No part of this documentation may be reproduced in any form or by any means (including electronic storage and retrieval or translation into any other
language) without prior agreement and written consent from Chauvin Arnoux
Chauvin Arnoux
®
, Inc. d.b.a. AEMC® Instruments
®
, Inc., as governed by United States and International copyright laws.
15 Faraday Drive • Dover, NH 03820 USA
Tel: (800) 945-2362 or (603) 749-6434 • Fax: (603) 742-2346
®
This documentation is provided “as is,” without warranty of any kind, express, implied, or otherwise. Chauvin Arnoux
effort to ensure that this documentation is accurate; but does not warrant the accuracy or completeness of the text, graphics, or other information contained in this documentation. Chauvin Arnoux
®
, Inc. shall not be liable for any damages, special, indirect, incidental, or inconsequential; including (but
, Inc. has made every reasonable
not limited to) physical, emotional or monetary damages due to lost revenues or lost prots that may result from the use of this documentation, whether
or not the user of the documentation has been advised of the possibility of such damages.
Chauvin Arnoux
®
, Inc, AEMC®, and DataView® are registered trademarks of AEMC® Instruments.
2
Megohmmeter Models 6528/6529
Statement of Compliance
Chauvin Arnoux®, Inc. d.b.a. AEMC® Instruments certifies that this
instrument has been calibrated using standards and instruments
traceable to international standards.
We guarantee that at the time of shipping your instrument has met its
published specifications.
An NIST traceable certificate may be requested at the time of purchase,
or obtained by returning the instrument to our repair and calibration
facility, for a nominal charge.
The recommended calibration interval for this instrument is 12 months
and begins on the date of receipt by the customer. For recalibration,
please use our calibration services.
Serial #: ________________________________
Catalog #: 2126.54 / 2126.55
Model #: 6528 / 6529
Please fill in the appropriate date as indicated:
Date Received: _________________________________
Date Calibration Due: _______________________
Megohmmeter Models 6528/6529
3
MEASUREMENT CATEGORIES
770 V
Thank you for purchasing the TRMS Megohmmeter Model 6528 or 6529. For best results with your instrument:
■Read
■Respect
this user manual carefully
precautions for use
AC - Alternating currentDC - Direct Current
AC or DCDouble insulated
Shock HazardFuse
Battery
Instrument follows recycling
directives
Sorting for the recycling of electric
and electronic waste
Danger Hazard
Definition of Measurement Categories (CAT)
Instructions that must be read
and understood
Comply with US directives
Earth Ground
Voltage on the instrument should
not exceed 770V
RMS
■ CAT IV: Test and measurement circuits connected to the source of the building’s low-voltage network installation.
Example: Measur
■CAT III: Test and measurement circuits connected to parts of the building’s low voltage network installation.
Example: Measurement on distribution switchboards (including secondary meters), the circuit breakers, cabling including
cables, busbars, junction boxes, circuit breakers, power outlets in the xed installation and industrial instruments and
other equipment such as motors permanently connected to the xed installation.
■CAT II: Test and measurement circuits directly connected to points of use (power outlets and other similar points) on
the low voltage network.
Example: Measurement on circuits in network for household appliances, portable tools and other similar instruments.
ement on equipment installed upstream of the main fuse or building installation cut-off switch.
4
Megohmmeter Models 6528/6529
PRECAUTIONS FOR USE
Failure to comply with safety instructions can create a risk of electric shock, re, explosion and destruction of the instrument
or the installations. If the instrument is used other than as specied in this User Manual, the protection provided by the
instrument may be impaired.
■ The operator and/or the responsible authority must carefully read and clearly understand the various precautions to be
taken in use.
■ Before using the instrument, make sure it functions properly by measuring a known voltage and a known insulation
resistor, and check continuity by short circuiting both test leads.
■ Do not use the instrument in an explosive atmosphere or in the presence of inammable gas or smoke.
■ Do not use the instrument on networks with a rated voltage or category higher than those listed in this manual.
■ Respect the maximum rated voltages and currents between terminals and in relation to the earth.
■ Do not use the instrument if it seems damaged, incomplete, or incorr
■ Before each use, check the condition of the cable insulation, the instrument, and its accessories. All elements on which
the insulation is damaged (even partially) must be put out of service for r
■ Use cables and accessories for voltage accor
those of the instrument. If not, an accessory of a lower category reduces the category of the combined megohmmeter
plus the accessory to that of the accessory.
■ Respect the envir
■ Strictly comply with the fuse specications. Disconnect all cables befor
■ Do not modify
be performed by qualied, approved personnel.
■ Replace the battery as soon as the symbol appears on the display. Disconnect all cables before opening the
battery access cover.
■ Use personal protection equipment (PPE) when conditions require it.
■ Keep your hands and ngers away from the unused instrument terminals. When handling sensors or test probes, do
not place ngers beyond the physical nger guard.
onmental conditions of use.
the instrument, and, do not replace components using equivalent parts. Repairs and adjustments must
ding to IEC 61010-031 and measurement categories at least equal to
1.1 Receiving Your Shipment ...................................................................................................................................... 7
Ordering Information .............................................................................................................................................7
6.1 General Condition of Reference ..........................................................................................................................26
6.2 AC/DC Voltage Measurement ............................................................................................................................. 26
6.3 Ground Bond Continuity Measurement ...............................................................................................................26
Ground Bond Resistance Measurement ............................................................................................................. 26
Upon receiving your shipment, make sure that the contents are consistent with the packing list. Notify your distributor of any
missing items. If the equipment appears to be damaged, le a claim immediately with the carrier and notify your distributor
at once, giving a detailed description of any damage. Save the damaged packing container to substantiate your claim.
Both models include soft carrying case, set of (2) 5 ft color-coded silicone leads, (2) color-coded alligator clips, (2) colorcoded test probes (red/black), {Rated 1000V CAT IV, UL V2}, (6) 1.5V AA batteries and a user manual.
(1) of the following Megohmmeter Models:
Model 6528 Cat. #2126.54 or
Model 6529 Cat. #2126.55
Lead – set of (2), 5 ft silicone color-coded (red/
black) with 4mm straight/right angle banana plugs
(Rated 1000V CAT IV, UL)
Cat. #5000.94
Also includes 6 AA batteries and a user manual.
Megohmmeter Models 6528/6529
(1) Pouch – replacement for Models 1026, 6528, & 6529
(7 x 8.5 x 2”)
Cat. #2117.73
Probe – black test probe (Rated
1000V CAT IV, 15A, UL V2)
Cat. #5000.97
Probe – red test probe (Rated
1000V CAT IV, 15A, UL V2)
Cat. #5000.98
Clip – safety alligator – black
(Rated 1000V CAT IV, 15A, UL V2)
Cat. #5000.99
Clip – safety alligator – red (Rated
1000V CAT IV, 15A, UL V2)
Cat. #5100.00
7
1.3 Accessories
Continuity Probe for use with Megohmmeters and Installation Testers ........................... Cat. #2138.54
1.4 Replacement Parts
Pouch – replacement for Models 1026, 6528, & 6529 (7 x 8.5 x 2”) ................................................................Cat. #2117.73
Lead – set of (2), 5 ft silicone color-coded (red/black) with 4mm straight/right angle banana plugs
(Rated 1000V CAT IV, UL) .................................................................................................................................Cat. #5000.94
Fuse – set of 2, FF, 200mA, 1000V, 10kA, 6x32mm .........................................................................................Cat. #2971.04
Probe – black test probe (Rated 1000V CAT IV, 15A, UL V2) ........................................................................... Cat. #5000.97
Probe – red test probe (Rated 1000V CAT IV, 15A, UL V2) ..............................................................................Cat. #5000.98
Clip – safety alligator – black (Rated 1000V CAT IV, 15A, UL V2) ....................................................................Cat. #5000.99
Clip – safety alligator – red (Rated 1000V CAT IV, 15A, UL V2) ........................................................................ Cat. #5100.00
For accessories and replacement parts, visit our store at www.aemc.com.
2.OVERVIEW
2.1 Description
The Megohmmeter Models 6528 and 6529 are portable measuring instruments with digital displays. They are powered
by batteries. These instruments can check the safety of electrical insulation. For example, they can be used to test new
insulation before they are powered up, check existing insulation in a power-off condition, or troubleshoot potential faulty
insulation.
Features include:
Model 6528Model 6529
Insulation test voltages250, 500, and 1000V50, 100, 250, 500, and 1000V
Insulation Resistance
PI and DAR ratios calculation
Continuity measurement
Resistance measurement
Programmable alarms
–
8
Megohmmeter Models 6528/6529
2.2 Insulation Resistance Testing Principal of Operation
Insulation resistance measurement is based on Ohm’s Law. By applying a known DC voltage and then measuring the
current owing, the instrument can determine the value of the resistance. In principle, the value of the insulation resistance
is very high, but not innite. Therefore, by measuring the low current owing, the instrument indicates the insulation
resistance value, providing a result in kΩ, MΩ, or GΩ. This resistance characterizes the quality of the insulation and
provides a good indication of the risks of leakage currents.
2.3 Megohmmeter Models 6528 & 6529 — Front View
HOLD
HOLD
V
Ω
kΩ
REL
MEGOHMMETER
Ω
0
OFF
MODEL 6528
P
V%
mA
AC
+
V
DC
kΩ
GΩ
MΩ
NOISE
SET-UP
Ω
0
250 V
V
500
1000 V
MΩ
Backlit LCD
display unit
TEST button
Four
Function Keys
Test Modes
Six (Model 6528)
Eight (Model 6529)
HOLD
HOLD
V
Ω
kΩ
PI
DAR
REL
MEGOHMMETER
Ω
0
OFF
MODEL 6529
P
V%
mA
AC
+
V
DC
kΩ
GΩ
MΩ
NOISE
SET-UP
Ω
0
50 V
V
100
250 V
500 V
1000 V
M
600 V CAT IV
FUSED
Megohmmeter Models 6528/6529
Ω
700 V
max
Fused
Test Lead Terminals
600 V CAT IV
Ω
FUSED
700 V
max
9
2.4 TEST Button and Function Keys
0
Ω
0
SET-UP
2.4.1 TEST Button
Pressing the TEST button starts an insulation measurement. It also serves to conrm a programmed threshold.
• Performs battery voltage check (§3.2)
• In resistance measurement, it is used to enter/exit the DMR mode and record the reference
measurement (§3.7.1)
• Starts insulation resistance test (§3.8)
2.4.2 Function Keys
Key
HOLD
Ω
▲, ►
2.5 Instrument Display
1. Indicates the measurement is on hold.
2. Indicates the Dierential Mode Resistance
(DMR) or relative mode function is active in a
resistance measurement.
3. Indicates the resistance of the leads is
compensated in a continuity measurement.
4. Indicates auto-o is deactivated.
5. Indicates the remaining battery life.
6. Indicates that the measurement is within or
outside of the alarm threshold set point.
7. Indicates the presence of a hazardous voltage.
8. Indicates the alarm is active in the insulation
or DMR measurement mode.
9. Indicates the audible buzzer is activated.
10. Indicates a voltage interference in a continuity
or resistance measurement.
DESCRIPTION
Freezes/unfreezes the measurement on the display. To unfreeze the display press HOLD key again.
In SET-UP mode, the function of the key
Toggle between AC+DC and DC (§3.5)
Activate/deactivate audible continuity indicator (§3.6) and insulation alarm (§3.7)
(long press) Perform lead compensation for continuity checks (§3.6.1)
SET-UP (>2s press): congure settings (§2.7)
(short press): select type of insulation test ( , , PI, DAR (Model 6529)) (§3.8)
Navigate the conguration menu (§2.7)
is to cycle through parameters.
12345
HOLD
REL
Ω
P
0
V%
mA
AC
V
DC
kΩ
GΩ
MΩ
NOISE
Secondary Display
+
Primary Display
10
678910
Megohmmeter Models 6528/6529
2.6 Terminals
SET-UP
The instrument has two measurement terminals: +(positive) and COM (common).
Ω
FUSED
700 V
max
2.7 Configuration Mode
Conguration mode enables you to set several options for the instrument. To open conguration mode, turn the rotary
switch to OFF. Then press and hold down the SET-UP
symbol ConF appears on the LCD.
key for >2s while turning the rotary switch to any position. The
V
OFF
Ω
kΩ
250 V
500
1000 V
V
MΩ
>2s
6528 rotary switch shown
Pressing the ▲ key cycles through parameters, and the ► key selects the setting for the displayed option. Available
parameters:
Insulation test mode buzzer enabled/disabled (§2.8)
P
1st press on ▲
Permanent mode enabled/disabled (§2.9)
2nd press on ▲
Backlighting enabled/disabled (§2.10)
3rd press on ▲
Firmware version number (display only, §2.11)
4th press on ▲
Return to the rst screen.
Switch the instrument off by turning the rotary switch to OFF. All of your changes are applied the next time the
instrument is turned ON.
Megohmmeter Models 6528/6529
11
2.8 Alarm Buzzer
P
To enable/disable the audible Alarm buzzer in Insulation mode:
1. Turn the rotary switch to OFF.
2. Press and hold down the SET-UP key for >2s turning the rotary switch to any position. The LCD momentarily
displays ConF to indicate the insturment is in conguration mode.
3. The symbol appears on the display. If this is set to ON, press ► to select OFF. When you restart the instrument,
the Alarm Buzzer will be disabled.
2.9 Permanent Mode
To help prolong battery life, by default the instrument’s auto-off is activated and turns off the display if it is idle for 10
minutes. Press the TEST button to awaken the instrument.
To deactivate auto-off:
1. Turn the rotary switch to OFF.
2. Press and hold down the SET-UP key for >2s turning the rotary switch to any position. The LCD momentarily
displays ConF to indicate the insturment is in conguration mode.
3. Press ▲ until
P
appears. If this is set to OFF, press ► to select ON.
When you restart the instrument, the auto-off is deactivated and the
mode is active.
appears on the LCD indicating the Permanent
2.10 Backlight (BL)
To help prolong battery life, by default the instrument’s Backlighting mode is deactivated (or auto-off of backight is
activated) and after 2 minutes the backlighting is switched off.
If you want the backlighting to stay on at all times:
1. Turn the rotary switch to OFF.
2. Press and hold down the SET-UP
displays ConF to indicate the insturment is in conguration mode.
3. Press▲until
on, automatic switching off of the backlighting will be deactivated and the backlighting will stay on at all times once
the
key has been pushed while the instrument is actively on.
appears. If this is set to OFF, press ► to select ON. The next time the instrument is switched
key for >2s while turning the rotary switch to any position. The LCD momentarily
2.11 Firmware Version
To display the rmware version number currently running on the instrument:
1. Turn the rotary switch to OFF.
2. Press and hold down the SET-UP key for >2s while turning the rotary switch to any position. The LCD will display ConF.
3. Press ▲ until SOFT appears; the rmware version appears in the primary display.
12
Megohmmeter Models 6528/6529
2.12 Alarm Functions and Indicators
SET-UP
2.12.1 Alarm Functions
The alarm functions make it possible to rapidly conrm that the readings are OK, without looking at the display. You can
adjust the alarm function default settings by adjusting the thresholds.
Function
Insulation 50V*50kΩfrom 10kΩ to 399.9MΩ
Insulation 100V*100kΩfrom 20kΩ to 399.9MΩ
There are 7 alarm thresholds:
To program a threshold, set the switch to the desired function, long press the SET-UP
the beep sounds. The instrument displays the current threshold with the rst digit blinking.
Use the ▲ key to set the digit and the ► key to go to the next digit. When all 4 digits have been set, choose the unit number.
Conrm by pressing the TEST button.
V
Ω
kΩ
OFF
250 V
500
1000 V
MΩ
V
Insulation 250V250kΩfrom 50kΩ to 3.999GΩ
Insulation 500V500kΩfrom 100kΩ to 3.999GΩ
Insulation 1000V1000kΩfrom 200kΩ to 9.99GΩ
Continuity1Ω1Ω or 2Ω
Resistance DMR
* Model 6529 only
>2s
Default Threshold Programmable Threshold
5%
kΩ
HOLD
from 0.1 to 3999.9%
key for >2s and release it when
6528 rotary switch shown
To enable Alarm Mode, press the Alarm Buzzer
is below the alarm threshold, the meter emits a continous beep, the backlighting lights red and the × symbol is displayed.
Megohmmeter Models 6528/6529
key. The symbols and are displayed. If the measurement
13
3.OPERATION
3.1 Battery Installation
Place the batteries in the instrument as follows:
1. Use a Phillips screwdriver to remove the battery cover screw
on the back of the instrument (see illustration on right).
2. Install the batteries, respecting polarity. Requires 6 AA (LR6) batteries.
3. Replace the battery cover and tighten the screw.
3.2 Instrument Check
We recommend performing the following steps when using
the instrument for the rst time, or after a long period without use.
1. Start the instrument. Ensure all display segments are displayed.
2. Perform a Continuity check (§3.6). Ensure the LCD displays >42.0Ω with no input connected. Then short circuit both
terminals, the display should show 0Ω.
3. Turn the rotary switch to V, check a known voltage (for example a battery) and ensure that the measured voltage is correct.
4. Disconnect from the voltage source, and press the TEST button. The instrument’s battery voltage is displayed.
Ensure this is above 6.6V.
5. Turn the rotary switch to an insulation voltage setting (§3.7), and check a known resistor that is within the
measurement range for the test voltage selected.
6. When all the steps above produce expected results, you can start using the instrument.
3.3 Power Supply and Battery Life
The power supply is six 1.5V alkaline batteries (type AA or LR6). Nominal voltage is 6.6 to 9.6V. Below 6.6V the instrument
will not turn on.
■Battery life for insulation measurements: Average battery life for insulation measurement (according to
IEC61557-2 clause 6.7) is 2000 measurements with new batteries at room temperature (5s ON, 25s OFF).
■Battery life for voltage and resistance measurements: Average battery life for voltage and resistance with
new batteries at room temperature @ 400kΩ is > 300h.
■Battery life for continuity checking: Average battery life for continuity checking (according to IEC 61557-4
clause 6.6) is > 6000 measurements with new batteries at room temperature (5s ON, 25s OFF). For shorter
continuity tests (0.8s ON, 10s OFF) battery life is > 40,000 tests.
■Battery voltage: To check the battery voltage, press and hold the TEST button with the switch set to the V position.
14
Megohmmeter Models 6528/6529
3.4 Stand, Door or Magnetic Mount
For convenience, the instrument can be used in
different positions:
■With the stand - simply pull the back stand
down to free it from its slot, then fold it and
insert the end in the upper available slot.
■The stand can be placed on the top of a door
or other similar surface by positioning the stand
over the top edge as shown.
■The stand is equipped with magnets allowing
the instrument to be attached to a metal surface.
3.5 AC+DC or DC Voltage
The instrument measures AC+DC or DC voltage, to minimize risk when measuring an unknown voltage.
Measure AC+DC voltage rst. Be sure to respect the rated voltage and category.
If the measurement departs from the measurement range, the instrument displays OL.
1. Turn the rotary switch to the V position.
2. Toggle between AC+DC or DC voltage measurement by pressing
the
3. Connect the red test lead to the + terminal and the black test
lead to the COM terminal.
4. Measure the voltage by attaching the test leads to the
desired test points of the circuit.
Note: When the rotary switch is set to V, you can display the
instrument’s battery voltage by pressing and holding down the
TEST key with no voltage present on the test leads. Releasing
the key returns to voltage measurement mode.
key.
AC
+
V
DC
V
DC
Toggle between
AC + DC and DC
Megohmmeter Models 6528/6529
15
3.5.1 Voltage Error Indicators
■If the measurement is outside of the measurement range, the instrument displays OL.
3.6 Continuity
To avoid electrical shock and damage to the instrument when measuring resistance or continuity
in a circuit, ensure power to the circuit is turned OFF and all capacitors are discharged.
The measurement can be aected by (1) impedances of additional operating circuits connected in
parallel, or (2) transient currents.
In Continuity mode the instrument outputs a current above 200mA @ 2Ω between the + and COM terminals. The voltage
across the circuit under test is measured by the instrument; this voltage is then used to calculate resistance (R = V/I).
1. Turn the rotary switch to
2. The continuity threshold can be either 1Ω or 2Ω. To change this, press SET-UP
setting. Press ▲ to change this, then press TEST to exit SET-UP
3. Connect the red test lead to the
4. To compensate leads see §3.6.1.
5. Check continuity by connecting the leads to the desired point of the circuit. If resistance is under the threshold, the
buzzer will sound, and
backlight will light up red, the buzzer won’t sound, and the LCD displays ×. The LCD displays the measured continuity
on its primary display and test current on the secondary display (§2.5). To activate the audible alarm signal, press
the
lets you check that the continuity
key. The symbol is displayed
Secondary Display
Ω.
for >2s. This displays the threshold
mode.
+
terminal and the black test lead to the COM terminal.
will display on the LCD designating a short circuit. If the resistance is above the threshold, the
and a beep is emitted when the measurement is below the threshold. This
measurement is OK just by listening, without looking at the display.
REL
%
Ω
REL
%
Ω
16
Primary Display
If you change the leads without redoing the compensation, the reading can become negative.
Megohmmeter Models 6528/6529
3.6.1 Compensation of the Measurement Leads
Ω
0
Ω
0
Short the input terminals by touching them together. Then press the
key and hold it down for >2s. The LCD should
display 0Ω. If current is 0 mA when the terminals are shorted, verify the fuse is good.
NOTE: If the resistance of the leads is >5Ω, compensation is not possible.
To remove test lead compensation, press
the alarm
key.
for >2s with the test leads open. The buzzer can be disabled by pressing
3.6.2 Continuity Alarm Indicators
■ The alarm is always active in continuity measurement.
■ The instrument gives you a choice of two alarm thresholds: 1Ω or 2Ω.
■ If the measurement is below the threshold, the
■ If the measurement exceeds the threshold, the backlighting lights turn red and the
symbol is displayed.
symbol is displayed.
×
3.6.3 Continuity Error Indicators
■ If the measurement is beyond the measurement range, the instrument displays >42.00Ω.
■ If during the measurement the
measurement. Ensure the circuit under measurement is de-energized.
■When the measurement current is <200mA, the measurement is still correct but is no longer compliant with the
IEC standard continuity measurements.
■The instrument displays the AC+DC voltage. If it is >30V, the
voltage on the terminals is hazardous and the instrument emits a pulsed beep.
■If there is a voltage of more than 0.4V on the object to be tested, the instrument displays .
symbol appears on the LCD, transient currents are affecting the
symbol is displayed to warn the user that the
3.7 Resistance
To perform a resistance test, the instrument outputs a DC voltage and measures the current across the circuit under test.
The instrument then calculates the resistance.
To avoid electrical shock and damage to the instrument when measuring resistance or continuity in a circuit,
ensure power to the circuit is turned OFF and all capacitors are discharged.
3.7.1 Differential Mode Resistance (DRM) Mode
In the resistance measurement mode, there is a very useful feature called the Differential Resistance Mode (DRM). It can
be used as an incoming inspection tool, a radiant heat element qualifying tool or a quality check on production lines to
name a few.
This is a valuable feature to compare successive measurements to a previously
measured
threshold percentage
Start by setting the threshold in %
1. Turn the rotary switch to the kΩ position.
2. Press and hold the SET-UP
“reference” resistance. If the measurement falls outside a congured
, a visual and audible alarm will occur.
key for >2s. The threshold percentage
appears in the LCD secondary display area, with the rst digit blinking.
kΩ
V
OFF
Ω
250 V
500
1000 V
MΩ
V
3. You can adjust the percentage by pressing the ▲ to increment the blinking
digit and the ► to advance to the next digit.
Megohmmeter Models 6528/6529
6528 rotary switch shown
17
4. The threshold percentage can be set from 0.1% to 399.9%; default is 5%.
When nished setting the threshold percentage, press TEST to exit SET-UP mode.
5. Connect the red lead to the + terminal and the black lead to the COM terminal.
6. Connect the test leads to the device to be used as the reference resistance.
The resistance measurement will appear in the primary display area.
7. Press the TEST button to accept that as the reference value. The ΔRel
symbol will appear and the percentage
of the secondary display indicating the instrument is in DRM mode.
As you take subsequent measurements, the difference between the measurement
and the reference is displayed both in ohms (primary display (§2.5)) and as a
percentage (secondary display (§2.5)).
If the difference exceeds the threshold percentage, an alarm activates with a red
back lit display and an audible buzzer (if enabled) indicating the out of tolerance
condition.
difference will appear on the top line
DRM
Mode
REL
%
M
%
Ω
Ω
3.7.2 Resistance Mode Alarm Indicators
■Before starting an insulation measurement, pressing the key
activates the alarm. The alarm threshold is displayed, along with the
and symbols.
■If the measurement exceeds the threshold, the
■If the measurement is below the threshold, the instrument emits a
continuous beep, the backlight turns red and the
symbol is displayed.
symbol is displayed.
×
REL
%
3.7.3 Resistance Mode Error Indicators
■If there is a voltage of more than 0.4V on the object to be tested, the instrument displays .
■If the object to be tested is at a hazardous voltage, >30V, the symbol is displayed and the instrument emits a
pulsed beep.
■If the measurement is outside of the measurement range, the instrument displays >420.0kΩ.
3.8 Insulation Resistance Tests
To avoid electrical shock and damage to the instrument when measuring resistance or continuity
in a circuit, ensure power to the circuit is turned OFF and all capacitors are discharged.
In Insulation Resistance mode the instrument outputs a DC voltage. This voltage depends on the resistance to be
measured. The device measures the voltage and current present between the two terminals and determines the value of
R=V/I.
Ω
You can perform up to ve types of resistance tests, depending on instrument model. By default, the instrument is in
“unlocked” untimed test mode. In this mode, the test begins when you press the TEST button, and ends when you release
it. This mode is typically used to run short “spot” tests.
18
Megohmmeter Models 6528/6529
The other test types are selected via the SET-UP
key:
KEY
1st press
2nd press
3rd press
Exit
PI
KEY
4th press DAR
5th press TEST
DESCRIPTION
Lock the TEST button. In this mode, after you start the measurement, it continues without
requiring you to keep the TEST button pressed. The test will run until you stop it, or when 40
minutes have passed.
Activate timed test mode. You can set the test duration between 1 second and 39:59 minutes.
Exit timed text mode (Model 6528).
Enable the PI function (Model 6529). This is used to calculate the polarization index (the ratio of
the measurement at T2 [default 10 minutes] to the measurement at T1 [default 1 minute]).
DESCRIPTION
Enable the DAR function (Model 6529). This is used to calculate the dielectric absorption ratio (the
ratio of the measurement at T2 [default 1 minute] to the measurement at T1 [default 30 seconds]).
Exit timed test mode (Model 6529).
3.8.1 Untimed Test (Including Unlocked and Locked)
1. Turn the rotary switch to an insulation voltage setting, Choices are 250 V, 500 V, 1000 V (available on both the Models
6528 and 6529), and 50 V and 100 V (available on Model 6529 only). The test voltage to select depends on the operating
voltage of the device to be tested.
2. To run an unlocked test, skip this step and continue with step 3. To lock the TEST button, press the SET-UP key
once. The icon appears on the LCD.
3. To set an Alarm (§2.12).
4. Use the leads to connect the system to be tested to the instrument’s terminals. The test sample must be powered
down and discharged. When testing insulation, the typical connection is negative (black) lead to conductor and positive
(red) lead to ground or the outer insulation of the test sample.
5. Press the TEST button to start the test.
If you are running an unlocked test, hold down the TEST button until the displayed measurement is stable.
If you are running a
Note that if the instrument detects a voltage greater than 30V in the system under test, pressing the TEST button
has no effect because the test will be prohibited.
When the insulation measurement stabilizes, it appears on the primary display (§2.5). The secondary display
(§2.5) shows the test voltage.
V
Ω
kΩ
test, release the TEST button.
OFF
250 V
500
1000 V
MΩ
V
V
Megohmmeter Models 6528/6529
6528 rotary switch shown
MΩ
19
6. When performing insulation measurement, you can press the key to enable the alarm feature. If the
measurement falls below the alarm threshold, the alarm will activate. To set alarm threshold (§2.12).
7. At the end of the test, either release the TEST button (for unlocked tests) or press the TEST button a second time
(for
tests). The instrument stops generating the test voltage and discharges the device being tested. During this
process the discharging voltage value is displayed to indicate residual voltage. The measured value will remain on the
display until you make another measurement, press the HOLD key, or turn OFF the instrument. You can also start
another measurement immediately by a long press on the TEST button.
MEGOHMMETER
MODEL 6528
P
V
Press TEST to start
measurement
HOLD
V
Ω
kΩ
600 V CAT IV
OFF
Ω
FUSED
700 V
max
SET-UP
MΩ
Ω
0
250V
500
1000V
MΩ
Timer
V
Model 6528 shown
Do not disconnect the instrument while the
symbol is still displayed.
3.8.2 Timed Test (Including PI and DAR)
Both the Model 6528 and Model 6529 can perform timed tests that stop automatically after a dened duration. In addition,
the Model 6529 can perform polarization index (PI) and dielectric absorption ratio (DAR) tests.
Polarization index is a useful indicator of motor insulation health. It can identify the accumulation of contaminants as well
as any physical damage to the insulation itself. The test involves applying a positive charge to a motor’s conductors and
a negative charge to the motor frame therefore polarizing the insulation. This test essentially measures the change over
time. Healthy clean insulation will charge over time resulting in the decrease of induced current and an increase in insulation
resistance. Unhealthy insulation will result in a at to decreasing insulation resistance over time.
The PI test compares the measurement of insulation resistance at two points in time. The most prominent used time points
are at 1 and 10 minutes. Model 6529 defaults to these two test points labeled as T1 and T2. The instrument allows user
programming of both time points if necessary. Acceptable values can be programmed between 1 second and 30 minutes
for each test point. The time for T2 must be at least 1 second longer than T1.
The polarization index is calculated by dividing the insulation resistance value at point in time T2 (e.g., 10-min) by the
insulation resistance value taken at point in time T1 (e.g., 1-min).
20
Megohmmeter Models 6528/6529
IEEE-43 recommends that good insulation will exhibit an increase in resistance at the 10-minute point that should be at
least two times greater than the measurement taken at the one-minute point. Therefore, healthy insulation will exhibit a PI
ratio of 2 or higher. Newer insulation material can progress through the absorption phase very quickly and therefore display
a PI ratio of 1.
3.8.2.1 Conducting a Polarization Index (PI) Test (Model 6529)
1. Select the desired insulation resistance test voltage using the rotary switch.
2. Turn the rotary switch to an insulation voltage setting. Choices are 50V, 100V, 250V, 500V, and 1000V. The test voltage
to select depends on the voltage of the installation to be tested.
3. Press the SET-UP
key three times until you see PI on the secondary display (§2.5) along with the test run time.
4. Press and hold the TEST button until the test begins.
5. The secondary display will then indicate a countdown timer showing the remaining time to complete the test. The
primary display (§2.5) will show a real time value of the insulation resistance measurement in progress.
6. At the conclusion of the test the secondary display will show PI and the elapsed test time.
7. The primary display will show the PI ratio.
PI
PI
M
Ω
PI
3.8.2.2 Programming the Polarization Index T1 and T2 Test Time Points
1. Select the desired insulation resistance test voltage using the rotary switch.
2. Press the SET-UP
key three times until you see PI on the top line of secondary
display along with the test run time.
3. Press and hold the SET-UP
key >2s until the Alarm threshold is shown. Press
SET-UP key again until T1 appears and the leading time digit begins to blink. You can
adjust the time for T1 by using the ▲ to increment the blinking digit and ► to advance
to the next digit.
4. Once completed press the SET-UP
key again.
5. The time T2 time will now appear in the display with the rst digit blinking repeat the
▲ and ► arrow sequence to adjust the desired time for T2.
6. Press the test button TEST button to lock-in the new times for T1 and T2.
Megohmmeter Models 6528/6529
PI
PI
T1
T2
21
3.8.2.3 Conducting a Dielectric Absorption Ratio (DAR) Test
The dielectric absorption ratio test is typically accomplished by measuring the
insulation
insulation resistance
resistance taken at the 30 second time interval and comparing it to the
measurement at the 60 second time interval. The instrument
allows user programming of both time points if necessary.
As with the PI ratio, these two time points are labeled T1 and T2.
The DAR ratio is then calculated by dividing the T2 measurement (e.g., 60 seconds)
by the T1 measurement (e.g., 30 seconds).
Generally, ratios of 1.25 or less are considered questionable. Ratios between
1.25 and 1.6 are considered acceptable and ratios higher than 1.6 are considered
good.
1. Turn the rotary switch to an insulation voltage setting. Choices are 250 V, 500
V, 1000 V (available on both the Models 6528 and 6529) and 50 V and 100 V
(available on the Model 6529 only). The test voltage to select depends on the
voltage of the installation to be tested.
DAR
DAR
2. Press the SET-UP
key four times until you see DAR in the secondary
display along with the test run time.
3. Press and hold the TEST button until the test begins.
The top line of the secondary display will indicate a countdown timer showing the
remaining time to complete the test. The bottom primary display will show a real
time value of the insulation resistance measurement in progress.
At the conclusion of the test the secondary display will show DAR and the elapsed
test time.
The bottom primary display will show the DAR ratio.
3.8.2.4 Programming DAR the T1 and T2 test time points
The instrument allows for user programming both time points if necessary.
Acceptable
values can be programmed between 1 second and 30 minutes for
each test point. The time for T2 must at least 1 second longer than T1.
1. Select the desired insulation resistance test voltage using the rotary switch.
2. Press the SET-UP key four times until you see DAR on the top line of
display along with the test run time.
DAR
T1
3. Press and hold the SET-UP
Press SET-UP key again until T1 appears and the leading time digit
to blink. You can adjust the time for T1 by using the ▲ arrow to increment
key >2s until the Alarm threshold is shown.
begins
the
blinking digit and the ► arrow to advance to the next digit.
4. Once completed press the SET-UP
key again.
5. The time T2 will now appear in the secondary display with the rst digit blinking;
repeat the ▲ and ► arrow sequence to adjust the desired time for T2.
6. Press the TEST button to lock-in the new times for T1 and T2.
22
DAR
Megohmmeter Models 6528/6529
T2
3.8.2.5 Performing a Timed Test
SET-UP
SET-UP
SET-UP
SET-UP
1. Turn the rotary switch to an insulation voltage setting. Choices are 250V, 500V, 1000V (available on both the Models
6528 and 6529), and 50V and 100V (available on the Model 6529 only). The test voltage to select depends on the voltage
of the installation to be tested.
2. Press the SET-UP key two times to enter duration mode.
V
OFF
Ω
kΩ
6528 rotary switch shown
3. To set an Alarm, see §2.12
4. Press and hold the SET-UP
timer duration, with the leading time digit blinking. You can adjust the duration by using the ▲ arrow to increment the
blinking digit and the ► arrow to advance to the next digit.
>2s
250 V
500
V
1000 V
MΩ
key >2s until the Alarm threshold is shown. Press SET-UP key again to display
HOLD
5. The insulation measurement time of duration can be programmed from 00:01 to 39:59 (from 1 second to 40 minutes)
in the duration mode .
6. Press the TEST button to lock-in the new time duration for a timed test.
7. In the Timed Test
automatically at the end of the programmed duration.
■ The instrument displays the AC+DC voltage. If it is >30V, the symbol is displayed to warn the user that the
voltage on the terminals is hazardous, the instrument emits a pulsed beep and pressing the TEST button is
restricted from performing a test.
■If the measurement is outside of the measurement range, the instrument displays LO (if the insulation resistance
is too low to allow generation of the voltage) or >4200MΩ (for a test voltage of 50V,* 100V*, 250V or 500V)or
>11.00GΩ (for a test voltage of 1000V). (*Model 6529 only.)
■If the instrument fails to generate a voltage, check the fuse (§4.3).
Megohmmeter Models 6528/6529
mode, simply long-press the TEST button to start the timed measurement. Test will stop
23
4. MAINTENANCE
The instrument has no parts that can be replaced by personnel who are not trained and approved. Any non-approved
repair or other work, or replacement of a part by an “equivalent,” may severely compromise safety.
4.1 Cleaning
Periodically wipe the case with a damp cloth and mild detergent. Do not use abrasives or solvents. Dirt or moisture in the
terminals can affect the readings. Dry the instrument thoroughly after cleaning and before use.
4.2 200mA Fuse Test
To avoid electrical shock, remove the test leads and any input signals before replacing the fuse.
1. Turn the rotary switch to
2. Short circuit both terminals.
3. If the current measurement is 0, the fuse needs to be replaced.
Ω.
4.3 Battery and Fuse Replacement
To avoid false readings, which could lead to possible electric shock or personal injury, replace the batteries
with LR6 batteries as soon as the battery indicator appears. To prevent damage or injury, install
ONLY replacement fuses with the specied amperage, voltage, and interrupt ratings. Disconnect test leads
before removing the battery cover.
Replace the batteries and fuse as follows:
1. Using a Phillips screw driver, remove the battery cover screw on the back of
the instrument.
2. Remove the old batteries.
3. Install the batteries, respecting polarity.
4. Locate the fuse in its slot near the base of the instrument (indicated by the red
square in the illustration to the right).
5. Using a thin athead screwdriver or similar implement, carefully remove the
fuse.
6. Insert the replacement fuse (Fast Fuse 200mA 1000V 10kA size: 6.3 x 32mm),
ensuring both ends are rmly pressed into the fuse holding clips.
7. Replace battery door and tighten the screw.
24
Megohmmeter Models 6528/6529
5. GENERAL SPECIFICATIONS
I
S
T
E
USC
Maximum Voltage applied to terminals700VRMS or DC
Storage temperature-40 to 140°F (-40 to 60°C)
Operating temperature14 to 122°F (-10 to 50°C)
Temperature coefficient
Relative humidity10% to 90%
Operating Altitude0 to ~6500’ (0 to 2000m)
Electromagnetic compatibilityEN 61326-1, EN 61326-2-2
Insulation 4000PPM/°C, Ohm/Cont 2%/10°C+2D, V
0.3%/10°C+D
Safety
600V CAT IV, IEC 61010-2-034, IEC 61010-031,
L
Battery type6x AA, NEDA 15A, IEC LR6
Battery lifeSee §3.3
Size (H x W x L)8.54 x 3.54 x 2.44” (217 x 90 x 62mm)
Vibration1mm 25Hz as per IEC61557
Weight1.68 lbs (760g)
ConformityCE, IEC61557-2, IEC61557-4, IEC61557-10
IP40 with test leads connected: EN60529
Mechanical conformity
IP20 without test leads connected: EN60529
Drop test according to IEC61010-1
Shock 0.5 J (IK04): IEC 68-2-27
FuseF1: Fast Fuse 200mA 1000V 10kA size : 6.3 x 32
®
D
Megohmmeter Models 6528/6529
25
6. ELECTRICAL SPECIFICATIONS
6.1 General Condition of Reference
Influencing QuantityConditions of Reference
Ambient temperature73.4°F ± 5.4°F (23°C ± 3°C)
Relative humidity[45%; 75%]
Electric eld≤ 0.1V/m AC
Power supply (batteries voltage)8 to 9V
6.2 AC/DC Voltage Measurement
RangeResolution
AC+DC700V1V±(1.2%+1ct)
V
DC700V1V±(1%+1ct)
V
Input impedance: 25MΩ
Common Mode Rejection Ratio: > 60dB
Overload protection: 700VRMS or DC
Frequency:30 to 440Hz
6.3 Ground Bond Continuity Measurement
RangeResolution
Continuity40Ω0.01Ω±(1.2%+3cts)
Resistance of test leads: ≤ 0.01Ω (compensated)
Overload fuse protection by fuse: 700VRMS
Open circuit voltage: >6V and <9V
Short circuit current:
0.02-2.00Ω: ≥ 200mA
2.01-39.99Ω: between 100-200mA
Uncertainty
±(% of reading +Digits)
Uncertainty
±(% of reading +Digits)
6.4 Ground Bond Resistance Measurement
RangeResolution
400Ω0.1Ω
kΩ
Resistance of test leads: ≤ 0.01Ω (compensated)
Overload fuse protection by fuse: 700VRMS
Open circuit voltage: 4.5V
26
4kΩ1Ω
40kΩ10Ω
400kΩ100Ω
Uncertainty
±(% of reading +Digits)
±(1.2%+3cts)
Megohmmeter Models 6528/6529
6.5 Insulation Resistance Specification
Measurement range:0.01MΩ to 10GΩ
Test voltages:
Model 6529: 50, 100, 250, 500, 1000V
Model 6528: 250, 500, 1000V
Test voltage accuracy:+25%, 0%
Nominal current:1mA
Auto discharge:Discharge time 2s for C=2µF
Live circuit detection:Inhibit test if terminal voltage >30V prior to initialization of test
Maximum capacitive load:2µF at nominal voltage and nominal current.
Output VoltageRangeDisplay RangeResolutionTest current
4MΩ0.01 to 4.000MΩ0.001MΩ
50V*
(0% to +25%)
40MΩ3.60 to 39.99MΩ0.01MΩ
1mA @ 50kΩ
400MΩ36.0 to 399.9MΩ0.1MΩ
0.020 to
4.000MΩ
0.001MΩ
1mA @ 100kΩ
100V*
(0% to +25%)
4MΩ
40MΩ3.60 to 39.99MΩ0.01MΩ
400MΩ36.0 to 399.9MΩ0.1MΩ
0.050 to
3.999MΩ
0.001MΩ
1mA @ 250kΩ
250V
(0% to +25%)
4MΩ
40MΩ3.60 to 39.99MΩ0.01MΩ
400MΩ36.0 to 399.9MΩ0.1MΩ
4GΩ360 to 3,999MΩ1MΩ±(4%+10cts)
0.100 to
3.999MΩ
0.001MΩ
1mA @ 500kΩ
500V
(0% to +25%)
4MΩ
40MΩ3.60 to 39.99MΩ0.01MΩ
400MΩ36.0 to 399.9MΩ0.1MΩ
4GΩ360 to 3,999MΩ1MΩ±(4%+10cts)
40MΩ0.20 to 39.99MΩ0.01MΩ
1000V
(0% to +25%)
400MΩ36.0 to 399.9MΩ0.1MΩ
4GΩ360 to 3,999MΩ1MΩ±(4%+5cts)
1mA @ 1MΩ
10GΩ3.60 to 9.99GΩ.01GΩ±(10%+10cts)
* Model 6529 only
Uncertainty
±(% of reading
+ Digits)
±(3%+10cts)
±(1.5%+10cts)
±(1.5%+10cts)
±(1.5%+10cts)
Capacitance in parallel: <1nF
External voltage in series: 0.0V
Common mode voltage: 0.0V
Megohmmeter Models 6528/6529
27
7. REPAIR AND CALIBRATION
To ensure that your instrument meets factory specications, we recommend that it be scheduled back to our factory
Service Center at one-year intervals for recalibration, or as required by other standards or internal procedures. For
instrument repair and calibration:
You must contact our Service Center for a Customer Service Authorization Number (CSA#). This will ensure that when
your instrument arrives, it will be tracked and processed promptly. Please write the CSA# on the outside of the shipping
container. If the instrument is returned for calibration, we need to know if you want a standard calibration, or a calibration
traceable to N.I.S.T. (Includes calibration certicate plus recorded calibration data).
Ship To: Chauvin Arnoux
15 Faraday Drive • Dover, NH 03820 USA
Phone: (800) 945-2362 (Ext. 360)
(603) 749-6434 (Ext. 360)
Fax: (603) 742-2346 or (603) 749-6309
E-mail: repair@aemc.com (Or contact your authorized distributor)
Costs for repair, standard calibration, and calibration traceable to N.I.S.T. are available.
NOTE: You must obtain a CSA# before returning any instrument.
®
, Inc. d.b.a. AEMC® Instruments
8. TECHNICAL AND SALES ASSISTANCE
If you are experiencing any technical problems, or require any assistance with the proper operation or application of your
instrument, please call, fax, or e-mail our technical support team:
Chauvin Arnoux
Phone: (800) 343-1391
(508) 698-2115
Fax: (508) 698-2118
E-mail: techsupport@aemc.com
®
, Inc. d.b.a. AEMC® Instruments
28
Megohmmeter Models 6528/6529
9. LIMITED WARRANTY
The instrument is warranted to the owner for a period of two years from the date of original purchase against defects in
manufacturing. This limited warranty is given by AEMC® Instruments, not by the distributor from whom it was purchased.
This warranty is void if the instrument has been tampered with, abused, or if the defect is related to service not performed
by AEMC® Instruments.
The warranty does not apply in the following cases:
■Inappr
■Modications made to the equipment without the explicit permission of the manufactur
opriate use of the equipment or use with incompatible equipment
er’s technical staff
■Work done on the device by a person not approved by the manufacturer
■Adaptation to a particular application not anticipated in the denition of the equipment or not indicated in this user’s manual
■Damage caused by shocks, falls, or oods
Full warranty coverage and product registration is available on our website at: www.aemc.com/warranty.html
Please print the online Warranty Coverage Information for your records.
®
What AEMC
Instruments will do:
If a malfunction occurs within the warranty period, you may return the instrument to us for repair, provided we have your
warranty registration information on le or a proof of purchase. AEMC
®
Instruments will, at its option, repair or replace the
faulty material.
Megohmmeter Models 6528/6529
29
10/20
99-MAN 100491 v1
Test Equipment Depot - 800.517.8431 - 5 Commonwealth Ave, MA 01801
TestEquipmentDepot.com
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.