Chauvin Arnoux®, Inc. d.b.a. AEMC® Instruments
certifies that this instrument has been calibrated
using standards and instruments traceable to
international standards.
We gu
arantee that at the time of shipping your
instrument has met its published specifications.
An NIST traceable certificate may be
requested at the time of purchase, or obtained
by returning the instrument to our repair and
calibration facility, for a nominal charge.
recommended calibration interval for this
The
instrument is 12 months and begins on the date of
receipt by the customer. For recalibration, please
use our calibration services. Refer to our repair
and calibration section at .
These instruments comply with safety standard EN 61010-1 (Ed 2-2010) or EN
61010-2-032 (2002) for voltages and categories of installation, at an altitude
below 2000m and indoors, with a degree of pollution at most equal to 2.
•
Do not use in explosive atmosphere or in the presence of flammable gases
or fumes.
•
Do not use on voltage networks greater than listed on the instrument label.
•
Observe the maximum voltages and intensities assigned between terminals
and earth.
•
Do not use it if appears damaged, incomplete or improperly closed.
•
Before each use, check the condition of the insulation of cables, case and
accessories. Anything which appears damaged (even partially) must be
reported for repair or scrapping.
•
Use leads and accessories of voltages and categories at least equal to those
of the instrument.
•
Observe the environmental conditions of use.
•
Do not modify the instrument and use only original replacement parts.
Repairs or adjustments must be performed by authorized personnel.
•
Replace the batteries when the “Low Bat” LED is blinking.
•
Disconnect all cables from the instrument or remove the clamp on from the
cable before opening the access door to the batteries.
•
Use protective equipment when security conditions are required.
•
Keep your hands away from unused terminals of the device.
•
When handling probes, probe tips, current sensors and alligators clips,
keep fingers behind the guards
•
For measurements of dangerous voltages, first connect the black lead to the
black terminal of the instrument and then connect this lead to the low voltage point of the source to measure (potential closer to Earth). Then connect
the red lead to the red terminal of the instrument and connect the lead to
the hot source to measure. Disconnection must be done in reverse order:
first disconnect the hot lead and then the black lead.
Upon receiving your shipment, make sure that the contents are consistent
with the ordering information. Notify your distributor of any missing items.
If the equipment appears to be damaged, file a claim immediately with the
carrier and notify your distributor at once, giving a detailed description of
any damage. Save the damaged packing container to substantiate your
claim.
1.4Ordering Information
Simple Logger
(4-Channel, TRMS, Bluetooth, 0-1V
Includes Bluetooth USB adapter, mounting hardware, DataView® software, 4x1.5V C-cell
alkaline batteries and user manual. Probes purchased separately.
®
II Model L104 ........................................... Cat. #2126.10
AC)
Simple Logger® II Model AL834 ....................................... Cat. #2126.14
(4-Channel, TRMS, Bluetooth, Captive AmpFlex
Includes Bluetooth USB adapter, mounting hardware, DataView® software, 4x1.5V C-cell
alkaline batteries and user manual.
®
300/3000AAC)
Simple Logger® II Model L264 ......................................... Cat. #2126.23
(4-Channel, TRMS, Bluetooth, 600V
Includes Bluetooth USB adapter, set of 5 color-coded (4-red/1-black) 5 ft leads with colorcoded alligator clips, twelve color-coded input ID markers, mounting hardware, DataView
software, 4x1.5V C-cell alkaline batteries and user manual.
AC/DC)
Simple Logger® II Model L564 ......................................... Cat. #2126.36
(4-Channel, TRMS, Bluetooth, Voltage & Current)
Includes Bluetooth USB adapter, set of 3 color-coded (red/black/white) 5 ft leads with colorcoded alligator clips, twelve color-coded input ID markers, mounting hardware, DataView
software, 4x1.5V C-cell alkaline batteries and user manual. Probes purchased separately.
Simple Logger® II Model ML914 ...................................... Cat. #2126.40
(4-Channel, TRMS, Bluetooth, Captive MiniFlex
Includes Bluetooth USB adapter, mounting hardware, DataView® software, 4x1.5V C-cell
alkaline batteries and user manual.
Throughout the content of this user manual, the Simple Logger® II may be
referred to as the “SLII”.
2.1 Description
The Simple Logger® II Bluetooth Series of loggers are four channel
recording devices powered by alkaline batteries.
Frequency tracking is performed such that 64 samples over one line cycle
are taken. Frequency tracking is performed around the nominal frequency
of 50/60Hz. Harmonic measurements are calculated from these 64
samples (harmonics are only available from the Simple Logger II Control
Panel within the DataView
The Simple Logger
Samples are taken at a user defined storage rate. This means that the
input(s) are ignored between measurement intervals.
®
®
software).
II records at a rate of up to eight times per second.
The SLII’s battery operation allows operation without the need for external
power. A series of front panel LEDs provide a quick status of the logger’s
state, memory usage and low battery indication.
The main advantage of the logger is its ability to perform a variety of
recording tasks with easy and intuitive setup from a computer using
DataView
Analog information on the input is sampled and converted to a digital
value. This digital signal is processed and stored along with scale and
time information. An integrated Bluetooth module provides for the transfer
of data from the instrument’s internal memory to the computer for analysis.
8
®
software, Bluetooth communication and long battery life.
1. Reset Switch
The RESET switch resets the CPU. To access this switch, disconnect all inputs and remove the battery compartment cover. The reset
switch is located on the exposed PCB (refer to § 3.10).
2. Flash Upgrade Switches
These three switches (accessible from under the battery cover), the
RESET switch and upgrade software are used to recover from a failed
flash upgrade procedure (refer to § 3.1).
3. Female Type Mini-B USB Connector
This connector is located under the battery cover and is used for performing firmware upgrades when available
4. Low Battery Indicator
The instrument contains protection circuitry to prevent it from being
turned on when the battery voltage is below 1.7V.
• The first is used to indicate low battery. The low battery indicator
(Red LED single blink) will blink when the battery voltage drops below
2.2V.
• The second is used to determine when to terminate recordingand turn the unit off. The shutdown threshold is when the battery
voltage drops below 1.7V.
Communication from the instrument to DataView
5. Five LED Indicators
The LEDs on the logger serve two functions: Control Operation and
Status Indication.
• The control operation (when holding down the button) is indicated
with text above each LED.
• The status indication (when the button is not being held down)
associated with each LED is indicated with text below each LED.
• Refer to § 3.1 for detailed descriptions of each LED.
6. CONTROL Button
This button selects the mode of operation. Use this button to start
or stop recordings, erase the memory, clear alarms and turn the
Bluetooth ON/OFF.
®
is disabled below 2.2V.
7. Input (model dependent)
AL834: Four captive AmpFlex® flexible sensor
ML914: Four captive MiniFlex® flexible sensor
L264: Five 4mm banana plugs
L564: Three 4mm banana plugs and two custom CA connectors for
current inputs
L104: Four custom CA connectors for current inputs
The ON/SLEEP state of the SLII can be determined by pressing the
CONTROL button for less than 0.5 seconds. If the instrument is ON, the
status of the instrument will be shown by the LEDs. If the instrument is in
the SLEEP mode, all the status LEDs will light until the CONTROL button
is released. Once released, the status indication resumes.
Control of the instrument is performed by pressing and holding the
CONTROL button until the control LED corresponding to the desired
operation lights. Releasing the CONTROL button when the desired control
LED is illuminated results in the corresponding operation being performed.
When pressing the CONTROL button each LED will light in sequence.
Continuing to press the CONTROL button will result in all LEDs being
off after the last LED lights. If you continue to hold the CONTROL
button down, the sequence will repeat with the first LED. Releasing the
CONTROL button after the last LED turns off and before the first LED turns
on will result in no action being taken by the instrument. This provides a
mechanism to cancel (or ignore) the button press.
The Instrument will turn ON from the SLEEP mode if a recording is scheduled to start.
Single-blinkLow Battery Condition. Batteries should be replaced
www.GlobalTestSupply.com
13
Simple Logger® II (Bluetooth Edition) Series
INSTALL DATAVIEW® BEFORE CONNECTING TO THE COMPUTER.
Overload occurs when any input is 10% above its input range. When the
battery voltage goes below 1.7 volt the instrument will shut down (terminating and saving the recording, if it is recording).
•SLEEP mode: The instrument enters the low power state if the button
is not pressed for one minute. It will remain in this state until either
the button is pressed or the internal clock reaches the start time for a
scheduled recording.
•RECORD mode: The instrument enters the low power state between
sample sets. The slower the storage rate, the greater the portion of
time the instrument is in the low power state. Thus, the slower the
storage rate, the longer the instrument can record.
3.2 Connecting the Simple Logger® II to a Computer
The SLII comes with both USB and Bluetooth connectivity.
The USB connector is located underneath the battery cover and is used
for firmware updates and may not be used for realtime measurements.
WARNING: The USB is not isolated from the inputs for safety.
Connecting to a computer while the inputs are connected may damage
your computer
operated with the
and/or your person.
battery cover removed.
The instrument must never be
The logger can be connected to the computer via Bluetooth during a
recording session, however, an additional drain on the batteries will occur
to support the communications.
3.2.1
The SLII is designed to connect to a computer wirelessly via Bluetooth for
real-time measurements, data downloads and instrument setup.
The instrument is supplied with a Bluetooth USB adapter for computers
without Bluetooth capability. If this is the case, connect it to an available
USB port on your computer. The default Windows driver will automatically
install the device.
The pairing procedure varies depending on your operating system,
Bluetooth equipment and driver software.
NOTE: The information in the following sections are only necessary the
first time an instrument is connected, or when the Bluetooth USB adapter
has been removed then reconnected.
3.2.2 Pairing
1. Make sure Bluetooth is enabled on the instrument.
2. Open the Bluetooth Devices dialog on your computer to pair the
instrument with your computer. Different operating systems have different steps for opening this dialog, so consult your computer’s documentation for instructions.
3. Once the dialog is displayed, click Add a Device. A dialog box appears
listing the locally available Bluetooth devices. There may be several
devices of varying types listed, depending on the location of your
computer.
4. Find the instrument, which will appear listed by its Bluetooth name.
Click the name. Depending on the operating system, you may be
prompted to enter a pairing code. If so, enter 1234.
5. Click Next. A screen appears informing you that the instrument has
been successfully connected with the computer. Click Close to exit
the screen.
6. To complete the connection, open the Simple Logger II Control
Panel. In the menu bar at the top of the screen, select Help. In the
drop-down menu that appears, click the option Help Topics. This
opens the Simple Logger II Control Panel Help system.
7. Use the Contents window in the Help system to locate and open the
topic “Connecting to an Instrument.” This topic provides instructions
explaining how to connect your instrument to the computer.
To conserve battery power, Bluetooth can be turned OFF.
To turn Bluetooth OFF, press and hold the Control button until the Blue
LED turns on solid, then release the button. Repeat this step to turn Bluetooth back ON.
If all LEDs light instantaneously, the instrument was in SLEEP mode.
Releasing the Control button will show the status (every 5 seconds).
3.4 Recording Data
The Simple Logger® II must first be configured before a recording can be
performed (see § 4).
Once a configuration is written to the instrument, it will no longer need to
be connected to DataView
When data is stored in the memory, the user may download the information onto a hard-disk (see § 4).
A scheduled recording will still start even if the logger is in SLEEP mode.
3.4.1
1. Connect the instrument to the measurement source.
2. Press and hold the CONTROL button. When the START (GREEN)
3. The GREEN LED double-blinks every 5 seconds when the logger is
Starting a Recording Session
A new recording cannot be started if the memory is full.
LED lights up, release the button.
recording.
®
to start the scheduled recording.
It may take a few seconds before the instrument starts recording. The
recording status LED will single blink during this period.
1.Press and hold the CONTROL button. When the STOP (AMBER) LED
2.The GREEN LED will change from a double-blink to a single-blink,
The data will be retained, even if the instrument is in SLEEP mode. The
recorded data is stored in Flash memory (maintained even in the absence
of batteries). The recorded data may be downloaded to a computer.
Stopping a Recording Session
lights up, release the button.
indicating STANDBY mode.
3.5 Downloading Recorded Data
Recorded measurements stored in the instrument are transferred to
a computer via the download command in the Simple Logger II Control
Panel.
3.6 Clearing Alarm Indication
Clearing alarms can be performed in the STANDBY or RECORD mode.
1.Press and hold the CONTROL button. When the ALARM (YELLOW)
LED lights up, release the button. The YELLOW LED will blink at a fast
rate for a period of five seconds
2.Press the CONTROL button for another 0.5 second to complete the
operation.
This does not clear any stored alarms, only indications. Stored alarms can
only be cleared when memory is erased (see § 3.7 - Erasing Data from Memory).
Erasing data from the instrument’s memory can only be performed while
in the STANDBY mode.
There are two ways to erase the memory:
Erasing the Memory using the Control Button:
1. Press and hold the CONTROL button. When the ERASE (RED) LED
lights up, release the button. This will arm the instrument for an erase
operation (when not in record mode). While armed to erase memory,
the RED LED will blink at a fast rate for a period of five seconds
2. Press the CONTROL button for another 0.5 seconds to start the erase
operation. During the erase operation the LED will blink once every
second. Erasure takes about 1 minute to complete.
e button is not pressed within five seconds of arming, the erase oper-
If th
ation will automatically disarm and memory will be maintained. For this
reason, if you do not intend to erase memory, simply wait until the RED
LED stops blinking at the fast rate.
Erasing the Memory using the Simple Logger II Control Panel:
1. Connect the instrument to the computer and launch the Simple
Logger II Control Panel.
2. Select Erase Memory from the Instrument menu.
3. A dialog box will appear asking to confirm the erasure of the instru- ment’s memory. Select Yes to confirm or No to cancel the operation.
Erasing the memory will also clear any stored alarms.
The logger captures Trend measurements. The following are definitions
of terms used in this section:
Input Channel: Source for the measurement channel of the instrument.
Measurement Channel: Measurement of input. This can be a simple
direct measurement, the result of complex mathematical operations on a
single or multiple input, or other channels.
Sample Rate: The rate at which the instrument measures inputs.
Storage Rate: The rate at which channel measurements are stored.
3.8.1
The logger stores the measurement of each of the inputs. In addition, the
user can define the storage rate, recording period and measurement
format using the Configure Instrument dialog box in the Simple Logger II
Control Panel software. Control Panel trend measurements are stored at
this fixed storage rate.
Trend Measurements
3.9Logger Operation
When the instrument is in use, the following occurs (provided there is sufficient battery voltage and no data is stored in the instrument s memory):
•The GREEN LED single-blinks every 5 seconds (STANDBY mode
is active and the logger is not recording).
•The RED LED is OFF, indicating there is no data in memory.
•The CONTROL button is used to Start/Stop a Recording Session.
•If the CONTROL button is not pressed for a period of one minute,
the instrument will enter SLEEP mode and wait for either another
button press or the recording start time to arrive (if a recording is
scheduled). While in SLEEP mode, the LEDs will not blink.
•A button press of 0.5 seconds will return the unit back to the
normal STANDBY mode.
When a recording starts, the logger will continue to record until one of the
following occurs:
Recording with Memory Cleared
•The Session is complete.
•The Memory is full and the recording mode is Start/Stop or Alarm.
•The CONTROL button is pushed until the STOP (AMBER) LED
lights up and is released before the next LED lights.
•The Stop Recording command from the Simple Logger II Control
Panel is sent to the unit.
•The battery voltage falls below 1.7V.
3.9.2
If the RED LED is double-blinking every 5 seconds, the memory is full and
must be erased before any further recording can be performed.
If the RED LED is single-blinking every 5 seconds prior to starting a New
Recording Session, the memory is partially full.
To save, clear or check memory availability, use the Simple Logger II
Control Panel software.
There may be instances where the GREEN LED is also double-blinking
every 5 seconds indicating that the logger is still recording. The user can
choose to stop the Recording Session and download the session and/or
erase the memory.
Recording with a Partial or Full Memory
The logger memory cannot be erased while in the Record mode. The
recording must be stopped first.
3.9.3
If the logger is recording using the Start/Stop or Store on Alarm mode and
the memory is filled before the Recording Session has finished, the session will end.
The following happens after the CONTROL button is pushed for 0.5 seconds:
Memory Filled During Recording Session
•The GREEN LED single-blinks (standby mode).
•The RED LED double-blinks (full memory).
(Start/Stop Mode)
At this time:
•The memory can be downloaded and erased.
•A new recording can be started or scheduled once memory is erased.
When a recording starts, the logger will continue to record until one of the
following occurs:
Recording with Memory Cleared
• The Session is complete.
• The Memory is full and the recording mode is Start/Stop or Alarm.
• The CONTROL button is pushed until the STOP (AMBER) LED lights up
and is released before the next LED lights.
• The Stop Recording command from the Simple Logger II Control Panel is
sent to the unit.
• The battery voltage falls below 1.7V.
3.9.5
If the RED LED is double-blinking every 5 seconds, the memory is full and
must be erased before any further recording can be performed.
If the RED LED is single-blinking every 5 seconds prior to starting a New
Recording Session, the memory is partially full.
To save, clear or check memory availability, use the Simple Logger II
Control Panel software.
There may be instances where the GREEN LED is also double-blinking every
5 seconds indicating that the logger is still recording. The user can choose to
stop the Recording Session and download the session and/or erase the
memory.
Recording with a Partial or Full Memory
The logger memory cannot be erased while in the Record mode. The
recording must be stopped first.
3.9.6
If the logger is recording using the Start/Stop or Store on Alarm mode and the
memory is filled before the Recording Session has finished, the session will
end.
The following happens after the CONTROL button is pushed for 0.5 seconds:
Memory Filled During Recording Session
• The GREEN LED single-blinks (standby mode).
• The RED LED double-blinks (full memory).
(Start/Stop Mode)
At this time:
• The memory can be downloaded and erased.
• A new recording can be started or scheduled once memory is erased.
If in either the XRM or FIFO mode, the recording will continue even after
memory becomes full. Memory will be freed to make room for new samples.
The method of freeing memory will depend on the recording mode.
3.9.7
If the battery voltage drops below 1.7V, the following will occur:
The logger continues to record until the battery voltage drops below 1.7V.
Pressing the CONTROL button will have no effect. The battery voltage
may rise slightly after the unit turns itself off. In this event, the unit may turn
on momentarily as a result of a button press.
The batteries must be replaced before the recorded session(s) can be
downloaded from the instrument (refer to § 5.1).
Battery Power Insufficient for a Full Recording Duration
•The Recording Session will terminate.
•The Data will be saved.
•The GREEN and RED LED will turn OFF.
Replacing the batteries while the unit is in SLEEP mode will not result in
the loss of data memory. The internal backup capacitor will maintain the
clock and memory while the batteries are being replaced. If the battery
level falls below the usable level or if the batteries have been removed
for an extended period of time, the clock time will be lost. However, the
recorded memory will be maintained.
3.9.8
The logger will be in STANDBY mode if one of the following occurs:
Recording Session has Ended
•The session terminates due to recording end time being reached.
•The recording in START/STOP or Store on Alarm mode fills the
memory.
•The user terminates the session by pressing the CONTROL
button until the STOP (AMBER) LED lights up and releasing the
button before the next LED lights, or issues a Stop Recording
command from the Simple Logger II Control Panel.
The logger is now ready for a New Session or Download. Pressing the
CONTROL button until the START (GREEN) LED lights up and releasing
the button before the next LED lights, will start a New Session depending
on the available memory.
CAUTION: Risk of electric shock. Disconnect the unit from any input
source before opening the battery cover.
Resetting the instrument takes about 10 seconds, at which time the instrument will be non-responsive.
The RESET button resets the CPU and is located under the battery cover.
To access it, remove the battery compartment cover. The reset button is
located on the exposed PCB (see § 2.2 for location).
It is recommended to only press the RESET switch when the logger stops
responding to a normal press button control when not connected to Data-
®
. It is not recommended to reset the instrument when the logger is
View
recording, downloading or being configured.
If the logger is not responding to a button press, make sure the unit has
sufficient battery power. If the battery voltage is below 1.7V, the unit will not
respond to a button press. In this condition, pressing the RESET button
will not restart the unit.
It is recommended to download any desired session and erase the memory
before starting a new recording.
22
The resumption of the logger operation in the above situations assumes
that the RESET button cleared the fault(s). The logger will not resume
normal operation if the fault condition still exists. The instrument will try
to recover normally. However, under certain conditions, the clock and
memory full state may reset, meaning that the memory is marked as full
and the clock date/time is lost.
CAUTION: Risk of electric shock. Disconnect the unit from any input
source before opening the battery cover.
The Flash Upgrade Switches are used to recover from a failed flash
upgrade.
To access them, remove the battery compartment cover. The Flash
Upgrade Switches are located on the exposed PCB (see § 2.2 for location).
All three switches should be turned to the ON position in order to flash
upgrade the instrument using the “Fail Safe” method of upgrade from
within the “SLIIFlashUp1xx” utility program.
All switches MUST be set to the OFF position once the upgrade has completed.
Failure to set the Flash Upgrade Switches OFF after the upgrade will prevent the instrument from operating properly and deplete the batteries.
DO NOT CONNECT THE INSTRUMENT TO THE PC BEFORE INSTALLING
THE SOFTWARE AND DRIVERS.
When you purchase an AEMC instrument supported by DataView, the
software is included as part of the product package. DataView program files
are stored on a USB stick.
1.Insert the DataView thumb drive into an available USB port on your
computer. If Autorun is enabled, an AutoPlay window appears on your
screen. Click “Open folder to view files” to display the DataView folder.
NOTE: If Autorun is not enabled or allowed, use Windows Explorer to
locate and open the USB drive labeled “DataView.”
When the DataView folder is open, find the file Setup.exe located in
2.
the root directory and doubleThe DataView setup screen appears.
3.In the upper left corner of the screen, choose the language version of
the Setup interface. (All Setup screens and dialogs will immediately
appear in the selected language.)
In the lower left corner are the available installation options. In
addition to the DataView software, you can select “Adobe Reader.”
This links to the Adobe web site where you can download the latest
version of Reader. This program is required to view DataView .pdf
documents. The option Firmware Upgrades links to the website
where you can check for new firmware updates for the instrument.
Finally, User Manuals displays a list of .pdf files contained in the USB
drive that accompanies DataView. (DataView also comes with a Help
system that is installed with the program files.)
click it to run the installation program.
To install DataView, select DataView in the Options list and click
Install.
4.Select the language version of DataView you want to install (English,
French, or Spanish) then click Next. (By default, the language
selected in step 3 is highlighted.).
5.You are now prompted to select the software you want to install.
Each AEMC product family has its own specially designed Control
Panel. By default all available Control Panels are selected (a check
mark next to the Control Panel indicates it is selected). Control
Panels take up disk space on the computer; so unless you have
other types of AEMC instruments, we recommend that you select
Simple Logger II and deselect the rest. You should also check the
option DataView Core, which is a requirement if you plan to create
DataView reports.
After you finish selecting and deselecting Control Panels and/or
DataView Core, click Next.
The Setup program now informs you that it is ready to install
6.
DataView. If you want to review any of your previous selections,
click the Previous button to return to earlier screens. Otherwise,
click Install to begin installation.
7.The InstallShield program installs the selected software. If an earlier
version of the software is already installed on your computer, for
each selected program the InstallShield program will:
(a) Ask you to confirm the installation of the program. Click Next.
(b) Display a status bar indicating the progress of the installation.
(c) Inform you when the program is installed. Click Finish to install
the next selected program.
If the software is not installed (or if the installed software is the same
version as the selected software), the software is installed without
requesting confirmation.
When all programs are installed, a message appears informing you
of this. Click Finish to return to the Setup screen.
8.You can now select additional Setup options to install (see step 3
above). When finished, click Exit.
After you finish selecting and deselecting Control Panels, click Next.
The DataView folder now appears on your computer desktop,
9.
within which is the Simple Logger II icon and the icon(s) for any
other Control Panel(s) you have installed.
After you install DataView, the DataView folder appears on your desktop. This folder contains the DataView icon, which opens the core DataView program. The folder also contains the Simple Logger II Control
Panel icon. Clicking this icon opens the Simple Logger II Control Panel.
In general, core DataView features are for creating, viewing, editing,
and storing DataView reports; while the Control Panel is for connecting
to, configuring, viewing real-time measurements, and downloading data
from the instrument. You can access all DataView features through
either the DataView icon or the Control Panel icon. For users who interact with a single type of AEMC instrument, we recommend primarily
using the Control Panel. However, there are situations where using the
core DataView icon may be more convenient for some users, such as
when viewing multiple archived reports from different AEMC product
families.
NOTE: For further information about DataView and its
capabilities, or for information about using the Simple
Logger II Control Panel, consult the Help system that
comes with the product. Access this Help by clicking the
option Help in the Control
the screen.
Use only factory specified replacement parts. AEMC® will not be held
responsible for any accident, incident, or malfunction following a repair
done other than by its service center or by an approved repair center.
5.1Replacing the Batteries
CAUTION: Risk of electric shock. Disconnect all input(s) from the unit
or remove the current sensors from any conductor before opening the
battery cover to change the batteries. Terminate any recordings before
changing the batteries or loss of recorded data may occur.
• Remove the two screws from the battery cover.
• Slide and lift up the cover to remove.
• Replace the four 1.5V C-cell (LR14) alkaline batteries (the backup
capacitor provides backup power for approximately 10 seconds while
the main batteries are being changed).
• Replace the cover and the screws.
NOTE: When changing the batteries, always replace both sets of batteries.
If the instrument is stored without the batteries installed, the internal
clock will need to be reset using the Simple Logger II Control Panel
software once the batteries are installed. Only store the unit with the
batteries installed for short periods of time. For prolonged storage of the
unit, it is recommended to remove the batteries.
5.2Cleaning
CAUTION: Risk of electric shock. Remove the clamp from any conductor
or disconnect all input(s) from the unit.
The body of the logger should be cleaned with a damp and soapy cloth.
Do not submerge the logger in water. Do not use solvent.
NOTE: The native measurement of the Model L104 is voltage. This voltage is scaled to
displace current by a factor that corresponds to the users probe selection. This accuracy
specification does not account for probe errors.
Start/Stop, FIFO and Extended Recording Mode* (XRM™), Store on Alarm
Recording Length
Memory
even if battery is low or removed.
Communication
Power Source**
4x1.5V C-cell alkaline batteries
Battery Life
MECHANICAL
Dimensions
Weight (with battery)
2.1 lbs (0.95kg)
Cover/Case
Vibration
Shock
IEC 68-2-27 (30G)
Drop
ENVIRONMENTAL
Operating Temperature
Storage Temperature
-4° to 140°F (-20° to 60°C)
Relative Humidity
Up to 85% at 95°F (35°C), Non-condensing
Altitude
SAFETY & ELECTRO-MAGNETIC COMPATIBILITY
Safety Rating
Protection Degree
IP50 (cover closed)
Electro-Magnetic
Compatibility
CE Approved
Five Recessed Banana Jacks
0 to 600V
0.1V
0 to 5V unspecified
5 to 50V: ±(0.5% of Reading + 1V)
50 to 600V: ±(0.5% of Reading + 0.5V)
1.2 x 600V
64 samples/cycle
Programmable from 125ms to 1 per day
15 minutes to 8 weeks, programmable using DataView
®
1,000,000 measurements (2MB).
Recorded data is stored in non-volatile memory and retained
Bluetooth (Class 2)
>200hrs at 125ms up to 180 days
(dependent on storage rate/recording length)
5.904 x 5.904 x 3.568" (150 x 150 x 91mm) w/o Sensor
Cover: UL94-V0 / Case: UL94-5VB
IEC 68-2-6 (1.5mm, 10 to 55Hz)
IEC 68-2-32 (1m)
14° to 122°F (-10° to 50°C)
2000m
EN61010-1; 600V CAT IV, 1000V CAT III; Pollution Degree 2
EN 61326-1; 07/1997 (+A1 10/1998, +A2 09/2001, +A3 05/2004)
Pending
*This unique recording mode provides the opportunity to continuously record over long periods of time by reducing the storage
resolution of the stored data and maintaining matching resolution for the newest data. Each time the memory fills up using
™
XRM
, every other of the oldest stored samples is discarded making room for newer samples. This process continues until the
recording is manually stopped.
**A memory backup capacitor provides backup power while batteries are being changed. This backup capacitor will maintain the
instrument for up to 10 sec. without batteries installed.After 10 seconds the date and time will need to be reset (data and configuration will be maintained). If the unit is connected to DataView
***Voltages beyond this range may damage the instrument.
®
via a PC, the battery life is 100 hours regardless of the storage rate.
NOTE: When installing probes, face the arrow on the probe in the direction of the load.
Model SR193
Nominal Range
Measurement Range
Probe Output Signal
Maximum Clamping Diameter
2"
Safety
Model MN93
Nominal Range
Measurement Range
Probe Output Signal
Maximum Clamping Diameter
Safety
Model MN193
Nominal Range
Probe Output Signal
Maximum Clamping Diameter
Safety
1000AAC for f ≤1kHz
1A to 1000AAC max (I >1000A not continuously)
1mVAC/AAC
(52mm)
EN 61010-2-032, Pollution Degree 2, 600V CAT IV, 1000V CAT III
200AAC for f ≤1kHz
2A to 200AAC max (I >200A not permanent)
5mVAC/AAC
0.8" (20mm)
EN 61010-2-032, Pollution Degree 2, 300V CAT IV, 600V CAT III
NOTE: Currents <0.5A will be displayed as zero. Neutral current measures down to 0A.
5A and 100AAC
Measurement Range
5A: 0.005A to 6A
EN 61010-2-032, Pollution Degree 2, 300V CAT IV, 600V CAT III
AC max (1A to 1000A with ratio 1000/5 selected)
100A: 0.1A to 100A
5A: 200mV/AAC; 100A: 10mV/AAC
0.8" (20mm)
AC max
The 5A range of the MN193 is designed to work with secondary current transformers.
Best accuracy is available when entering the transformer ratio (e.g. 1000/5A). When used to
measure 5A direct, the resolution will be limited to 0.1A max.
Specifications are subject to change without notice.
Symptom: After being in a damp, cold environment, the logger does not
function.
Cause, Correction: Condensation may have formed inside the logger,
shorting out the circuitry and discharging the battery. Allow the circuit
board to dry thoroughly in a warm location.
Symptom: Simple Logger® II does not start recording.
Cause, Correction: Make sure battery power is present. Make sure
the PRESS button is pushed long enough to light the GREEN LED and
released before the next LED lights. Make sure the RED LED is not
double- blinking. If it is, memory is full and you need to erase the data (see
§ 3.7). Make sure the SLII is properly configured so that you have Storage
Rate, Recording Period and at least one Measurement Channel specified.
Symptom: Cannot find the instrument when pairing the Bluetooth connection.
Cause, Correction: Make sure the Blue LED is single blinking, indicating
that the instrument Bluetooth module is active.
Symptom: Even though the pairing of the Bluetooth is established and
the LED indication is active, the instrument will not connect to the Simple
Logger II Control Panel.
Cause, Correction: Make sure the Low Battery LED is not blinking. If the
batteries are below 2.2V the instrument will have enough power to run the
circuit but not enough to maintain a stable connection. Replace the batteries.
Some general terminology associated with the data collection process
is listed here for convenience.
Bps: Bits Per Second, a unit of signal transfer speed equal to the num-
®
ber of elements per second. The Simple Logger
II transfers data at the
rate of 115200 bps.
Button: An actual key on the logger or computer keyboard or a soft key
in the program on the computer screen.
Data logger: A device used to sample and store electrical signals that
may be representative of physical phenomena such as temperature,
pressure and flow for long periods of time in an unattended environment.
Download: The process of transferring data from the logger to the
computer.
Hz: Hertz, a unit of measure of frequency equivalent to cycles per second.
I/O: Input/output, a device or port capable of sending or receiving digital
information.
Port: A name given to any connector allowing input or output of information.
Processor: A computing device used to calculate and run a set of
instructions.
Recording session: A recording session is defined as the time and data
contained within the starting and ending of a recording.
Resolution: The last significant digit of the measurement.
Zoom: The ability to select a section of the graph and magnify it for
better readability.
Bluetooth: Wireless communications device used to access the logger
®
via a computer program (Dataview
).
USB: Universal Serial Bus, a communications port used to access the
To ensure that your instrument meets factory specifications, we recommend
that it be sent back to our factory Service Center at one-year intervals for recalibration, or as required by other standards or internal procedures.
For instrument repair and calibration:
You must contact our Service Center for a Customer Service Authorization
Number (CSA#). This will ensure that when your instrument arrives, it will be
tracked and processed promptly. Please write the CSA# on the outside of the
shipping container. If the instrument is returned for calibration, we need to
know if you want a standard calibration, or a calibration traceable to N.I.S.T.
(Includes calibration certificate plus recorded calibration data).
(O
r contact your authorized distributor)
Cost for repair, standard calibration, and calibration traceable to N.I.S.T. are
available.
NOTE: You must obtain a CSA# before returning any instrument.
Technical and Sales Assistance
If you are experiencing any technical problems, or require any assistance with
the proper operation or application of your instrument, please call, mail, fax or
e-mail our technical support team:
The Simple Logger® II is warranted to the owner for a period of two year from
the date of original purchase against defects in manufacture. This limited warranty is given by AEMC
®
Instruments, not by the distributor from whom it was
purchased. This warranty is void if the unit has been tampered with, abused
®
or if the defect is related to service not performed by AEMC
Instruments.
Full warranty coverage and product registration is available on our
website
Pl
ease print the online Warranty Coverage Information for your records.
What AEMC
®
Instruments will do:
If a malfunction occurs within the warranty period, you may return the instrument
to us for repair, provided we have your warranty registration information on
®
file or a proof of purchase. AEMC
Instruments will, at its option, repair or
replace the faulty material.
Warranty Repairs
What you must do to return an Instrument for Warranty Repair:
First, request a Customer Service Authorization Number (CSA#) by phone
or by fax from our Service Department (see address below), then return the
instrument along with the signed CSA Form. Please write the CSA# on the
outside of the shipping container. Return the instrument, postage or shipment
pre-paid to:
ion: To protect yourself against in-transit loss, we recommend you insure
Caut
your returned material.
NOTE: You must obtain a CSA# before returning any instrument.