932 DIAGNOSTIX™DESKTOP MERCURY UNIT
18
19
20
36
35
34
33
32W
31W
30
21
22
23
29
24
28
25
37
15
13
14
6
5
4
3
16
17W
15A
26
27
2
1
2

Remove red safety cap which seals the reservoir containing mercury (on the 952,
the red safety cap is located on the air inlet beneath swivel base). Attach free end
of reservoir tube securely to air inlet. Save red safety cap for later use. When transporting mercury instrument we recommend placing the red safety cap on reservoir.
Please Note:
accumulate around the air inlet beneath the red safety cap during transportation.
Ref # Item # Description
1 932-101
2 932-102
3 932-103
4 932-104
5 932-105
6 932-106
13 952-109
14 952-110
15 952-111
15A 952-112
16 952-113
17W 952-114
18 952-115
19 952-116
20 952-117
21 952-118
22 952-119
23 952-120
24 952-121
25 952-122
26 952-123
27 952-124
28 952-125
29 952-126
30 952-127
31W 952-128
32W 952-129
33 952-130
34 952-132
35 952-131
36 952-133
37 952-134
CONNECTING THE INFLATION SYSTEM TO MANOMETER
it is normal for negligible amounts of residual mercury droplets to
Base Hex Bolt
Base Washer
Base Felt
Base
Basket Screws (3)
Basket
Reservoir Connector
Cotton Filter
Reservoir Connector Gasket
Reservoir Feeder Tube
Reservoir Cover Screws (3)
Reservoir Cover
Scale Plate Holder
Cartridge Top Holder
Rubber Gasket
Reservoir Holder for Top Plate
Rods (2)
Springs (2)
Locking Washer (2)
Locking Lever Handle
Reservoir Screws (2)
Reservoir Tank
Reservoir Rod
Bushing
Locking screw (1)
Scale Plate
Graduated Plastic Cartridge Tube
Rubber Gasket
Filter Plug
Diaphragm
Screw Cap
Reservoir Base Screws (2)
3

MAINTENANCE
To Lock Mercury Within the Reservoir: Tilt the entire
unit back 45° towards the reservoir to permit mercury
to flow out of cartridge tube and into reservoir. When
cartridge tube is completely emptied of ALL mercury
(and while it is still tilted 45°), move locking lever to the
right “OFF”. Mercury should be locked within the reservoir during maintenance or transport. To release mercury, move lever to “ON” position. If necessary, tilt unit
towards tube to allow mercury to flow out of reservoir.
Lift up the spring loaded EZ-Tube™mechanism located at the top of the instrument. Tilt instrument forward to free
tube. If necessary use a small flat head
screw driver to help dislodge the top of
the tube. To replace tube, lift EZ-Tube
mechanism and inset base of tube into
main unit, being sure graduation marks
face out. Next position top of tube and
release EZ-Tube
™
mechanism.
™
To Register Your Product, visit us at
www.adctoday.com
and follow the links
55 Commerce Drive
Hauppauge, NY 11788
Customer Service: 1-800-ADC-2670
Voice: 1-516-273-9600
Fax: 1-516-273-9659
email: info@adctoday.com
http://www.adctoday.com
NEW YORK • LONDON • TAIPEI • TOKYO
©2017 ADC IFU p/n: 9356A-00
4