Warnings and cautions
Do not exceed the safety pressure limit(60 psi).
Be careful of any problem by some pressure media against the
exploisive or corrupsive conditon.
Do not tighten connectors over to avoid any damage.
Store the pump in the dry and non-corruptive environment.
Any safety problems or damages caused by incorrect operation,
is beyond Additel's responsibility.
Specification
Low Pressure Test Pump User's Manual
Configuration & Air routine
ADT901
[Version number:1103V10]
Pressure range: 6 psi ( 0.4 bar ) vacuum to
6 psi ( 0.4 bar ) pressure
Temperature: (0 ~ 50)℃
Humidity: < 95%RH
Adjusting fineness: 0.01Pa ( 0.0001 mbar )
Safety pressure: < 60 psi
Pressure media: Air
Size: Height: 5.71'' ( 145 mm )
Base: 9.65'' ( 245 mm ) x 6.50'' ( 165 mm )
Weight: 3.5 lb ( 1.6 kg )
1
2
3 4
Troubleshooting
Problems
Hard to coarsely adjust
1-1/4 NPT quick connector
2-Shut off valve
(clockwise to close/ anticlockwise to open)
5
3-Fine adjust handle
(clockwise to increase pressure)
4-1/4 NPT quick connector
5-Coarse adjustment handle
(clockwise to increase pressure)
Causes
A. The isolating valve is not closed.
B. The sealing ring is broken or loosen.
Air routine
2
3
5
Solutions
Close the isolating valve.
Replace sealing ring.
Hard to fine adjust
Hard to turn the fittings
A. The gauges are not tightened.
B. The sealing ring is aged or frayed.
C. The thread surface is not smooth.
D. The connector type is unmatched.
A. Too tightly turned last time!
B. The assembly is not fitted.
C. The threads have no lubrication.
Tighten the reference gauge or the gauge under test.
Replace the sealing ring.
Put in a Teflon seals and turn it tightly.
Use the right and suitable adapter.
Do not apply much force to close the fittings and valves.
New assembly needs time to abrode.
Lubricate the thread.
![](/html/4f/4fbd/4fbd6850d3d66ec8882d74b486615694bdc5486b7c555ec749148ff8a5dd348c/bg2.png)
Basic Operation
Connection
D
I
GTA
I
L
P
R
E
S
S
U
R
E
CALIB
RAT
OR
Units
k
SW V mA COM 24VSW V mA COM 24V
flexible hose
Pressure GaugePressure Gauge
kk
dial gauge
SW V mA COM 24VSW V mA COM 24V
pressure switch
pressure transmitter
A
TAL
GI
I
D
nits
U
①open
Open to atmosphere
Pressure preparation
Zeroing should be
O
T R
A
BR
A I
C L
E
UR
SS
E
R
P
00
res
res
③set it to the mid position of its travel
operated under
this condition only.
eu
e
GaP
GaP
re
re
u
s ug
s ug
kk
00
②turn it to the
maximum
degree
B
Increase pressure process
OR
T
R
B A
I
AL
E
SUR
PRE
L
A S C
T
I
IG
D
s
it
n
U
③fine adjust
①Close
00
e
e
g
g
a
a
G
G
ure
ure
s
s
s u
s u
e
e
r
r
P
P
kk
00
00
C
②operate it to
approach the
desired
value
Decrease pressure process
T
BRA
L OR
A I
C
E
U
S R
S
A PRE
T
I L
IG
D
ts
ni
U
②fine adjust
00
e
e
ug
ug
Ga
Ga
re
re
su
su
s
s
re
re
P
P
kk
00
00
D
①operate it to
approach the
desired
value
P
L
TA
GI
DI
iU
nts
①open
Prepare for vacuum
R
TO
BRA
I
CAL
E
UR
SS
E
R
1010
15 15
u ecu
u ecu
G
G
m a gVa u
m a g
Va u
k
k
2020
25
25
-
-
③set it to the mid position of its travel
55
00
②turn it to the
maximum
degree
E
RES
P
GT
I I AL
D
t
U i sn
①close
R
O
LIB AT
A
R
E C
U
S R
③fine adjust
Vacuum process
1010
1515
G u eV cu
G u eV cu
m a g
m a g
a u
a u
kk
2020
-25-25
55
00
②operate it to
approach the
desired
value
F
Decrease vacuum process
O
AT
LIBR
A
R
E C
U
S R
RES
P
AL
IGIT
D
ts
n
U i
②fine adjust
1010
515
1
g
g
au e
au e
G
G
m
m
cu
cu
Va u
Va u
kk
20
20
-25-25
55
00
①operate it to
approach the
desired
value
G
Remark: A: Additel has made a concerted effort to provide complete and current information for the proper use of the equipment. The product
specifications and other information contained this manual are subject to change without notice.
B: Above pictures are just for reference.