ADMAG AXW User Manual

User’s Manual
ADMAG TI Series AXW Magnetic Flowmeter [Size: 500 to 1800 mm (20 to 72 in.)] Installation Manual [Style: S1, S2]
Integral Flowmeter
(AXW###G)
Remote Transmitter
(AXG1A)
Remote Transmitter
(AXFA11G)
This manual outlines the basic guidelines for installation and wiring procedures. For the items which are not covered in this manual, see the user’s manuals and the general speci ca-
tions as listed in Table 1.1.
Remote Sensor
(AXW###G, AXW###W)
Remote Transmitter
(AXW4A)
IM 01E25A01-01EN

Contents

1. Introduction
1.1 For Safe Use of Product ................................................3
1.2 Warranty .........................................................................6
1.3 Combination for Remote Sensor and Remote
Transmitter .....................................................................6
2. Receiving and Storage
2.1 Model and Speci cations Check ...................................7
2.2 Storage Precautions ......................................................8
3.1 Piping Design Precautions ............................................9
3.2 Handling Precautions .................................................. 12
3.2.1 General Precautions ...................................... 12
3.2.2 Flowmeter Piping ........................................... 13
3.3 Integral Flowmeter and Remote Sensor Installation .. 13
3.4 Remote Transmitter Installation .................................. 19
3.4.1 Installation Location ....................................... 19
3.4.2 Mounting of AXW4A Transmitter ................... 19
3.4.3 Mounting of AXG1A Transmitter .................... 19
3.4.4 Mounting of AXFA11 Transmitter ................... 20
3.5 Changing Direction of Electrical Connection .............. 20
3.5.1 Integral Flowmeter ......................................... 20
3.5.2 Remote Sensor (Style: S1) ............................ 21
3.5.3 Remote Sensor (Style: S2) ............................ 22
3.6 Changing Direction of Display Unit ............................. 23
4. Wiring
4.1 Wiring Precautions ...................................................... 24
4.2 Cables ......................................................................... 25
4.3 Electrical Connections ................................................ 26
4.4 Connecting to External Instruments of Integral
Flowmeter and Remote Transmitter ........................... 30
4.5 Connecting to Remote Sensor and Remote
Transmitter (Sensor Side) ........................................... 36
4.6 Input and Output ......................................................... 39
5. Basic Operating Procedures
5.1 Operation by Display unit ............................................ 42
5.2 Display and Basic Con guration ................................. 42
5.3 Display Mode and Setting Mode ................................. 45
5.4 Parameter Setting from Display Panel ....................... 46
5.5 microSD Card Setting ................................................. 48
5.6 BRAIN Con guration Tool ........................................... 48
5.7 HART Con guration Tool ............................................ 49
5.8 Modbus Con guration Tool ......................................... 50
5.9 F
OUNDATION
 eldbus Con guration Tool ..................... 51
6. Operation
6.1 Pre-operation Zero Adjustment ................................... 52
6.2 Zero Adjustment from Display Unit ............................. 52
6.3 Hardware Switch Setting ............................................ 53
7. Errors and Countermeasures (Display unit)
1
2
3
4
5
6
7
IM 01E25A01-01EN
6th Edition
<1. Introduction>
NOTE
IMPORTANT
NOTE

1. Introduction

1
This manual provides the basic guidelines for installation,
wiring procedures and basic operation of ADMAG TI
(Total Insight) Series AXW magnetic owmeters (size: 500 to 1800 mm (20 to 72 in.)) with BRAIN, HART and
Modbus protocol.
For the items which are not covered in this manual, read the applicable user’s manuals and general specications as listed in Table 1.1. These documents can be downloaded from the website of YOKOGAWA. To ensure correct use of the instrument, read these manuals thoroughly and fully understand how to operate the instrument before operating it. For method of checking the model and specications, read Chapter 2 and general specications as listed in Table 1.1. Website address: http://www.yokogawa.com/d/doc/ These manuals can be downloaded from the website of YOKOGAWA or purchased from the YOKOGAWA representatives.
Table 1.1 Manual and General Specications List
Model Document Title Document No.
ADMAG TI Series
AXW###G AXW###W AXW4A AXG1A AX01C
AXG###, AXW###, AXG4A, AXW4A, AXG1A Magnetic Flowmeter Read Me First
ADMAG TI Series AXG/AXW Magnetic Flowmeter Safety Manual
ADMAG TI Series AXW Magnetic Flowmeter [Size: 500 to 1800 mm (20 to 72 in.)] Installation Manual
ADMAG TI Series AXW Magnetic Flowmeter [Size: 25 to 1800 mm (1 to 72 in.)] Maintenance Manual
ADMAG TI Series AXG, AXW Magnetic Flowmeter BRAIN Communication Type
ADMAG TI Series AXG, AXW Magnetic Flowmeter HART Communication Type
ADMAG TI Series AXG, AXW Magnetic Flowmeter Modbus Communication Type
ADMAG TI Series
AXG/AXW Magnetic Flowmeter
OUNDATION Fieldbus
F Communication Type
ADMAG TI Series AXG1A Magnetic Flowmeter BRAIN Communication Type
ADMAG TI Series AXG1A Magnetic Flowmeter HART Communication Type
ADMAG TI Series AXW Magnetic Flowmeter
[Size: 500 to 1800 mm (20 to 72 in.)] General Specications
ADMAG TI Series AXG1A Magnetic Flowmeter Remote Transmitter
General Specications
IM 01E21A21-01Z1
IM 01E21A21-02EN
IM 01E25A01-01EN (this manual)
IM 01E24A01-02EN
IM 01E21A02-01EN
IM 01E21A02-02EN
IM 01E21A02-05EN
IM01E21A02-03EN
IM 01E22C02-01EN
IM 01E22C02-02EN
GS 01E25D11-01EN
GS 01E22C01-01EN
Model Document Title Document No.
AXF Series
AXFA11G AXFC
Magnetic Flowmeter Read Me First
AXFA11G Remote Converter [Hardware Edition/Software Edition]
AXFA11G Remote Converter General Specications
IM 01E20A21-01Z1
IM 01E20C01-01E
GS 01E20C01-01E
When describing the model name like AXW###G in this manual, “###” means any of the following.
500, 600, 700, 800, 900, 10L, 11L, 12L, 13L, 14L, 15L, 16L, 18L
Precautions Related to the Protection,
Safety, and Alteration of the Instrument
The following safety symbol marks are used in this
manual and instrument.
WARNING
A WARNING sign denotes a hazard. It calls attention to procedure, practice, condition or the like, which, if not correctly performed or adhered to, could result in injury or death of personnel.
CAUTION
A CAUTION sign denotes a hazard. It calls attention to procedure, practice, condition or the like, which, if not correctly performed or adhered to, could result in damage to or destruction of part or all of the product.
An IMPORTANT sign denotes that attention is required to avoid damage to the instrument or system failure.
A NOTE sign denotes information necessary for
essential understanding of operation and features.
1
Introduction
All Rights Reserved, Copyright © 2017, Yokogawa Electric Corporation
IM 01E25A01-01EN6th Edition: Nov. 2020 (KP)
<1. Introduction>
2
The following symbols are used in the Instrument and the manual to indicate the accompanying safety precautions:
Protective grounding terminal
Functional grounding terminal (This terminal should not be used as a protective grounding terminal.)
Alternating current
Direct current
Caution
This symbol indicates that the operator must refer to an explanation in the user’s manual in order to avoid the risk of injury or death of personnel or damage to the instrument.
• For the protection and safe use of the instrument and the system in which this instrument is incorporated, be sure to follow the instructions and precautions on safety that is stated in this manual whenever you handle the instrument. Take special note that if you handle the instrument in a manner that violated these instructions, the protection functionality of the instrument may be damaged or impaired. In such cases, YOKOGAWA does not guarantee the quality,
performance, function, and safety of instrument.
• When installing protection and/or safety as lighting protection devices and equipment for the instrument
and control system or designing or installing separate
protection and/or safety circuits for fool-proof design and fail-safe design of the processes and lines that use the instrument and the control system, the user should implement these using additional devices and equipment.
• Should use the parts specied by YOKOGAWA when replacing. Please contact YOKOGAWA’s service o󰀩ce for fuse replacement.
• This instrument is not designed or manufactured to be used in critical applications that directly a󰀨ect or threaten human lives. Such applications include nuclear power equipment, devices using radioactivity, railway facilities, aviation equipment, air navigation facilities, aviation facilities, and medical equipment. If so used, it is the user’s responsibility to include in the system additional equipment and devices that ensure
personnel safety.
• Do not modify this instrument.
• YOKOGAWA will not be liable for malfunctions or damage resulting from any modication made to this instrument by the customer.
• The instrument should be disposed of in accordance with local and national legislation/regulations.
Regarding This User’s Manual
• This manual should be provided to the end user.
• The contents of this manual are subject to change without prior notice.
• All rights reserved. No part of this manual may be reproduced in any form without YOKOGAWA’s written
permission.
• YOKOGAWA makes no warranty of any kind with regard to this manual, including, but not limited to, implied warranty of merchantability and tness for a
particular purpose.
• If any question arises or errors are found, or if any information is missing from this manual, inform the nearest YOKOGAWA sales o󰀩ce.
• The specications covered by this manual are limited to those for the standard type under the specied model number break-down and do not cover custom-
made instruments.
• Note that changes in the specications, construction, or component parts of the instrument may not immediately be reected in this manual at the time of change, provided that postponement of revisions will not cause di󰀩culty to the user from a functional or
performance standpoint.
• This manual is intended for the following personnel;
Engineers responsible for installation and wiring of the
instrument.
Personnel responsible for normal daily operation of
the instrument.
• To ensure correct use, read this manual and the applicable manuals as listed in Table 1.1 thoroughly before starting operation. Read the general specications as listed in Table 1.1 for its specication.
Trademarks:
• HART is a registered trademark of FieldComm Group.
• Modbus is a registered trademark of AEG Schneider.
• F
OUNDATION
Group.
• All the brands or names of Yokogawa Electric’s products used in this manual are either trademarks or registered trademarks of Yokogawa Electric
Corporation.
• All other company and product names mentioned in this manual are trade names, trademarks or registered trademarks of their respective companies.
• In this manual, trademarks or registered trademarks are not marked with ™ or
is a registered trademark of FieldComm
®
.
IM 01E25A01-01EN
<1. Introduction>
IMPORTANT
3

1.1 For Safe Use of Product

For the protection and safe use of the instrument and the system in which this instrument is incorporated, be sure to follow the instructions and precautions on safety that is stated in this manual whenever you handle the instrument. Take special note that if you handle the instrument in a manner that violated these instructions, the protection functionality of the instrument may be damaged or impaired. In such cases, YOKOGAWA shall not be liable for any indirect or consequential loss incurred by either using or not being able to use the
Instrument.
(1) General
• This instrument conforms to EN61326-1, EN61326­2-3, EN61326-2-5, EN61000-3-2, and EN61000-3-3 (EMC standard).
• This instrument is an EN61326-1 (EMC standard),
Class A (for use in commercial, industrial, or business
environments).
• This instrument (General-purpose type) is complied with IP66 and IP67 in the IP Protection Grade.
YOKOGAWA assumes no liability for the customer’s
failure to comply with these requirements.
• This instrument (Submersible type) is complied with IP68 in the IP Protection Grade.
YOKOGAWA assumes no liability for the customer’s
failure to comply with these requirements.
• This instrument is designed for indoor and outdoor
use.
• This instrument is Micro Pollution Degree: 2.
• This instrument is Macro Pollution Degree: 4.
• This instrument is designed for altitude at installation Site Max. 2000 m above sea level.
• This instrument is installation Category (Overvoltage
category): II.
CAUTION
This instrument is a Class A instrument in the EN61326-1(EMC standard). Operation of this
instrument in a residential area may cause radio
interference, in which case the user is required to take appropriate measures to correct the interference.
The minimum ambient temperature is limited by the minimum uid temperature of the sensor (the lining). For more information, read the applicable general specications as listed in Table 1.1. The owmeter may be used in an ambient humidity where the relative humidity ranges from 0 to 100%. However, avoid long-term continuous operation at relative humidity above 95%.
WARNING
• Purpose of use
This instrument is the Magnetic Flowmeter for use of measuring the liquid ow. Do not use this instrument for other purposes.
WARNING
• Installation, wiring and maintenance of the magnetic owmeter must be performed by expert engineer or skilled personnel. No operator shall be permitted
to perform procedures relating to installation, wiring and maintenance.
• Wiring work should be done adequate wire, sleeve crimp and torque force. Use terminal with insulating cover for the power supply wiring and protective grounding wiring. Do not pull the wires too much strongly in order to prevent electric shocks caused by their damage.
• Do not open the cover in wet weather or humid environment. When the cover is open, stated
enclosure protection is not applicable.
• Ensure that the power supply is o󰀨 in order to prevent electric shocks.
• When opening the cover, wait for more than 20 minutes after turning o󰀨 the power. Only expert engineer or skilled personnel are permitted to open the cover.
• When opening and closing the transmitter cover, be sure to handle the transmitter cover carefully so that there are no damage and foreign matter adhesion at its threads and O-ring.
• This instrument employs the parts which are a󰀨ected by a function damage caused by static electricity. Thus, you should do the antistatic work
using an anti-static wrist band for it and be careful
to avoid touching each electrical parts and circuitry
directly.
• When connecting the wiring, check that the supply voltage is within the range of the voltage specied for this instrument before connecting the power cable. In addition, check that no voltage is applied to the power cable before connecting the wiring.
• To prevent electric shocks, ensure the electrical wiring cover is completely attached after the wiring work.
• To prevent electric shocks, do not impress over rated voltage to each input/output terminals.
• If there is any unused electrical connection, use the blanking plug to cover which comes with this instrument or which is supplied by YOKOGAWA. The blanking plug should be fastened into the unused electrical connection without any mistake. If
not, stated enclosure protection is not applicable.
1
Introduction
IM 01E25A01-01EN
<1. Introduction>
4
WARNING
• For AXG1A, impact resistance rating of glass on the display cover is IK06, metal housing is IK08. In the test method, the steel ball is dropped from a height of 200 mm after pre-cooling the housing to
-40 degree C. (Impact on horizontal surface)
• The magnetic owmeter is a heavy instrument.
Be careful that no damage is caused personnel
through accidentally dropping it, or by exerting excessive force on the magnetic owmeter. When moving the magnetic owmeter, always use a trolley and have at least two people carry it.
• Do not apply excessive weight, for example, a person stepping on the magnetic owmeter.
• The magnetic owmeter must be installed within the specication conditions.
Connect the Protective Grounding Terminal
Ensure to connect the protective grounding to
prevent electric shock before turning on the power.
Do Not Impair the Protective Grounding
Never cut o󰀨 the internal or external protective
grounding wire or disconnect the wiring of the protective grounding terminal. Doing so invalidates the protective functions of the instrument and poses a potential shock hazard.
Do Not Operate with Defective Protective
Grounding
Do not operate the instrument if the protective
grounding might be defective. Also, ensure to check them before operation.
Do Not Operate in an Explosive Atmosphere
Do not operate the instrument in the presence
of ammable gas, vapors, or combustible dust. Operation in such an environment constitutes a safety hazard. Prolonged use in a highly dense corrosive gas (H
malfunction.
Ground the Instrument before Making External Connections
Connect the protective grounding before connecting
to the item under measurement or control unit.
Damage to the Protection
Operating the instrument in a manner neither
described in this manual nor the manuals as listed in Table 1.1 may damage the instrument’s protection.
• The owmeter should be installed away from electrical motors, transformers, and other power sources in order to avoid interference with
measurement.
S, SOx, etc.) will cause a
2
WARNING
• Install an external switch or circuit breaker as a means to turn the power o󰀨 (capacitance: 15A, conforming to IEC60947-1 and IEC60947-3).
Locate this switch either near the instrument or in other
places facilitating easy operation. A󰀩x a "Power O󰀨 Equipment" label to this external switch or circuit breaker.
All procedures relating to installation must comply
with the electrical code of the country where it is used.
(3) Wiring
WARNING
• In cases where the ambient temperature exceeds 50°C, use external heat resistant wiring with a maximum allowable temperature of 70°C or above.
• When wiring the conduits, pass the conduit through the wiring connection port, and utilize the waterproof gland to prevent water from owing in. Install a drain valve at the low end of the vertical pipe, and open the valve regularly.
Do not connect cables outdoors in wet weather in order to prevent damage from condensation and to protect the insulation, e.g. inside the terminal box of the owmeter.
• The transmitter case should be removed by YOKOGAWA’s qualied personnel only. Opening the transmitter case is dangerous, because some areas inside the instrument have high voltages.
• The protective grounding must be connected securely at the terminal with the
danger to personnel.
mark to avoid
(4) Operation
WARNING
Be sure to enable the write protect function to prevent the overwriting of parameters after nishing parameter
setting.
In rare cases, the infra-red switches may respond unexpectedly in such conditions as sticking ball of water or extraneous substances on the surface of display panel glass according to the principle of infra-red switch operation. Its probability rises in such cases as sticking rain water by storm or other similar situation and washing up work near owmeter installation place. Blinking light from a ashlight etc. to the infra-red switches may result in the malfunction. Read Section 6.3 for the hardware write protect function, and the user's manual of applicable communication type as listed in Table 1.1 for the
software write protect function.
IM 01E25A01-01EN
<1. Introduction>
IMPORTANT
5
(5) Maintenance
WARNING
• When maintaining the instrument, read the
maintenance manual as listed in Table 1.1. Do not
perform the maintenance that is not described in the manual. If necessary, contact YOKOGAWA.
• When the magnetic owmeter is processing hot uids, the instrument itself may become extremely hot. Take su󰀩cient care not to get burnt.
• Where the uid being processed is a toxic substance, avoid contact with the uid and avoid inhaling any residual gas, even after the instrument has been taken o󰀨 the piping line for maintenance and so forth.
• If dirt, dust or other substances surfaces on the glass of display cover, wipe them clean with a soft dry cloth.
• Maintenance of this owmeter should be implemented in a maintenance service shop where the necessity tools and environment condition are provided.
The necessity of this environmental condition is that
ambient temperature is 5 to 40°C (the maximum relative humidity is 80 % for temperature 5 to 31°C, and decreasing linearly to 50 % relative humidity at
40°C).
(6) Modication
• Do not modify this instrument.
• YOKOGAWA will not be liable for malfunctions or damage resulting from any modication made to this instrument by the customer.
(7) Product Disposal
The instrument should be disposed of in accordance
with local and national legislation/regulations.
(8) Power Supply
Ensure that the source voltage matches the voltage of the power supply before turning on the power.
Power Supply Code 1:
• AC Type: Rated Power Supply: 100 to 240 V AC, 50/60 Hz
• DC Type:
Rated Power Supply: 100 to 120 V DC Power Supply Code 2:
• AC Type: Rated Power Supply: 24 V AC, 50/60 Hz
• DC Type:
Rated Power Supply: 24 V DC Power Consumption:
Integral Type: 13 W
Remote Type (with AXG4A): 13 W Remote Type (with AXG1A): 32 W
Note: Power Consumption is independent of communication
and I/O specication.
Note: For AXFA11, read the applicable user’s manual as
listed in Table 1.1.
(9) microSD Card
• Do not store or use the microSD card in places with static electricity, near electrically charged objects, or where electrical noise is present. Doing so can result in shock or damage.
• Do not disassemble or modify the microSD card.
• Do not physically shock, bend, or pinch the
microSD card.
• During reading/writing of data, do not turn o󰀨 the power, apply vibration or shock, or pull out the
card. Data can corrupt or be permanently lost.
• Use only micro SD cards sold by YOKOGAWA. Operation cannot be guaranteed when other cards
are used.
• When inserting the microSD card into the instrument, make sure to orient the microSD card
correctly (face up or down) and insert it securely. If
not inserted correctly, the microSD card will not be recognized by the instrument.
• Do not touch the microSD card with wet hands.
• Do not use the microSD card if it is dusty or dirty.
• The microSD card comes formatted. If you want to format the microSD card, use the instrument's
Format function.
• YOKOGAWA provides no warranty for damage to, or loss of data recorded on the microSD card, regardless of the cause of such damage or loss.
We recommend making backup copies of your
data.
1
Introduction
IM 01E25A01-01EN
<1. Introduction>
IMPORTANT
NOTE
6

1.2 Warranty

• The warranty shall cover the period noted on the quotation presented to the purchaser at the time of purchase. Problems occurred during the warranty period shall basically be repaired free of charge.
• In case of problems, the customer should contact the YOKOGAWA representative from which the instrument was purchased, or the nearest YOKOGAWA o󰀩ce.
• If a problem arises with this instrument, please inform us of the nature of the problem and the circumstances under which it developed, including the model specication and serial number. Any diagrams, data and other information you can include in your communication will also be helpful.
• Responsible party for repair cost for the problems shall be determined by YOKOGAWA based on our investigation.
• The Purchaser shall bear the responsibility for repair costs, even during the warranty period, if the
malfunction is due to:
- Improper and/or inadequate maintenance by the
purchaser.
- Failure or damage due to improper handling, use
or storage which is out of design conditions.
- Use of the product in question in a location
not conforming to the standards specied by YOKOGAWA, or due to improper maintenance of the installation location.
- Failure or damage due to modication or repair
by any party except YOKOGAWA or an approved representative of YOKOGAWA.
- Malfunction or damage from improper relocation of
the product in question after delivery.
- Reason of force majeure such as res,
earthquakes, storms/oods, thunder/lightening, or other natural disasters, or disturbances, riots, warfare, or radioactive contamination.

1.3 Combination for Remote Sensor and Remote Transmitter

• The AXW remote sensor (sizes 500 to 1800 mm (20 to 72 in.)) should be combined with a remote
transmitter according to Table 1.2.
• If the transmitter combined with the AXW magnetic owmeter’s remote sensor is changed from the original transmitter which was delivered the meter factor of the remote sensor must be readjusted according to its ow calibration.
Table 1.2 Combination for sensor and transmitter
Remote Sensor
Model
AXW500G to AXW10LG AXW500W to AXW10LW
AXW500G to AXW18LG AXW500W to AXW18LW
AXW500G to AXW18LG AXW500W to AXW18LW
Converter,
Communication,
I/O
-W AXW4A
-B AXG1A
-N AXFA11G
Contact YOKOGAWA before using it in combination with transmitters other than those listed above.
In case of combination with AXFA11 remote transmitter, select “ADMAG AXF” in the parameter “C30” of the
AXFA11 remote transmitter.
Combined with the
Remote Transmitter
Model
IM 01E25A01-01EN
<2. Receiving and Storage>
F0201.ai
F0202.ai
Style: S1

2. Receiving and Storage

7
When the instrument is delivered, check visually that no damage has occurred during transportation. Also check that all owmeters mounting hardware shown below is
included.
Integral Flowmeter
Model Part name Qty.
AXW###G Blanking Plug (*1) 0 to 2 pcs.
Remote Transmitter
Model Part name Qty.
AXW4A Mounting Bracket 1 set
Blanking Plug (*1) 0 to 2 pcs.
AXG1A Mounting Bracket 1 set AXFA11G Mounting Bracket 1 set
*1: When the following code is specied for “Power Supply”
and “Converter, Communication, I/O”, the following quantity of blind plug is attached.
Power Supply code
-1
-2
Communication and I/O code
-D, -E, -M, -F 1 pc.
Other than those above 0 pc.
-D or -E 2 pcs.
-P 0 pc.
Other than those above 1 pc.
Qty
2.1 Model and Specications
Check
As shown in Figure 2.1 to Figure 2.5, the model, su󰀩x code, serial number, meter factor, uid specication, and device information are found on the name plate located on the outside of the housing. And, this instrument can check their information from parameters. Read the user’s manual
of applicable communication type as listed in Table 1.1 for
checking device information from parameters. When checking the matching of model and specication you ordered, see the applicable general specications as listed in
Table 1.1.
Be sure you have the model code and serial number available when contacting YOKOGAWA.
Note: Description on the nameplate
- Made in _______: Country of origin
- COMB No.: Serial number of the combined remote
sensor or remote transmitter
2
Receiving and Storage
Figure 2.1 Name Plate (AXW Integral Flowmeter)
Style: S2
Figure 2.2 Name Plate (AXW Remote Sensor)
IM 01E25A01-01EN
<2. Receiving and Storage>
F0204.ai
F0205.ai
F0203.ai
Figure 2.3 Name Plate (AXW4A Remote Transmitter)
Figure 2.4 Name Plate (AXG1A Remote Transmitter)
8
Figure 2.5 Name Plate (AXFA11 Remote Transmitter)

2.2 Storage Precautions

If the instrument is to be stored for a long period of time after delivery, observe the following points. The instrument should be stored in its original packing
condition in the storage location.
Select a storage location that fulls the following
conditions:
• A place where it will not be exposed to rain or water
• A place subject to minimal vibrations or shocks
Temperature and humidity levels should be as follows:
Temperature: -10 to 70°C
Humidity: 5 to 80% RH (no condensation) The preferred ambient temperature and humidity
levels are 25°C and approximately 65% RH.
If the instrument is transferred to the installation site
and stored without being installed, its performance may be impaired due to the inltration of rainwater and so forth. Be sure to install and wire the instrument as soon as possible after transferring it to the
installation location.
IM 01E25A01-01EN
<3. Installation>
IMPORTANT
NOTE
IMPORTANT
IMPORTANT

3. Installation

9
WARNING
Installation of the magnetic owmeter must be performed by expert engineer or skilled personnel. No operator shall be permitted to perform procedures
relating to installation.
Installation Location Precautions
Select the installation location with consideration to the
following items to ensure long-term stable operation of
the instrument.
Ambient Temperature:
Avoid installing the instrument in locations with constantly uctuating temperatures. If the location is subject to radiant heat from the plant, provide heat insulation or improve ventilation.
Atmospheric Condition:
Avoid installing the instrument in a corrosive atmosphere. In situations where this is unavoidable, consider ways to improve ventilation and to prevent rainwater from entering and being retained in the conduit pipes.
Vacuum:
In the case of PTFE lining, avoid the negative pressure inside the measuring pipe.
Vibrations or Shocks:
Avoid installing the instrument in a place subject to shocks or vibrations.

3.1 Piping Design Precautions

Design piping correctly, referring to the following to prevent damage to sensors and to assure accurate
measuring.
This section describes the remote sensor as an example. The same attention must be paid to the integral owmeter.
(1) Location
Install the owmeter in a location where it is not exposed to direct sunlight. The minimum ambient temperature is limited by the minimum uid temperature of the sensor (the lining). For more information, read the applicable general specication as listed inTable 1.1. The owmeter may be used in an ambient humidity where the relative humidity ranges from 0 to 100%. However, avoid long-term continuous operation at relative humidity 95% or higher.
(2) Noise Avoidance
3
Installation
The owmeter should be installed away from electrical motors, transformers, and other power sources in order to avoid interference with measurement. When installing two or more magnetic owmeters, provide a distance of at least 5D (D is size of model code) each other. If diameters of them are di󰀨erent, let D be the
larger one.
IM 01E25A01-01EN
F0301.ai
D: Sensor Size
2D or more
IMPORTANT
F0302.ai
(Incorrect) Upstream side
(Correct) Downstream side
IMPORTANT
F0303.ai
Bypass valve
Block valve
<3. Installation>
10
(3) Required Straight Pipe Length
Based on JIS B 7554 “Electromagnetic Flowmeters” and our piping condition test data, we recommend the piping conditions as shown in the following gures. This is not always enough when the piping line incorporates multiple conditions at the same time.
Gate valve fully open
5D or more2Dor more
Tee
0 is allowable.5D or more 0 is allowable.5D or more
Figure 3.1.1 Required Straight Pipe Length
*1: Do not install anything in the vicinity that may interfere
with the magnetic eld, induced signal voltages, or ow velocity distributions of the owmeter.
*2: A straight run may not be required on the downstream
side of the owmeter. However, if a downstream valve or other tting causes irregularity or deviation in ows, provide a straight run of 2D to 3D on the downstream side.
*3: The valves shall be mounted on the downstream side so
that deviated ows do not occur in the sensor and to avoid
startup from an empty condition.
*4: In case the piping conditions are compounded, install
on the straight pipe section where the upstream part is su󰀩ciently rectied.
Reducer pipe
0 is allowable. 0 is allowable.
90-degree bend
Expander pipe
10D or more
Various valves
10D or more
2D
or more
(4) Maintaining Stable Fluid Conductivity
(5) Precautions for Use of Liquid Sealing
Compounds
Care must be taken in using liquid sealing compounds on the piping, as it may have a negative inuence on the ow indications by owing out and covering the
surfaces of an electrode or grounding ring. In particular,
care must be taken if a liquid sealing compound is used in the case of vertical piping.
(6) Service Area
Select locations where there is adequate space to service installing, wiring, overhauling, etc.
(7) Bypass Line
It is recommended to install a bypass line to facilitate
maintenance and zero adjustment.
Block valve
Do not install the owmeter where uid conductivity tends to become uneven. If chemicals are fed near the upstream side of a magnetic owmeter, they may a󰀨ect the ow rate’s indications. To avoid this situation, it is recommended that the chemical feed ports be located on the downstream side of the owmeter. If it is unavoidable that chemicals must be fed on the upstream side, provide a su󰀩cient straight pipe length (approximately 50D or more) to ensure the proper mixture of uids.
Figure 3.1.2 Chemical Injection
Figure 3.1.3 Bypass Line
IM 01E25A01-01EN
<3. Installation>
F0304.ai
Abutment Abutment
F0305.ai
IMPORTANT
F0306.ai
(Incorrect)
IMPORTANT
F0307.ai
(Incorrect)
IMPORTANT
F0308.ai
Correct
the terminal
11
(8) Supporting the Flowmeter
CAUTION
Do not secure the owmeter separately to prevent the vibrations, shocks, and expansion and contraction forces of the piping from a󰀨ecting it. Fix the pipes rst, then support the owmeter with the pipes.
Magnetic flowmeter
To release the load from installing the owmeter to the pipeline, set a expansive pipe in the downstream side.
Magnetic flowmeter
Abutment Abutment
For vertical mounting, make sure that the weight of piping does not weigh on the owmeter.
(9) Mounting Positions
Expansive pipe
Avoid air bubbles.
If air bubbles enter a measurement pipe, ow rate indications may be a󰀨ected and measurement errors
may be caused.
In cases where uids contain air bubbles, piping must be designed to prevent them from accumulating in the
measurement pipe of a sensor.
If a valve exists near the owmeter, try to mount the owmeter on the valve’s upstream side in order to prevent a possible reduction of pressure inside the pipe, thereby avoiding the possibility of air bubbles.
(Correct)
(Incorrect)
Figure 3.1.5 Avoiding Air Bubbles
(Correct)
Valve
Mounting orientation
Install the magnetic owmeter so that the electrodes position is not perpendicular to the ground. Otherwise it may cause the measuring errors because air bubbles at upper side or slurry at downside covers the electrode. Mount the terminal box of a remote sensor and the transmitter of an integral owmeter above the piping to prevent water from entering them.
3
Installation
Pipes must be fully lled with liquids.
It is essential that pipes remain fully lled at all times, otherwise ow rate indications may be a󰀨ected and
measurement errors may be caused.
Piping shall be designed so as to maintain the interior of the sensor lled with uids. Vertical mounting is e󰀨ective in such cases as when uids tend to separate or solid matter may be precipitated. When employing vertical mounting, direct the uids from the bottom to the top to ensure that the pipes remain fully lled.
(Correct)
Figure 3.1.4 Mounting Positions
(Incorrect)
h
h>0
(Correct)
h>0
h
Incorrect
Electrode
Figure 3.1.6 Mounting Orientation
Electrode
Air bubble
Precipitate
Incorrect
Water can seep into
box.
IM 01E25A01-01EN
NOTE
F0309.ai
IMPORTANT
IMPORTANT
IMPORTANT
<3. Installation>
12

3.2 Handling Precautions

WARNING
The magnetic owmeter is a heavy instrument. Be careful that no damage is caused to personnel through accidentally dropping it, or by exerting excessive force on the magnetic owmeter. When moving the magnetic owmeter, always use a trolley and have at least two people carry it.
This section describes the remote sensor as an example. The same attention must be paid to the integral owmeter.

3.2.1 General Precautions

(1) Precaution during Transportation
The magnetic owmeter is packed tightly. When it is unpacked, pay attention to prevent damaging the owmeter. To prevent accidents while it is being transported to the installing location, transport it to the site in its original packing.
CAUTION
(2) Avoid Shocks from Impact
CAUTION
Care should be taken not to drop the owmeter or expose it to excessive shock. In particular, be careful not to subject the ange surface to shock. This may lead to lining damage which will result in inaccurate
readings.
(3) Flange Protection Covers
Keep the protective covering (i.e. the corrugated cardboard or other cushioning material) in place over the ange except when mounting the owmeter to the
pipe.
(4) Terminal Box Cover
As it is possible that the insulation will deteriorate, do not open the terminal box cover until it is time to wire it.
(5) Long-term Non-use
When lifting the magnetic owmeter, use the lifting rings
(eye bolts or eye plates) as in Figure 3.2.1.
Never lift it using a bar passed through the magnetic owmeter as this damages the liner severely.
90 degrees or less
Eye bolt Eye plate
Figure 3.2.1 Lifting Flowmeter
90 degrees or less
It is not desirable to leave the owmeter unused for a long term after installation. If this situation is unavoidable, take care of the owmeter by observing the following.
Conrmation of sealing conditions for the
owmeter
Conrm that the terminal box screw and electrical connections are well sealed. Equip the conduit piping with drain plugs or waterproof glands to prevent moisture or water from penetrating into the owmeter through the
conduit.
Regular inspections
Inspect the sealing conditions as mentioned above, and the inside of the terminal box at least once a year. Also, due to rain, etc. when it is suspected that water may have penetrated into the inside of the owmeter, perform
supplementary inspections.
IM 01E25A01-01EN
<3. Installation>
F0310.ai
NOTE
NOTE
IMPORTANT
13

3.2.2 Flowmeter Piping

CAUTION
Misaligned or slanted piping can lead to leakage and damage to the anges.
(1) Correct any misaligned or slanted piping, and any
gaps that may exist between mounting anges before installing the owmeter (see Figure 3.2.2).
Slanted
Figure 3.2.2 Slanted and Misaligned Flowmeter Piping
(2) Inside a newly installed pipeline, there may be some
foreign substances such as residue from welding or wood chips. Remove them by ushing the piping before mounting the owmeter. This prevents the lining from being damaged, as well as the occurrence
of erroneous measured signals resulting from foreign
substances passing through the sensor during
measurement.
Misaligned

3.3 Integral Flowmeter and Remote Sensor Installation

WARNING
• All gaskets used for piping of Magnetic Flowmeters should be prepared by customers except in some
cases.
With the option of grounding ring, if the lining is
natural hard rubber or PTFE, gaskets must be placed between each the grounding ring and the sensor to avoid uid leakage. These gaskets are to be supplied by customer. Do not forget those gaskets also when ordering and attaching the
grounding rings later additionally.
• Use the gasket of the equivalent hardness for gasket (customer pipe) and gasket (sensor).
• The tightening torque of gaskets varies by the type and external dimensions of the lining and the gasket. The tightening torque values should be
decided by referring to Table 3.3.1.
• For uids capable of potentially permeating PTFE linings (such as nitric acid, hydrouoric acid, or sodium hydrate at high temperatures), do not use the PTFE lining type.
3
Installation
• The PTFE lining has a structure adhering PTFE to the metal inner face of the sensor. When tightening bolts and nuts to install to the piping, be careful not to bring unequal stress or torque to the PTFE lining.
• For the PTFE lining, it is recommendedo be installed with grounding rings, or to be installed with short
pipes at upstream and downstream sides.
Use bolts and nuts in compliance with the ange ratings. When choosing the gaskets, be sure to choose sheet gaskets designed to t for ange standard.
IM 01E25A01-01EN
IMPORTANT
IMPORTANT
A
A
<3. Installation>
14
(1) Mounting Direction
Mount the owmeter so that the ow direction of the uid to be measured is in line with the direction of the arrow mark on the owmeter. It may be especially di󰀩cult to move large size sensors after bringing them into the pit. Check directions before bringing.
If it is impossible to match the direction of the arrow mark, the direction of the electrical connection can be changed. Read Section 3.5. In case the uid being measured ows against the arrow direction, change the value from “Forward” to “Reverse” at the parameter “Flow direct”. Read the user’s manual of the applicable communication type (for AXW/AXW4A/AXG1A) or the hardware/software
edition (for AXFA11) as listed in Table 1.1.
Display Menu Path (AXW/AXW4A/AXG1A):
Device setup ► Detailed setup ► AUX calculation ► Flow direct
(2) Lifting
90 degrees or less
90 degrees or less
(4) Applying Gasket and optional Grounding
Rings
Gasket:
Necessary gaskets for piping connection are as below depending on the choice of grounding rings. Use sheet gaskets designed to t for the ange ratings ad uid specications. The GF type-1 gaskets by JIS G 3443-2 should be used for the JIS F12 ange models in sizes 1100 mm (44 in.) and above. In this case, the gasket groove is required on the customer pipe anges. The thickness of gasket should be 2 mm (0.08 in.) to 5 mm (0.2 in.) for sizes up to 1000 mm (40 in.), and 5 mm (0.2 in.) or more for larger sizes. Use soft rubber gasket, PTFE-sheathed non-asbestos gasket, or its equivalent in hardness.
1. Standard (no grounding rings)
Customer
pipe
A A
Gaskets A supplied by customer
Figure 3.3.2 Installation without Grounding Rings
Customer
pipe
F0312.ai
Eye bolt
Figure 3.3.1 Lifting
Eye plate
F0311.ai
CAUTION
• When carrying the magnetic owmeter, use the
lifting rings (eye bolts or eye plates).
• To assure safety, keep lifting angle less than 90 degrees as shown in Figure 3.3.1
(3) Transportation and Positioning
Bring in the magnetic owmeter, place it and use a jack to
adjust its position.
When using the GF type-1 gaskets, the gasket groove is required as mentioned above.
2. With optional grounding rings (code GR1)
Optional grounding rings
Customer
pipe
B
Gaskets A and B supplied by customer
Figure 3.3.3 Installation with Grounding Rings GR1
B
Customer
pipe
F0313.ai
It is recommended to use the same gasket for A and B. Both gaskets A and B are to be supplied
by customer.
When polyurethane or natural soft rubber lining, the gaskets B are not necessary.
Apply the jack to the anges of the magnetic owmeter. In addition, adjust any misalignment when the magnetic owmeter is brought in, as the jack can adjust vertical position, but not horizontal one.
IM 01E25A01-01EN
<3. Installation>
15
3. With optional grounding rings (code GR2)
Optional grounding rings
with integrated gaskets B included
B
Customer
pipe
A A
Gaskets A supplied by customer
Figure 3.3.4 Installation with Grounding Rings GR2
B
Customer
pipe
F0314.ai
This is available for the models with process connection JIS F12 (JIS 75M) in sizes 1100 mm (44 in.) and above. Only gaskets A are to be supplied by customer. Use soft rubber gasket, PTFE-sheathed non-asbestos gasket, or its equivalent in hardness. The grounding rings are installed and locked onto the magnetic owmeter with gaskets B when shipped from factory.
Mounting Procedure (no Grounding Rings):
Connect the magnetic owmeter’s ange and the customer’s pipe which contact process uid by
some wire supplied by customer.
Mounting Procedure (with optional Grounding Rings GR1 for sizes up to 1000 mm (40 in.)):
1. Handles of the grounding ring have some holes which correspond to outer diameter of each ange type. There are printings near each hole. The printings show types of ange. See the
table below.
5. Install the magnetic owmeter into the customer’s pipe with the gaskets A.
Centering Pin
Magnetic Flowmeter
Note : Gasket A and B are also placed concentrically with
Figure 3.3.5 Mounting Procedure with Grounding Rings
Screw & Nut : Connect the wire from the
grounding ring or the customer’s pipe
Centering Pin
Clockwise
Gasket B
Grounding Ring
the magnetic owmeter.
GR1 for sizes up to 1000 mm (40 in.)
Handle
Printing of Flange Type
F12
10K
Hole
Handle
Printed Surface
F0315.ai
3
Installation
Printing
ASME -CA1
AWWA -CB1 AWWA C207 Class D PN10 -CE1 EN1092-1 PN10 PN16 -CE2 EN1092-1 PN16 10K -CJ1 JIS B2220 10K
AS -CS1, -CS2, -CT1
F12 -CG1 JIS F12 (JIS 75M)
Process Connection Code
Flange Standard
ASME B16.5 Class 150, ASME B16.47 Class 150
AS2129 table D, E
AS4087 PN16
2. Conrm the centering pin is xed to the hole corresponding to ange or x the centering pin to the correct hole.
3. Hang the grounding rings with their ange type printings outer side of the magnetic owmeter. Set the angle of both handles symmetrically to be 45 degree from top. If there are any bolt­holes under the handles, turn the grounding rings clockwise in order to locate handles between bolt-holes. Center the grounding ring to the center of the magnetic owmeter.
4. Connect the wire from the grounding ring to the screw of the magnetic owmeter’s ange and x the wire by the nut. This procedure (item 1 to 4) must be done for the both sides of the magnetic owmeter.
Mounting Procedure (with optional Grounding Rings GR1 for sizes 1100mm (44 in.)) and above :
1. The grounding ring is temporarily xed onto the magnetic owmeter by four bolts. Hang the grounding ring up using a crane or a hoist so that it would not fall down from the magnetic owmeter while setting the gasket B.
2. Unfasten the bolts and remove the grounding ring o󰀨 from the magnetic owmeter.
3. Place the gasket B to the magnetic owmeter. Cut out holes on the gasket B if necessary, so that the bolts can go through. Fix the grounding ring to the magnetic owmeter by the four bolts
again.
4. Connect the wire from the grounding ring to the screw at the ange and x the wire by the nut. This procedure (item 1 to 4) must be done for the both sides of the magnetic owmeter.
IM 01E25A01-01EN
<3. Installation>
5. Install the magnetic owmeter into the customer’s pipe with the gaskets A.
Screw & Nut : Connect the wire from the
grounding ring or the customer’s pipe.
Magnetic Flowmeter
Hang the grounding ring up with this point.
Unfasten the bolts and remove the grounding rings. Cut out some holes on the gasket B when necessary.
Wire
16
Grounding Ring
F0316.ai
Note : Gasket A and B should be placed concentrically with
the magnetic owmeter.
Figure 3.3.6 Mounting Procedure with Grounding Rings
GR1 for sizes 1100 mm (44 in.) and above
Mounting Procedure (with optional Grounding Rings GR2):
The grounding rings are installed and locked onto the magnetic owmeter with gaskets B when shipped from factory. Install the magnetic owmeter into the customer’s pipe with gaskets A supplied by
customer.
(5) Tightening Nuts
Pass the bolts from pipe line side, not magnetic owmeter side, and tighten the nuts according to the torque values for the metal piping in Table 3.3.1 or Table 3.3.2.
CAUTION
• Be sure to tighten the nuts according to the prescribed toeque values. Tighten them diagonally with the same torque values, step by up to the prescribed torque value.
IM 01E25A01-01EN
<3. Installation>
Table 3.3.1 Tightening Torque Values for Metal Piping (N-m)
Lining type
Gasket type Soft rubber gasket, or the equivalent in hardness (supplied by customer)
Process
Size mm (inch)
Size mm (inch)
connection
500 (20) 99 to 166 85 to 140 143 to 238 151 to 233
600 (24) 144 to 240 108 to 175 218 to 363 239 to 331
700 (28) 168 to 281 176 to 266 180 to 300 176 to 293 240 to 333
800 (32) 259 to 432 212 to 302 236 to 393 246 to 410 400 to 612
900 (36) 274 to 457 206 to 338 224 to 373 258 to 430 399 to 604
1000 (40) 274 to 457 277 to 422 307 to 512 286 to 477 422 to 587
Lining type Polyurethane rubber
Gasket type Soft rubber gasket, or the equivalent in hardness (supplied by customer)
Process
connection
500 (20) 71 to 118 60 to 99 74 to 123 76 to 123 131 to 218
600 (24) 97 to 162 82 to 126 101 to 168 118 to 171 124 to 206
700 (28) 125 to 208 135 to 202 146 to 244 120 to 200 167 to 242 202 to 337
800 (32) 176 to 294 160 to 229 190 to 316 169 to 281 218 to 348 197 to 328
900 (36) 170 to 283 151 to 251 178 to 296 161 to 268 200 to 317 221 to 369
1000 (40) 175 to 291 206 to 314 220 to 367 165 to 275 253 to 370 218 to 363
B16.5 (500, 600)
B16.47 (700-1000)
ASME
B16.5 (500, 600)
B16.47 (700-1000)
Class 150
ASME
EN1092-1 PN10
JIS B2220 10K
Class 150
PTFE / Natural hard rubber / Natural soft rubber
EN1092-1 PN10
JIS B2220 10K
EN 1092-1 PN16
EN 1092-1 PN16
AWWA C207
Class D
AWWA C207
Class D
AS2129 Table D, E
AS4087 PN16
AS2129 Table D, E
17
Unit: N-m
AS4087 PN16
Unit: N-m
JIS F12
(JIS 75M)
3
Installation
Lining type Natural hard rubber Polyurethane rubber
Gasket type Soft rubber gasket, or the equivalent in hardness (supplied by customer)
Process
Size
connection
mm (inch)
1100 (44) 285 to 476
1200 (48) 180 to 300 327 to 546 298 to 497 400 to 582 281 to 468
1350 (54) 397 to 662
1400 (56) 233 to 388 379 to 632
1500 (60) 421 to 702
1600 (64) 238 to 397 532 to 886 532 to 887
1800 (72) 291 to 485 552 to 919 468 to 779
EN1092-1 PN6 EN1092-1 PN10
AWWA C207
Class D
AS2129 Table D, E
AS4087 PN16
JIS F12
(JIS 75M)
Unit: N-m
IM 01E25A01-01EN
<3. Installation>
Table 3.3.2 Tightening Torque Values for Metal Piping (lbf-in)
Lining type
Gasket type Soft rubber gasket, or the equivalent in hardness (supplied by customer)
Process
Size mm (inch)
Size mm (inch)
connection
500 (20) 880 to 1467 748 to 1239 1266 to 2110 1332 to 2062
600 (24) 1275 to 2125 955 to 1546 1928 to 3214 2113 to 2932
700 (28) 1490 to 2483 1561 to 2357 1594 to 2657 1558 to 2596 2123 to 2943
800 (32) 2292 to 3820 1880 to 2671 2085 to 3475 2176 to 3627 3538 to 5413
900 (36) 2426 to 4043 1821 to 2991 1980 to 3301 2285 to 3809 3535 to 5346
1000 (40) 2426 to 4043 2450 to 3739 2717 to 4528 2534 to 4224 3734 to 5265
Lining type Polyurethane rubber
Gasket type Soft rubber gasket, or the equivalent in hardness (supplied by customer)
Process
connection
500 (20) 626 to 1044 527 to 876 652 to 1087 674 to 1089 1159 to 1931
600 (24) 862 to 1436 727 to 1113 894 to 1490 1043 to 1512 1095 to 1825
700 (28) 1106 to 1843 1193 to 1791 1295 to 2159 1061 to 1769 1478 to 2142 1788 to 2980
800 (32) 1562 to 2603 1420 to 2024 1680 to 2800 1494 to 2491 1926 to 3076 1739 to 2899
900 (36) 1505 to 2508 1340 to 2218 1571 to 2619 1422 to 2370 1773 to 2808 1958 to 3263
1000 (40) 1546 to 2577 1824 to 2781 1951 to 3251 1459 to 2432 2239 to 3274 1928 to 3213
B16.5 (500, 600)
B16.47 (700-1000)
ASME
B16.5 (500, 600)
B16.47 (700-1000)
Class 150
ASME
EN1092-1 PN10
JIS B2220 10K
Class 150
PTFE / Natural hard rubber / Natural soft rubber
EN1092-1 PN10
JIS B2220 10K
EN 1092-1 PN16
EN 1092-1 PN16
AWWA C207
Class D
AWWA C207
Class D
AS2129 Table D, E
AS4087 PN16
AS2129 Table D, E
18
Unit: lbf-in
AS4087 PN16
Unit: lbf-in
JIS F12
(JIS 75M)
Lining type Natural hard rubber Polyurethane rubber
Gasket type Soft rubber gasket, or the equivalent in hardness (supplied by customer)
Process
Size
connection
mm (inch)
1100 (44) 2527 to 4211
1200 (48) 1594 to 2657 2898 to 4830 2639 to 4398 3539 to 5149 2484 to 4139
1350 (54) 3517 to 5862
1400 (56) 2060 to 3434 3358 to 5597
1500 (60) 3729 to 6215
1600 (64) 2109 to 3515 4705 to 7842 4713 to 7854
1800 (72) 2576 to 4293 4883 to 8138 4138 to 6897
EN1092-1 PN6 EN1092-1 PN10
AWWA C207
Class D
AS2129 Table D, E
AS4087 PN16
JIS F12
(JIS 75M)
Unit: lbf-in
IM 01E25A01-01EN
<3. Installation>
IMPORTANT
F0317.ai
F0318.ai
Mounting fixture
4-ø6 Hole or
F0319.ai
Unit: mm
(approx. inch)
IMPORTANT
19

3.4 Remote Transmitter Installation

WARNING
Installation of the magnetic owmeter must be performed by expert engineer or skilled personnel. No operator shall be permitted to perform procedures
relating to installation.

3.4.1 Installation Location

Install the instrument in a location where it is not exposed to direct sunlight. For ambient temperature, read the applicable general specications as listed in
Table 1.1.
The instrument may be used in an ambient humidity where the RH ranges from 0 to 100%. However, avoid long-term continuous operation at relative humidity above 95%.
Horizontal Pipe Mounting
Nut
Washer
Screw
Figure 3.4.2 2-inch Pipe Horizontal Mounting
1. Fix the instrument on the mounting xture using four
screws.
2. Fix the mounting xture with the instrument installed on a 2-inch pipe using a U-bolt.
2-inch pipe
U-bolt
3
Installation

3.4.2 Mounting of AXW4A Transmitter

AXW4A Remote Transmitter can be mounted on a 2-inch pipe in a vertical or horizontal position depending on the
installation site.
Vertical Pipe Mounting
2-inch pipe
U-bolt
Nut
Mounting fixture
Washer
Screw
Figure 3.4.1 2-inch Pipe Vertical Mounting
1. Fix the instrument on the mounting xture using four
screws.
2. Fix the mounting xture with the instrument installed on a 2-inch pipe using a U-bolt.

3.4.3 Mounting of AXG1A Transmitter

AXG1A Remote Transmitter can be mounted using surface
mounting, 2-inch pipe mounting, or panel mounting.
Surface Mounting (Wall Mounting)
For surface mounting, use the mounting fixture provided, using M6 screws. These M6 screws must be provided by the user.
M6 Screw
65
(2.56)
194 (7.64)
Figure 3.4.3 Surface Mounting
Mounting xture on equipment intended to be mounted on a wall or ceiling shall withstand a force of four times the weight of the equipment (AXG1A: 3.5 kg (7.7 lb)).
IM 01E25A01-01EN
Loading...
+ 44 hidden pages