Additel 227 Data sheet

Page 1
Additel 227, 227Ex Documenting Multifunction Process Calibrator
Sourcing, Simulating and Measuring Pressure, Temperature and Electrical Signals
Built-in Full Hart Communicator (ADT227-HART)
Built-in Barometer
Intrinsically Safe Models Available (Ex)
Large Smartphone Like Touchscreen User Experience
USB Type-C and Bluetooth Communications
Pressure / Process Calibration Equipment
IP67 Rated
High Voltage Measurement Capability (300V AC)
True RMS Voltage Meter Capability
Dual Channel Pressure Module Ports
High Static Differential Pressure Measurement 0.002% FS
ISO 17025-accredited Calibration w/data Included
OVERVIEW
Additel’s new Multi-functional Documenting Process Calibrator series takes portability, functionality, and accuracy to a whole new level and packages it with an intuitive and easy to use color touchscreen display. This series includes an advanced documenting pressure calibrator (ADT227) and an advanced documentation process calibrator with a built-
in HART communicator (ADT227-HART). Additionally, each calibrator has an ATEX certied intrinsically safe option (ADT227Ex) allowing you to perform calibration in the harshest of environments. We’re condent these new tools will not
only meet your calibration requirements but will make metrology simple for you!
Features
Easy-to-use Cellphone Like Interface
The ADT227 series brings an all new user interface to the world of process calibrators. With a menu driven interface and a small size/weight, the ADT227 is the industry's smallest advanced multifunctional process calibrator with an intrinsically safe version to boot (ADT227Ex). It adopts advanced human hand
engineering design for the most convenient eld handheld process calibrator
available.
The ADT227 has been developed with a powerful embedded operating system which solves common problems of other designs including slow response, cumbersome key operation, high power consumption and overall slow processing.
Accuracy
01
Rev # 20220825
Additel's new and improved ADT227 series provides much improved accuracies including an electrical accuracy of 0.005% RD + 0.005% FS, high-static differential pressure mode accuracy to 0.002% FS and across the board improvements in temperature measurement accuracies.
Corporate Headquarters
2900 Saturn St #B Brea, CA 92821, USA
Salt Lake City Office
1364 West State Rd. Suite 101 Pleasant Grove, UT 84062, USA
Page 2
Features
Thermocouple Measurement Performance
The ADT227 series delivers highly improved thermocouple measurement capabilities by vastly improving the cold junction compensation(CJC)
specications and a much improved stabilization time.
Portable and Robust
The demands of remote calibration work can be challenging. The ADT227series is lightweight and highly portable and utilizes an advanced color LCD screen to help ensure you can easily see, even in the (Ex) intrinsically safe versions.
Pressure / Process Calibration Equipment
The Additel 227Ex series calibrators have passed the most stringent testing by
certied organizations to acquire intrinsically safe certicates, ATEX, IECEX, CSA
and UKCA. The explosion-proof grade (Ex ia IIC T4 Ga), can be widely used in
potentially explosive environments, such as oil and gas platforms, oil reneries,
chemical and petrochemical plants, pharmaceutical industries, energy and gas processing industries.
Each intrinsically safe calibrator has an advance transective color LCD display
which has enhanced visibility when viewed in direct sunlight. No matter where your work takes you, these calibrators are up to the task.
All models in the ADT227 family have been designed with ruggedness and dependability in mind and meet IP67 standards with a 1-meter drop test, 4G vibration, xenon exposure and 130g steel ball drop testing of the display.
Other environmental conditions have also been considered, such as temperature and humidity. To combat these external elements, Additel has designed a unique internal circuit
design and process technology to allow for the utmost condence in your critical calibration
and measurement work.
Intrinsically Safe Option
Voltage Meter (RMS)
The Additel 227 non-Ex version is equipped with “true effective value” RMS measuring function, which can measure the RMS of various waveforms with no need to consider distortion or waveform parameters and other errors caused by various waveforms
02
Automated Tasks for Paperless Calibration Management
Additel 227 Series Calibrators come with a powerful documenting calibration task application which provides a turnkey solution for automation and paperless calibration management.
Tasks are easily created for temperature, pressure, ow and loop instruments. Up
to 10,000 documented tasks for ADT227 and up to 1,000 documented tasks for ADT227Ex can be stored in the extensive on-board memory. Many tasks, when executed, are fully automated in data collection and performance validation, such as pass/fail and hysteresis calculations. All information can be integrated into Additel’s ACal software for additional calibration management.
Full HART Communicaton ( For ADT227-HART only)
The built-in full HART communicator will work with most HART transmitters . The ADT227-HART contains an extensive DD library to meet the needs of your smart transmitter. Our DD library is updated on a regular basis and at no additional cost. The ADT227Ex-HART is integrated with the HART communication functions, permitting users to monitor, control, and calibrate HART instrumentations. It's an ideal device for calibrating, maintaining, and troubleshooting HART instrumentations
Rev # 20220825
Corporate Headquarters
2900 Saturn St #B Brea, CA 92821, USA
Salt Lake City Office
1364 West State Rd. Suite 101 Pleasant Grove, UT 84062, USA
Page 3
Features
Targeted application features
The onboard applications provide a useful selection of features including HART communicator, high static differential pressure mode, pressure leak test, safety valve test, analog transmitter calibration, unit converter, thermal calculator, and snapshots to name a few.
High static differential pressure mode uses two sensors, unique calculation technology to achieve a differential pressure measurement to 0.002% FS at high static pressures. The leak test will automatically calculate the pressure drop to determine a leak condition. The safety valve test is a specialized task which captures the exact pressure release point by taking 10 readings per second during a valve crack test.
Pressure / Process Calibration Equipment
The ADT227 calibrator can record pressure, temperature and electrical signals. Recorded values can be displayed numerically or graphically to identify trending. The ADT227 & ADT227-HART can store up to 500 results. each result can include up to 100,0000 recordings and each recording can record a maximum of 7 channel values. These results can easily be exported to Additel’s application software. Each log
session is easily congured at a set interval and provides a date and time stamp with
each reading.
In addition to all the great features mentioned above, the ADT227series is loaded with time saving features like our builtin pressure and temperature converter, thermal calculator, wiring diagram guide for assisting with electrical connections, a built-in diagnostic center including intelligent alarm messaging and a real time error report and comprehensive selftesting to help our customers get the very most out of their investment in Additel calibration tools.
You will nd this and much more as we continue to develop new apps at Additel.
Data Logger
( For ADT227 & ADT227-HART only)
Connectivity & Battery
Users can remotely connect mobile devices to the ADT227 via Bluetooth with an unobstructed distance up 20 meters. The included USB type-C comm port and cable provide a hard wired communication option as well as charging for the removeable Li-ion battery, which provides up to 35 hours of run time.
Time Saving Features
03
SPECIFICATIONS
Electrical Specification
Source Accuracy
Specifications
Voltage DC 0 to 15 V 0.25 mV 0.005%RDG + 0.75 mV 0 to 10.5 V 0.2 mV 0.01%RDG + 0.5 mV
Current DC 0 to 25 mA 0.5 uA 0.01%RDG + 1.2 μA 0 to 25 mA 0.5 uA 0.01%RDG + 1.2 μA
Resistance
Frequency
Voltage mV
(TC)
Pulse
Loop power (max 25mA)
Rev # 20220825
Range Resolution Accuracy Range Resolution Accuracy
0 to 400 Ω 10 mΩ 0.005%RDG + 20 mΩ 0 to 400 Ω 10 mΩ 0.01%RDG + 20 mΩ
0 to 4000 Ω 100 mΩ 0.01%RDG + 200 mΩ 0 to 4000 Ω 100 mΩ 0.01%RDG + 200 mΩ
0.01 to 5 Hz
5 to 50 Hz 0.002% RD + 0.0002 Hz 5 to 50 Hz 0.002% RD + 0.0002 Hz
50 to 500 Hz 0.002% RD + 0.002 Hz 50 to 500 Hz 0.002% RD + 0.002 Hz
500 to 5000 Hz 0.002% RD + 0.02 Hz 500 to 5000 Hz 0.002% RD + 0.02 Hz
5000 to 50000 Hz 0.002% RD + 0.2 Hz 5000 to 50000 Hz 0.002% RD + 0.2 Hz
-10 to 75 mV 1.5 uV 0.008%RDG + 3.0 μV -10 to 75 mV 1.5 uV 0.01%RDG + 3.0 μV
0 to 9999999 1 N/A 0 to 9999999 1 N/A
Optional rising edge and falling edge, minimum threshold voltage: 2.5V
24 V N/A ±1 V 20 V N/A ± 10%
ADT227 ADT227Ex
0.002% RD + 0.00002 Hz 0.01 to 5 Hz
Auto range,
6-digit
Corporate Headquarters
2900 Saturn St #B Brea, CA 92821, USA
0.002% RD + 0.00002 Hz
Auto range,
6-digit
Salt Lake City Office
1364 West State Rd. Suite 101 Pleasant Grove, UT 84062, USA
Page 4
SPECIFICATIONS
Measurement Accuracy Cont.
Specifications
-300 to 300 mV 1 μV 0.005% RDG + 15 μV -300 to 300 mV 1μV
Voltage DC
Pressure / Process Calibration Equipment
DC High Voltage
AC High Voltage
-30 to 30 V 0.1 mV 0.005% RDG + 1.5 mV -30 to 30 V 0.1 mV 0.01% RDG + 1.5 mV
Temperature Coefcient:
±5 ppm FS/°C (-10°C to 10°C and 30°C to 50°C)
Impedance: -300 mV to 300 mV = > 100 MΩ
-30 V to 30 V = >1 MΩ
-3 to 3 V
-30 to 30 V 0.05% RDG + 3 mV
-300 to 300 V 0.05% RDG + 30 mV
Temperature coefcient: ±0.0025% FS/°C
(-10°C to 10°C and 30°C to 50°C)
Maximum input voltage = 300 V, IEC61010 300V CATII
Commong mode rejection: >100 dB (at 50 or 60 Hz)
Impedance: > 4 MΩ, DC coupling
3V (40 to 500 Hz)
30V (40 to 500 Hz) 0.5% RDG + 15 mV
300V (40 to 500 Hz) 0.5% RDG + 150 mV
Temperature coefcient: ±(0.025% RD + 0.0025% FS)/°C
(-10°C to 10°C and 30°C to 50°C)
Maximum input voltage = 300 V, IEC61010 300V CATII
9% to 100% of range is suitable for the above accuracy indicators
ADT227 ADT227Ex
Range Resolution Accuracy Range Resolution Accuracy
0.01% RDG + 15 μV
Temperature Coefcient:
±5ppm FS/°C (-20°C to -10°C)
0.05% RDG + 0.3 mV
10 mV
N/A
0.5% RDG + 1.5 mV
10 mV
N/A
Current DC
Resistance
(4-Wire)
Voltage mV (TC)
Frequency
Pulse
Impedance: >4 MΩ, <100pF, AC coupling
-30 to 30 mA 0.1 μA 0.01% RDG + 1.5 μA -30 to 30 mA 0.1 μA 0.01% RDG + 1.5 μA
Temperature Coefcient: ±5ppm FS/°C (-10°C to 10°C and 30°C to 50°C), Impedance: < 40 Ω
0 to 400 Ω 1 mΩ 0.005% RDG + 20 mΩ 0 to 400 Ω 1 mΩ
0 to 4000 Ω 10 mΩ 0.01% RDG + 200 mΩ 0 to 4000 Ω 10 mΩ
Temperature coecient:
±5 ppm FS/°C (-10°C to 10°C and 30°C to 50°C)
2-Wire + 50 mΩ, 3-wire+ 10 mΩ
Excitation current: 0.2 mA
-10 to 75 mV 0.1uV 0.008% RDG + 3.0 μV -10 to 75 mV 0.1uV 0.01% RDG + 3.0 μV
Temperature Coefcient:
±5ppm FS/°C (-10°C to 10°C and 30°C to 50°C)
Impedance: >100 MΩ
0.01 to 5 Hz
5 to 50 Hz
50 to 500 Hz 0.002% RD + 0.002 Hz 50 to 500 Hz 0.002% RD + 0.002 Hz
500 to 5000 Hz 0.002% RD + 0.02 Hz 500 to 5000 Hz 0.002% RD + 0.02 Hz
5000 to 50000 Hz 0.002% RD + 0.2 Hz 5000 to 50000 Hz 0.002% RD + 0.2 Hz
Minimum threshold voltage: 2.5 V
Supported units: Hz, kHz, MHz, CPM, CPH, s, ms, μs
0 to 9999999 1 N/A 0 to 9999999 1 N/A
Optional rising edge and falling edge, minimum threshold voltage: 2.5V
Auto range,
6-digit
0.002% RD + 0.00002 Hz 0.01 to 5 Hz
0.002% RD + 0.0002 Hz
Temperature Coefcient:
±5ppm FS/°C (-20°C to -10°C)
Temperature Coefcient:
±5ppm FS/°C (-20°C to -10°C)
Temperature Coefcient:
±5ppm FS/°C (-20°C to -10°C)
5 to 50 Hz
Auto range,
6-digit
0.01% RDG + 20 mΩ
0.01% RDG + 200 mΩ
0.002% RD + 0.00002 Hz
0.002% RD + 0.0002 Hz
04
Switch Supports dry or wet switches. Voltage range of 3 to 30 V. Response speed < 10 ms
Rev # 20220825
Corporate Headquarters
2900 Saturn St #B Brea, CA 92821, USA
Salt Lake City Office
1364 West State Rd. Suite 101 Pleasant Grove, UT 84062, USA
Page 5
SPECIFICATIONS
Genreal Specification
Specifications
Operating Temperature -10°C to 50°C -20°C to 50°C
Specification guaranteed
temperature range
Storage Temperature -30°C to 70°C -30°C to 70°C
Humidity <95%, non-condensing <95%, non-condensing
Power supply
Pressure / Process Calibration Equipment
User interface Icon drive menus Icon driven menus with navigation buttons
Ports protection voltage
Display
Data logger
Maximum altitude 3000 meters
European Compliance CE, UKCA
Electrical Connection Ø4mm sockets and at mini-jack thermocouple socket
Size 6.97" x 4.13" x 2.04" (177 mm x 105 mm x 52 mm)
Weight 1.6 lb (0.7 kg) 1.65 lb (0.75Kg)
Battery Rechargeable Li-ion battery (included)
Battery Life Typically 16 hours Typically 35 hours
Battery Charge 110V/220V external power adapter included. Battery can be charged external to the unit. Typically charge time is 6-8 hours.
External pressure module Dual channel aerial plug, can connect two digital pressure modules
Warm-up time Full specication performance is achieved after a 10 minute warm-up time.
ROHS compliant
Display rate
Barometric Accuracy
(Built-in barometer)
IP protection level IP67, 1 meter drop test
Communication Isolate USB-TYPEC (slave), Bluetooth BLE
User Interface
Localization
Calibration ISO 17025 accredited calibration with data
Warranty 3 years
6600mAh, 23.8Wh lithium battery, charging time 4~6
hours, battery pack can be charged independently
5.0 inch 480 x 800 mm TFT LCD capacitive screen 4.4 inch 640 x 480 mm color display capacitive screen
500 results, each result x 100,0000 recordings,
each recording records a maximum of 7 channel values
English, German, French, Italian, Spanish, Portuguese,
Simplied Chinese, Traditional Chinese, Japanese,
ADT227 ADT227Ex
10°C to 30°C 10°C to 30°C
4000mAh 14.4Wh Explosion-proof lithium battery packcharging time
6~8 hours, battery pack can be charged independently
50V max
(Only for the top ports)
Rohs II Directive 2011/65/EU, EN50581:2012
3 readings per second
55Pa
English, Simplied Chinese, Traditional Chinese, Japanese
Russian, Czech, Slovak
30V max
N/A
CE
05
Pressure Specification
Pressure Specification (ADT227 & ADT227Ex)
The ADT161 and ADT161Ex series Intelligent Digital Pressure Modules are available for gauge, vacuum and absolute pressure from -15 psi to 60,000 psi (-1 bar to 4200 bar).
Accuracy from 0.02% FS includes operation over 14°F to 122°F (-10°C to 50°C), one year stability and calibration uncertainty. For detailed specications,
please refer to the pressure modules datasheet.
Rev # 20220825
Corporate Headquarters
2900 Saturn St #B Brea, CA 92821, USA
Salt Lake City Office
1364 West State Rd. Suite 101 Pleasant Grove, UT 84062, USA
Page 6
SPECIFICATIONS
Temperature Specification
Thermocouple Measurement and Source Accuracy
Type Standard Temperature Range (oC)
S IEC 584 -50 to 1768
R IEC 584 -50 to 1768
Pressure / Process Calibration Equipment
B
K
N IEC 584 -270 to 1300
E IEC 584 -270 to 1000
C ASTM E988 0 to 2315
D ASTM E988 0~2315
G ASTM E1751 0 to 2315
U DIN43710 -200 to 600
Note: Internal CJC is ±0.15°C (-10°C to 50°C ambient temperature) Accuracy with external cold junction only, for internal cold junction add 0.15 ℃(k=2)
IEC 584 0 to 1820
IEC 584 -270 to 1372
J
IEC 584 -210~1200
T IEC 584 -270 to 400
L DIN43710 -200 to 900
ADT227 ADT227Ex
Accuracy (oC)
Measure / Source Measure / Source
-50~0 0.77
0~100 0.56 100~1000 0.56
100~1768 0.44 1000~1768 0.47
-50~0 0.82
0~200 0.57 0~200 0.57
200~1768 0.38 200~1768 0.42
200~300 1.51
300~500 1.00 300~500 1.00
500~800 0.62 500~800 0.62
800~1820 0.43 800~1820 0.43
-250 to -200 0.72
-200 to -100 0.23 -200 to -100 0.24
-100 to 600 0.12 -100 to 600 0.13
600 to 1372 0.22 600 to 1372 0.25
-250 to -200 1.14
-200 to -100 0.33 -200 to -100 0.34
-100 to 1300 0.19 -100 to 1300 0.22
-250~-200 0.39
-200~-100 0.15 -200~-100 0.15
-100~700 0.09 -100~700 0.10
700~1000 0.12 700~1000 0.14
-210~-100 0.19
-100~700 0.10 -100~700 0.11
700~1200 0.15 700~1200 0.17
-250~-100 0.55
-100~0 0.12 -100~0 0.12
0~400 0.08 0~400 0.08
0 to 1000 0.24
1000 to 1800 0.40 1000 to 1800 0.45
1800 to 2315 0.65 1800 to 2315 0.73
0~100 0.31
100~1200 0.25 100~1200 0.27
1200~2000 0.42 1200~2000 0.47
2000~2315 0.65 2000~2315 0.74
50~100 0.90
100~200 0.57 100~200 0.57
200~400 0.35 200~400 0.36
400~1500 0.25 400~1500 0.27
1500~2315 0.49 1500~2315 0.55
-200 to -100 0.11
-100 to 400 0.08 -100 to 400 0.09
400 to 900 0.10 400 to 900 0.12
-200 to 0 0.21 0 to 600 0.08 0 to 600 0.09
Standard Temperature Range (oC)
IEC 584 -50 to 1768
IEC 584 -50 to 1768
IEC 584 0 to 1820
-250 to -200 0.75
IEC 584
IEC 584 -270 to 1300
IEC 584 -270 to 1000
IEC 584 -210~1200
IEC 584 -270 to 400
ASTM E988 0 to 2315
ASTM E988 0~2315
ASTM E1751 0 to 2315
DIN43710 -200 to 900
DIN43710 -200 to 600
-270 to 1372
-250 to -200 1.17
-200 to -100 0.12
-50~100 0.77
-50~0 0.82
200~300 1.51
-250~-200 0.41
-210~-100 0.20
-250~-100 0.57
0 to 1000 0.26
0~100 0.31
50~100 0.90
-200 to 0 0.21
Accuracy (oC)
06
Rev # 20220825
Corporate Headquarters
2900 Saturn St #B Brea, CA 92821, USA
Salt Lake City Office
1364 West State Rd. Suite 101 Pleasant Grove, UT 84062, USA
Page 7
SPECIFICATIONS
RTD Measurement and Source Accuracy
Measure and Simulate Temperature Range (°C)
PT10(385)
PT25(385) -200 to 850
Pressure / Process Calibration Equipment
PT50(3916) -200 to 850
PT100(385)
PT100(391) PT100(3916) PT100(3926)
PT200(385) -200 to 850
PT400(385) -200 to 850
PT500(385) -200 to 850
PT1000(385) -200 to 850
Cu10(427) -200~260 -200~260 0.54 0.56
Cu50(428) --200~260 -200~260 0.11 0.13
Cu100(428) -200~260 -200~260 0.07 0.08
Ni100(617) Ni100(618)
Ni120(672) --80~260 -80~260 0.04 0.05
Ni1000 -50~150 -50~150 0.07 0.07
-200 to 850
-200 to 850
-60~180
Accuracy (°C)
ADT227 ADT227Ex
-200~200 0.57 0.59
200~600 0.67 0.72
600~850 0.75 0.82
-200~200 0.24 0.27
200~600 0.30 0.35
600~850 0.34 0.41
-200~200 0.13 0.16
200~600 0.17 0.22
600~850 0.20 0.27
-200~200 0.08 0.10
200~600 0.11 0.16
600~850 0.14 0.20
-200~200 0.32 0.32
200~300 0.34 0.34
300~600 0.41 0.41
600~850 0.48 0.48
-200~0
0~200 0.18 0.18
200~600 0.25 0.25
600~850 0.30 0.30
-200~200 0.16 0.16
200~600 0.22 0.22
600~850 0.27 0.27
-200~200 0.10 0.10
200~600 0.16 0.16
600~850 0.20 0.20
-60~0 0.05 0.06
0~180 0.05 0.05
0.15 0.15
07
*Note: Ambient temperature of 20°C±10°C.
4-wire accuracy. For 2-wire add 50 mΩ, for 3-wire add 10 mΩ
Rev # 20220825
Corporate Headquarters
2900 Saturn St #B Brea, CA 92821, USA
Salt Lake City Office
1364 West State Rd. Suite 101 Pleasant Grove, UT 84062, USA
Page 8
ORDERING INFORMATION
Model Number
ADT227
•••
HART
ADT227 ADT227Ex: Intrisically Safe
Pressure / Process Calibration Equipment
HART = HART Capabilities
Accessories (included)
Model number Description QTY
9811-X
9811Ex-X
9704
9704Ex
9023 Test leads 1 set (6 pcs)
9027
9060
9052
9052Ex
9040 Hanging strap with magnet 1 pc
110V/220V external power adapter
(Only for ADT227 & ADT227-HART)
110V/220V external power adapter
(Only for ADT227Ex & ADT227Ex-HART)
Chargeable Li-ion battery
(Only for ADT227 & ADT227-HART)
Chargeable Li-ion battery
(Only for ADT227Ex & ADT227Ex-HART)
Right angle test leads (Non-Ex models only)
Pressure module connection cable
USB Cable type A to type C (Non-Ex models only) 1 pc
Ex USB Cable type A to type C (For Ex models only) 1 pc
ISO 17025 accredited calibration certicate 1 pc
Optional Accessories
Model number Description
ADT161 - XXX Digital Pressure Modules
1 pc
1 pc
1 pc
1pc
1 set (2 pcs)
1 pc
ADT161Ex - XXX Intrinsically Safe Digital Pressure Modules
ADT129-X Differential Pressure Manifold, -15 to 3,000 psi
9062
AM1602-6FT
9080
9704
9704Ex
9811-X
9811Ex-X
9906A
9918-SC
9530-BASIC
9530-NET
* Additel/Land software can be downloaded for free at www.additel.com
Connection adapter cable for Fluke style pressure modules to non-explosion-proof Additel readouts
Class A, PT100/385 Industrial RTD, -40°C to 160°C, 3/16 (4.76 mm) inch x 2 inch (50 mm) with 6 foot (1.8 Meters) cable w/ banana jack connectors
Cable kit (including TC plug, compensation cable, S,R,K,J,T,E,N)
Spare chargeable Li-ion battery for multifunction calibrator ADT227 & ADT227-HART
Spare chargeable Li-ion battery for multifunction calibrator ADT227Ex & ADT227Ex-HART
110 V/220 V external power adapter for handheld models ADT227 & ADT227-HART 110 V/220 V external power adapter for Ex handheld models ADT227Ex & ADT227Ex-HART
Hard carrying case for handheld instrument with accessories
Soft carrying case, with space for handheld instrument, test leads, and accessories
Additel/Acal Task management software for multifunction calibrator
Additel/Acal Automated calibration software with asset management, network version, Includes server installation and 1 user license
08
Rev # 20220825
Corporate Headquarters
2900 Saturn St #B Brea, CA 92821, USA
Salt Lake City Office
1364 West State Rd. Suite 101 Pleasant Grove, UT 84062, USA
Loading...