By registering online within 30 days from the date of
purchase, your warranty will be extended to 3 years
User Manual
Thank you for purchasing a PowerPad® III Model 8435
For best results from your instrument and for your safety, read the enclosed operating instructions carefully and comply
with the precautions for use. These products must be only used by qualied and trained users.
WARNING, risk of DANGER! The operator must refer to
these instructions whenever this danger symbol appears.
CAUTION! Risk of electric shock. The voltage at the
parts marked with this symbol may be dangerous.
Equipment is protected by double insulation.
USB socket.
SD Card.
Ground/Earth.
Important instructions to read and to fully understand.
Useful information or hint to read.
The CE marking guarantees conformity with European directives and with regulations covering EMC.
The trash can with a line through it means that in the European Union, the product must undergo selective disposal for the
recycling of electric and electronic material, in compliance with Directive WEEE 2002/96/EC.
Denition of Measurement Categories (CAT)
■ CAT IV Measurement category IV corresponds to measurements taken at the source of low-voltage installations.
Example: power feeders, counters and protection devices.
■ CAT III Measurement category III corresponds to measurements on building installations.
Example: distribution panel, circuit-breakers, machines or xed industrial devices.
■ CAT II Measurement category II corresponds to measurements taken on circuits directly connected to low-voltage
installations. Example: power supply to domestic electrical appliances and portable tools.
PRECAUTIONS FOR USE
This instrument and its accessories comply with safety standards IEC 61010-1, IEC 61010-031, and IEC 61010-2-032 for
voltages of 600V in category IV or 1000V in category III.
Failure to observe the safety instructions may result in electric shock, re, explosion, and destruction of the instrument
and of the installations.
● The operator and/or the responsible authority must carefully read and clearly understand the various precautions to be
taken in use. Sound knowledge and a keen awareness of electrical hazards are essential when using this instrument.
● If you use this instrument other than as specied, the protection it provides may be compromised, thereby
endangering you.
● Do not use the instrument on networks of which the voltage or category exceeds those mentioned.
● Do not use the instrument if it seems to be damaged, incomplete, or poorly closed.
● Before each use, check the condition of the insulation on the leads, housing, and accessories. Any item of which the
insulation is deteriorated (even partially) must be set aside for repair or scrapping.
● Use only the leads and accessories supplied. Using leads (or accessories) of a lower voltage or category reduces the
voltage or category of the combined instrument + leads (or accessories) to that of the leads (or accessories).
● Use personal protection equipment systematically.
● Keep your hands away from the terminals of the device.
● When handling the leads, test probes, and crocodile clips, keep your ngers behind the physical guard.
● Use only the mains power adaptor and battery pack supplied by the manufacturer. They include specic safety
features.
● Some current sensors must not be placed on or removed from bare conductors at hazardous voltages: refer to the
sensor manual and comply with the handling instructions.
READ CAREFULLY BEFORE
USING FOR THE FIRST TIME
Your instrument is equipped with a NiMH battery. This technology oers several advantages:
● Long battery charge life for a limited volume and weight.
● Possibility of quickly recharging your battery.
● Signicantly reduced memory eect: you can recharge your battery even if it is not fully discharged.
● Respect for the environment: no pollutant materials such as lead or cadmium, in compliance with the applicable
regulations.
After prolonged storage, the battery may be completely discharged. If so, it must be completely recharged.
Your instrument may not function during part of this recharging operation.
Full recharging of a completely discharged battery may take several hours.
NOTE: In this case, at least 5 charge/discharge cycles will be necessary for your battery to recover 95% of its capacity.
To make the best possible use of your battery and extend its eective service life:
● Only use the charger supplied with your instrument. Use of another charger may be dangerous.
● Only charge your instrument at temperatures between 0° and 40°C.
● Comply with the conditions of use dened in the operating manual.
● Comply with the storage conditions specied in the operating manual.
NiMH technology allows a limited number of charge/discharge cycles depending signicantly on:
● The conditions of use.
● The charging conditions.
Do not dispose of the battery pack with other solid waste. Used batteries must be entrusted to a qualied recycling
company or to a company specialized in processing hazardous materials.
1.1 RECEIVING YOUR SHIPMENT ............................................................................................................................................... 7
1.2 ORDERING INFORMATION ..................................................................................................................................................... 7
1.2.2 REPLACEMENT PARTS ............................................................................................................................................... 7
2. PRODUCT FEATURES .................................................................................................................................... 8
2.2 KEY FEATURES ........................................................................................................................................................................ 8
2.3 CONTROL FUNCTIONS ......................................................................................................................................................... 10
3.2.3.1 ENERGY UNIT .................................................................................................................................................... 19
3.2.3.2 K FACTOR ........................................................................................................................................................... 19
3.2.5 PROBES AND RATIOS ............................................................................................................................................... 25
3.2.5.1 CURRENT PROBES ........................................................................................................................................... 25
3.2.5.2 VOLTAGE RATIOS .............................................................................................................................................. 26
3.2.10 ABOUT ........................................................................................................................................................................ 34
4.1.1.1 PROGRAMMING AND STARTING A SEARCH ................................................................................................ 35
4.1.1.2 STOPPING A SEARCH....................................................................................................................................... 36
4.1.1.3 DISPLAYING A TRANSIENT .............................................................................................................................. 37
4.1.1.4 DELETING A TRANSIENT ................................................................................................................................. 38
4.1.2 INRUSH CURRENT MODE
4.1.2.1 PROGRAMMING A CAPTURE .......................................................................................................................... 39
4.1.2.2 STARTING A CAPTURE ..................................................................................................................................... 40
4.1.2.3 MANUALLY STOPPING A CAPTURE ............................................................................................................... 40
4.1.2.4 DISPLAYING THE PARAMETERS OF A CAPTURE ........................................................................................ 40
4.1.3 TRUE RMS CURRENT AND VOLTAGE .................................................................................................................... 41
4.2.1.2 PHASE VOLTAGE HARMONICS DISPLAY SCREEN (L1) .............................................................................. 45
4.2.2 CURRENT .................................................................................................................................................................... 46
4.2.2.1 CURRENT HARMONICS DISPLAY SCREEN (3L) .......................................................................................... 46
4.2.2.2 CURRENT HARMONICS DISPLAY SCREEN (L1) .......................................................................................... 47
4.2.3 APPARENT POWER ................................................................................................................................................... 48
4.2.3.1 APPARENT POWER HARMONICS DISPLAY SCREEN (3L) .......................................................................... 48
4.2.3.2 APPARENT POWER HARMONICS DISPLAY SCREEN (L1) .......................................................................... 48
4.2.4 PHASE-TO-PHASE VOLTAGE ................................................................................................................................... 49
4.2.4.1 PHASE-TO-PHASE VOLTAGE HARMONICS DISPLAY SCREEN (3L) ......................................................... 49
4.2.4.2 PHASE-TO-PHASE VOLTAGE DISPLAY SCREEN (L1) ................................................................................. 50
4.2.5 HARMONIC ANALYSIS IN EXPERT MODE .............................................................................................................. 51
4.4.2 STARTING AN ALARM ................................................................................................................................................ 62
4.4.3 MANUALLY STOPPING AN ALARM .......................................................................................................................... 62
4.4.4 DISPLAYING AN ALARM LOG ................................................................................................................................... 62
4.4.5 DELETING AN ALARM LOG ....................................................................................................................................... 63
4.5 TREND MODE
4.5.1 PROGRAMMING AND STARTING A RECORDING ................................................................................................. 63
4.5.2 MANUALLY STOPPING A RECORDING ................................................................................................................... 64
4.5.3 DISPLAYING THE RECORDING LIST ...................................................................................................................... 64
4.5.4 DELETING A RECORDING ........................................................................................................................................ 64
4.6 POWER AND ENERGY MODE .............................................................................................................................................. 64
5.1.1 USB FLASH DRIVE INSTALL ..................................................................................................................................... 73
5.2 CONNECTING THE MODEL 8435 TO YOUR COMPUTER ................................................................................................. 76
5.3 OPENING THE CONTROL PANEL ........................................................................................................................................ 76
5.4 CONFIGURING THE INSTRUMENT ..................................................................................................................................... 78
5.5 REAL-TIME DATA .................................................................................................................................................................... 85
5.5.4 POWER ....................................................................................................................................................................... 86
5.5.5 ENERGY ....................................................................................................................................................................... 87
5.6 DOWNLOADING DATA ........................................................................................................................................................... 87
6.2.1 VOLTAGE INPUTS ...................................................................................................................................................... 91
6.2.2 CURRENT INPUTS ..................................................................................................................................................... 91
6.2.3 ACCURACY SPECIFICATIONS (EXCLUDING CURRENT PROBES) ................................................................... 91
6.2.4 CURRENT PROBE ACCURACY (TO BE ADDED TO THE MODEL 8435’S ACCURACY) .................................... 99
6.2.5 CURRENT PROBES AND SENSORS ..................................................................................................................... 100
6.2.6 THREE-PHASE 5A ADAPTER BOX (3-CHANNEL USE ONLY) ............................................................................ 102
6.2.7 POWER SUPPLY ....................................................................................................................................................... 103
APPENDIX A .................................................................................................................................................... 109
A.1 MATHEMATICAL FORMULAS FOR VARIOUS PARAMETERS ........................................................................................ 109
A.1.1 NETWORK FREQUENCY AND SAMPLING ........................................................................................................... 109
A.1.4.1 DISTRIBUTION SYSTEM WITH NEUTRAL ....................................................................................................11 6
A.1.4.2 THREE-PHASE SYSTEM WITHOUT NEUTRAL ............................................................................................117
A.1.4.3 TWO-PHASE SYSTEMS WITHOUT NEUTRAL .............................................................................................119
A.1.5 POWER RATIOS (EXCLUDING NEUTRAL – OVER ONE SECOND) .................................................................. 120
A.1.5.1 DISTRIBUTION SYSTEM WITH NEUTRAL ................................................................................................... 120
A.1.5.2 THREE-PHASE SYSTEM WITHOUT NEUTRAL ........................................................................................... 121
A.1.5.3 TWO-PHASE SYSTEM WITHOUT NEUTRAL ............................................................................................... 121
A.1.6 ENERGY .................................................................................................................................................................. 122
A.1.6.1 DISTRIBUTION SYSTEM WITH NEUTRAL ................................................................................................... 122
A.1.6.2 DISTRIBUTION SYSTEM WITHOUT NEUTRAL ........................................................................................... 124
A.1.10 MECHANISM FOR TRIGGERING TRANSIENT SENSORS .............................................................................. 127
A.1.11 CAPTURE CONDITIONS IN INRUSH CURRENT MODE ................................................................................... 128
A.2 GLOSSARY OF TERMS ....................................................................................................................................................... 129
A.3 SYMBOLS AND ABBREVIATIONS ...................................................................................................................................... 130
6
Power Quality Analyzer PowerPad® III Model 8435
1. INTRODUCTION
1.1 RECEIVING YOUR SHIPMENT
NOTE: After receiving your PowerPad® III Model 8435, charge and discharge the instrument one or two cycles to ensure the
proper level display of the battery indicator.
Make sure the contents shipped are consistent with the packing list. Notify your distributor of any missing items. If the
equipment appears damaged, le a claim immediately with the carrier and notify your distributor at once. Save the damaged
packing container to substantiate your claim. Do not use equipment which is damaged or appears to be damaged.
1.2 ORDERING INFORMATION
PowerPad® III Model 8435 (No Sensors - Waterproof IP67) ............................................................................ Cat. #2136.41
PowerPad
Includes extra large tool bag, accessory pouch, 5 ft USB cable, five 10 ft black voltage leads with alligator clips, 110V
US power cord, four water-tight AmpFlex® A196-24-BK (included with Cat. #2136.42 only), NiMH battery, SD-Card,
twelve color-coded input ID markers, quick start guide, and a USB stick containing DataView® software and user manual.
®
III Model 8435 (AmpFlex®- Waterproof IP67 ) ............................................................................... Cat. #2136.42
Only the SR, AmpFlex® and MiniFlex® sensors are rated 600V CAT IV. The MN, MR and SL probes are 300V CAT IV,
600V CAT III. Only AmpFlex® 196-18-BK and 196-24-BK sensors are rated IP67.
1.2.1 ACCESSORIES
10 ft USB cable.................................................................................................................................................. Cat. #2136.80
AC/DC Current Probe Model SL261* (10A-100mV/A, 100A-10mV/A, BNC) .................................................... Cat. #1201.51
* Adapter - BNC Adapter (mandatory if using an SL261 w/ the 8435) ........................................................... Cat. #2140.40
AC Current Probe Model MR193-BK................................................................................................................. Cat. #2140.28
AC Current Probe Model MN93-BK................................................................................................................... Cat. #2140.32
AC Current Probe Model SR193-BK ................................................................................................................. Cat. #2140.33
AC Current Probe Model MN193-BK ................................................................................................................ Cat. #2140.36
MiniFlex® Sensor 10" Model MA193-10-BK ...................................................................................................... Cat. #2140.48
MiniFlex® Sensor 14" Model MA193-14-BK ...................................................................................................... Cat. #2140.50
AmpFlex® Sensor 24" Model A196-24-BK ......................................................................................................... Cat. #2140.75
1.2.2 REPLACEMENT PARTS
Extra Large Classic Tool Bag ............................................................................................................................ Cat. #2133.73
Battery 9.6V NiMH
Set of 12, Color-coded Input ID Markers ........................................................................................................... Cat. #2140.45
5 ft USB cable.................................................................................................................................................... Cat. #2140.46
Accessory Pouch (accessories not included) .................................................................................................... Cat. #2140.72
®
AmpFlex
Lead – One 10 ft (3M) Black Lead (Waterproof cap) {Rated 1000V CAT IV} &
One Black Alligator Clip {Rated 1000V CAT IV, 15A} ....................................................................................... Cat. #2140.73
Power Cord 110V .............................................................................................................................................. Cat. #5000.63
Sensor 24" Model A196-24-BK ......................................................................................................... Cat. #2140.75
The easy-to-use, compact and shock-resistant PowerPad® III Model 8435 is a three-phase power quality analyzer
equipped with four current inputs and ve voltage inputs. It is intended for technicians and engineers to measure and
carry out diagnostic work and power quality work on one, two or three phase low voltage networks.
The PowerPad® III Model 8435 has SD Card memory for storing trend data, and internal Flash memory for storing
transient, alarms, and snapshot data. You can store up to 50 screen snapshots, up to 210 captured transients that contain
four cycles for each active input, and 10,000 alarm events from up to 40 dierent parameters. You can also record trend
data for days, weeks, or even months. The Model 8435 is IP67 rated.
Six access buttons quickly display the following functions. Note that these functions can be accessed during a recording.
Waveforms (
inputs on one screen, phase-to-phase or phase-to-neutral. Real-time phasor diagrams can be displayed for volts and
amps, also by phase or for all phases including phase unbalance.
Harmonics () displays harmonics out to the 50th for volts, amps, and VA. Individual harmonics are displayed as a
percentage and value for volts, amps and VA. Harmonic direction and sequencing can also be displayed.
Transients or Inrush Current () sets, captures, and displays transients. You select the threshold and the number of
transients to capture. The Model 8435 then captures four waveforms for each transient; the triggering waveform as well
as one pre-triggered and two post-triggered waveforms. As many as 210 transients, each consisting of 4 waveforms per
channel for up to 8 channels, can be captured. Inrush current is measured when the power is rst turned on. Depending
on the type of load, the Inrush current may be very high for some time when compared with steady state current later on.
Trend () records and displays trend data at user selectable sample rates from 1/sec to 1/15 minutes and user
programmable recording periods into 2GB of memory.
Alarm Events (
Power/Energy (
tangent).
) displays volts, amps, THD, and Crest Factor by phase or for all phases. You can display all the voltage
) provides a list of the alarms recorded according to the thresholds programmed during conguration.
) records and displays power levels and associated parameters (power factor, displacement, and
Recordings can only be downloaded through the USB port. Extremely long recordings will take a long time to download (> 2H
per 100MB).
2.2 KEY FEATURES
■ Voltage ratios: 1V to 500kV individual or grouped conguration of channels
■ Mixed sensors: 10mA to 60kA depending on sensor. Combine as needed. The measurements are performed
simultaneously using several dierent current sensors
■ Transient adjustment level: Adjustment of 1V or 1A increments. Individual or grouped channels
■ Measurement of TRMS voltages up to 1000Vrms AC/DC for two, three, four, or ve-wire systems
■ Measurement of TRMS currents from 1mA to 10kA (sensor dependent)
■ Measurement of DC current up to 1200ADC (with MR193 probe)
■ Automatic probe detection and scaling
■ Frequency measurement (40 to 69Hz systems)
8
Power Quality Analyzer PowerPad® III Model 8435
■ Direct measurement of neutral current for WYE congurations
■ Record and display trend data as fast as once per second for one month for up to 25 variables. It is recommended to
limit recordings to under 100MB to reduce downtime
■ Energy assessments
■ Transient detection on all V and I inputs
■ Inrush current detection and measurement
■ Calculation of Crest Factors for current and voltage
■ Calculation of the K Factor for transformers
■ Calculation of short-term icker for voltage
■ Calculation of the phase unbalance for voltage and current (3 phase only)
■ Measurement of harmonic angles and rates (referenced to the fundamental or RMS value) for voltage, current or
power, up to the 50th harmonic
■ Display of harmonic sequencing and direction
■ Calculation of overall harmonic distortion factors
■ Real time display of phasor diagrams including values and phase angles
■ Monitoring of the average value of any parameter, calculated over a period running from 1 sec to 2 hrs
■ Measurement of active, reactive and apparent power per phase and their respective sum total
■ Calculation of power factor, displacement power factor and tangent factor
■ Total power from a point in time, chosen by the operator
■ Recording, time stamping and characterization of disturbance (swells, sags and interruptions, exceedence of power
and harmonic thresholds)
■ Detection of transients and recording of associated waveforms
■ Color-coded input ID markers identify voltage and current inputs
■ Internal SD-Card
■ DataView® analysis software. This includes the PowerPad® III Control Panel for con
guring the instrument and viewing
measurement data in real time, as well as DataView® report generation capabilities.
Power Quality Analyzer PowerPad® III Model 8435
9
2.3 CONTROL FUNCTIONS
2
3
4
5
1
?
6
7
8
9
10
11
12
1. Protective cover
2. Battery charging connector
3. Six variable function buttons (yellow)
4. Four function buttons (gray)
5. ON/OFF button (gray)
6. Four current inputs and ve voltage inputs
2.4 CONNECTION TERMINALS
N/D
N/DL3/C
L3/C
Figure 2-1
7. LCD Display
8. USB port
9. Conrm/Enter button (gray)
10. Navigation buttons (gray)
11. Six mode buttons (see § 2.6) (purple)
12. Battery compartment and SD-Card slot cover
L2/BL1/A
1
2
L2/BL1/A
E/GN
Figure 2-2
1. Four current inputs on the top of the instrument to enable the use of current sensors (MN, SR, AmpFlex®, MiniFlex®,
and MR probes).
2. Five (5) voltage inputs.
The instrument ships with protective plugs that are designed to be inserted into the connection terminals. The plugs must
be removed to connect the leads, then stored in the pouch inside the front cover. Insert the plugs into unused terminals to
keep the instrument water/air-tight and the terminals clean.
10
Power Quality Analyzer PowerPad® III Model 8435
2.5 DISPLAY
4
The Model 8435 includes a 320 x 240 pixel LCD display.
1
2
3
1. Top bar on the display, including:
● Symbol of the tested mode (, , , etc.)
● Frequency of measured signal
● Memory capacity status bar (only available in certain modes)
● Current date and time
● Battery charge status (see § 7.1)
02/06/14
5
Figure 2-3
2. Measured RMS values associated with waveforms.
3.
Values of signals at an instant “t” at the intersection of cursor and the waveforms. The cursor can be moved along the time
scale by pressing the
◄ and ► buttons
.
4. Measurement selection (refer to chart below)
Calculation of DPF, Tan, KF, Φ, UNB, Min, Max, VAR, Harmonics, PST, and DF parameters and the frequency measurement can
only be performed if voltage or current with a frequency of 40 to 70Hz is applied to the Ch1 voltage input or the A1 current input.
5. Selection of waveforms to be displayed (use the ▲ and ▼ buttons to select):
U: Signies phase-to-phase voltage
V: Signies phase-to-neutral voltage
A: Displays current for each phase
L1, L2, L3: Refer to the phases (A,B,C)
Icon
RMS
THD
CF
V
A
VA
Description
True RMS Measurement
Total Harmonic Distortion
Crest Factor
Display of current values and their min/max extremes
Display all voltage and current measurements simultaneously (RMS, DC, THD, CF, PST, KF, DF) in tabular format
Phasor diagram
Phase-to-neutral voltage mode
Phase-to-neutral current mode
Power mode
Power Quality Analyzer PowerPad® III Model 8435
11
Icon
1
2
3
4
1
2
3
4
U
Description
Voltage mode (phase-to-phase)
Zoom in
Zoom out
< >Left/right key prompt
^
^
PF...
W...
Up/down key prompt
Display of PF, DPF and Tan.
Active power
Recording mode
Recording display and selection mode
OK
Validation prompt
Stop function in progress prompt
Display of energy consumed
Display of energy generated
Screen 1 of the help function
Screen 2 of the help function
Screen 3 of the help function
Screen 4 of the help function
Screen 1 of the selected recording parameter
Screen 2 of the selected recording parameter
Screen 3 of the selected recording parameter
Screen 4 of the selected recording parameter
>t=0<
>5=-T<
Moves the cursor to transient triggering time
Moves the cursor to one signal period before the transient triggering date
Activates/deactivates the selection of the transients list display filters
Following page screen
Previous page screen
Trash to delete or remove elements
2.6 BUTTON FUNCTIONS
IconDescription
Transients or Inrush Current:
• Sets and views transient and Inrush current waveforms associated with rapid changes in input
Harmonics Mode:
• Displays the harmonics in percent and value ratios for voltage, current, and power for each harmonic through the 50th
• Determines harmonic current produced by non-linear loads
• Analyzes the problems caused by harmonics according to their order (heating of neutrals, conductors, motors, etc.)
Waveforms Mode:
• Displays voltage and current waveforms or vector representation
• Identifies signal distortion signatures
• Displays of amplitude and phase unbalance for voltage and current
• Checks connections for correct phase order
Alarm Events:
• Provides a list of the alarms recorded according to the thresholds programmed during configuration
• Logs interruption with half-cycle resolution
• Determines energy consumption exceedances
• Stores value, duration, date, time and set point for up to 4096 events
12
Power Quality Analyzer PowerPad® III Model 8435
Trend Mode:
• Lists all recording trends and views them on the display (Urms, Vrms, Arms, etc.)
Power / Energy:
• Displays power levels and the associated parameters (power factor, displacement, and tangent)
• Energy monitoring
• Four quadrant measurement to discern produced/consumed active energy and inductive/capacitive reactive energy
Return to the choice of measurement view
Configure the PowerPad® III Model 8435 (see § 3.2)
Take a snapshot of the current screen or access snapshots already stored in the memory. Record associated waveform and
power measurement data.
Get help on the current display functions, in the language chosen by the user.
Power Quality Analyzer PowerPad® III Model 8435
13
3. OPERATION
3.1 GETTING STARTED
Fully charge the battery before the rst use.
NOTE: A full recharge of a completely discharged battery takes approximately 5 hours.
120V ± 10%, 60Hz
230V ± 10%, 50Hz
3.1.1 CONNECTING
1. Start the instrument by pressing the button.
2. Congure the unit to obtain the required results and type of network (see § 3.2).
3. Connect the current leads and sensors to the Model 8435.
4. Connect the ground and/or neutral lead to the network ground and/or neutral (when distributed), as well as the
corresponding current sensor.
5. Connect the L1 phase lead to the network L1 phase, as well as the corresponding current sensor.
To recharge the battery:
1. Unscrew the cover of the power charging connector.
2. Connect the supplied power cord to the instrument and AC power.
3. The button lights and will go out when the power cord is
disconnected.
Repeat the procedure for phases L2, L3 and N.
NOTE: Complying with this procedure limits connection errors to a minimum and conserves time.
3.1.2 DISCONNECTING
■ Proceed in the reverse order to connecting, always nishing by disconnecting the ground and/or neutral (when
distributed).
■ Disconnect the unit leads and power it down.
■ Recharge the battery and recover the data registered where necessary.
■ USB sockets can be used to connect the unit to any type of network.
The following precautions for use must be complied with:
■ Do not connect any voltages exceeding 1000Vrms in relation to the ground/earth.
■ When connecting and disconnecting the batteries, check that measuring leads are disconnected.
The instrument operates on batteries. The batteries are charging when the instrument is connected to a 120/240; 60/50Hz
line.
The current probes connected are identified every second. The scaling will automatically reset when a new probe is connected. When installing probes, face the arrow on the probe in the direction of the load.
For the SL261, MN193, and 5A Box, the ratio needs to be programmed from the instrument menu or software.
14
Power Quality Analyzer PowerPad® III Model 8435
3.2 INSTRUMENT CONFIGURATION (SET-UP MODE)
NOTE: All congurations can also be modied through DataView® software.
The instrument must be configured the first time it is used. The configuration is saved in memory when the instrument is turned
OFF.
1. Press the
2. Set the display language by pressing the yellow button corresponding to the screen language icons.
3. The parameter that is ready to be congured will be highlighted in yellow. To move to a dierent parameter, use the ▲
and ▼ buttons.
4. Press the Enter button to select a parameter.
5. Use the ▲, ▼, ◄, and ► buttons to change a value or setting; and the button to save the setting.
6. When nished, return to the Conguration menu by pressing the
PARAMETERFUNCTION
Date / Time
Display
Calculation
Methods
Electrical Hook-Up
button to congure the unit. The following screen appears:
Figure 3-1
button.
Sets the date and time format (see § 3.2.1)
Adjusts the contrast and brightness of the display
Defines the color of the voltage and current curves (see § 3.2.2)
Determines if harmonics are used or not used in calculations of reactive quantities (power and energy) - (see § 3.2.3)
• With harmonics: Harmonics are taken into account when calculating reactive parameters
• Without harmonics: Only the fundamental part is used for the calculation of reactive parameters
Determines the type of connection to the network (see § 3.2.4)
• 1-Phase 2-Wire
• 1-Phase 3-Wire
• 2-Phase 2-Wire
• 2-Phase 3-Wire
• 2-Phase 4-Wire
• 3-Phase 3-Wire
• 3-Phase 4-Wire
• 3-Phase 5-Wire
Power Quality Analyzer PowerPad® III Model 8435
15
Probes and Ratios
Defines the type of current probe to connect (see § 3.2.5)
• MN93: 200A
• MN193: 100A or 5A (with variable ratio)
• SR193: 1000A
• SL261: 10A and 100A range
• AmpFlex® Sensors: 3000A (measures up to 10kA)
• MiniFlex® Sensors: 3000A
• MR193: 1000AAC/1200ADC
• 5A three-phase adapter (3-channel only)
Capture Mode
Trend Mode
Configures the voltage and current thresholds (see § 3.2.66)
Selects the parameters to record (see § 3.2.77)
(Up to four configurations)
Alarm Mode
Erase Memory
About
Defines the parameters of an alarm (see § 3.2.88)
Deletes configurations, alarm settings, snapshots, and recordings (see § 3.2.9)
Displays the serial number, software and hardware version
3.2.1 DATE/TIME
The parameter denes the system date and time. The display is as follows:
Figure 3-2
The Date/Time eld is highlighted in yellow.
■ To change the date/time, press the button.
■ To change a value or move from one eld to another, press the ▲ or ▼ button.
■ Press thebutton to conrm conguration.
NOTE: 12/24: Display of time in 24-hour format. AM/PM: Display of time in 12-hour format. The time is followed by AM or PM.
To return to the Conguration menu, press the
button.
16
Power Quality Analyzer PowerPad® III Model 8435
3.2.2 DISPLAY
3.2.2.1 CONTRAST/BRIGHTNESS
The menu is used to dene the contrast and brightness of the display unit.
Figure 3-3
The selected eld is highlighted in yellow.
■ To modify the contrast, press ◄ or ►
■ To move to the next eld, press ▲ or ▼
■ To change the brightness, press◄ or ►
■ To return to the Conguration menu, press
3.2.2.2 COLORS
The menu is used to dene the colors of the voltage and current traces on the graph. The colors available are: green,
dark green, yellow, orange, pink, red, brown, blue, turquoise blue, dark blue, light grey, grey, dark grey, and black.
Figure 3-4
The selected eld is highlighted in yellow.
■ To select the color of the voltage and current curves, press ◄ or ►
■ To move to the next eld, press ▲ or ▼
■ To return to the Conguration menu, press
Power Quality Analyzer PowerPad® III Model 8435
17
3.2.2.3 DISPLAY SHUTDOWN (AUTO POWER OFF)
X
=
The Display Shutdown button displays a menu that denes the screen shutdown process.
Figure 4-5
Use the up and down arrow buttons to choose Automatic or Never.
■ In Automatic mode the display screen switches OFF automatically after ve minutes of inactivity when the instrument is
powered by the battery and a recording is in progress, and after ten minutes if no recording is in progress.
■ The ON/OFF button blinks to indicate that the instrument is still in operation. Press any button to relight the
screen. To return to the Conguration menu, press
.
3.2.3 CALCULATION METHODS
determines a number of variables used in calculating the reactive parameters (powers and energy).
Figure 3-6
To select SeparatedorCombined, press ▲ or ▼.
■ Separated: Harmonics are not taken into account when calculating reactive quantities.
■ Combined: Harmonics are included in the reactive values calculations.
Press thebutton to save the setting and return to the Conguration menu.
18
Power Quality Analyzer PowerPad® III Model 8435
3.2.3.1 ENERGY UNIT
This Wh screen determines the units to be used to display calculation results.
Figure 3-7
To select the units, press ▲ or ▼. Choices are:
■ Wh
■ Joule (Watt/second)
■ Nuclear toe (Tonne Oil Equipment)
■ Non-nuclear toe
■ BTU
Press to save the setting and return to the Conguration menu.
3.2.3.2 K FACTOR
This screen denes the transformer factor K. This value is used for weighting the harmonic load currents in accordance
with their eects on transformer heating. A higher factor K indicates larger harmonic heating eects.
Figure 3-8
This consists of two values, q and e:
■ q (allowable values are 1.5, 1.6, and 1.7)
■ e (allowable values are 0.05, 0.06, 0.07, 0.08, 0.09, and 0.10)
Press ▲ or ▼ to select q or e; press ◄ or ► to select values for these parameters. Press to save the settings and
return to the Conguration menu.
Power Quality Analyzer PowerPad® III Model 8435
19
3.2.3.3 PHASE HARMONIC RATIOS
You can choose to calculate, display, and report harmonics as a ratio of the fundamental value, or in absolute value.
Figure 3-9
Press ▲ or ▼ to toggle between Fundamental Value as Reference (%f) and Total Value as Reference (%r). Press
to save the setting and return to the Conguration menu.
3.2.3.4 LONG-TERM FLICKER
This screen denes whether a xed or sliding window is used to calculate long-term icker (rapid uctuations in the power
supply).
Figure 3-10
Press ▲ or ▼ to toggle between Sliding Window and Fixed Window. Press to save the setting and return to the
Conguration menu.
20
Power Quality Analyzer PowerPad® III Model 8435
3.2.4 ELECTRICAL HOOK-UP
L1
L2
L3
N
V1V2
3V
V2V3
V3V1
ELECTRICAL HOOK-UP
3-phase 5-wire
3Φ
03/10/14 10:26
am
L1
L1
L1
The menu is used to dene how the instrument is connected, according to the type of network.
Figure 4-11
Several electrical diagrams can be selected. Use the arrow buttons to choose a connection. One or more types of network
correspond to each distribution system.
These connection choices allow you to connect the instrument to all existing networks.
Press to conrm the selection and return to the Conguration menu.
NOTE: This setting cannot be modified when the instrument is recording, and/or searching for alarms.
Distribution SystemSource
Single-phase 2-wire (L1 and N)
L1
N
Single-phase 3-wire (L1, N and ground)
L1
N
Split-phase 2-wire (L1 and L2)
L1
L2
Single-phase 2-wire non-grounded neutral
Single-phase 3-wire grounded neutral
Split-phase 2-wire
N
N
GND
L2
Power Quality Analyzer PowerPad® III Model 8435
3-phase open star 2-wire
L1
L2
21
Distribution SystemSource
L1
N
L1
N
Split-phase 3-wire (L1, L2 and N)
L1
L2
N
Split-phase 3-wire non-grounded neutral
3-phase open star 3-wire non-grounded neutral
3-phase high leg delta 3-wire non-grounded neutral
3-phase open high leg delta 3-wire non-grounded neutral
N
L2
L1
L2
L1
N
L2
L1
N
L2
Split-phase 4-wire (L1, L2, N and ground)
L1
L2
N
Split-phase 4-wire grounded neutral
3-phase open star 4-wire grounded neutral
3-phase high leg delta 4-wire grounded neutral
3-phase open high leg delta 4-wire grounded neutral
N
GND
L2
L1
GND
L2
L1
N
GND
L2
L1
N
GND
L2
22
Power Quality Analyzer PowerPad® III Model 8435
Distribution SystemSource
L3
L3
L3
L3
3-phase star 3-wire
3-phase delta 3-wire
3-phase 3-wire (L1, L2 and L3)
L1
L2
L1
L2
L1
L2
L3
3A
A1A2
A2A3
A3A1
Indicate which 2 current sensors will be
connected: A1 and A2, or A2 and A3,
or A3 and A1.
Two-wattmeter method or two-element
method or Aron method.
The third sensor is not necessary if the
other two are of the same type, same
range, and same ratio. Otherwise, the third
sensor must be connected to make current
measurements.
L3
3-phase open delta 3-wire
L1
L2
3-phase open delta 3-wire grounded junction of phases
L1
L2
3-phase open delta 3-wire grounded corner of phase
L1
L2
L3
3-phase high leg delta 3-wire
L1
Power Quality Analyzer PowerPad® III Model 8435
3-phase open high leg delta 3-wire
L2
L3
L1
L2
23
Distribution SystemSource
L3
L3
L3
L3
L3
L3
3-phase 4-wire (L1, L2, L3 and N)
L1
L2
L3
N
3V
V1V2
V2V3
V3V1
Indicate which voltages will be connected:
all 3 (3V) or only 2 (V1 and V2, or V2 and
V3, or V3 and V1).
If only two of the three voltages are con-
nected, the three phase voltages must be
balanced (2½-element method)
3-phase star 4-wire non-grounded neutral
3-phase open high leg delta 4-wire non-grounded neutral
3-phase high leg delta 4-wire non-grounded neutral
N
L1
L2
L1
N
L2
L1
N
L2
3-phase 5-wire (L1, L2, L3, N and ground)
L1
L2
L3
N
Indicate which voltages will be connected:
all 3 (3V) or only 2 (V1 and V2, or V2 and
V3, or V3 and V1).
If only two of the three voltages are con-
nected, the three phase voltages must be
balanced (2½-element method)
.
3V
V1V2
V2V3
V3V1
3-phase star 5-wire grounded neutral
3-phase open high leg delta 5-wire grounded neutral
3-phase high leg delta 5-wire grounded neutral
N
L1
GND
L2
L1
N
GND
L2
L1
N
GND
L2
24
Power Quality Analyzer PowerPad® III Model 8435
3.2.5 PROBES AND RATIOS
3.2.5.1 CURRENT PROBES
The screen denes the current probes and ratios. It automatically displays the current probe models detected by the
instrument. It can also be used to dene the transformation ratio (sensitivity) of certain current sensors (SL261 clamp). To
select current probes and current ratios settings, press the A button to display the following screen:
Figure 3-12
The probe choices that are available are:
MN93 clamp: 200AAC
MN193 clamp: 100 or 5AAC
SR193 clamp: 1000AAC
J93 clamp: 3500AAC/5000ADC
AmpFlex® A193: 6500AAC/10,000AAC
MiniFlex® MA193: 6500AAC
MR193 clamp: 1000A/1200AAC/DC
SL261 clamp: 100A (sensitivity 10mV/AAC/DC)
SL261 clamp: 10A (sensitivity 100mV/AAC/DC)
Three phase adapter: 5AAC
If an MN193 clamp, 5A range, or an Adapter is used, the current ratio setting is proposed automatically. The conguration
is done as follows:
■ To select the channel, press the ▲ or ▼ buttons.
■ To congure the primary circuit current (1 to 60,000A) / secondary circuit current (1A, 2A or 5A) transformation ratio,
press. To select the elds, use ◄ or ►.
■ To adjust the values, use ▲ or ▼. Proceed in the same way for the primary and secondary circuit currents.
■ To validate, press (the parameters are applied only if conrmed).
The primary current cannot be less than the secondary current.
Power Quality Analyzer PowerPad® III Model 8435
25
3.2.5.2 VOLTAGE RATIOS
The screen also enables you to dene the voltage ratios. To do this, press the V button to display the following
screen.
Figure 3-13
The ratios can be programmed to be the same in all channels, or dierent in one or more of them. The following steps
describe a 3-phase 5-wire hook-up; the available channels are determined by connection type and may be dierent for
your connection.
1. Highlight Ratio Set-up and press. Then use the ▲ or ▼ buttons to display the available options:
V1+V2+V3+VN: each channel and the neutral has a different ratio. You are prompted to enter a separate ratio for
each channel (1, 2, and 3) and the neutral (N) as shown in Figure 3-13.
3V + VN: all channels have the same ratio and the neutral has a different ratio. You are prompted to enter a ratio for
the channels and another ratio for the neutral.
4V: all channels have the same ratio. You are prompted to enter the ratio that will apply to all.
4V 1/1: all channels have the same 1/1 ratio.
2. Select the set-up and validate by pressing .
3. To select channels, press the ▲ and ▼ buttons to highlight the channel eld and press. Then use the ◄ and ►
buttons to navigate within the field, and the ▲ and ▼ buttons to adjust values. When you are finished with the field,
press to validate (the parameter must be validated to be applied).
4. To return to the Conguration menu, press
.
For the primary voltage (in kV) and the secondary voltage (in V), it is possible to specify the use of the multiplier 1/√3. If
the phase-to-neutral voltage ratios of phases 1, 2, and 3 are not identical; then all measurements and curves concerning
the phase-to-phase voltages are suppressed.
26
Power Quality Analyzer PowerPad® III Model 8435
3.2.6 CAPTURE MODE
The
Capture Mode scr
een denes voltage and current thresholds for transient and Inrush captures.
3.2.6.1 TRANSIENT VOLTAGE THRESHOLDS
By default, the Capture Mode screen opens with the Transient Voltage Thresholds screen displayed.
Figure 3-14
1. Select Threshold Set-up, highlighted in yellow, by pressing . The ▲ and ▼ buttons appear in the eld to enable
you to select the threshold conguration.
2. Use the ▲ or ▼ button to go from one type of conguration to another.
3. Press to validate the choice of conguration.
4. Select the eld of the rst threshold using the ▲ or ▼ button. The selected eld is highlighted in yellow. Press to
displays the arrows in the eld.
5. Use the ▲ or ▼ button to increment or decrement a value and ◄ or ► to go to the next digit in the value.
6. Press to validate the programming of the threshold.
To return to the Conguration screen, press
.
The voltage thresholds can be congured in V or in kV.
Power Quality Analyzer PowerPad® III Model 8435
27
3.2.6.2 TRANSIENT CURRENT THRESHOLDS
To display the Transient Current Thresholds screen, open the Capture Mode screen and press the A button.
Figure 3-15
1. The process for setting the transient current threshold is similar to setting the transient voltage threshold. Select
Threshold Set-up, highlighted in yellow, by pressing . The ▲ and ▼ buttons appear in the eld to enable you to
select the threshold conguration.
2. Use the ▲ or ▼ button to go from one type of conguration to another.
3. Press to validate the choice of conguration.
4. Select the eld of the rst threshold using the ▲ or ▼ button. The selected eld is highlighted in yellow. Press to
displays the arrows in the eld.
5. Use the ▲ or ▼ button to increment or decrement a value and ◄ or ► to go to the next digit in the value.
6. Press to validate the programming of the threshold.
To return to the Conguration screen, press
.
You can congure the current thresholds in mA, A, or kA.
28
Power Quality Analyzer PowerPad® III Model 8435
Loading...
+ 105 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.