Accuracy to 0.01% of reading Small and rugged handheld design Measure, source, or simulate loop current Measure DC volts Simultaneously mA and % span display Switch functionality Selectable ramp and step functions Easy to read display and user interface
HART 250Ω resistor in series with 24V loop
OVERVIEW
The new Additel 209 and 210 loop calibrator series combine ease of use and functionality, making them the ideal tools to troubleshoot your process loop. The ADT209 has an accuracy of 0.03% of reading whereas the ADT210 holds an accuracy of 0.01% of reading. If you want to source, simulate or simply measure, the Additel Loop calibrator series will fit your need. The ADT209 and ADT210 allow for measurement of current, voltage and a switch. You can also simulate or source mA or a process transmitter. With a push of a button, you can switch to zero and span values, auto ramp, and auto step throughout the range. Each loop calibrator has a large, easy to read screen which simultaneously displays the measurement with the % of span.
ELECTRICAL MEASURE SPECIFICATIONS
Range Resolution ADT209 Accuracy ADT210 Accuracy
1
Voltage DC
Current DC
Switch test
[1] 1MΩ input resistance [2] Loop transmitter current measure: 700Ω maximum
0 to 30 V 1 mV 0.03%RD + 2mV 0.01%RD + 2mV
2
0 to 24 mA 1 µA 0.03%RD + 2µA 0.01%RD + 2µA
Input resistance more than 500 MΩ
Trigger level: low level <0.3V; high level: >2V
ELECTRICAL SOURCE SPECIFICATIONS
Range Resolution ADT209 Accuracy ADT210 Accuracy
Current DC 0 to 24 mA 1 µA 0.03%RD + 2µA 0.01%RD + 2µA
Source mode 700Ω/20 mA maximum
Sink mode External loop voltage nominal 24V, maximum 30V, minimum 12V
01
Phone: 714-998-6899 Fax: 714-998-6999 Email: sales@additel.com
Corporate Headquarters
2900 Saturn St #B Brea, CA 92821, USA
Salt Lake City Office
1364 West State Rd. Suite 101 Pleasant Grove, UT 84062, USA
GENERAL SPECIFICATIONS
Voltage limit 30V between terminals or between terminals and ground
Measurement Functions Auto step, auto ramp, span step
Display VA LCD display. 2.04 x 2.04 in (52 x 52 mm)
Loop power 24V
Over-voltage protection 30V DC (240V AC)
Overload current protection 33 mA DC
Storage temperature -20°C to 70°C
Working Environment -10 to 50°C, 95%RH
Working Altitude <3,000 m
Vibration/shock
Power
Battery life (typical)
Size (LxWxH) 163 x 83 x 41 mm
Weight 350 g
Calibration Certification ISO17025 accredited certicate of calibration with data
Compliance Certification
Warranty 3 years
Random 2G 5 to 500Hz 1 meter drop
One 9V alkaline battery (ANSI/NEDA 1604A or IEC)
DC9V optional adapter available
Output mode: 18 hours (12 mA/500Ω)
Measure mode: 50 hours
ADT209: CSA, CE, DNV ADT210: CSA, CE, DNV
ORDERING INFORMATION
Model Number
ADT209
02
ADT210
Accessories included
Test lead set 1 set
Alligator clips 2 pcs
User manual 1 pc
9V Alkaline battery 1 pc
Optional Accessories
9812 110V/220V external power adapter (DC 9V)
Phone: 714-998-6899 Fax: 714-998-6999 Email: sales@additel.com