How to Contact Additel ......................................................................................................................................... 1
Safety Information ................................................................................................................................................ 2
1.1 Model Information .................................................................................................................................. 5
1.3 Features ................................................................................................................................................... 8
1.5 Power .................................................................................................................................................... 12
1.6 Standard Equipment ............................................................................................................................. 13
2. Get Started ...................................................................................................................................................... 15
2.3 Starting the Calibrator ........................................................................................................................... 19
2.4 Electric Connections .............................................................................................................................. 19
3.2.4 Service ........................................................................................................................................ 44
3.2.5 Process ....................................................................................................................................... 47
5.1 Control Settings ..................................................................................................................................... 62
5.2 24V Power ............................................................................................................................................. 62
5.3 Communication ..................................................................................................................................... 63
5.4 Head Correction .................................................................................................................................... 64
5.7.4 Safety Data ................................................................................................................................. 73
5.8 Power Management .............................................................................................................................. 73
5.9 System Information ............................................................................................................................... 74
V
Page 7
6. Task (Only available on 760-X-DL) ................................................................................................................... 75
6.1 New Task ............................................................................................................................................... 75
6.4 End of Task ............................................................................................................................................ 90
7. Data Logger (Only available on 760-X-DL) ....................................................................................................... 91
8.1 Leak Test ................................................................................................................................................ 93
8.2 Pressure Unit Converter ........................................................................................................................ 94
9. Service ............................................................................................................................................................. 95
9.1 Inlet Port Filter ...................................................................................................................................... 95
9.2 Pressure Outlet Port Filter and O-ring (Same as REF port) ................................................................... 96
VI
Page 8
Appendix 1: SCPI List ........................................................................................................................................... 97
A1.5 System ............................................................................................................................................... 109
A1.6 Status ................................................................................................................................................ 116
A1.7 Unit ................................................................................................................................................... 117
A1.8 Data Record....................................................................................................................................... 118
Appendix 2: Unit ............................................................................................................................................... 125
A3.3 Additel Land ...................................................................................................................................... 137
A3.4 Additel Log II ..................................................................................................................................... 140
A3.5 How to install ADT760 Driver by using Zadig .................................................................................... 143
Table 2 Model Information ............................................................................................................................ 5
Table 5 Standard Equipment ....................................................................................................................... 13
Table 6 Electric Connections ....................................................................................................................... 20
Table 7 Pressure Output Control Settings ................................................................................................... 25
Table 8 Auto Step ........................................................................................................................................ 28
Table 9 Min / Max / Avg .............................................................................................................................. 31
Table 16 Pressure Gauge Test Settings ........................................................................................................ 50
Table 17 Current / Voltage Transmitter Test Settings .................................................................................. 52
Table 18 HART Device Test Settings............................................................................................................. 54
Table 19 Pressure Switch Test Settings ........................................................................................................ 56
Table 20 I/P Converter Test Settings............................................................................................................ 59
VIII
Page 10
Table 21 Test Information ............................................................................................................................ 60
Table 22 Test Save Options .......................................................................................................................... 61
Table 23 Communication Settings ............................................................................................................... 63
Table 24 Head Pressure Correction ............................................................................................................. 64
Table 25 Date and Time Settings ................................................................................................................. 72
Table 26 Power Management ..................................................................................................................... 73
Table 27 Dial Pressure Gauge Information Settings .................................................................................... 76
Table 28 Pressure Transmitter Information Settings ................................................................................... 78
Table 29 Pressure Switch Information Settings ........................................................................................... 79
Table 30 I/P Converter Information Settings ............................................................................................... 80
Figure 16 Jog Function ................................................................................................................................ 50
Figure 18 Inlet Port Filter ............................................................................................................................ 95
Figure 19 REF and Outlet Port Filter ............................................................................................................ 96
X
Page 12
Welcome
The Additel 760 Automatic Handheld Pressure Calibrator is a hand held controller, as well as a documenting process
calibrator, all rolled into one with high precision and control stability. Each unit comes with an internal pressure module
ranged to cover the full scale range of the unit. The internal module is located at the back of the calibrator and is
removable allowing for the selection of lower range pressure to improve accuracy. In addition, the calibrator can work
with two external pressure modules at the same time. With high precision electrical signal measurement function, this
calibrator can be used for dial pressure gauges, digital pressure gauges, differential pressure gauges, pressure
transmitters, I/P converters, or pressure switch calibration. For the most up-to-date manual, please visit
www.additel.com.
How to Contact Additel
Additel Corporation
Phone: +1-714-998-6899
Fax:+1-714-998-6999
E-mail: sales@additel.com or service@additel.com
Website: www.additel.com
1
Page 13
Safety Information
WARNINGS identify action or condition that may be hazards to the user;
CAUTIONS identify action or condition that may damage the calibrator or the equipment under test.
WARNINGS:
To prevent personal injury, please follow this user manual.
To prevent possible electrical shock, fire, or personal injury, please:
Check product exterior before use
Read and follow all instructions carefully
Do NOT apply more than 30V AC or DC between any two electric jacks
Correctly place and lock the battery before use
Avoid close contact to the vent port when venting
Charge the battery immediately with supplied power adapter when the low battery indication occurs. This will
prevent potential pressure release due to a loss of electrical power
Do NOT use the product if it is damaged or operates incorrectly
Do NOT use in flammable, high humidity, or dusty environments
Only use accessories, test leads, and probes that have the same measurement specification
2
Page 14
CAUTIONS:
Do NOT shake, drop, or bump the calibrator while in use
If condensation occurs, thoroughly dry out the calibrator before startup
Do NOT use any power adapter other than 9816-X
Charge the battery immediately when low battery symbol indicates
Do not remove the battery while it is in charging or when the calibrator is in use
Release the system pressure before turning off the calibrator
3
Page 15
Product Symbols Table 1.
Symbol
Meaning
Important information
Danger
See user manual
Risk of electrical shock
Hazardous voltage
Conforms to European Union directives
Please recycle
Lithium-ion battery
Table 1 Symbols
4
Page 16
Model
Model No.
Pressure Range
Pressure Type
Low Differential Pressure
LLP
±30 inH2O (±75 mbar)
D, G
Differential Pressure
D
-12.5 to 35 psi (-0.86 to 2.5 bar)
D, G
General Pressure
MA
-12.5 to 300 psi (-0.86 to 20 bar)
G, A
1. Introduction
1.1 Model Information
Table 2 Model Information
NOTE: D - Differential; G - Gauge; A - Absolute.
5
Page 17
1.2 Basic Structure
Figure 1 Basic Structure
6
Page 18
No.
Name
(1)
Battery Lock
(2)
Power Switch
(3)
Touch Screen
(4)
Electrical Measurement Panel
(5)
REF Port (-MA: Barometer Sensor Calibration Port)
(6)
Pressure Outlet Port
(7)
Power Adapter and USB Port Panel
(8)
Battery
(9)
Inlet Port
(10)
Exhaust Port
(11)
VENT Port
(12)
Internal Pressure Module
(13)
External Pressure Module Connection Ports (A & B)
(14)
Product Nameplate
Table 3 Basic Structure
7
Page 19
1.3 Features
Fully automatic calibrator with built-in pump and controller
Switchable internal pressure modules for expandable ranges
Accuracy (1 year) of 0.02%FS
External pressure modules available (measurement only)
Less than 4 lbs. (1.8 kg) for handheld operation
Source pressure, measure pressure and measure electrical
4 channels
Optional HART communication
Optional data logging and documentation
USB, Wi-Fi, and Bluetooth communications
8
Page 20
1.4 Technical Specification
1.4.1 General Specification
◆Media:Clean Air
◆Internal module temperature compensation range:32 ~ 122℉ (0 ~ 50℃)
◆Stability:0.005%FS or 0.05 pa whichever is greater, based on FS of the switchable internal
pressure module.
◆Display:5 inch 480 x 800 TFT color touch screen
◆Size:9.37" x 4.33" x 2.76" (238mm × 110mm × 70mm)
◆Weight:less than 4 lbs. (1.8 KG)
◆Connection:
1) -LLP:
Two Barb fittings
2) -D:
One Barb fitting
One Hose, 5 ft (1.5m), with built-in filter to 1/4 BSPF, 1/4 NPTF, and M20F adapters
3) -MA:
One Hose, 5 ft (1.5m), with built-in filter to 1/4 BSPF, 1/4 NPTF, and M20F adapters
One barometer sensor calibration port
NOTE: Filters are installed in the pressure connection ports. By passing the filter will void the
product warranty.
9
Page 21
◆Inlet filter:Dedicated filter,50μm
◆Electrical Measuring Jacks:Φ4 mm electrical jacks
◆USB: Standard USB port, type A
◆External Pressure Module Port: Two 5-pin connections (A & B)
1.4.2 Environment Specification
◆Working Temperature: 32 ~ 122℉ (0 ~ 50℃)
◆Storage Temperature: -4 ~ 158℉ (-20 ~ 70℃)
◆Humidity:0 ~ 90% (32 ~ 122℉ or 0 ~ 50℃), Non condensation
◆Atmosphere Pressure:Less than 9,800 ft. (3,000 m)
◆Protect Level:IP54
10
Page 22
Signal Type
Range
Resolution
Accuracy
Note
Current Measurement
±30 mA
0.0001 mA
±(0.01%RD + 0.005%FS)
Input:<10 Ω
Voltage Measurement
±30 V
0.0001 V
±(0.01%RD + 0.005%FS)
Input:>1M Ω
Current Output
24 mA
0.001 mA
±(0.01%RD + 0.005%FS)
20 mA @ 1K
Loop Power Source
24 V
N/A
±1 V
Max Load: 50 mA
Pressure Switch
Open, close. Support for mechanical switch and NPN / PNP digital switches.
Temperature
Compensation
41℉ to 95 ℉ (5℃ to 35 ℃)
Temperature Coefficient
<(±0.001%RD + 0.001%FS) / ℃ outside of 5℃ to 35 ℃
1.4.3 Electrical Specification
Table 4 Electrical Specification
11
Page 23
1.5 Power
Power Supply:Power supplied by rechargeable Li-Ion battery or an external 10 V power
adapter
Li-Ion rechargeable battery is removable and can be charged directly using the external
power adapter
The battery should be removed from the calibrator and charged directly for the fastest charge
time
Charge Time:Full charge, less than 4 hours when removed from the calibrator and charged
directly
The charging time will be extended when battery is charged through calibrator
Battery Working Time:More than 10 hours or 100 cycle of full scale pressure
12
Page 24
Model
Quantity
-LLP
-D
-MA
760 Calibrator
1
● ● ●
9724 Rechargeable Li-ion Battery
1
● ● ●
9816-X External Power Adapter
1 pc
● ● ●
9025 Test Leads
1 set (2 red, 1 black)
● ● ●
Alligator Clip
2 sets
(2 red, 1 black for each set)
● ● ●
Silicone Tube
Depends on 760 model
●
(2 pcs)
●
(1 pc)
ADT100 Hose
1 pc
● ●
ADT100-760-KIT Adapter Set
1 set
(1/4BSPF, 1/4NPTF, and M20F)
● ●
Barb Fitting Adapter
1 pc
●
O-ring
1 pack
● ● ●
Filter for Outlet Port
2 pcs
● ● ●
Filter for Inlet Port
2 pcs
● ● ●
CD Manual
1 pc
● ● ●
NIST-Traceable Certificate
1 pc
● ● ●
1.6 Standard Equipment
Table 5 Standard Equipment
13
Page 25
Figure 2 Standard Equipment
14
Page 26
2. Get Started
2.1 Battery
WARNINGS:
Lithium-ion batteries may get hot or ignite and cause serious injury if not treated properly
Do NOT incinerate or heat the battery
Do NOT storage the battery in direct sunlight
Do NOT try to disassemble the battery
The battery contains dangerous chemical substances
Please clean with water and look for medical treatment if exposure to chemicals occurs
CAUTIONS:
Please remove the battery and store in a cool, dry, ventilated area if it will not be used for a long period
Do not store batteries with hazardous or combustible materials
Discontinue use and contact Additel or an authorized distributor immediately if the battery leaks
Do not short circuit the battery
Only use the dedicated power adapter (9816-X) to charge the battery
Be sure to properly dispose of the battery if it is no longer usable
15
Page 27
The Additel 760 Automatic Handheld Pressure Calibrator uses a dedicated rechargeable, Li-Ion
battery (model 9724). Each battery has a built-in protection circuit allowing the battery to be charged
independent of being attached to the calibrator.
Each battery contains an illuminated charge indicator. Push button to see the level of charge.
The battery is fully charged when 5 blue LEDs are illuminated.
Figure 3 Battery Install and Charging
◆ To Install the battery:
1) Align to alignment markings of the calibrator and battery
2) Push the battery
3) Use a flat-head screwdriver to lock the battery
16
Page 28
◆ To Remove the battery:
1) Turn off the calibrator
2) Turn the battery lock to the unlock position with a flat-head screwdriver
3) Remove the battery
◆9816-X External Power Adapter:
1. Input: AC 85V ~ AC 265V, 50Hz / 60Hz
2. Output: DC 10V, 2A, 40W MAX
17
Page 29
2.2 Internal Pressure Module
The Additel 760 Calibrator comes with an internal pressure module which covers the full span of the
unit. This pressure module is user-switchable and Additel offers a large variety of different range
pressure modules.
◆To switch the internal pressure module:
1)VENT the system pressure (see section 5.5.3)
2)Turn off the calibrator, disconnect the power adapter if it is connected, and remove the battery
3)Unscrew both screws on the pressure module until the screw pin is loose and the screw pin can
be extended out
4)Firmly pull the sensor out of the calibrator
5)Insert a new internal pressure module (please pay attention to module direction)
6)Tighten both screws on the pressure module until the screws reach the bottom
Figure 4 Internal Pressure Module Switching
18
Page 30
2.3 Starting the Calibrator
Push the power button to turn on the calibrator. The display will illuminate and the Additel logo will
display. It will take a few seconds for the operating system to load after which the main screen will
display.
◆Date and Time: To change date and time setting, please see section 5.7.1
◆Language: The default language is English. To change the language please refer to section 5.7.2
2.4 Electric Connections
Electric connection instruction:
Figure 5 Electric Connection
19
Page 31
Main Jacks
Power Supply Jack
Function
Black(1)
Red(2)
Red(3)
Function
±30 V Voltage
±30 mA Current
2-wire 4 ~ 20 mA Loop Transmitter
3-wire Voltage/Current Transmitter
Non-isolated 24 V
power output
Mechanical Switch
NPN/PNP Switch
Non-isolated 24 V
power output
HART Transmitter
(Internal resistor, Internal power supply)
HART Device
(External resistor, External power supply)
0~20 mA Loop Current Output
(Internal power supply)
0~20 mA Loop Current Output
(External power supply)
Table 6 Electric Connections
20
Page 32
2.5 Pressure Connection
Figure 6 Pressure Connection
21
Page 33
3. Operation
3.1 Display and Operation
3.1.1 Main Screen
The main screen normally contains the Status Bar, Title Bar, Electrical Measurement Window,
Pressure Window, and Control Bar.
①Status Bar:Includes date and time, external pressure module status , communication status,
24V internal power status , and battery status.
NOTE: Information on Status Bar is read only. All other icons, pressure units, and output
values can be selected for further operation.
②Title Bar:Includes screen lock, abnormal notification,and system menu.
◆Abnormal notification signcan be cleared by rebooting the calibrator. Specific information about
the abnormal condition can be reviewed in the system log, please see section 5.5.3.
③Electrical Measurement Window:Includes measurement value, data analysis (current/voltage
measurement and transmitter test only), current output step (current output only), and loop current
value (transmitter test only).
④Pressure Window:Includes pressure set point value , current pressure value, pressure unit, step
control, and step size.
⑤Control Bar:Including Vent, Measure, and Control button.
◆Do not point the VENT port toward personal during venting
22
Page 34
◆When using one or two external pressure modules, the main screen will add additional windows to
display the added module or modules
◆Every window can be collapsed
◆Press on the top right corner of the window and select collapse
Press on the top right corner of a collapsed window and select expand to open that window
Figure 7 Main Screen
23
Page 35
Factor
Unit Name
Reference
2
USER1
1*kPa
3.1.2 Pressure Unit
Press the pressure unit on Pressure Window to change system unit. There are 25 default pressure
units which can be selected: kPa, Pa, MPa, hPa, bar, mbar, torr, atm, psi, gf/cm2, kgf/cm2,
In addition, there are 5 customizable pressure units which can be set up.
◆ To set up a custom pressure unit:
1) Press in pressure unit selection screen and set up three parameters:
①Factor:This is the multiplier against the reference pressure unit.
②Unit Name:Custom unit name.
③Reference:Reference pressure unit, should be selected from one of the 25 default pressure
units.
◆The calculation equation is: Factor * Unit Name = Reference
For example:
In this way, 1 USER1 pressure equals to 0.5 kPa.
24
Page 36
Subject
Valid Value
Comment
Pressure Type
Diff. / Gauge / Absolute
The pressure type displayed will depend on the ADT760 model
Resolution
4 / 5 / 6
Display digit
Slew Rate
0~100%
Pressurization speed, depends on internal pressure module
Stability Limit
±(0.003~1)% FS
Pressure control stability condition
Vent Limit
Depends on 760
model
Pressure at which the vent valves will open to atmosphere
Auto Zero
On / Off
Pressure measurement will auto zero every time the unit is
vented (not available with all models)
Set Point Limits
On / Off
Limits the pressure output ranges
Set Point Limits (on)
Depends on internal
pressure module
User editable for 0% and 100% of span ranges
Module Information
N/A
Information is read only and includes range, pressure type,
accuracy, SN, version, and manufacture date
Running Information
N/A
Information is read only and includes supply pressure, vacuum
pressure, and atmosphere
3.1.3 Pressure Output
1. Pressure Output Control Settings:
Press on the Pressure Window to enter the control settings menu:
Table 7 Pressure Output Control Settings
25
Page 37
2. Connection:
Figure 8 Output Connection
26
Page 38
3. Pressure Set Point:
①Manual Input:
1)Press pressure set point value or current pressure value on the Pressure Window and input the
target pressure.
2)Press the enter key or .
NOTE: Pressure set point should not exceed pressure range limit shown above the keyboard.
②Manual Step:
1)Press step size on the Pressure Window and input the target step size.
2)Pressto decrease or increase pressure by the target step size.
27
Page 39
Subject
Valid Value
Comment
Range
Depends on internal pressure module
The start and end points of auto step
Step Mode
By Points/ Percent / Size
Calculation method for pressure
change of each step
Points
Pressure change of each step
= Range ÷ Points
Pressure change calculation
Percent
Pressure change of each
= Range × Percent ÷ 100
Step Size
Pressure change of each
= Step Size
Stroke
Round Trip / One Way
Auto Step stroke type
Dwell Time
0~100,000,000(s)
Retention time after each step
Repeat
1 / +∞
Repeating criteria
③Auto Step:Press on Pressure Window, then select Auto Step:
1) Auto Step settings:
2) Press , then select Start.
4. Discontinue Control or Pressurization:
◆Manual Input and Manual Step:
Press Measure to pause and press Control to restart.
Press Vent and the pressure will decrease and release.
Table 8 Auto Step
28
Page 40
◆Auto Step:
Press Pause to pause.
Press Exit to exit and the calibrator will release the pressure.
Press Vent to release the pressure then press Exit to stop the test or Continue to resume the test.
5. Pressure Stabilization:
If the difference between current output pressure value and set point is smaller than Stability Limit,
the current pressure value on the Pressure Window will turn to green with a beep to indicate the
output pressure has reached the set point stability criteria.
29
Page 41
3.1.4 Current/Voltage Measurement
1. Press on Electrical Measurement option ( ) on the top left corner of Electrical
Measurement Window and select:(-30~30) mA or (-30~30) V.
◆Do not apply current / voltage that exceeds current / voltage measuring range
◆Apply a short circuit to zero
◆If the measurement value falls outside of the current/voltage measuring range, then it will turn red
with an alarm
◆If measure value exceeds measure limit of the calibrator, it will display "--------" with alarm
2. Connection
Figure 9 Current / Voltage Measurement
30
Page 42
Subject
Valid Value
Comment
Samples
1~100
Sample collection quantity
Subject
Valid Value
Comment
Measuring Range
Depends on measurement type
Electrical measurement range
Scale Range
Number, letter, or pressure,
temperature, electric unit
Scale unit
-999999~999999
Zero point value
-999999~999999
100% point value
Resolution
4, 5, 6
Display digit
Transfer Function
Linear / Square Root
Transfer function type
3. Data Analysis:
Press and select one of the functions below:
① Min / Max / Avg:Shows the minimum, the maximum, and the average
Table 9 Min / Max / Avg
② Scale:
Table 10 Scale
◆To turn off data analysis function press and select or .
31
Page 43
3.1.5 Mechanical Switch
1. Press the Electrical Measurement option ( ) on the top left corner of Electrical
Measurement Window and select Switch.
2. Connection
Figure 10 Mechanical Switch Measurement
3. Trip Values
◆The calibrator records the trip values and the state (open / close) of the switch.
32
Page 44
3.1.6 NPN /PNP Switch
1. Press the Electrical Measurement option ( ) on the top left corner of Electrical
Measurement Window and select Switch NPN or Switch PNP.
2. Connection
3. Trip Values
◆The calibrator records the trip values and the state (open / close) of the switch.
Figure 11 Electrical Switch Measurement
33
Page 45
3.1.7 Current Output
1. Press the Electrical Measurement option ( ) on the top left corner of Electrical
Measurement Window and select Simulate mA or (0 ~ 24) mA.
2. Connection:
Figure 12 Current Output
3. 24V Internal Power:
◆Simulate mA:
Internal 24 V power should be turned off automatically, for more information please see section 5.2
◆(0~24) mA:
Internal 24 V power should be turned on automatically, for more information please see section 5.2
34
Page 46
4. Current Set Point:
①Manual Input:
1)Click Measure Value on the Electrical Measurement Window, and input the target value
2)Press .
②Manual Step:
1)Press step size on the Electrical Measurement Window and input the current output range
2)Press/ each time to decrease or increase current output by 25% of range
NOTE: Current output range shall not exceed 0~24 mA.
35
Page 47
3.1.8 External Pressure Module
1. Connection:
Figure 13 External Pressure Module Connection
36
Page 48
Subject
Valid Value
Comment
Pressure Type
Gauge / Absolute / Diff.
External pressure module pressure type
Resolution
4 / 5 / 6
Display digit
Unit
Up to 25 default units or 5
custom units
External pressure module unit is independent of
the system unit, see section 3.1.2
Module Information
N/A
This information is read only: including range,
pressure type, accuracy, SN, version, and
manufacture date
2. Display
◆Symbols indicate the associated external pressure modules are connected.
The main screen will add one/two additional windows when connecting one/two external pressure
module.
◆The external pressure module can only be used for pressure measurement, not for pressure
control.
◆If the measurement value falls outside of the range then it will turn red with an alarm.
◆If measurement values exceed the limit of the calibrator then it will display "--------" with an alarm.
3. Settings:
Press on the external pressure module window to enter external pressure module setting
menu:
On top of Electrical Measurement Window please press (or ) and select:
1). Internal Power and Resistor
2). External Power and Resistor
39
Page 51
2. Search
1) Poll
Press on Electrical Measurement option ( ) on the top left corner of the Electrical
Measurement Window and select HART. The calibrator will poll address 0 if no connection detected.
Then it will scan addresses 1 to 15 automatically.
2) Manual Search
HART device manual search is available if necessary.
1)Press on Electrical Measurement Window, then select Search. The calibrator will list all
HART device detected.
2)Select the device which needs to be tested. If still no required HART device are detected then
press to scan address from 1 to 15 automatically.
3.2.2 Operation Window
The HART measurement value will display on the Electrical Measurement Window. Other process
variables can be displayed, see section 3.2.5. Loop current and HART device name are displayed on
the bottom of the window.
◆Press the pressure unit to change HART device pressure measuring unit
◆Changing HART device pressure measure unit will not change the calibrator system unit
40
Page 52
Subject
Type
Comment
Manufacturer
Read Only
Manufacturer of HART device
Device Type
Read Only
HART device type
Device ID
Read Only
HART device ID number
Tag
Editable
Alphanumeric content (8 max length)
Date
Editable
HART device date setting
Write Protect
Read Only
Write protection for HART device
Message
Editable
Alphanumeric content (32 max length)
Descriptor
Editable
Alphanumeric content (12 max length)
Final Assembly Number
Editable
Numeric content (12 max length)
Preambles
Editable
Numeric content (12 max length)
Universal Revision
Read Only
Universal version number
Software Revision
Read Only
Software version number
Hardware Revision
Read Only
Hardware version number
Device Revision
Read Only
Device version number
3.2.3 Settings
Press and select Settings to enter HART device settings menu:
1.Device Information: HART device information
Table 12 HART Device Information
41
Page 53
Subject
Type
Comment
Sensor Serial Number
Read Only
HART device sensor SN
Sensor Unit
Read Only
HART device pressure measurement unit (user selectable, see
section 3.2.2
Sensor Lower Limit
Read Only
HART device sensor measurement range lower limit
Sensor Upper Limit
Read Only
HART device sensor measurement range upper limit
Sensor Min Span
Read Only
HART device sensor measurement resolution
2. Sensor: HART device sensor information
Table 13 HART device sensor information
42
Page 54
Subject
Type
Valid Value
Comment
PV / Range Unit
Editable
Select from 25 default units
or from 5 custom units
HART device pressure measurement
unit, see section 3.1.2
PV LRV
Editable
Within HART device sensor
range
Process variable lower range
PV URV
Editable
Within HART device sensor
range
Process variable upper range
Transfer function
Editable
Linear / Square Root
Transfer function type
Alarm State
Read Only
N/A
Alarm State
Damping
Editable
1~100
Damping time, only support one digit
after decimal point
Polling Address
Editable
0~15
HART device polling address
Burst Mode
Editable
On / Off
Burst mode status
Burst Command
Editable
1 / 2 / 3
Burst command, depends on HART
device
3. HART Device Output
Table 14 HART Device Output
43
Page 55
3.2.4 Service
Pressand select Service to enter HART device service menu:
1. Loop Test:
This test allows you to compare the HART device current output value with the ADT760 calibrator
current measurement value.
◆The loop test function available only when HART device address is 0
1). Input a current value or select Fetch to capture the calibrator current measurement value and
then press the enter key or select to apply.
2). After few seconds, the calibrator will send a signal to HART device comparing the HART device
current output value (on the top left window) with the calibrator's current measurement value. If
the difference is out of tolerance it is recommended to trim the HART device.
2. Pressure Zero Trim:
This test is designed to do a zero point trim on the HART device.
1). If using an external pressure module ensure the correct module is selected.
2). The calibrator will automatically vent the pressure. Wait until the measurement stabilizes and
turns green.
3). Press Zero to trim HART device pressure zero point.
44
Page 56
3. D/A Trim:
This test is designed to trim the HART device current output value at zero and span.
◆D/A Trim function available only when HART device address is 0.
①D/A Zero:
1). Input a current value (4 mA is typical) or select Fetch to capture the calibrator measurement
value (on the top right window). Then press the enter key or select to apply.
2). After few seconds, the calibrator will send a signal to HART device to trim the current value.
②D/A Gain:
1). Input a current value (20 mA is typical) or select Fetch to capture the calibrator measurement
value (on the top right window). Then press the enter key or select to apply.
2). After few seconds, the calibrator will send a signal to HART device to trim the current value.
4. Sensor Trim:
This test is designed to trim the HART device sensor and the lowest and highest points of the range.
①Low trim:
1). Calibrator will control pressure to the lowest point of HART device range automatically.
2). Input calibrator pressure measuring value (on the top left window) through soft key board or
select Fetch to catch 760 Calibrator current measuring value to trim HART device pressure
measuring value.
3). Press the enter key or select to apply.
②High trim:
1). The calibrator will automatically control the pressure to the highest point of the HART device
45
Page 57
range.
2). Input the calibrator pressure measurement value (on the top left window) or select Fetch to
capture the calibrator value to trim the HART device.
3). Press the enter key or select to apply.
③Trim Reset:
This feature will reset the HART device back to factory parameters.
46
Page 58
Subject
Comment
PV
Process variable
PVAO
Digital current value
%range
Percentage of pressure range
SV
Secondary variable
TV
Tertiary variable
QV
Quaternary variable
Loop Current
Loop current value
3.2.5 Process
Press and select Process to enter the HART device process variable setting menu:
Table 15 HART Device Variable
◆Selected process variable will be displayed at the top left corner of Electrical Measurement Window,
after the HART- icon
47
Page 59
3.3 Zeroing
The zeroing function is available for current/voltage measurement and pressure measurement either
internal or external pressure modules.
◆To zero: Press and select
◆Apply a short circuit for current / voltage zeroing
◆Allowable zeroing range:
Pressure: within 1%FS;
Current / Voltage: within 0.1%FS.
3.4 Vent
Press Vent on Control Bar and the calibrator will run the vent process.
◆The vent process will end pressurization during a regular test, typical application, task, and leak
test
◆The VENT port should not face the operator during venting process
◆Dust or contaminants may blow out through the VENT port
◆Any moisture or liquid may also blow out through the vent port
48
Page 60
4. Typical Applications
The Additel 760 Automatic Handheld Pressure Calibrator supports six typical applications.
To start, please select Calibrator under Main Menu, then select the desired application.
◆In the operation interface, press to re-select from the application list
◆In the operation interface, press to enter the new test set up
◆Zeroing:If current / voltage measurement or internal pressure module requires zeroing then press
and select to zero the measurement, see section 3.3
◆Custom applications can be built with several simple selections
49
Page 61
Subject
Valid Value
Comment
Range
Depends on UUT
UUT range
4.1 Pressure Gauge (Includes dial and digital pressure gauges)
In pressure gauge mode, pressure is the only output function available.
1. Pressure Output
◆Ensure proper pressure connections
◆ For pressure output, please see section 3.1.3
◆Jog:Press and select Jog, which allows for a digit to be selected and numerically
incremented, see Figure 16
Figure 16 Jog Function
◆Numeric Keyboard:Press and select to allow for quick entry of pressure values
2. New Test:
2.1 New Test Settings:①UUT (Unit Under Test):
Table 16 Pressure Gauge Test Settings
②Test Method:
See section 3.1.3-3-③Auto Step.
50
Page 62
2.2 Test Execution:
After test settings are complete, press to execute the test:
1). Press to zero pressure module if necessary;
2). Press to start the test;
◆The calibrator will pressurize to the first set point (which may be zero)
◆Press to switch calibration interface
◆Pressto check or change test settings
3). Standard test steps:
(1). After the pressure shows it is stable, the gauge pressure value can be entered in the box on
the right. Press to switch the calibration interface.
(2). Press the enter key or for next test point and the calibrator will start to pressurize again
(3). Wait for the pressure value to be generated and stabilize and then enter the UUT values and
proceed to the next point.
(4). Repeat step (2) and (3) until all test points passed.
2.3 Report:
Once the test is complete, the test report interface will appear. For more information, please see
section 4.6.
51
Page 63
Subject
Valid Value
Comment
Input
Depends on UUT
UUT pressure measuring range
Output
Depends on UUT
UUT electrical signal output range
Transfer Function
Linear
Transfer function type
4.2 Current / Voltage Transmitter
In the Current / Voltage Transmitter mode, only current, voltage, and pressure are available functions.
1. Current / Voltage Measurement:
◆Correct electrical connection please see Figure 9
◆Data Analysis:Please see section 3.1.4-3
2. Pressure Output
◆Correct output connection please see Figure 8
◆Pressure output: Please see section 3.1.3
3. New Test:
3.1 New Test Settings:
①UUT (Unit Under Test):
②Test Method: See section 3.1.3-3-③Auto Step
Table 17 Current / Voltage Transmitter Test Settings
52
Page 64
3.2 Test Execution:
After test settings are complete, press to execute the test:
1). Press to zero the pressure if necessary.
2). Press to run a manual test, or press to run an automatic test.
(1).In manual mode, run next test point by pressing, and the calibrator will automatically record
the electrical signal and pressure value.
(2).In automatic mode, the calibrator will run the test and record the electrical and pressure values
automatically without the need to advance the calibrator to the next set point.
◆Press to switch calibration interface
◆Pressto check or change test settings
3.3 Report:
Once the test is complete the test report interface will appear. For more information, please see
section 4.6.
53
Page 65
Subject
Valid Value
Comment
Input
Depends on UUT
UUT pressure measuring range
Transfer Function
Linear / Square Root
Transfer function type
4.3 HART Transmitter
In HART Transmitter mode, only the Process Variable function and the Pressure function are
available.
1. HART Function:
◆Correct electrical connection and HART function, please see section 3.2
2. Pressure Output:
◆Correct output connection please see Figure 8
◆Pressure output: Please see section 3.1.3
3. New Test:
3.1 New Test Settings:①UUT (Unit Under Test):
Table 18 HART Device Test Settings
②Test Method:
See section 3.1.3-3-③Auto Step
3.2 Test Execution:
After test settings are complete, press to execute the test:
54
Page 66
1).Press to zero the pressure if necessary.
2).Press to run a manual test, or press to run a automatic test.
(1).In manual mode, run the next test point by press calibrator and the calibrator will
automatically record the electrical signal and pressure value.
(2).In automatic mode, the calibrator will run the test and record the electrical and pressure values
automatically without the need to advance the calibrator to the next set point.
◆Press to switch calibration interface
◆Pressto check or change test settings
3.3 Report:
Once the test is complete the test report interface will appear. For more information, please see
section 4.6.
55
Page 67
Subject
Valid Value
Comment
Range
Depends on UUT
UUT pressure range
Accuracy
0.01 ~ 100
UUT accuracy
Switch Type
NC (normal close) / NO (normal open)
Pressure switch type
Repeatability
On / Off
Pressure switch repeatability
4.4 Pressure Switch
In the Pressure Switch mode only switch and pressure measurement functions are available.
◆Presson the top left corner of Switch Measurement window to select the switch type:
Mechanical Switch, Switch NPN, or Switch PNP.
1. Switch Measurement:
◆Correct electrical connection please see Figure 9
2. Pressure Output
◆Correct output connection please see Figure 8
◆Pressure output: Please see section 3.1.3
3. New Test:
3.1 New Test Settings:①UUT (Unit Under Test):
Table 19 Pressure Switch Test Settings
56
Page 68
3.2 Test Execution:
After test settings are complete, press to execute the test:
1). Press to run test automatically
◆Pressto check or change test settings
3.3 Report:
Once the test is complete the test report interface will appear. For more information, please see
section 4.6.
57
Page 69
4.5 I/P Converter
In the I/P Converter mode only pressure and current functions are available.
◆Press on the top left corner of Current Output window to select the switch type: (0~24) or
Simulate mA.
1. Pressure Output
◆Correct output connection please see Figure 8
◆Do not exceed the pressure range of the calibrator
◆Please select Measure on Control Bar
2. Current Output:
◆Correct electrical connection and Current Output function, please see section 3.1.7
◆Jog:Press and select Jog, current output value could be adjusted by each digit in this
mode, see Figure 16
◆Scale:Press and select Scale for scale function, more detail please see section 3.14-3-②
3. New Test:
3.1 New Test Settings:①UUT (Unit Under Test):
58
Page 70
Subject
Valid Value
Comment
Input
Depends on UUT
UUT current measuring range
Output
Depends on UUT
UUT pressure output range
Transfer Function
Linear / Square Root
Transfer function type
②Test Method:
See section 3.1.3-3-③Auto Step
3.2 Test Execution:
After test settings are complete, press to execute the test:
1). Press to zero the pressure if necessary.
2). Press to run a manual test, or press to run an automatic test.
(1).In manual mode, run the next test point by press and the calibrator will automatically
record the electrical signal and pressure value.
(2).In automatic mode, the calibrator will run the test and record the electrical and pressure values
automatically without the need to advance the calibrator to the next set point.
◆Press to switch calibration interface
◆Pressto check or change test settings
3.3 Report:
Once the test is complete the test report interface will appear. For more information, please see
section 4.6.
Table 20 I/P Converter Test Settings
59
Page 71
4.6 Documenting
Subject
Valid Value
Comment
Name
Alphanumeric content (16 max length)
Test ID
Serial Number
Alphanumeric content (16 max length)
Test serial number
Location
Alphanumeric content (16 max length)
Test location
Notes
Alphanumeric content (16 max length)
Note
Accuracy
0.001~100
Test accuracy requirement
Range
N/A
Under test pressure gauge range
Once the test is complete the test report interface will appear.
◆Report interface can be switched between chart view and data view by pressing
◆The test results will be canceled and the test will restart by pressing
◆Report results are saved by pressing and entering the information below
1. Save to task:
Table 21 Test Information
60
Page 72
Subject
Valid Value
Comment
Operator
Alphanumeric content (16 max length)
Test operator ID
Execute Date
2000/01/01 ~ 2099/12/31
Execution date
Temperature
-20 ~ 100
Environment temperature during test ,
unit: ℃
Humidity
0 ~ 100
Environment humidity during test
2. Save Options:
Table 22 Test Save Options
◆Results can be recorded as As Found, As Left or Both (As Found and As Left)
61
Page 73
5. Setup
To enter the setup menu, please select Setup under Main Menu.
◆Any changes made in the Setup will become the default values after the calibrator is rebooted.
5.1 Control Settings
See section 3.1.3-1
5.2 24V Power
To turn on or turn off internal 24V power supply.
◆For safety and operation convenience, 24V power will be turned on or off automatically depending
on the calibration operation.
62
Page 74
Subject
Valid Value
Comment
WLAN Settings
WLAN
On / Off
Enable or disable Wi-Fi communication function
SSID
Depends on network environment
(only available when WLAN is on)
Select Wi-Fi router
Network Port
N/A
760 calibrator network port, read only
MAC
N/A
760 calibrator MAC address, read only
Advanced
DHCP / Static
Network address acquisition mode
USB Settings
Host / Slave Mode
Host / Slave
USB communication mode:
Host: For firmware update
Slave (device):For communication with computer
5.3 Communication
Communication settings:
Table 23 Communication Settings
63
Page 75
Subject
Valid Value
Comment
Correction Type
Auto / Fixed
Head pressure correction value calculation type
Correction Type: Fixed
Correction Value
Depends on internal
pressure module
Fixed correction value
Correction Type: Auto
Correction Value
N/A
Calculated value, read only
Unit
SI / BS
Unit standard:
SI: Metric unit
BS: Imperial unit
Density
Air / Nitrogen / Custom
Density of gas in used
Height
-1000 ~ 1000 cm (4 digit
after decimal point)
Altitude difference between UUT pressure inlet port and
760 calibrator pressure outlet port
Gravity
9 ~ 10 m/s2 (4 digit after
decimal point)
The gravity acceleration at the location of 760 calibrator
Temperature
0 ~ 100℃(2 digit after
decimal point)
Temperature at the location of 760 calibrator
5.4 Head Correction
Head pressure correction can be enabled or disabled.
◆If head pressure correction is enabled:
Table 24 Head Pressure Correction
64
Page 76
NOTE: Default air and nitrogen density is 0°C @ 1 013,25 hPa. After inputting your height and
temperature the ADT760 will calculate the correction value.
5.5 Services
5.5.1 Calibration
To calibrate 760 calibrator system.
CAUTIONS:
◆The calibrator should be powered to warm up 30 minutes prior to starting the calibration
◆Current / voltage measurement, current output, and internal pressure module need to be calibrated periodically
◆Carefully read user manual and follow the instruction before 760 calibrator system calibration operation
◆Incorrect system calibration operation will decrease the accuracy of 760 calibrator, even shutting down
◆During system calibration operation, pressing will restore the factory calibration and change the calibration
date to "----/--/--"
◆A higher accuracy standard should be used to calibrate the ADT760 calibrator
◆The calibration operation password is: 123456
◆Once the calibration is completed and saved the new data will be immediately loaded and used. A new
calibration date will also be uploaded
◆Once a new calibration has been finished and saved any previous calibrations will be permanently deleted
except for the factory calibration
65
Page 77
5.5.1.1 Current / Voltage Measurement Calibration
1. Connection:
◆Correct electrical connection please see Figure 9
2. Calibration point setup:
Select the calibration point which needs to be changed, then input value
◆3 points for current / voltage measurement calibration
◆Default current measurement calibration points are -30 mA, 0 mA, 30 mA
Default voltage measurement calibration points are -30 V, 0 V, 30 V
◆Additel does not recommend changing calibration points.
3. Calibration Execution:
1). Press to start calibration process
2). Output current / voltage from standard to 760 calibrator and the standard output value needs to
be the same as the corresponding calibration set point
3). When Measured Value is stable, pressto record data and move to the next calibration
point
◆If the difference between Calibration Point and Measured Value is unacceptable, the Measured
Value will display red.
5.5.1.2 Current Output Calibration
1. Connection:
◆Correct electrical connection please see Figure 3-3.
2. Calibration point setup:
66
Page 78
Select the calibration point which needs to be changed, then input value
◆3 points for current output calibration
◆Default current output calibration points are 1mA, 10mA, 20mA
◆Additel does not recommend changing calibration points.
3. Calibration Execution:
1). Press to start calibration process
2). Output current from 760 calibrator to standard
3). When standard measuring value is stable, press correspondent Measured Value and input
measuring value displayed on the standard then select Enter button or to confirm and then
pressto record data and move to the next calibration point.
5.5.1.3 Internal Pressure Module Calibration
1. Output Connection:
Select Internal / External pressurizing;
◆For Internal pressurizing: Please connect 760 calibrator pressure outlet port to the standard
ensuring the standard is in measurement mode
◆For External pressurizing: Please connect 760 calibrator pressure outlet port with the standard,
ensuring the standard is in pressure generation mode
2. Calibration point setup:
Select the calibration point which needs to be changed, then input value
◆3 points for internal pressure module calibration
◆Default calibration points are the lower limit, zero point, and the upper limit of internal pressure
67
Page 79
module
◆Additel does not recommend changing calibration points
NOTE: The lower limit of some internal pressure modules may exceed the lower limit of 760
calibrator pump, please use External pressurizing to calibrate these internal pressure
modules.
3. Calibration Execution:
1). Press to start calibration process
2). Pressurization:
①Internal pressurizing:
Generate pressure using the 760 internal pump. When the pressure is stable press Reference
value and input the value and select the Enter key or to confirm, then press to record
data and move to the next calibration point.
②External pressurizing:
Output pressure from standard to 760 calibrator, and not the standard output value shall be the
same as correspondent calibration set point. When Measured Value is stable, pressto record
data and move to the next calibration point.
5.5.1.4 Auto Tune
This function is only used during when recommended by Additel Corporation to service the calibrator
5.5.1.5 Pressure Sensor
This function is for positive and negative pressure sensor check with atmosphere.
68
Page 80
5.5.1.6 Barometer (Only for -MA & -MA-DL)
This function is for barometer sensor calibration.
Figure 17 Barometer Sensor Calibration
1. Connection: Please connect 760 calibrator barometer sensor calibration port with standard
pressure outlet port.
2. Calibration point setup:
Select the calibration point which needs to be changed, then input value
◆2 points for current output calibration
◆Default calibration points are 60kPa and 110kPa
◆Additel does not recommend changing calibration points
3. Pressure Outlet: The output pressure from the standard needs to be the same as the
corresponding calibration set points. When the Measured Value is stable, enter the Reference
69
Page 81
value displayed on the standard and press the Enter key or to confirm, then pressto
record data and move to the next calibration point.
5.5.1.7 Touch Screen
To check the touch screen, please press the center of black cross displayed on the screen five times.
◆If pressed on the deviation from the center of black cross, the test will repeat until it is precisely
pressed five times on the center of the black cross.
5.5.2 Diagnosis
This menu provides diagnostic information from the main board, control board, and electrical board
which may help indicate if anything is working improperly.
◆If a problem is detected then please contact Additel for further assistance
◆More information:
1. Password: 123456
5.5.3 System Log
Lists all abnormal information. Powering off or rebooting will not clear the information. The information
can be cleared by an update, please see section 5.5.4, or factory reset, please see section 5.5.5.
70
Page 82
5.5.4 Maintenance
This menu is used to depressurize the unit when switching internal pressure models.
◆Press to start.
5.5.5 Factory Reset
Resets all data to factory default settings.
◆Password:123456
5.5.6 Updates
To update 760 calibrator firmware, please:
1). Copy update file into an USB root directory
2). Insert USB into 760 calibrator USB port
3). Select Updates by USB on calibrator
4). Wait for the update to complete in which a notification will be given
◆Password:123456
◆The USB has to be in FAT16 or FAT32 type
◆The calibrator will switch USB port to Host type automatically
71
Page 83
Subject
Valid Value
Comment
Time
0:00 ~ 23:59
Time
Date
2000-1-1 ~ 2099-12-31
Date
Date Format
YYYY-MM-DD / MM-DD-YYYY / DD-MM-YYYY
Date format
Date Separator
-, /, .
Date format separator
5.6 Sound
To enable or disable Touch Beep and Over range alarm beep.
5.7 Personalization
5.7.1 Date and Time
Table 25 Date and Time Settings
5.7.2 Language
The calibrator is equipped with a multi-language user interface. Use this menu to change from the
offered languages.
5.7.3 Theme
Switch to Light Theme or Dark Theme.
72
Page 84
5.7.4 Safety Data
Subject
Valid Value
Comment
Brightness
1~100%
Adjusting this setting may impact the battery life.
Battery Info
N/A
Battery Information: remaining capacity, full capacity,
temperature, voltage, cell voltage I & II, Charging /
Discharging Current
Auto Backlight Off
Never, 15 sec, 30 sec, 1
min, 5 min, 15 min, 30 min
Touch the screen to turn the backlight on
Auto Off
Never, 5 min, 15 min, 30
min, 1 hour, 2 hour
Auto power off after backlight is turned off. There will
be a 30s countdown
Switch ON/OFF a warning before data deletion.
5.8 Power Management
Table 26 Power Management
73
Page 85
5.9 System Information
1. Main:
Main information of 760 calibrator: including serial number, application, system version, Wi-Fi, BLE
(Bluetooth), control board, electric board information
2. Battery:
Battery information: including serial number, version, manufacture date information.
3. Internal module:
Internal pressure module information: including range, pressure type, accuracy, serial number,
version, manufacture date information.
74
Page 86
6. Task (Only available on 760-X-DL)
The Additel 760 calibrator provides task functionality for automated calibration while storing and
archiving the calibration data. To use the task function, please select Task under Main Menu
6.1 New Task
Press and select UUT (Unit Under Test) type to enter a New Task interface and add UUT
information.
75
Page 87
Subject
Valid Value
Comment
Name
Enter an alphanumeric content (16 max length)
UUT ID
Serial Number
Enter an alphanumeric content (16 max length)
UUT serial number
Location
Enter an alphanumeric content (16 max length)
Task location
Note
Enter an alphanumeric content (16 max length)
Task note
Pressure Type
Gauge / Absolute / Diff.
UUT pressure type
Range
Depends on UUT
UUT pressure measuring range
Accuracy
0.06%, 0.1%, 0.16%, 0.25%, 0.4%, 0.6%, 1%, 1.6%,
2.5%, 4%,Custom
UUT accuracy
Resolution
0.0001~999999999999
The pressure difference represented
by nearest two scale on UUT.
6.1.1 Dial Pressure Gauge
UUT (Unit Under Test) information:
Table 27 Dial Pressure Gauge Information Settings
Press to complete UUT information set up and the calibrator will switch back to the task list. For
further instruction, please see section 6.2.1.
76
Page 88
6.1.2 Digital Pressure Gauge
Please see section 6.1.1.
Press to complete UUT information set up and the calibrator will switch back to the task list. For
further instruction, please see section 6.2.2.
77
Page 89
Subject
Valid Value
Comment
Name
Enter an alphanumeric content (16 max length)
UUT ID
Serial Number
Enter an alphanumeric content (16 max length)
UUT serial number
Location
Enter an alphanumeric content (16 max length)
Task location
Note
Enter an alphanumeric content (16 max length)
Task note
Pressure Type
Gauge / Absolute / Diff.
UUT pressure type
Input
Depends on UUT
UUT pressure measuring range
Output
Analog:
4~20 mA / 0~10 mA / 0~20 mA / 1~5 V / 0~5 V / 0~10 V /
Press to complete UUT information set up and the calibrator will switch back to the task list. For
further instruction, please see section 6.2.5.
80
Page 92
Subject
Valid Value
Comment
Set Points
2~17;
Set point values
depend on UUT
range
Set point values will be calculated automatically.
◆To edit the number of set points: press/ or press on set point number at the
bottom of the screen and input value through soft keyboard
◆To edit value: select and input value through soft keyboard
Repeat
1 / 2 / 3
Task cycle times
Stroke
Round Trip / One
Way
Task stroke type
Readout
format
Value / Scale
1.Value:Input dial pressure gauge indicated value
2.Scale: Input scale number indicated on dial pressure gauge. For example,
Resolution of UUT is 0.1 kPa, input 2 as scale number means UUT measurement
value is 0.2 kPa
Tapping
On / Off
To enable or disable tapping after reach each set point during the test
6.2 Task Execution
Please select UUT(Unit Under Test) from the task list, and check UUT information. Next select to
view and or edit the task information and execute the test.
◆To edit UUT information, please select UUT from the task list then select .
6.2.1 Dial Pressure Gauge
1. Task Execution Setup (Run Information):
Table 31 Dial Pressure Gauge Task Settings
81
Page 93
2. Task Execution:
Press to execute the task:
1). Press to zero pressure module if necessary
2). Press to start the test;
◆The calibrator start pressurizing to the first test point
◆Press to switch calibration interface
3). Standard test steps:
(1). After pressure is stable , enter the UUT pressure value on the top right UUT window or press
to switch the calibration interface.
(2). Press the Enter key or for the next test point in which the calibrator will pressurize again.
Note: if the tapping function is enabled, you should tap the dial gauge at each set point.
(3). After the next pressure is stable then enter the UUT value in the UUT window or press
to switch the calibration interface.
(4). Repeat step (2) and (3) until all test points are completed.
3. Report:
Once test is done, the reporting interface will appear. For more information please see section 6.4.
82
Page 94
Subject
Valid Value
Introduction
Set Points
2~17;
Set point values depend on
UUT range
Set point values will be calculated automatically.
◆To edit the number of set points press: press/or press on Set
point number at the bottom of the screen and input value
◆To edit value: select and input value
Repeat
1 / 2 / 3
Task cycle times
Stroke
Round Trip / One Way
Task stroke type
6.2.2 Digital Pressure Gauge
1. Task Execution Setup (Run Information):
Table 32 Digital Pressure Gauge Task Settings
2. Task Execution:
Press to execute task:
1). Press to zero pressure module if necessary.
2). Press to start the test;
◆The calibrator start pressurization to the first test point
◆Press to switch calibration interface
3). Standard test steps:
(1). After the pressure is stable at the first test point then enter the UUT pressure value on the top
right UUT window or press to switch the calibration interface.
(2). Press the Enter key or for the next test point in which the calibrator will begin pressurizing
again.
83
Page 95
(3). After the pressure is stable enter the UUT value.
(4). Repeat step 2 and 3 until all test points are completed.
3. Report:
Once a test is complete, the test report interface will appear. For more information please see section
6.4.
84
Page 96
Subject
Valid Value
Comment
Set Point
2~17;
Set point values depend
on UUT range
Set point values will be calculated automatically.