Additel 672 Datasheet

Additel 672 Digital Pressure Calibrators
Pressure ranges to 40,000 psi (2,800 bar)
HART Communication capability
Measure mA or V, and with 24V loop power
Easy-to-use, inexpensive pressure calibrator with uncertainty better than 0.025%FS
Pressure / Process Calibration EquipmentAdditel Datasheet 2014 / 2015
OVERVIEW
With advanced microprocessor technology and state-of-the-art silicon pressure sensors, the 672 series precision pressure calibrators provide a pressure calibration solution for gauges, transmitters, and switches over a wide pressure range. The 672 is a very portable device and can measure pressure precisely with a built-in pressure sensor, as well as read the current or mV produced by a transducer. It can even supply an excitation voltage to power sensors or transmitters during calibration. In order to reach 0.025% F. S. accuracy up to 10,000 psi (700 bar) and 0.1% F. S. accuracy up to 40,000 psi (2,800 bar), every silicon pressure sensor in our gauges has been specially aged, tested and screened before assembly. The 672 series precision pressure calibrators are unmatched in performance and reliability.
Gauge pressure Differential pressure
FEATURES
Pressure ranges to 40,000 psi (2800 bar)
Measure mA with 0.01% RD + 1.5 μA accuracy
Measure V with 0.01% RD + 1.5 mV accuracy
Power transmitters during test using 24V loop supply
Pressure switch test
HART Communication capability
Advanced temperature compensation
Dual readout
Min/Max/Hold to capture changing measurements
Data logging
Large, easy to read display with 6-digit resolution
Backlit display
Rechargeable battery or AC adapter
NIST traceable calibration with data (Included)
SPECIFICATIONS
Accuracy
ADT672-02: 0.025% of full scale ADT672-05: 0.05% of full scale >15,000 (1,000 bar): 0.1% FS
Gauge Types
Gauge pressure Compound Pressure Absolute pressure Differential pressure
Display
Description: Dual-line 6 full digit FSTN LCD with LED Backlight Display rate: 3.5 readings per second (Default setting) Numeral display height: 16.5mm (0.65")
Pressure Units
Pa, kPa, MPa, psi, bar, mbar, inH mmH
O@4°C, inHg@0°C, mmHg@0°C, kgf/cm
2
Environmental
Compensated Temperature: 14°F to 122°F (-10°C to 50°C) Operating Temperature: 14°F to 122°F (-10°C to 50°C) Storage Temperature: -4°F to 158°F (-20°C to 70°C) Humidity: <95%
O@4°C,
2
2
Corporate Headquarters
22865 Savi Ranch Parkway Ste F Yorba Linda, CA 92887, USA
Phone: 714-998-6899 Fax: 714-998-6999 Email: sales@additel.com
Salt Lake City Office
802 Bamberger Dr, Suite C American Fork, UT 84003, USA
Phone: 801-756-5570 Fax: 714-998-6999 Email: sales@additel.com
SPECIFICATIONS
Pressure Ranges
Gauge Pressure
P/N
Pressure / Process Calibration EquipmentAdditel Datasheet 2014 / 2015
V15 -15 -1.0 G 0.025 (0.05)
Pressure Range
(psi) (bar)
GP2 2 0.16 G 0.05
GP5 5 0.35 G 0.025 (0.05)
GP10 10 0.7 G 0.025 (0.05)
GP15 15 1.0 G, L
GP30 30 2.0 G, L
GP50 50 3.5 G, L 0.025 (0.05)
GP100 100 7.0 G, L 0.025 (0.05)
GP300 300 20 G, L 0.025 (0.05)
GP500 500 35 G, L 0.025 (0.05)
GP600 600 40 G, L 0.025 (0.05)
GP1K 1,000 70 G, L 0.025 (0.05)
GP2K 2,000 140 G, L 0.025 (0.05)
GP3K 3,000 200 G, L 0.025 (0.05)
GP5K 5,000 350 G, L 0.025 (0.05)
GP10K 10,000 700 G, L 0.025 (0.05)
GP15K 15,000 1,000 G, L 0.05(0.1)
GP25K 25,000 1,600 G, L 0.1 1.5×
GP30K 30,000 2,000 G, L 0.1 1.5×
GP36K 36,000 2,500 G, L 0.1 1.5×
GP40K 40,000 2,800 G, L 0.1 1.1×
Note: [1]. Sealed gauge pressure for above 1000 psi
[2]. G=Gas, L=Liquid
[3]. 0.025% FS for gas media only
Compound Pressure
P/N
Pressure Range
(psi) (bar)
CP2 ±2 ±0.16 G 0.05
CP5 ±5 ±0.35 G 0.025 (0.05)
CP10 ±10 ±0.7 G 0.025 (0.05)
CP15 ±15 ±1 G 0.025 (0.05)
CP30 -15 to 30 -1 to 2 G 0.025 (0.05)
CP100 -15 to 100 -1 to 7 G, L 0.025 (0.05)
CP300 -15 to 300 -1 to 20 G, L 0.025 (0.05)
[1]
Media
[2]
Accuracy(%FS)
Burst
Pressure
Absolute Pressure
P/N
Pressure Range
(psi) (bar)
Media
[2]
Accuracy(%FS)
Burst
Pressure
AP5 5 0.35 G 0.1
AP10 10 0.7 G 0.1
AP15 15 1.0 G 0.1
AP30 30 2.0 G 0.1
[3]
0.025 (0.05)
[3]
0.025 (0.05)
AP50 50 3.5 G 0.1
AP100 100 7.0 G , L 0.1
AP300 300 20 G , L 0.1
AP500 500 35 G , L 0.1
AP1K 1,000 70 G , L 0.1
AP3K 3,000 200 G , L 0.1
AP5K 5,000 350 G , L 0.1
Differential Pressure
P/N
Pressure Range
(inH
O) (mbar)
2
Media
Accuracy
(%FS)
Burst
Pressure
Static
Pressure
Range
DP1 ±1 ±2.5 G 0.05 100× ±10 psi
DP2 ±2 ±5.0 G 0.05 100× ±10 psi
DP5 ±5 ±10 G 0.05 50× ±10 psi
DP10 ±10 ±25 G 0.05 20× ±10 psi
DP20 ±20 ±50 G 0.05 20× ±10 psi
DP30 ±30 ±75 G 0.05 20× ±10 psi
DP50 ±50 ±160 G 0.05 ±10 psi
DP150 ±150 ±350 G 0.025 50 psi
DP300 ±300 ±700 G 0.025 50 psi
* One year accuracy (including yearly stability) except DP1 and DP2 range
Media Accuracy(%FS)
Burst
Pressure
which is 0.05%FS calibration accuracy and 0.05%FS yearly stability.
Electrical Measurement Accuracy
Range Accuracy
Voltage DC ±30.0000V ± (0.01%RD + 1.5 mV)
Current DC ±30.0000mA ± (0.01%RD + 1.5 μA)
DC 24V 24V±0.5V max:50mA Protect at: 120mA
Switch
[1]
Status OPEN/CLOSED
Note: [1] 1V~12V if switch has detective voltage