ABB EXGP Installation Manual

Page 1
ABB Instrumentation
EXGP Oxygen Analyzer System
Installation Guide
EXGP Probe
Page 2
ABB INSTRUMENTATION
Note.
Clarification of an instruction or additional information.
Information.
Further reference for more detailed information or technical details.
Although Warning hazards are related to personal injury, and Caution hazards are associated with equipment or property damage, it must be understood that operation of damaged equipment could, under certain operational conditions, result in degraded process system performance leading to personal injury or death. Therefore, comply fully with all Warning and Caution notices.
Information in this manual is intended only to assist our customers in the efficient operation of our equipment. Use of this manual for any other purpose is specifically prohibited and its contents are not to be reproduced in full or part without prior approval of Technical Communications Department, ABB Instrumentation.
The Company
ABB Instrumentation is an established world force in the design and manufacture of instrumentation for industrial process control, flow measurement, gas and liquid analysis and environmental applications.
As a part of ABB, a world leader in process automation technology, we offer customers application expertise, service and support worldwide.
We are committed to teamwork, high quality manufacturing, advanced technology and unrivalled service and support.
The quality , accuracy and performance of the Company’s products result from over 100 years experience, combined with a continuous program of innovative design and development to incorporate the latest technology.
The NAMAS Calibration Laboratory No. 0255 is just one of the ten flow calibration plants operated by the Company, and is indicative of ABB Instrumentation’s dedication to quality and accuracy.
BS EN ISO 9001
St Neots, U.K. – Cert. No. Q5907 Stonehouse, U.K. – Cert. No. FM 21106
EN 29001 (ISO 9001)
Lenno, Italy – Cert. No. 9/90A
Use of Instructions
Warning.
An instruction that draws attention to the risk of injury or death.
Caution.
An instruction that draws attention to the risk of damage to the product, process or surroundings.
Health and Safety
To ensure that our products are safe and without risk to health, the following points must be noted:
1. The relevant sections of these instructions must be read carefully before proceeding.
2. Warning labels on containers and packages must be observed.
3. Installation, operation, maintenance and servicing must only be carried out by suitably trained personnel and in accordance with the information given.
4. Normal safety precautions must be taken to avoid the possibility of an accident occurring when operating in conditions of high pressure and/or temperature.
5. Chemicals must be stored away from heat, protected from temperature extremes and powders kept dry. Normal safe handling procedures must be used.
6. When disposing of chemicals ensure that no two chemicals are mixed.
Safety advice concerning the use of the equipment described in this manual or any relevant hazard data sheets (where applicable) may be obtained from the Company address on the back cover, together with servicing and spares information.
0255
R
E
G
I
S
T
E
R
E
D
Stonehouse, U.K.
Page 3
1
CONTENTS
1.1 Documentation – Fig. 1.1
Documentation for the EXGP Oxygen Analyzer System is shown in Fig. 1.1.
1.2 Certification
The EXGP Probe is classed as a 'Simple Apparatus' and may therefore be installed in Zone 1 Hazardous Areas as an intrinsically safe product.
✶
Note. A 'Simple Apparatus' is a device that does not
generate more than 1.2V, 0.1A, 20mJ and 25mW.
Documented evidence confirming the compliance of EXGP Probes is available and is endorsed by BASEEFA.
Fig. 1.1 System Documentation
SPECIFICATION SHEET
Full Specification■
Part No.
SS/EXGP
PROBE
■
■
■
■
Part No.
IM/EXGP–PB
Product Identification Mechanical Installation Connections Fault Finding
INTERFACE UNIT
Certification Product Identification Mechanical Installation Electrical Installation
■
■
■
■
Part No.
IM/EXGP–INT
TRANSMITTER
Product Identification Mechanical Installation Electrical Connections Controls & Displays Operation Programming
■
■
■
■
■
■
Part No.
IM/EXGP–4600
1 INTRODUCTION
1 INTRODUCTION ..........................................................1
1.1 Documentation ..................................................1
1.2 Certification .......................................................1
1.3 System Hardware..............................................2
2 PREPARATION ............................................................4
2.1 Unpacking .........................................................4
2.2 Checking the Code Number..............................4
2.3 Assembling the Probe .......................................5
3 INSTALLATION............................................................6
3.1 Siting..................................................................6
3.2 Mounting............................................................7
3.2.1 Mounting Configurations ..................... 8
4 CONNECTIONS ...........................................................9
4.1 Cable and Gland Specifications........................9
4.2 Access to Electrical Connections......................9
4.3 Electrical Connections ......................................9
4.4 Reference Air Connection .................................9
4.5 Routine Maintenance ........................................9
5 FAULT FINDING.........................................................10
5.1 In Situ Checking Using a Test Gas .................10
5.2 Comparison with Another O2 Probe ...............10
5.3 Returning the Probe
to the Manufacturer for Checking ...................10
5.4 Continuity Check .............................................10
APPENDIX...........................................................................11
A1 Principles of Operation....................................11
A2 Range of Operation.........................................11
Page 4
2
1.3 System Hardware – Fig. 1.2
The EXGP Oxygen Probe is an 'in situ' method of measuring oxygen concentration designed for use in hazardous areas (Zone 1). When used in conjunction with the EXGP Interface Electronics Unit (also Zone 1 Certified), it enables signals to be retransmitted up to 1km into a safe area.
The oxygen probe provides a mV signal that relates directly to oxygen concentration in the range 0.25 to 25% O
2
, over a process temperature range of 600°C to 1250°C. An integral 'R' type thermocouple provides accurate continuous measurement of the process temperature.
A calibration gas inlet port is provided to enable the probe to be checked using test gas mixtures without removing it from the process.
Reference air supplied at a rate of 100 to 1000cm
3
/hr (0.2 to
2ft
3
/hr) is required for accurate operation and can be provided
by a pump or regulator unit – see Fig. 1.2.
…1 INTRODUCTION
Page 5
3
Mains Powered Pump
Unit (003000240)
NON HAZARDOUS AREAHAZARDOUS AREA
(Zone 1, Class IIB)
Process
600°C to 1400°C
(1250°C
continuous)
EXGP Probe
Simple Device Conforming to BS5345 Part 4: 1977
Mains Supply
R Type Thermocouple mV Signal (0 to 1400°C)
Two 4 to 20mA Output Signals (0 to 1400°C and 0.25 to 25%O2)
or
4680 Transmitter
Mains Supply
Flameproof Interface Electronics Unit
EExd. IIB T6 Conforming to BS5501 Part 5: 1977
Flue
4685 Transmitter
Retransmission
Output
Reference Air Line Entry
Regulated Reference Air
Unit (003000241)
Mains Supply
O2 mV Signal (–20 to +180mV)
or
or
1 INTRODUCTION
Fig. 1.2 System Schematic
Page 6
4
2.1 Unpacking
Caution. The probe is very fragile. Unpack and
handle with care.
Each probe is despatched with its components individually placed within a common pack. Remove the probe from its packing as follows:
a) Place the polystyrene pack, top uppermost, on a flat
surface and cut the tapes holding the two halves together.
b) Carefully lift the top half of the pack to gain access to the
probe components which are individually located within the bottom half.
c) When removing the components from the pack take great
care to support the probe head and not to twist or bend the electrode assembly attached to it.
d) Reassemble the packing and retain for future use.
2.2 Checking the Code Number – Fig. 2.1
Before installing the equipment, check the code number on the probe – see Fig. 2.1.
Code No.
TEST GAS
KEEP TIGHTLY CLOSED WHEN NOT IN USE
Serial No.
AIR REF
THC
I/P
Remove cover retaining screws (x3)
Rotate cover
Check probe code number against Table 2.1
1
2
3
eborPPGXE /PGXE XXX000
htgneL enoN
mm006
mm0001 laicepS
0 1 2 9
htaehS enoN
008yolocnI
nialecroPsuonimulA
laicepS
0 1 2 9
smetIdradnatS enoN
59CepyTdnalgniwT
)7394SBhR%31tP/tPelpuocomrehtepyt’R‘,detnuomegnalF(
0 1
2 PREPARATION
Fig. 2.1 Checking the Code Number
Table 2.1 Code Number Interpretation
Page 7
5
2.3 Assembling the Probe – Fig. 2.2
Caution. Take care to avoid damage to the probe when fitting the sheath assembly.
Wadding
Filter Washer
Vent Holes
Secure countersunk-head screws (x3) in fixing flange using the lock nuts (x3).
Position the clamp ring over the boss on the probe head…
Position the sealing ring over the boss, ensuring it is located correctly in the centre taper.
Slide the fixing flange over the sheath, ensuring it is seated over the locating step.
Drill the fixing flange to suit fixing requirements
Secure the sheath to the boss (x3)
…ensuring the wadding at the probe end is not damaged and the filter washer is in position at the end of the sheath
…ensuring correct orientation of the ring's center taper
Locate the clamp ring over the 3 screws on the fixing flange and secure using nuts (x3).
Tighten the nuts down equally so that the fixing flange is pulled down tightly against the locating step on the sheath and the sealing ring is compressed until the adjacent faces of the clamp ring and sheath are 1 to 1.5mm apart. Use a 1mm drill or shim to set the gap.
Probe Head
1 to 1.5 mm
Carefully insert the probe into the ceramic sheath…
1
2
3
4
5
6
7
8
9
5
3
8
2
2 PREPARATION
✶
Note. To reduce its diameter the wadding
may be compressed slightly.
To increase its diameter, apply slight end
pressure on the wire bonding at both ends of the wadding. Mould the wadding by hand to ensure a light push fit into the sheath.
Fig. 2.2 Assembling the Probe
Page 8
6
Warning. The probe operates at high temperatures – take all necessary precautions to avoid injury through burns.
Caution. Thermal shock may damage the zirconia cell if the flue is cleaned using a high pressure water hose. If this
method of cleaning is used, remove the probe from the flue prior to cleaning. Never use the probe without the sheath.
3.1 Siting – Fig. 3.1
Select the position for the probe avoiding obstructions which may inhibit insertion or subsequent removal. Dimensions for the probe are shown in Fig. 3.2. A clearance of at least 1in (25mm) in excess of the overall probe length is necessary for installation or removal procedures.
Caution. The probe must not be sited:
• Where it would be subjected to mechanical or thermal shock.
• In the presence of aggressive compounds, e.g. molten slags, molten silicates, metals and vapors of lead, zinc, silicon and vanadium.
1000°C Max. Incoloy Sheath 1250°C Max. Porcelain Sheath
600°C
Min.
B – Within Process Temperature Limits
Maximum Distance 230ft (70m)
A – Maximum Distance Between EXGP Probe and Interface Electronics Unit
ATMOSPHERE IS PRESENT
DO NOT OPEN WHEN FLAMMABLE
3 INSTALLATION
Caution. Reduced probe life may result if probe is operated above 1250°C.
Fig. 3.1 Siting
Page 9
7
3.2 Mounting – Figs. 3.2 and 3.3
Reference Air Line Entry
Connector Head
Mounting Flange
1.08
(27.5)
21.89 (556) or
37.52 (953)
4.80 (122)
0.20 (5)
Probe Outer Sheath
Fixing holes in mounting flange to be drilled by customer
Dimensions in inches (mm)
4.92
(125)
4.53
(115)
3.46 (88)
3 INSTALLATION…
Fig. 3.2 Overall Dimensions
Page 10
8
3.2.1 Mounting Configurations – Fig. 3.3
The probe may be fitted horizontally or, preferably, vertically for prolonged high temperature operation.
Mount the probe through the furnace wall using the preferred method shown in Fig. 3.3A. Ensure that a sufficiently large hole is provided through the mounting – min. bore 1.65ins (42mm) diameter – to avoid damage to the probe during insertion or use.
Drill the fixing flange to suit the local fixing requirements before assembling the probe – see Section 2.3.
Caution. To prevent furnace gas from contaminating the reference air, always ensure that the clamp ring, fixing flange and sealing ring are fitted to the probe as described in Section 2.3.
With applications such as high temperature incinerators, fluid bed boilers and ore roasters, where the level of water vapor in the waste gases to be measured can be extremely high, it is important that the mounting flange of the probe, including any standoff which may be used, is thermally insulated to minimize condensation within the probe – see Fig. 3.3A.
Condensation within the probe sheath, particularly in outdoor installations, can be sufficient to allow water to come into contact with the hot ceramic (zirconia) tube leading to thermal shock and failure of the sensor. This problem is most likely to occur on such installations where the plant is shut down regularly over the weekend. Additionally, take care to protect the head of the probe from the elements in outdoor installations.
Raising the temperature of the probe as rapidly as possible assists in reducing the level of condensation. This is not possible on some processes where the plant temperature is gradually raised to the normal operating level. The temperature rise at the head of the probe can be assisted by having a large clearance hole in the refractory allowing hot gases to get up to the mounting flange/standoff – see Fig.
3.3B. Horizontal installations are more likely to exhibit the problems
described. Where it is not possible to mount the probe in the vertical position, install the probe with a slope of at least 10 to 15° downwards from the head of the probe – see Fig. 3.3C.
Clamp Bolt
Fixing Flange
Support Tube and Flange
5.90 (150) min.
Furnace Wall
Thermal Insulation
Packing
A – Standard Configuration
Support Tube and Flange
Thermal Insulation
Furnace Wall
5.90 (150) min.
B – Improved Probe Heating
Support Tube and Flange
Thermal Insulation
Furnace Wall
10° Min.
C – Inclined Configuration
Dimensions in inches (mm)
…3 INSTALLATION
Fig. 3.3 Probe Mounting
Page 11
9
Reference Air Input
2 Cable Entries, Threaded M16
+ Thermocouple – Thermocouple
+ Cell Input (Inner) – Cell Input (Outer)
4.2 Access to Electrical Connections – Fig. 4.1
For access to the terminal block – see Fig.4.1.
4.3 Electrical Connections – Fig. 4.2
Make connections as shown in Fig. 4.2 and on the label inside the probe head cover. The connecting cable must enter the probe head via the two M16 cable glands.
4.4 Reference Air Connection – Fig. 4.2
The reference air connector is 1/8in. BSP fitted with an adaptor to accept
1
/4in i.d. x 3/8in o.d. nylon or PVC tubing (100°C
ambient maximum). The probe requires clean, dry air supplied at 100 to 1000cm
3
/hr
(0.2 to 2ft
3
/hr). The air may be regulated instrument air or preferably atmospheric air from a small pump unit. Suitable air supply units are available from the Company:
a) Model 003000240 – mains powered pump unit with flow
gauge for use in safe areas only.
b) Model 003000241 – regulated air supply unit for use in
both safe and hazardous areas.
4.5 Routine Maintenance
No maintenance is necessary, or possible, on the probe and any attempt to dismantle it could cause irreparable damage. The probe can be tested for accuracy whilst connected to its measuring system but without removal from its operating position.
It is recommended that at least one spare probe is held for replacement or comparison purposes.
Remove cover retaining screws (x3)
Remove gasket
Identify terminal block
Test Gas Connector (with screwed plug)
1
2
3
4 CONNECTIONS
Table 4.1 Cable and Gland Specifications (electrical requirements only)
eborP–noitacificepSdnalG )slangiSefaSyllacisnirtnI(noitacificepSelbaC ecafretnI–noitacificepSdnalG
deifitrec-noN61M
llarevo,eroc-2,2.0/61:tupnI)Vm(elpuocomrehTepyTR
7394SBotelbacgnitasnepmocelpuocomrehtepyt-R,neercs
3ecnaticapaC µ 82oitaRR/L,Hm72.0ecnatcudnI,f µ /H Ω
'd'dnalgreirrabdeifitreCAFEESAB02M
)foorpemalf(
deifitrec-noN61M
deneercsllarevo,reppoceroc-2,2.0/61:tupnI)Vm(negyxO
3ecnaticapaC µ 82oitaRR/L,Hm72.0ecnatcudnI,f µ /H Ω
'd'dnalgreirrabdeifitreCAFEESAB02M
)foorpemalf(
ATMOSPHERE IS PRESENT
DO NOT OPEN WHEN FLAMMABLE
Fig. 4.1 Access to Electrical Connections
Fig. 4.2 Connections
4.1 Cable and Gland Specifications – Table 4.1
Caution. Cables and glands are not supplied with the probe and must conform to the following requirements:
• The Capacitance and Inductance or Inductance to Resistance (L/R) ratio of the cables connected to the output (hazardous area terminals) of the zener barriers mounted in the EXGP Interface Electronics Unit must not exceed the values shown.
• All cables must conform to BS5345 for flameproof 'd' type enclosures for mechanical construction.
• EEx d glands used on the Interface Electronics Unit must be of the EEx d 'Barrier Gland' type with BASEEFA certification because the enclosure is over 2 litres volume, has a source of ignition within and is designed for use in Zone 1 areas (reference BS5345 part 3).
Page 12
10
5.2 Comparison with Another O2 Probe
The probe may be checked by comparison with another of known reliability , either by replacing the suspect probe with the known probe or by mounting the known probe in close proximity and monitoring both probes continuously for a short period.
5.3 Returning the Probe to the Manufacturer for Checking
If it is not possible to carry out the above tests on site and failure or malfunction is suspected, the probe may be returned to the Company for checking.
If returning the probe, carefully dismantle the probe components by reversing the procedure detailed in Fig. 2.2. Repack in the original packing to ensure safe carriage.
5.4 Continuity Check
Connect a 100kΩ resistor across the probe output. If the output drops to near zero millivolts and then drifts when the resistance is removed, this indicates that the probe may be open circuit or have a high impedance.
Warning. The probe operates at high temperatures. Take all necessary precautions to avoid injury through burns.
During its working life, under normal recommended conditions, the probe output remains accurate and without drift. However, probe malfunctions can result from a fault in the probe or from incorrect operating conditions.
The testing of a suspect probe can only be carried out satisfactorily in its working position or in a furnace controlled within the normal probe operating temperature range.
If a probe failure is suspected, first perform a test gas check as described in the following sections.
5.1 In Situ Checking Using a Test Gas
This procedure introduces a test gas of known O
2
concentration around the outer electrode of the probe in the space between the filter washer, at the bottom of the sheath, and the wadding around the lower end of the probe. Four vent holes (see Fig. 2.2) prevent the gas from being trapped in the upper part of the sheath.
a) Gain access to the interior of the probe head by
unscrewing the lid fixing screws – see Fig. 4.1.
b) Remove the screwed plug from the test gas connector –
see Fig. 4.1.
c) Fit
1
/4in i.d. x 3/8in o.d. plastic or similar tubing and supply a
test gas of known oxygen between 0.25 and 25% O
2
concentration to the probe at a steady flow rate of 800 to 1000ml/min. Allow at least 5 minutes for the system to stabilize before making a measurement.
d) Check that the oxygen concentration measured by the
4680 or 4685 transmitter (mounted in a safe area), compares with the specification for the test gas used.
✶
Note. The O2 concentration measured by the probe can be checked locally by taking readings from the probe thermocouple (for temperature) and the cell O
2
mV input and entering the results in to the
Nernst Equation.
e) Disconnect the test gas and ensure that the screwed plug
is replaced in the test gas connector. Failure to do this may result in serious measurement errors due to the entry of air into the probe.
f) Proceed to Table 5.1.
5 FAULT FINDING
tseTfotluseR
kcehCsaG
foesuaCelbissoP
noitcnuflaMnoitarepOnoitcA
tcerroC—enoN
tcerrocnIylppusriaecnerefeR
Ohgih(eruliaf
2
)gnidaer
ecnerefereborpkcehC
ylppusria
tinuscinortceleytluaF
rotacidniro
fonoitarepokcehC
denrecnocstinu
ebutainocriznikaeL
nekorbroeborpnihtiw
ebutainocriz
Ohgih(
2
)gnidaer
htiwnoitarepoerapmoC
Orehtona
2
–ebut
yrassecenfiecalper
gnitfirdrooreZ
htiwsgnidaer
otesnopseron
sagtset
lleceborptiucricnepO
snoitcennoc
ytiunitnocatuoyrraC
4.5noitceSees–kcehc
tinuscinortceleytluaF
snoitcennocrorotacidni
fonoitarepokcehC
denrecnocstinu
Table 5.1 Fault Finding
Page 13
11
A1 Principles of Operation
The EXGP probe comprises a ceramic detector cell housed in a protective sheath. A thermocouple is fitted within the probe to enable the process temperature to be monitored or for automatic temperature compensation to be provided.
The sheath material may be either aluminous porcelain (recommended for oxidizing atmospheres) or Incoloy 800 (recommended for reducing atmospheres) for use at maximum temperatures of 1250°C and 1000°C respectively. Special sheaths are available for use up to 1400°C. Connections for the cell output, thermocouple and reference air are all made at the probe connector head.
The probe contains a high temperature oxygen concentration cell using zirconium oxide as a solid electrolyte. Inner and outer electrodes are attached to the cell which is specific to oxygen.
Air is supplied to the inner reference electrode to provide a constant partial pressure of oxygen while the process gas to be measured is in contact with the outer electrode.
A voltage is generated across the electrodes which is a function of the ratio of the oxygen partial pressures at the two electrodes. This voltage output represents 'oxygen potential' and can be used to control the atmosphere of annealing processes etc.
For the measurement of 'oxygen concentration' it is necessary to correct for absolute temperature of the probe. The temperature term in the Nernst equation, on which the voltage output depends, illustrates this point:
Nernst Equation
E = 0.0496 T (log10 P0/P1) + C mV
Where:
T = Absolute Temperature (°K) P
0
= Partial Pressure Reference Gas
P
1
= Partial Pressure Sample Gas
C = Cell Constant
A2 Range of Operation – Fig. A1
The internal resistance of the cell, i.e. the resistance of the electrolyte between the electrodes, decreases approximately exponentially with increasing temperature and for this reason it is recommended that the probe is used at temperatures greater than 600°C.
The maximum operating temperature is limited by two factors: a) The onset of electronic conduction through the electrolyte
which reduces the measured output below its theoretical
value. b) Evaporation of the outer electrode. Electronic conduction is a function of both temperature,
oxygen partial pressure and, for stabilized zirconia, it occurs at low levels of oxygen and high temperature. For the above reasons it is recommended that the probe is used at temperatures between 600°C and 1250°C – see Fig. A1.
Electronic Conduction
Errors
High
Resistance
Normal Operation
Non-carburising Atmospheres only
400 600 800 1000 1200
1250
1400
150
100
50
Kilocalories
Probe
Output
mV
1500
500
1000
Degrees Celsius
Pt/Pt 13% Rh Thermocouple
Probe
Life
Reduced
APPENDIX
Fig. A1 Range of Operation
Page 14
12
NOTES
Page 15
PRODUCTS & CUSTOMER SUPPORT
A Comprehensive Instrumentation Range
Analytical Instrumentation
• Transmitters
On-line pH, conductivity, and dissolved oxygen transmitters and associated sensing systems.
• Sensors
pH, redox, selective ion, conductivity and dissolved oxygen.
• Laboratory Instrumentation
pH and dissolved oxygen meters and associated sensors.
• Water Analyzers
For water quality monitoring in environmental, power generation and general industrial applications including: pH, conductivity, ammonia, nitrate, phosphate, silica, sodium, chloride, fluoride, dissolved oxygen and hydrazine.
• Gas Analyzers
Zirconia, katharometers, hydrogen purity and purge-gas monitors, thermal conductivity.
Controllers & Recorders
• Controllers
Digital display, electronic, pneumatic. Discrete single­loop and multi-loop controllers which can be linked to a common display station, process computer or personal computer.
• Recorders
Circular and strip-chart types (single and multi-point) for temperature, pressure, flow and many other process measurements.
Electronic Transmitters
• Smart & Analog Transmitters
For draft, differential, gauge and absolute pressure measurement. Also, liquid level and temperature.
• I to P Converters and Field Indicators
Flow Metering
• Magnetic Flowmeters
Electromagnetic, insertion type probes and watermeters.
• Turbine Flowmeters
• Wedge Flow Elements
• Mass Flow Meters
Transmitters, sensors, controllers and batch/display units.
Level Control
• Submersible, Capacitance & Conductivity.
Pneumatic Instrumentation
• Transmitters
• Indicating Controllers
• Recording Controllers
Customer Support
ABB Instrumentation provides a comprehensive after sales service via a Worldwide Service Organization. Contact one of the following offices for details on your nearest Service and Repair Centre.
United Kingdom
ABB Instrumentation Ltd Tel: +44 (0)1480 475321 Fax: +44 (0)1480 470787
United States of America
ABB Automation Inc. Instrumentation Division Tel: +1 215-674-6000 Fax: +1 215-674-7183
Italy
ABB Instrumentation SpA Tel: +39 (0) 344 58111 Fax: +39 (0) 344 58278
Client Warranty
Prior to installation, the equipment referred to in this manual must be stored in a clean, dry environment, in accordance with the Company's published specification. Periodic checks must be made on the equipment's condition.
In the event of a failure under warranty, the following documentation must be provided as substantiation:
1. A listing evidencing process operation and alarm logs at time of failure.
2. Copies of operating and maintenance records relating to the alleged faulty unit.
Page 16
IM/EXGP–PB Issue 3
The Company’s policy is one of continuous product improvement and the right is reserved to modify the information contained herein without notice.
© ABB 2000 Printed in UK (05.00)
ABB Instrumentation Ltd
St. Neots Cambs. England, PE19 8EU Tel: +44 (0) 1480 475321 Fax: +44 (0) 1480 217948
ABB Automation Inc. Instrumentation Division
125 E. County Line Road Warminster, PA 18974 USA Tel: +1 215-674-6000 Fax: +1 215-674-7183
ABB Instrumentation SpA
22016 Lenno Como Italy Tel: +39 (0) 344 58111 Fax: +39 (0) 344 58278
Loading...