AccuMac AM1730-12 Specifications

AM1730 Secondary Reference PRT
OVERVIEW
HIGHLIGHTS
FEATURES
Affordable reference probe Accuracy ±0.015 °C at 0.01 °C Temperature range: -200 °C to
420 °C
Customized dimensions available
AM1730 series Secondary Reference PRT provides our customers an affordable reference probe for precision temperature measurement and calibration in labs and fields. The PRT features accuracy of
± ±
available at 9-inch and 12-inch.
To reach the best performance in stability and repeatability, the sensing element is specially designed to protect the platinum sensing wire from contamination at high temperature. A unique support structure and filling material provide the best balance among the hysteresis effect, mechanical shock and thermal shock performance. This high performance probe meets ITS-90 criteria of reference thermometer fully.
0.015 °C and short term stability of
0.007 °C. Two different lengths of PRTs are
Temperature range: -200 °C to 420 °C Accuracy: ±0.015 °C
 
Long term drift: ±0.01 °C Short term stability: 0.007 °C
Durable and shock resistance Temperature Coefficient 0.003925 Ω/ Ω/°C
W(Ga)>=1.11807
Inconeltm sheath Customized dimensions available
Address: 90 N William Dillard Drive C-107, Gilbert, AZ 85233 For Technical Help: info@accumac.com Phone: 480-634-0603 For Sales: sales@accumac.com
© 2017 Accumac Corporation, All Rights Reserved www.accumac.com
AM1730 Secondary Reference PRT
tm
SPECIFICATIONS
Temperature Range 1730-12: -200 °C to 420 °C
1730-9: -60 °C to 300 °C Resistance at 0 °C Nominal 100 Ω Temperature Coefficient 0.003925 Ω/ Ω/°C Accuracy ±0.025 °C at -196 °C
±0.015 °C at 0.01 °C
±0.035 °C at 420 °C
Drift ±0.01 °C at TPW after 100 hours at 420 °C Short Term Stability ±0.007 °C Thermal Shock ±0.005 °C after 10 times thermal cycles from minimum to
maximum temperatures
Hysteresis <=0.005 °C Self-heating 50 mW/°C Response Time 9 seconds for 63% response to step change in water moving
at 3 feet per second
Measurement Current 0.5 mA or 1 mA Sensor Length 32 mm Sensor Location 5 mm from tip Insulation Resistance >1000 MΩ at room temperature Sheath Material Inconeltm Dimension 1730-12: 0.25 inch X 12 inch (6.35 mm X 305 mm)
1730-9: 0.187 inch X 9 inch (4.75 mm X 229 mm)
External Leads Teflon Handle Dimension 1730-12: 15 mm (OD) x 65 mm (L)
1730-9: 10 mm (OD) X 50 mm (L)
Handle Temperature Range* -50 °C to 180 °C Optional Calibration NIST traceable calibration and data available per request:
1730-12: Ordering # 5014
1730-9: Ordering # 5015
*Handle temperature outside this range will cause damage to the probe.
– insulated copper wire, 4 leads, 2.5 meters
OPTIONAL ACCESSORIES
Model Description
9001 Wooden Carrying Case
Address: 90 N William Dillard Drive C-107, Gilbert, AZ 85233 For Technical Help: info@accumac.com Phone: 480-634-0603 For Sales: sales@accumac.com
© 2017 Accumac Corporation, All Rights Reserved www.accumac.com
Loading...