AccuMac AM8060 User Manual

Page 1
AM8060 Precision Thermometer
User’s Guide
Page 2
AM8060 Precision Thermometer User’s Guide
1/43
Table of Contents
1.1 Main Application ................................................................................................ 3
1.2 Main Features...................................................................................................... 3
2.1 Specifications ...................................................................................................... 5
2.2 Probe Types ........................................................................................................ 5
4.1 Connecting the probes......................................................................................... 7
4.2 Instrument connection with AC power ............................................................... 7
4.3 Instrument front panel display and touchpad. ..................................................... 7
4.4 Selecting Units .................................................................................................... 9
4.5 Selecting Sensors/Probes and Adjusting Parameters .......................................... 9
5 Instrument Calibration And Accuracy Adjustments ................................................. 17
5.1 100 Ω calibration .............................................................................................. 18
5.2 25 Ω calibration ................................................................................................ 20
5.3 Reset the measurement/calibration switch after calibration ........................... 23
6 Wireless Data Acquisition......................................................................................... 24
7 Using a Flash Drive(U disk)for Data Acquisition............................................... 30
8 Using the USB Interface for Data Acquisition ......................................................... 35
9 Appendix: Installation of USB Driver....................................................................... 39
Please read the
Important Notice”
information on the next page before turning on the instrument!
Page 3
AM8060 Precision Thermometer User’s Guide
2/43
Important Notice ---- Warning
Check the position of the “Measurement/Calibration
Switch” on the rear panel before turning on the device. Please refer to Figure 2. The switch must be at the measurement (left) position. Loss of calibration data will occur if the switch is at the calibration (CAL) position when the instrument is used in measurement application.
The instrument is calibrated before shipment. Only
qualified calibration agencies should use the switch when the instrument is under calibration. Please refer to 5.
Instrument Calibration And Accuracy Adjustments
Page 4
AM8060 Precision Thermometer User’s Guide
3/43
1 Introduction
The AM8060 precision thermometer is a dual-channel temperature readout. It can be used as either a temperature measurement device or a temperature reference standard. It is capable of working together with various types of platinum resistance thermometers (PRTs) and can measure temperatures from two probes simultaneously. The measured temperature is calculated by the embedded microcontroller by following ITS-90 standard. The device can also be equipped with a wireless data acquisition function, a USB data acquisition and a flash drive interface. With these flexible peripherals, temperature data can be easily captured and saved in real time.
1.1 Main Application
Temperature measurement with standard platinum
resistance thermometers (SPRTs), secondary platinum resistance thermometers (Secondary PRTs) and precision industrial platinum resistance thermometers (IPRTs).
Either temperature reference standard or temperature
measurement.
Temperature reference for dry-wells or temperature baths. Wireless/wire/flash real time temperature measure ment/data
capturing.
1.2 Main Features
High accuracy: 0.006 ºC at 0 ºC High resolution0.001 ºC Concurrent two PRTs measurement and data capturing. Display of two PRTs temperature in Celsius ºCºF,
resistance Ω and temperature difference.
2.7” OLED display. Touchpad for function control.
Page 5
AM8060 Precision Thermometer User’s Guide
4/43
Flexible probe/sensor parameter adjustments and settings. USB interface for transmitting and saving real time
temperature data.
Wireless real time temperature data capturing with optional
external wireless module.
Real time temperature data capturing and saving to flash
drive.
Page 6
AM8060 Precision Thermometer User’s Guide
5/43
2 Specifications
2.1 Specifications
Temperature range-200 ºC ~ 850 ºC Resolution0.001 ºC Accuracy
25Ω standard platinum temperature probe
100Ω standard or industrial platinum temperature probe
±0.01 ºC @ -196 ºC
±0.010 ºC @ -200 ºC
±0.008 ºC @ 0 ºC
±0.008 ºC @ 0 ºC
±0.009 ºC @ 232 ºC
±0.009 ºC @ 232 ºC
±0.015 ºC @ 420 ºC
±0.015 ºC @ 420 ºC
±0.025 ºC @ 660 ºC
±0.020 ºC @ 660 ºC
±0.025 ºC @ 850 ºC
Short term stability
o ±0.001 ºCwith 100Ω platinum probe o ±0.003 ºCwith 25Ω platinum probe
Working environment15ºC ~ 35 ºC Power supplyAC 100V ~ 220V Power consumption< 15 W Dimensions 180 mm x 65 mm x 200 mm Weight 0.7 lbs (300 g)
2.2 Probe Types
SPRT or PRT with temperature coefficient α=0.003925
(WGa≥1.11807).
IPRT with temperature coefficient α=0.00385.
Page 7
AM8060 Precision Thermometer User’s Guide
6/43
3 User Interface
Front Panel
Display Main Menu U Drive recording
Start/Stop button
Left key Value Adj Right/Return Key
Figure 1: Front Panel.
Rear Panel
AC Power Inlet/Switch Probe 1(S1) terminals
U Drive Interface USB Measurement Probe 2(S2) terminals
/Calibration Switch
Figure 2Rear Panel
Warning: unless calibration is needed, do NOT put the switch on ‘Cal’
Page 8
AM8060 Precision Thermometer User’s Guide
7/43
4 General Operations
The AM8060 precision thermometer is a dual-channel temperature readout. It can measure temperature of two probes concurrently. It can also be used with just one probe for single channel temperature measurement.
4.1 Connecting the probes
The AM8060 can work with one or two probes for temperature measurement. Figure 3 shows the rear panel of the AM8060. Probe 1 can be connected to terminals marked with S1, and Probe 2 to S2.
Figure 3: Rear Panel and Probe Terminals.
4.2 Instrument connection with AC power
AM8060 operation requires power of AC100V~220V.
4.3 Instrument front panel display and touchpad.
After the power is turned on and the probes are connected, the device should show initialization information, including the instrument interface status, measured temperatures of the
Set to CAL when doing Calibration.
Probes
Page 9
AM8060 Precision Thermometer User’s Guide
8/43
probe/probes and temperature difference. This information is shown in 4 lines, as seen below:
Line 1 is the instrument interface status, including USB flash drive, wireless communication status and USB to computer
interface status. If the “Udris shown in Line 1, a USB flash drive is connected to the USB host interface; if “ZB-xx (where “xx” are numbers) displays on Line 1 as shown above, it means the Zigbee wireless data capturing is working; if the display showsUSBon Line 1, then AM8060 is connected via its USB client interface.
Line 2, which is indicated with ‘t1’, shows the measured temperature of Probe 1 (S1). TheSPRT” at the end of the Line 2 tells the type of probe of Probe 1 (S1). The AM8060
supports two types of probes. “SPRT” indicates a 25Ω or 100 Ω standard platinum resistance thermometers. “IPRT” indicates a 100Ω industrial platinum resistance thermometer. The “C” shown at the end of Line 2 indicates that the temperature is in Celsius.
Line 3, which is indicated with ‘t2’, shows the measured temperature of Probe 2 (S2).
Page 10
AM8060 Precision Thermometer User’s Guide
9/43
A “>” will toggle in front of “t1” or ”t2” to indicate which probe is being measured. Each toggling instance is one measurement.
Line 4, which is indicated with “t1-t2”, shows the temperature difference of the two probes.
4.4 Selecting Units
Users can select to display ºC, ºF or Ω by scrolling touch keys up
or down ( or )
4.5 Selecting Sensors/Probes and Adjusting Parameters
4.5.1 Probe Types
AM8060 can work with different types of PRTs. Users can choose probe types and input the parameters of the probes.
The menu key
MENU
and return/right shift key can be used
for probe selection and parameter adjustment.
The probes compatible with AM8060 are:
25 Ω or 100 Ω SPRTs, or 25 Ω or 100 Ω Secondary PRTs, or 100 Ω IPRTs.
The temperature coefficient of these probes can either be
0.00385 Ω/Ω/°C or 0.003925 Ω/Ω/°C.
The probe types working with AM8060 are classified and displayed as IPRT and SPRT
Page 11
AM8060 Precision Thermometer User’s Guide
10/43
If the probe is an industrial or precision platinum
resistance thermometer probe with a temperature coefficient of 0.00385 Ω/Ω/°C, AM8060 will categorize the probe as “IPRT”. AM8060 calculates temperature according to CVD (IEC751) equation.
If the probe is a precision platinum resistance
thermometer probe or secondary PRT with temperature coefficient of 0.003925 Ω/Ω/°C (WGa>=1.11807), AM8060 will categorize the probe as “SPRT”. AM8060 calculates temperature according to ITS-90 standard platinum resistance thermometer equations.
If the probe is a standard platinum resistance thermometer,
the probe will be displayed as “SPRT”. AM8060 calculates temperature according to ITS-90 standard platinum resistance thermometer equations.
4.5.2 Selecting Sensors/Probe and Parameters Input
4.5.2.1 To select the type of sensor, press
MENU
key and then
hold key. After 3 seconds, AM8060 will display the parameter adjustment main menu:
There are four options for parameter adjustments:
Page 12
AM8060 Precision Thermometer User’s Guide
11/43
Sensor 1 Setting. This is for Sensor 1 (S1) parameter adjustment. Users can select the Sensor 1 type and do parameter adjustment.
Sensor 2 Setting. This is for Sensor 2 (S2) parameter adjustment. Users can select the Sensor 2 type and do parameter adjustment.
U-Drive Setting. Users can select this to change settings for saving data to an external drive.
Wireless Setting. Users can select this to change the settings for wireless communication.
Although only “Sensor 1 Setting” is illustrated below, the procedure of Sensor 1 and Sensor 2 adjustments are identical.
Page 13
AM8060 Precision Thermometer User’s Guide
12/43
4.5.2.2 Press or key to move the selection line with the
“>” symbol to “Sensor 1 Setting”. Press key to enter the Sensor 1 setting adjustment program as shown below. The probe type can either be “IPRT” or “SPRT”. Users can select either one by moving the “>” to the
sensor type, and then press key. Users can exit this
selection by pressing
MENU
key. The figure below shows
that SPRT is selected.
4.5.2.3 Press key to go to the parameter adjustment menu:
Page 14
AM8060 Precision Thermometer User’s Guide
13/43
4.5.2.4 If the sensor type is SPRT, the adjustable parameters are Rtp, a, b, c, a4 and b4. The details of these parameter definitions can be found from the ITS-
90.
As an example, calibration agencies provide following parameters after SPRT is calibrated:
Rtp: probe’s resistance at triple point of water. a4 and b4low temperature range parameters. a, b and chigh temperature range parameters (b and c can be “0” for some instances. (Please refer to ITS-90 for more
information.)
If the sensor type is IPRT, the adjustable parameters are R0, A, B and C:
The details of the parameter definitions can be found from the industrial platinum resistance thermometer calibration procedure.
As an example, calibration agencies provide following parameters after IPRT is calibrated:
R0: the probe’s resistance at 0°C. A and B: the temperature coefficients at temperature range above 0°C.
Page 15
AM8060 Precision Thermometer User’s Guide
14/43
C: the temperature coefficient at the temperature range below 0°C.
4.5.2.5 The
MENU
and keys can be used
together to adjust these parameters. The parameters can also be saved into the instrument. The following is the procedure of adjusting and saving the parameters of Rtp and a4 for an SPRT type probe. The other parameter adjustments follow the same process.
4.5.2.6 To adjust Rtp (continuing from step 4.5.2.4) press
or key to move the cursor to “>SET Rtp”, and
then press key to go into the Rtp adjustment program.
The instrument will display the current Rtp value. The digit underlined with the blinking bar indicates the adjustable
position. Users can use and keys to move this
position and use and keys to adjust the value.
4.5.2.7 After the Rtp is adjusted, press
MENU
key, the display
will remind user to save the adjusted value by flashing
the adjusted Rtp on screen. Press key to save the
Page 16
AM8060 Precision Thermometer User’s Guide
15/43
value or press
MENU
key to exit the program without
saving and go back to the parameter adjustment menu.
4.5.2.8 The following is a4 adjustment procedure
(Continuing from step 4.5.2.4). Pressing and keys
to move the cursor to “>SET a4”, and then press key to go into the a4 adjustment program, as seen below:
The second line (a4 = -4.129934e-05) displays the current a4 value, while the next two lines show the adjustment selections. The a4 adjustment is different from Rtp, as the value of a4 can be “0”. If a4 is “0”, the user can move the
cursor by using and keys to “SET A4=0” line and
press the key to confirm the selection. If a4 is a value other than “0”, the user can move the cursor to “ADJUST
a4” line and then press key to go into the a4 adjustment program.
Page 17
AM8060 Precision Thermometer User’s Guide
16/43
4.5.2.9 The current a4 value will be displayed. To adjust
this value, press or key to move the cursor to the position where the value will be adjusted and then adjust
it with or key.
4.5.2.10 After adjustment, users can choose to save or quit
by pressing the
MENU
key. The a4 line will be flashing to
remind the user that a change has been made. To save
the changes, press key; otherwise, press
MENU
key and
the program will go back to the previous screen.
4.5.2.11 Other parameter adjustments are similar to the above Rtp and a4 adjustment.
4.5.2.12 After all of the desired adjustments are made, users
can go back to the main menu by pressing
MENU
key multiple times until the AM8060 goes back to the main screen.
Page 18
AM8060 Precision Thermometer User’s Guide
17/43
5 Instrument Calibration And Accuracy
Adjustments
AM8060 is designed for temperature measurement by measuring the probe resistance and converting it into a temperature. Consequently, the resistance measurement is very critical, and the instrument must be calibrated correctly. Every AM8060 is calibrated before shipment.
AM8060 needs regular calibration performed by qualified professionals. One year calibration interval is recommended.
Users can calibrate AM8060 using standard resistors with calibrated values.
Warning: It is required to set correct R0 (Rtpw) values in Sensor 1 (S1) before calibrations. To calibrate AM8060 in the range from 10 Ω to 50 Ω, R0 (Rtpw) needs to be set to 25 Ω. To calibrate AM8060 in the range from 50 Ω to 375 Ω R0 (Rtpw) needs to be set to 100 Ω.
In case of AM8060 out of tolerance, its accuracy can be adjusted. Only qualified agencies, labs or institutes (who must be equipped with calibrated 25 Ω and 100 Ω standard resistors) should adjust the accuracy of AM8060.
The calibration should be done in an environment with room temperature of 22±2°C.
Instrument calibration procedure: (No special sequence for 100 Ω and 25 Ω calibration)
Page 19
AM8060 Precision Thermometer User’s Guide
18/43
5.1 100 Ω calibration
5.1.1 Turn off the power, put the measurement/calibration switch
to the “CAL”side (right side). Please see Figure 3 for details.
5.1.2 After turning on the powerAM8060 shows the calibration
main menu. It can be seen that there are two calibrations required: 100 Ohm CALIBRATION and 25 Ohm CALIBRATION.
The following is an illustration of 100 Ohm CALIBRATION.
* CALIBRATION MENU *
> 100 Ohm CALIBRATION
25 Ohm CALIBRATION
Enter = Yes
-------------------------------------------
5.1.3 Use or key on the keypad to select the calibration. A “>” sign will be displayed in front of the selected line. For example, the above figure shows that 100 Ohm
CALIBRATION has been selected. Pressing the key will start the calibration program, and the following display will
appear
Page 20
AM8060 Precision Thermometer User’s Guide
19/43
* 100 Ohm CALIBRATION
Connect a 100 Ohm Standard Resistor and Press Enter
Menu= Exit Enter = Yes
-------------------------------------------
5.1.4 The instrument will show the phrase “Connect a 100 Ohm Standard Resistor and Press Enter” to indicate that a 100
must be connected to sensor1 (S1) terminals as Figure3.
Users can exit the calibration by pressing
MENU
key to go back
to the main menu (*CALIBRATION MENU*).
5.1.5 After the probe is correctly connected, press key to start the calibration. The instrument will display “Waiting…” while the resistance is calibrated. When the process is finished, the screen will display the result as seen below:
* 100 Ohm CALIBRATION
100.0238
Rt = 111.0000 OHM
Menu= Exit <^v>=Adj
100R Rs = OHM
5.1.6 To calibrate the instrument, users must adjust the value (100.0238 as shown above) to match the standard resistor’s resistance. For example, if the resistance of the standard resistor is 100.0038 ohm, users can move the underline
cursor to the desired position using and keys, and
Page 21
AM8060 Precision Thermometer User’s Guide
20/43
users can adjust the value by using and keys. The adjusted result will appear like the picture below:
* 100 Ohm CALIBRATION
100.0038
Rt = 100.0038 OHM
Menu= Exit <^v>=Adj
100R Rs = OHM
5.1.7 It is required to wait for 10 seconds for the data to be stable. Each second is shown with a blinking dot at the top left hand corner of the display. Once this is completed, the next step involves saving the calibrated data. To save the data, press
the
MENU
key and the third line of data will blink to indicate
the option to save.
5.1.8 To save this calibration value, press and hold the key and
then press the key. This will store the calibration data to the instrument, and the program will go back to the main menu (*CALIBRATION MENU*). If users DO NOT want
to save the data, press the
MENU
key and bring the program
back to main menu ( *CALIBRATION MENU*).
5.2 25 Ω calibration
5.2.1 The process of 25 Ω calibration is similar to that of 100 Ω
calibrationn. First, turn off the power of the instrument and
Page 22
AM8060 Precision Thermometer User’s Guide
21/43
ensure the measurement/calibration switch at the rear panel is moved to the “CAL” side (right side), as shown in Figure 3.
5.2.2 Choose the 25 Ohm CALIBRATION under main menu
(*CALIBRATION MENU*) by touching or key to move the “>” to the “25 Ohm CALIBRATION” line
* CALIBRATION MENU
100 Ohm CALIBRATION
> 25 Ohm CALIBRATION
Enter = Yes
-------------------------------------------
5.2.3 Touch key to start the program of 25 Ohm CALIBRATION. The instrument will display the following message:
* 25 Ohm CALIBRATION
Connect a 25 Ohm Standard Resistor and Press Enter
Menu= Exit Enter = Yes
-------------------------------------------
5.2.4 Follow the “Connect a 25 Ohm Standard Resistor and Press Enter” instructions to connect a 25
Ω to
Probe 1 (S1). A standard resistor of 25Ω must be have resistance close to 25Ω, otherwise the calibration will not be accurate.
5.2.5 Touch key to start the 25
Ω calibration.
The instrument
will display “Waiting…” for a few seconds while the
Page 23
AM8060 Precision Thermometer User’s Guide
22/43
resistance is calculated, and the following message will appear:
25.9977
Rt =25.0000 OHM
Menu= Exit Enter = Yes
25R Rs = OHM
* 25 Ohm CALIBRATION
5.2.6 The value on the third line is the parameter to be adjusted and this value must match the standard resistance of the 2 5Ω resistor. For example, the above figure shows the current value as 25.9977, but the standard resistor’s value is 24.9977.
This value can be adjusted by using or key to move
the “_” (underscore) to the desired, and using or key to adjust the value. After the adjustment, the instrument will
display
24.9977
Rt =24.9977 OHM
Menu= Exit Enter = Yes
25R Rs = OHM
* 25 Ohm CALIBRATION
5.2.7 It is required to wait for 10 seconds to ensure the reading is stable. Each second is shown with a blinking dot at the top left hand corner of the display. The next step is to save the calibrated data to the instrument. To start the data saving,
Page 24
AM8060 Precision Thermometer User’s Guide
23/43
touch the
MENU
key, and the third line of data will blink to
indicate the option to save.
5.2.8 To save this calibration, press and hold the key and then
press the key. This will save the calibration data to the instrument, and the program will go back to the main menu
(*CALIBRATION MENU*)If users DO NOT want to
save the data, press the
MENU
key will bring the program back
to the main menu ( *CALIBRATION MENU*).
5.3 Reset the measurement/calibration switch after
calibration
5.3.1 After the calibration, turn off the AC power
5.3.2 Set the Measurement/Calibration switch at the rear panel to
the left side (away from “CAL”)
Page 25
AM8060 Precision Thermometer User’s Guide
24/43
6 Wireless Data Acquisition
The AM8060 precision thermometer can transmit temperature data through a wireless transmission module. Free software for this feature is included with the AM8060. To transmit data wirelessly, users will need the followings:
AM8060-W, precision thermometer with embedded
wireless function.
Wireless Transceiver . USB cable with length of 2m to 5m. Computer with USB interface and Windows operating
system.
Figure 4AM8060 Wireless Data Acquisition and Data Storage
Directions for setting up the wireless communication system:
1) The AM8060 precision thermometer must be on and connected to a probe.
Page 26
AM8060 Precision Thermometer User’s Guide
25/43
2) The computer and Wireless Transceiver must be within 50m (indoors) or 80m (outdoors) from the AM8060. The computer must be powered on.
3) Connect the Wireless Transceiver to the computer with a USB cable. Place it at a high spot for best performance. The system setup is shown in Figure4.
4) After connecting the Wireless Transceiver to the computer, the user will be required to install USB drivers unless the PC is running on Windows7 operating system. The detailed instructions for this are shown in the Appendix.
5) After the installation of the USB driver the user will need to install the free software on the CD included with the AM8060. Insert the CD into the computer’s CDROM drive and copy the file named CDrN_2xAir_Reader.exe to the computer.
6) Once the file has finished copying, double-click it to begin the installation. Follow the installation steps on screen to complete the process. After it has finished installing, double-click the icon “CDrN_2xAir_Reader”, and the following screen will display:
Page 27
AM8060 Precision Thermometer User’s Guide
26/43
7) “The Base Status” and “Remote” wil display “Wireless Base RF OK” and “Wireless Remote OK” to confirm the wireless connection. The value after “ADDR” shows the wireless address of the AM8060. The value after “RSSI” is the wireless signal strength, which will not display until the user clicks on “Start”. The RSSI should be greater than -99dBm. Otherwise, the wireless signal is too weak and the link is not stable. To increase the strength of the wireless connection, users can move the AM8060 closer to the computer or increase the height of the CoolDragon Air Base. If the signal is still too weak, the environment is not suitable for the wireless transmission of data.
8) To begin the wireless data acquisition, click on “Start”. If the wireless signal is good enough, the AM8060 should display “ZB-xx” where “xx” indicates the wireless signal strength. At the same time,
Page 28
AM8060 Precision Thermometer User’s Guide
27/43
CDrN_2xAir_Reader should also show the signal strength, as seen below:
9) The wireless data acquisition can be halted by clicking on “Stop”, or by clicking on “Exit” to quit the program. The default setting for a sampling period is 1 second. The data acquisition process can be restarted by clicking on “Start”.
10) CDrN_2xAir_Reader can also save the captured data to a file, entitled “Air_data.txt”. The data sampling rate can also be adjusted. The minimum sampling period is 1 second.
Users can adjust the data sampling rate by typing in an integer into the box “Every x Sec” below “Data Saving”. X indicates the sampling period in the unit of seconds. The sampling period can be adjusted only
Page 29
AM8060 Precision Thermometer User’s Guide
28/43
when the sampling is stopped. The data sampling can be restarted after the modification.
Users can save data by clicking on “Start”and then clicking on the “REC” button. The captured temperature data is saved into “Air_data.txt”, and the phrase “Recording…” will appear under “Data Saving” to indicate the recording process has begun. To stop data saving, clicking on “END” or “Stop” will cease the temperature data capture.
The following is an example with a 4 second sample period:
11) The data saved in the file Air2x_data.txt will be in the following format:
Page 30
AM8060 Precision Thermometer User’s Guide
29/43
Time t1(C) t2(C) t1-t2(C) 16:23:19 -4.526 -84.456 79.930 16:23:20 -4.526 -84.456 79.930 16:23:21 -4.526 -84.456 79.930 end recording
Time t1(C) t2(C) t1-t2(C) 10:48:45 -4.526 -84.456 79.930 10:48:49 -4.526 -84.456 79.930 end recording
The first column indicates the sampling time, which should be same as the computer time. The second column lists the Probe 1 temperatures, while the third column lists the Probe 2 temperatures. The fourth column is the difference between Probe 1 and Probe 2. Each time “Stop” is pressed, the data saving will be halted, and the phrase “end recording” will be added to the file. To continue data saving, users can click on “REC”, and new data will be added to the last line of the previous data. The above example shows data stored with two different sample periods: the first time with 1 second, the second time with 4 seconds.
Page 31
AM8060 Precision Thermometer User’s Guide
30/43
7 Using a Flash Drive
U disk
for Data
Acquisition
The flash drive recording start/stop key
REC
can be used for
saving data to the USB flash drive.
There is a USB port on the rear panel of the AM8060 which can be used to store the measured temperature data to a flash drive. Any flash drive with size lower than 4 gigabytes can be connected to the “UDrive” connector at the rear panel. The following steps explain how to transfer data from the AM8060 to a flash drive:
1) Connect a flash drive to the “UDrive” USB port at the rear panel of the AM8060.
2) When the flash drive is connected, the display will say “Device Detected” “D:/p” to indicate that the flash drive is detected. The top left hand corner of the display should display “UDr” to remind users that a flash drive is connected and is available for data capturing.
3) If the flash drive is connected while the AM8060 is turned off, AM8060 will display “Device Detected P2” during initialization to indicate that the flash drive is detected, and the upper left hand corner of the display will show “UDr” to remind users that a flash drive is connected and available for data capturing.
4) To save the temperature data measured by the AM8060
to the flash drive, just press the
REC
key on the front
panel. AM8060 will emit a short beep and the LED
Page 32
AM8060 Precision Thermometer User’s Guide
31/43
above the
REC
key will be turned on. This is to inform the user that the temperature data is being written to the flash drive. The saved file name is U2DATA.txt.
5) If users want to stop the data saving, just press
REC
key
on the front panel. AM8060 will produce a short beep
and the LED above
REC
key will be turned off to
indicate the process has been halted.
The following list is an example of data saved in the U2DATA.txt file on a flash drive. Time column indicates time in unit of seconds; t1, t2 and t1-t2 indicate temperature readings from sensor 1, sensor2 and difference of two sensors in Celsius. An “end” is
written to the file when
REC
key has been turned off.
Time t1 t2 t1-t2
0.23 -4.525 +529.475 -534.000
1.44 -4.524 +529.475 -534.000
2.64 -4.524 +529.475 -533.999
3.85 -4.524 +529.475 -533.999
5.05 -4.524 +529.475 -534.000
6.26 -4.524 +529.475 -534.000
7.46 -4.524 +529.476 -534.000
8.67 -4.524 +529.476 -534.000 end
6) The above operations can be repeated to do multiple time data saving.
7) The temperature data file U2DATA.txt in the flash drive can be transferred to a computer. Just unplug the flash drive when it is stopped (when the LED above the
REC
key is offand plug it into a computer’s USB
connector. Once the flash drive is detected by the computer’s OS, users can copy the file over normally.
Page 33
AM8060 Precision Thermometer User’s Guide
32/43
While using the flash drive, the data saving period can be adjusted with the front panel. The default saving period is 1.21 seconds and it can be set as a multiple of this value, with the multiple
ranging between 1 and 65,535. The instructions for this are as follows:
A. When AM8060 is in a normal working state, press
MENU
key
and then hold key for 3 seconds. The AM8060 will display the main menu as below:
B. Press key or key to move the cursor line with ”>”
sign to “ > U-Drive Setting”, then press key to go to the “Record Cycle adjustment” program, as seen below. “N” represents how often the temperature will be saved as a multiple of 1.21 seconds. The current value is “1”.
C. To set a sample frequency other than “1”, press and hold
key while pressing key, and a display similar to the one below will appear.
Page 34
AM8060 Precision Thermometer User’s Guide
33/43
It should be noted that this operation is a two-key
operation. Don’t release the key before touching
key.
D. Press key or key to adjust the “N” value (which must
be equal or greater than 1).
E. After the adjustment, press
MENU
key. The new adjusted value
will blink to indicate whether the user wishes to save this new N value. For examplethe above picture shows the new N value as 9, which means the AM8060 will save data to the
file every 9x1.21 seconds. To save this value, press and hold
key and then press key. If users DON’T want to save,
just press
MENU
key to exit.
Note: saving the N value is a two-key operationdo
not release key before pressing key. Otherwise the new value will not be saved.
Page 35
AM8060 Precision Thermometer User’s Guide
34/43
F. After the adjustment, press
MENU
multiple times until the
AM8060 returns to its normal working state.
Note: The parameter adjustments should only be performed while the flash drive is not in the data saving process.
Page 36
AM8060 Precision Thermometer User’s Guide
35/43
8 Using the USB Interface for Data
Acquisition
AM8060 precision thermometer is designed with a USB interface for temperature data acquisition and saving temperature data to a file.
The following section describes the process of setting up a USB temperature data acquisition system:
1) Connect a USB cable to the USB (USB B type) port at the rear panel of the AM8060 digital thermometer to a computer.
Cool Dragon
Digital Thermometer
PC with USB interface
Running CoolDragon USB
Reader
USB Cable
Figure 5AM8060 USB Temperature Data Acquisition System.
2) The first time the AM8060 digital thermometer is connected to a computer, the computer will require the installation of USB drivers. Detailed instructions for this
process are illustrated in the Appendix. After the
installation of these USB drivers, no further drivers will need to be installed.
AM8060 Precision Thermometer
Page 37
AM8060 Precision Thermometer User’s Guide
36/43
a. Copy the file “CDRN_USBx2_Reader.exe” from the
attached CD to the computer’s desktop.
b. Double click on “CDRN_USBx2_Reader” to start
the software. The following screen will appear
c. The measurement can be started by clicking on
“Start”, stopped by clicking on “Stop”, or quit by clicking on “Exit”. The default sampling period is 1 second.
The free software “CDRN_USBx2_Reader” also has a data saving function to save the captured temperature to a file, named “U_x2data.txt”. Users can also adjust the sampling period for their requirements. The shortest sampling period is 1 second. The instructions below detail how to adjust the sampling period.
Find the “Every x Sec” section of the “Data Saving” section at the right side of the software’s GUI. The unit of the sampling period is seconds. If a sampling period other than 1 second is
Page 38
AM8060 Precision Thermometer User’s Guide
37/43
required, users can adjust the value in the box. Any adjustment must be done when the measurement is in the “Stop” state as shown in the figure above.
To save the captured temperature data, click on “Start”. The temperature data should be displayed with a blinking tick mark to indicate the data capturing, then users can click on “REC” to start the data saving. The phrase “Recording…” will be displayed on the screen as the data is saved to the file, named “U_x2data.txt”. To stop the data saving, just click on “END” or “Stop”, and the temperature data saving will be stopped.
The following figure shows that the data is saved to file with a sampling period of 3 seconds.
Data saved in “U_x2data.txt” will appear in the following format:
Page 39
AM8060 Precision Thermometer User’s Guide
38/43
Time t1(C) t2(C) t1-t2(C) 23:50:57 -4.524 +529.476 -534.000 23:51:00 -4.524 +529.475 -533.999 23:51:03 -4.524 +529.475 -533.999 end recording
Page 40
AM8060 Precision Thermometer User’s Guide
39/43
9 Appendix: Installation of USB Driver
Important Update: if your PC is running Windows7 operation system, you can skip USB driver installation process. Otherwise, please follow the instructions below.
The installation of USB drivers are required the first time these instruments are connected to a computer.
The USB drivers for the AM8060 precision thermometer and Wireless Transceiver are the same. The following steps describe the installation process:
1. When AM8060 precision thermometer and Wireless
Transceiver are first connected to a computerthe OS will display the “Found New Hardware” wizard.
Page 41
AM8060 Precision Thermometer User’s Guide
40/43
2. Put the attached CD into the CDROM driveand then
select the “No, not this time” button and click on “Next”.
3. Select “Install from a list or specific location (Advanced)” and then click “Next”.
Page 42
AM8060 Precision Thermometer User’s Guide
41/43
4. Click “Browse” to select the CD drive containing the attached CD (in this example, the CDROM drive is E:), then click “Next”.
5. Select “Continue Anyway”. The computer will search for another file, called “FTD2XX.sys”.
6. Click “Browse” to find and select the file named “FTD2XX.sys”and then click “OK”. After the installation, the computer will display the following screen
Page 43
AM8060 Precision Thermometer User’s Guide
42/43
7. Click “Finish” to end the installation.
8. The computer will display a message saying, “Your new hardware is ready to use” to indicate the USB drivers are correctly installed.
Loading...