AccuMac AM1642 Specifications

AM1642 Full Immersion PRT
Address: 3150 N Arizona Ave STE 103, Chandler, AZ 85225 Phone: (480)634-0603 Email: sales@accumac.com
SPECIFICATIONS
HIGHLIGHTS
Transition junction and lead wire
can withstand high temperature up to 420°C
Accuracy ±0.04°C at 0°CTemperature range: 0°C to 420°CCustomized dimensions available
OVERVIEW
AM1642 full immersion PRT features a unique design that allows for the users to apply the probe together with its lead wire in a high temperature environment such as ovens or furnaces etc. It covers a wide range of temperature from 0 °C to 420 °C with amazing
accuracy of ±0.04°C at 0 °C, short term stability of ±0.02°C and fast respond time of 5 seconds.
The standard length is 4-inch but customized dimensions are available per request. Customized length can range from 2 inches to 12 inches.
To reach the best performance in stability and repeatability, the wire-wound sensing elements are specially designed to protect the platinum sensing wire from contamination at high temperature. A unique support structure and filling material provide the best balance among the hysteresis effect, mechanical shocks and thermal shock performance. This probe conforms to the DIN/IEC-751 curve precisely.
FEATURES
Temperature range: 0 °C to 420 °CAccuracy: ±0.04 °C at 0 °CLong term drift: ±0.04 °CShort term stability: 0.02 °CTransition junction and lead wire can
withstand the full temperature range of the PRT
Temperature Coefficient 0.00385Follow DIN/IEC-751 preciselyInconeltmsheath  Quick response timeCustomized dimensions available
AM1642 Full Immersion PRT
Address: 3150 N Arizona Ave STE 103, Chandler, AZ 85225 Phone: (480)634-0603 Email: sales@accumac.com
SPECIFICATIONS
Temperature Range
0°C to 420°C
Resistance at 0 °C
Nominal 100 Ω
Temperature Coefficient
0.00385 Ω/ Ω/°C
Accuracy*
±0.04°C at 0°C ±0.05°C at 200°C ±0.07°C at 420°C
Drift
±0.04°C at 0 °C after 100 hours at 420 °C
Short Term Stability
±0.02°C
Thermal Shock
±0.02°C after 10 times thermal cycles from minimum to maximum temperatures
Hysteresis
<=0.01°C
Self-heating
50 mW/°C
Response Time
5 seconds for 63% response to step change in water moving at 3 feet per second
Measurement Current
1 mA
Minimum immersion depth
50 mm
Maximum immersion depth
Dry medium: full immersion Liquid medium: use protective tube if immersion depth reaches transition junction
Sensor Length
30 mm
Insulation Resistance
>500 MΩ at room temperature
Sheath Material
Inconel
tm
Dimension
0.125 inch X 4 inch (3 mm X 100 mm) (length of the probes can be customized per request, range from 2 inches to 12 inches)
External Leads
Four fiberglass insulation Ni-plated copper wires, 2.5 meters (longer lead wires are available per request)
Handle Dimension
6.4mm (OD) X 25 mm (L)
*With optional calibration.
OPTIONAL ACCESSORIES
Model
Description
9001
Wooden Carrying Case
Loading...