A and D UA-1030 User Manual [es]

+
Premier Talking Blood Pressure moniTor
Models UA-1030T + UA-1030TCN
Instruction Manual
Manuel d'instructions
Manual de instrucciones
englishFrançaisesPañol
THIS MONITOR IS CLINICALLY VALIDATED
IMPORTANT INFORMATION
This blood pressure monitor was clinically validated by an independent organization. Blood pressure measurements determined with this device are equivalent to those obtained by a trained observer using the cuff/stethoscope auscultation method within the limits prescribed by the American National Standard Institute (ANSI/AAMI SP10) for electronic sphygmomanometers. The published study was performed by a reviewing committee consisting of physicians and/or nurses. Among blood pressure manufacturers, A&D Medical is proud to have the highest number of Clinically Validated monitors with a positive recommendation from professionals that are available to consumers.
Please read this important information before using your monitor.
Please remember that only a medical practitioner is qualified to interpret
☞
your blood pressure measurements. Use of this device should not replace regular medical examinations.
Consult your physician if you have any doubt about your readings.
☞
Should a mechanical problem occur, contact A&D Medical in the U.S.A. at 1-888-726-9966 or contact Auto Control in Canada at 1-800-461-0991.
Do not attempt to service, calibrate, or repair this monitor.
☞
Because your UA-1030T monitor contains delicate, high-precision parts,
☞
avoid exposing it to extremes in temperature or humidity or to direct sunlight, shock and dust. A&D Medical guarantees the accuracy of this monitor only when it is stored and used within the temperature and humidity ranges noted on page E-27.
Clean the monitor and cuff with a dry, soft cloth or a cloth dampened with
☞
water and a mild detergent. Never use alcohol, benzene, thinner or other harsh chemicals to clean monitor or cuff.
Remove and replace batteries if monitor is not used for more than six
☞
months. Alkaline batteries recommended.
PRECAUTIONS
The UA-1030T is designed to be used at home, by those who are eighteen (18) years and older, to monitor blood pressure (systolic and diastolic) and pulse rate. It is not designed for ambulatory use.
IMPORTANT SAFETY INSTRUCTIONS - SAVE THESE INSTRUCTIONS
DANGER - TO REDUCE THE RISK OF FIRE OR ELECTRIC SHOCK,
CAREFULLY FOLLOW THESE INSTRUCTIONS.
Use appropriate AC adapter and converter(s) when outside the U.S.A.
☞
E-1
A&D Medical (Toll-Free): 1-888-726-9966
In Canada - Auto Control Medical (Toll-Free): 1-800-461-0991
E-2
+
Fast. Easy. Accurate.
Congratulations on purchasing a state-of-the-art A&D Medical blood pressure monitor—one of the most technologically advanced monitors available today. Designed for ease of use and accuracy, this monitor will facilitate your daily blood pressure regimen.
Physicians agree that daily self-monitoring of blood pressure is an important step individuals can take to maintain cardiovascular health and prevent the serious consequences associated with undetected and untreated hypertension.
?
TABLE OF CONTENTS
VALIDATION STATEMENT. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .E-1
IMPORTANT INFORMATION ...........................................E-2
MANUFACTURER'S STATEMENT. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . E-3
TABLE OF CONTENTS ...............................................E-4
WHAT DISPLAY SYMBOLS MEAN ..................................E-5 – E-6
MONITOR COMPONENTS .............................................E-7
HOW UA-1030T WORKS .............................................E-8
BEFORE YOU START ................................................E-8
SELECTING THE CORRECT CUFF SIZE ....................................E-9
ADJUSTING THE CLOCK .............................................E-10
ADJUSTING THE PRESSURE .........................................E-10
TRICHECK™ MODE ...............................................E-11
TAKING YOUR BLOOD PRESSURE ................................ E-12 – E-13
MEASUREMENT USING TRICHECK .....................................E-14
MEASUREMENT WITH SET PRESSURE ..................................E-15
MEASUREMENT WITH DESIRED SYSTOLIC PRESSURE ......................E-15
WHAT IS AN IRREGULAR HEARTBEAT ..................................E-16
ABOUT THE PRESSURE RATING INDICATOR™ ............................E-17
ABOUT MEMORY .................................................E-18
ADJUSTING THE SPEAKER VOLUME ....................................E-19
SELECTING THE LANGUAGE ..........................................E-19
A&D Medical has been manufacturing high quality healthcare products for over 20 years. We are committed to providing you and your family with monitoring devices that provide the utmost in accuracy and convenience.
A&D Medical - Designed for Life.
E-3
A&D Medical (Toll-Free): 1-888-726-9966
USING THE MONITOR WITH AC ADAPTER ...............................E-20
ABOUT BLOOD PRESSURE ..................................... E-21 – E-25
TROUBLESHOOTING ...............................................E-26
SPECIFICATIONS .................................................E-27
CONTACT INFORMATION ............................................E-28
FRENCH INSTRUCTIONS .............................................F-1
SPANISH INSTRUCTIONS ............................................S-1
LOGBOOK PAGES ..................................................L-1
In Canada - Auto Control Medical (Toll-Free): 1-800-461-0991
E-4
WHAT DISPLAY SYMBOLS MEAN
WHAT DISPLAY SYMBOLS MEAN
Symbols that are printed on the device case
Symbols Function / Meaning Recommended Action
Standby and turn the device on.
R6(AA)
Battery installation guide Direct current
SN Serial number
2010
Date of manufacture Type BF: Device, cuff and tubing are
designed to provide special protection against electrical shock.
0366
EC directive medical device label WEEE label
Manufacturer
EC REP
EU-representative
Symbols that appear on the display
Symbols Function / Meaning Recommended Action
Appears while measurement is in progress. It blinks when the pulse is detected.
Irregular Heartbeat symbol (I.H.B.) Appears when an irregular heartbeat is detected during measurement.
Measurement is in progress. Remain as still as possible.
TriCheck mode Takes three consecutive measurements automatically and displays the average values of the three measurements.
The reading may yield an incorrect value. Take another measurement. Remain still during measurement.
A&D Medical (Toll-Free): 1-888-726-9966
E-5
Appears when a body or arm movement is detected.
Symbols Function / Meaning Recommended Action
The reading may yield an Appears during measurementwhen the cuff is applied loosely.
incorrect value. Apply the
cuff correctly, and take
another measurement. Previous measurements stored in memory
Average data
The battery power indicator during
Full Battery
Low Battery
measurement.
The battery is low when it blinks.
Unstable blood pressure due to movement during measurement.
Replace all batteries with new
ones when the mark blinks.
Take another measurement.
Remain still during
measurement. Thesystolic and diastolic values are
within 10 mmHg of each other. The pressure value did not increase
during the inflation.
The cuff is not applied correctly.
PUL. DISPLAY ERROR
The pulse is not detected correctly.
Apply the cuff correctly,
and take another
measurement.
Remove the batteries and press theSTART button,
Blood pressure monitor internal error
SYS.
DIA.
PUL./min.
Systolic blood pressure in mmHg Diastolic blood pressure in mmHg Pulse per minute Data taken between 4:00 and 9:59
AM
Data taken between 18:00 and 1:59
PM
Pressure settings Indicates the pressure value as set
by the user.
In Canada - Auto Control Medical (Toll-Free): 1-800-461-0991
and then install the batteries again. If the error still appears, contact the A&D.
E-6
MONITOR COMPONENTS
Battery Compartment
Display
Air Hose
AC Adapter Jack
START Button
Air Connector Plug
Air Socket
Display
1.5V Batteries
(AA, LR6 or R6P)
SET Button
Button Button
Battery Cover
Button
VOLUME
Arm Cuff
Speaker
HOW UA-1030T WORKS
Battery Compartment
WHO Classification Indicator
E-7
Air Connector Plug Arm Cuff
Air Hose
(Irregular heartbeat symbol)
Cuff Fit Error Symbol
Movement Error Symbol
TriCheck
and Cuff Inflation Meter
MEMORY
I.H.B. Symbol
TM
Symbol
Heart Mark
Battery Indicator
AM / PM Mark
AC Adapter Jack
Battery Cover
Display
Air Socket
Average Pressure Settings
A&D Medical (Toll-Free): 1-888-726-9966
START Button
1.5V Batteries
(AA, LR6 or R6P)
Clock Display
Speaker
Button Button
SET Button
VOLUME
Button
Systolic Pressure
Diastolic Pressure
Pulse Rate
A&D blood pressure monitors are easy to use, accurate and digitally display full measurement readouts. Our technology is based on the "oscillometric method" – a noninvasive blood pressure determination. The term "oscillation" refers to any measure of vibrations caused by the aterial pulse. Our monitor examines the pulsatile pressure generated by the aterial wall as it expands and contracts against the cuff with each heartbeat.
The cuff is inflated until the artery is fully blocked. The inflation speed is maximized and the pressure level is optimized by the device. The monitor takes measurements while the cuff is both inflating and deflating. This results in a faster measurement providing greater comfort to the user.
BEFORE YOU START
You must install 4 type A (1.5 volt) batteries (alkaline batteries recommended), or use the AC Adapter (see page E-20 for using the Monitor with AC Adapter) and attach the cuff to the monitor before using it. To install batteries (or replace them if the "Low Battery" symbol appears on display), proceed as follows:
1. Remove battery compartment cover by gently pushing down on arrow and sliding cover forward.
2. Put in bottom row of batteries first. Place the batteries in compartment with positive (+) and negative (–) terminals matching those indicated in the compartment. Be sure batteries make contact with compartment terminals.
3. Replace cover by sliding it into the compartment and gently pressing into place.
NOTE: Rechargeable batteries are not recommended
for use with this monitor.
NOTE: Please dispose of batteries following local
regulations.
In Canada - Auto Control Medical (Toll-Free): 1-800-461-0991
Step 1
Step 3
Step 2
E-8
SELECTING THE CORRECT CUFF SIZE
22 - 32cm
8.6" - 12.6"
1 - 2cm
(0.5'')
1 2
ADJUSTING THE CLOCK
Using the correct cuff size is important for an accurate reading. A cuff that is too large will produce a reading that is lower than the correct blood pressure; a cuff that is too small will produce a reading that is higher than the correct blood pressure. With your arm hanging at the side of your body, measure the circumference of your upper arm at the midpoint between the shoulder and elbow.
6.3
Using the right sized cuff is essential to an accurate measurement. Our unique cuff is designed to accommodate a wide range of arm sizes, from 9 – 14.6" (23 – 37 cm). The SlimFit™ cuff is contoured to provide you with a quick and comfortable fit. Just slide the cuff onto your upper arm and secure – it’s that simple! For more detail, please read the section Taking Your Blood Pressure (see pg. E-12).
If your arm circumference is greater than 14.6" (37 cm) or smaller than 9" (23 cm), you will need to purchase a different size cuff than the one that is included in this box. Please call A&D Medical at 1-888-726-9966 (in Canada please call 1-800-461-0991) for more information on our alternate cuff sizes.
1 2 3
1 - 2cm
(0.5'')
SlimFit™ Small Cuff
SlimFit™ Medium Cuff
SlimFit™ Large Cuff
SmoothFit™ Cuff
BRAND
The UA-1030T blood pressure monitor has a built-in clock that gives you the time and date and can tell you when a blood pressure reading stored in memory was taken. We recommend that you adjust the clock prior to use. To adjust the clock, follow these simple steps:
1. Press the clock button
located to the right of the display.
2. Use the arrow buttons to set the year.
3. After reaching the desired year, press the clock button again.
4. Repeat the process for the month, day and hour.
5. The next setting will be for pressure.
ADJUSTING THE PRESSURE
1. After setting the year, month, day and hour, the next mode is the pressure setting.
2. Default is , which stands for automatic.
3. For even more rapid readings, you can adjust the default to about 30 mmHg or more above your expected systolic pressure.
4. Use the arrow buttons to set the pressure.
5. After reaching the desired pressure, press the clock button again.
6. The next setting will be for TriCheck™ Mode.
E-9
A&D Medical (Toll-Free): 1-888-726-9966
In Canada - Auto Control Medical (Toll-Free): 1-800-461-0991
E-10
TRICHECK™ MODE TAKING YOUR BLOOD PRESSURE
The UA-1030T blood pressure monitor has an optional feature to automatically take three consecutive readings one minute apart and then provide you with the average. This complies with the American Heart Association's recommended measuring method.
1. After setting the year, month, day, hour and pressure, the next mode is the TriCheck™ Mode.
2. Default is off.
3. To turn on, use the arrow
4. To turn off again, use the arrow
5. After reaching the desired setting, press the clock
6. This will end your settings session.
7. To adjust any values, push the clock
buttons.
buttons.
button again.
button and begin again.
TIPS FOR BLOOD PRESSURE MONITORING:
Relax for several moments before measurement.
☞
Do not smoke or ingest caffeine at least 30 minutes prior to measurement.
☞
Remove constricting clothing and place cuff on bare arm.
☞
Unless your physician recommends otherwise, use left arm to measure
☞
pressure. Do not talk during measurement.
☞
Do not cross legs and keep your feet flat on floor during measurement.
☞
NOW YOU ARE READY. FOLLOW THESE SIMPLE STEPS:
1. Sit comfortably with your left arm resting on a flat surface so that the center of your upper arm is at the same height as your heart.
2. Lay left arm on the table, palm up and wrap the cuff around the upper arm, about 0.5" (1-2 cm) above the inside of the elbow, as shown, Place the cuff directly against the skin, as clothing my cause a faint pulse, and result in a measurement error.
3. Make sure cuff is aligned at heart level
Cuff
Artery position mark
At heart level
0.5” or 1-2 cm
Air Hose
E-11
A&D Medical (Toll-Free): 1-888-726-9966
In Canada - Auto Control Medical (Toll-Free): 1-800-461-0991
E-12
TAKING YOUR BLOOD PRESSURE MEASUREMENT USING TRICHECK
5. Press the segments are displayed. Zero (0) is displayed blinking briefly. The display changes, as indicated in the figure at the right, as the measurement
START
button. All of the display
All of the display
segments displayed
Press
begins. The cuff starts to inflate. It is normal for the cuff to feel very tight. A pressure bar indicator is displayed, on the left edge of the display, during the inflation.
Note: If you wish to stop inflation at any time, press the
button again.
6. When inflation is complete, deflation starts automatically and
blinks, indicating that the
START
Zero display
Starts inflation
Pressurizing
measurement is in progress. Once the pulse is detected, the mark blinks with each pulse beat.
Note: If an appropriate pressure is not obtained, the device
starts to inflate again automatically. To avoid re-inflation, see “Measurement with the SET Pressure” on the next page.
7. When the measurement is complete, the systolic and diastolic pressure readings and pulse
Measurement
in progress
rate are displayed. The device announces the measurement results. The cuff exhausts the remaining air and deflates completely.
8. Press the
START
button to turn the device off. After one minute of non-operation, the device will turn off automatically.
9. The reading is then stored into memory. See page
Systolic pressure
Diastolic pressure WHO classification
Pulse rate
Exhausts remaining
air automatically
E-18 for more details about the memory function.
10. Remove cuff and make a note of blood pressure and pulse rate on the chart in the back of this manual, indicating date and time of measurement. Additional logbook charts can be downloaded and printed for free at www.andmedical.com or in Canada at www.autocontrol.com. We advise that you record the date and time after each measurement becaue an accurate blood pressure history relies not on single or sporadic readings, but on a pattern over time.
START
1. Press the All of the display segments are displayed.
START
button.
At heart level
Zero (0) is displayed blinking briefly and the
START
first measurement starts.
2. When the measurement is complete, the systolic and diastolic pressure readings and pulse rate are displayed, then a one-minute countdown starts for the second measurement.
Press
All of the display segments displayed
First measurement
3. After one minute, the second measurement starts.
4. When the measurement is complete, the average readings of the first and second measurements are displayed, then a one-minute countdown starts for the third
One minute after
Results of the first measurement
Second measurement
measurement.
5. After one minute, the third measurement starts.
6. When the measurement is complete, the average readings of the three measurements are displayed and stored in memory.
Notes:
One minute after
Average of the first and second measurements
Third measurement
During the measurement, the TriCheck
symbol is displayed.
To cancel the measurement, press the START
button. In this case, no data is stored in
Average of the three measurements
memory.
When the cuff fit error symbol appears during the first measurement, cancel
the measurement, reapply the cuff correctly and start a new measurement.
After measurement, the average data of the three measurements is stored in
memory. No data will be stored, when canceling the measurements, or before completion of the third measurement.
E-13
A&D Medical (Toll-Free): 1-888-726-9966
In Canada - Auto Control Medical (Toll-Free): 1-800-461-0991
E-14
MEASUREMENT WITH SET PRESSURE WHAT IS AN IRREGULAR HEARTBEAT
Turn off
TriCheck setting mode
or
1932-1AIc UA787EXC•06 illust•03,12p symbols.FH7
The UA-1030T is designed to detect the pulse and to inflate the cuff to a systolic pressure level automatically. If re-inflation occurs repeatedly, use the following
methods. During the blood pressure measurement, re-inflation may occur. A fixed pressure
value can be set to avoid re-inflation.
1. At step 6 of the clock adjustment procedure (see page E-10), press the
SET
button to go to the pressure setting
mode. The current setting blinks.
2. Press the or button to select a pressure value about 30 mmHg or more above your expected systolic pressure from the following.
AUTO
3. Press the
Press the
: Automatic pressurization (default value)
180
: Pressure value of 180 mmHg (fixed)
210
: Pressure value of 210 mmHg (fixed)
240
: Pressure value of 240 mmHg (fixed)
SET
button to go to the TriCheck setting mode.
START
button to turn the device off. After three minutes of
Turn off
non-operation, the device will turn off automatically. The next measurement will be performed with the new pressure value.
MEASUREMENT WITH DESIRED SYSTOLIC PRESSURE
Use this method when re-inflation occurs repeatedly even if the pressure value is
240
set to
in the procedure above or when the results are not displayed even if
the pressure decreases to 20 mmHg or less.
1. Place the cuff on the arm (preferably the left arm).
2. Press and hold the
START
button until a number about 30 to 40 mmHg higher than your expected systolic pressure appears.
3. When the desired number is reached, release the
START
button to start measurement. Continue to measure your blood pressure as described on the previous page.
E-15
A&D Medical (Toll-Free): 1-888-726-9966
At heart level
Press and hold the button
Release the button
at the desired
systolic pressure
Refer to previous page
for measurement
or
TriCheck setting mode
START
The UA-1030T blood pressure monitor provides a blood pressure and pulse rate measurement even when an irregular heartbeat occurs. The Irregular Heartbeat symbol
will appear in the display window in the event an Irregular Heartbeat has occured during measurement. An irregular heartbeat is defined as a heartbeat that varies by 25% from the average of all heartbeat intervals during the blood pressure measurement. It is important that you relax, remain still and refrain from talking during measurements.
Note: We recommend contacting your physician if you see this symbol frequently.
ABOUT CUFF INFLATION METER
The Cuff Inflation Meter is located on the left side of the display screen to tell you when the blood pressure monitor is inflating and deflating the cuff. The Cuff Inflation Meter moves up during inflation and moves down during deflation.
Inflation in Progress Inflation Complete
In Canada - Auto Control Medical (Toll-Free): 1-800-461-0991
Deflation/Measurement
in Progress
E-16
ABOUT THE PRESSURE RATING INDICATOR™ ABOUT MEMORY
JNC7 Classification Indicator
JNC7 Classification Indicator
Hypertension Stage 2 Hypertension Stage 2 Hypertension Stage 1 Prehypertension Prehypertension Normal
JNC7 Classification Indicator
Stage 2 Hypertension Stage 1 Hypertension Prehypertension Normal
Example
Stage 2
Hypertension
The Pressure Rating Indicator™ is a feature which provides a snapshot of your blood pressure classifi­cation based on your measurements. This will let you quickly know what your blood pressure readings mean. Each segment of the bar indicator corresponds to the Seventh Report of the Joint National Committee (JNC7) on Prevention, Evaluation and Treatment of High Blood Pressure from the National Heart, Lung and Blood Institute - May 2003. For a more detailed look at this blood pressure classification, please refer to “Assessing High Blood Pressure” on page E-22.
: The indicator displays a segment, based on the current measurement,
corresponding to the JNC7 Classification.
NOTE: Due to other risk factors (e.g. diabetes, obesity, smoking, etc.) in addition to your blood
pressure measurement, the Pressure Rating Indicator is approximate. Please consult with your physician for interpretation and diagnosis of your blood pressure measurements.
NOTE: Residents outside of the United States (e.g.. Canada and Mexico) should refer to the WHO
Classification Table on page E-22 for assessment of their blood pressure measurement.
E-17
Example
Hypertension
A&D Medical (Toll-Free): 1-888-726-9966
Stage 2 Hypertension Stage 1 Hypertension Prehypertension Normal
Stage 2
Note: This device stores the last 90 measurements in memory. The device announces the memory
data values as they are displayed.
1. Press the or button. The average of all
measurements and the number of data are displayed. (If no data, “0” is displayed. Press the
, or
START
button to turn the
device off.)
2. Each time the
button (or the button to display the data in the reverse order) is pressed, the memory data is displayed as follows.
Average of all AM (morning) measurements
Press or
Average of all the data
Average of AM data
Average systolic Average diastolic Average pulse
Average systolic Average diastolic Average pulse
taken between 4:00 and 9:59. (In the example, 10 measurements. If no data, “--” is displayed.)
Average of all PM (evening) measurements
Average of PM data
Average systolic Average diastolic Average pulse
taken between 18:00 and 1:59. (In the example, 9 measurements. If no data, “--” is displayed.)
Most recent data (See No. 35). Three
Most recent data
Systolic Diastolic Pulse
seconds after the data number display, the measurement data is displayed.
Last data (No.1). Three seconds after the
data number display, the measurement data is displayed.
Last data (Oldest)
Systolic Diastolic Pulse
3. After the last data is displayed, press the button to return the average display of all measurements.
4. Press the
START
button to turn the device off. After one minute of
non-operation, the device will turn off automatically.
In Canada - Auto Control Medical (Toll-Free): 1-800-461-0991
E-18
ADJUSTING THE SPEAKER VOLUME USING THE MONITOR WITH AC ADAPTER
Volume off
The current setting appears
or to change the volume
1. Press the
VOLUME
button to enter the volume adjustment mode. Six stages of volume, which are represented by bars, and OFF are available.
2. Press the or press the
button to increase the volume,
button to decrease the
volume.
3. Press the
START
button to turn the device off. After three minutes of non-operation, the device will turn off automatically.
SELECTING THE LANGUAGE
1. In the volume adjustment mode, press the
VOLUME
button to enter the language selection
mode. The current setting appears.
2. Press the
or button to select a language
from the following. L1: English (default value)
L2: French L3: Spanish
3. Press the
START
button to turn the device off. After three minutes of non-operation, the device will turn off automatically.
High
Low
The current setting appears
or to change the volume
press
The current setting appears
or to change the language
Volume off
The UA-1030T has an AC adapter jack to allow you to supply power from an outlet in your home. We recommend you only use the exclusive AC adapter to avoid potential damage to the monitor. Please contact A&D Medical at 1-888-726-9966 or if you are in Canada contact AutoControl at 1-800-461-0991, if you would like to purchase an additional AC adapter.
CONNECTING THE AC ADAPTER TO THE MONITOR:
1. Gently insert the AC adapter plug into a 120V AC outlet.
2. Connect the AC adapter plug into the jack on the back of the monitor.
DISCONNECTING THE AC ADAPTER FROM THE MONITOR:
1. Turn the unit off by pressing the START button.
2. Disconnect the plug from the jack of the monitor quickly.
3. Gently remove the AC adapter from the outlet.
NOTE: If monitor has batteries and an AC adapter connected, the measurements stored
in memory may be lost if AC adapter is unplugged from the wall first. To ensure that measurements remain in memory, make sure the unit is off and that the jack plugged into the monitor is removed quickly before unplugging adapter from the wall.
Jack
E-19
A&D Medical (Toll-Free): 1-888-726-9966
Plug
In Canada - Auto Control Medical (Toll-Free): 1-800-461-0991
AC Adapter
TB-233
E-20
ABOUT BLOOD PRESSURE
ABOUT BLOOD PRESSURE
WHAT IS BLOOD PRESSURE?
Blood pressure is the force exerted by blood against the walls of the arteries. Systolic pressure occurs when the heart contracts; diastolic pressure occurs when the heart expands. Blood pressure is measured in millimeters of mercury (mmHg).
WHAT AFFECTS BLOOD PRESSURE?
Blood pressure is affected by many factors: age, weight, time of day, activity level, climate, altitude and season. Certain activities can significantly alter one’s blood pressure. Walking will likely raise an individual’s blood pressure. Sleeping will likely decrease blood pressure. Not relaxing for several moments before measurements can influence readings as well.
In addition to these factors, beverages containing caffeine or alcohol, certain medications, emotional stress and even tight-fitting clothes can make a difference in the readings.
WHAT CAUSES VARIATIONS IN BLOOD PRESSURE?
An individual’s blood pressure varies greatly from day to day and season to season. For hypersensitive individuals, these variations are even more pronounced. Normally, blood pressure rises during work or play and falls to its lowest levels during sleep.
ASSESSING HIGH BLOOD PRESSURE FOR ADULTS
The following standards for assessing high blood pressure (without regard to age or gender) have been established as a guideline. Please note that other risk factors (e.g. diabetes, obesity, smoking, etc.) need to be taken into consideration and may affect these figures. Consult with your physician for an accurate assessment.
JNC7 Classification Table – for adults within the U.S
BP Classification Systolic (mmHg) Diastolic (mmHg)
Normal less than 120 and less than 80 Prehypertension 120-139 or 80-89 Stage 1 Hypertension 140-159 or 90-99 Stage 2 Hypertension 160 or 100
SOURCE: The Seventh Report of the Joint National Committee on Prevention, Evaluation
and Treatment of High Blood Pressure for adults. National Heart, Lung and Blood Institute - May 2003.
greater than or equal to
WHO Classification Table – for adults outside of the U.S. (e.g. Canada, Mexico)
BP Classification Systolic (mmHg) Diastolic (mmHg)
greater than or equal to
Optimal less than 120 and less than 80 Normal less than 130 or less than
85 High-Normal 130-139 or 85-89 Grade 1 Hypertension 140-159 or 90-99 Grade 2 Hypertension 160-179 or 100-109 Grade 3 Hypertension 180 or 110
SOURCE: Standards to assess high blood pressure, without regard to age or gender, have
been established by the World Health Organization (WHO).
greater than or equal to
greater than or equal to
E-21
A&D Medical (Toll-Free): 1-888-726-9966
In Canada - Auto Control Medical (Toll-Free): 1-800-461-0991
E-22
ABOUT BLOOD PRESSURE
ABOUT BLOOD PRESSURE
WHAT IS HYPERTENSION?
Hypertension (high blood pressure) is the diagnosis given when readings consistently rise above normal. It is well known that hypertension can lead to stroke, heart attack or other illness if left untreated. Referred to as a “silent killer” because it does not always produce symptoms that alert you to the problem, hypertension is treatable when diagnosed early.
CAN HYPERTENSION BE CONTROLLED?
In many individuals, hypertension can be controlled by altering lifestyle and minimizing stress, and by appropriate medication prescribed and monitored by your doctor. The American Heart Association recommends the following lifestyle suggestions to prevent or control hypertension:
Don’t smoke. Have regular Reduce salt and fat intake. physical checkups. Maintain proper weight. Monitor your blood Exercise routinely. pressure at periodic
intervals.
WHY MEASURE BLOOD PRESSURE AT HOME?
It is now well known that, for many individuals, blood pressure readings taken in a doctor’s office or hospital setting might be elevated as a result of apprehension and anxiety. This response is commonly called “white coat hypertension.”
In any case, self-measurement at home supplements your doctor’s readings and provides a more accurate, complete blood pressure history. In addition, clinical studies have shown that the detection and treatment of hypertension is improved when patients both consult their physicians and monitor their own blood pressure at home.
ANSWERS TO WHY YOUR READINGS ARE DIFFERENT BETWEEN HOME AND AT THE DOCTOR’S OFFICE
Why are my readings different between home and at a doctor’s office?
Your blood pressure readings taken in a doctor’s office or hospital setting may be elevated as a result of apprehension and anxiety. This response is known as white coat hypertension.
When I bring my monitor to a doctor’s office, why do I get a different measurement from my monitor to that taken by a doctor or nurse?
The healthcare professional may be using a different sized cuff. The size of the bladder inside the cuff is critical for the accuracy of the measurement. This may give you a different reading. A cuff too large will produce a reading that is lower than the correct blood pressure; a cuff that is too small will produce a reading that is higher than the correct blood pressure. There may also be other factors that can cause the difference in measurements.
KEYS TO SUCCESSFUL MONITORING:
Blood pressure fluctuates throughout the day. We recommend that you are consistent in your daily measurement routine:
• Measure at the same time every day.
• Sit in the same chair/position.
• Do not cross legs and keep your feet flat on the floor.
• Relax for several minutes before measurement.
• Use the correct cuff size to get an accurate reading.
• Sit still during measurement – no talking, eating or sudden movements.
• Record your measurement in a logbook.
E-23
A&D Medical (Toll-Free): 1-888-726-9966
In Canada - Auto Control Medical (Toll-Free): 1-800-461-0991
E-24
ABOUT BLOOD PRESSURE TROUBLESHOOTING
ESTABLISHING BASELINE MEASUREMENTS
The most important method to get an accurate blood pressure measurement is consistency. To get the most benefit out of your monitor, it is important to establish a “baseline measurement.” This helps build a foundation of measurements that you can use to compare against future readings. To build this baseline measurement, devote two weeks for consistent blood pressure monitoring. This involves doing everything the same way when you measure (e.g. measuring during the same time of day, in the same location, sitting in the same chair, using the same cuff, etc.). Once you establish your baseline measurement, you can start evaluating if your measurement has been affected based on things like lifestyle changes or medication treatment.
HOW DO I RECORD MY BLOOD PRESSURE?
Blood pressure readings are typically recorded with the systolic pressure written first, followed by a slash mark and the diastolic pressure. For example, 120 mmHg systolic and 80 mmHg diastolic measurements are written as 120/80. Pulse is simply written with the letter “P” followed by the pulse rate—P 72, for example. Please see the back of the manual for the blood pressure tracking record.
CUSTOMER SUPPORT TOOLS ONLINE
Additional tools are available on www.andmedical.com to help you get the most out of your blood pressure monitoring. These include:
• Large Print Instruction Manuals
• Animated Operating Instructions
• Additional Logbook Sheets
Problem Probable Cause
Nothing appears in the display, after I press the START button.
The cuff does not inflate.
The unit does not measure. Readings are too high or too low.
The value is different from that measured at a clinic or doctor’s office.
Batteries are drained.
Battery terminals are not in the correct position.
Battery voltage is too low. Low battery symbol blinks. [If the batteries are drained completely, the mark does not appear.]
The cuff is not fastened properly.
You moved your arm or body during the measurement.
The cuff position is not correct.
You are using the wrong size cuff.
Taking too many readings on the same arm in a short period of time.
The healthcare professional may be using a different sized cuff.
Your measurements may be elevated by white coat hypertension
Corrective Action
Replace all batteries with new ones. Alkaline batteries recommended.
Reinstall the batteries with negative and positive termi­nals matching those indicated in the battery compartment.
Replace all batteries with new ones. Alkaline batteries recommended.
Fasten the cuff correctly.
Make sure you remain very still and quiet during the measurement.
Sit comfortably and still. Make sure the cuff is at the same level as your heart.
See Pg. E-9 "Selecting the Correct Cuff."
Relax several minutes before each measurement.
See Pg. E-24 "Answers to Why Your Readings Are Different Between Home and the Doctor’s Office."
See Pg. E-23 "Why Measure Blood Pressure at Home."
E-25
A&D Medical (Toll-Free): 1-888-726-9966
NOTE: If the actions described above do not solve the problem, call 1-888-726-9966.
Do not attempt to repair the device yourself.
In Canada - Auto Control Medical (Toll-Free): 1-800-461-0991
E-26
SPECIFICATIONS CONTACT INFORMATION
R6(AA)
2010
R6(AA)
2010
Model
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . .UA-1030T / UA-1030TCN (Canada)
Type
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .Oscillometric
Display
. . . . . . . . . . . . . . . . . . . . . . . . . . . . .Digital character height
Pressure (20.0 mm)/Pulse (11.5 mm) displayed simultaneously
Memory
. . . . . . . . . . . . . . . . . . . . . . . . . . . . .90 readings
Measurement range
. . . . . . . . . . . . . . . . . . . .Pressure: 0 - 299 mmHg
Systolic pressure: 60 - 279 mmHg Diastolic pressure: 40 - 200 mmHg Pulse: 40 - 180 beats per minute
Accuracy
. . . . . . . . . . . . . . . . . . . . . . . . . . . .Pressure: ± 3 mmHg or ± 2%, whichever is
greater, Pulse: ± 5%
Power source
. . . . . . . . . . . . . . . . . . . . . . . . .4 x 1.5 volt batteries (AA, LR6 or R6P)
or 120 volt AC adapter (TB-233)
Classification Clinical test EMC Recommended operating environment
. . . . . . . . . . . . . . . . . . . . . . . . .Type BF
. . . . . . . . . . . . . . . . . . . . . . . . . .According to ANSI / AAMI SP-10 1992
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .IEC 60601-1-2: 2007
. . . . . .50ºF to 104ºF (10ºC to 40ºC)
Less than 85% relative humidity
Recommended storage environment
Dimensions
. . . . . . . . . . . . . . . . . . . . . . . . . .Length: 4.1” (105 mm)
. . . . . . . .-4ºF to 140ºF (-20ºC to 60ºC)
Less than 95% relative humidity
Width: 5.5” (140 mm) Height: 2.4” (60 mm)
Weight
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . .10.6 oz. (285 g) without batteries
(U.S.A.)
0366
Blood pressure measurements determined by the UA-1030T are equivalent to those obtained by a trained observer using the cuff/stethoscope auscultation method within the limits prescribed by the American National Standards Institute for electronic or automated sphygmomanometers.
Our products are designed and manufactured using the latest scientific and technological methods, and offer accurate, easy to use, home monitoring and treatment options. Our full product line includes:
■■■ Digital■Blood Pressure Monitors
■■■ Manual■Blood Pressure Monitors
■■■ Blood■Pressure Cuffs
■■■ Stethoscopes
■■■ Personal■Health Scales
■■■ Digital■Thermometers
■■■ Activity■Monitors
■■■ Wireless■Products
This blood pressure monitor and cuff carries different warranties in different countries of purchase. Please see below for warranty length and contact information where you purchased this product.
In the United States of America:
5 year warranty A&D Medical A division of A&D Engineering, Inc. 1756 Automation Parkway San Jose, CA 95131 Toll-Free: 1-888-726-9966 www.andmedical.com
In Canada:
7 year warranty Auto Control Medical 6695 Millcreek Drive, Unit 6 Mississauga, Ontario L5N 5R8 Canada Toll-Free: 1-800-461-0991
In Latin America:
No warranty A&D Medical A division of A&D Engineering, Inc. 1756 Automation Parkway San Jose, CA 95131 Toll-Free: 1-888-726-9966 www.andmedical.com
All other countries:
Various warranties - see packaging A&D Company, Limited 1-243 Asahi, Kitamoto-shi, Saitama 364-8585 Japan [81] (48) 593-1119 www.aandd.jp
www.autocontrol.com
E-27
A&D Medical (Toll-Free): 1-888-726-9966
In Canada - Auto Control Medical (Toll-Free): 1-800-461-0991
E-28