Fluidized Temperature Baths
AP PL IC A T I O N S
Nitinol shape setting
General heat treatment of
devices & materials
Temperature sensor and
system calibration
Heating of reactors
Thermal analysis of
circuits & components
BE NE FI T S
Large working volume –
7.3”
diameter
24” depth
4000 and 6200 watt heat
capacity
Stability and uniformity
better than 0.3C
throughout operating
range
Designed and
manufactured in USA
CE marked
by 12” and
Fluidized Temperature Bath
• Fully automatic Fluidizing
air control
• Advanced 4 temperature zone PID
controller
• Cover and lid design that controls and
minimizes media loss
• Independent over-temperature limit
protection
• Included RS485 computer interface
with Windows PC control and
scheduling software
• Optional interface for control from any
PC on the corporate network or
internet
With over 25 years of experience with Fluidized Bath
technology we have developed a range of products that offer
outstanding performance, safety, features and value that are
unmatched. Unlike competing Fluidized baths our systems
have a smaller footprint, include advanced features and cost
thousands of dollars less and are much safer than salt baths
with thermal response that is 2 to 3 times faster than ovens.
Fluidized Temperature Baths have been the heat source of
choice for over 20 years by many leading Fortune 500
medical device manufacturers and research companies who
require fast heat up of their immersed devices and materials
with minimal quenching.
Unique Solutions for Thermal Applications
WH AT ’ S I N CL UD ED
-Fluidized Bath
-100 pounds of media
-Bath cover and Lid
-RS485 interface, cable and
software
-Instruction manual
WH AT D O I N EE D TO
RU N TH E S YS TE M
-240 VAC mains, 50/60hz 20 amp supply (
-30 amp supply
Clean dry air supply at a fixed
40 PSI, max flow of 3.5 CFM
FTBSL12)
(FTBSL26) -
WI NDOW S S OFTW ARE
The AccuTherm FTB
software allows the bath to
be operated automatically
on a weekly schedule with
daily set temperatures and
nightly down for operational
efficiency and energy
savings. The operating
setpoint can be changed at
any time from the PC.
SER VIC E S A VAI LAB LE
-Technical support
-Application support
-Installation and setup
-Maintenance
V7 715
Accurate Thermal Systems have developed an
innovative cover and lid system that prevents
media from escaping out of the bath opening
keeping the operating area cleaner and media
loss to a minimum. Media is either blocked by
the lid or collected on the cover flange which
drops back into the system.
The lid that is placed onto the cover contains a
handle that is elevated above the lid away
from heat. The lid can be readily modified by
customers to suspend devices or apparatus
for immersion of devices into the bath or can
be used with a parts basket. Thermo- wells
can also be added for calibration of
temperature sensors and systems.
Specifications
FTBSL12 FTBSL26
Temperature range 50 to 605°C
Working volume, inches 7.25 x 12 7.25 x 24
Typical stability at 500°C ±0.3 ±0.5
Calibrated accuracy - ±3.0 ±3.0
Heat up time to 600°C, 240V supply 150 minutes 180 minutes
Cool down time – max to 200°C 150 minutes 210 minutes
Power – 240 V, single phase, 50/60 hz. 4000 watts 6200 watts
Air pressure & flow required, max 40 PSI, 3.5 CFM
Overall footprint, H x W x D - inches 27 x 24 x 21 41 x 24 x 21
Total weight with aluminum oxide 210 lbs 320 lbs
Warranty 1 year
Windows PC software Included
Catalog number ATS2012 ATS2014
Recommended spare lid ATS1040 ATS1040
100 pound pail of Aluminum oxide ATS1027
USB to RS485 converter & Cable Included
Hainesport, NJ 08036 USA
Ph: 609-326-3190
Fax: 609-479-5124
Email: sales@accuthermal.com