DigiTeo-2/5/10/20
General
Thank you for selecting the DigiTeo Electronic Theodolite of ADA Instruments. For the best performance of the instrument, please
read this operator’s manual carefully and keep it in a convenient location for future reference.
Precautions
1. Avoid heavy shock
For long-distance transportation, pay attention to external package and shock proof.
2. Setting and moving
When placing the instrument on the tripod head, hold the instrument and rotate the tripod screw until it is xed securely on top of the
tripod head. Repeat the above mentioned procedure for removing the instrument from tripod head. If the instrument must be carried
with tripod attached, never carry it horizontally over the shoulder, always keep it in vertical direction when carried. The instrument
must be kept in container for long-distance transportation.
3. Keep it clean
Clean dust of the instrument surface with cotton wool or small brush after using the instrument. Dry the instrument on time after
exposed in the rain. Make sure not to use chemicals to clean battery case and plastic parts. If necessary, damp soft cloth is permissible.
High absorbent cotton and lens-cleaning paper are used for exposed optics. Never use handkerchief and clothes.
4. Avoid the long-time irradiation
Never leave the instrument in extreme heat longer than necessary. It could adversely affect its performance.
5. Check the batterie
Be sure to check the battery for voltage level before using the instrument.
6. Notice
Store of the instrument should have good air circulation and low humidity. Temperature is kept under 45ºC. Often change drier in
the instrument container.
3
MEASUREMENT FOUNDATION
MEASUREMENT FOUNDATION
DigiTeo-2/5/10/20
Application
DigiTeo electronic theodolite adopts incremental digit angle measurement system. The resolution of horizontal angle reading and verti-
cal angle reading is 1″, 5″(0.2mgon, 1mgon ). The angle precision is 2″, 5″(0.5mgon, 1mgon ).
Meanwhile microcomputer techniques adopted in the instrument realizes automatic calculation, storage, and display. The instrument
can display the readings of horizontal angle and vertical angle simultaneously. It can use with the PDA and EDM . Then you can get the
electric speed measurement instrument. It can display, put down the angle, distance and coordinate data. It can correct the instrument
error. Many measuring modes as angle, slope etc can be fullled. DigiTeo electronic theodolite is used for the control surveying, mine,
railway, and irrigation etc projects surveys. Still capable of topographic surveys and general projects surveys.
Construction features
Objective lens
1.
Sighting collimator
2.
Vertical tangent screw and motion clamp
3.
Late level
4.
Display windows
5.
Battery
6.
Operating key
7.
Base
8.
Circular level
9.
Handle screw
10.
Carrying handle
11.
Focusing knob
12.
Eyepiece
13.
Horizontal x clamp and motion clamp
14.
RS-232C communication interface
15.
Base locking lever
16.
Foot screw
17.
4
DigiTeo-2/5/10/20
2
10
1
12
11
4
3
13
6
5
14
15
7
9
8
Pic.1
16
5
MEASUREMENT FOUNDATION
17
MEASUREMENT FOUNDATION
DigiTeo-2/5/10/20
Display and display mark
DisplayFunctionDisplayFunction
SDSlope distanceHOLDHold the horizontal angle
HDLevel distanceTILTOver the compensation range
VDHeight distanceBattery level
VAVertical angle %Percent grade
HALHorizontal angle leftmDistance unit, #m
HARHorizontal angle rightgonAngle unit
SFTThe second function08-06-09 Date
REPRepeat the horizontal angle14:38Time
CRNCompensateAuto power off
6
DigiTeo-2/5/10/20
Operating keyboard and operating key
KeysFunction 1Function 2
OSETSetting horizontal 0Distance measurement
HOLDHold the horizontal angleRepeat horizontal angle
SFTIlluminateSelect the second
R/LHorizontal angle right or leftSetting the mode
V%Percent grade of vertical
angle
Power switch
(Connected with rangender)
measurement
function
Output the data
Preparative before measurement
1. Level the instrument.
1.1. Level and center the instrument correctly to insure
the best performance.
1.2. Place the tripod.
First, put the tripod leg in the proper position and
tighten the locking screws.
1.3. Leveling the instrument with circular level.
Use leveling screws 1,2 to move the bubble of the
circular level until the bubble of the circular level
until the bubble is now located on a line perpendicular to a line running through the centers of the two
leveling screws being adjusted. Revolve the leveling screw 3 to shift the bubble to the center of the
circular.
bubble
7
MEASUREMENT FOUNDATION
bubble
MEASUREMENT FOUNDATION
DigiTeo-2/5/10/20
1.4. Further leveling the instrument with plate level
1.4.1. Loosen horizontal motion clamp and revolve
the instrument. By adjusting leveling screws 1,2 , the
plate level vial is parallel to a line running through the
centers of two leveling screws, and place the bubble
in the center of the level vial.
1.4.2. Next, revolve the instrument 90° (100gon)
around its vertical axis and use the remaining screw
3 to center the level bubble once more.
1.4.3. Repeat the above procedure for each 90°
(100gon) revolution of the instrument and check
whether the level bubble is correctly centered for
all points.
2. Centering the instrument with laser plummet
Turn on the laser plummet by pressing and holding down the SFT for
2 seconds. Move the tool by loosening the screws on the tripod. Align
the point of the laser plummet with the required point on the terrain
(object).
3. Final levelling of the instrument
Repeat procedure of (1) and check whether the level bubble is in the
center of the level vial. Finally tighter adjusting screw.
8
Power switch on
Press , all segments of the display will light on. The
display shows that Vertical angle should be zero. Rotate
the telescope to set the instrument to a vertical angle reading of 0. Press over 2 seconds, it can be power off.
● In order to make sure instrument work continuously,
pay attention to battery display. If battery is insufcient,
replace battery.
● For setting the vertical angle at 0, a datum 0 is provided
on the vertical angle scale circumference. If the telescope
is turned and the sensor passes the datum 0, angle measurement begins.
Power display
Sufcient battery power
Effective battery power
Effective battery power
Need to replace battery
Measurement is impossible. The power will
be cut off in one minute.
Change the batteries
For removing: Push the two locking levers to the opposite direction and pull out the battery box . Then take
the battery from the box.
DigiTeo-2/5/10/20
Installation: First Put the battery into the battery box , then Push the two
locking levers to the opposite direction and put the battery box to the
instrument until it clicks into the position.
Angle measurement
1. Measuring a HAR and vertical angle
OperatingDisplay
Collimate the rst target.
Press [OSET] twice and make
the target horizontal angle A is
0°00′00″.
Collimate the second target B
and the horizontal and vertical
angles are displayed.
How to collimate
- Point the telescope towards the light. Turn the diopter ring and adjust
the diopter so that the cross-hair is clearly observed. (Turn the ring
coward you rst and then backward to focus.)
9
MEASUREMENT FOUNDATION
MEASUREMENT FOUNDATION
DigiTeo-2/5/10/20
- Observe the target with sighting collimator. Allow a
certain space between the collimator and yourself, if for
collimating.
- Focus the target with the focusing knob (pic.2).
Pic.2
Note:
If the focus of the telescope is incorrect the parallax is
created between the cross-hair and target when viewing
vertically or horizontally. This will affect precision in
measurement or survey. The parallax should be eliminated by carefully focusing.
2. Alternation of HAR and horizontal angle left HA
Conrm in the angle mode
OperatingDisplay
Collimate the target A.
Press [L/R], and it will
change between the HAR
and HAL.
Measure in the H
● Press [L/R] key, HAR / HAL is alternation. Every time press
the key, HAL and HAR mode can alternate.
10
mode.
L
L
DigiTeo-2/5/10/20
3. Horizontal angle set
Conrm the angle mode
OperatingDisplay
Turn tangent
screw and set the
horizontal angle
needed.
Press [HOLD] twice.
The horizontal angle is
set and ashes.
Collimate target.
Press [HOLD] key
again the horizontal
angle stops
ashing and is not set
any more.
4. Repeat the angle measure
Conrm in the angle mode
OperatingDisplay
1. Press [SFT] and then
press [HOLD], the
instrument in the repeat
measure mode
2. Collimate the target A.
3. Press [OSET], and
make the horizontal
angle of A is 0°00′00″.
4. Collimate the
second target B
5. Press [HOLD], and
put the horizontal angle
in the instrument .
11
MEASUREMENT FOUNDATION
MEASUREMENT FOUNDATION
DigiTeo-2/5/10/20
OperatingDisplay
6. Collimate the rst
target A, through the
under plate level x
screw and the jiggle
screw.
7. Press [OSET] ,
and make the rst
target horizontal angle
0°00′00″.
8. Collimate the rst
target B, through the
under plate level x
screw and the jiggle
screw.
9. Press [HOLD], it
will display the
average.
OperatingDisplay
10. Repeat 2-8,
you can get the
measure you need.
● In the repeat measure mode, the times should in nine times,
if it is over the times, the display is error..
● In the repeat measure mode, the measure error is ≥30″,
If the instrument display error information. You must do
from 2.
● Press [SFT], exit from this mode
12
5. Measuring a percent of grade (slope measurement)
Conrm in the angle mode
OperatingDisplay
Press [V%], the display of vertical angle
changes to percent
grade.
Press [V%]. Percent
grade changes to angle
display.
Every time press V% , the displayed mode will change
alternately. When measured angle is exceeding ±100% to
horizontal position,“EEEEE.EEE”displays.
DigiTeo-2/5/10/20
1. RS-232 interface
DigiTeo-2 has the RS-232 interface joined. The instrument with the
computer or the PDA through the cable. It can transport the data of
the measure to the computer or the data collection equipment. It also
can transport the computer data to the DigiTeo-2. Remember the interface is under the vertical knob.
2. Record the data
DigiTeo-2 in the different measuring mode, press [SFT], press [V%];
you can output the data from the instrument to the computer or the
PDA (select the cluster correspond), or you can record the data in the
memory of the instrument (select the memory of the instrument).
Output the data and serial communication (DigiTeo-2;5)
This instrument can record the data, you can record the angle
data and the distance data in the memory (500 data) or you can
output the data through COM. The data will include the time,
you should set it before you record or output it. (please see
“function setting ”)
ModeOutput (record)
Angle modeVA, HAR or HAL(vertical angle, horizontal
Distance mode VA, HAR or SD (vertical angle, horizontal
13
angle)
angle, slope distance)
MEASUREMENT FOUNDATION
MEASUREMENT FOUNDATION
14
DigiTeo-2/5/10/20
Memory mode (DigiTeo-2)
OperatingDisplay
Press [V%], power on, come in
the memory mode.
● The rst line will display the
effective data in the memory.
Press [REC], the second line will
glint , and the instrument output
the data to the interface, until it
nished, it will not glint.
Press [HOLD], the rst line
will glint, you should press
[HOLD] in 5 seconds , you
will delete all the data, after
this , the instrument exit from
the memory mode.
● In the memory mode, press [SFT], exit from the memory mode,
return to the angle mode.
Function setting
This series instrument have many functions. You can select
what you need, and you can select and set the mode.
1. the compensator of the vertical angle: ON, *OFF
2. the vertical level 0: *90° (OFF), 0° (ON)
3. power off automatic: ON, OFF (if there is no operation
after 20 minutes, it will power off).
4. the minimum angle display: *1”, 5”, 10”
5. setting the baud rate: 1200, 2400, 4800, *9600
6. select the record: *interface (OFF), memory (ON)
7. correction the level: (ON), *(OFF)
8. buzzer: OFF, *ON
9. select the angle unit: 00°00’00’’ *(360), gon (400 gon)
Options marked with “*” are the factory default settings.
1. Function setting method
The function keys:
[OSET] - select the item circularly
[HOLD] - select the time item circularly (month, date, year,
hour, minute)
[L/R] - select the upwards item or the time add 1
[V%] - select the downwards item or the time minus 1
[SFT] - conrm the setting, exit the setting mode, return to
the angle mode
OperatingDisplay
Press [SFT] , and then
press
[L/R], come in the setting
mode.
Press 0 [OSET], select
the (1-7)
Press [L/R] or [V%],
change the selected item.
DigiTeo-2/5/10/20
2. Time setting
OperatingDisplay
Press [SFT] , and then
press
[L/R], come in the setting
mode.
Press [HOLD], select the
time setting item (month,
data, year, hour, minute,
second), when you select
it, it will glint.
Press [L/R] or [V%], add
or minus it.
Settint the item as you need.
Press [SFT], memory the
setting and return to the
angle mode.
Setting it as the current
time.
Press [SFT], it will save
it and return to the angle
mode.
15
MEASUREMENT FOUNDATION
MEASUREMENT FOUNDATION
16
DigiTeo-2/5/10/20
Setting the vertical collimation error and the telescope axis error and
the compensator error
In this mode, you can observe : the plate on the right and on the left and you can
get the error. Use this error you can adjust the telescope axis when you observe in a
same position of the horizontal plate. You can set the vertical plate error, and avoid
the error in measuring.
OperatingDisplay
Press [R/L] and power, it will dis-
play “SETUP” and “SET 0”. Rotate the telescope, the rst line will
display “SET F1” and will glint.
Level the instrument and make
the plate at left, collimate
the target at innitude, press
[OSET], the rst line will glint
and display “SET F2”
Make the plate at right and then
collimate the same target, press
[OSET], the rst line will glint
and display “SET”.
Press [OSET], the instrument perform the
new data of the vertical
error, the telescope axis
error and the compensator 0 error, and return to
the angle mode.
● If you want to exit at any time , you can press [SFT].
● DigiTeo have no the function of setting the telescope
axis error and the compensator 0 error.
● After you change the “minimum unit”, “compensator
function” mode, you should repeat the above steps.
● After you adjustment it, you should check the instrument OK. Collimate the same target when the plate
at left and at right , the vertical angle summation is
360°±″20″. The difference of the vertical angle should
less than 20″. If there display the E03, you should ad-
just it again or follow “collimation of the telescope
axis”.
DigiTeo-2/5/10/20
The other function
1. Measuring distance stadia hairs
So scale station pole is needed, for example horizontal measuring staff and apparent distance staff. By
viewing through the telescope, the length between
upper and under stadia hairs which multiplies 100 is
the distance from instrument center to station pole.
(The length refers to the reading from station pole
between two stadia hairs).
a. First x the station pole at the measuring point.
stadia hairs
b. Level instrument. By viewing through the telescope,
make sure the reading “ ” between two stadia lines.
c. The distance from instrument plumb bob center to
station staff “L” is 100 times of “l”: L = 100Xl
2. Correction function (DigiTeo-2; 5)
DigiTeo-2 has a vertical axis incline compensate. It can self-motion compensate the incline
angle. When the incline sensor is switch on,
the instrument can display the vertical axis
incline angle. We can adjust the instrument
level exactly by hand through the data. When
it incline over the compensate arrange, the
instrument display “TILT”. We must level the
instrument.
Note:
● In gale day or the quiver surrounding, the angle display is not stabilization. You should close the compensate before you measure it.
● Set or close the compensate function, please read the “setup”.
3. Illumination and timing close
DigiTeo have a display and a illumination setting on the reticle. When
you press [SFT] over 2 seconds, the display and the illumination setting
will be open or closed.
If you don’t operate the instrument for 20 minutes or 30 minutes, the
power will be closed. About this function, you can reference the setup.
17
MEASUREMENT FOUNDATION
MEASUREMENT FOUNDATION
18
DigiTeo-2/5/10/20
Check and adjustment
a. Adjust the eyepiece of the telescope properly prior
to any checking operation which involves sighting
through the telescope. Remember to focus properly,
with parallax completely eliminated.
b. Carry out the adjustment in the order listed, as
the adjustments are dependent one upon another.
Adjustments carried out in the wrong sequence may
even nullify previous adjustments
c. Conclude adjustments by tightening the adjustment
screws securely ( but do not tightening them more than
necessary, as you may strip the threads, twist off the
screw necessary, as you may strip the threads, twist off
the screw or place undue stress on the parts.)
d. The attachment screws must also be tightened suf-
ciently upon completion of adjustments.
e. Always repeat checking operations after adjustments
are made in order to verify results.
1. Check and adjust the plate level.
Check:
a. Place the plate level parallel to aline running through
the centers of two leveling screws(e.g. A, B). Use these
two screws to place the bubble in the center of the plate
level vial.
b. Next, revolve the instrument 180° or 200gon
around the vertical axis and check bubble movement
of the plate level. If the bubble has been displaced,
then proceed with the following adjustment.
Adjustment:
a. Adjust the level adjustment capstan screw with the
accessory adjusting pin and return the bubble towards
the center of the plate level vial. However, correct only
one-half of the displacement by this method.
adjustment screw
displacement with the leveling screws.
c. Revolve the instrument 180°or 200gon around the vertical axis once more and
check bubble movement if the bubble is still displaced, then repeat the adjustment.
adjustment screws
tribrach
2. Check and adjust vertical cross-hair.
Check: Carefully level the instrument with the plate level. If the bubble of
the circular level is centered properly at this time, adjustment is not required.
Otherwise, proceed with the following adjustment.
Adjustment: Shift the bubble to the center of the level by adjusting three
capstan adjustment screws on the bottom surface of the circular level, with
the accessory adjusting pin.(see diagram)
b. Correct the remaining 1/2 amount of the bubble
bubble