Accuhard A101 User Manual

Page 1
Accuhard A101 LCR Meter accuhard.com

Accuhard A101

USB LCR Meter User Manual
1
Page 2
Accuhard A101 LCR Meter accuhard.com

Getting Started

1. Install A101 software from installation CD or download from www.accuhard.com.
2. Connect A101 device to PC.
4. If driver installation procedure did not start, open Device Manager, find A101, right click on it and choose “Update driver”. Than do as described in #3.
5. Run A101 software.
6. Connect leads or probe to A101 device.
7. Calibrate A101 device with connected probe or leads “Main menu > Calibrate > Calibrate”.
8. After calibration save calibration file “Main menu > Calibrate > Save calfile as..”. It is better to include test leads’ type into calfile name. If you will change test leads’ type, do calibration again.
9. Save profile Main menu > Profile > Save profile as..”.

Safety and Warnings

Do not apply voltage to input terminals.
Discharge capacitor before testing.
Do not measure powered circuits and connected to ground.

Connections

Figure 1: Connection of test leads, tweezers and USB cable to A101.
HC means high side current connector, positive polarity. HP means high side potential connector, positive polarity. LP means low side potential connector. LC means low side current connector.
2
Page 3
Accuhard A101 LCR Meter accuhard.com
3
Page 4
Accuhard A101 LCR Meter accuhard.com

Display, Panels and Menus

Display

Full list of measured parameters may be found in “Specifications” chapter.
Main value is basic measured element of equivalent schematic. This indicator displays current
type of measurement, measured value and unit of measurement. See CLR panel for more information.
Auxiliary value is second element of equivalent schematic.
Figure 2: Display and panels, normal view.
It may be:
Rs (ESR) – equivalent series resistance Rs = Z.im * D;  Rp – equivalent parallel resistance Rp = Z.im / D; Cp – equivalent parallel capacitance for Rac Cp = 1 / (2
“Auto start” button runs continuous measurements.“Manual start” button runs single measurement or stops continuous measurements. Impedance value indicator displays either real (Z.re) and imagery (Z.im) parts of impedance
or module (Z.md) and phase angle (Z.Θ), see view menu for more information about view options.
4
*p*F*
Rac / D);
Page 5
Accuhard A101 LCR Meter accuhard.com
Admittance value indicator displays either real (Y.re) and imagery (Y.im) parts of impedance
or module (Y.md) and phase angle (Y.Θ), see view menu for more information about view options.
Reference value indicator displays “main” value used for sorting or relative measurement.
It uses the same units as main value.
Delta value indicator displays result of relative measurement.CLR panel is used to choose necessary type of measurement.Range panel is used for auto or manual range choice.Frequency panel is used for auto or manual frequency choice.Sorting panel is used to display results of sorting and for fast access to tolerance parameters.Nominal frequency indicator displays rounded value of real measuring frequency.
All displayed impedance and admittance values are recalculated from real frequency to nominal frequency.
Signal level indicator displays currently used output signal level. Signal level may be set in
work menu.
Status indicator displays connection state of A101 device. It must be “Ready” for
measurements.
DC voltage indicator displays Direct Current (DC) bias voltage on high side connectors (HC
and HP), see Figure 1.
DUT voltage indicator displays voltage on Device Under Test (DUT).Real frequency displays exact value of measuring frequency.Rst voltage displays voltage on standard resistor, see Figure 56.
To start work, run A101 application, connect device or vice verse, choose type of measurement, connect leads to DUT and press “Auto start” or “Manual start” button. “Auto start” means continuous measurements until “Manual start” button will be pressed. “Manual start” means single measurement with fixed range choice or a few measurements with auto range choice.
If device already was calibrated and calfile was saved and settings were saved in profile, they will be auto loaded and you will see them on caption of application form.

Panels

CLR Panel

CLR panel provides fast access and visual control of current measurement.
Figure 3: CLR panel.
“Cs” means capacitance, series model. “Cp” means capacitance, parallel model. “Ls” means inductance, series model.
5
Page 6
Accuhard A101 LCR Meter accuhard.com
“Lp” means inductance, parallel model. “Rac” means Alternate Current (AC) resistance, parallel model.
All buttons have hints with equivalent models for every button.
Figure 4: CLR panel’s equivalent schematics.

Range Panel

Range panel provides fast access to auto or manual choice of range and visual control of current range.
Capacitance:
Figure 5 Capacitance, range panel with auto choice.
with auto range choice, current range’s value is highlighted in green color
Figure 6: Capacitance, range panel with manual choice.
and with manual range choice, recommended range’s value is highlighted in green color.
Inductance:
Figure 7: Inductance, range panel with auto choice.
with auto range choice, current range’s value is highlighted in green color
Figure 8: Inductance, range panel with manual choice.
and with manual range choice, recommended range’s value is highlighted in green color.
AC resistance:
6
Page 7
Accuhard A101 LCR Meter accuhard.com
Figure 9: AC resistance, range panel with auto choice.
with auto range choice, current range’s value is highlighted in green color
Figure 10: AC resistance, range panel with manual choice.
and with manual range choice, recommended range’s value is highlighted in green color.

Frequency Panel

Figure 11: Frequency panel with auto choice.
Frequency panel with auto range choice, current frequency’s value is highlighted in green color
Figure 12: Frequency panel with manual choice.
and with manual range choice, recommended frequency’s value is highlighted in green color.

Sorting Panel

Figure 13: Sorting panel.
Sorting panel shows tolerance relatively Reference value. Click on a button will open “Options”
window with “Sorting” page to adjust tolerance parameters for the button and the bin.

Menus

Profile Menu

Figure 14: Profile menu.
Profile menu allows all current settings to be saved to or loaded from a profile file. Profile is useful when you need periodically repeat a few
7
Page 8
Accuhard A101 LCR Meter accuhard.com
different measurements like tolerance sorting for different values and different tolerance settings.

View Menu

In “View” menu you can choose:
- normal or compact application size,
Figure 15: Compact view.
Figure 16: View menu.
- panels’ visibility,
8
Page 9
Accuhard A101 LCR Meter accuhard.com
Figure 17: Panels’ visibility menu.
- format of values – Cartesian with real and imagery parts or Polar with module and phase angle
parts,
Figure 18: Format menu.
Figure 19: Example of formats.
- format of numbers - General or Scientific,
Figure 20: Number menu.
Figure 21: Example of numbers.
- set “Stay on top” view mode,
9
Page 10
Accuhard A101 LCR Meter accuhard.com
- number of digits,
Figure 22: Digits menu.
and
- open “Options” window to set DC bias voltage scan parameters and tolerance parameters for
sorting.
Figure 23: Options window.

Work Menu

In “Work” menu, which is useful if CLR, range and frequency panels are hidden,
10
Page 11
Accuhard A101 LCR Meter accuhard.com
Figure 24: Work menu.
you can choose:
- type of measurement: capacitance with series model “Cs”, capacitance with parallel model “Cp”, inductance with series model “Ls”, inductance with parallel model “Lp” and AC resistance with parallel model “Rac”.
- range of value: with auto choice (“Auto” on range panel or menu) or manual choice by user,
Figure 25: CLR and range menus.
- measuring frequency: with auto choice (“Auto” on frequency panel or menu) or manual choice by user,
- set averaging value from 1 to 50,
11
Page 12
Accuhard A101 LCR Meter accuhard.com
Figure 26: Frequency and averaging menus.
- set signal level 0.5V or 0.05V,
Figure 27: Signal level menu.
- open direct current (DC) voltage scan parameters page (“DCV scan”) to start scanning,
Figure 28: DCV scan menu and Options window.
- set last main value as reference (“Set as reference”) or double click on the main value,
12
Page 13
Accuhard A101 LCR Meter accuhard.com
Figure 29: Set as reference menu.
- set arbitrary value as reference on “Options” window or double click on reference value,
Figure 30: Set reference menu.
- open sorting parameters page,
Figure 31: Sorting menu and window.
Also click on “Sorting” panel will open “Options” window to edit a value of this bin. Names of
bins on sorting panel are values of maximum limits.
13
Page 14
Accuhard A101 LCR Meter accuhard.com
- start recording of data in new worksheet (Alt + N) or add last data (Space),
Figure 32: Record data menu.
- set power net frequency and thereby do 100Hz (50Hz) or 120Hz (60Hz) frequency choice.
Figure 33: Net frequency menu.
“CLR”, “Range” and “Frequency” menus in “Work” menu are duplicated with “CLR”, “Range”
and “Frequency” panels.

Calibrate Menu

Figure 34: Calibrate menu.
“Calibrate” menu allows to start new calibration and than save it as calibration file (calfile) and
later open it to save time if maximum accuracy is not necessary.

Charts& Menu

“Charts&” menu provides access to all created charts and data worksheets.
Figure 35: Find and show result window.

Help Menu

14
Page 15
Accuhard A101 LCR Meter accuhard.com
“Help” menu gives access to this User Manual and “About” window with link to www.accuhard.com.
Figure 36: Help menu and “About” window.
15
Page 16
Accuhard A101 LCR Meter accuhard.com

Making C, L and Rac Measurements

Connect leads to LCR meter as shown on Figure 1. Connect LCR meter to computer, run “A101” software. Connect leads to Device Under Test (DUT) as shown on figures 37 – 39.

Measuring Capacitance (C)

Connect leads to Device Under Test (DUT) as shown on Figure 37.
All capacitors must be discharged before connection to LCR meter.
Select “Cs’ or “Cp” type of measurement on CLR panel or in CLR menu. Set “Auto” on range and frequency panels or in range and frequency menus. Press “Auto start” or “Manual start” button. Read the displays. If the main value is colored in red and status bar has a label “Out of range”, check the range and frequency and also check leads’ connection, it may be shorted.
If auto range choice cannot find right range, set the range manually.
Since test leads have significant uncompensated inductance, up to 0.5µH and even more, and this inductance very strong depends from the leads’ position, it is hard to keep the inductance stable to compensate by calibration. For measuring frequency 100kHz the uncompensated portion of the inductance will give reactance comparable with reactance of the capacitors with capacitance of a few µF and higher. So you need to be very careful with measuring frequency 100kHz. LCR meter must be freshly calibrated before such measurements take place. Leads’ position must be unchanged after calibration.
Figure 37: Through hole, SMD and low ESR SMD capacitors' connection.
16
Page 17
Accuhard A101 LCR Meter accuhard.com
17
Page 18
Accuhard A101 LCR Meter accuhard.com

Measuring Inductance (L)

Figure 38: Inductor’s connection.
Connect leads to an inductor as shown on Figure 38. Select “Ls’ or “Lp” type of measurement on CLR panel or in CLR menu. Set “Auto” on range and frequency panels or in range and frequency menus. Press “Auto start” or “Manual start” button. Read the displays. If the main value is colored in red and status bar has a label “Out of range”, check the range and frequency and also check leads’ connection, it may be open.
If auto range choice cannot find right range, set the range manually.
Since test leads have significant uncompensated inductance, up to 0.5µH and even more, and this inductance very strong depends from the leads’ position, it is hard to keep the inductance stable to compensate by calibration. For measuring frequency 100kHz the uncompensated portion of the inductance will give reactance comparable with reactance of low value inductors with the inductance less than tens µH. So you need to be very careful with measuring frequency 100kHz. LCR meter must be freshly calibrated before such measurements take place. Leads’ position must be unchanged after calibration.
18
Page 19
Accuhard A101 LCR Meter accuhard.com

Measuring AC Resistance (Rac)

Figure 39: Resistor’s connection.
Connect leads to a resistor as shown on Figure 39. Select “Rac” type of measurement on CLR panel or in CLR menu. Set “Auto” on range and frequency panels or in range and frequency menus. Press “Auto start” or “Manual start” button. Read the displays. If the main value is colored in red and status bar has label “Out of range”, check the range and frequency and also check leads’ connection, it may be open.
If auto range choice cannot find right range, set the range manually.
19
Page 20
Accuhard A101 LCR Meter accuhard.com

DC Bias Voltage Scan

Click “DCV scan” in “Work” menu to begin DC bias voltage scan.
Figure 40: DC bias voltage scan settings.
Use “DCV scan” page of “Options” window to set parameters and start DCV scanning. Better to use fixed range, not Auto range, for DCV scan.
DCV scan page on Options window also gives possibility to measure capacitance with any fixed DC bias voltage in range 0 – 30V.
20
Page 21
Accuhard A101 LCR Meter accuhard.com
Figure 41: Diode’s connection.
Connect positive high side leads to diode’s cathode as shown on Figure 41. Use 0.05V signal level for diode’s capacitance measurement. For capacitors, inductors and resistors 0.5V level is preferred.
Figure 42: DC bias voltage scan results for ceramic capacitor with high capacitance density.
21
Page 22
Accuhard A101 LCR Meter accuhard.com
You can save the DCV scanned data in file and charts as pictures by the main menu or pop up
menu.
To edit the chart titles use the pop up menu.
Figure 43: Data page of DCV scan.
22
Page 23
Accuhard A101 LCR Meter accuhard.com
Figure 44: Main and Rs (ESR) chart example.
23
Page 24
Accuhard A101 LCR Meter accuhard.com
Figure 45: Main, D and Q chart example with pop up menu.
24
Page 25
Accuhard A101 LCR Meter accuhard.com
Figure 46: Main chart with title edit window.

Sorting and Relative Measurement

To start sorting, first you need to set “Main reference” parameter and, if necessary, change the limits of bins. Zero value of the main reference means no sorting. After editing the reference value click “OK” button or press “Enter”.
Choose “Set reference” in “Work” menu or double click reference value.
To edit thresholds of bins click the bin’s button if “Options’ window is not visible or choose the required bin page. Click “OK” button.
25
Page 26
Accuhard A101 LCR Meter accuhard.com
Figure 47: Set reference example.
Figure 48: Relative measurement and sorting.
26
Page 27
Accuhard A101 LCR Meter accuhard.com
You will see relative difference or delta and only one highlighted bin with limits wider than the delta.
To set the current main value as reference just double click it.
27
Page 28
Accuhard A101 LCR Meter accuhard.com

Performing Calibration

Start calibration as shown on Figure 49 and Figure 50.
Figure 49: Calibrate menu.
Figure 50: “Open probe” message.
Open test leads, tweezers or probe as shown on Figure 51 and press OK button.
Figure 51: Kelvin test leads and tweezers in open state for calibration.
28
Page 29
Accuhard A101 LCR Meter accuhard.com
Figure 52: Calibration progress window.
Wait until the “Short probe” message (Figure 53), short test leads, tweezers or probe as shown on Figure 54 then press OK button. Wait until the end of calibration (Figure 55).
Figure 53: “Short probe” message.
Figure 54: Kelvin test leads and tweezers in short state for calibration.
29
Page 30
Accuhard A101 LCR Meter accuhard.com
Figure 55: End of calibration.
Save results of calibration in a calibration file (calfile) and if you created a new calfile and planning to use the same test leads later, save the profile as well. It is better to include test leads’ type into the calfile name. On the next A101 application start these files will be auto loaded.
30
Page 31
Accuhard A101 LCR Meter accuhard.com

Theory of Operation

Figure 56: I-V schematics for low and high DUT impedance.
I-V method is used in A101 LCR meter. Standard resistor’s (Rst) and DUT’s (Zdut) positions are changed by type of measurement and range to increase accuracy. Left schematic is used for low impedance DUTs and right for high impedance DUTs.
31
Page 32
Accuhard A101 LCR Meter accuhard.com
Figure 57: Simplified A101 schematic’s configuration for high impedance DUTs.
32
Page 33
Accuhard A101 LCR Meter accuhard.com
Figure 58: Simplified A101 schematic’s configuration for low impedance DUTs.
Simplified schematic of A101 LCR meter is shown on Figure 57 and Figure 58. Different configurations allow to apply additional gain of signal ADC to increase signal to noise ratio. To minimize capacitance of cables, to shields the cables the same voltage as V2 is applied. In this case shields work as guards.
Signal ADC measures all voltages V0 – V4 one by one and software on PC calculates all parameters of a DUT.
DC ADC periodically measures DC bias voltage on high side to provide visual control of really applied to DUT DC bias voltage.
Maximum DC current through DUT is limited by 2k resistor.
33
Page 34
Accuhard A101 LCR Meter accuhard.com

Specifications

Parameters Measured

Cs – capacitance, series model;Cp – capacitance, parallel model;Ls – inductance, series model;Lp – inductance, parallel model;Rac – alternate current (AC) resistance, parallel model;Rs – equivalent series resistance (ESR), series model, Rs = Z.im * D;Rp – equivalent parallel resistance, parallel model, Rp = Z.im / D;D – dissipation factor;Q – quality factor;Z.re – impedance, real part, series model, Cartesian coordinates;Z.im – impedance, imagery part, series model, Cartesian coordinates;Z.md – impedance, module, polar coordinates;Z.Θ – impedance, phase angle, polar coordinates;Y.re – admittance, real part, parallel model, Cartesian coordinates;Y.im – admittance, imagery part, parallel model, Cartesian coordinates;Y.md – admittance, module, polar coordinates;Y.Θ – admittance, phase angle, polar coordinates;Delta – relative delta, Delta = (Main value – Reference) / Reference * 100%;

Ranges

C: 300pF, 3nF, 30nF, 300nF, 3µF, 30µF, 300µF, 3mF, 30mF, 100mF L: 30µH, 300µH, 3mH, 30mH, 300mH, 3H, 30H, 300H, 1kH Rac: 30Ω, 300Ω, 3kΩ, 30kΩ, 300kΩ, 3MΩ, 10MΩ

Measurement Rate

>120Hz: 5 per second ≤120Hz: 1.1 per second

Test Signal Level

≤0.5Vrms ±10% ≤0.05Vrms ±10%
34
Page 35
Accuhard A101 LCR Meter accuhard.com

Consumption

Voltage: 5V ±10% Current: <400mA

Environmental

Operating temperature: 5°C to 40°C, <70%RH Storage temperature: -20°C to 70°C, <80%RH

Mechanical

Device dimensions: 14.5 cm W x 10.5 cm D x 4.1 cm H (5.7” x 4.4” x 1.6”) Summary weight: < 700 g (1.6 lb)

Accuracy Specifications

Accuracy is specified at 23°C ± 5°C, <70% RH. All accuracy is specified as ±[(% of reading) + (Offset as % of current range’s value or Full Scale (FS))]. For the most precise measurement results, the meter has to be freshly calibrated before taking a reading.
Table 1: Impedance/Resistance accuracy AZ with auto range choice and 0.5V level.
Range Resolution 20Hz 100/120Hz 1kHz 10kHz 100kHz
30Ω 0.1mΩ 0.3%+0.01% 0.3%+0.01% 0.3%+0.01% 0.3%+0.01% 1.0%+0.03%
300Ω 1mΩ 0.3%+0.01% 0.3%+0.01% 0.3%+0.01% 0.3%+0.01% 1.0%+0.03%
3kΩ 10mΩ 0.3%+0.01% 0.3%+0.01% 0.3%+0.01% 0.3%+0.01% 1.0%+0.03%
30kΩ 100mΩ 0.3%+0.01% 0.3%+0.01% 0.3%+0.01% 0.3%+0.01% 2.0%+0.03%
300kΩ 0.3%+0.01% 0.3%+0.01% 0.3%+0.01% 1.0%+0.03% -
3MΩ 10Ω 1.0%+0.03% 1.0%+0.03% 1.0%+0.03% - -
10MΩ 100Ω 2.0%+0.05% 2.0%+0.05% - - -
Table 2: Capacitance accuracy AC with auto range choice and 0.5V level.
35
Page 36
Accuhard A101 LCR Meter accuhard.com
Range Resolution 20Hz 100/120Hz 1kHz 10kHz 100kHz
*
300pF
3nF 0.01pF - 0.3%+0.01% 0.3%+0.01% 0.3%+0.01% 1.0%+0.03%
30nF 0.1pF 0.3%+0.01% 0.3%+0.01% 0.3%+0.01% 0.3%+0.01% 1.0%+0.03%
300nF 1pF 0.3%+0.01% 0.3%+0.01% 0.3%+0.01% 0.3%+0.01% 1.0%+0.1%
3µF 10pF 0.3%+0.01% 0.3%+0.01% 0.3%+0.01% 1.0%+0.03% 3.0%+0.3%
30µF 100pF 0.3%+0.01% 0.3%+0.01% 0.3%+0.01% 3.0%+0.3% 10.0%+0.5%
300µF 1nF 0.3%+0.01% 0.3%+0.01% 1.0%+0.03% 10.0%+0.5% -
3mF 10nF 1.0%+0.03% 1.0%+0.03% 3.0%+0.3% - -
30mF 100nF 2.0%+0.1% 3.0%+0.3% - - -
100mF 1µF 3.0%+0.3% - - - -
0.001pF - - 0.5%+0.03% 0.5%+0.03% 2.0%+0.05%
Table 3: Inductance accuracy AL with auto range choice and 0.5V level.
Range Resolution 20Hz 100/120Hz 1kHz 10kHz 100kHz
*
30µH
300µH 1nH - - 0.3%+0.01% 0.3%+0.01% 1.0%+0.03%
3mH 10nH - 0.3%+0.01% 0.3%+0.01% 0.3%+0.01% 1.0%+0.03%
30mH 100nH - 0.3%+0.01% 0.3%+0.01% 0.3%+0.01% 1.0%+0.03%
300mH 1µH - 0.3%+0.01% 0.3%+0.01% 0.3%+0.01% 1.0%+0.03%
3H 10µH - 0.3%+0.01% 0.3%+0.01% 1.0%+0.03% -
30H 100µH - 0.3%+0.01% 1.0%+0.03% 3.0%+0.1% -
300H 1mH - 1.0%+0.03% 3.0%+0.1% - -
1kH 10mH - 3.0%+0.1% - - -
0.1nH - - - 1.0%+0.03% 1.0%+0.03%
*
Table 4: AC Resistance accuracy ARac with 0.5V level.
Range Resolution 20Hz 100/120Hz 1kHz 10kHz 100kHz
30Ω 0.1mΩ - 0.3%+0.01% 0.3%+0.01% 0.3%+0.01% 1.0%+0.03%
300Ω 1mΩ - 0.3%+0.01% 0.3%+0.01% 0.3%+0.01% 1.0%+0.03%
3kΩ 10mΩ - 0.3%+0.01% 0.3%+0.01% 0.3%+0.01% 1.0%+0.03%
30kΩ 100mΩ - 0.3%+0.01% 0.3%+0.01% 0.3%+0.01% -
300kΩ - 0.3%+0.01% 0.3%+0.01% 1.0%+0.03% -
3MΩ 10Ω - 1.0%+0.03% 1.0%+0.03% - -
10MΩ 100Ω - 2.0%+0.05% - - -
* After calibration with the same leads/probes position as for measurement.
Table 5: Phase angle of impedance accuracy Θe.
Range Resolution Accuracy = Θ
-180° to 180° 0.001°
(AZ + Offset / Zx)* 180 /
e
p
Condition
D < 1 or Q > 1
Notes:
*
36
Page 37
Accuhard A101 LCR Meter accuhard.com
1. The AZ and Offset variables are the accuracy specified at Table 1.
37
Page 38
Accuhard A101 LCR Meter accuhard.com
Table 6: Dissipation/Quality factor accuracy.
Range Resolution Accuracy = D
e
Condition
C 0.00001 to 99999 AC + (Offset / Cx)*100% + 0.01% D < 1 or Q > 1
L 0.00001 to 99999 AL + (Offset / Lx)*100% + 0.01% D < 1 or Q > 1
Z 0.00001 to 99999 AZ + (Offset / Zx)*100% + 0.01% D < 1 or Q > 1
Rac 0.00001 to 99999 ARac + (Offset / Rx)*100% + 0.01% D < 1 or Q > 1
Notes:
1. The AC, AL, AZ, Arac and Offset variables are the accuracy specified at Table 1, Table 2, Table 3 and Table 4 respectively.
2. The quality factor is the reciprocal of the dissipation factor.
Frequency accuracy: 0.01%. DC bias voltage accuracy: 0.5% * V ± 0.1V. DC current limiting resistance 2k ± 5%.

What is included

USB LCR meter device 1pc.Kelvin alligator leads 1set.Kelvin tweezers 1pc.USB AB male cable 1pc.CD with software 1pc.
38
Page 39
Accuhard A101 LCR Meter accuhard.com

Warranties

FCC Compliance

This device is exempt from FCC Certification under 47 CFR Part 15.103(c).

CE Compliance

This device complies with the Class B limits of CISPR 11 and all relevant requirements of EN 61326-1: 2006.

Warranty

The information contained in this manual is subject to change without notice. Accuhard makes no warranty of any kind with regard to this material, including, but not limited to, the implied warranties or merchant-ability and fitness for a particular purpose. Accuhard shall not be liable for errors contained herein or for incidental or consequential damages in connection with the furnishing, performance, or use of this material. This Accuhard product has a warranty against defects in material and workmanship for a period of one year from date of shipment. During the warranty period, Accuhard will, at its option, either repair or replace products that prove to be defective.

Warranty Service

For warranty service or repair, this product must be returned to Accuhard. The Buyer shall pay shipping charges to Accuhard and Accuhard shall pay UPS Ground, or equivalent, shipping charges to return the product to the Buyer. However, the Buyer shall pay all shipping charges, duties, and taxes, to and from Accuhard, for products returned from another country.

Limitation of Warranty

The foregoing warranty shall not apply to defects resulting from improper use by the Buyer, Buyer-supplied software or interfacing, unauthorized modification or misuse, operation outside of the environmental specifications for the product. No other warranty is expressed or implied. Accuhard specifically disclaims the implied warranties or merchant-ability and fitness for a particular purpose.

Certification

Accuhard certifies that, at the time of shipment, this product conformed to its published specifications.
39
Loading...