2. MICROS 60 OT HYDRAULIC................................................................................................................. 4
2.1. With bottles..................................................................................................................................... 4
2.1.1. Transmission tubes list ........................................................................................................... 4
3.1. With bottles................................................................................................................................... 15
3.1.1. Transmission tubes list ......................................................................................................... 15
4.1. Micros 60 CT bottle version .......................................................................................................... 26
4.2. Micros 60 CT pack version ........................................................................................................... 26
4.3. Micros 60 OT bottle version.......................................................................................................... 26
4.4. Micros 60 OT pack version........................................................................................................... 26
RAA 009 A Ind.A
Page 1/118/06/98
1. HYDRAULIC & PNEUMATIC PRINCIPLES
1. GENERALITIES
MICROS 60 instrument has been designed for simple mechanical operations.
4 stepper motors provide movements to mechanical assemblies.
Pressure and vacuum are provided by the vacuum/waste syringe up and down movements
(diag 1).
Liquid movements are achieved either by means of mechanical assembly movements (liquid
Diag.1
syringes diag 2) or pressure/vacuum and simultaneous action of specific valves.
• Dilution chambers (Diag .3)
WBC/HGB and RBC chambers are made of GRILAMID TR55 LY injected.
The diode and the cell of the spectrophotometer are glued on the WBC/HGB chamber.
Chamber positions can be modified in order to obtain the best sampling position possible.
Dilutions :
First dilution is carried out in the WBC/HGB chamber (with a bubbling phasis).
The RBC blood sample is aspirated from this dilution.
Lyse is sent from the drain nipple of the WBC/HGB chamber.
Counts have a duration of 2 x 6 seconds.
(see procedures RAS 188 A and RAS 187 A for cycle hydraulic details)
Rinse :
To obtain the best rinse in the counting heads, diluent is sent
from the liquid syringes. This is carried out before, between
and after the two counts.
Diag.2
RAA 009 A Ind.A
Page 1/2
18/06/98
1. HYDRAULIC & PNEUMATIC PRINCIPLES
IMPORTANT
A window on the HGB/WBC chamber allows the needle to move down into the chamber
and to inject reagents. As important light or variation of light can cause HGB result
drifts, close instrument cover and door before running blood analyses.
Bubbling
Insulators avoid polluted liquid overflows during bubbling phasis. they also allows an accurate
adjustment of the bubbling volume.
MICROS 60 CT specifics :
- The piercing needle is equipped with two injectors to obtain a homogeneous diluent flow
during needle rinsing phasis (see procedures RAS 188 A and RAS 187 A).
- Atmosphere is provided to sample tubes to allow a correct aspiration of blood.
:
RAA 009 A Ind.A
Page 1/318/06/98
1. HYDRAULIC & PNEUMATIC PRINCIPLES
2. MICROS 60 OT HYDRAULIC
Liquid circuits, hydropneumatic connections, as well as the transmission tubes used, are
described in the following chart tables.
3.2. Required tools and products .......................................................................................................... 7
RAA 009 A Ind.A
Page 3/120/06/98
3. MAINTENANCE
1. MAINTENANCE
1.1. Introduction
WARNING !
Customer maintenance has to be carried out according to the recommended
frequency chart table and after having performed an ABX approved customer training course.
The system warranty may be affected if damage occurs after a non trained technician
intervenes or if replaced spare parts and consumables do not come from an ABX
approved origin.
1.2. Daily customer maintenance
No special adjustments or maintenance has to be done on your equipment if the recommended
startup and shutdown procedures are explicitly respected.
See the
ABX MICROS 60
User Manual for the daily rinsing and cleaning of the system.
1.3. Weekly customer maintenance
An overall check for cleanliness of the system is recommended every week.
All traces of blood or reagent have to be wiped off as soon as possible using a piece of cloth
and distilled water.
CAUTION
Never use solvent or abrasive cleaning material to clean the system.
RAA 009 A Ind.A
Page 3/2
20/06/98
3. MAINTENANCE
2. MAINTENANCE KITS
O RING KIT : XEA 328 AS
FAA 036 A
FAA 029 A1 Diluent dispenser O ring
FAA 055 A2 Sampling syringe O rings
FAA 049 A2 Aperture O rings
FAA 046 A2 Coaxial O rings
FAA 054 A2 Sampling needle O rings (CT)
FAA 053 A1 Sampling needle O ring (OT)
FAA 017 A1 Vacuum/waste syringe O ring
Maintenance and adjustments that need to be done on
NOTE
CONCERNING PARTS :
- Hydraulic maintenance and adjustments.
- Pneumatic maintenance and adjustments.
- Electrical maintenance and adjustments.
- Power supply maintenance and adjustments.
- Electronic maintenance and adjustments.
- Printer maintenance and setup.
are divided into «procedures» according to concerned assemblies. This
should make any updating easier as all interventions can be done with the
corresponding «procedure» on its own.
Each procedure has to be read entirely before beginning the intervention.
NOTE
ABX MICROS 60
WARNING !
When cleaning instruments, disposable gloves should be worn.
RAS 174 ADR AIN DETECTION ADJUSTM ENT Drain detection sensor adjustment
RAS 175 APCB VOLTAGE CHECKS
MICROS 60 OT IN STALLATION
AND STARTUP
SAMPLING NEEDLE
MAINTENANCE
POWER SUPPLY
CHECK/REPLACEMENT
TECHNICIAN FUNCTIONS
MICROS 60 OT
MICROS OT
Unpacking - Working conditions - Visual check s - Reagent
connection - Printer & instrument connections - Priming &
Startup
Needle replacement - O ring replacement
RBC, WBC/HGB chamber cleaning - Aperture O ring
replacement - Coaxial O ring replacement
Voltage supply chec k - Pow er supp ly mod ule replac ement
Fan operation check.
Version display. - Adjustments : HGB photometer calibration,
Aperture voltage, pressure check, W BC gain , RBC & PLT
gain, Sensor, Needle heigth and motion , bubbling. Temperature sensor adjustment - Run mode - Reagent pack Serial number - Cycle number - Burning
RAS 174 ADR AIN DETECTION ADJUSTM ENT Drain detection sensor adjustment
RAS 175 APCB VOLTAGE CHECKS
MICROS 60 CT IN STALLATION
AND STARTUP
SAMPLING NEEDLE
MAINTENANCE
POWER SUPPLY
CHECK/REPLACEMENT
TECHNICIAN FUNCTIONS
MICROS 60 CT
MICROS CT
Unpacking - Working conditions - Visual check s - Reagent
connection - Printer & instrument connections - Priming &
Startup
Needle O ring replacement - Sampling needle replacement
Piercing needle repla cement
RBC, WBC/HGB chamber cleaning - Aperture O ring
replacement - Coaxial O ring replacement
Voltage supply chec k - Pow er supp ly mod ule replac ement
Fan operation check.
Version display. - Adjustments : HGB photometer calibration,
Aperture voltage, pressure check, W BC gain , RBC & PLT
gain, Sensor, Needle heigth and motion , bubbling. Temperature sensor adjustment - Run mode - Reagent pack Serial number - Cycle number - Burning
RAS 197 ADRAIN & RINSEInstrument rinse and drain for an extended shutdown
RAS 198 APIERCING BLOCKDescription - Maintenance - Sampling position
VACUUM/WASTE SYRINGE
MAINTENANCE
CHANGING THE INSTRUMENT
LANGUAGE
HYDRAULIC CYCLE CHECKU P
M
ICROS 60 C T
Configuration - Contro l panel - Control LEds and ke ys Description
Lyse dispenser O ring replacement - Diluent dispenser O ring
replacement - Samplin g needle dispen ser O ring replact. Lubrification
O ring replacement
Changing the instrum ent language
Step by step control of the hy draulic cycle
RAA 009 A Ind.A
Page 3/6
20/06/98
3. MAINTENANCE
3.2. Required tools and products
TOOLSPRODUCTS
DESIGNATIONPART NUMBERDESIGNATIONPART NUMBER
HEXAGONAL KEYS
DYNAMOMETRIC
SCREW DRIV ER
A302
DYNAMOMETRIC
SCREW DRIV ER
A301
DYNAMOMETRIC
SCREW DRIV ER
A300
CLAMPSLIQUID SOAP
SCALPELDISTILLED WATER
CUTTING PLIERSMICROPIPETTE TIP
PAIR OF SCISORS
VOLTMETERLATEX WBCLAD 001 AS
MAG 019 A
MAG 020 A
MAG 013 A
EMPTY SAMPLE
TUBES
SILICONE GREASELAM 004 A
GREASE FOR
MECHANICAL
ASSEMBLIES
SOFT TISSUE
FLAT PIECE OF
STIFF PLASTIC
XEA 381 AS
FLAT SCREWDRIVERLATEX RBCLAD 002 AS
BARFLEXFELT PEN
THERMOMETERSYRINGE 5ML
TORX KEYS
RAA 009 A Ind.A
Page 3/720/06/98
INSTALLATION
CONCERNS
1 - Unpacking
2 - Working conditions
3 - Visual checks
4 - Reagent connections
5 - Printer and instrument connections
6 - Priming and startup
REQUIRED TOOLS
Hexagonal keys
01/10/98
REQUIRED PRODUCTS
- MICROS 60 Reagents : Bottles or Pack.
- Waste container (for bottle model).
INTERVENTION TIME
- 30 minutes
FREQUENCY
SPECIFIC KIT OR CONSUMABLES
- PACK installation kit : XEA 314 A
or
- Bottle installation kit : XEA 332 A
MICROS 60 OT
RAS 165 A Ind.B
PROCEDURE
1 - Unpacking
INSTALLATION
The instrument is enveloped in a special,
protective foam before being placed in a
cardboard box. Cut the four angles of the box
to unpack the system.
Remove the cardboard box containing the ABXMICROS 60 installation kit from its location (see
Diag.1).
Diag.1
2 - Working conditions
• Environment
ABX MICROS 60-OT should be operated in an indoor location only. Operation at an altitude
over 2000 meters is not recommended. Instrument is designed to be safe for transient voltages according to INSTALLATION CATEGORY II and POLLUTION DEGREE 2.
• Location
ABX MICROS 60-OT should be placed on a clean and level table or work station. Please note
that ABX MICROS 60-OT, printer and reagents weigh approximately 30 kilograms (66 lbs).
Avoid exposure to sunlight. Proper ventilation requires that a space of at least 20 cm (8 inches)
must be left behind the apparatus.
• Grounding
Proper grounding is required. Check that the wall ground (earth) plug is correctly connected to
the laboratory grounding electricity installation. If there is no ground then use a ground stake.
Current electricity norms must be applied.
• Humidity and temperature conditions
ABX MICROS 60-OT can function between 18 to 32°C (65 to 90°F), with relative humidity,
meaning less than 80% with no condensation. If it is kept at a temperature less than 10°C
(50°F), the instrument should be allowed to sit for an hour at the correct room temperature
before use.
MICROS 60 OT
Page 2/13RAS 165 A Ind.B
INSTALLATION
3 - Visual checks
Diag.2
MICROS 60 OT
Using the key contained into the installation kit, turn
the locker as shown on the Diag. 2 to open the
pneumatic protection door.
Unscrew the 3 cover fixation screws (Diag. 3) and
loosen the 2 tightening screws under the reagent
flap (Diag 4).
Remove the cover : pull it backward and lift it up to
the rear of the instrument .
Diag.3
Diag.5
A
B
Diag.4
Push the black plastic carriage locking clip
(A) as far as possible to the left and place
the sample needle carriage (B) as far
forward as possible to the right-hand side,
as shown in Diag. 5.
Check that the needle is not bent and make
sure it is in its upper position.
RAS 165 A Ind.BPage 3/13
INSTALLATION
Unscrew slightly the 2 screws of the WBC/HGB chamber protection cover (diag 6). Remove
the cover and check that both chambers (RBC/PL T, WBC/HGB) are fixed properly in their clips
and the electrode blocks are attached firmly to the chambers (Diag. 7).
Diag.6
Re-install the HGB/WBC chamber cover.
Diag.8
Diag.7
Check that the connectors on the
printed circuit board are securely in
place (Diag. 8).
Re-install the instrument cover.
Remove the fuse holder from its location on
the rear panel pressing on the holder lock
(Diag.9) and check the fuse characteristics :
they should be 1 Ampere, 250 Volts SlowBlow.
MICROS 60 OT
Diag.9
Page 4/13RAS 165 A Ind.B
INSTALLATION
4 - Reagent connections
• Bottle connections
Diag.10
IMPORTANT
MICROS 60 OT
Lyse and cleaning reagents are placed inside the
reagent compartment as shown in the Diag. 10.
Install the reagent straws and the bottle stoppers.
Connect the blue tube to the MINICLEAN bottle
and the white tube to the MINILYSE bottle.
Close the compartment cover.
When the ABX MICROS 60-OT is set up with the 16 or 18 parameters mode, it is mandatory
to use specific MINILYSE LMG and MINIDIL LMG reagents.
CAUTION
The Diluent container will be located on the bench at the same level than the instrument.
• Waste connection
Install the male connectors included in the
installation kit at the liquid input and output located at the bottom of the instrument rear panel (Diag.11).
Connect the diluent container (see CAUTION above) using the diluent straw and
a 3x6 cristal tube (1 meter maximum) on
the diluent input located at the bottom of
the instrument rear panel (Diag.11).
Connect the waste container using the
cristal tube 3 x 6 on the waste output,
and place the waste container below the
instrument level (under the bench).
Diag.11
WARNING !
Always follow the recommended procedures for waste disposal. Never connect
the instrument wastes directly to the laboratory drain pipes. For each waste container, follow the neutralization procedure as described in the user manal.
RAS 165 A Ind.BPage 5/13
INSTALLATION
• Reagent pack connection
Remove the reagent output protections, as well as the waste input protection (Diag.12 & 13)
Diag.12
Install the pack directly into the compartment of the instrument as shown on the Diag. 14, 15 &
16. Push the pack down in order to plug correctly the pack on the male connectors.
Diag.13
Diag.14Diag.15
Diag.16
CAUTION
MICROS 60 OT
Page 6/13RAS 165 A Ind.B
In order to avoid leak problems it is recommended not to unplug several times the
same reagent pack.
The free male connector (see Diags.
14, 15 & 16) must be plugged on the
pack upper valve in order to receive
the waste liquids.
INSTALLATION
5 - Printer and instrument connections
• Instrument connection.
If the instrument has to be connected to a
laboratory computer, use the plug RS232.
Connect the power cable to the plug located
on the rear left-hand side of the device (Diag.
17).
• Setting up the printer
- Remove all the package protections
- Install the paper feed-knob
MICROS 60 OT
Diag.17
Diag.19
Diag.18
Open the ink ribbon access door at the top of the printer and install the ribbon as shown in diag
20 & 21 :Slide the printer head to the middle of the printer.
Insert the ribbon cartridge into the printer
Guide the ribbon between the print head and ribbon guide.
Slide the printer head from side to side to make sure it moves smoothly.
Diag.20
Diag.21
RAS 165 A Ind.BPage 7/13
INSTALLATION
- Install the paper supports (Diag 22) for single sheets paper use only.
Diag.22
- The printer is connected to ABX MICROS 60-OT with the cable delivered with the instrument.
Lock the connector in place by tightening the 2 screws on each end of the connector to the
ABX MICROS 60-OT. Attach the other end of the cable to the printer and lock the printer
connector in place by the means of the 2 clips located on the connector itself (Diag. 23).
Diag.23
- For continuous paper, introduce it in the slot at the back of the printer and use the sprocket
covers to load paper, or else feed the paper frontward when using single sheets (see printer
user manual).
MICROS 60 OT
Diag.24
Diag.25
Page 8/13RAS 165 A Ind.B
INSTALLATION
• Printer command keys
Diag.26
LED PAUSE : The orange LED PAUSE lights when the printer stops printing. During each
power ON, this LED blinks for few seconds and 4 audible beeps occur. When the printer runs
out of paper, the LED blinks and 3 audible beeps occur. This LED lights also when the paper is
in its tear off position. When a problem occurs, this LED lights ON and 5 audible beeps occur.
LEDS FONT 1 and FONT 2 : These 2 green LEDS indicate the selected font. Refer to the
printer user's manual to select the font.
MICROS 60 OT
Key FONT : During normal operation, the FONT key allows the font selection. For each pressure on this key , the selection is modified. Refer to the printer user's manual to select the font.
When this key is pressed during the printer power ON, the printer setup menu is entered.
Key LF/FF : During normal operation, a quick pressure on this key allows a ligne feed of the
paper. Keep the pressure on this key to feed a whole page. This key can be used to load or
eject the paper.
When this key is pressed during the printer power ON, the printing test starts. Key PAUSE :
When this key is pressed during the printing, the printout stops. Press again on this key to
restart the printout.
PAPER PARK : If Z folded paper is used, the paper can be driven to its parking position when
pressing simultaneously on the keys LF/FF and FONT.
MICRO ADJUST : This function allows to adjust the loading paper position. See the user's
manual for details.
• Printer Configuration
Switch on the printer when pressing the <FONT>. The configuration should be the following :
Diag.29
RAS 165 A Ind.BPage 9/13
INSTALLATION
Each parameter can be modified by the corresponding parameter chart.
Each chart is accessible using the keys <PAUSE>, <FONT> and <LF/FF> according to the
control LED combinations (Diag 30).
Diag.30
6 - Instrument startup
• Reagent priming
When the ABX MICROS 60-OT is first installed, it contains no reagents. All the reagents have
to be primed now. Turn ON instrument by pressing the ON/OFF switch located on the rear
panel. When the instrument turns on, the display shows :
PLEASE WAIT FOR 3 MIN
ESCAPE : ESC
This time is required at the startup for the instrument initialization and stabilization, specifically
for the HGB diode to reach its operationnal temperature. Press ESC several times in order to
abort the cycle : the LED of the front panel turns from red to green and the display shows the
following :
STARTUP NOT INITIATED
PRESS A KEY TO CONTINUE...
This message appears when the instrument is setup with the manual startup cycle to prevent
any analysis cycle before running a startup cycle.
Press any key, the main menu is displayed :
MAIN MENU1 RESULTS
HH : MM2 QC
MICROS 60 OT
Page 10/13RAS 165 A Ind.B
INSTALLATION
• Bottles and containers set :
Install the reagent bottle and carry out a PRIME cycle to clear the reagent line of air bubbles.
This procedure should be done whenever a new bottle of reagent is installed.
From the MAIN MENU, move the cursor to the function 4 SERVICE and press ENTER. The
service menu is displayed :
SERVICE1 BACK FLUSH
HH : MM2 DRAIN CHAMBERS
Move the cursor to function 3 PRIME REAGENTS and press ENTER. The PRIME menu is
displayed :
PRIME1 ALL REAGENTS
HH : MM2 DILUENT
MICROS 60 OT
Select either the function 1 prime ALL REAGENTS or move the cursor next to the required
reagent and press ENTER.
The priming cycle starts while the following menu is displayed :
ALL REAGENTSWAIT FOR 2 MN 3 S
******************
IMPORTANT
Before analyzing samples, visually inspect reagent lines and pumps for air bubbles.
Repeat priming if air bubbles are still present. Call the
department
Never initiate two Lyse prime cycles back-to-back. This causes excessive foaming in
the waste chamber. Run a blank cycle between each Lyse prime cycle.
- Run a STARTUP cycle.
if priming does not eliminate air bubbles.
ABX representative service
RAS 165 A Ind.BPage 11/13
INSTALLATION
• Reagent pack
From the MAIN MENU, move the cursor to the function 4 SERVICE and press ENTER. The
service menu is displayed :
SERVICE> 2 DRAIN CHAMBER
HH : MM 3 PRIME REAGENTS
Move the cursor to 3 PRIME REAGENTS and press the ENTER key. Select the function
1
CHANGE PACK and follow the instructions given by the LCD in order to install the pack.
REAGENT PACK> 1 CHANGE PACK
HH : MM 2 CBC LEFT < 150>
Once the new PACK is installed a priming cycle will be automatically carried out and the
following menu is displayed.
PRIMEWAIT FOR 2 MIN 3 S
* * * * * * * * * *
IMPORTANT
Before analyzing samples, visually inspect reagent lines and pumps for air bubbles.
Repeat priming if air bubbles are still present. Call the
department
From the REAGENT P ACK menu, the function 2 "CBC LEFT" displays the number of analysis
cycles left to run with the same pack.
It is also possible to run a priming cycle at any time using the selection 3 "PRIME REAGENTS"
of the SERVICE menu.
if priming does not eliminate air bubbles.
ABX representative service
IMPORTANT
It is recommended not to remove the pack several times before the reagents are
totally used in order to avoid leak problems.
- Run a STARTUP cycle.
MICROS 60 OT
Page 12/13RAS 165 A Ind.B
INSTALLATION
Once the instrument ready for the analyses, remove the adhesive protection from the front
panel (diag 31)
Diag.31
MICROS 60 OT
RAS 165 A Ind.BPage 13/13
INSTALLATION
CONCERNS
1 - Unpacking
2 - Working conditions
3 - Visual checks
4 - Reagent connections
5 - Printer and instrument connections
6 - Priming and startup
REQUIRED TOOLS
Hexagonal keys
01/10/98
REQUIRED PRODUCTS
Reagents : Bottles or Pack.
Waste container (for bottle model).
INTERVENTION TIME
30 minutes
FREQUENCY
SPECIFIC KIT OR CONSUMABLES
MICROS 60 CT
PACK installation kit : XEA 317 A or
Bottle installation kit : XEA 335 A or
RAS 166 A Ind.B
PROCEDURE
1 - Unpacking
INSTALLATION
The instrument is enveloped in a special,
protective foam before being placed in a
cardboard box. Cut the four angles of the box
to unpack the system.
Remove the cardboard box containing the instrument installation kit from its location (see
Diag.1).
Diag.1
2 - Working conditions
• Environment
ABX MICROS 60-
over 2000 meters is not recommended. Instrument is designed to be safe for transient voltages according to INSTALLATION CATEGORY II and POLLUTION DEGREE 2.
• Location
ABX MICROS 60-CT should be placed on a clean and level table or work station. Please note
that ABX MICROS 60-CT, printer and reagents weigh approximately 30 kilograms (66 lbs).
Avoid exposure to sunlight. Proper ventilation requires that a space of at least 20 cm (8 inches)
must be left behind the apparatus.
• Grounding
Proper grounding is required. Check that the wall ground (earth) plug is correctly connected to
the laboratory grounding electricity installation. If there is no ground then use a ground stake.
Current electricity norms must be applied.
• Humidity and temperature conditions
ABX MICROS 60-CT can function between 18 to 32°C (65 to 90°F), with relative humidity,
meaning less than 80% with no condensation. If it is kept at a temperature less than 10°C
(50°F), the instrument should be allowed to sit for an hour at the correct room temperature
before use.
CT
should be operated in an indoor location only. Operation at an altitude
MICROS 60 CT
Page 2/13RAS 166 A Ind.B
INSTALLATION
3 - Visual checks
Diag.2
MICROS 60 CT
Using the key contained into the installation kit, turn
the locker as shown on the Diag. 2 to open the
pneumatic protection door.
Unscrew the 3 cover fixation screws (Diag. 3) and
loosen the 2 tightening screws under the reagent
flap (Diag 4).
Remove the cover : pull it backward and lift it up to
the rear of the instrument.
Diag.3
Diag.5
Diag.4
Push the black plastic carriage locking clip
(A) as far as possible to the left and place
the sample needle carriage (B) as far
forward as possible to the right-hand side,
as shown in Diag. 5.
Check that the needle is not bent and make
sure it is in its upper position.
RAS 166 A Ind.BPage 3/13
INSTALLATION
Unscrew slightly the 2 screws of the WBC/HGB chamber protection cover (diag 6). Remove
the cover and check that both chambers (RBC/PL T, WBC/HGB) are fixed properly in their clips
and the electrode blocks are attached firmly to the chambers (Diag. 7).
Re-install the HGB/WBC chamber cover.
Diag.8
Diag.7Diag.6
Check that the connectors on the
printed circuit board are securely in
place (Diag. 8).
Re-install the instrument cover.
Remove the fuse holder from its location on
the rear panel pressing on the holder lock
(Diag.9) and check the fuse characteristics :
they should be 1 Ampere, 250 Volts SlowBlow.
MICROS 60 CT
Diag.9
Page 4/13RAS 166 A Ind.B
INSTALLATION
4 - Reagent connections
• Bottle connections
Lyse and cleaning reagents are placed inside the
reagent compartment as shown in the Diag. 10.
Install the reagent straws and the bottle stoppers.
Connect the blue tube to the MINICLEAN bottle
and the white tube to the MINILYSE bottle.
Close the compartment cover.
Diag.10
IMPORTANT
When the ABX MICROS 60-CT is set up with the 16 or 18 parameters mode, it is mandatory
to use specific MINILYSE LMG and MINIDIL LMG reagents.
MICROS 60 CT
CAUTION
The Diluent container will be located on the bench at the same level than the instrument.
• Waste connection
Install the male connectors included in the
installation kit at the liquid input and output located at the bottom of the instrument rear panel (Diag.11).
Connect the diluent container (see CAUTION above) using the diluent straw and
a 3x6 cristal tube (1 meter maximum) on
the diluent input located at the bottom of
the instrument rear panel (Diag.11).
Connect the waste container using the
cristal tube 3 x 6 on the waste output,
and place the waste container below the
instrument level (under the bench).
Diag.11
WARNING !
Always follow the recommended procedures for waste disposal. Never connect
the instrument wastes directly to the laboratory drain pipes. For each waste container, follow the neutralization procedure as described in the user manal.
RAS 166 A Ind.BPage 5/13
INSTALLATION
• Reagent pack connection
Remove the reagent output protections, as well as the waste input protection (Diag.12 & 13)
Diag.12
Install the pack directly into the compartment of the instrument as shown on the Diag. 14, 15 &
16. Push the pack down in order to plug correctly the pack on the male connectors.
Diag.13
Diag.14Diag.15
Diag.16
MICROS 60 CT
Page 6/13RAS 166 A Ind.B
CAUTION
In order to avoid leak problems it is recommended not to unplug several times the
same reagent pack.
The free male connector (see Diags.
14, 15 & 16) must be plugged on the
pack upper valve in order to receive
the waste liquids.
INSTALLATION
5 - Printer and instrument connections
• Instrument connection.
If the instrument has to be connected to a
laboratory computer, use the plug RS232.
Connect the power cable to the plug located
on the rear left-hand side of the device (Diag.
17).
• Setting up the printer
- Remove all the package protections
- Install the paper feed-knob
MICROS 60 CT
Diag.17
Diag.19
Diag.18
Open the ink ribbon access door at the top of the printer and install the ribbon as shown in diag
20 & 21 :Slide the printer head to the middle of the printer.
Insert the ribbon cartridge into the printer
Guide the ribbon between the print head and ribbon guide.
Slide the printer head from side to side to make sure it moves smoothly.
Diag.20
Diag.21
RAS 166 A Ind.BPage 7/13
INSTALLATION
- Install the paper supports (Diag 22) for single sheets paper use only.
Diag.22
- The printer is connected to ABX MICROS 60-CT with the cable delivered with the instrument.
Lock the connector in place by tightening the 2 screws on each end of the connector to the
ABX MICROS 60-CT. Attach the other end of the cable to the printer and lock the printer
connector in place by the means of the 2 clips located on the connector itself (Diag. 23).
- For continuous paper, introduce it in the slot at the back of the printer and use the sprocket
covers to load paper, or else feed the paper frontward when using single sheets (see printer
user manual).
MICROS 60 CT
Diag.24
Diag.23
Diag.25
Page 8/13RAS 166 A Ind.B
INSTALLATION
• Printer command keys
Diag.26
LED PAUSE : The orange LED PAUSE lights when the printer stops printing. During each
power ON, this LED blinks for few seconds and 4 audible beeps occur. When the printer runs
out of paper, the LED blinks and 3 audible beeps occur. This LED lights also when the paper is
in its tear off position. When a problem occurs, this LED lights ON and 5 audible beeps occur.
LEDS FONT 1 and FONT 2 : These 2 green LEDS indicate the selected font. Refer to the
printer user's manual to select the font.
Key FONT : During normal operation, the FONT key allows the font selection. For each pressure on this key , the selection is modified. Refer to the printer user's manual to select the font.
When this key is pressed during the printer power ON, the printer setup menu is entered.
MICROS 60 CT
Key LF/FF : During normal operation, a quick pressure on this key allows a ligne feed of the
paper. Keep the pressure on this key to feed a whole page. This key can be used to load or
eject the paper.
When this key is pressed during the printer power ON, the printing test starts. Key PAUSE :
When this key is pressed during the printing, the printout stops. Press again on this key to
restart the printout.
PAPER PARK : If Z folded paper is used, the paper can be driven to its parking position when
pressing simultaneously on the keys LF/FF and FONT.
MICRO ADJUST : This function allows to adjust the loading paper position. See the user's
manual for details.
• Printer Configuration
Switch on the printer when pressing the <FONT>. The configuration should be the following :
Diag.29
RAS 166 A Ind.BPage 9/13
INSTALLATION
Each parameter can be modified by the corresponding parameter chart.
Each chart is accessible using the keys <PAUSE>, <FONT> and <LF/FF> according to the
control LED combinations (Diag 30).
Diag.30
6 - Instrument startup
• Reagent priming
When the ABX MICROS 60-CT is first installed, it contains no reagents. All the reagents have
to be primed now. Turn ON instrument by pressing the ON/OFF switch located on the rear
panel. When the instrument turns on, the display shows :
PLEASE WAIT FOR 3 MIN
ESCAPE : ESC
This time is required at the startup for the instrument initialization and stabilization, specifically
for the HGB diode to reach its operational temperature. Press the ESC key in order to abort the
cycle : the LED of the front panel turns from red to green and the display shows :
STARTUP NOT INITIATED
PRESS A KEY TO CONTINUE...
This message appears when the instrument is setup with the manual startup cycle to prevent
any analysis cycle before running a startup cycle.
Press any key, the main menu is displayed :
MAIN MENU1 RESULTS
HH : MM2 QC
MICROS 60 CT
Page 10/13RAS 166 A Ind.B
INSTALLATION
Bottles and containers set :
Install the reagent bottle and carry out a PRIME cycle to clear the reagent line of air bubbles.
This procedure should be done whenever a new bottle of reagent is installed.
From the MAIN MENU, move the cursor to the function 4 SERVICE and press ENTER. The
service menu is displayed :
SERVICE1 BACK FLUSH
HH : MM2 DRAIN CHAMBERS
Move the cursor to function 3 PRIME REAGENTS and press ENTER. The PRIME menu is
displayed :
PRIME1 ALL REAGENTS
HH : MM2 DILUENT
Select either the function 1 prime ALL REAGENTS or move the cursor next to the required
reagent and press ENTER.
The priming cycle starts while the following menu is displayed :
MICROS 60 CT
ALL REAGENTSWAIT FOR 2 MN 3 S
******************
IMPORTANT
Before analyzing samples, visually inspect reagent lines and pumps for air bubbles.
Repeat priming if air bubbles are still present. Call the
department
Never initiate two Lyse prime cycles back-to-back. This causes excessive foaming in the
waste chamber. Run a blank cycle between each Lyse prime cycle.
- Run a STARTUP cycle.
if priming does not eliminate air bubbles.
ABX representative service
RAS 166 A Ind.BPage 11/13
INSTALLATION
Reagent pack
From the MAIN MENU, move the cursor to the function 4 SERVICE and press ENTER. The
service menu is displayed :
SERVICE> 2 DRAIN CHAMBER
HH : MM 3 PRIME REAGENTS
Move the cursor to 3 PRIME REAGENTS and press the ENTER key. Select the function
1
CHANGE PACK and follow the instructions given by the LCD in order to install the pack.
REAGENT PACK> 1 CHANGE PACK
HH : MM 2 CBC LEFT < 150>
Once the new PACK is installed a priming cycle will be automatically carried out and the
following menu is displayed.
PRIMEWAIT FOR 2 MIN 3 S
* * * * * * * * * *
IMPORTANT
Before analyzing samples, visually inspect reagent lines and pumps for air bubbles.
Repeat priming if air bubbles are still present. Call the
department
From the REAGENT P ACK menu, the function 2 "CBC LEFT" displays the number of analysis
cycles left to run with the same pack.
It is also possible to run a priming cycle at any time using the selection 3 "PRIME REAGENTS"
of the SERVICE menu.
if priming does not eliminate air bubbles.
ABX representative service
IMPORTANT
It is recommended not to remove the pack several times before the reagents are
totally used in order to avoid leak problems.
- Run a STARTUP cycle.
MICROS 60 CT
Page 12/13RAS 166 A Ind.B
INSTALLATION
Once the instrument ready for the analyses, remove the adhesive protection from the front
panel (diag 31)
Diag.31
MICROS 60 CT
RAS 166 A Ind.BPage 13/13
CONCERNS
SAMPLING NEEDLE
MAINTENANCE
- Needle O ring replacement
- Sampling needle replacement
- Piercing needle replacement
REQUIRED TOOLS
- Hexagonal keys
- Dynamometric screw driver A302 : MAG 019 A
REQUIRED PRODUCTS
- Silicone grease : LAM 004 A
18/06/98
INTERVENTION TIME
- 15min
FREQUENCY
- See frequency chart table for cleaning.
- O ring replacement : 1/year
- Needle replacement : On request only
SPECIFIC KIT OR CONSUMABLES
- O ring kit : XEA 328 AS
- Sampling needle : GBC 052 A
- Piercing needle : GBC 189 A
MICROS 60 CT
RAS 167 A Ind.A
PROCEDURE
SAMPLING NEEDLE MAINTENANCE
A - Needle or O ring replacement
- Disconnect the tube
1
from the
Diag.1
top of the needle
Remove the clip
manually lift up the sampling needle
(Diagram 1).
- Unscrew the 2 screws 3(Diagram
1) in order to freed the needle rinsing
block
- Remove the rinsing block/needle
assy from the carriage taking care not
to bend the needle.
- Lift up the O ring holder
4
from the carriage frame
5
.
2
(diagram 1).
3
(diagram 2) and
1
and re-
Diag.2
Proceed the same way to replace the needle if necessary
- Reassemble in the reverse order. Use a dynamometric screw driver to tighten the screws
(Diag.1) to 700 mN.m(99.4 Ozf.in).
MICROS 60 CT
place the O rings
previously greased. Wipe all excess of
grease away .
- If necessary clean the inner surface
of the rinsing block with a little piece of
paper.
2
by new ones
3
Page 2/3RAS 167 A Ind.A
SAMPLING NEEDLE MAINTENANCE
B - Piercing needle replacement
- Disconnect the tube 1from the piercing needle
4
(Diag.3).
- Lift the needle 2in the upper position.
- Loosen the 2 screws
enough to enable the rinsing block
to be lifted up of about 5 mm.
- Pull the piercing needle
(foreward) and replace it by a new
one if necessary .
- Replace the piercing needle O ring
5
(FAA 036 A) by a new one.
- Reassemble in the reverse order.
Apply the same torque : 700mN.m(99.4 Ozf.in) to tighten the two
screws 3.
MICROS 60 CT
3
just
4
Diag.3
NOTE
Diag.4
It is recommended to check the correct motion of the needle.
Proceed as follows :
Enter the «SERVICE» menu and then the sub menu «MECHANIC»
and perform a «NEEDLE U/D» cycle.
Blockage problems may occur on some
instruments during the needle or carriage
motions giving some motor error messages.
Before replacing the concerned motor, it is
necessary to check the correct rotation of
the free puley located at the end of the
notched belt.
Remove the axle screw of the puley and
clean its 2 parts and the washer. Reinstall
the puley assy, the rounded edge facing
the puley. Tighten the screw with a torque
of 400mN.m (56.8 Ozf.in).
Check that the puley turns freely after the tightening. Add a drop of oil (LAM 007 A) if necessary .
1 - FAG 011 A : Autolub. axle
2 - GBC 146 A : Free pulley
3 - GBC 147 A : Pulley holder
RAS 167 A Ind.APage 3/3
CONCERNS
SAMPLING NEEDLE
MAINTENANCE
- Needle replacement
- O ring replacement
REQUIRED TOOLS
- Hexagonal keys : 2,5
- Dynamometric screw driver A300 : MAG 013 A
REQUIRED PRODUCTS
- Silicone grease : LAM 004 A
10/04/98
INTERVENTION TIME
- 15 min
FREQUENCY
- Needle replacement : On request only
- O ring replacements : 1/year
SPECIFIC KIT OR CONSUMABLES
- O ring kit : XEA 328 AS
- Spare parts kit : XEA 458 AS
MICROS 60 OT
RAS 168 A Ind.A
PROCEDURE
SAMPLING NEEDLE MAINTENANCE
- Disconnect the tube
(Diagram 1).
- Unscrew the 2 screws
frame. Remove the clip
7
from the top of the needle. Manually lift up the sampling needle
5
in order to freed the needle rinsing block
6
.
2
from the carriage
Diag.1
- Remove the rinsing block/needle assy from the carriage taking care not to bend the needle.
- If necessary clean the inner surface of the rinsing block by means of a little piece of soft paper.
Spread a little amount of grease in between the rinsing block
- Lift up the O ring holder
Wipe all excess of grease away .
NOTE
4
and replace the O rings
Proceed the same way to replace the needle if necessary
MICROS 60 OT
Page 2/3RAS 168 A Ind.A
2
and its support.
3
by new ones previously greased.
SAMPLING NEEDLE MAINTENANCE
MICROS 60 OT
- Reassemble in the reverse order. Use a dynamometric screw driver to tighten the screws
(Diag.1) to 100 mN.m.
It is recommended to check the correct motion of the needle.
NOTE
Blockage problems may occur on some instruments during the needle or carriage motions
giving some motor error messages.
Before replacing the concerned motor, it is necessary to check the correct rotation of the free
pulley located at the end of the notched belt.
Remove the axle screw of the pulley and clean its 2 parts and the washer.Reinstall the pulley
assy, the rounded edge facing the pulley. Tighten the screw with a torque of 400mN.m (99.4Ozf.in).
Check that the pulley turns freely after the tightening. Add a drop of oil (LAM 007 A) if necessary .
Proceed as follows :
Enter the «SERVICE» menu and then the sub menu «MECHANIC»
and perform a «NEEDLE UP/DOWN» cycle.
- Run a drain chamber cycle
(SERVICE menu, DRAIN
CHAMBERS Sub menu).
- Record the tube positions
before dismantling the
chambers.
- Disconnect the chamber tubes.
- Unclip the RBC chamber.
- Dismantle the electrode
6
loosening the 2 fixation
Diag.1
screws
holding screw 3(Diagram 1).
• Electrode O ring replacement :
- Use a previously cut
micropipette tip to replace the
electrode O ring as shown on
Diag.2.
• Aperture O ring replacement :
- Install the chamber over a piece
of white paper or cloth.
- Carefully remove the counting
head
7
and the terminal
4
and plunge the aper-
Diag.2
- Clean the chamber and the counting head with liquid soap, do not introduce any sharp instru-
MICROS 60
Page 2/3RAS 169 A Ind.A
ments inside so as to avoid damaging the inside of the chamber and the aperture.
ture 5in distilled water.
Replace the O ring 2by a new
one.
CHAMBER MAINTENANCE
CAUTION
Do not manipulate the aperture using hard instruments. Clean the aperture
with a piece of soft paper or preferably, in between 2 fingers.
- Rinse thoroughly with distilled water
- Dry the exterior of the chamber with a soft paper.
CAUTION
- Do not apply too much pressure on the electrode fixation screws, as it can
break the aperture (tightening torque = 100mN.m / 14.2 Ozf.in).
- It is recommended to reconnect the tubes on the counting head before
reassembling the "electrode/chamber/counting head" assy in order to avoid
applying constraint on the chamber.
MICROS 60
- Position the chamber in its fixation clips.
- Reconnect the tubes
2 - WBC/HGB chamber cleaning :
Diag.3
- Run a drain chamber cycle
(SERVICE menu, DRAIN
CHAMBERS Sub menu).
- Loosen the cover screws of
the WBC/HGB chamber and
remove the cover.
- Record the tube positions
before dismantling the
chambers.
- Disconnect the chamber tubes.
- Unclip the RBC chamber.
- Dismantle the electrode
6
loosening the 2 fixation
screws
nal holding screw
- Proceed as described in
RBC chamber
chamber and to replace the
lectrode and aperture O
rings.
7
and the termi-
3
to clean the
.
1 -
NOTE
The spectrophotometer can not be dismantled from the chamber. If this one
has been dammaged it is necessary to replace the whole chamber assy .
When cleaning the spectrophotometer, make sure to thoroughly rinse it in
order to obtain a correct HGB blank measure.
RAS 169 A Ind.APage 3/3
CONCERNS
LIQUID VALVE
MAINTENANCE
Liquid valve assy replacement
V alve body replacement
REQUIRED TOOLS
Hexagonal keys
Pair of pliers
REQUIRED PRODUCTS
Soft paper
14/04/98
INTERVENTION TIME
15 min
FREQUENCY
On request only
SPECIFIC KIT OR CONSUMABLES
6 Valve assembly (MICROS 60 CT) : XDA 579 CS
6 V alve assembly (MICROS 60 OT) : XDA 578 CS
5 V alve assembly : XDA 580 CS
2 ways NC liquid valve without solenoïd : XDA 481 B
3 ways liquid valve without solenoïd : XDA 483 B
Solenoïd 24V 4W : EAZ 004 A
MICROS 60
RAS 170 A Ind.A
LIQUID VALVE MAINTENANCE
PROCEDURE
- Several reasons can unable the correct operations of the valves :
- Leaks on the valve bodies
- Defective Solenoïds
- Liquid discharge on valves
- Corrosion traces on the axis, ect...
- Run a «DRAIN CHAMBERS» cycle.
- Switch off the instrument.
- Note the tube positions on the valve assembly that requires to be dismantled.
- Unscrew the valve assembly fixation screws (Diag.1).
Diag.1
- Disconnect the supplying flat cable from the valve assembly (diag.2).
When leaks occur on valves it is recommended to replace the
NOTE
- If only one valve has been damaged it is possible to dismantle the valve body on its own as
shown on the diagram 3 : use a pair of pliers to disconnect the valve holder clip and remove the
body.
entire valve assembly by a new one.
CAUTION
When replacing one valve only , it is important to check if the «neighboured»
MICROS 60
Page 2/3RAS 170 A Ind.A
valves have not been damaged too.
Diag.2
LIQUID VALVE MAINTENANCE
When replacing one valve only, it is important to check if the «neighboured» valves have not
been damaged too.
Diag.3
MICROS 60
CAUTION
The solenoïd can not be dismantled unless unsoldering it.
If this one is suspected solder a new one or replace the entire valve assembly .
- Reconnect the tubes on the nipples (with the sleeves).
- Re-install in the reverse order. Switch on the instrument.
- Control the watertightness of the valves and check for the correct operations :
Go to «SERVICE» menu, then to «MECHANIC» sub menu (5) and require a «V AL VES» (6) test
(see Mechanic functions : RAS 173 A).
- Check the calibration too.
• REMARKS
On MICROS 60 - CT, when piercing several times the same tube cap, some pieces of cork may
be dragged along towards the WBC/HGB chamber and then the liquid valve .
This may damaged the operation of the valve.
RAS 170 A Ind.APage 3/3
CONCERNS
POWER SUPPLY CHECK
OR REPLACEMENT
- Voltage supply check
- Power supply module replacement
- Fan operation check
REQUIRED TOOLS
- Hexagonal keys
- Flat screw driver
- Volmeter
REQUIRED PRODUCTS
- None
10/04/98
INTERVENTION TIME
- 15 min
FREQUENCY
- 1/year
SPECIFIC KIT OR CONSUMABLES
- None
MICROS 60
RAS 171 A Ind.A
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.