This Manual contains confidential and proprietary information owned by Accelerated Care Plus ("ACP") which is protected by copyright. This
Manual or any portion thereof may not be photocopied, reproduced or translated to another language without the express prior written consent of
ACP. This Manual may only be used by entities who have purchased the equipment or have implemented the ACP program and are covered by
an executed Lease Agreement with ACP containing a Confidentiality Agreement, which is incorporated by reference in its entirety. This Manual
may not be used for any other purpose.
Any additional copies of the Manual shall be ordered from ACP. No changes or modifications shall be made to the Manual without prior review
and written authorization from ACP. No authorization is given to market, sell, disclose, or exploit this Manual except as for purposes of using the
Equipment as contemplated by the Lease Agreement.
ACCELERATED CARE PLUS MAKES NO WARRANTY OF ANY KIND WITH REGARD TO THIS MANUAL, INCLUDING, BUT NOT
LIMITED TO, THE IMPLIED WARRANTIES OF MERCHANTABILITY AND FITNESS FOR A PARTICULAR PURPOSE. Accelerated
Care Plus shall not be liable for errors contained herein or for incidental or consequential damages in connection with the furnishing, performance
or use of this Manual.
The information contained in this document is subject to change without notice.
500A
Revised 070101
Part No. 290500A
COPYRIGHT 2001 - 2007, ACCELERATED CARE PLUS CORP., ALL RIGHTS RESERVED
CONFIDENTIAL AND PROPRIETARY
2 OMNISTIM
®
500A USER MANUAL
OMNISTIM® 500A
ACP manufactures a premier line of electromedical devices to assist health care professionals in the rehabilitation of
their patients. The ACP product line includes Pain Control Systems, Muscle Stimulators, Interferential Therapy, and
Ultrasound devices. Our MEGAPULSE
most recent world wide technological advances available for therapeutic application of electromedical devices.
ACP is internationally recognized for its contribution to research in the development of medical applications for
therapeutic rehabilitation. The Company sponsors and conducts research at leading health care institutions and major
universities throughout the world. This new medical frontier holds great promise and opportunity, which will result
in substantial advancements in the health care industry and for ACP.
®
, NEUROPROBE®, OMNISTIM®, and OMNISOUND
®
represent the
COPYRIGHT 2001 - 2007, ACCELERATED CARE PLUS CORP., ALL RIGHTS RESERVED
THE OMNISTIM® 500 .................................................................................................................................................8
Delivery of the OMNISTIM® 500 ......................................................................................................................8
Controls and Functions .......................................................................................................................................9
Use of Barriers – Intermediate Level Disinfection ...........................................................................................18
MODES OF OPERATON: Protocol Reference Sheets...............................................................................................19
Interferential Current (IFC)...............................................................................................................................19
Medium Frequency Alternating Current (MFAC)............................................................................................20
Low Volt Pulsed Current (LVPC) ....................................................................................................................21
High Volt Pulsed Current (HVPC) ...................................................................................................................22
Interferential Current Therapy (IFC) ................................................................................................................23
Medium Frequency Alternating Current (MFAC)............................................................................................30
Low Volt Pulsed Current (LVPC) ....................................................................................................................35
High Volt Pulsed Current HVPC......................................................................................................................39
Service Center...................................................................................................................................................43
Infection Control Supplies ................................................................................................................................47
Return of Defective Products............................................................................................................................49
Procedure for Quick Start-Up ...........................................................................................................................51
Adjusting Global Factory Settings....................................................................................................................59
COPYRIGHT 2001 - 2007, ACCELERATED CARE PLUS CORP., ALL RIGHTS RESERVED
CONFIDENTIAL AND PROPRIETARY
4 OMNISTIM
®
500A USER MANUAL
INDICATIONS & CONTRAINDICATIONS
CAUTION: Federal law restricts this device for sale or use by, or on the order of, a Practitioner
licensed by the laws of the state in which he/she practices to use or order the use of the device.
Please note that Accelerated Care Plus cannot provide medical advice. If you have specific
medical questions, please contact your healthcare professional.
Indications
The OMNISTIM
1. Symptomatic relief and management of chronic intractable pain and as an adjunctive treatment in the
management of acute pain, post surgical pain and pain associated with post-traumatic injury.
2. Relaxation of muscle spasms.
3. Re-education of muscle action.
4. Prevention or retardation of disuse atrophy.
5. Immediate post surgical stimulation of calf muscles to prevent venous thrombosis.
6. Increased local blood circulation.
®
500 is indicated for:
7. Maintaining or increasing range of motion.
Electrical muscle stimulator devices should be used under medical supervision for adjunctive therapy for the
treatment of medical diseases and conditions.
Contraindications
1. Transcutaneous electrical stimulation, or powered muscle stimulators should not be used on patients with
2. Never connect lead wires to the power line or electro-surgery equipment. Use only the lead wires
recommended or approved by the manufacturer.
NOTE: There is no contraindication to the application of Transcutaneous Electrical Stimulation or
Powered Muscle Stimulation over metal implants.
Adverse Reactions
Skin irritation and burns beneath the electrodes have been reported with the use of powered muscle stimulators.
COPYRIGHT 2001 - 2007, ACCELERATED CARE PLUS CORP., ALL RIGHTS RESERVED
CONFIDENTIAL AND PROPRIETARY
OMNISTIM® 500A USER MANUAL 5
WARNINGS & PRECAUTIONS
CAUTION: Federal law restricts this device for sale or use by, or on the order of, a Practitioner
licensed by the laws of the state in which he/she practices to use or order the use of the device.
Please note that Accelerated Care Plus cannot provide medical advice. If you have specific
medical questions, please contact your healthcare professional.
Warnings
•Do not operate this device until the User Manual, including all Indications for Use, Contraindications, Warnings
and Precautions, have been carefully read and understood.
•Operation of this device or placement of lead wires, probes, pads and electrodes in close proximity (less than 5
feet) to an operating shortwave or microwave diathermy unit may produce instability in the device output or
burns at the treatment site. Lead wires and device can pick up the magnetic field output of the diathermy and
through induction convert it into an electrical field, transmit the energy into the patient increasing the current
density at the electrodes of applicators. Since the patient may not feel the 27 MHz frequency, they lack the
protective sensation and tissue burns could result. Short-wave field could potentially damage or reset medical
devices in close proximity to the drum applicator.
•Treatment should not be applied over the carotid sinus nerves, (located in the anterior neck triangle), including,
stellete ganglion, vagus nerve, or laryngeal or pharyngeal muscle. Particular care should be taken for patients
with a known sensitivity to the carotid sinus reflex, as carotid sinus stimulation may alter blood pressure and
cardiac contractility.
•Do not apply treatment over testes, heart or eyes. Thermal effects or electrical stimulation may affect organ
function.
•Do not apply over or in close proximity to active cancer (except in terminal / palliative / hospice care), as
therapy may increase blood flow to the tumor.
•Treatment should not be applied when high fever is present, over swollen, severe infection, (osteomyelitis,
sepsis, tuberculosis, etc.), or inflamed areas or skin eruptions, (e.g., phlebitis, thrombophlebitis, varicose veins,
etc.).
•Do not apply over the lumbar or abdominal region, or over the uterus during pregnancy (to prevent uterine
contraction), or during menstruation as therapy may temporarily increase menstrual flow.
•The long-term effects of chronic electrical stimulation are unknown. To date, the only negative effects reported
are based on skin and tissue irritation under the electrode sites associated with chronic estim use.
•Stimulation should not be applied transthoracically in the vicinity of the heart, as introduction of electrical
current into the heart may cause cardiac arrhythmias.
• Treatment should not be applied transcranially.
• Stimulation should not be applied to patients connected to patient monitoring equipment, as the stimulation may
have an affect on the proper operation of the monitoring equipment.
•Stimulation should not be applied directly over external or implanted stimulator systems, such as shunts,
infusion pumps, stomach, bladder, brain, spinal cord, bone growth, or other implanted stimulators.
•Neuromuscular electrical stimulation (NMES) should not be applied directly over or in close proximity to Deep
Vein Thrombosis (DVT), as it activates the muscle and causes muscle contractions. This should be avoided in
tissue following an acute DVT when the thrombosis is not completely resolved. Therapists should follow the
guidelines provided by the referring physician on recommended activity level and modality use. If the patient is
not permitted exercise, NMES therapy should be avoided. Generally, NMES over a DVT of six weeks or less
should be avoided altogether.
COPYRIGHT 2001 - 2007, ACCELERATED CARE PLUS CORP., ALL RIGHTS RESERVED
CONFIDENTIAL AND PROPRIETARY
6 OMNISTIM
®
500A USER MANUAL
Precautions
• Application site and settings should be based on the guidance of the prescribing practitioner.
• All equipment and accessories should be kept out of the reach of children or unqualified persons.
• Do not connect this device to any wall outlet that has not been properly grounded, or to any electrically non-
isolated medical device.
•Use only ACP specified accessories and/or supplies with ACP devices. Do not use any power cords, or power
supplies, other than the ones provided or specified for this device. Use of any other power supply could
seriously damage the device and will void the warranty.
•When cleaning the device, never immerse them or wash them with water. See the Infection Control section in
this manual for cleaning instructions. Devices should not be submerged in water or other liquids.
•Failure to follow the manufacturer’s prescribed maintenance for this device may lead to device failure and
transient or unreliable performance. State and federal survey and JCAHO require all equipment to be
maintained and calibrated according to the manufacturer recommended schedules.
•A potential electric shock hazard exists once the device outer casing has been in part, or fully, removed. Only
qualified service personnel should perform Service and repairs. Warranty will be voided if the outer casing has
been removed or tampered with.
•Do not apply over areas of hemorrhage or active bleeding. Modalities should be applied following cessation of
hemorrhage or active bleeding as they may promote increased circulation and inhibit coagulation.
•Inspect and cleanse the skin prior to application. Following treatment check the skin for evidence of irritation
or burns, and if present, treat as appropriate. If the patient has, or complains of, skin irritation following
treatment, shorten the treatment time on the next treatment session, or use an alternative type of therapy or
electrode placement.
•Gradually increase the output intensity/power to required dose or patient tolerance while monitoring the device
display.
•Caution should be taken with patient exhibiting psychological or physical hypersensitivity to the therapeutic
treatment. Several attempts should be made to place them at ease so that their confidence and cooperation can
be gained during the treatment.
•The treatment area should be checked from time to time, and if there is evidence of, or if the patient complains
of, pain during treatment, adjust the output downward until it is tolerated by the patient. If the patient continues
to complain of pain, discontinue the treatment and shorten the treatment time on the next treatment session, or
use an alternative type of therapy or electrode placement.
•Do not apply treatment directly over/under hot or cold packs. Caution is recommended when treatment follows
the application of hot or cold therapy, which may alter the patient’s sensation. Application of thermal agents
over areas of impaired circulation should be performed with caution as the circulation may be insufficient to
heat or cool the tissue, altering the patient’s perception of warmth and pain, and burns or tissue necrosis may
result from subsequent treatment.
•Caution is recommended when treatment follows the application of medicated patches, salves, or creams which
may alter the patient’s sensation. If there is a medical necessity to perform such treatments, these patients
should be monitored diligently during application. The presence of electrical stimulation may be altered by the
presence of these materials on the patient’s skin.
•Caution should be used over areas of body where circulation is impaired, or which lack normal sensation.
Absent or diminished sensation should be avoided or, if unavoidable, treated with caution. Establishment of
acceptable intensity levels for desensitized areas may be related to the intensity levels tolerated on normal skin
in opposite or related body parts.
•Caution should be used in the presence of recent surgical procedures, fractures or healing bone and soft tissue
when muscle contraction may disrupt the healing process.
•Caution should be used for persons with suspected or diagnosed heart problems.
COPYRIGHT 2001 - 2007, ACCELERATED CARE PLUS CORP., ALL RIGHTS RESERVED
CONFIDENTIAL AND PROPRIETARY
OMNISTIM® 500A USER MANUAL 7
• Caution should be used for patients with suspected or diagnosed epilepsy.
• Electrodes should not be placed in direct contact or in close proximity (one inch or less) of each other during
treatment. Electrodes placed in contact or in close proximity can lead to high energy density and skin burns
under or between the electrodes.
•Care should be used when removing electrodes after treatment, in order to minimize the potential for skin
tearing. Skin should be inspected after removal of electrodes for any signs of tearing or irritation.
•Do not connect the stimulator to any electrical equipment for combination therapy except the Omnisound
family of ultrasounds.
•ACP does not recommend the use of the Denervated Muscle or Low Intensity DC protocols, and does not
supply the DC or Microcurrent adapter required for use with these protocols. If this device is used with a DC or
Microcurrent adapter, follow the specific treatment guidelines, indications, contraindication, warning and
precautions specified in the DC or Microcurrent adapter manual. Do not use any protocols requiring the use of
the DC or Microcurrent adapter without first having plugged the adapter into the appropriate channel.
®
COPYRIGHT 2001 - 2007, ACCELERATED CARE PLUS CORP., ALL RIGHTS RESERVED
CONFIDENTIAL AND PROPRIETARY
8 OMNISTIM
®
500A USER MANUAL
THE OMNISTIM® 500
Delivery of the OMNISTIM® 500
Upon receipt of your OMNISTIM
damage. All ACP products are packaged carefully for rapid, safe delivery. We guarantee delivery in perfect
condition to the postal or delivery services. However any damage or loss incurred during transportation or delivery
is the Postal or Delivery Company’s responsibility. If damage or loss to the product and/or container is obvious or
suspected, appropriate notation must be made on the signed freight bill at the time of delivery. All damage claims
should be promptly filed with the delivering carrier and must be initiated by the addressee where the package was to
be delivered. Retain the original shipping container and inserts for validation of damage claim or use at a later date.
Unpack and check all accessories. A list of enclosed accessories is provided with each unit to assist you in
identification of the type and number of accessories.
NOTE: The purpose of this manual is to acquaint you with the OMNISTIM® 500’s operating features and
functionality. Please read the manual carefully before attempting to operate the OMNISTIM
remain unanswered, contact your ACP sales representative or call 1-800-350-1100. Outside the USA call 1-775685-4000.
®
500, inspect the shipping container and contents for any obvious or concealed
®
500. If questions
Introduction
The OMNISTIM
Currents (MFAC), Low Voltage Pulsed Current (LVPC) and High Voltage Pulsed Current Therapy (HVPC). Its
two separate generators produce medium frequency (2000, 2500, 4000 or 5000 Hz) alternating current in continuous
or modulated modes. Two isolated output circuits with independent intensity controls are provided. The output of
each circuit is easily determined in milliamps through the display screen. The digital timer allows the operator to
select the length of the total treatment time and to monitor the time remaining in minutes.
The OMNISTIM
interferential vector scanning mode can be turned off or on for stationary or continuous movement of the
interferential field.
The OMNISTIM
provides a wide variety of uses for muscle re-education and muscle spasm reduction protocols for innervated
muscle.
The OMNISTIM
operation with fully adjustable ON and OFF times and ramps allow applications of HVPC therapy. Increasing
circulation, pain control and muscle re-ed can be set up in this mode.
®
500 is designed to provide Interferential Current Therapy (IFC), Medium Frequency Alternating
®
500 offers Frequency Difference (FD) and Full Field (FF) Interferential Current Therapy. An
®
500 MFAC and LVPC modes with fully adjustable ON and OFF times and ON and OFF ramps
®
500 provides High Voltage Pulsed Current (HVPC) on one channel. Continuous or surged
COPYRIGHT 2001 - 2007, ACCELERATED CARE PLUS CORP., ALL RIGHTS RESERVED
CONFIDENTIAL AND PROPRIETARY
OMNISTIM® 500A USER MANUAL 9
Controls and Functions
A. Main power switch. Press for power ON, press again for power OFF. The display illuminates with SELECT
MODALITY/PROGRAM.
B. Parameter select button provides adjustment of stimulation parameters, screen contrast, audio on/off and button
speed and sensitivity. Press to enter SETECT mode. It is possible to adjust most stimulation parameters at any
time during treatment.
C. START/STOP Set output button. Having selected the type of treatment and program press to set the output or
start or stop the treatment. START/STOP also cancels the SET mode.
D. Increase/Decrease OUTPUT A. Press rocker switch to the right to increase output current. Press to the left to
decrease output current.
E. Increase/Decrease OUTPUT B. Press rocker switch to the right to increase output current. Press to the left to
decrease output current.
F. Increase/Decrease Beat frequency or TIME OFF Press rocker switch to the right to increase the duration of the
TIME OFF period; press to the left to decrease the duration of the TIME OFF period during MFAC, LVPC or
HVPC modes. Press rocker switch to the right to increase BEAT (or base) frequency; press to the left to
decrease BEAT frequency during INTERFERENTIAL modes.
G. Increase/decrease SWEEP frequency or TIME ON. Press rocker switch to the right to increase duration of
TIME ON period; press to the left to decrease the duration of the TIME ON period during MFAC, LVPC or
HVPC modes. Press rocker switch to the right to increase SWEEP (or spectrum) frequency; press to the left to
decrease SWEEP frequency during INTERFERENTIAL mode.
H. Timer/parameter switch. Increases/decreases treatment time or changes parameters in set mode.
I. Select single channel HVPC protocol when pressed.
J. Selects LVPC protocol when pressed.
K. Selects MFAC (Russian style stimulation) protocol when pressed.
L. Select Interferential (IFC) protocol mode.
M. Patient Safety Input Switch
N. Charging/line operation indicator LED.
O. Battery charger input jack.
P. Output for HVPC.
A
B
C
S
T
D
Q
R
E
P
F
O
N
M
L
K
J
I
G
H
COPYRIGHT 2001 - 2007, ACCELERATED CARE PLUS CORP., ALL RIGHTS RESERVED
CONFIDENTIAL AND PROPRIETARY
10 OMNISTIM
®
500A USER MANUAL
Q. Output for channel “B” in IFC, MFAC and LVPC modes.
R. Output for channel “A” in IFC, MFAC and LVPC modes.
S. Lead wire adapter – 3 pin.
T. Super twist LCD graphics display for all functions and parameters. The graphics screen allows display of:
• Modality selected
• Program selected
• Time remaining
• Bar graph for timing
• On-Off time, delay and ramps
• Beat frequency and sweep ranges
• Sweep rate and modulation selection
• Burst or pulse frequency
• Phase duration and interphase interval
• Output voltage or current
• Vector selection
• Battery condition
• Display contrast
• Button speeds and delay
Factory Settings
The OMNISTIM
These are generally suitable for most clinicians. To adjust these default settings proceed as follows:
1. When the main screen comes up and states “SELECT MODALITY/PROGRAM” push the SELECT
PARAMETERS button repeatedly to display the factory settings. You will then be able to adjust the
selected factory settings by pressing the TIME/PARAMETER up/down button. The new settings will be
saved from use to use as long as you do not remove the batteries or allow a full discharge of the unit.
®
500 comes with the following factory settings:
Button Speed, A & B Output 87
Button Speed, Beat, etc. 93
Button Delay, all 40
Audio ON
Screen Contrast 5
2. The screen contrast sets the appropriate viewing angle for easy screen legibility.
3. The button speed for output A & B sets the speed at which output will increase and decrease when you
push the buttons. This should generally be set at a slower speed (80 - 90) so as not to startle the patient.
4. The Button speed BEAT Etc. controls the scrolling speed of all of the remaining buttons on the system. A
speed of 90 to 95 is typical if you have good coordination and hand control.
5. The button delay represents the amount of time that you have to push and hold the buttons down before
they will begin to increment automatically at the button speed. The smaller the number the faster the
response time.
6. Audio on or off disables or enables the audio system that beeps at the end of the treatment and during the
NMES programs.
NOTE: The factory settings cannot be accessed during operation and are only available following the start
up screen. The settings will reset to their initial d efault parameters if the ba tteries are removed or the unit
is fully discharged. To restore the new settings follow the procedure in Appendix I.
COPYRIGHT 2001 - 2007, ACCELERATED CARE PLUS CORP., ALL RIGHTS RESERVED
CONFIDENTIAL AND PROPRIETARY
OMNISTIM® 500A USER MANUAL 11
Battery Charger / Power Supply Operation
Before using the OMNISTIM
1. Verify that rechargeable batteries are installed within the OMNISTIM
2. Check to see that the battery type selector switch located inside the battery compartment has been switched
to the rechargeable position.
If you plug in the charger, after two seconds, the charge LED on the OMNISTIM
batteries are fully charged, the LED blinks. If you plug in the charger during treatment or when the unit is ON the
LED turns off. (The charging continues but at a trickle and does not cause the LED to glow.) When you turn OFF
the unit after two seconds the LED comes on to announce charging.
NOTE: If non-rechargeable batteries are used:
1. Do not use the charger.
2. Make sure that the rechargeable/non-rechargeable selector switch in the bottom of the battery
compartment is in the non-rechargeable position.
Life expectancies of batteries under nominal (20 mA IFC, 40 mA MFAC, 100v HVPC) load conditions:
COPYRIGHT 2001 - 2007, ACCELERATED CARE PLUS CORP., ALL RIGHTS RESERVED
CONFIDENTIAL AND PROPRIETARY
12 OMNISTIM
Operational Sequence
SET OUTPUT A/B
A:00 mA B:00 mA r2.0
®
500A USER MANUAL
1. Press the ON/OFF button to turn on the power. The display panel will illuminate. There will be no
stimulation current and the unit will be in the SELECT MODALITY/PROGRAM mode.
2. At this time select the treatment modality desired and the appropriate treatment protocol by using the
modality buttons (2) to select HVPC, LVPC, MFAC or IFC protocols.
3. Now position the electrodes on the patient, connect the electrode plug(s) and press the START/STOP
button (3/5). The program parameters will disappear and be replaced with the screen seen above.
The bar graph on the lower right of the display shows the relative outputs of the two channels primarily
with respect to Time On and Time Off and Vector operation. The numeric displays on the lower left show
the output intensities into a 500-ohm load. Each time a treatment is commenced, the output current level is
reset to zero; this ensures that your patient cannot receive an initial shock due to the current output,
inadvertently, being left at a high setting from a previous treatment. The patient safety switch is a remote
duplicate of the Start/Stop switch and therefore resets the outputs to zero.
4. Adjust output by pressing the up/down (4) for channel A and channel B. Once desired output intensity is
selected, press START/STOP button (3/5) to display timer and to initiate treatment.
5. Press the SELECT PARAMETERS button to modify any settings. Press the MODIFY TIME / PARAM up or
down buttons to adjust the desired treatment time, or other settings, if you wish to change from the default
settings. Set the timer for the desired treatment time. Pressing the switch up or down, increases or decreases the
duration from a minimum of 1 minute to a maximum of 60 minutes. The timer will count down to zero during
treatment; when zero is reached a warning tone is emitted from the unit, the treatment current is switched off and
Treatment Off is displayed. Pressing the START/STOP button (3/5) will shut off the warning tone. If the unit is
left unattended, the warning tone is emitted for ten seconds before power is automatically switched off. Note that
the treatment time is preset for each program but may be adjusted at will during the treatment.
NOTE: When selecting the MFAC or LVPC alternate or delayed modes the system will allow first set up of
output A followed by set up of output B. Pushing the START/STOP SET output again will start the u nit
timing cycles if applicable in the program.
6. During operation the parameters may be altered within the program at will by selecting the set mode. See
set mode operation and adjustable parameters for each modality.
7. Once the treatment is completed, remove the electrodes from the patient and disconnect the lead wires from
the electrodes.
8. Turn the unit off by pressing the power button.
COPYRIGHT 2001 - 2007, ACCELERATED CARE PLUS CORP., ALL RIGHTS RESERVED
CONFIDENTIAL AND PROPRIETARY
OMNISTIM® 500A USER MANUAL 13
TREATMENT GUIDELINES
Introduction to Medium Frequency Currents
Medium frequency (MF) currents may be described as electrical currents applied to the body for therapeutic
purposes, which fall in the range of 1000 to 10,000 cycles per second (Hz). This is in contrast to low frequency
currents (0.1 to 1000 Hz) such as LVPC, HVPC, and high frequency currents (1 million Hz and beyond), which
include ultrasound, shortwave and microwave diathermy. Medium frequency currents are very advantageous for
clinical use due to their ease of skin penetration at lower intensities.
Normal human skin reacts differently to different frequencies of current. Specifically, there is an inverse
relationship between the frequency of the applied current and the skin’s resistance to it. Medium frequency
alternating currents in the range of 1000 or 5000 Hz provide markedly lower resistance to penetration than low
frequency electrical stimulation commonly used in TENS, LVPC, and HVPC stimulation. Medium frequency
currents can be used in Bipolar or in Quadripolar Interferential mode for patient treatment.
With medium frequency currents, the energy of each individual pulse is low providing for stimulation of only one or
two neurons. Since the pulses are coming in very rapid succession, stimulation of surrounding neurons occurs prior
to completion of the previous neurons refractory period. This allows for asynchronous activation of individual
sensory neurons, mimicking the natural physiologic process of the intact nervous system. This is not the case with
low frequency stimulators (0.1 to 1000Hz), which are capable of only stimulus synchronous neural activation.
Medium frequency currents provide rapid analgesic effects. This occurs due to rapid depolarization of nonmyelinated pain-transmitting fibers, which block pain transmission, further contributing to high muscle contraction
capabilities. Additionally, medium frequency currents have been shown to alter the vascular dynamics affecting
local and possibly systemic blood flow to the muscle(s) being stimulated. The unique characteristics of medium
frequency currents, (higher percent duty cycle, higher average current intensity, and wider pulse widths), can
significantly increase blood flow by altering the metabolic activity of muscles.
PLOT IMPEDANCE AS A FUNCTION OF FREQUENCY USING EPIDUCTIVE SYSTEMS
COPYRIGHT 2001 - 2007, ACCELERATED CARE PLUS CORP., ALL RIGHTS RESERVED
CONFIDENTIAL AND PROPRIETARY
14 OMNISTIM
®
500A USER MANUAL
Electrode Application Techniques
The following electrode placement diagrams are a theoretical representation of treatment set-ups using the
OMNISTIM
Electrode placement is dependent on the etiology of the condition.
Monopolar (Mono-Polar) Technique
This technique may use two electrodes of different sizes. The smaller, or “active” electrode can be positioned over
the segmental innervation or peripheral nerve path of the involved tissue, or over a distal location overlying any
muscle that is not an antagonist to the muscle being stimulated.
®
500. Monopolar, bipolar and quadripolar techniques are illustrated for various areas of the body.
CERVICAL MONOPOLAR
PLACEMENT OF
ELECTRODES
Bipolar (Bi-Polar) Technique
This is the most commonly used technique for muscle stimulation. This technique utilizes two electrodes but not
exclusively of the same size. One electrode should be applied over the motor point and the other electrode over the
belly of the muscle as far away from the motor point as possible. This technique allows for more effective muscle
and nerve fiber recruitment since the entire neural innervation of the muscle is furnished with current.
CERVICAL BIPOLAR
PLACEMENT OF
ELECTRODES
COPYRIGHT 2001 - 2007, ACCELERATED CARE PLUS CORP., ALL RIGHTS RESERVED
CONFIDENTIAL AND PROPRIETARY
OMNISTIM® 500A USER MANUAL 15
Quadripolar (Quadri-Polar) Technique
This technique requires the use of two output channels and four electrodes usually, but not exclusively of the same
size. The two electrodes from one channel are usually placed diagonally across the tissue area or joint to be treated
with the second channel electrodes placed on the opposite diagonal. This ensures that the current will intersect and
thus provide and interferential pattern.
CERVICAL QUADRIPOLAR
PLACEMENT OF
ELECTRODES
COPYRIGHT 2001 - 2007, ACCELERATED CARE PLUS CORP., ALL RIGHTS RESERVED
CONFIDENTIAL AND PROPRIETARY
16 OMNISTIM
®
500A USER MANUAL
Treatment Preparation
Skin Inspection
Thoroughly cleanse the treated area with soap and water to remove oils, creams, dirt, and sweat; this will ensure
uniform current conduction across the skin. After cleansing, inspect and evaluate the skin’s integrity and sensation
prior to treatment. Avoid absent or diminished sensation; if unavoidable, treat with caution. Establishment of
acceptable intensity levels for desensitized areas may be related to the intensity levels tolerated on normal skin in
opposite or related body parts. Frequently monitor the intensity level and skin response during all treatments.
Stinging, burning or other painful sensation under the electrodes on normal or desensitized areas is an indication of
increased current density under part or all of the electrode surface. In this case, slowly but immediately reduce the
current intensity to zero; remove the electrodes to inspect the surface skin. Recheck your application techniques.
Immediately after treatment, clean and thoroughly inspect the skin under the electrode. Peripheral vasodilation along
with systemic vasomotor responses can lead to redness (hyperemia) directly under both electrodes. Inform the
patient of this normal after effect and that the redness will disappear within an hour or two. Apply topical agents to
the reddened area under the electrodes if needed to decrease post-treatment irritation. Persistent skin irritation could
be due to repeated stimulation of the same electrode site or a possible allergic reaction to the conductive mediums,
tapes, elastic wraps, and/or cleaning and disinfectant solutions. Therefore, use additional electrode stimulation sites
to decrease or eliminate skin irritation on electrically sensitive patients. If skin irritation persists with alternate site
applications, decrease the treatment times and lower the intensities; if necessary, discontinue treatment. If an
allergic reaction is suspected, attempt to identify and change the allergic substance(s). If skin irritation persists,
discontinue treatment until the source of irritation is determined.
By far the most common causes which lead to machines being incorrectly reported as faulty are inadequate or
improper conductive medium interface or lead wire breakage. Because of the increased current density available
with pulsed or continuous medium frequency currents, a proportionally greater degree of conductive medium
interface problems exists and should be monitored by the clinician.
ACP Reusable Electrodes
Remove the electrodes from their foil packaging. Cleanse the skin then apply the electrodes over the treatment site
points according to the electrode placements techniques described in this manual. Various sizes of electrodes are
available dependent upon muscle size of the area to be treated. Follow the enclosed infection control procedures.
Review the warnings and application directions on the electrode packaging.
NOTE: The use of conductive mediums other than specifically approved pre-gelled or self adhering electrodes such
as ultrasound gel or lotion, hand or body lotion, electrolyte spray mist, paper towels, non-approved reusable or
disposable pre-gelled or self-adhering electrodes—are contraindicated for use with OMNISTIM
Lead Wires
Inspect the full length of the lead wires for signs of frayed or cut wires and loose connections where the lead wires
join the stereo jack plug and tip pins. Insert the stereo plug completely. Allow the lead wires to hang freely with no
excessive strain on the stereo plug insulator.
Periodically check the lead wires by checking their conductivity with an Ohmmeter. With the lead wire pin tips
together, measure the resistance between the tip and ring conductors of the stereo plug. The resistance should
measure very near to zero. If the resistance remains near to zero when the pin tips are disconnected from each other,
a short exists in the stereo plug insulator; replace the lead wire set.
®
systems.
COPYRIGHT 2001 - 2007, ACCELERATED CARE PLUS CORP., ALL RIGHTS RESERVED
CONFIDENTIAL AND PROPRIETARY
OMNISTIM® 500A USER MANUAL 17
INFECTION CONTROL EQUIPMENT AND PRINCIPLES OF USE
Definitions
•Barrier Film – One-time use, disposable plastic film for use over touch/operator surfaces of equipment to
reduce risks of cross-contamination and need for high level disinfection of equipment between patients.
wipe for use on electrotherapeutic devices and accessories.
•Plastic Lead Wire Sleeve – Barrier to be used on electrical stimulation lead wires, covering the junction of
lead wire and electrode wire.
Universal Precautions – Body Substance Isolation
Universal Precautions (UP) must be implemented in the care of all patients to protect employees from occupational
exposure to bloodborne pathogens. Personal protective equipment (gloves, masks, gowns) should be available and
worn by staff when occupational exposure to blood, body fluids containing blood, semen and vaginal secretions is
likely to occur. Health care workers with exudating lesions or weeping dermatitis should refrain from all direct
patient care and from handling patient care equipment until the condition resolves. Equipment must be
cleaned/disinfected and protective barriers used when appropriate.
Cleaning/Disinfecting of the OMNISTIM® 500
Modality equipment shall be cleaned/disinfected per facility infection control policy. ACP recommends the
following guidelines:
Cleaning and Low Level Disinfection
This is a recommended daily housekeeping practice to keep the equipment clean and free of contaminants which
could contribute to transmission of infection. The following practices are recommended for use when treating intact
skin without the presence of physiologic fluids such as blood and urine.
•Clean equipment daily with ACP germicidal wipes. At the end of the day, wipe common contact surfaces,
such as control panel and lead wires, with germicidal disposable wipe and allow to air dry. This technique
will inactivate M. tuberculosis as well as most bacteria and viruses. This will also facilitate removal of
organic material contaminants from equipment.
•Disposable/reusable electrodes are for individual patient use only and should not be used on multiple patients.
Intermediate Level Disinfection and Barriers
This method is recommended to keep the equipment clean and free of contaminants when used between patients for
treatment of non-intact skin or incontinence management, where there is an increased risk of patient crosscontamination. The following are the recommended practices:
•After each use, clean common contact surfaces, such as control panel and lead wires with ACP germicidal
wipes.
•With a second ACP germicidal cloth, wipe again leaving surfaces wet for at least 5 minutes. Allow the
surface to air dry before patient use.
•Barriers should also be used on the equipment for treatment of non-intact skin or incontinence
management. This technique will inactivate M. tuberculosis as well as most bacteria and viruses.
COPYRIGHT 2001 - 2007, ACCELERATED CARE PLUS CORP., ALL RIGHTS RESERVED
CONFIDENTIAL AND PROPRIETARY
18 OMNISTIM
®
500A USER MANUAL
Use of Barriers – Intermediate Level Disinfection
The use of an all-purpose barrier film provides surface protection from cross-contamination resulting from a variety
of applications. This procedure should be used whenever dealing with non-intact skin or the chance of coming in
contact with bodily fluids. Barrier film is designed to cover any surface that may be touched during a patient
treatment to help prevent cross-contamination. Barrier film is for single-use only. The film is discarded after each
patient treatment. The procedure for use is as follows:
1. Wash hands.
2. Apply intermediate level disinfection prior to barrier application.
3. Select, tear and place a length of barrier film to fit over the
operatory surfaces of the OMNISTIM
4. Select and cut with clean scissors a 2-foot length of plastic sleeve and fit
over the lead wire and the electrode cabling.
5. Prepare any items which may come in contatct with the therapist during
treatment, such as pens, assessment tools, cart handles, etc.
6. Set up the patient per guidelines for the procedure.
7. Provide treatment as appropriate.
8. Discard all disposables.
9. With clean gloves, remove the plastic film from the OMNISTIM
and discard.
10. Remove the plastic sleeve from the lead wire by sliding it toward the
electrode. Remove the electrode and discard with the sleeve.
11. Intermediate disinfect the OMNISTIM
treatment application.
®
500 unit.
®
500 unit prior to the next
®
500
COPYRIGHT 2001 - 2007, ACCELERATED CARE PLUS CORP., ALL RIGHTS RESERVED
CONFIDENTIAL AND PROPRIETARY
®
PROTOCOL / CLINICAL
INDICATION
MODES OF OPERATON: Protocol Reference Sheets
Interferential Current (IFC)
STIMULATION PARAMETERS MECHANISM OF ACTION APPLICATION TECHNIQUE OPERATIONAL SEQUENCE
OMNISTIM
500A USER MANUAL 19
COPYRIGHT 2001 - 2007, ACCELERATED CARE PLUS CORP., ALL RIGHTS RESERVED
CONFIDENTIAL AND PROPRIETARY
ANALGESIA - SENSORY
SEGMENTAL SUB ACUTE
Symptomatic relief of pain from
localized dermatome or
segmental origin.
ANALGESIA - MOTOR
NON - SEGMENTAL CHRONIC
Symptomatic relief of pain with
inflammation, and pain of
generalized or multi-segmental
nature.
ANALGESIA / MOTORSENSORY
SUB-ACUTE CHRONIC
Symptomatic relief of pain with
inflammation, and pain of local,
generalized single or multisegmental nature.
ANALGESIA SENSORYMOTOR
ACUTE SUB-ACUTE
Symptomatic relief of pain with
Inflammation, pain of local,
generalized single or multisegmental nature.
• 5000Hz 80/120 BPS continuous
• Vector Fast 90
• Treatment time of 15 minutes
• 2500Hz – 2/6 BPS continuous
• Vector Fast 90
• Treatment time of 15 minutes
• 5000Hz – 15/2/100 BPS
• Vector Fast 90
• Treatment time of 30 minutes
• 5000Hz – 100/2 BPS continuous
• Vector Fast 90
• Treatment time of 30 minutes
o
o
o
o
Sensory stimulation activates A-beta fibers
causing the release of spinal Enkephalin and
Dynorphin to block pain at the segmental
level (Gate Control). Duration of relief is
typically from 30 min to 2 hrs. Fast onset of
relief usually within 20 minutes.
Motor level stimulation activates motor and
A-delta fibers causing the release of
Bendorphin systemically. Duration of relief is
typically from 2-6 hours. Slow Onset of relief
usually within 30 minutes to 1 hour
Combines motor and sensory stimulation.
Starts with motor and ends with sensory.
More aggressive protocol. Duration of relief
is typically from 2-6 hours. Slower onset of
relief usually within 30 minutes.
Combines sensory and motor stimulation.
Starts with sensory and ends with motor.
Less aggressive protocol. Duration of relief
is typically from 2-6 hours. Faster onset of
relief usually within 20 minutes.
Target tissue - superficial and deep. Bilateral,
bipolar or quadripolar through the painful area or
over the involved spinal segments. Apply parallel
to incision line for post op pain management. Set
intensity to elicit a pleasant tingling sensation, just
below muscle contraction.
Target tissue - superficial and deep. Bipolar
placement over local and distal acu / trigger
points; quadripolar placement over area of local
pain; or quadripolar placement at spinal segment.
Set intensity to elicit a moderate muscle twitch.
Target tissue - superficial and deep. Bipolar
placement over local and distal acu / trigger
points; quadripolar placement over area of local
pain; or quadripolar placement at spinal segment.
Set intensity to elicit a moderate muscle twitch.
Target tissue - superficial and deep. Bipolar
placement over local and distal acu / trigger
points; quadripolar placement over area of local
pain; or quadripolar placement at spinal segment.
Set intensity to elicit a strong tingling sensation
just under muscle contraction.
• Press IFC button until screen reads "ANALGESIA - SENSORY"
• Press START / STOP until screen reads " SET OUTPUTS A-B"
• Set Outputs A and B until the patient feels the desired sensation.
• Press START / STOP and T15 will display in upper right corner.
• Adjust treatment time if necessary using the TIME/PARAMETER button
• The treatment will end at the end of the preset time or if the patient safety
switch is pressed.
• Press IFC button until screen reads "ANALGESIA - MOTOR"
• Press START / STOP until screen reads " SET OUTPUTS A-B"
• Set Outputs A and B until the patient feels the desired sensation.
• Press START / STOP and T15 will display in upper right corner.
• Adjust treatment time if necessary using the TIME/PARAMETER button
• The treatment will end at the end of the preset time or if the patient safety
switch is pressed.
• Press IFC button until screen reads "ANALGESIA/MOTOR - SENSORY"
• Press START / STOP until screen reads " SET OUTPUTS A-B"
• Set Outputs A and B until the patient feels the desired sensation.
• Press START / STOP and T30 will display in upper right corner.
• Adjust treatment time if necessary using the TIME/PARAMETER button
• The treatment will end at the end of the preset time or if the patient safety
switch is pressed.
• Press IFC button until screen reads "ANALGESIA/SENSORY - MOTOR"
• Press START / STOP until screen reads " SET OUTPUTS A-B"
• Set Outputs A and B until the patient feels the desired sensation.
• Press START / STOP and T30 will display in upper right corner.
• Adjust treatment time if necessary using the TIME/PARAMETER button
• The treatment will end at the end of the preset time or if the patient safety
switch is pressed.
ANALGESIA NERVE
BLOCK
LOCAL ACUTE
Symptomatic relief of pain from
localized dermal or segmental
origin.
• 5000Hz Continuous
• Vector OFF
• Treatment time of 10 minutes
Blocks pain by causing a temporary nerve
block through reactive depolarization
(Conduction block) of the pain signal on its
way to the spinal input. Also known as
Wedensky inhibition. Duration of relief is
typically from 1-2 hours Faster onset of relief
usually within 5-10 minutes.
Target tissue - superficial and deep. Bipolar
placement over local nerve; quadripolar placement
over area of local pain; quadripolar placement at
spinal segment. Set intensity to a numb-gripping
sensation just under muscle contraction.
• Press IFC button until screen reads "ANALGESIA NERVE BLOCK"
• Press START / STOP until screen reads " SET OUTPUTS A-B"
• Set Outputs A and B until the patient feels the desired sensation.
• Press START / STOP and T10 will display in upper right corner.
• Adjust treatment time if necessary using the TIME/PARAMETER button
• The treatment will end at the end of the preset time or if the patient safety
switch is pressed.
Loading...
+ 43 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.