AccuMac AM1643 Specifications

AM1643 Full Immersion PRT
HIGHLIGHTS
OVERVIEW
FEATURES
Transition junction and lead wire
can withstand high temperature up to 420°C
Short term stability: 0.02 °C Temperature range: 0 °C to 420 °C
AM1643 full immersion PRT features a unique design that allows for the users to apply the probe together with its lead wire in a high temperature environment such as ovens or furnaces etc. It covers a wide range of temperature from 0 °C to 420 °C with amazing
accuracy of ±0.04 °C at 0 °C, short term stability
of ±0.02 °C and fast respond time of 5 seconds.
To reach the best performance in stability and repeatability, the wire-wound sensing elements are specially designed to protect the platinum sensing wire from contamination at high temperature. A unique support structure and filling material provide the best balance among the hysteresis effect, mechanical shocks and thermal shock performance. This probe conforms to the DIN/IEC-751 curve precisely.
Temperature range: 0 °C to 420 °C
Accuracy: ±0.04 °C at 0 °C Long term drift: ±0.04 °C
Short term stability: 0.02 °C
Transition junction and lead wire can withstand the full temperature range of the PRT Temperature Coefficient 0.00385
Follow DIN/IEC-751 precisely
Inconeltm sheath Quick response time
PECIFICATIONS
Address: 90 N William Dillard Drive C-107, Gilbert, AZ 85233 For Technical Help: info@accumac.com Phone: 480-634-0603 For Sales: sales@accumac.com
© 2017 Accumac Corporation, All Rights Reserved www.accumac.com
AM1643 Full Immersion PRT
tm
SPECIFICATIONS
Temperature Range 0 °C to 420 °C Resistance at 0 °C Nominal 100 Ω Temperature Coefficient 0.00385 Ω/ Ω/°C Accuracy* ±0.04 °C at 0 °C
±0.05 °C at 200 °C ±0.07 °C at 420 °C
Drift ±0.04 °C at 0 °C after 100 hours at 420 °C Short Term Stability ±0.02 °C Thermal Shock ±0.02 °C after 10 times thermal cycles from minimum to
maximum temperatures
Hysteresis <=0.01 °C Self-heating 50 mW/°C Response Time 5 seconds for 63% response to step change in water moving
at 3 feet per second
Measurement Current 1 mA Minimum immersion depth 50 mm Maximum immersion depth Dry medium: full immersion
Liquid medium: use protective tube if immersion depth reaches transition junction
Sensor Length 30 mm Insulation Resistance >500 MΩ at room temperature Sheath Material Inconel Dimension 0.125 inch X 2 inch (3 mm X 50 mm) External Leads Four fiberglass insulation Ni-plated copper wires, 2.5 meters
(longer lead wires are available per request)
Handle Dimension 7 mm (OD) X 30 mm (L) Optional Calibration NIST traceable calibration and data available per request:
Ordering # 5012
*With optional calibration.
OPTIONAL ACCESSORIES
Model Description
9001 Wooden Carrying Case
Address: 90 N William Dillard Drive C-107, Gilbert, AZ 85233 For Technical Help: info@accumac.com Phone: 480-634-0603 For Sales: sales@accumac.com
© 2017 Accumac Corporation, All Rights Reserved www.accumac.com
Loading...