Additel 917 Instruction Manual

Warningsand cautions
Low pressure gauges can very easily be over pressured if not careful. Please take caution when applying pressures. Do not exceed the safet y limit of 1500 psi (100 bar) for 917 or 2000 psi (140 bar) for 918. Do not switch between pressure and vacuum mode under pressure. Open the vent valve and isolation valve before switching between positive to vacuum pressures. Over tightening connectors may cause damage. Light hand tightening is sufficient. Store the pump in a dry non-corrosive environment. Additel is not responsible for problems or damage resulting from improper use or operation of the equipment. All moisture and contaminants should be cleaned out of the liquid isolator before creating a vacuum. This is accomplished by venting the pump when at high pressure. For the least risk in contamination of the reference gauge, the DUT should be installed to the pressure port which is closest to the vent valve.
Specification
917/918
Pneumatic Pressure Test Pump User Manual's
[Version number:1407V02]
Configuration & Air routine
31
2
5
67
4
8
Under Vacuu
1 - Hand pump 2 - Transport handle 3 - Quick connector 4 - Vent valve
(clockwise to close/ counterclockwise to open)
5 - P/V selector lock 6 - Pressure/vacuum (P/V) selector
Positive pressure: pull it out; vacuum: push it in
DO NOT switch under pressure or vacuum to
!
avoid damage)
7 - Isolation valve (Bypass valve)
(Isolated the calibration volume from the hand pump, piston, and check valve)
8 - Fine adjust handle
clockwise to increase pressure)
Pressure range:
Temperature: Humidity:
Adjustment resolution
Maximum pressure:
Pressure media:
Height: 5.5'' ( 140 mm );
Size:
917: 95% vacuum to 1000 psi ( 70 bar )
8
91 : 95% vacuum to 1500 psi ( 100 bar )
0~50
)℃
< 85%RH
10 Pa ( 0.1 mbar )
:
917: 1500 100 918
psi ( bar )
: 2000 psi ( 140 bar )
Air
Base: 12.4'' ( 315 mm ) x 7.8'' ( 198 mm )
Weight:
RemarkIf local atmosphere pressure is 1 bar, the vacuum can reach to -0.95 bar;
5.7 lb ( 2.6 kg )
If local atmosphere pressure is P, the vacuum can reach to - 0.95P.
Air routine
7
3
6
1
4
8
Troubleshooting
Problem
Hand pump will not generate pressure
It is difficult to increase pressure
Hard to use the fine adjust
It is difficult to turn the quick connector
Cause
The isolation valve is not open
A. Vent valve is not closed.
B. The O-ring seal is loose or damaged.
C. P/V selector is in the wrong position
A. The isolation valve is not closed
B. The gauges are not tightened.
C. The O-ring seal is damaged.
D. The thread surface is not smooth.
E. The connector type mismatched to the gauge pressure port.
F. Dust or contaminants are blocking the pressure flow
A. Too much force was previously applied.
B. The threads have no lubrication..
.
.
Solution
Open the isolation valve
Close the vent valve.
Replace O-ring seal
Pressure: pull the P/V selector out; Vacuum: push the P/V selector in.
Close the isolation valve
Tighten the reference gauge or the gauge under test
Replace the O-ring seal.
Use Teflon tape on the thread and turn it in tight
Use the correct adapter
Pressurize the pump several times and then suddenly vent which will push any
contaminants out.
Secure the quick connect valves with less force.
Apply lubrication to the threads..
.
.
.
.
Basic Operation
open
DIGITALPRESSURE CALIBRATOR
Units
DIGITALPRESSURE CALIBRATOR
Units
A
0255075100
A
0255075100
Pressure GaugePressure Gauge
open
pressure position is shown. push in for vacuum.
Must vent before switching between
!
vacuum and positive pressure.
Vent
Zeroing should be operated under this condition only
B
.
Pressure GaugePressure Gauge
k
dial gauge
push down to
pre pressurize
-
to desired value or until hard to push down
open
SW V mA COM 24V
SW V mA COM 24VSW V mA COM 24V
pressure switch pressure transmitter
DIGITALPRESSURE CALIBRATOR
Units
0255075100
A
set it to the mid position of its travel
Pre-pressure
Pressure Gauge
close,do not
use too much
C
force
close,do not
use too much force
Connection
flexible hose
Pressurizing process and fine adjustment
DIGITALPRESSURE CALIBRATOR
Units
A
0255075100
Pressure Gauge
pressurize to
higher pressure by turning the handle and make fine adjustment
A
D
Decrease pressure process
DIGITALPRESSURE CALIBRATOR
Units
A
0255075100
Pressure G
Pressure Gauge
auge
open it slowly
to get the desired value, and close!
fine adjustment
Decrease vacuum process
DIGITALPRESSURE CALIBRATOR
Units
A
0255075100
close
0 8.00 8.0
-0.1
fine adjustmentt
0.06
Vacuum
Vacuum Gauge
0.04
Gauge
0.02
0
open it slowly to get the desired value, and close!
E
H
operate it to get the desired value
open
Vacuum process
DIGITALPRESSURE CALIBRATOR
Units
A
0255075100
push in,select the vacuum position
m②
ake sure to complete the step
!
before operating step .
F
DIGITALPRESSURE CALIBRATOR
Units
close,do not
use too much force
0255075100
0.06
0.04
Vacuum
Vacuum Gauge
Gauge
0 8.00 8.0
0.02
-0.1 0
close
open
Vacuum and
fine adjustment
A
0.06
Vacuum
Vacuum Gauge
0 8.00 8.0
-0.1
fine adjustment
0.04
G
auge
0.02
0
Remarks:
A Additel has made a concerted effort to provide complete and current
information for the proper use of this equipment. The product specifications and other information contained in this manual are subject to change without notice.
B:: Above pictures are just for reference.
Additel Corporation, USA
www.additel.com