Accumet 13636AP74A Operating Instructions Manual

OPERATING INSTRUCTIONS
accumet®13636AP74A
Portable Waterproof
Dissolved Oxygen
Printed 1/02
I
P
67
2
Table of Contents
1. Introduction ............................................................................................4
2. Display and keypad functions .............................................................5-6
2.1 Display...................................................................................................................................5
2.2 Keypad...................................................................................................................................6
3. Preparation..........................................................................................7-8
3.1 Inserting the batteries..........................................................................................................7
3.2 Connecting the probe...........................................................................................................8
4. Calibration.........................................................................................9-14
4.1 Preparing the meter for calibration ...................................................................................9
4.2 Dissolved oxygen calibration in % saturation..........................................................10-11
4.3 Dissolved oxygen calibration in mg/l (ppm) mode ...............................................12-13
4.4 Temperature calibration....................................................................................................14
5. Measurement...................................................................................15-18
5.1 Taking DO readings......................................................................................................15-16
5.2 Taking pressure/salinity compensated DO measurements...................................17-18
6. Hold function........................................................................................19
7. Memory functions............................................................................20-21
7.1 Memory input.....................................................................................................................20
7.2 Memory recall.....................................................................................................................21
8. Advanced setup functions...............................................................23-39
Advanced setup mode overview......................................................................................23-25
mg/l (ppm) measurement mode overview...................................................................24
% saturation measurement mode overview .................................................................25
8.1 Pr
ogram Clr: Memory clear..............................................................................................26
8.2 Pr
ogram dPr: Dissolved oxygen parameters............................................................27-28
Pressure adjustment mode................................................................................................27
Salinity adjustment mode .................................................................................................28
8.3 Pr
ogram OFS: Offset for % saturation measurement ...................................................29
8.4 Pr
ogram CAL: Previous calibration information ..........................................................30
3
8.5 Program ELE: Electrode properties.................................................................................31
View probe slope................................................................................................................31
View % saturation offset....................................................................................................31
View 100% saturation mV value ......................................................................................31
View 0% saturation mV value ..........................................................................................31
8.6 Pr
ogram COF: Unit configuration..............................................................................32-34
Ready indicator and auto endpoint function .................................................................32
Selecting mg/l or pm units ...............................................................................................33
Selecting °C or °F temperature readout..........................................................................34
8.7 Pr
ogram LCd: Adjusting LCD brightness......................................................................35
8.8 Pr
ogram CLO: Setting the real-time clock ................................................................36-37
8.9 Pr
ogram rST: Resetting to factory default settings .......................................................38
9. Probe care and maintenance...........................................................39-42
9.1 Probe care............................................................................................................................39
9.2 Membrane replacement ...............................................................................................40-41
9.3 Electrolyte solution ............................................................................................................41
10. Troubleshooting ..................................................................................42
11. Error Messages....................................................................................43
12. Specifications......................................................................................44
14. Appendix 1: Meter theory ..................................................................45
15, Appendix 2: Pressure vs. Altitude table.............................................46
16. Appendix 3: Factory default settings.................................................47
4
1. Introduction
Thank you for selecting an accumet AP74 Dissolved Oxygen meter. This portable meter is a microprocessor-based instrument that measures dissolved oxygen and temperature. It’s completely waterproof—and it floats! Your meter has many user-friendly features, all of which are accessible through the membrane keypad.
Your meter includes a dissolved oxygen/temperature probe, electrolyte solution, and batteries. Please read this manual thoroughly before operating your meter.
5
2. Display and Keypad Functions
Display
The LCD has a primary and secondary display.
• The primary display shows the measured dissolved oxygen reading.
• The secondary display shows the temperature of the reading. The display also shows error messages, keypad functions and program functions.
2.1
Primary display
5.
6.
Secondary display
3.
2.
17.
18.
16.
10.
14.
13.
12.
8.
1.
15.
4.
11.
7.
1. SETup mode indicator
2. MEASurement mode
indicator
3. CALibration indicator
4. MEMory recall mode
indicator
5. % saturation indicator
6. Salinity input indicator
7. Oxygen solubility
indicators
8. Temperature
indicators
9.Automatic Temperature
Compensation indicator
10. Clock indicator
11. ERRor indicator
12. MEMory location
indicator
13. Low battery indicator
14. Probe indicator
15. Calibration data
indicator
16. Function ON/OFF
indicator
17. HOLD indicator
18. READY indicator
9.
6
Keypad
The large membrane keypad makes the instrument easy to use. Each button, when pressed, has a corresponding graphic indicator on the LCD.
ON/OFF............Powers and shuts off the meter.
HOLD ..............Freezes the measured reading. To activate, press HOLD while in
measurement mode. To release, press HOLD again.
NOTE:
When auto endpoint feature is switched on, it automatically holds reading after 5 seconds of stability. The HOLD indicator appears on the display. Press HOLD to release auto endpoint feature.
MODE..............Selects the measurement parameter. Press MODE to toggle between
DO % saturation; DO mg/l (ppm); and date/time. In DO calibration mode, press MODE to access temperature calibration.
NOTE:
To switch between mg/l and ppm modes, see page 33.
CAL/MEAS......Toggles user between Calibration and Measurement mode.
• If you were in DO % saturation mode, press CAL/MEAS to enter
DO % saturation calibration mode.
• If you were in DO mg/l (ppm) mode, press CAL/MEAS
to enter DO mg/l (ppm) calibration mode.
NOTE: Temperature calibration is available from DO calibration mode; see page 14 for directions. In advanced set-up mode: Press CAL/MEAS to return to main menu from sub menus. Press CAL/MEAS again to return to measurement mode from main menu.
ENTER ............Press to confirm values in Calibration mode and to confirm
selections in Setup mode.
▲▼ ..........Press in Setup mode to scroll through subgroups. Also lets you
increment/decrement the values in the mg/l and temperature calibration modes.
..........................MI/MR work in the
measurement mode. MI (memory input) stores the measured value into memory. MR (memory recall) recalls the sets of values stored in the memory.
SETUP..............Press to enter SETUP
mode. SETUP mode lets you customize meter preferences and defaults, and view calibration and probe data.
..........................LIGHT Press to activate
backlit display.
MI MR
2.2
7
Battery compartment
Remove these two screws to access battery compartment
3. Preparation
Inserting the Batteries
Four AAA batteries are included with your meter.
1. Use a Phillips screwdriver to remove the two screws holding the battery cover.
See figure below.
2. Lift off battery cover to expose batteries.
3. Insert batteries. Follow the diagram inside the cover for correct polarity.
4. Replace the battery cover into its original position. Screw cover back into place.
3.1
8
Connecting the probe
Your meter includes a dissolved oxygen/temperature probe. The probe cable has a notched 6-pin connector to attach the probe to the meter.
NOTE: Do not substitute other probes. NOTE: Keep connector dry and clean. Do not touch connector with soiled hands. To connect the probe:
1. Line up the notch and 6 pins on the top of the meter with the holes in the probe
connector. Push down and turn the locking ring to lock into place.
See figure below.
2. To remove probe, turn the locking ring on the probe connector.
Pull probe away from the meter.
CAUTION: Do not pull on the probe cord or the probe wires might disconnect.
3.2
locking ring
top view of 6-pin
connector
9
4. Calibration
Probe guard
Preparing the meter for calibration
Before calibrating your meter, make sure to rinse the probe well with de-ionized water or rinse solution.
NOTE: Do not let membrane surface of the probe
touch any other surface. The probe guard (the piece with holes fitted over the end of the probe) protects the membrane; make sure this is always attached to the probe while it is in use.
You can calibrate for dissolved oxygen in either % saturation or mg/l (ppm). All new calibrations
automatically override existing calibration data.
Before calibration
Before starting calibration, make sure you are in the correct measurement mode. When you switch on the meter, the meter starts up in the units last used (either mg/l, ppm, % air saturation, or clock). For example, if you shut the meter off in “mg/l” units, the meter will read “mg/l” units when you switch the meter on.
NOTE: Most users will calibrate to 100% saturation
even when working in mg/l. When calibrat­ing the meter in mg/l mode, you can make fine adjustments, typically to a midrange value between 10 and 100%. If you are calibrating to a midrange value, you need to set the 100% saturation value first.
4.1
To calibrate 100 % saturation:
1. Rinse the probe well with deionized
rinse water or rinse solution. For best accuracy, wrap the end of the probe in a damp cloth. Do not touch the membrane.
2. Press the MODE key to select the
% saturation mode.
See figure
3. Press the CAL key. The CAL indicator
will appear above the primary display. The primary display shows the current value of the measurement and the secondary display will show 100.0 (see NOTE above).
See figure
4. Hold the probe in the air (or in damp
cloth). Wait for the reading to stabilize. If the Ready indicator feature is enabled, it will appear when the reading is stable (see page 32).
5. Press the ENTER key. The meter auto-
matically calibrates to 100.0% air satura­tion and returns to Measurement mode.
See figure
10
4.2
A
A
B
C
B
C
Dissolved Oxygen calibration in percent saturation
You can calibrate this meter quickly and easily in air. The exact calibration value depends on barometric pressure. The meter is set to a factory default of 760 mm Hg, which results in a calibration value of 100% saturation in air.
NOTE: If the barometric pressure setting has been changed from 760 mm Hg, the
calibration value in air will automatically adjust to a value other than 100%. The adjusted value will be correct for the new barometric pressure setting.
• See page 27 to change the pressure setting.
• See Appendix 2 on page 46 for a chart of adjusted % saturation values.
Notes
The reading in the primary display in step 3 must read at 50% or above for the calibration to work correctly. Whenever an error occurs during calibration, the ERR indicator appears in the lower left hand corner of the display.
You can offset your % DO calibration: see page 29 for directions.
To calibrate 0 % saturation:
1. Rinse the probe well with deionized
rinse water or rinse solution. Do not touch the membrane.
2. Press the MODE key to select the
% saturation mode.
See figure
2. Place the probe in zero oxygen
solution. Stir the probe gently to homogenize the sample. Make sure that the sample is continuously flowing past the membrane sensor.
3. Press the CAL key. The CAL indicator
will appear above the primary display. The primary display shows the current value of the measurement and the secondary display will show 0.0.
See figure
4. Wait for the reading to stabilize. If the
Ready indicator feature is enabled, it will appear when the reading is stable (see page 32).
5. Press the ENTER key. The meter auto-
matically calibrates to 0.0% saturation and returns to Measurement mode.
See figure
Notes
The reading in the primary display in step 3 must read at 10% or under for the calibra­tion to work correctly. Whenever an error occurs during calibration, the ERR indicator appears in the lower left hand corner of the display.
You can offset your % DO calibration: see page 29 for directions.
11
D
D
E
E
F
F
12
To calibrate in mg/l (ppm) mode:
1. Calibrate 100% saturation per section 4.1.
2. Rinse the probe well with deionized
rinse water or rinse solution. Wipe the outside of the probe carefully (do not touch the membrane).
3. Dip the probe into a sample of known
oxygen concentration. Wait for the reading to stabilize. If the Ready indicator feature is enabled, it will appear when the reading is stable (see page 32).
NOTE: The sample must continuously flow
past the membrane for accurate readings. Astirrer can keep the solution agitated; order a magnetic stirrer on page 47.
4. Press the MODE key to select the
mg/l (ppm) mode.
See figure
5. Press the CAL key. The CAL indicator
will appear above the primary display. The primary display shows the current value of the measurement and the sec­ondary display shows the temperature.
See figure
6. Press the and keys to adjust
the reading.
7. Press the ENTER key. The meter
automatically calibrates to the value you entered and returns to Measurement mode.
See figure
4.3
A
A
B
C
C
B
Dissolved Oxygen calibration in mg/l (ppm) mode
Calibrating the meter to 100 % in saturation mode will also calibrate the mg/l mode at the value in mg/l corresponding to 100% saturation. This should produce acceptable results in most applications.
This meter also lets you make a calibration adjustment in mg/l (ppm) mode without affecting your % saturation calibration.
To select between mg/l and ppm units, see page 33.
13
Notes on mg/l (ppm) mode calibration
During mg/l (ppm) calibration, the meter adjusts to the barometric pressure value that is programmed into the meter. It also adjusts to the salinity value that is pro­grammed into the meter.
You can change the barometric pressure value and salinity value in the mg/l (ppm) Set Up mode (see pages 27-28 for directions).
Whenever an error occurs during calibration, the ERR indicator appears in the lower left hand corner of the display.
14
Temperature Calibration
The built-in temperature sensor included in the probe is factory calibrated. Calibrate your sensor only if you suspect sensor drift that may have occurred over a long period of time or if you have a replacement probe.
1. Switch the meter on. Press MODE to
select mg/l (ppm) Measurement mode.
2. Press the CAL/MEAS key to enter
mg/l (ppm) calibration mode. The CAL indicator will appear above the primary display.
See figure
3. While in mg/l (ppm) calibration mode,
press the MODE key to enter tempera­ture calibration mode. The primary
display shows the last set temperature reading and the secondary display shows you the factory default temperature value.
See figure
4. Compare the primary display reading
to a NIST-traceable thermometer or another thermometer known to be accurate.
5. Press the or keys to adjust
the primary display reading to agree with your temperature standard.
See figure
6. Press the ENTER key to confirm
temperature calibration and return to Measurement mode.
Notes
To exit from Temperature Calibration mode without confirming calibration, DO NOT
press ENTER in step 6. Press CAL/MEAS instead.
Temperature calibration is restricted to ±5°C from the factory default value displayed during calibration (shown in the
secondary display).
A
B
C
4.4
A
B
C
15
5. Measurement
Taking DO readings
During measurement, the probe can be:
• fully immersed in the solution
• partially immersed in the solution Do not allow the probe’s membrane surface to touch any-
thing! The probe guard (the piece with holes fitted over the end of the probe) protects the membrane; you should leave this piece attached to the probe at all times.
IMPORTANT: since the DO probe consumes oxygen from the sample, the sample must constantly flow past the membrane to achieve more accurate readings. You can use a stirrer to accomplish this.
To take measurements:
1. Rinse the probe well with deionized
rinse water or rinse solution.
2. Select the appropriate measurement
mode. Press the MODE key to toggle between modes:
• %saturation.
mg/l (ppm)**,
• date/time clock
3. Dip the probe into the sample. Stir the
probe gently to homogenize the sample. Make sure that the sample is continuously flowing past the membrane sensor.
4. Note the reading on the display. If the
READY indicator is switched on, it will appear when the reading is stable.
See figure
5. Press the key to activate the backlit
LCD. This helps you read measurements in dimly lit areas. The backlit display automatically turns off 20 seconds after releasing the key to conserve batteries.
5.1
A
A
Probe guard
**To select between mg/l and ppm units, see page 33. The ATC indicator should appear in the lower right hand corner of the display.
If it does not, this indicates an error.
CONTINUED ON NEXT PAGE
16
Taking measurements with READY indicator selected on
If the READY indicator has been activated, the READY annunciator lights when the reading is stable. Switch the READY indicator on or off in the Set up program— see page 32 for directions.
Taking measurements with the auto endpoint feature selected on
When a reading is stable for more than 5 seconds, the auto endpoint feature will automatically “hold” the reading. The “hold” indicator appears on the left side of the display. Press the HOLD key to release the reading. Switch the Auto endpoint feature on or off in the Set up program—see page 32 for directions.
17
Taking pressure/salinity compensated DO measurements
If necessary, you can adjust the pressure and salinity values of your measurements in the Set Up mode. The DO meter will automatically compensate for salinity and pressure based on the values entered in the setup functions. The meter is factory set at 760 mm Hg (101.3 Pascals) pressure adjustment and a factor of 0.0 salinity adjust­ment. See Appendix 2 on page 46 for a “Pressure vs Altitude” table.
Pressure adjustment
1.
Press the SETUP key to enter Set Up mode.
2. Press the key once. The upper display
shows “dPr”.
See figure
3. Press the ENTER key until the upper
display shows a number and the lower display shows “Hg” or “PA”.
4. Use the and keys to enter the
barometric pressure. The upper display will show the value you have entered.
See figure
5. Press ENTER to confirm pressure value.
6. Press CAL/MEAS to return to
Measurement mode, or continue with step 4 on page 18 to make a salinity adjustment [available from mg/l (ppm) mode only].
5.2
A
B
A
B
18
Salinity adjustment
NOTE: this mode appears in ppm (mg/l)
measurement mode only.
1. Press the Mode key to select mg/l (ppm)
mode.
2. Press the SETUP key to enter Set Up
mode.
3. Press the key until the upper display
shows “dPr”.
See figure
4. Press the ENTER key until the upper
display shows a number and the lower display shows “SAL”.
5. Use the and keys to enter the
salinity of your solution in ppt. The upper display will show the value you have entered.
See figure
6. Press the ENTER key to confirm value.
7. Press CAL/MEAS to return to
Measurement mode.
A
B
A
B
19
This feature lets you freeze the dissolved oxygen and temperature readings for a delayed observation. HOLD can be used any time when in MEAS mode.
1. To hold a measurement, press the
HOLD key while in measurement mode. "HOLD" will appear on the display.
See figure
2. To release the held value, press HOLD
again. Continue to take measurements.
NOTE: This meter shuts off automatically
after 20 minutes of nonuse. If the meter is shut off either automatically or manually, the HOLD value will be lost. For longer storage, use the memory functions (see pages 20-21).
NOTE: This meter has an auto endpoint
feature. When this feature is switched on, the display will automatically “hold” a reading that has been stable for more than 5 seconds. The “hold” indicator appears. Press the HOLD key to release the reading. To switch on or off the auto endpoint feature, see page 32.
A
A
6. HOLD function
20
7. Memory functions
Memory Input
Your meter stores data in sets that include:
• dissolved oxygen with corresponding temperature reading
• time and date reading was taken
You can store up to 50 sets of data in any combination of %, mg/l, or ppm readings.
7.1
A
To store a reading:
1. During any measurement function
(MEAS), press the MI key to input any data into the memory.
2. MEM, “Sto” and memory number
will flash. The meter then returns to measurement mode.
See figure
NOTE: If the memory is full, the first value
stored will be erased to create space for the new value.
A
21
To recall readings:
1. Press the MR key once to retrieve the
last reading stored. The memory location screen—MEM, “Loc” and the memory number—will flash on the display.
See figure
2. Press the ENTER key to recall the read-
ing stored under that memory number.
See figure
3. Press the ENTER key again to view the
date and time the reading was taken.
See figure
4. Press the ENTER key again to return to
the “memory location” screen. The display automatically moves to the next memory location screen.
See figure
5. If necessary, press the key to select
the next “memory location” screen; press the key to select the previous “memory location” screen.
6. Repeat steps 2-5 to review additional
stored data sets.
7. To exit Memory Recall, press the MEAS
key to return to the Measurement mode.
Memory Recall
This function recalls the previous readings stored in the memory. You can only access MR from the MEASurement mode. Memory recall is in “Last In First Out” order.
7.2
B
C
C
D
E
E
D
Notes
Readings stored in memory are retained even if the unit is turned off. To erase all readings stored in memory, use the SETUP mode Clr on page 26.
B
22
23
8. Advanced set up functions
The advanced set up mode lets you customize your meter ’s preferences and defaults. Your AP74 waterproof meter features different sub groups that organize all set-up parameters.
This meter blanks out subgroups that do not apply to the measurement mode [% saturation or mg/l (ppm)] you are in when you enter Setup mode.
The full selection of available sub groups are:
1. CLR: Memory clear
2. dPr: Dissolved oxygen parameters
3. OFS: % saturation offset adjustment—
% saturation mode only
4. CAL: Calibration data
5. ELE: Probe data
6. COF: Configuration
7. LCd: Backlit display
8. CLO: Setting clock
9. rSt: Reset meter to factory default
IMPORTANT: If a parameter is not available in a particular mode, the meter automatically skips past it. The parameter numbers in the lower display adjust accordingly.
If you enter Set up mode from the date/time measurement screen, you will see % saturation set up parameters.
See pages 24-25 for more a detailed overview on the different parameters available in % saturation mode and mg/l (ppm) mode.
available in
% saturation
mode only
24
mg/l (ppm) measurement mode
Advanced set-up mode detailed overview
CLr: Memory clear
• Clear all stored readings
dPr: Dissolved oxygen parameters
• Select Hg or Pa barometric pressure units
• Select barometric pressure
• Select salinity adjustment factor
CAL: Viewing previous calibration data
• View previous calibration data, including date/time
ELE: Viewing probe data
• View probe slope
• View mV value equivalent to 100% saturation
• View mV value equivalent to 0% saturation
Instructions on page 26
8.1
COF: Unit configuration
• Ready indicator on or off / auto endpoint on or off
• Select mg/l or ppm units
• Select °F or °C
LCd: Backlit display
• Adjust brightness of backlit LCD
CLO: Set clock
• Setting year
• Setting date (month/day)
• Setting time (hour/minute/second)
rSt: Reset to factory default
• Reset to factory default settings
Instructions on pages 27-28
Instructions on page 30
Instructions on page 31
Instructions on pages 32-34
Instructions on page 35
Instructions on pages 36-37
Instructions on page 38
Press the SETUP key to enter Set up mode. Press the ▲ and ▼ keys to scroll through sub groups.
% saturation measurement mode
CLr: Memory clear
• Clear all stored readings
dPr: Dissolved oxygen parameters
• Select Hg or Pa barometric pressure units
• Select barometric pressure
CAL: Viewing previous calibration data
• View previous calibration data, including date/time
OFS: % saturation offset adjustment
• Set % saturation offset adjustment
ELE: Viewing probe data
• View probe slope
• View % saturation offset
• View mV value equivalent to 100% saturation
• View mV value equivalent to 0% saturation
Instructions on page 26
COF: Unit configuration
• Ready indicator on or off / auto endpoint on or off
• Select °F or °C
LCd: Backlit display
• Adjust brightness of backlit LCD
CLO: Set clock
• Setting year
• Setting date (month/day)
• Setting time (hour/minute/second)
Instructions on pages 27-28
Instructions on page 29
Instructions on page 30
Instructions on page 31
Instructions on pages 32-34
Instructions on page 35
Instructions on pages 36-37
rSt: Reset to factory default
• Reset to factory default settings
Instructions on page 38
25
Clr: Memory Clear
Use this sub group to clear all memory values when you need to store a new series of values. This lets you avoid confusing the old values with the new ones. NO is th e default setting.
NOTE: Selecting YES will wipe out all memory.
8.2
A
A
B
B
From measurement mode:
1. Press the Set up key to enter Set Up
mode.
2. Press the and keys to scroll through
subgroups until you view parameter ”Clr”in the upper display.
See figure
3. Press the ENTER key to enter parameter.
4. Press the and keys to toggle
between NO and YES.
• NO retains current memory
• YES clears all memory
See figure
5. Press the ENTER key to confirm selec-
tion and return to the subgroup menu. Press the CAL/MEAS key to return to measurement mode.
26
27
dPr: Dissolved Oxygen Parameters
This sub group lets you adjust the barometric pressure and salinity.
Pressure adjustment mode
Barometric pressure is vital to correct dissolved oxygen measurements. You need to enter the correct barometric pressure of the area you are measuring. This mode lets you perform two functions:
• Select either mm Hg or Pascal barometric pressure units
• Adjust the barometric pressure. See Appendix 2 on page 46 for a “Pressure vs Altitude” table.
8.3
From measurement mode:
1. Press the Set up key to enter Set Up
mode.
2. Press the and keys to scroll
through subgroups until you view parameter “dPr” in the upper display.
See figure
3. Press the ENTER key twice. The upper
display shows either Hg or PA and the lower display shows bAr.
4. Press the and keys to toggle
between mm Hg and Pascal units.
See figure
5. Press the ENTER key to confirm
selection and move to the next screen. The upper display shows the barometric pressure and the lower display shows the units selected in step 3.
See figure
6. Press the and keys to adjust
the barometric pressure. The pressure
adjustment range is 500 to 1499 mm Hg
(66.6 to 199.9 kPa).
7. Press the ENTER key to confirm
selection and move to step 5 on page 28 (salinity adjustment). If instead you want to return to measurement mode, press CAL/MEAS twice.
NOTE: Salinity adjustment appears in mg/l (ppm) measurement mode only.
A
A
B
B
C
C
28
Salinity adjustment mode
NOTE: this mode appears in mg/l (ppm)
measurement mode only.
Salinity correction mode lets you correct for the variations in oxygen solubility due to salt concentration in the sample.
From measurement mode:
1. Press the Mode key to select mg/l
(ppm) mode.
2. Press the Setup key to enter Set Up
mode.
3. Press the and keys to scroll
through subgroups until you view parameter “dPr” in the upper display.
See figure
4. Press the ENTER key five times. The
upper display shows the salinity value and the lower shows SAL.
See figure
5. Press the and keys to enter the
correct salinity adjustment factor. The salinity adjustment factor range is
0.0 to 50.0 ppt.
6. Press the ENTER key to confirm
selection and to move back to subgroup “dPr”. If you want to return to measure­ment mode, press CAL/MEAS.
D
D
E
E
29
OFS: Offset for % saturation measurement
NOTE: this sub group appears in % saturation measurement mode only.
Use the offset adjustment to act as an offset at values between 0 and 100%. It is useful to match against end-user standards (i.e. a test kit value).
8.4
From measurement mode:
1. Press the Mode key to select %
saturation mode.
2. Press the Set up key to enter Set Up
mode.
3. Press the and keys to scroll
through subgroups until you view parameter “OFS” in the upper display.
See figure
4. Press the ENTER key. The upper
display shows the current measurement in % saturation and the lower shows OFS.
See figure
5. Press the and keys to offset the
% saturation measurement.
6. Press the ENTER key to confirm
selection and to move back to subgroup “OFS”. If you want to return to measure­ment mode, press CAL/MEAS.
A
A
B
B
30
CAL: Previous calibration information
This sub group shows you the previous calibration data, along with date and time of calibration. This is a “view only” parameter.
In % saturation mode: calibration information is shown in % saturation units. In mg/l (ppm) mode: calibration information is shown in mg/l (ppm) units.
8.5
From measurement mode:
1. Press the Mode key to select the
calibration data you want to view: % or mg/l (ppm).
2. Press the Set up key to enter Set Up
mode.
3. Press the and keys to scroll
through subgroups until you view parameter “CAL” in the upper display.
See figure
4. Press the ENTER key. The upper
display shows the calibration data.
See figure
5. Press the ENTER key again. The display
shows the date and time of the last cali­bration.
See figure
6. Press the ENTER key to move back to
subgroup “CAL”. If you want to return to measurement mode, press CAL/MEAS.
NOTE: If you did not calibrate this meter in a particular mode, the screen will show “– – –”.
See figure
A
A
B
B
C
C
D
D
31
From measurement mode:
1. Press the Mode key to select the meas-
urement mode for the electrode proper­ties you want to view: % or mg/l (ppm).
2. Press the Set up key to enter Set Up
mode.
3. Press the and keys to scroll through
subgroups until you view parameter “ELE” in the upper display.
See figure
4. Press the ENTER key. The upper
display shows the probe slope.
See figure
5. Press the ENTER key. The upper display
shows the % saturation offset. See direc­tions for setting this offset on page 29. NOTE: this appears only in % saturation measurement mode. If you are in mg/l (ppm) mode, the meter skips to step 6.
See figure
6. Press the ENTER key. The upper display
shows the 100 % saturation mV value.
See figure
7. Press the ENTER key. The upper display
shows the 0 % saturation mV value.
See figure
8. Press the ENTER key to move back to
subgroup “ELE”. If you want to return to measurement mode, press CAL/MEAS.
ELE: Electrode properties
These “view only” parameters show the electrode properties for diagnostic purposes:
A. probe slope:
lets you view the calibration slope of the probe. This mode displays
slope from 0.5 to 1.999 % of slope (1.0 = 100%).
B. % saturation of
fset (available in % saturation mode only): lets you see the value of
the % saturation offset entered in parameter “OFS” (see page 29 for instructions).
C. 100% saturation mV value:
lets you view the sensor ’s mV output corresponding
to 100% saturation.
D. 0% saturation mV value:
lets you view the sensor ’s millivolt output corresponding to 0% saturation.
8.6
B
B
C
C
A
A
D
E
D
E
From measurement mode
1. Press Setup key to enter Set Up mode.
2. Press the and keys to scroll through
subgroups until you view parameter COF in the upper display.
See figure
3. Press the ENTER key to select
parameter rdY (Ready).
See figure
4. Press the and keys to select the
configuration you require.
•ON switches the READY indicator on.
•OFF switches the READY indicator off.
• ON and HOLD together switches the auto endpoint feature on.
5. Press the ENTER key to confirm selection
and to proceed to:
• in % mode: step 4 on page 34.
• in mg/l (ppm) mode: step 3 on page 33.
Or, press the CAL/MEAS key twice to return to measurement mode.
Notes
Meter default is set for Ready indicator on, and auto endpoint function off.
32
COF: Unit configuration
Unit configuration mode lets you select the following parameters:
A. Ready indicator and auto endpoint function B. mg/l or ppm units (available from mg/l or ppm mode only) C. Temperature in °C or °F
READY indicator and auto endpoint function
The first program lets you select:
“READY
indicator on” to indicate when the reading is stable.
“READY
indicator off” for faster meter response.
Auto endpoint function on.
Select auto endpoint on to “hold” the reading when it is stable for more than 5 seconds. The display automatically freezes, and the HOLD indicator appears on the left side of the display. Press the HOLD key to release the display and access other functions.
8.7
A
A
B
B
33
Selecting mg/l or ppm units
NOTE: this mode appears in mg/l (ppm) measurement mode only.
This mode lets you select between mg/l or ppm dissolved oxygen units.
From measurement mode
1. Press the Mode key to select mg/l
(ppm) mode.
2. Press Setup key to enter Set Up mode.
3. Press the and keys to scroll through
subgroups until you view parameter “COF” in the upper display.
See figure
4. Press the ENTER key until “dO”
appears in the upper display.
See figure
5. Press the and keys to toggle
between mg/l or ppm units.
6. Press the ENTER key to confirm
selection and move to step 4 on page
34. Press the CAL/MEAS key twice to return to measurement mode.
C
D
C
D
34
Selecting °C or °F temperature readout
This meter lets you select between °C and °F units for temperature readings.
From measurement mode
1. Press Setup key to enter Set Up mode.
2. Press the and keys to scroll through
subgroups until you view parameter “COF” in the upper display.
See figure
3. Press the ENTER key until “C” or “F’
appears in the upper display.
See figure
4. Press the and keys to toggle
between °C and °F.
5. Press the ENTER key to confirm
selection and to return to the subgroup menu. Press the CAL/MEAS key to return to measurement mode.
E
E
F
F
35
LCd: Adjusting LCD brightness
This mode lets you adjust the brightness of the backlit LCD. Selecting a dimmer backlighting level helps conserve batteries.
8.8
From measurement mode
1. Press Setup key to enter Set Up mode.
2. Press the and keys to scroll through
subgroups until you view parameter “LCd” in the upper display.
See figure
3. Press the ENTER key. A number (0-8)
appears in the upper display and “LCd” appears in the lower display.
See figure
4. Press the and keys to select
from level 0 (dimmest light) to level 8 (brightest light).
5. Press the ENTER key to confirm selec-
tion and to return to the subgroup menu. Press the CAL/MEAS key to return to measurement mode.
A
A
B
B
36
CLO: Setting the real-time clock
Your meter features a real-time calendar and clock. This helps you meet GLP (Good Laboratory Practice) standards.
8.9
A
B
From measurement mode
1. Press Setup key to enter Set Up mode.
2. Press the and keys to scroll through
subgroups until you view parameter “CLO” in the upper display.
See figure
3. Press the ENTER key to enter
parameter “CLO”. The meter lets you select the century: “19-” or “20-”. The century digits will flash.
See figure
4. Press the and keys to toggle to the
correct century.
5. Press the ENTER key to confirm the
century and move to “year” selection. The “year” digits will flash.
See figure
6. Press the and keys to toggle to the
correct year.
7. Press the ENTER key to confirm the
year and move to “month” selection. The “month” digits will flash.
See figure
8. Press the and keys to toggle to the
correct month.
9. Press the ENTER key to confirm the
month and move to “date” selection. The “date” digits will flash.
See figure
10. Press the and keys to toggle to
the correct date.
CONTINUED ON NEXT PAGE
A
B
C
C
D
D
E
E
37
11. Press the ENTER key to confirm the
date and move to “hour” selection. The “hour” digits will flash.
See figure
12. Press the and keys to toggle to
the correct hour. Note the “AM” and “PM” indicator on the lower portion of the display.
13. Press the ENTER key to confirm the
hour and move to “minute” selection. The “minute” digits will flash.
See figure
14. Press the and keys to toggle to
the correct minutes.
15. Press the ENTER key to confirm the
minutes and move to “second” selec­tion. The “second” digits will flash.
See figure
16. Press the and keys to toggle to
the correct seconds.
17. Press the ENTER key to confirm selec-
tion and to return to the subgroup menu. Press the CAL/MEAS key to return to measurement mode.
Notes
Press the CAL/MEAS key at any point while setting the time to return to the subgroup menu.
F
F
G
G
H
H
38
P6.0: Resetting to factory default settings
This program lets you reset all parameters to factory default settings. This clears all calibration data, memory, and any other setup functions you might have changed. However, the following settings will remain as you have set them:
• temperature unit of measure (°C or °F)
• The temperature offset calibration value
• the LCD contrast setting
• the clock setting.
A
B
8.8
From measurement mode
1. Press Setup key to enter Set Up mode.
2. Press the and keys to scroll through
subgroups until you view parameter “rSt” in the upper display.
See figure
3. Press the ENTER key.
See figure
4. Press the and keys to toggle
between NO and YES.
• NO retains current settings
• YES resets to factory default settings
5. Press the ENTER key to confirm
selection and to return to measurement mode.
Notes
See page 47 for a table of factory default settings.
A
B
39
9. Probe care and maintenance
Probe care
The probe is a galvanic measuring element which produces an output proportional to the oxygen present in the medium in which it is placed. The galvanic probe design lets you take measure­ments immediately—without the typical 15 minute wait of other dissolved oxygen probes.
The probe consists of two parts:
• an upper part consisting of an anode, a cathode, and cable
• a lower part consisting of a membrane cap, membrane, and electrolyte solution.
Under typical operating conditions, the probe should last for several years. Proper care and
maintenance will help you receive the maximum probe life and ensure more accurate readings.
Since any deposits on the membrane surface act as a barrier to oxygen diffusing through the mem­brane, the membrane must be cleaned at regular intervals to assure maximum reliability.
After using the probe, rinse the probe in clean water and wipe it with a soft cloth or paper to avoid any hardening of deposits. If growth develops on the probe, use a disinfecting chemical to clean.
NOTE: Although the membrane is strong and
not easily damaged, wipe it gently while cleaning it. If the membrane is damaged or torn, the probe will no longer function.
There are no special probe storage requirements.
Anode
Membrane cap
(holds membrane
and electrolyte
solution)
Probe guard
Cathode
9.1
40
Temperature
sensor
Anode
Membrane
lock
Membrane
O-ring
Membrane
cap
Probe guard
Cathode
Membrane replacement
Replacement of the membrane is required only when you cannot calibrate the probe, or if the membrane is damaged.
Typical membrane damages are punctures or wrinkles caused during measurements or cleaning.
To replace the probe membrane:
1. Pull off the probe guard.
2. Unscrew the membrane cap from the probe.
3. Hold the probe under a water tap and brush
away the white oxide on the cylindrical anode with a stiff plastic brush—do not use metal cleaning material.
4. It the cathode has any deposits, remove them
with a light scouring powder. Do not polish the cathode.
NOTE: If you have purchased a replacement
membrane module with preinstalled membrane, skip to step 12.
CONTINUED ON NEXT PAGE
9.2
A
41
Installation tool
Carefully place O-ring, then membrane, into membrane cap.
Insert notches on installation tool into slots on membrane lock. Then, unscrew membrane lock from membrane cap.
B
C
5. Using the installation tool, unscrew and
remove the membrane lock from the membrane cap.
See figure
6. Remove the membrane and O-ring.
Discard both.
7. Rinse the membrane cap and membrane lock
in tap water.
8. Install a new O-ring inside the
membrane cap.
9. Install a new membrane. Make sure the
membrane covers the O-ring all around its circumference.
See figure
10. Using the installation tool, screw the
membrane lock back into the cap. Tighten the lock firmly over the membrane and O-ring, but do not overtighten.
11. Inspect the membrane for wrinkles.
If wrinkles exist, remove the membrane and repeat steps 8-11.
12. Fill the membrane cap with water and
inspect the bottom for leaks. If water drops are leaking from the membrane, re-seat the membrane on the O-ring (repeat steps 8-11).
13. If the assembly is leak-free, empty the water
and fill the membrane cap with electrolyte to the brim.
14. Screw the cap onto the probe.
Excess electrolyte will drain out.
15. Replace probe guard.
16. Calibrate the probe (see section 4) after the
% saturation readings have stabilized.
B
C
42
10. Troubleshooting
Problem Cause Solution Power on but a) Batteries not in place. a) Check that batteries are in
no display place and making good
contact.
b) Batteries not in correct b) Reinsert batteries with
polarity (+ and –). correct polarity.
c) Weak batteries. c) Replace batteries.
Unstable a) Insufficient electrolyte a) Fill probe with reference readings in probe. electrolyte.
b) Dirty probe. b) Clean the probe and
recalibrate.
c) External noise pickup c) Move or switch off
or induction caused by interfering motor. nearby electric motor.
e) Broken probe. d) Replace probe.
Slow response a) Dirty membrane. a) Clean probe. See "Probe
Care & Maintenance", page 40.
Not responding a) HOLD mode in operation. a) Cancel HOLD mode. to key press
b) Damaged key pad. b) Return to dealer. c) Internal program error. c) Reset all internal programs
by reinserting batteries.
43
11. Error Messages
LCD
Indicates Cause Solution
Display
Err Unrecognized input Wrong input in Release key. Select annunciator from keypad selected mode. valid operations
depending on mode.
CAL & Err Calibration error Wrong value input Check your input value, annunciators at calibration. clean probe. blink Dirty probe. See Calibration sections or
Probe Maintenance section.
Battery Low battery level Need new batteries Clean battery contacts. indicator blinks or battery Replace batteries with fresh
connection is bad. ones, noting polarity.
Err. 1 Memory write Hardware failure. Turn meter on and off (in primary error again. If message persists, display) return unit*.
If an error message appears in the primary display (the upper row of larger digits), switching off the meter and switching it on again may eliminate the error message.
See figure at right. If error persists, or the meter shows incorrect
values, return the meter.
For a complete diagram of the display, see page 4.
ERR 1 in primary display
44
12. Specifications
Mode mg/l (ppm) % saturation Temperature
Range 0.00 to 19.99 mg/l (ppm) 0.0 to 199.9%
0.0 to 50.0˚C /
32.0 to 122.0°F
Resolution 0.01 mg/l (ppm) 0.1% 0.1˚C or °F Accuracy ±1.5% of full scale ±1.5% of full scale ±0.3˚C / ±0.6 °F
Temperature compensation: automatic from 0 to 50°C Salinity correction:
Range: 0.0 to 50.0 ppt Resolution: 0.1 ppt Method: Key in manually and automatic correction
Barometric pressure correction:
Range: 500 to 1499 mm Hg / 66.6 to 199.9 kPa Resolution: 1 mm Hg / 0.1 kPa Method: Key in manually and automatic correction
Probe: galvanic probe Memory: stores and recalls up to 50 points with date/time Real-time clock: stamps calibration data and stored data with time and date
(month and day)
Display: dual LCD Operating temperature: 0 to 50°C Power: four 1.5 V AAA batteries (included) Battery life: up to 100 hours continuous use ( depends on use of backlit display) Dimensions:
Meter: 7.5"L x 3.5"W x 1.75"H (19.1 cm x 8.9 cm x 4.5 cm) Boxed: 9.2"L x 9.2"W x 2.75"H (23 cm x 23 cm x 7 cm) Probe: 6.8"L x 1.3" dia (17.3 cm L x 3.2 cm dia)
Weight:
Meter: 1.0 lb (0.5 kg) Boxed: 2.0 lbs (0.9 kg)
45
14. Appendix 1: Meter Theory
Dissolved oxygen levels in natural and waste waters depend on the physical, chemical, and biochemical activities in the water body.
This meter uses a galvanic probe. It consists of a cell that contains electrolyte and that is enclosed by a selective membrane, and two metallic electrodes. The mem­brane is practically impermeable to water and ionic dissolved matter, but is perme­able to oxygen and a few other gases.
The cathode consumes the oxygen passed through the membrane, and produces an electric current in the probe. This current is proportional to the partial pressure of oxygen in the sample.
Since the cathode consumes the oxygen in the sample, it is essential that the fluid must flow past the sensor to maintain accurate readings.
The solubility of oxygen in water varies with barometric pressure and temperature, and decreases as salinity increases. For the most accurate DO readings, you need to compensate for these factors. This meter automatically compensates for tempera­ture readings. It also allows you to enter a salinity correction factor and the baromet­ric pressure to correct for this variability.
46
15. Appendix 2: Pressure vs. Altitude table
Barometric pressure affects DO readings, therefore this meter lets you enter the cor­rect barometric pressure at your altitude. If you do not have equipment that lets you measure the exact barometric pressure at your altitude, you can estimate it using the chart below.
If you change the barometric pressure setting from its factory setting (760 mm Hg), the % saturation calibration value in air will automatically adjust to a value other than 100% (see “corrected % saturation value” column below). The adjusted value is correct for the new barometric pressure setting.
See page 27 for information on how to adjust the barometric pressure.
Pressure vs Altitude
Altitude (ft) Pressure (mm Hg) Corrected % saturation value
0 (Sea level) 760 100
500 746 98.1 1000 732 96.3 1500 720 94.7 2000 707 93.0 2500 694 91.3 3000 681 89.6 3500 668 87.8 4000 656 86.2 4500 644 84.6 5000 632 83.0 5500 621 81.6 6000 609 80.0
47
16. Appendix 3: Factory Default Settings
Resetting the meter to factory default settings clears all calibration data and memory, and returns other setup functions to the default settings shown in the table below. However, the following settings will remain as you have set them:
• temperature unit of measure (°C or °F)
• The temperature offset calibration value
• the LCD contrast setting
• the clock setting. See page 38 for directions on setting the meter to factory default settings.
Program Function Options/settings Default
Clr Memory clear yes/no no dPr Pressure units Hg or Pa Hg
Barometric pressure adjust from 500-1499 mm Hg 760 mm Hg
Salinity adjustment adjust from 0-50 ppt 0 ppt OFS Set % saturation offset up to ±10% no offset CAL View calibration data – – – ELE View probe slope 1.000
View % saturation offset 0 % offset
View mV = 100% saturation 37.0 mV
View mV = 0% saturation 0.3 mV COF Ready indicator/auto endpoint on/off Ready on only
Select mg/l or ppm units mg/l or ppm mg/l
Select temperature units °C or °F retains settings LCd Adjust backlit display levels 0 to 8 (brightest) retains settings CLO Setting clock year, date, time retains settings rST Reset to factory default yes/no no
Loading...