Additel 209, 210 Operating Manual

Page 1
ADT209/210 Loop Calibrator
Page 2
Page 3
ADT209/210 Loop Calibrator
User Manual
(Version number: 1609V02)
Additel Corporation.
Page 4
Statement
This manual is applicable to the ADT209/210 loop calibrator designed and produced by the
Additel Corporation. This manual is written in English. If there are any discrepancy in the content
without prior notice.
Page 5
CONTENTS
1. Warning Notice 1……………………………………………………………………………………………………………
2. Product Description 1………………………………………………………………………………………………………
3. Basic Structure 2……………………………………………………………………………………………………………
4. Keypad function 3……………………………………………………………………………………………………………
5. Display 4………………………………………………………………………………………………………………………
6. Boot display 5………………………………………………………………………………………………………………
7. Main Function Operation 6…………………………………………………………………………………………………
7.1 Measurement/Output Switch 6………………………………………………………………………………………………
7.2 Backlight Brightness Adjustment 6…………………………………………………………………………………………
7.3 Keypad Lock 6………………………………………………………………………………………………………………
7.4 Voltage Measurement 6……………………………………………………………………………………………………
7.5 Current Output/Simulate Transmitter Output 8………………………………………………………………………………
7.5.1 Range Switching 9………………………………………………………………………………………………………
7.5.2 Function Selection 9……………………………………………………………………………………………………
7.5.3 Output Value Adjustment 9………………………………………………………………………………………………
7.5.4 Automatic Ramp/step Output 10………………………………………………………………………………………
7.6 Current Measurement/Loop Transmitter mA Measurement 10………………………………………………………………
7.6.1 Range Switching 10……………………………………………………………………………………………………
7.6.2 Function Selection 11……………………………………………………………………………………………………
8. Menu Operation 12…………………………………………………………………………………………………………
8.1 Menu Structure 12……………………………………………………………………………………………………………
8.2 Menu Operation 13…………………………………………………………………………………………………………
I
Page 6
8.2.1 Enter/Return To Menu 13………………………………………………………………………………………………
8.2.2 Switch Menu Items 13…………………………………………………………………………………………………
8.2.3 Modify Setting Menu Parameter 13……………………………………………………………………………………
8.2.3 Data Entry 13……………………………………………………………………………………………………………
8.3 Auto Shutdown Setting 13……………………………………………………………………………………………………
8.4 Automatic Shutdown Backlight Setting 13……………………………………………………………………………………
8.5 Measurement Filter Parameter Setting 14…………………………………………………………………………………
8.6 Automatic Ramp/Step Output Setting 14……………………………………………………………………………………
8.6.1 Ramp/Step Switch 14……………………………………………………………………………………………………
8.6.2 Direction Setting 14……………………………………………………………………………………………………
8.6.3 Ramp/Step Time Setting 14……………………………………………………………………………………………
8.7 Clear Menu Setting 15………………………………………………………………………………………………………
8.8 Software/hardware Version 15………………………………………………………………………………………………
8.9 Self-test 15…………………………………………………………………………………………………………………
8.9.1 Self-test Description 15…………………………………………………………………………………………………
8.9.2 Self-test Item Switch 15…………………………………………………………………………………………………
8.9.3 Self-test Item Description 16……………………………………………………………………………………………
8.10 Calibration 18………………………………………………………………………………………………………………
8.10.1 Calibration Condition 18………………………………………………………………………………………………
8.10.2 Calibration Item 18……………………………………………………………………………………………………
8.10.3 Calibration Process 18…………………………………………………………………………………………………
8.10.4 Restore Factory Calibration Data 19…………………………………………………………………………………
9 Specification 19………………………………………………………………………………………………………………
9.1 Main Specification 19………………………………………………………………………………………………………
II
Page 7
9.1.1 DC Voltage Measurement 19……………………………………………………………………………………………………
9.1.2 DC mA Measurement/Loop Transmitter Current Measurement 19……………………………………………………………
9.1.3 Source/sink DC Current Output 20………………………………………………………………………………………………
9.1.4 Mechanical/live Switch Measurement 20………………………………………………………………………………………
9.2 General Specifications 20………………………………………………………………………………………………………………
10. Replace The Battery 22……………………………………………………………………………………………………
11. Manufacturer Information 23……………………………………………………………………………………………
III
Page 8
I. Warning Notice
Before using this instrument, please make sure you have read and understand the user manual.
Inspect the calibrator for any damages before use.
Do not apply more than 30V between any two terminals or between any terminal and ground.
When using the calibrator to measure or output the correct wiring, mode and range must be used.
To avoid any potential damage to other devices when testing, please adjust the calibrator to the correct mode before
connecting the test leads.
Replace the batteries as soon as the battery symbol flashes.
Before removing the battery door, remove all test leads from the calibrator.
2. Product Description
ADT209/210 is a multi-function, durable, hand-held loop calibrator. This calibrator offers six primary functions: current
output, analog transmitter output, voltage measurement, switch measurement, transmitter loop measurement, and current
measurement. Each unit comes in a compact, design such that the calibrator can be operated easily with oneergonomic
hand. The large, high contrast VA screen clearly displays the results. Each unit comes standard with an IP54 rating.
The ADT209 0.03%RD accuracy, while the ADT210 provides higher accuracy as 0.01%RD.provides
1
Page 9
3. Basic Structure
Model label
Keypads
210 Loop Calibrator
25%
100%
Range
Math
Auto
30V
Hold
+
Mode
COM
Case
Display
Protective
case
Measuring
jack
Adapter
jack
Loop Calibrator
ADT210
Powered by 9V Alkaline battery or specified adapter.
Additel Corporation USA,Additel Corporation USA,
www additel com..www additel com..
Battery
compartment
Nameplate
2
Page 10
4. Keypad function
measure/output
interface
menu setting
interface
measure/output
interface
menu setting
interface
measure/output
interface
menu setting
interface
measure/output
interface
menu setting
interface
Long press -> on/off
Short press -> Cancel/ Return to previous menu
Short press -> Move left Long press -> Auto ramp/step
Short press -> Enter last menu
/ Move left
Short press -> Change function
mode
Long press -> Enter the menu
Short press -> Switch modes
Short press -> Decrease by 1
Short press -> Decrease/Switch parameter level of menu
Auto
Mode
25%
Range
100%
Math
Hold
measure/output
interface
menu setting
interface
measure/output
interface
menu setting
interface
measure/output
interface
menu setting
interface
measure/output
interface
menu setting
interface
Short press -> 25% step Long press -> Switch range
N/A
Short press -> 100% Span Long press -> Switch function
Short press -> Swith calibration point/ Set ramp start point
Short press -> Increase by 1
Short press -> Increase by 1 /
Switch menu parameter level
Short press -> Move right Long press -> Adjust backlight
Short press -> Enter next menu
/ Move right
measure/output
interface
menu setting
interface
Short press ->Confirm Long press -> Lock/Unlock
Short press ->Confirm
3
Page 11
5. Display
mA output
Range
Keypad lock
Automatic ramp
Automatic step
Main display
Sub display
Function
indication
Menu
Simulate transmitter
output
mA
measurement
Loop transmitter
mA measurement
Voltage
measurement
Battery indicator
Engineering unit
Switch
measurement
Engineering
function
Filter icon
Calibration
standard
Progress bar
Auto off
Auto light Filtering
Ramp/Step
Clear
Software/hardware
version
Self-test
Calibration
4
Page 12
6. Boot display
Full Screen
Voltage Measurement
Software version
/hardware version
5
Page 13
7. Main Function Operation
7.1 Measurement/Output Switch
ADT209 and ADT210 provide the following functions:
Voltage Measurement
Switch measure
Current output
Simulate transmitter output
Current measurement
Loop transmitter current measurement
In the measurement/output interface, you can press to switch through above functions.
Mode
7.2 Backlight Brightness Adjustment
In any non-input screen, you can press and hold to adjust backlight brightness(3-level brightness adjustment).
7.3 Keypad lock
In the measurement/output interface, you can press and hold to lock/unlock the keypad.
Hold
7.4 Voltage Measurement
Voltage measurement interface can simultaneously display , youvoltage measurement value and percentage of range
can press and hold to switch the current range, the voltage measurement has 3 ranges:
25%
Range
0 ~ 30V (0% ~ 100%) [default]
1V ~ 5V (0% ~ 100%)
0 ~ 10V (0% ~ 100%)
6
Page 14
210 Loop Calibrator
25%
100%
Range
Math
MEASURE V
Press
Mode
Auto
to display
30V
Hold
+
Mode
COM
The maximum
voltage of 30V DC
- +
Fig 7.1 Voltage Measurement
7
Page 15
7.5 Current Output/Simulate Transmitter Output
In current output mode ,()As Fig 7.2 the calibrator supplies DC current with loop power.In simulate transmitter mode
As Fig 7.3 ), the calibrator simulates a two-wire transmitter that has external power supply.
DC output
Each press
25%
Range
can increase or reduce by 25% of the current full scale in stepwise manner
Long press
Auto
to
switch the automatic step/ramp output
Press
Mode
to display
210 Loop Calibrator 210 Loop Calibrator
Simulate transmitter output
Each press
25%
Range
can increase or reduce by 25% of the current full scale in stepwise manner
Auto
25%
Range
100%
Math
Long press
Auto
to
Auto
25%
Range
100%
Math
switch the automatic
Mode
Hold
COM
30V
+
step/ramp output
Press
Mode
Mode
COM
COM
Hold
30V
30V
+
+
to display
- +
Fig 7.2 Current Output
- +
Fig 7.3 Simulate Transmitter Output
8
Page 16
7.5.1 Range Switching
In the interface, you can press and hold key to switch the currentcurrent output/simulate transmitter output
range. has 3 ranges:Current output/simulate transmitter output
4 ~ 20mA (0% ~ 100%) [default]
0 ~ 20mA (0% ~ 100%)
0 ~ 24mA (0% ~ 100%)
7.5.2 Function Selection
In the interface, you can long press to switch the percentagecurrent output/simulate transmitter output
calculation. provides three calculation functions:Current output/simulate transmitter output
Linear, [default]
Square
Valve
7.5.3 Output Value Adjustment
100%
Math
25%
Range
In the interface, you can press to adjust the output value by 25% ofcurrent output/simulate transmitter output
current range. You can press key to output 0% and 100% of full scale. Press key or key to adjust by an
100%
Math
25%
Range
integer by 1mA.
You can press key or key, the cursor will blink in the main display. Press or to adjust the position of
Auto
Auto
the cursor. Press or to increase or decrease the value.
9
Page 17
7.5.4 Automatic Ramp/Step Output
In the function,current output/simulate transmitter output
you can press and hold key to select auto ramp/automatic
step output. Press andhold key to switch output function.
Long press key to switch range and press key to
25%
Range
Auto
100%
Math
100%
Math
select a starting point.
You can press key to start output, and press again
Hold
Hold
key to pause the output. Press to exit automatic ramp/
automatic step menu.
7.6 Current Measurement/Loop Transmitter mA Measurement
Loop transmitter mA measurement provides 24V loop power, and
also can measure the current value in the loop. In this mode As Fig
7.4 , the circuit power has series resistor greater than 250Ω, which
can be compatible with HART devices without using external resistor.
7.6.1 Range Switching
Loop transmitter
current measurement
Press
Mode
to display
Max. 30V DC voltage
210 Loop Calibrator
25%
100%
Range
Math
Auto
Mode
COM
Hold
30V
+
Dual Cable Transmitter
- +
In the current measuring/loop transmitter mA measurement
interface, long press key to switch current range. The current
25%
Range
Fig 7.4 Loop Transmitter Current
Measurement Configuration
10
Page 18
4 ~ 20mA (0% ~ 100%) [default]
0 ~ 20mA (0% ~ 100%)
0 ~ 24mA (0% ~ 100%)
7.6.2 Function Selection
has three ranges:measurement/loop transmitter mA measurement
In the interface, long press key to switch thecurrent measuring/loop transmitter mA measurement
100%
Math
percentage calculation function. The provides twocurrent measurement/Loop transmitter mA measurement
calculation functions:
Linear, [default]
Square
11
Page 19
8. Menu Operation
8.1 Menu Structure
Automatic shutdown
time(min)
No automatic shutdown
Backlight shutdown
()time min
Always on
First order filtering coefficients
(the greater the value
,is the more obvious
the filtering effect is)
Close filter
1 Automatic ramp parameter settings.
.2 Automatic step parameter settings
See 8.6 automatic ramp/step
output settings
15 30 60
120
OFF
10
ON
10
OFF
Automatic shutdown settings
1
Automatic shutdown backlight settings
5
2 3
Filter parameter settings
……
Ramp/step parameter settings
Calibration
System self test-
Software hardware/
version
Measurement clear
setting
1 Calibration.
.2 Restore factory settings
See 8.10 Calibration
1 Battery voltage detection.
.2 Board voltage detection .3 Loop Power Supply Voltage
Detection
./4 Current output measurement .-5 Switching measuring self test .6 LCD segment code complete
detection
1 SV software version.-
.-2 HV hardware version
1 Clear voltage measurement.
.2 Clear current measurement
12
Page 20
8.2 Menu Operation
8. 2.1 Enter/Return to menu
In the measurement/output display interface, long press key to enter the settings menu.
In any menu button, you can press key to return to the previous short menu until you exit.
8.2.2 Switch Menu Items
In the menu interface, you can press or to toggle the items along the menu bar:
8.2.3 Modify Setting Menu Parameter
You can press or to switch the setup parameters of current menu.
8.2.3 Data Entry
Hold
Confirmation/ enter data
Cancel data input
Auto
Move the cursor left/right
Increase/decrease by 1
8.3 Auto Shutdown Setting
Mode
Auto
In the automatic shutdown menu, press or to select auto shutdown time, 15/30/
60/120 minutes are optional, or OFF (no auto shutdown).
8.4 Automatic Shutdown Backlight Setting
In the auto shutdown backlight menu, you can press key or key to select the auto
13
Page 21
backlight shutdown t time, 1/5/10 minutes are optional, or ON (the backlight is always on).
8.5 Measurement Filter Parameter Setting
In the filter adjustment menu if the measurement is voltage, current or loop transmitter
current, the filter coefficients will be displayed directly. If there are other items, you can press key to switch between
Mode
measurement items.
Press key or key to select first order filtering coefficients, 2…10 are optional (the larger the value, the more
obvious the filtering effect) or OFF (no filter).
8.6 Automatic Ramp/Step Output Setting
8.6.1 Ramp/Step switch
In the menu press key or key to switch betweenautomatic ramp/step setup automatic ramp output
Mode
Auto
setting automatic step output settingand .
8.6.2 Ramp Direction Setting
In the menu press to set .automatic ramp/step setting automatic ramp/step direction
100%
Math
This icon means the direction is 0%→100%→0% or 100% →0%→100%, and means the 0%→100%
or 100%→0%.
8.6.3 Ramp/Step Time Setting
In automatic ramp/step setting menu press key to enter the time setting. Press key to move the cursor
and press key to increase or decrease by 1 second.By pressing the key again will confirm the entry or
Hold
Hold
Auto
14
Page 22
press key to cancel the entry.
8.7 Clear Menu Setting
In the clear setting menu ,In the voltage measurement clear or current measurement clear
menu, you can press key to preform clear operation, the cleared measurement value will be observed in the main
Hold
display interface.
8.8 Software/hardware Version
In the software/hardware version browse menu , main display area indicates that the software
version is 1-01, sub display area indicates the hardware version is 1-01.
8.9 Self-test
8.9.1 Self-test Description
When performing self-test, you will need to short the electrical measuring port. Self Test menu item contains the
following self-tests: supply voltage detection, system voltage detection, loop power/voltage measurement self-test, current
output/current measurement self-test, switch measurement self-test, and LCD test.
8.9.2 Self-test Item Switch
In the self-test menu , you can press key to toggle self-test items (e.g. supply voltage
Mode
15
detection, system voltage detection, etc.).
Page 23
8.9.3 Self-test Item Description
Test Item Interface Schematic
Supply voltage detection
System Power Detection
Loop supply voltage /voltage
measurement self-test
Range
Supply voltage: 6V ~ 10V
a.3.3V Voltage: 3.2V ~ 3.4V b 5V Voltage: 4.8V ~ 5.2V.
22V~25V
Notes
a If the power is supplied only by.
the battery, the displayed voltage will be the battery voltage
b If you insert the power adapter,.
the displayed voltage will be the adapter voltage
_
If the range is exceeded additional troubleshooting or calibration may be required.
Current Output/Current
Measurement Self test -4mA
3.9mA~4.1mA
If the range is exceeded calibration is required.
16
Page 24
Current Output/Current
Measurement Self test -20mA
19.9mA~20.1 mA
If the range is exceeded calibration is required.
Switch measurement self-
test - OPEN
Switch measurement
self-test - CLOSE
LCD Detection
The main display area displays OPEN
The main display area displays CLOSE
All pixels will illuminate
_
_
_
17
Page 25
8.10 Calibration
We recommend your calibration standards to be at least four times more accurate than the ADT210.
8.10.1 Calibration Condition
Note: Ensure the following critera are meet during the calibration:
Ambient temperature relative humidity ; atmospheric pressure . ℃± ℃; 20 2 45%-75% 86kPa-106kPa
Allow at least a 15 min warmup time for the ADT210.
8.10.2 Calibration Item
ADT210 2-point calibration:
Voltage measurement: calibration points are 0 and 30V
Current measurement: calibration point are 4 and 20mA
Current output: calibration point are 4 and 20mA
8.10.3 Calibration Process
1) Select the calibration menu, you can press key, and enter the password “310” and
press key to confirm.
Hold
2) Press key to toggle between (the first screen is default to ).calibration items current output calibration item
3) Press to browse the calibration data, and toggle between calibration points.
4) Press to enter the calibration value.
Mode
100%
Math
Hold
: enter the value displayed by the current calibrator by pressing key to confirm entry Current output calibration
or pressing key to cancel the entry.
Hold
Hold
18
Page 26
 Current measurement/voltage measurement calibration: When sourcing the standard value press the key
Hold
to confirm the entry or press key to cancel the entry.
5) Press to switch to the next calibration point, repeat step 4).
100%
Math
8.10.4 Restore Factory Calibration Data
when restoring factory calibration data, all other settings parameters will be reset to factory settings.Note:
Select the calibration menu, you can press to enter, and enter password “310”. Press
Hold
key to confirm, the system will reboot.
Hold
9 Specification
9.1 Main Specification
Specifications are based on 1 year and an environmental temperature of 10°C to 30°C .
Temperature coefficient: -10°C~10°C and 30°C~50°C: ±0.005%FS/°C
9.1.1 DC Voltage Measurement
Range:0~30V
Resolution:1mV
Input impedance:1MΩ
Accuracy:ADT209 ±(0. 03%RD + 2mV) ADT210 ±(0.01%RD + 2mV) ::
9.1.2 DC mA Measurement /Loop Transmitter Current Measurement
Range:0~24mA
Resolution: 1μA
19
Page 27
Accuracy: ADT209 ±(0. 03%RD + 2μA) ADT210 ± (0.01%RD + 2μA) ::
Loop transmitter current measurement: maximum load 700Ω
9.1.3 Source/Sink DC Current Output
Range: 0~24mA
Resolution:1μA
Accuracy: ADT209 ±(0. 03%RD + 2μA) ADT210 ±(0.01%RD+2μA) ::
Source mode:Maximum load 700Ω/20mA
Sink mode: External loop voltage nominal 24V, maximum 30V, minimum 12V
9.1.4 Mechanical /Live Switch Measurement
Equipment:
*
Input impedance: greater than 500MΩ
Charged:
*
Input impedance: greater than 500MΩ
Trigger level: low <0.3V, high> 2V*
9.2 General Specifications
Pressure: Maximum voltage between any terminals and ground or between any two terminals:30V
Storage Temperature: -20 70◆℃~
℃~ Operating Temperature: -10 50
20
Page 28
Relative Humidity:
*
95%/30
* 75%/40
Operating altitude: Maximum 3000m
Vibration Shock:
Power Requirements: A 9V alkaline battery (ANSI/NEDA 1604A or IEC6LR61) or standard DC9V adapter
Battery Life:
Weight: 350g (with holster)
Protection class: IP54
Certification: CE
* 45%/50
*
Random 2g, 5 to 500Hz
1m drop*
*
Output mode:18 hours(12mA /500Ω)
Measurement mode:50 hours*
Dimensions: 163mm(L) × 83mm(W) × 41mm(H) (with holster)
21
Page 29
10. Replace the battery
Warming:
To avoid false readings, the battery should be replaced immediately when the display indicates that the battery is low or
does not start properly
Use a 9V alkaline battery (ANSI/NEDA 1604A or IEC6LR61)
Battery replacement:
1) Disconnect test lead from the input signal;
2) Long press key to turn off the calibrator
3) Remove the protective cover;
4) Open the battery door behind the calibrator;
5) Remove the battery;
6) Insert the replacement battery to make sure the
battery polarity is correct;
7) Replace the battery door and determine that the
batteries are in place;
8) Place the calibrator into the leather.
10.1 Replace the battery
22
Page 30
11. Manufacturer Information