Chauvin Arnoux®, Inc. d.b.a. AEMC® Instruments certifies that
this instrument has been calibrated using standards and
instruments traceable to international standards.
We guarantee that at the time of shipping your instrument
has met its published specifications.
An N.I.S.T. traceable certificate may be requested at
the time of purchase, or obtained by returning the
instrument to our repair and calibration facility, for a
nominal charge.
The recommended calibration interval for this instrument is
12 months and begins on the date of receipt by the
customer. For recalibration, please use our calibration
services. Refer to our repair and calibration section at
RECEIVING YOUR SHIPMENT ....................................................................................................................................................... 5
ORDERING INFORMATION ............................................................................................................................................................ 5
1. GETTING STARTED .................................................................................................................................................................... 6
1.4. Turning the Instrument ON/OFF ........................................................................................................................................ 7
1.5. Function Buttons ................................................................................................................................................................ 7
2.1. Making a Measurement ..................................................................................................................................................... 9
2.1.1. Selecting Temperature Units ............................................................................................................................9
2.3. 1D and 3D Modes .............................................................................................................................................................. 9
2.3.1. 1D and 3D Modes Activation ............................................................................................................................9
2.3.2. Visible and Audible Warnings .......................................................................................................................... 10
2.6. MIN MAX Function ........................................................................................................................................................... 12
2.7. HOLD Function ................................................................................................................................................................ 12
2.8. M_REC (Manual Recording) Function ............................................................................................................................. 13
2.11.1. OL (Overload) Symbol .................................................................................................................................. 14
2.11.2. Err (Error) Symbol ......................................................................................................................................... 14
2.11.3. MEM_FULL (Memory Full) Symbol ............................................................................................................... 14
3.2. Connecting the Instrument to a Computer ....................................................................................................................... 15
4.2.1. CO2 Measurements ........................................................................................................................................ 16
4.2.2. Influences on the measurements of CO2. ....................................................................................................... 16
4.2.3. Temperature Measurements ............................................................................................................................ 16
4.4. Power Supply ................................................................................................................................................................... 17
4.8. Compliance with International Standards......................................................................................................................... 18
REPAIR AND CALIBRATION .......................................................................................................................................................... 20
TECHNICAL AND SALES ASSISTANCE ..................................................................................................................................... 20
Thank you for purchasing the Air Quality Logger Model 1510. For best results from your instrument:
read these operating instructions carefully
comply with the precautions for use
WARNING, risk of DANGER! The operator must refer to these instructions whenever this danger
symbol appears.
Information or useful tip.
Battery.
Magnet.
The product has been declared recyclable after analysis of its life cycle in accordance with the
ISO14040 standard.
AEMC has adopted an Eco-Design approach in order to design this appliance. Analysis of the
complete lifecycle has enabled us to control and optimize the effects of the product on the
environment. In particular this appliance exceeds regulation requirements with respect to recycling
and reuse.
Indicates conformity with European directives and with regulations covering EMC.
Indicates that, in the European Union, the instrument must undergo selective disposal in compliance
with Directive WEEE 2002/96/EC. This instrument must not be treated as household waste.
Definition of measurement categories
CAT IV corresponds to measurements taken at the source of low-voltage installations.
Example: power feeders and protection devices.
CAT III corresponds to measurements on building installations.
Example: distribution panel, circuit-breakers, machines or fixed industrial devices.
CAT II corresponds to measurements taken on circuits directly connected to low-voltage
installations.
Example: power supply to electro-domestic devices and portable tools.
Precautions
This instrument is compliant with safety standard IEC 61010-2-030, for voltages up to 5V with respect to
ground. Failure to observe the following safety instructions may result in electric shock, fire, explosion, and
damage to the instrument and/or the installation in which it is located.
The operator and/or the responsible authority must carefully read and clearly understand all
precautions to be taken prior to using the instrument. Thorough knowledge and awareness of
electrical hazards are essential when using this instrument.
Observe the conditions of use, including temperature, relative humidity, altitude, pollution degree,
and location of use.
Do not use the instrument if it appears damaged, incomplete, or improperly closed.
Before each use, check the condition of the housing and accessories. Any item on which the
insulation is deteriorated (even partially) must be set aside for repair or disposal.
Never make measurements in atmospheres contaminated by solvents, which could damage the
sensor.
All troubleshooting and metrological checks must be done by accredited personnel.
4 Air Quality Logger Model 1510
Receiving Your Shipment
Upon receiving your shipment, make sure that the contents are consistent with the packing list. Notify your
distributor of any missing items. If the equipment appears to be damaged, file a claim immediately with the
carrier and notify your distributor at once, giving a detailed description of any damage. Save the damaged
packing container to substantiate your claim.
Ordering Information
Air Quality Logger Model 1510 (gray)………..….….………..………….…..…..….......…….... Cat. #2138.08
Air Quality Logger Model 1510 (white)….…………....….….…………………..….….........….. Cat. #2138.09
Both models include soft carrying pouch, two AA batteries, 6 ft. USB cable, quick start guide, USB drive
with DataView
®
software and user manual.
Replacement Parts:
Cable – Replacement 6 ft. (1.8m) USB…………….…..……….….............................…........ Cat. #2138.66
Adapter, Bluetooth USB for Simple Logger II 4-Channel Models,
Models 6417, 1510 & L452 ….………………………………………..………………………… Cat. #2126.45
Wall Mount for Models 1510 & L452 (gray or black) ....………..…..……….........………....... Cat. #2138.61
Desk Stand for Model 1510 (white) .………………..…..……………….…….…….................. Cat. #2138.62
Calibration Kit for Model 1510 (includes calibration software CD)………............……….......Cat. #2138.63
Metal Carrying Case …………………..……………..…..….…………….………….................. Cat. #2138.65
Wall Mount for Model 1510 (white) .…………….…..….…………….………………................ Cat. #2138.67
Adapter – USB to Wall Plug …….….…….……………….…...….….……….............……....... Cat. #2153.78
For the accessories and replacement parts, visit our web site: www.aemc.com
Air Quality Logger Model 1510 5
1. GETTING STARTED
1.1. Battery Installation
The instrument accepts two AA or LR6 alkaline batteries.
2 1 3
4
To change the batteries:
1. Press the tab of the battery compartment cover and lift it clear.
2. Remove the battery compartment cover.
3. Insert the new batteries, ensuring correct polarity.
4. Close the battery compartment cover; ensuring it is completely and correctly closed.
1. Non-skid pads
2. “Tear-drop” notch to hang
instrument
3. Battery compartment cover
4. Magnets for mounting to a
metallic surface
1.2. Instrument Front Panel
1. Temperature and humidity sensors
1
7
2
3
8
4
9
5
6
10
2. LCD display
3. MODE/temperature units button
4. Backlight/ECO mode button
5. Record button
6. ON/OFF button
7. CO
8. Activate/deactivate buzzer
9. HOLD/Bluetooth enable/disable
10. MAX/MIN button
sensor
2
button
6 Air Quality Logger Model 1510
1.3. Instrument Functions
you cannot turn OFF the instrument when it is in HOLD or in recording mode.
Starts manual recording.
buzzer.
Long press activates/deactivates Bluetooth.
Long press exits MAX MIN mode.
The Model 1510 measures:
Concentration of carbon dioxide (CO2) in the air
Ambient temperature
Relative humidity
The instrument uses these measurements to determine and display the current comfort level.
DataView with the Data Logger Control Panel software can be installed on a computer to allow you to
configure the instrument, view measurements in real time, download data from the instrument, and create
reports.
1.4. Turning the Instrument ON/OFF
ON: Press the button.
OFF: Press the button for >2 seconds when the instrument is ON. Note that
In Portable mode, the instrument automatically turns OFF after 15 minutes of inactivity. This
feature is de-activated in other modes (1D or 3D, ECO, P_REC, MIN MAX, and M_REC), if
the instrument is actively communicating via Bluetooth or USB, or is running on USB power.
1.5. Function Buttons
On each button, a short press activates the top function and a long press (>2 seconds) activates the lower
function.
Button Function
Short press enables 1D and 3D modes.
Long press changes temperature units.
Pressing this button while holding down activates/deactivates alarm buzzer.
REC
MAX
MIN
Short press turns on the backlight.
Long press enables/disables energy-saving mode.
Holding down this button and then pressing activates/deactivates alarm
Short press freezes the display.
Short press enters MAX MIN mode; measurement values continue to be
displayed.
Second press displays the maximum value.
Third press displays minimum value.
Fourth press returns to normal measurement operation.
Air Quality Logger Model 1510 7
1.6. Display
Steady: recording
MIN MAX function: detection of minimum and maximum values
Blinking: Batteries low
or of the USB connection.
1
2
Symbol Function
3
1. CO2 measurement
2. Temperature
3. Relative humidity
MEM_FULL
1D
3D
MAX
MIN
ECO
P_REC
M_REC
HOLD
MIN MAX
ppm
Memory full
Monitor CO2 level
Monitor CO2, temperature, and humidity comfort zone thresholds
Maximum value
Minimum value
Energy-saving operating mode
Programmed recording
Blinking: waiting to start recording
Manual recording
Hold of the display
Unit of concentration of CO2 in the air in parts per million
Buzzer active
Bluetooth
Blinking: waiting for connection
Steady:
connected.
Steady: Indicator of external power supply by connection of the mains adapter
High threshold passed
Low threshold passed
Air quality/comfort level indicator
8 Air Quality Logger Model 1510
2. OPERATION
2.1. Making a Measurement
When the instrument is turned ON, the message CO2 is displayed on the screen (and function buttons are
disabled) until the first CO
If the instrument is exposed to large variations of temperature, after the temperature
stabilizes, wait 20 minutes before starting measurements.
The instrument has four operating modes: Portable mode, 1D and 3D modes, ECO mode, and P_REC
mode. It also has several functions that can be used in certain modes: MIN MAX, HOLD, M_REC, and
Backlight.
2.1.1. Selecting Temperature Units
Long-press the button to toggle between °F and °C. The setting remains in effect when the instrument
is turned OFF.
2.2. Portable Mode
When you turn ON the instrument, it automatically starts up in Portable mode. In this mode the instrument
makes a measurement every 15 seconds, more frequently than in other modes. It is therefore highly
responsive to changes in air quality. This mode is used to inspect several rooms one after another. Install
the instrument in a room and wait for the CO
measurement appears. Updated measurements then appear automatically.
2
measurement to stabilize (approximately 10 minutes).
2
2.3. 1D and 3D Modes
1D mode: surveillance of CO2 level.
3D mode: surveillance of CO2 level, temperature, and humidity.
The audible and/or visual warnings alert you to overages. The level of CO2 is measured every minute.
2.3.1. 1D and 3D Modes Activation
1. Press to activate 1D mode. The symbol 1D appears on the
display.
2. Press a second time to activate 3D mode. The symbol 3D
appears on the display.
3. Press a third time to return to normal operation.
Air Quality Logger Model 1510 9
2.3.2. Visible and Audible Warnings
Air quality good (CO2 level below 1000ppm)
Air quality poor (CO2 level over 1700ppm)
Air quality good (no overages of CO2, temperature, and humidity comfort zones)
Air quality poor (overage of temperature and humidity thresholds)
1D Mode
Optimum comfort zone indicator:
Backlight OFF.
Air quality average (CO2 level between 1000 and 1700ppm)
Average comfort zone indicator:
Arrow indicates overage of low CO
Backlight blinks orange.
threshold.
2
"Outside optimum comfort zone" indicator:
Arrow indicates overage of high CO
Buzzer sounds (if enabled).
Backlight blinks red.
3D mode
Optimum comfort zone indicator:
Backlight OFF.
"Outside optimum comfort zone" indicator:
Arrow in the direction of the overage for each quantity outside
the comfort zone.
Buzzer sounds (if enabled).
Backlight blinks red.
Backlight is OFF in ECO mode
threshold.
2
Backlight is OFF in ECO mode
Backlight is OFF in ECO mode
.
.
.
For 1D and 3D modes, if the buzzer is active it will sound
displayed.
intermittently when is
10 Air Quality Logger Model 1510
The Porcher diagram defines the hygrothermal comfort zones:
Relative Humidity (%RH)
Absolute humidity (g/kg)
Dry-bulb temperature (°C)
2.4. ECO (Energy Saving) Mode
ECO mode saves energy when the Model 1510 measures air quality over a long period. The backlight and
the buzzer are deactivated, and CO
addition, the instrument automatically activates “sleep” mode between 4:30 p.m. to 8:30 a.m. These times
can be changed using DataView software (see § 3).
is measured only once every 10 minutes. This saves the batteries. In
2
1. Long-press (>2 seconds) to activate ECO mode. The symbol
ECO appears on the display.
2. Long-press a second time to return to normal operation.
In ECO mode, measurements are more sensitive to instantaneous variations of CO
. For
2
example, breathing near the instrument can affect measurement. It will be necessary to wait
for the next measurement to recover the true concentration of CO
in the room.
2
2.5. P_REC (Programmed Recording) Mode
When the instrument is connected to a computer, you can program two types of recordings (see §3):
Locked recording, in which the instrument only displays the P_REC symbol (blinking before
recording starts, then steady while recording) and the keys are inactive. At the end of recording,
the instrument automatically turns OFF.
Unlocked recording, in which the instrument displays the measurements. The P_REC symbol is
displayed, blinking before recording starts, then steady while recording. The instrument
otherwise operates normally. It is not possible to change the mode, but the MIN MAX, HOLD,
and backlight functions can be used.
In P_REC mode, pressing the button has no effect.
Air Quality Logger Model 1510 11
2.6. MIN MAX Function
This function displays the maximum and minimum measured values in addition to the current value: the
instrument compares each new measurement to those displayed. If the new measurement is greater than
the current MAX value or less than the current MIN value, it replaces them in the display.
The MIN MAX function cannot be used in the 1D and 3D modes
1. Press to activate the MIN MAX function. The symbol MIN MAX
appears on the display.-
2. Press a second time to display maximum and minimum
measurements for CO
.
2
3. Press a third time to display maximum and minimum
measurements for temperature.
4. Press a fourth time to display maximum and minimum
measurements for humidity. Additional presses of repeat this
cycle.
.
5. Long-press to exit MIN MAX and return to normal measurement
display.
2.7. HOLD Function
Pressing HOLD freezes the displayed measurement on the screen. This function does not interrupt
recording or operation of the current mode; but access to the other functions is disabled.
1. Press to activate HOLD. The symbol HOLD appears and the
current measurement remains on the screen.
2. Press a second time to return to normal measurement display.
12 Air Quality Logger Model 1510
2.8. M_REC (Manual Recording) Function
You can start a recording session manually. All measurement data is recorded in the instrument at the rate
determined by the current mode (see § 4.3).
You can start a manual recording session in any mode except P_REC. After recording is in
progress, you cannot change modes
.
1. Long-press REC (>2 seconds) to start a manual recording. The
symbol M_REC appears on the display.
2. Long-press REC a second time to stop recording.
2.9. Backlight
Press to turn on the backlight. The backlight automatically turns OFF after 10 seconds.
The backlight function is disabled in ECO mode.
2.10. Alarm Buzzer
1. To activate the alarm buzzer, press the button; and
while holding it down press . The icon appears on
the screen.
2. To deactivate the alarm, press and while holding it
down press .
The alarm buzzer function is disabled in ECO mode.
Air Quality Logger Model 1510 13
2.11. Error Messages
2.11.1. OL (Overload) Symbol
The OL symbol appears when the measurement is outside the range measured by the instrument:
CO2 > 5000ppm
14 < T < 140°F (-10 < T < 60°C)
5% < RH < 95%
2.11.2. Err (Error) Symbol
If an Err error is displayed, turn the instrument OFF and back ON. If the error persists, the instrument must
be sent in for repair (see Repair and Calibration at the end of this manual).
2.11.3. MEM_FULL (Memory Full) Symbol
When the memory is full, the MEM_FULL symbol appears on screen.
If a recording is in progress, it automatically stops when the memory becomes full. The memory must be
erased before starting a new recording.
You can erase memory manually or via DataView. To do this in DataView, install the software (see § 3)
and consult the Help system.
To erase memory manually, do the following:
1. Connect the instrument to a computer via the USB cable.
2. Open the computer’s file system.
3. Locate and open the folder named CA1510.
4. Delete any unnecessary files to free memory space.
14 Air Quality Logger Model 1510
3. DATAVIEW
DataView is required to perform several basic setup tasks including connecting the instrument to a
computer and setting the time and date on the instrument. In addition, DataView allows you to:
Configure and schedule a recording session on the instrument.
Download recorded data from the instrument to the computer.
Generate reports from downloaded data.
View instrument measurements in real time on the computer.
3.1. DataView Installation
1. Insert the USB drive that comes with the instrument into a USB port on your computer.
2. If Autorun is enabled, an AutoPlay window appears on your screen. Click “Open folder to view files”
to display the DataView folder. If Autorun is not enabled or allowed, use Windows Explorer to locate
and open the USB drive labeled “DataView.”
3. When the DataView folder is open, double-click Setup.exe in the USB drive root directory.
4. The Setup screen appears. This enables you to select the language for the setup UI. You can
also select additional install options (each option is explained in the Description field). Make your
selections and click Install.
5. Click OK to confirm setup. The InstallShield Wizard screen appears. This leads you through
DataView install. As you complete these screens, be sure to check Data Loggers when prompted to
select features to install.
6. When the InstallShield Wizard finishes installing DataView, the Setup screen appears. Click Exit to
close. The DataView folder appears on your computer desktop.
3.2. Connecting the Instrument to a Computer
You can connect the instrument to a computer either through USB cable (provided with the instrument) or
Bluetooth
USB:
Bluetooth:
Connecting the instrument via Bluetooth requires a Bluetooth USB adapter, Cat. #2126.45 (sold
separately). When the dongle is installed, do the following:
After the USB cable is connected or Bluetooth is activated, proceed as follows:
®
. The first two steps of the connection procedure depend on the connection type:
1. Connect the instrument to an available USB port using the supplied cable.
2. Turn ON the instrument. If this is the first time this instrument has been connected to this
computer, drivers will be installed. Wait for this to finish before proceeding with step 3 below.
1. Turn ON the instrument by pressing the button.
2. Activate Bluetooth on the instrument by pressing the button until the symbol appears in the
LCD.
3. Open the DataView folder on your desktop. This displays a list of icons for the Control Panel(s)
installed with DataView.
4. Open the DataView Data Logger Control Panel by clicking the icon.
5. In the menu bar at the top of the screen, select Help. In the drop-down menu that appears, click the
option Help Topics. This opens the Data Logger Control Panel Help system.
6. Use the Contents window in the Help system to locate and open the topic "Connecting to an
Instrument." This provides instructions explaining how to connect your instrument to the computer.
7. When the instrument is connected, its name appears in the Data Logger Network folder in the left
side of the Control Panel. A green check mark appears next to the name indicating it is currently
connected.
For instructions about using DataView with the instrument, consult the Data Logger Control Panel Help
system.
Air Quality Logger Model 1510 15
4. SPECIFICATIONS
Supply voltage
3 ± 0.5V
Air pollution
No pollution (CO, solvents, etc.)
Type of sensor:
Dual-beam infrared cell
Measuring principle:
Non-dispersive infrared (NDIR) technology
Measurement range:
0 to 5000ppm
Intrinsic uncertainty:
±3% ± 50ppm at 77°F (25°C) and 1013mbar
In ECO mode: ±3% ± 80ppm at 77°F (25°C) and 1013mbar
Response time at 63%:
195 seconds
Resolution (R):
1ppm
Type of sensor:
CMOS
Measurement range:
14 to 140°F (-10 to +60°C)
Intrinsic uncertainty:
±0.5°C at 50% RH
Influence of relative
humidity
±0.5°C ± R from 10 to 40% RH
Outside this
Resolution (R):
0.1°C or 0.1°F
Type of sensor:
Capacitive
Measurement range:
5 to 95% RH
Intrinsic uncertainty:
±2% RH ± R from 10 to 90% RH
±3% RH ± R outside of the range stated above.
Resolution (R):
0.1% RH
Measurement
hysteresis
±1% RH
N
produce
between
Rate of increase of
intrinsic
< 0.5% RH/year
4.1. Reference Conditions
Quantity of influence Reference conditions
The intrinsic uncertainty is the error specified for the reference conditions.
4.2. Measurement Characteristics
4.2.1. CO2 Measurements
4.2.2. Influences on the Measurements of CO2
The influence of the temperature is 1 ppm/°C from -10 to +45°C (14 to 113°F).
The influence of the atmospheric pressure is:
2 real
= CO
2 measured
CO
where P = pressure in mbar
x (1 + (1013 - P) x 0.0014)
4.2.3. Temperature Measurements
:
range: ±0.032 x (T - 25°C) ± R
4.2.4. Humidity Measurement
:
OTE: Prolonged exposure to humidity outside the 10% to 80% range may
measurement bias up to ±3% RH. This bias disappears after 5 days
68 to 86°F (20 and 30°C) and 40 and 75% RH.
uncertainty:
16 Air Quality Logger Model 1510
±5
±5
±5
±4
±4
±3
±4
±4
±4
±4
±4
±4
±4
±5
±5
±5
±4
±3
±2
±2
±2
±3
±3
±3
±3
±4
±4
±4
±5
±4
±4
±3
±2
±2
±2
±3
±3
±3
±3
±4
±4
±4
±4
±4
±4
±3
±2
±2
±2
±3
±3
±3
±3
±3
±3
±4
±4
±4
±4
±3
±2
±2
±2
±3
±3
±3
±3
±3
±3
±3
±4
±4
±4
±3
±2
±2
±2
±3
±3
±3
±3
±3
±3
±3
±4
±4
±3
±3
±2
±2
±2
±2
±3
±3
±3
±3
±3
±3
±4
±3
±3
±2
±2
±2
±2
±2
±2
±2
±2
±2
±2
±3
±4
±2
±2
±2
±2
±2
±2
±2
±2
±2
±2
±2
±2
±3
±4
±2
±2
±2
±2
±2
±2
±2
±2
±2
±2
±2
±2
±3
±4
±2
±2
±2
±2
±2
±2
±2
±2
±2
±2
±2
±2
±3
±4
±2
±2
±2
±2
±2
±2
±2
±2
±2
±2
±2
±2
±3
±4
±3
±3
±2
±2
±2
±2
±2
±2
±2
±2
±2
±2
±3
±4
±3
±3
±2
±2
±2
±2
±2
±2
±2
±2
±2
±2
±3
±4
±4
±3
±2
±2
±2
±2
±2
±2
±2
±2
±2
±2
±3
±4
±4
±4
±3
±2
±2
±2
±3
±3
±3
±3
±3
±3
±3
±5
±4
±4
±3
±2
±2
±2
±3
±3
±3
±3
±3
±4
±4
±8
±5
±5
±4
±3
±2
±3
±4
±4
±4
±4
±4
±4
±4
±8
±8
±8
±6
±5
±3
±5
±5
±5
±5
±5
±5
±5
±5
4.2.5. Influence of the temperature on the relative humidity measurement
ECO
Instantaneous, not
Every 10 minutes
Every 5 seconds
Relative humidity (%)
100
90
80
70
60
50
40
30
20
10
0
0 10 20 30 40 50 60 70
Temperature (°C)
4.3. Measurement Modes
Measurement
mode
Portable
Displayed value Display rate of CO2
values
Mean of 11 successive
Every 15 seconds Every 2 seconds
Polling rate for temperature
and RH sensors
measurements
1D and 3D Mean Every minute Every 2 seconds
averaged
P_REC Mean User programmable User programmable
4.4. Power Supply
Batteries: 2 x 1.5V AA/LR6
Mean battery life (without backlighting and Bluetooth):
Portable mode: 15 days
1D or 3D mode: 45 days
ECO mode: approximately 1 year
P_REC 10 minutes mode: 45 days
Rechargeable storage batteries can be used, but the period between charges will be shorter. The batteries
are not used while the instrument is connected to USB power.
Air Quality Logger Model 1510 17
4.5. Recording
Memory: 1,000,000 measurements (8MB) FAT12
format
4.6. Environment Conditions
Use indoors
Operating range: 14 to 140°F (-10 to +60°C) and 5 to 95% RH
Range in storage
(without battery): -4 to 140°F (-20 to +60°C)
Altitude: <2000m
4.7. Mechanical Characteristics
Dimensions
(L x W x H): 125 x 65.5 x 32mm
Weight: Approximately 190g
Protection class: IP 40 per IEC60529
IK 04 per IEC50102
Free fall test: Per IEC61010-1
4.8. Compliance with International Standards
Compliant with IEC61010-1 safety standard for voltages of 50V in category II.
4.9. Electromagnetic Compatibility
Emissions and immunity in an industrial setting compliant with IEC 61326-1.
18 Air Quality Logger Model 1510
Except for batteries, the instrument contains no parts that can be replaced by
personnel who have not been specially trained and accredited. Any unauthorized
repair or replacement of a part by an “equivalent” may significantly impair safety.
5.1. Cleaning
5. MAINTENANCE
Disconnect the instrument from all sensors, cable, etc. and turn it OFF.
Use a soft cloth, dampened with soapy water. Rinse with a damp cloth and dry
rapidly with a dry cloth or forced air. Do not use alcohol, solvents, or
hydrocarbons.
Be sure to keep the two sensor intakes perfectly clean (indicated by arrows in
illustration on left).
5.2. Battery Replacement
The symbol indicates the remaining battery life. When the symbol is empty, all the batteries must
be replaced (see §1.1).
Spent batteries must not be treated as ordinary household waste. Take them to an appropriate
recycling facility.
Air Quality Logger Model 1510 19
REPAIR AND CALIBRATION
To ensure that your instrument meets factory specifications, we recommend that it be scheduled to be sent
back to our factory Service Center at one-year intervals for recalibration, or as required by other standards
or internal procedures.
For instrument repair and calibration:
You must contact our Service Center for a Customer Service Authorization Number (CSA#). This will
ensure that when your instrument arrives, it will be tracked and processed promptly. Please write the CSA#
on the outside of the shipping container. If the instrument is returned for calibration, we need to know if you
want a standard calibration or a calibration traceable to N.I.S.T. (Includes calibration certificate plus
recorded calibration data).
For North / Central / South America, Australia and New Zealand:
Ship To: Chauvin Arnoux
Costs for repair, standard calibration, and calibration traceable to N.I.S.T. are available.
NOTE: You must obtain a CSA# before returning any instrument.
TECHNICAL AND SALES ASSISTANCE
If you are experiencing any technical problems, or require any assistance with the proper operation or
application of your instrument, please call, fax, or e-mail our technical support team:
Your AEMC instrument is warranted to the owner for a period of two years from the date of original
purchase against defects in manufacture. This limited warranty is given by AEMC
distributor from whom it was purchased. This warranty is void if the unit has been tampered with, abused,
or if the defect is related to service not performed by AEMC
®
Instruments.
®
Instruments, not by the
Full warranty coverage and product registration is available on our website at:
www.aemc.com/warranty.html.
Please print the online Warranty Coverage Information for your records.
®
What AEMC
If a malfunction occurs within the warranty period, you may return the instrument to us for repair, provided
we have your warranty registration information on file or a proof of purchase. AEMC
Instruments will do:
®
Instruments will, at its
option, repair or replace the faulty material.
Warranty Repairs
What you must do to return an Instrument for Warranty Repair:
First, request a Customer Service Authorization Number (CSA#) by phone or by fax from our Service
Department (see address below), then return the instrument along with the signed CSA Form. Please write
the CSA# on the outside of the shipping container. Return the instrument, postage or shipment pre-paid to: