AccuMac AM1610 Specifications

AM1610 PRECISION INDUSTRIAL PRT
FEATURES
HIGHLIGHTS
OVERVIEW
Sensor Length: 30 mm Accuracy: ±0.05°C Durable and shock resistance Customized dimensions
available
With 30 mm long sensor, AM1610 series precision industrial PRT provides users an excellence temperature probe with extremely low stem conduction (heat loss along sheath) in precision temperature measurement. It is a great tool for measuring and testing the vertical temperature uniformity of heat sources.
AM1610 uses a special aged and tested 30-mm long sensing element to achieve the best performance in stability and repeatability. A unique assembly procedure provides the best balance among the hysteresis effect, mechanical shock and thermal shock performance.
Temperature range: -60 °C to 160 °C Sensor Length: 30 mm Accuracy: ±0.05 °C Long term drift: ±0.04 °C Short term stability: 0.01 °C Durable and shock resistance Temperature Coefficient 0.00385 SST sheath Quick response time Customized dimensions available
Address: 90 N William Dillard Drive C-107, Gilbert, AZ 85233 For Technical Help: info@accumac.com Phone: 480-634-0603 For Sales: sales@accumac.com
© 2017 Accumac Corporation, All Rights Reserved www.accumac.com
AM1610 PRECISION INDUSTRIAL PRT
tm
SPECIFICATIONS
Temperature Range -60 °C to 160 °C Resistance at 0 °C Nominal 100 Ω Temperature Coefficient 0.00385 Ω/ Ω/°C Accuracy ±0.05 °C over the full range Drift ±0.04 °C at 0 °C after 100 hours at 160 °C Short Term Stability ±0.01 °C Thermal Shock ±0.01 °C after 10 times thermal cycles from minimum to
maximum temperatures
Hysteresis <=0.01 °C Self-heating 50 mW/°C Response Time 5 seconds for 63% response to step change in water moving
at 3 feet per second
Measurement Current 0.5 mA or 1 mA Sensor Length 30 mm Sensor Location 5 mm from tip Insulation Resistance >1000 MΩ at room temperature Sheath Material 316 Stainless Steel Dimension 1610-12: 0.25 inch X 12 inch (6.35 mm X 305 mm)
1610-9: 0.187 inch X 9 inch (4.75 mm X 229 mm)
External Leads Teflon Handle Dimension 1610-12: 15 mm (OD) X 65 mm (L)
1610-9: 10 mm (OD) X 50 mm (L)
Handle Temperature Range* -50 °C to 160 °C Optional Calibration NIST traceable calibration and data available per request:
1610-12: Ordering # 5017
*Handle temperature outside this range will cause damage to the probe.
– insulated copper wire, 4 leads, 2.5 meters
OPTIONAL ACCESSORIES
Model Description
9001 Wooden Carrying Case
Address: 90 N William Dillard Drive C-107, Gilbert, AZ 85233 For Technical Help: info@accumac.com Phone: 480-634-0603 For Sales: sales@accumac.com
© 2017 Accumac Corporation, All Rights Reserved www.accumac.com
Loading...