POWER QUALITY ANALYZERS, METERS & LOGGERS
Power & Energy Loggers PEL 100 Series
S
Model PEL 102
D
C
K
A
R
T
E
N
$
E
R
G
Y
O
M
T
E
R
V1V2 V3 N
I1I2 I3
1000V CAT III600V CAT IV
MODEL PEL 103
M
POWER & ENERGY LOGGER
ON/OFF
START/STOP
O
N
I
T
O
Monitor your energy usage & costs locally
S
T
S
O
C
E
G
N
I
R
Visit the PEL 100 Series
website for more information
on software, specications
and more!
Android™ App Available
for PEL 102, 103 & 105 !
IP 54
Rated
Y
S
R
G
ETHERNET
Configure Measurements and Recordings
Display Data in Real-Time
For Use on any Device with an Android Platform
KIT OPTION INCLUDES
PEL 102 Kit
Small classic tool bag, three MiniFlex® MA193-10-BK sensors, 5 ft USB cable, four black test leads and alligator clips, power cord, twelve color-coded ID markers, Multifix mounting system, safety data sheet, compliance sheet, SD-Card with USB-SD-Card reader, quick start user guide, and USB stick supplied with DataView user manual.
®
software and
FEATURES
Simple-to-use, single-, dual- (split-phase) and three-phase
(Y, ) power & energy loggers
Designed to work in 1000V CAT III and 600V CAT IV
environments and fits in many distribution panels
Power measurements: VA, W and var
Energy measurements: VAh, Wh (source, load) and varh
(quadrant indication)
®
Updated features in DataView
real-time communication with a PC and report generation with pre-defined or user defined templates
USB, LAN, Ethernet and Bluetooth
(Class 1 wireless communication, up to 300 ft away)
Satisfies the monitoring requirements of NEC Code 220.87
AC/DC Current Probe, Model J93, is available for measuring
current up to 5000A
dc & 3500Aac
Power adapter allows the PEL 102 to be powered from a
phase measurement input
Provides all the necessary functions for power and energy
data logging for 50Hz, 60Hz, 400Hz and DC distribution systems
Automatic recognition of the connected current
sensors/probes
software for configuring
www.aemc.com
Technical Assistance (800) 343-1391
Page 2
Model PEL 103
A
T
I
N
A
F
E
T
TM
Large Functional Displays
Measurement ModeInformation Mode
IP 54
Rated
Y
S
R
G
ETHERNET
FEATURES
Simple-to-use, single-, dual- (split-phase)
and three-phase (Y, ) power & energy loggers
Designed to work in 1000V CAT III and 600V CAT IV
environments and fits in many distribution panels
Power measurements: VA, W and var
Energy measurements: VAh, Wh
(source, load) and varh (quadrant indication)
®
Updated features in DataView
real-time communication with a PC and report generation with pre-defined or user defined templates
USB, LAN, Ethernet and Bluetooth
(Class 1 wireless communication, up to 300 ft away)
Satisfies the monitoring requirements
of NEC Code 220.87
PEL 103 can be configured from front panel, DataView
control panel or the FREE Android™ application
AC/DC Current Probe, Model J93, is available
for measuring current up to 5000A
Provides all the necessary functions for power
and energy data logging for 50Hz, 60Hz, 400Hz and DC distribution systems
Automatic recognition of the connected current
sensors/probes
software for configuring
dc & 3500Aac
Hook up, voltage and current ratios and aggregation period can be configured from the front panel of the PEL 103.
Max Mode Harmonic Mode
Max values for voltage, current (including neutral current), power and harmonics.
®
Real-time updates are displayed for voltage, current, power, frequency, power factor and tangent.
Total Harmonic Distortion (THD) can be displayed by phase or phase to phase. Neutral current THD can also be displayed
.
KIT OPTION INCLUDES
PEL 103 Kit
Small classic tool bag, three MiniFlex® MA193-10-BK
sensors, 5 ft USB cable, four black test leads and
alligator clips, power cord,
twelve color-coded ID
markers, Multifix mounting
system, safety data
sheet, compliance sheet,
SD-Card with USB-SD­Card reader, quick start
user guide, and USB stick
supplied with DataView
software and user
manual.
®
Technical Assistance (800) 343-1391
www.aemc.com
Page 3
POWER QUALITY ANALYZERS, METERS & LOGGERS
TM
POWER QUALITY ANALYZERS, METERS & LOGGERS
Power & Energy Loggers PEL 100 Series
Three-Phase Power Quality Analyzer
Model PEL 105
Waterproof...great for outdoor use!
IP 67
Rated
Phase
ETHERNET
Powered
KIT OPTION INCLUDES
PEL 105 Kit
Extra-large tool bag, accessory pouch, 5 ft USB cable, five 10 ft black voltage leads (waterproof cap) with alligator clips, power adapter 110/230V with US power cord, four water-tight AmpFlex #2137.59 only), 9.6V NiMH battery, SD-card, USB SD­card reader, twelve color-coded input ID markers, quick start guide, and a USB stick supplied with DataView software and user manual.
®
A196-24-BK (included with Cat.
®
ACCESSORIES
Pole Mounting Kit
Catalog #2137.82 (optional)
Set of two, with hardware
cover closed
FEATURES
Simple-to-use, single-,
dual- (split-phase) and three-phase (Y, ) power & energy loggers
Designed to work in 1000V CAT III and 600V CAT IV environments
Supports 17 different network connections
Power measurements: VA, W and var
Energy measurements: VAh, Wh (source, load) and varh
(four quadrant indication)
Includes DataView
real-time display, analysis and report generation
USB, LAN, Ethernet, Wi-Fi and Bluetooth communication
(Class 1 wireless communication, up to 300 ft away)
Satisfies the monitoring requirements of NEC Code 220.87
PEL 105 can be configured from front panel, DataView
or the FREE Android™ application
Supports AC/DC Current Probe Model J93
for measuring current up to 5000A
Provides all the necessary functions for power and energy data logging
for 50Hz, 60Hz, 400Hz and DC distribution systems
Automatic recognition of the connected current sensors/probes
Powers directly from phase measurement
Pole mountable
®
software for configuring,
dc & 3500Aac
NEW!
®
control panel
CATALOG NO. DESCRIPTION
2137.51 Power & Energy Logger Model PEL 102 (no LCD w/3 MA193-10-BK Sensors)
2137.52 Power & Energy Logger Model PEL 103 (with LCD w/3 MA193-10-BK Sensors)
2137.57 Power & Energy Logger Model PEL 105 w/o Sensors (Waterproof IP67)
2137.59 Power & Energy Logger Model PEL 105 w/4 A916-24-BK (AmpFlex
2137.61 Power & Energy Logger Model PEL 102 (no LCD or Sensors)
2137.62 Power & Energy Logger Model PEL 103 (with LCD, no Sensors)
Accessories (Optional)
2137.77 Adapter–Power Adapter for use with Models PEL 102 & PEL 103
2137.82 Pole Mounting Kit
www.aemc.com
®
– Waterproof IP67)
Technical Assistance (800) 343-1391
Page 4
SPECIFICATIONS
MODELS PEL 102, PEL 103 & PEL 105
GENERAL
Sampling Frequency Data Storage Rate Demand Period Storage Rate Recorded Parameters
(Single- and Poly-Phase) Event Log Front Panel Indicator LEDs
Bluetooth active, recording in progress, phase connection reversal, overload, battery charging and SD Card status
Individual harmonics (from 1 through 50 per phase); Crest Factor (CF), Cos f / DPF
Tracks and records status changes and error messages along with recorded data
Storage Capacity INPUTS Voltage
Current
PEL 102/103 - 3 current input channels ; PEL 105 – 4 input channels via custom 4 pin jacks that accept AEMC® probes and sensors
PEL 102/103 – 3 input channels; PEL 105 – 4 input channels via 4mm safety banana jacks
ELECTRICAL
VOLTAGE MEASUREMENT PEL 102/103 RANGE PEL 105 RANGE RESOLUTION/ACCURACY*
Single-Phase RMS Voltages Phase-to-Phase RMS Voltages
Single-Phase RMS Voltages Phase-to-Phase RMS Voltages
PT Ratios
50/60Hz
400Hz
DC
Programmable from 50V to 650,000V
42.5 to 69Hz 42.5 to 69Hz ±0.1Hz 10 to 1000Vrms 10 to 1000Vrms 0.1V/±0.2% Rdg ± 0.2V 17 to 1700Vrms 17 to 1000Vrms 0.1 to 1V/±0.2% Rdg ± 0.4V
340 to 460Hz 340 to 460Hz
10 to 600Vrms 10 to 600Vrms 0.1V/±1% Rdg ± 1V
17 to 1200Vrms 17 to 1200Vrms 0.1 to 1V/±1% Rdg ± 1V
100 to 1000V 10 to 1000V 0.1V/±1% Rdg ± 3V (typical)
CURRENT MEASUREMENT A193*** A196*** Nominal range for current probes
supplied with kit. See chart on Pages 26 and 27 for other probes
200mA to 3000Arms 200mA to 10,000A
CT Ratios POWER MEASUREMENTS Active Power (P)* Reactive Power (Q)* Apparent Power (S)* Power Factor Tangent φ (active/reactive power ratio)
-2 to 2GW 0.001W ±0.5% Rdg ± 0.005% Pnom
-2 to 2Gvar 0.001var ±1% Rdg ± 0.01% Qnom 0 to 2GVA 0.001VA ±0.5% Rdg ± 0.005% Snom
-1 to 1 0.001 ± 0.05
-3.2 to 3.2 0.001 ± 0.02
ENERGY MEASUREMENTS Active Energy (EP) Reactive Energy (EQ) Apparent Energy (ES)
0 to 4 x 10 0 to 4 x 10 0 to 4 x 10
"
"
"
THD Individual Harmonics External Supply Power From Phase Measurement
PEL 102/103 requires optional Power Adapter Cat #2137.77; PEL 105 Internal up to 1000Vac
Back-Up Power Source/Charge Time Battery Life
MECHANICAL
Communication Ports Dimension/Weight
Case Mounting/Security
Kensington anti-theft system (PEL 102 & PEL 103) / Embedded magnets on back side, keyhole slot on back side
DISPLAY
Display Type for Model PEL 103 / PEL 105
2.63 x 2.16" (67 x 55mm), four line, monochrome, backlit LCD with adjustable brightness and contrast
ENVIRONMENTAL / SAFETY
Operating Temperature/Relative Humidity Storage Temperature Safety Rating/CE Rating
Complies with IEC 61010-1:Ed3, and IEC 61010-2-030: Ed1 for 1000V CAT III / 600V CAT IV (PEL 102 & PEL 103)
-4° to 122°F (-20° to 50°C) with batteries; -4° to 158°F (-20° to 70°C without batteries)
Index of Protection
* Maximum value is current probe dependent. ** Computers with Class II Bluetooth will restrict range to 40 ft Computers without Bluetooth will require a Class I or Class II Bluetooth radio adapter. *** Maximum current reduced by a factor of 2 for 400Hz fundamental frequency.
128 samples per cycle; 50/60Hz (16 samples/cycle 400Hz)
1 per second (200ms also available on PEL 105)
User selectable (1, 2, 3, 4, 5, 6, 10, 12, 15, 20, 30 and 60 minutes)
V, I, W, VA, var, PF, Tan, Wh, VAh, varh, THD (V and I),
SD card included / SD cards up to 32 GB formatted FAT32 are supported
Programmable from 1:1 to 25,000:1 (probe dependent)
1Wh ±0.5% Rdg
1varh ±2% Rdg
1Vah ±0.5% Rdg
± 655%
1 to 50 displayed in percentage; 1 to 7 at 400Hz
110V/250V (10%) @ 50/60Hz; 400Hz
Rechargeable 8.4V NiMH battery pack / Approximately 5 hours
30 minutes minimum, 60 minutes typical
USB 2.0, Ethernet (RJ45), Wireless Bluetooth Class 1 **/ Wi-Fi (PEL 105 only)
10.08 x 4.92 x 1.46" (256 x 125 x 37mm) / 2.20lbs (1kg) (PEL 102 & PEL 103);
9.8 x 7.8 x 2.6" (250 x 200 x 67mm) / 8.8lbs (4kg) (PEL 105)
Double insulated, rubber over-molded, polycarbonate UL94 V1 rated
50° to 122°F (10° to 50°C) / up to 85%
& 1000V CAT IV (PEL 105), Pollution Degree 2 / Yes
IP54 non operating (PEL 102 & PEL 103) / IP67 with cover closed (PEL 105)
0.01 to 0.1V
Technical Assistance (800) 343-1391
www.aemc.com
Page 5
POWER QUALITY ANALYZERS, METERS & LOGGERS
Optional Accessories
SENSOR TYPE CURRENT RANGE
MiniFlex® MA193-BK *
10" or 14" sensor
MR193-BK
Battery operated
SR193-BK
100mA to 3000Aac
1 to 1000Aac 1 to 1300Adc
1 to 1200Aac
ACCURACY
(TYPICAL)
±1%
±2.5% -0.80°
±0.3% 0.2°
TYPICAL
ERROR ON Φ AT 50/60HZ
MAX
CONDUCTOR
SIZE
2.75"
(70mm)
1.6"
(41mm)
2.05"
(52mm)
USED WITH
MODEL
PEL 102 PEL 103 PEL 105
8333 8336
8220
8230 PEL 102 PEL 103 PEL 105
8333
8336
8435
8220
8230 PEL 102 PEL 103 PEL 105
8333
8336
8435
LIMITED
RANGE IF
USED WITH
MODEL
8220 8230 8435
N/A 2140.28
N/A 2140.33
CATALOG NUMBER
2140.48
(10" sensor)
2140.50
(14" sensor)
AmpFlex® 193-BK *
100mA to 12,000Aac
24" or 36" sensor
AmpFlex® A196
10 to 10,000Aac
24" sensor IP67
Maximum current reduced by a factor of 2 for 400Hz fundamental frequency.
*
±1%
±2%
7.64"
(190mm)
or
11.46"
(290mm)
5.73"
(145mm)
PEL 102 PEL 103 PEL 105
8333
8336
PEL 105
8435
8220 8230 8435
N/A
2140.34
(24" sensor)
2140.35
(36" sensor)
2140.75
(24" sensor)
www.aemc.com
Technical Assistance (800) 343-1391
Page 6
POWER QUALITY ANALYZERS, METERS & LOGGERS
Optional Accessories
MN93-BK
MN193-BK
SL261 **
J93
SENSOR TYPE CURRENT RANGE
0.5 to 240Aac
100A
5A
100A
10A
200mA to 120Aac ±1% 0.75°
5mA to 6Aac ±1% 1.7°
5 to 100Aac/dc ±4% ±0.5°
50mA to 10Aac/dc ±3% ±1°
Battery operated
50 to 3500Aac 50 to 5000Adc
Battery operated
ACCURACY
(TYPICAL)
±1% 0.8°
±1% ±1°
TYPICAL
ERROR ON Φ AT 50/60HZ
MAX
CONDUCTOR
SIZE
0.78"
(20mm)
0.78"
(20mm)
0.46"
(11.8mm)
2.83"
(72mm)
Busbar:
5 x 1.69"
(127 x 43mm)
USED WITH
MODEL
PEL 102 PEL 103 PEL 105
8333 8336 8435
8220
8230 PEL 102 PEL 103 PEL 105
8333
8336
8435
8220
8230 PEL 102 PEL 103 PEL 105
8333
8336
8435
PEL 102 PEL 103 PEL 105
8333
8336
8435
LIMITED
RANGE IF
USED WITH
MODEL
8220 8230
N/A 2140.36
N/A 1201.51
N/A 2140.49
CATALOG NUMBER
2140.32
Note: Refer to the Power Meter’s product user manual for complete specifications.
All current sensors can be used with the Models 8435 and PEL 105. However, only the A196 flexible sensors are waterproof.
CATALOG NO. DESCRIPTION
1201.51 AC/DC Current Probe Model SL261 (BNC)
2140.37 Adapter – US 110V Power Adapter (eliminates need for batteries) 8220 & 8230 only
2140.28 AC/DC Current Probe Model MR193-BK
2140.32 AC Current Probe Model MN93-BK
2140.33 AC Current Probe Model SR193-BK
2140.34 AmpFlex
2140.35 AmpFlex
2140.36 AC Current Probe Model MN193-BK
2140.48 MiniFlex
2140.49 AC/DC Current Probe Model J93-BK
2140.50 MiniFlex
2140.75 AmpFlex
®
Sensor 24" Model 193-24-BK
®
Sensor 36" Model 193-36-BK
®
Sensor 10" Model MA193-10-BK
®
Sensor 14" Model MA193-14-BK
®
Sensor 24" (waterproof - IP67) Model A196-24-BK
Technical Assistance (800) 343-1391
** AC/DC Current Probe BNC Adapter for Model SL261 only Catalog #2140.40
www.aemc.com
Page 7
DATAVIEW®
Software for Power & Energy Loggers
Data Analysis and Reporting Software
Configure all functions of the Power & Energy Loggers
Display and analyze real-time data on your PC
Configure functions and parameters from your PC
Customize views, templates and reports to your exact needs
Create and store a complete library of configurations that
can be uploaded as needed
Zoom in and out and pan through sections of the graph to
analyze the data
Download, display and analyze recorded data
Display waveforms, trend graphs, harmonic spectrums,
text summaries, transients, event logs and stored alarms
Print reports using standard or custom templates you design
Free updates are available on our website www.aemc.com
Typical DataView® Functional, Digital & Graphical Displays
Display all harmonics from 1
Clear and easy setup of all functions from one tabbed dialog box.
Display real-time waveforms by phase, parameter or total.
form for voltage, current and power.
st
to 50th in bar graph
Display real-time Phasor diagrams. Includes unbalance for both voltage and current.
www.aemc.com
Display power and energy parameters – both instantaneous and total.
Display harmonics in a text table from harmonic 0 (DC) through the 50
Technical Assistance (800) 343-1391
th
.
Loading...