AccuMac AM8040 Specifications

AM8040 PRECISION THERMOMETER
Address: 3150 N Arizona Ave Ste 103, Chandler, AZ 85225 Phone: (480) 634-0603 Email: sales@accumac.com
HIGHLIGHTS
Accuracy up to ±0.008°CResolution 0.001°CData storage into USB flash drive
OVERVIEW
AM8040 is a single channel bench top Precision Thermometer. It features high accuracy, fast readings and great stability. It operates with a wide range of Platinum Resistance Thermometers (PRTs). Users can choose to display temperature in °C, °F or
resistance values. All of the readings can be
stored in a USB flash drive or transferred to PC through an USB cable.
AM8040 Precision Thermometer allows users to choose ITS-90, IEC-751 (DIN), or Callendar-Van Dusen conversion methods to response to various PRTs. Users can also choose to key in PRT coefficients to ensure the best accuracy.
AM8040 provide a simply solution as a PRT readout to achieve high accuracy at very affordable cost for calibration labs.
FEATURES
High accuracy: up to ±0.008 °C at 0 °CHigh resolution: 0.001 °C over the full
range
Temperature range: -200 °C to 850 °CITS-90, CVD, IEC-751 conversion methods
or key in coefficients of PRTs
2.7 inch OLED displayFull-size Touchpad for function controlData storage into USB flash driveData transfer to PC using an USB cable
AM8040 PRECISION THERMOMETER
Address: 3150 N Arizona Ave Ste 103, Chandler, AZ 85225 Phone: (480) 634-0603 Email: sales@accumac.com
SPECIFICATIONS
Temperature Range
-200°C to 850°C
Accuracy (meter only)
±0.01°C @ -200°C ±0.008°C @ 0°C ±0.009°C @ 232°C ±0.01°C @ 420°C ±0.015°C @ 660°C
Resolution
0.001°C (0.0001 Ω) over full range
Probe
Nominal Rtpw: 25 Ω or 100 Ω RTD, PRT or SPRT
Characterizations
ITS-90 coefficients, Callender Van Dusen coefficients, IEC­751 (DIN 385)
Sample Interval
1 second
Display
2.7 inch OLED
Display Units
°C, °F, Ω
Excitation Current
1 mA, reversing
Operation Range
15 °C to 35 °C
Thermometer Input Connectors
Spade plug or bare wire
Power Requirements
100-240V
Dimension
180 mm (W) X 65 mm (H) X 200 mm (D)
Weight
0.3 kg (0.7 lbs)
OPTIONAL ACCESSORIES
Model
Description
1610/1620/1640/1660
Precision Industrial PRT, -200°C to 670°C
1730
Secondary Reference PRT, -200°C to 420°C
1750
Secondary Reference PRT, -200°C to 670°C
1760/1762
Secondary SPRT, -200°C to 670°C
1850
Metal-sheath SPRT, -200°C to 500°C
1860
Metal-sheath SPRT, -200°C to 670°C
1880
Metal-sheath SPRT with Bird-Cage sensor, -200°C to 670°C
1950
Quartz-sheath SPRT, -200°C to 500 °C
1960
Quartz-sheath SPRT, -200°C to 670 °C
Ordering Information
Model
Description
8040
Single Channel Precision Thermometer
9003
Carrying case included
Loading...