Warningsandcautions
949
Hydraulic Ultra-High Pressure Test Pump User's Manual
[Version number:1306V01]
Operate the pump in the rated pressure range ,500 psi ,
DO NOT over the safety pressure 40,000 psi .
(
When the system pressure is greater than ,
the re-pressurization shut off valve(#7)
close p and then
(
37 )
)
10,000 psi you must
pressurize the system.
Always tighten the gauges using two wrenches .
Close all valves and handles of the pump, and tighten the plugs
on the connectors during transportation.
Always keep the reservoir cover vent valve #4 open during
( )
operation.
Do not over tighten the valves, connectors and handles to avoid
any damage.
Change media immediately if it is contaminated.
Keep media level between 1/4 and 3/4 of the liquid reservoir filled.
Keep the threads clean and lubricious, and remove any dirt on
threads.
Used by Trained Personnel only.
Any security problems or damages caused by incorrect operations
are beyond Additel's responsibility.
Specification
Pressure range:
Remark:.If local atmosphere pressure is 1 bar, the vacuum can reach to 0 85 bar;
If local atmosphere pressure is P, the vacuum can reach to bar.( P 85% )
Temperature:
Adjusting fineness:
Safety pressure:
Pressure media:
Height: 10.43'' ( 265 mm )
Size:
Base: 20.71'' ( 526 mm ) x 9.65'' ( 245 mm )
Weight:
35.5 lb ( 16 kg )
12.5 psi ( 0.85 bar ) vacuum to
37,500 psi ( 2500 bar ) positive pressure
(
5~50)℃
0.015
psi ( 1 mbar )
< 40,000 psi
Diethylhexyl Sebacate
Views and Hydraulic Schematic
1 32
9
4
5
7
8
Troubleshooting
Problems
Hard to pressurize
through turning
Pre-pressurization
handle (#9)
Hard to pressurize through
turning high-pressure
pressurization handle (#8)
Hard to reach high vacuum
Hard to pressurize
large-cavity DUT
Pressure gauges do not
reach to zero
Hard to turn the
valves or handles
Reservoir shut off valve (#1) is not closed
Pre-pressurization shut off valve (#7) is closed
The media is not enough
Too much air in the pump
The pre-pressurization shut off valve (#7) is not closed completely
Reference gauge or devices under test (DUTs) are not connected tightly
The connector of DUT is not matched to connector
Too much air in the pump
#4 valveis not open Open the#4 valve
Large-cavity of DUT, and pressurize the system too slow
#4 valveis not open Open the#4 valve
Too tight in last operation
Hard to turn pre-pressurization handle (#9) at high pressure
Lack of lubrication on threads
1- Reservoir shut off valve
6
2- Over-flow reservoir
3-
Positional Autoclave F-250-C,
9/16"-18UNF female
or customized female connector
()
4- Reservoir cover vent valve
5- Reservoir
6- Liquid drain valve
7- Pre-pressurization shut off valve
8- High-pressure Pressurization and
fine pressure adjustment handle
9- Pre-pressurization handle
Causes
Hydraulic Schematic
5
1
9
8
Solutions
Close shut off valve (#1)
Open pre-pressurization shut off valve (#7)
Fill more media, and keep media level between 1/4 and 3/4 of the liquid reservoir filled
Turn the handle anticlockwise to vacuum the pump, open reservoir shut off valve (#1) to have air released
Close pre-pressurization shut off valve (#7)
Check connector, replace the PTFE type, re-tight using a wrench
Use proper adapter
Close pre-pressurization shut off valve (#7), open reservoir shut off valve (#1), turn pre-pressurization handle
(#9) clockwise fully, close reservoir shut off valve (#1), and then open pre-pressurization shut off valve (#7)
Step 1: Turn pre-pressurization handle (#9) clockwise fully, close pre-pressurization shut off valve (#7),
open reservoir shut off valve (#1).
Step 2: Turn pre-pressurization handle (#9) anticlockwise fully, close reservoir shut off valve (#1).
Step 3: Open pre-pressurization shut off valve (#7), pressurize the system.
Step 4: Repeat step 1 to 3.
Do not close shut off valves and handles too tightly
It is normal. Please use high-pressure pressurization handle to increase/reduce the pressure.
Lubricate the threads
7
![](/html/f5/f5cb/f5cbdd466c84cfedca3dc76149a01364afb11625b521666a4fd2a87b7629f421/bg2.png)
Basic Operation
⑤open
DIGITALPRESSURECALIBRATOR
Units
02550751000255075100
Units
③close
barbar
A
DIGITALPRESSURECALIBRATOR
b
bar
a
r
A
0255075100
④a few turns
Note: Gauge positions are exchangeable. If only one device under test (DUT) is connected, the third connector on pump should be closed by a plug provided.①
64 96
PressureGaugePressureGauge
Bar
32
128
160
② Connect the gauges on pump by fingers, tighten the connectors using two wrenches.
64 96
PressureGaugePressureGauge
Bar
32
128
160
pressure switch
③
calibration in a narrow
pressure range
64 96
PressureGaugePressureGauge
Bar
32
128
160
32
④high-pressure
Pressurizing
64 96
PressureGaugePressureGauge
Degas
①Always keep valve(#4)
open during operation
B
64 96
PressureGaugePressureGauge
Bar
32
128
160
②open
①close
dial gauge
Pressurizing procedure Depressurizing procedure
DIGITALPRESSURECALIBRATOR
Units
0255075100
②Pre-pressurizing
Note: Skip step for dial gauge
b
bar
a
r
A
Bar
128
160
SW V mA COM 24VSW V mA COM 24V
C
③close
SW V mA COM 24VSW V mA COM 24V
pressure transmitter
DIGITALPRESSURECALIBRATOR
Units
b
bar
a
r
A
0255075100025507510
0
④open
③Depressurizing
Connection
flexible hose
64 96
PressureGaugePressureGauge
Bar
32
128
160
①Depressurizing
A
D
64 96
PressureGaugePressureGauge
Bar
32
128
160
②open
Open to atmosphere (Venting)
and Zeroing
DIGITALPRESSURECALIBRATOR
Units
b
bar
a
r
A
④open
②close
0255075100
③Fully turn
0.6 0.4
PressureGauge
Bar
0.8
0.2
0
-1
⑤Fully turn
Maintenance
Fill media Drain and Clean
Note: Please degas after filling media liquid,
in case hard to generate pressure.
①Open reservoir
cover vent valve (#4)
E F
DIGITALPRESSURECALIBRATOR
Units
0.6 0.4
PressureGauge
Bar
0.8
0.2
0
-1
①close
Vacuum procedure Back to atmosphere
b
bar
a
r
A
0255075100
②Vacuum
0.6 0.4
PressureGauge
Bar
0.8
0.2
0
-1
①open
A
③ Open reservoir cover
vent valve (#4)
④ Unscrew with
②Fill media
Open drain valve (#6)
①
0.6 0.4
PressureGauge
Bar
0.8
0.2
0
-1
hex wrench
⑤Remove the
reservoir(#5)
B
G
DIGITALPRESSURECALIBRATOR
Units
b
bar
a
r
A
0255075100
②open
①back to atmosphere
Remark:
A Additel has made a concerted effort to provide
complete and current information for the proper
use of the equipment. The product specifications
and other information contained this manual are
subject to change without notice.
B:: Above pictures are just for reference.