1.0 INTRODUCTION ......................................................................................................3
2.0 SPECIFICATIONS....................................................................................................4
3.0 INSTALLATION ........................................................................................................6
3.1 UNPACKING.........................................................................................................6
3.2 LOCATING............................................................................................................6
3.3 SETTING UP.........................................................................................................7
4.0 KEYPAD ...................................................................................................................8
4.1 NUMERIC ENTRY METHOD................................................................................8
5.0 DISPLAY...................................................................................................................9
6.0 SYMBOLS AND INDICATORS.................................................................................9
7.0 BATTERY OPERATION .........................................................................................10
8.0 BACKLIGHT............................................................................................................10
9.0 AUTO POWER OFF...............................................................................................10
10.0 OPERATIONS .......................................................................................................11
10.1 ZEROING............................................................................................................11
10.2 TARING...............................................................................................................11
10.3 WEIGHING..........................................................................................................12
10.4 CHECK-WEIGHING............................................................................................12
10.5 LIMITS STORED IN MEMORY........................................................................13
11.0 RS-232 INTERFACE ..............................................................................................13
12.0 CALIBRATION........................................................................................................14
12.1 WBW-M APPROVED SCALES, CALIBRATION PROCEDURE........................14
12.2 SECURITY SEALS..............................................................................................14
12.3 CALIBRATION COUNTER..................................................................................15
12.4 WBW SERIES , NOT APPROVED .....................................................................15
13.0 PARAMETER SETTING.........................................................................................17
13.1 CHECK WEIGHING PARAMETERS...............................................................17
13.2 SCALE PARAMETERS ...................................................................................18
14.0 ERROR CODES .....................................................................................................19
15.0 REPLACEMENT PARTS AND ACCESSORIES.....................................................20
16.0 SERVICE INFORMATION......................................................................................20
17.0 WARRANTY INFORMATION .................................................................................21
© Adam Equipment Company 2013
1
© Adam Equipment Company 2013
2
1.0 INTRODUCTION
The WBW range provides accurate, fast and versatile general purpose weighing scales with check-weighing functions.
There are four series within the range- WBW..M, WBW.., WBW..a and WBW..aM
The WBW..M scales are configured at the factory for compliance with EN 45501, OIML R-76. Type Approval Number T7476/TC7477.
The WBW..aM are NTEP approved. NTEP certificate 10-082. Approved scales have different capacities and readabilities from the standard WBW series.
The WBW series are similar to the WBWa series except the WBW scales are usually set for metric units, whereas for the WBWa series are usually set for imperial units. However, the user
can change the setting as explained in section 13.2.
All have stainless steel weighing platforms on an ABS plastic base assembly which is sealed to IP 65, making it water-proof.
All scales have sealed keypads with colour coded membrane switches and the displays are large easy to read liquid crystal type displays (LCD) supplied with a backlight. The scales can be supplied with an optional rear display.
The scales include automatic zero tracking, audible alarm for pre­set weights and semi-automatic tare.
© Adam Equipment Company 2013
3
2.0 SPECIFICATIONS
EC Type Approved Models (Type Approval Number T7476/TC7477)
Kilograms Max 3 kg 6 kg 15 kg e = 0.001 kg 0.002 kg 0.005 kg Grams Max 3000 g 6000 g 15000 g e = 1 g 2 g 5 g
NTEP Type Approved Models (Type Approavl number 10-082)
Pounds Max 6 lb 15 lb 30 lb e = 0.002 lb 0.005 lb 0.01 lb Grams Max 3 kg 6 kg 15 kg e = 0.001 kg 0.002 kg 0.005 kg
WBW 2 / 5a WBW 4 / 9a WBW 8 / 18a WBW 16 / 35a Kilograms
Maximum Capacity 2.000 kg 4.000 kg 8.000 kg 16.000 kg Tare Range -2.000 kg -4.000 kg -8.000 kg -16.000 kg Readability 0.0002 kg 0.0005 kg 0.001 kg 0.002 kg Repeatability (S.D.) 0.0002 kg 0.0005 kg 0.001 kg 0.002 kg Linearity (±) 0.0004 kg 0.001 kg 0.002 kg 0.004 kg
Grams
Maximum Capacity 2000 g 4000 g 8000 g 16000 g Tare Range -2000 g -4000 g -8000 g -16000 g Readability 0.2 g 0.5 g 1 g 2 g Repeatability (S.D.) 0.2 g 0.5 g 1 g 2 g Linearity (±) 0.4 g 1 g 2 g 4 g
Pounds
Maximum Capacity 5 lb 9 lb 18 lb 35 lb Tare Range -5 lb -9 lb -18 lb -35 lb Readability 0.0005 lb 0.001 lb 0.002 lb 0.005 lb Repeatability (S.D.) 0.0005 lb 0.001 lb 0.002 lb 0.005 lb Linearity (±) 0.001 lb 0.002 lb 0.004 lb 0.01 lb
Ounces
Maximum Capacity 80 oz 144 oz 288 oz 560 oz Readability 0.01 oz 0.02 oz 0.05 oz 0.1 oz Repeatability (S.D.) 0.01 oz 0.02 oz 0.05 oz 0.1 oz Linearity (±) 0.02 oz 0.04 oz 0.1 oz 0.2 oz
WBW 3M WBW 6M WBW 15M
WBW 6aM WBW 15aM WBW 30aM
© Adam Equipment Company 2013
4
Pounds:Ounces
Maximum Capacity 5 lb: 0.00 oz 9 lb:0.00 oz 18 lb:0.0 oz 35 lb: 0.0 oz Readability 0.01 oz 0.02 oz 0.1 oz 0.1 oz Repeatability (S.D.) 0.01 oz 0.02 oz 0.1 oz 0.1 oz Linearity (±) 0.02 oz 0.04 oz 0.2 oz 0.2 oz
OTHER SPECIFICATIONS
Units of measure WBW..M kg, g
WBW..aM lb, kg, g, oz
WBW/WBW..a kg, g, lb, oz, lb:oz
Tare Full range Stabilisation Time 2 seconds typical Operating
Temperature Power supply 12 VDC, 800 mA
Through an external Power Supply Module Battery Internal rechargeable battery (~50 hours operation) Calibration Automatic External Display 6 digits LCD digital display
with capacity tracker and symbols for units
Scale Housing IP 65 Sealed ABS Plastic housing with
Pan Size 210 x 173 mm Overall Dimensions
(wxdxh) Net Weight 3.3 kg / 7.26 lb
Applications Weighing Scales Functions Weighing, Check weighing
WARNING: The WBW..M and WBW..aM scales should not be calibrated by the user. Calibration of the scales may make it illegal to use the scales. The scales are sealed to prevent unauthorised access to the circuit boards inside the scale. Any modifications done to the mechanisms inside by breaking the security seals may make it illegal to use the scale. If the seals are broken or tampered with, the scale needs to be re-verified by an authorised certification body and re-sealed, before it is used legally. Contact your local metrology standards office for further assistance.
-10°C to 40°C
Stainless Steel pan
231 x 265 x 153 mm
© Adam Equipment Company 2013
5
3.0 INSTALLATION
3.1 UNPACKING
Remove the scale from the packing carefully. Inside the box you will find everything needed to start using the scale-
3.2 LOCATING
Pre-assembled scale  Stainless steel top pan  Power Supply Module
Operators Manual
The scales should not be placed in a location that will reduce the accuracy.
Avoid extremes of temperature. Do not place in direct sunlight or near air conditioning vents.
Avoid unsuitable tables. The table or floor must be rigid and not vibrate.
Avoid unstable power sources. Do not use near large users of electricity such as welding equipment or large motors.
Do not place near vibrating machinery.
Avoid air movement such as from fans or
opening doors. Do not place near open windows or air-conditioning vents.
Keep the scales clean. Do not stack material on the scales when they are not in use.
© Adam Equipment Company 2013
6
3.3 SETTING UP
The WBW Series comes with a stainless steel pan. Place it on the top if already not installed.
Do not press with excessive force as this could damage the load cell inside.
Level the scale by adjusting the four feet. The scale should be adjusted such that the bubble in the spirit level is in the centre of the level and the scale is supported by all four feet.
Attach the power supply module to the bottom of the scale and plug into the mains.
Attach Power Supply module here
Press the [O/I] key to start. The scale will first display the software revision, followed by a self-test.
At the end of the self-test it will display zero weight, if the zero condition has been achieved and the weighing unit that was in use last. A stable and zero symbol indicators are also displayed.
© Adam Equipment Company 2013
7
Loading...