys for less than < 1s...................................................................................................................17
Pressing keys for more than > 1s.................................................................................................................18
Chapter IV
General commands ............................................................................................................................................. 19
in Normal mode ............................................................................................................................................19
in Version mode............................................................................................................................................19
in Calibration mode.......................................................................................................................................19
in Autotest mode...........................................................................................................................................19
Activating MAIN OUT ....................................................................................................................................... 20
Selection of the instrument function ................................................................................................................. 21
Display of the software version ........................................................................................................................ 22
Saving a configuration (GX 320) ...................................................................................................................... 28
Reloading a configuration (GX 320)................................................................................................................. 29
Clearing a configuration (GX 320).................................................................................................................... 30
I
a
p
t
e
r
I
I
a
p
t
e
r
I
I
I
I
I
C
h
a
p
t
e
r
V
C
h
a
C
h
a
Generation of simple CONTinuous periodical signals.................................................................................... 31
Available output signals.................................................................................................................................... 31
Signal selection ................................................................................................................................................ 31
Adjusting signal frequency ............................................................................................................................... 32
Adjusting signal duty cycle ............................................................................................................................... 35
Adjusting signal amplitude................................................................................................................................ 35
Adjusting offset and DC level ........................................................................................................................... 36
Adjusting signal logical levels........................................................................................................................... 36
Selection of FSK mode..................................................................................................................................... 37
Selection of PSK mode .................................................................................................................................... 37
Selection of piloting source .............................................................................................................................. 37
Adjusting jump frequencies (in FSK mode)...................................................................................................... 38
Adjusting jump phases (in PSK mode)............................................................................................................. 38
Other settings ................................................................................................................................................... 38
SWEEP Frequency scan function...................................................................................................................... 39
Selection of the sweep mode ........................................................................................................................... 39
Selection of the scan source ............................................................................................................................ 40
Adjusting START / END frequencies ............................................................................................................... 40
Adjusting scan period using an INTernal source.............................................................................................. 41
Other settings ................................................................................................................................................... 41
MODUL Modulation function (GX 320 only) ..................................................................................................... 42
Selection of AM / FM mode .............................................................................................................................. 42
Selection of the modulation source .................................................................................................................. 43
Adjustment of the FM START / END frequencies............................................................................................ 43
Other settings ................................................................................................................................................... 43
C
h
a
p
t
e
r
C
h
C
h
C
h
a
C
h
C
h
C
h
a
C
h
a
C
h
a
V
a
p
t
e
r
V
a
p
t
e
r
V
p
t
e
r
V
a
a
p
p
I
p
t
e
r
V
I
p
t
e
r
V
I
t
e
r
V
I
I
t
e
r
V
I
p
I
t
e
r
V
I
I
I
I
I
I
I
I
I
I
I
C
h
a
p
t
e
r
I
C
h
C
h
FREQ Frequency meter function ....................................................................................................................... 44
Activating signal generation (MASTER) .......................................................................................................... 47
Other settings .................................................................................................................................................. 48
C
h
a
C
h
C
h
GATE function (GX 320 only) ............................................................................................................................. 50
Selection of the BURST source....................................................................................................................... 51
Setting the number of pulses........................................................................................................................... 52
Setting the generation time for INTernal source.............................................................................................. 52
Manual triggering in EXTernal source ............................................................................................................. 52
Other settings .................................................................................................................................................. 52
Introduction You have just purchased a GX 305, GX 310 or GX 320 Function Generator
and we appreciate your confidence.
Content of the box
Precautions
Safety measures
•the generator
•the safety notice
•the power supply cable
•the USB A/B cable for the programmable versions
•the ETHERNET cable for the GX 320E
•the CD-ROM containing:
the operating guide in 5 languages
the programming in 2 languages
the USB ‘CP210x USB to UART Bridge Controller’ Drivers
the LabView and LabWindows Drivers
the USBxPress application (USB port identification)
the GX320E-Admin (IP address programming)
To obtain the best service:
- read this notice carefully,
- respect the safety instructions.
Failure to respect the warnings and/or usage instructions may damage the
device and/or installations and may be dangerous for the user.
This instrument complies with the NF EN 61010-1 - Ed. 2 (2001) safety
standard relating to the safety of electric measurement devices.
Definition of
installation
categories
• It is designed for indoor use in an level 2 pollution environment at an altitude
of less than 2000 m, a temperature between 0° C and 40° C and a RH
(relative humidity) of less than 80% up to 40° C.
• The MAIN OUT, SWEEP OUT, TTL OUT outlets are referenced t
o earth and
protected from accidental voltages that are not in excess of 60 V DC or
40 V AC.
• The FREQ EXT entry can only be used for measurements on Category 1
installations and for voltages not exceeding 300 V in relation to the earth.
• Mains power supply: 120 V
CAT II:
Category II corresponds to measurements on circuits that are directly
connected to low voltage installations.
Example: power supply for household appliances and portable tools
CAT III:
Category III corresponds to measurements on the building
installation.
Example: power supply for industrial machinery or devices.
CAT IV:
Category IV corresponds to measurements at the source of the low
Warning: potential hazard, refer to the operating guide.
Selective waste sorting for recycling electric and electronic waste. In
compliance with the WEEE 2002/96/EC directive:
the device should not be considered as household waste.
Earth terminal
Alternating signal
This equipment is guaranteed for all manufacturing and parts defects in
compliance with the general terms and conditions which are available on request
During the warranty period (3 years), the instrument may only be repaired by the
manufacturer who reserves the right to make the decision to either repair or
replace all or part of the appliance. In the event of a return of the equipment to
the manufacturer the shipping charge from the customer to the manufacturer is
at the customer’s expense.
Indication of a key double function when pressed for mo
second
USB symbol
re than 1
Maintenance,
repairs, metrological
checks
Cleaning
The guarantee does not apply in the following conditions:
• inappropriate use of the equipment or use with incompatible equipment
• one or more changes made to the equipment without prior explicit
authorisation from the manufacturer’s technical department
• an intervention is made on the instrument by a person n
manufacturer
• the adapting to a specific application that is not part of the definition of the
instrument or in the operating guide
• damage caused by a mechanical shock, by dropping the instrument or by
flooding.
The device includes no parts that can be replaced by the operator. All operations
must be carried out by competent approved personnel.
For checks and calibrations, contact one of our accredited metrology laboratories
(information and contact details available on request), at our Chauvin Arnoux
subsidiary or the branch in your country.
No interventions are authorised inside the instrument.
- Turn the instrument off (remove the power supply cable).
- Clean using a damp cloth and soap.
ot approved by the
- Never use abrasive products or solvents.
- Dry quickly using a dry cloth or an air blower at max. 80° C.
The GX 305 and GX 310 are alternating standard form wave generators, using the
DDS (Direct Digital Synthesis) technology. They may simulate the operation and
specifications of various electronic systems.
They also include a frequency meter input.
The GX 310P is a generator that can be programmed remotely via an USB link.
LOGIC waveform selected:
Adjustment of the high or low signal level using the wheel
FUNCTION keys:
Selection of one of the three available functions
SWEEP selection of the INTernal or EXTernal command signal
• SWEEP activated:selection of LIN or LOG sweep
• calibration:triggering of the selected adjustment step
• Autotest:run the selected test
SWEEP function activated in INT: assignment of the desired duration setting
for carrying out the sweep using the wheel.
Then, by pressing several times, selection of the digit on which to increment.
Division or multiplication by 10 of the current frequency value (decade change)
•Assignment of frequency adjustment to the wheel.
Then, by pressing several times, selection of the digit on which to
increment.
•SWEEP function activated: same functions with Freq
The GX 320 is a standard alternating signal generator, using the DDS (Direct Digital
Synthesis) technology. It may simulate the operation and the specifications of various
electronic systems. It also includes a frequency meter input. The GX 320E is remote
programmable via an USB or ETHERNET link.
Pressing for more than 1 second sets the GATE function.
Pressing for more than 1 second forces the duty cycle to 50 %.
Pressing the key for more than 1 second switches from a peak to peak
amplitude display to an RMS (root mean square) display.
Pressing the key for more than 1 second forces the offset value to 0.
LOGIC LEV EL
LOGIC LEVE L
LOGIC LEVE LLOGIC LEVE L
Pressing the key for more than 1 second assigns the LCD contrast adjustment
to the wheel.
GX 320 Description
BURST function activated, INTernal source. Pressing the key for more than 1
second is used to switch the number of pulses Num in the pulse generation
period Time, and vice versa.
These keys assign the selected frequency to the start or end of the current
range.
Ranges
[0.001 Hz ; 0.01 Hz] 0.001 Hz 0.01 Hz
[0.01 Hz ; 0.1 Hz] 0.01 Hz 0.1 Hz
[0.1 Hz ; 1 Hz] 0.1 Hz 1 Hz
[1 Hz ; 10 Hz] 1 Hz 10 Hz
[10 Hz ; 100 Hz] 10 Hz 100 Hz
[100 Hz ; 1 kHz] 100 Hz 1 kHz
[1 kHz ; 10 kHz] 1 kHz 10 kHz
[10 kHz ; 100 kHz] 10 kHz 100 kHz
Press > 1 Second
‘RANGE-’
Press > 1 second
‘RANGE+’
[100 kHz ; 1 MHz] 100 kHz 1 MHz
[1 MHz ; 10 MHz] 1 MHz 10 MHz
[10 MHz ; 20 MHz] 10 MHz 20 MHz
For the SWEEP or MODULFM or FSK functions pressing the key for more
than 1 second is used to switch between Freq
Check that your instrument is compatible with the mains network voltage
(see the label at the back of the instrument), that the power supply cable
is not damaged and that it is earthed.
The power supply cable plug is used as a cut off point, connect the device
to a mains outlet that is easily accessible and is earthed in order to
ensure safety.
Four start-up modes are possible depending on the key
keys - used:
1. Normal Mode:
The instrument starts up using the last used configuration. By default the
factory configuration is restored.
The key becomes:
2. Version Mode:
– or combination of
FUNCTION
+
MAIN OUT
+
The instrument starts up in Version mode and displays the current software
version number and date.
+
The key becomes: (See Display of the software version).
3. Calibration Mode
The instrument starts up in Calibration mode with the selection of the
calibration to be run: automatic mode CAL_AU, by default.
The key becomes: (See automatic calibration).
4. Autotest Mode:
The instrument starts up in Autotest mode with the selection of the test to
be run: automatic mode tSt_AU by default.
Whatever the mode, pressing this key puts the instrument on STANDBY.
When pressed while in Normal mode the context is saved:
- current settings in use for signal generation when the instrument was
stopped,
- settings for other functions that may have been changed.
The key becomes:
Each time Normal mode start-up is used all the settings are reloaded.
In the event of a power failure (or if the power cable is unplugged …), the
instrument restarts using the last backup (backup made the last time the device
was turned off using the ON/STANDBY key).
In the event of an error the default configuration is loaded:
• Signal sinusoidal
• Function CONTinuous
• Frequency 1 kHz
• Amplitude 1 Vpp
• Offset 0 V
• Output MAIN OUT ON not active
• No adjustments assigned to the wheel.
Activating the
N OUT terminal
MAI
MAIN OUT
MAIN OUT
The key becomes: .
At start-up the MAIN OUT terminal is always de-activated.
Pressing the key activates the terminal and the key lights: .
On the GX 320: the key may blink when the GATE function is activated (see
GATE function).
De-activation of the MAIN OUT terminal, the key is no longer lit: .
n the 4 seconds following the function validation the wheel is
General Command (contd.)
Setting the screen
contrast
LOGIC LEVEL
1s or 1s
Selection of the
instrument function
FUNCTION
The display shows:.
The key becomes:
Adjustment of the contrast value from 0 to 99 using the coding wheel.
Exiting from this mode is made by pressing another key. The frequency display
returns to the screen and the possible associated keys blink.
The key becomes: .
The contrast value is memorised in the device configuration once it is turned off
(see left margin) or when the configuration is saved (GX 320).
Pressing once displays the list of functions available on the device in the top
.
FUNCTION
right hand corner:
The cursor indicates the selected function.
Pressing again moves the cursor towards the top or bottom to select another
function.
If, after 2 seconds, no keys have been pressed or when another key is pressed,
the selected function is validated and is the only one remaining displayed:
When the function has been validated the keys that can be assigned to the
wheel blink until one of them is selected; the key then lights up.
If no keys are used i
automatically assigned to frequency setting
(Freq or Freq
The device has an automatic function that can be used to calibrate signal
generation.
This function can be triggered:
- automatically (all settings are run automatically) or
- manually (individual selection and run of settings).
No specific wiring is needed for this function.
For optimal calibration the device must be at operating temperature
(switched on for 30 minutes) before running calibration.
In addition, when in manual mode it is recommended to respect the
running order of the calibration steps.
Entry into this mode is the CAL.AU. automatic mode. The display is as below:
Switching to manual mode is done by turning the wheel and selecting the
calibration step to be run individually.
The device has an automatic electronics test function. This feature can be run
automatically (all tests run automatically) or manually (individual selection and
running of tests).
These tests require specific wiring of the device’s input/output terminals.
Two wirings are needed.
When needed they are indicated by the following messages:
for wiring n° 1:
(GX 305/310)
(GX 320)
(GX 305/310)
(GX 320)
for wiring n° 2:
(GX 305/310)
(GX 320)
Once the wiring has been made pressing the key continues the test.
The GX 320 can save and reload user configurations.
A total of up to 15 files can be saved.
This backup is permanent (the data is saved even if the instrument is powered
down).
Enters the configuration management mode.
is displayed on the screen with the current file number:
General commands
if file 3 is empty;
if file 3 already contains a configuration
the data (other than frequency) is displayed on the screen.
Pressing another key than or exits the mode without saving.
Selects files from SEt.01 to SEt.15. The screen is updated with the data from
the selected file.
Pressing the key again saves the current configuration in the selected file.
The display returns to its pre-backup status and the display is cleared.
When saving the content of the selected file is overwritten by the content
of the current configuration without any warning messages.
Each time a key is pressed the symbol is displayed on the screen and the keys
that can be assigned to the wheel blink.
Pressing once displays the list of available waveforms at the top left of the screen:
The cursor indicates the current waveform.
Pressing again moves the cursor up or down to select a new
waveform.
If the keys are not pressed for 2 seconds or if another key is pressed the
selected waveform is validated and remains displayed:
When the waveform is validated the keys that can be assigned to
the wheel blink until one of them is selected; this key is then lit.
If no key is pressed within 4 s of validation the wheel is
automatically assigned to frequency adjustment (Freq or
Freq
Linear rule between
the command signal
and the generated
frequency
Logarithmic rule
between the command
signal and the
generated frequency
= 1 kHz, F
start
= 100 kHz
end
A SWEEP OUT signal is available on the VCF IN BNC (GX 305/310)or VCG IN (GX 320). It is a proportional signal with a generated frequency, amplitude from 0 to 2V.
The generated output frequency is proportional (according to a linear or
logarithmic rule) to the voltage on VCF IN (GX 305/310) or VCG IN (GX 320).
The command signal is sampled on 8 bits using a frequency of 60 kHz.
For -10 V: the output frequency F ≅ FreqFor 10 V: F ≅ Freq
END
START
Selecting the scan
source
Setting the
START / END
frequencies
or
1s or1s
Pressing the key successively selects the INTernal source or the
EXTernal source .
Displays Freq
and assigns adjustment to the wheel.
START
The key lights: .
Displays Freq
and assigns adjustment to the wheel.
END
The key lights: .
Pressing successively selects the digit from which the increment will apply.
Selecting the FREQ function activates measurement of the frequency of the
signal input to the FREQ EXT terminal.
The frequency meter can measure frequencies from 5 Hz to 100 MHz with the
following precision:
< 50 mV sensitivity F ≤ 30 MHz
< 60 mV sensitivity for 30 MHz < F ≤ 80 MHz
< 90 mV sensitivity for 80 MHz < F ≤ 100 MHz
The maximum amplitude (*) of the measured signal is:
300 V sensitivity from 5 Hz to 5 kHz
30 V sensitivity from 5 kHz to 1 MHz
10 V sensitivity above this value
(*) signal with a 50% duty cycle.
Measurement stabilisation time depends on the input frequency:
≤ 1 s from 5 to 20 Hz (≥ 1 measurements per second)
≤ 100 ms from 20 to 400 Hz (2 measurements per second)
≤ 40 ms from 400 Hz to 100 MHz (2 measurements per second)
Connections
Indication of the 300 V protection (50 - 60 Hz) CAT I
The SYNC function is used to synchronise several GX 320 set up in a cascade
in order to create a variable phase multiple signal generator.
The frequency resolution of this function is: 37 mHz, the clock frequency of the
DDS is set at 10 MHz.
To limit the sampling effect the maximum frequency of the output signal is set at
100 kHz.
The Master generator supplies the Slave generators with the clock (Clk) used to
generate the signals (10 MHz) and a synchronisation signal (Ctrl). This allows
all the generators to start at the same time and control their phase offset.
Connections
Control signal (Ctrl): Connect the Slave VCG IN BNCs to the Master.
Clock signal (Clk): Connect the slave FREQ_EXT BNCs to the Maste
TTL OUT.
ring signal generation disconnecting one of the Ctrl or Clk cables
Du
desynchronises the generators.
To resynchronise them use the master’s ‘MAIN OUT ON/OFF’ key to
deactivate and then reactivate signal generation.
www.GlobalTestSupply.com
SYNC Function
SYNC Synchronisation Function (GX 320 only, contd.)
Selection of the
Slave / Master mode
Pressed successively selects S mode (Slave):
or M (Master):
Adjusting phase
offset
or
1s
Phase offset can be set on the master and on the slave (if it is not locked).
Whatever the mode selected (M/S) the phase offset is that of the slave(s)
related to the master.
The phase offset entered on the master is applied to all the slaves whereas the
phase offset entered on the slave is limited to that slave:
Displays phase offset and assigns the adjustment to the coding wheel.
The key lights.
Value adjustment.
The phase is in degrees and can have values between -180° C and
+180° C, varying by 1°.
The master mode phase is reversed in relation to the slave mode.
ster all adjustments are possible in real time because each
selecting waveform and adjusting frequency and phase are
SYNC Synchronisation Function (GX 320 only, contd.)
Activation of signal
generation (Master)
MAIN OUT
On the ma
change in the master commands resynchronisation of all the slaves.
As this is not possible on the slaves, changing waveform, frequency
or phase are not possible when signal generation is activated. On the
other hand as amplitude and offset have no effect on synchronisation
they remain adjustable at all times.
The slave is said to be locked: is displayed on the top right hand
corner of the slave screen(s).
In order to change the waveform, frequency or phase on the slave
you must stop signal generation on the master using its ‘MAIN OUT
ON/OFF’ key.
•on the Master:
- Activates MAIN OUT and signal generation on all devices on which MAIN
OUT is activated.
The master key lights: .
- Locking slaves:
no longer possible on the slaves.
The symbol is dispayed on the slave screen as below:
MAIN OUT
Other settings
•on the slaves:
- Activation of the associated MAIN OUT (effective signal output is only
possible if signal generation is activated on the master).
The slave key lights: .
•on the Master:
- Deactivation of MAIN OUT and halting of signal generation on all devices.
The master key is turned off: .
- The master frees the slaves: waveform selection and frequency and phase
adjustment are now possible.
The symbol disappears from the slaves.
•on the Slaves: the associated MAIN OUT is deactivated.
To synchronise multiple frequencies the values programmed in the DDS must
also be multiple.
We are here faced with the problem of calculation rounding and programming
resolution: it is highly probable that the direct entry on F on the master and n*F
on the slaves will not give synchronous signals.
The DDS is programmed using a 28 bit register and is piloted by a 10 MHz
clock (in the SYNC function).
The DDS frequency resolution for this function is therefore:
10 MHz / 228 = 0.037 Hz, which means that for a frequency F entered the
resulting frequency is F ± 18.5 MHz.
The formula relating the user entered frequency to the value programmed in the
DDS is the following:
Val
= ENT((Frequency
DDS
with: ENT( ) function returning the whole part of the value
SYNC Synchronisation Function (GX 320 only, contd.)
Thus when you programme a frequency of 100 Hz, the programmed value is:
ENT((100*228)/107 + 0.5) = 2684 which is the equivalent of a frequency of
99.987 Hz (obtained by reverse calculation).
If you wish to programme a synchronous n*100 Hz multiple frequency you must
enter a frequency which results in a DDS programmed value of n*2684, or a
true frequency equal to n*99.987 Hz.
In our example we will generate a square 100 Hz signal using its first three
harmonics: 3 sinusoids with a frequency of 100 Hz, 300 Hz and 500 Hz and an
amplitude of A, A/3 and A/5.
For this example 3 GX 320 generators are needed:
- one Master: on which the SINE waveform is selected, amplitude 20 Vpp,
a null offset, a null phase and a frequency of 100 Hz
(or 99.987 Hz).
- Slave 1: on which the SINE waveform is selected, amplitude 6.7 V, a null
offset, a null phase and a frequency of 3*99.987 = 299.96 Hz.
- Slave 2: on which the SINE waveform is selected, amplitude 4 V, a null
offset, a null phase and a frequency of 5*99.987 = 499.93 Hz.
Connect the generators as shown in the Connections paragraph, activate the
ve outputs and then the master output (to ensure synchronisation do a
sla
master MAIN OUT OFF and then ON).
On the oscilloscope connect the device MAIN OUT (respectively Master, Slave1
and Slave 2) outputs on channels 1, 2, 3.
Select the same sensitivity of 5 V/div. on each channel (choose the weakest
frequency signal as trigger: channel 1).
On channel 4 carry out the sum of Channel1 + Channel2 + Channel3, and
observe the result:
A square signal forms: the higher the number of odd harmonics the better the
signal quality obtained.
This function is only available with “CONT”, “SWEEP” and “MODUL”.
It superimposes a stop command for the alternating MAIN OUT signal
component over the current function, piloted by a TTL input to the BNC
FREQ EXTGate in terminal.
When the TTL signal is at the 1 logical level (5 V), the alternating
component of the MAIN OUT terminal is cut.
At 0 level it is generated freely.
The GATE has no effect on the direct component of the signal.
The command takes effect in approximately 100 ns.
GATE Function
Channel 1: Main Out (sine, 1 kHz, 10 Vpp)
Channel 2: TTL Out
Channel 3: Gate In (LOGIC, 300 Hz, 10 V - 0 V)
Connections
GATE activation
MAIN OUT
1s
MAIN OUT
1s
GATE de-activation
MAIN OUT
1s
Main output
Function activation, the
TTL OUT
output
indication is displayed, the MAIN OUT
GATE c
signal
terminal remains activated
The key blinks:
.
Pressing the key for more than 1 second does not activate the MAIN OUT
terminal but only the GATE function :the key remains unlit.
Function is de-activated and the is cleared, the MAIN OUT terminal
remains active.
Remote programming (programmable version only) (contd.)
The programming instructions respect the IEEE 488-2 standard and the
SCPI (Standard Commands for Programmable Instruments) protocol.
The user has the possibility of having complete remote control of the device.
For more information please consult the programming guide.
Communication
interfaces
ETHERNET
Connection
USB
Connecting the generator to a PC is done using either :
•an A/B type USB cable via a USB to UART converter,
•via ETHERNET Warn
If the CP210x driver is correctly installed on the PC the USB peripheral will
be recognised and a new COM port will appear in the PC’s system settings
(see the programming guide to install).
The new COM port is configured as follows:
- speed: 19200 bauds
- data bits: 8
- parity: none
- stop bit: 1
- protocol: hardware (RTS / CTS)
Once the IP address has been programmed through the GX320E-Admin
application, the GX320E can be accessed via this address.
Remote programming (programmable version only) (contd.)
Remark
LabViews Driver
LabWindows Driver
All the devices GX 310 or GX 320 – whether programmable or not –
respond to the IEEE488.2 *idn? command which returns the device
identification and version.
<programmable> ‘P’ if the GX 310 device is programmable
‘E’ if the GX 320 device is programmable
<firmware version> software version
<version date> software version date
<serial number> device serial number
<NL> CR character (code ASCII 13 or 0x0D)
The GX 310P and GX 320E device drivers for LabWindows and LabView
are available on the CD-ROM which contains this guide.
They can be used to interface SCPI commands for these development
environments.
triangle
square
logic pulses (programmable high and low levels)
positive pulses (TTL level)
continuous (DC: offset)
•GX 305 : from 0.001 Hz to 5 MHz in 10 ranges (decades)
GX
310 : from 0.001 Hz to 10 MHz in 10 ranges (decades)
GX 320 : from 0.001 Hz to 20 MHz in 11 ranges (decades)
•3 internal ranges, for DDS resolution:
F ≤ 1 kHz the DDS resolution is approx. 1 mHz
1 kHz < F ≤ 10 kHz the DDS resolution is approx. 10 mHz
10 kHz < F ≤ 20 MHz the DDS resolution is approx. 280 mHz
• Frequency display on the LCD: 5 digits (units: Hz, kHz, MHz)
• Settings: direct using the encoder, automatic range switching
• Precision: ± 30 ppm for F < 10 kHz
± 20 ppm for F ≥ 10 kHz
for sine, square, LOGIC and triangle (duty cycle 50 %)
• Temperature ratio: ± 20 ppm / ° C
• Long term derivation: ± 5 ppm / an
MAIN OUT signal • Adjustable amplitude in open circuit: from 0 to 20 Vpp
Precision: from 0.1 to 20 Vpp < 5 % from 1 mHz to 10 MHz
± 1,5 dB for F > 10 MHz (± 0.5 dB typical)
• Impedance : 50 Ω ±
• DC offset voltage: adjustable from -10 V to +10 V in open circuit (OFFSET).
•≤ 1 s from 5 Hz to 20 Hz(≥ 1 measurement per second)
•≤ 100 ms from 20 Hz to 400 Hz(2 measurements per second)
•≤ 40 ms from 400 Hz to 100 MHz(2 measurements per second)
• 1 MΩ // 22 pF approx.
• Max voltage. : 300 V (50 - 60 Hz) CAT I
www.GlobalTestSupply.com
General Specifications
General Specifications
Environment
• Reference temperature 23° C ± 5° C 45 to 65 % RH
• Nominal usage range 5° C to 35° C 45 to 65 % RH
• Operating temperature 0° C to 40° C 20 to 80 % RH
• Storage temperature -20° C to + 70° C 10 to 95 % RH
• Use indoor
• Altitude < 2000 m
• Relative Humidity < 80 % up to 31° C
Power supply
Mains
Safety
CEM
• Voltage 120 V
• Frequency 50 - 60 Hz
• Consumption 20 VA max.
• Removable power supply cable
This device has been designed in compliance with the current CEM standards
and its compatibility has been tested in compliance with the following
standards:
Operating instructions in 5 languages
Programming instructions in 2 languages
USB ‘CP210x USB to UART Bridge Controller’ USB Drivers
LabView and LabWindows Drivers
USBxPress application (USB port identification)
GX320E-Admin application (IP address application)