Aalborg XFM Digital User Manual

Technical Data Sheet No. TD0701M Rev. C Date of Issue: September 2009
OPERATING MANUAL
XFM DIGITAL MASS
FLOW METERS
TABLE OF CONTENTS
1. UNPACKING THE XFM MASS FLOW METER.........................................
1.1 Inspect Package for External Damage.................................................
1.3 Returning Merchandise for Repair.......................................................
2. INSTALLATION........................................................................................
2.1 Primary Gas Connections.................................................................
2.2 Electrical Connections......................................................................
2.2.1 Power Supply Connections..............................................................
2.2.2 Output Signals Connections..............................................................
2.2.3 Communication Parameters and Connections...................................
3. PRINCIPLE OF OPERATION..................................................................
4. SPECIFICATIONS...................................................................................
4.1 CE Compliance.................................................................................
5. OPERATING INSTRUCTIONS..................................................................
5.1 Preparation and Warm Up..................................................................
5.2 Swamping Condition.......................................................................
5.3 XFM Parameters Settings.................................................................
5.3.1 Engineering Units Settings...............................................................
5.3.2 Gas Table Settings..............................................................................
5.3.3 Totalizer Settings.............................................................................
5.3.4 Flow Alarm Settings......................................................................
5.3.5 Relay Assignment Settings............................................................
5.3.6 K Factors Settings...........................................................................
5.3.7 Zero Calibration...............................................................................
5.3.8 Self Diagnostic Alarm.......................................................................
5.4 Analog output Signals configuration...................................................
6. MAINTENANCE.........................................................................................
6.1 Introduction......................................................................................
6.2 Flow Path Cleaning...........................................................................
6.2.1 Restrictor Flow Element (RFE)........................................................
6.2.2 XFM 17 models................................................................................
6.2.3 XFM 37/47 models..........................................................................
6.2.4 XFM 57/67/77 models.....................................................................
1 1 1 1
1 1 3 3 3 4
6
7 8
11 11 12 13 13 14 14 15 16 17 17 19 19
20 20 21 21 21 21 22
7. CALIBRATION PROCEDURES.................................................................
7.1 Flow Calibration...............................................................................
7.2 Gas Calibration of XFM Mass Flow Meters........................................
7.2.1 Connections and Initial Warm Up.....................................................
7.2.2 ZERO Check/Adjustment Adjustment.................................................
7.2.3 Gas Linearization Table Adjustment.................................................
7.3 Analog output Calibration of XFM Mass Flow Meters.......................
7.3.1 Initial Setup.......................................................................................
7.3.2 Gas flow 0-5 Vdc analog output calibration....................................
7.3.3 Gas flow 4-20 mA analog output calibration...................................
8. RS-485/RS-232 SOFTWARE INTERFACE COMMANDS...........................
8.1 General............................................................................................
8.2 Commands Structure.........................................................................
8.3 ASCII Commands Set.........................................................................
9. TROUBLESHOOTING................................................................................
9.1 Common Conditions.......................................................................
9.2 Troubleshooting Guide.....................................................................
9.3 Technical Assistance.......................................................................
10. CALIBRATION CONVERSIONS FROM REFERENCE GASES....................
APPENDIX I AALBORG XFM EEPROM Variables........................................
APPENDIX II INTERNAL USER SELECTABLE GAS FACTOR TABLE
(INTERNAL "K" FACTORS)........................................................
APPENDIX III GAS FACTOR TABLE ("K" FACTORS)....................................
APPENDIX IV COMPONENT DIAGRAM......................................................
APPENDIX V DIMENSIONAL DRAWINGS.................................................
APPENDIX VI WARRANTY...........................................................................
22 22 23 23 24 24 26 27 27 28
28 28 28 30
36 36 37 39
39
40
43
44
48
50
53
1
1. UNPACKING THE XFM MASS FLOW METER
1.1 Inspect Package for External Damage
Your XFM Mass Flow Meter was carefully packed in a sturdy cardboard carton, with anti-static cushioning materials to withstand shipping shock. Upon receipt, inspect the package for possible external damage. In case of external damage to the package contact the shipping company immediately.
1.2 Unpack the Mass Flow Meter
Open the carton carefully from the top and inspect for any sign of concealed ship­ping damage. In addition to contacting the shipping carrier please forward a copy of any damage report to your distributor or Aalborg7 directly.
When unpacking the instrument please make sure that you have all the items indicated on the Packing List. Please report any shortages promptly.
1.3 Returning Merchandise for Repair
Please contact the customer service representative of your distributor or Aalborg7 if you purchased your Mass Flow Meter directly, and request a Return
Authorization Number (RAN). Equipment returned without an RAN will not be accepted. Aalborg7 reserves the right to charge a fee to the customer for
equipment returned under warranty claims if the instruments are tested to be free from warrantied defects.
Shipping charges are borne by the customer. Meters returned "collect" will not be accepted!
It is mandatory that any equipment returned for servicing be purged and neutral­ized of any dangerous contents including but not limited to toxic, bacterially infec­tious, corrosive or radioactive substances. No work shall be performed on a returned product unless the customer submits a fully executed, signed SAFETY CERTIFICATE. Please request form from the Service Manager.
2. INSTALLATION
2.1 Primary Gas Connections
Please note that the XFM Mass Flow Meter will not operate with liquids. Only clean gases are allowed to be introduced into the instrument. If gases are con­taminated they must be filtered to prevent the introduction of impediments into the sensor.
2
CAUTION: XFM TRANSDUCERS SHOULD NOT BE USED FOR MONITORING OXYGEN GAS UNLESS SPECIFICALLY CLEANED AND PREPARED FOR SUCH APPLICATION.
For more information, contact your distributor or Aalborg7.
Attitude limit of the Mass Flow Meter is ±15Ffrom calibration position (standard calibration is in horizontal position). This means that the gas flow path of the Flow Meter must be within this limit in order to maintain the original calibration accura­cy. Should there be need for a different orientation of the meter, re-calibration may be necessary. It is also preferable to install the XFM transducer in a stable envi­ronment, free of frequent and sudden temperature changes, high moisture, and drafts.
Prior to connecting gas lines inspect all parts of the piping system including fer­rules and fittings for dust or other contaminant’s.
When connecting the gas system to be monitored, be sure to observe the direc­tion of gas flow as indicated by the arrow on the front of the meter.
Insert tubing into the compression fittings until the ends of the properly sized tub­ing home flush against the shoulders of the fittings. Compression fittings are to be tightened to one and one quarter turns according to the manufacturer's instruc­tions. Avoid over tightening which will seriously damage the Restrictor Flow Elements (RFE's)!
CAUTION: For XFM 17/37/47/57/67/77 models, the maximum pressure in the gas line should not exceed 500 PSIA (34.47 bars). Applying pressure above 500 PSIA (34.47 bars) will seriously damage the flow sensor.
XFM transducers are supplied with either standard 1/4 inch (XFM 17 and 37), 3/8 inch (XFM 47/57), ½ inch (GDFM 67), or optional 1/8 inch inlet and outlet com­pression fittings which should NOT be removed unless the meter is being cleaned or calibrated for a new flow range. XFM 77 transducers are supplied with ¾ inch FNPT fittings.
3
2.2 Electrical Connections
XFM is supplied with a 15 pin "D" connector. Pin diagram is presented in Figure b-1.
2.2.1 Power Supply Connections
The power supply requirements for XFM transducers are: 11 to 26 Vdc, (unipo­lar power supply)
DC Power (+) --------------- pin 7 of the 15 pin "D" connector DC Power (-) --------------- pin 5 of the 15 pin "D" connector
CAUTION: Do not apply power voltage above 26Vdc. Doing so will cause XFM damage or faulty operation.
2.2.2 Output Signals Connections
CAUTION: When connecting the load to the output terminals, do not exceed the rated values shown in the specifications. Failure to do so might cause damage to this device. Be sure to check if the wiring and the polarity of the power supply is correct before turning the power ON. Wiring error may cause damage or faulty operation.
XFM series Mass Flow Meters are equipped with either calibrated 0-5 or cali­brated 4-20 mA output signals (jumper selectable). This linear output signal rep­resents 0-100% of the flow meter’s full scale range.
WARNING: The 4-20 mA current loop output is self-powered (non-isolated). Do NOT connect an external voltage source to the output signals.
Flow 0-5 VDC or 4-20 mA output signal connection:
Plus (+) --------------------------- pin 2 of the 15 pin "D" connector Minus (-)--------------------------- pin 1 of the 15 pin "D" connector
To eliminate the possibility of noise interference, use a separate cable entry for the DC power and signal lines.
2.2.3 Communication Parameters and Connections
The digital interface operates via RS485 (optional RS-232 or Profibus DP avail­able) and provides access to applicable internal data including: flow, CPU tem­perature reading, auto zero, totalizer and alarm settings, gas table, conversion factors and engineering units selection, dynamic response compensation and lin­earization table adjustment.
Communication Settings for RS-485/RS-232 communication interface:
Baud rate: ...................... 9600 baud
Stop bit: ...................... 1
Data bits: ...................... 8
Parity: ...................... None
Flow Control: ...................... None
RS-485 communication interface connection:
The RS485 converter/adapter must be configured for: multidrop, 2 wire, half duplex mode. The transmitter circuit must be enabled by TD or RTS (depending on which is available on the converter/adapter). Settings for the receiver circuit should follow the selection made for the transmitter circuit in order to eliminate echo.
RS-485 T(-) or R(-) ...................... pin 8 of the 15 pin "D" connector (TX-)
RS-485 T(+) or R(+) ...................... pin 15 of the 15 pin "D" connector (RX+)
RS-485 GND (if available) ...................... pin 9 of the 15 pin "D" connector (GND)
RS-232 communication interface connection:
Crossover connection has to be established:
RS-232 RX (pin 2 on the DB9 connector) ..... pin 8 of the 15 pin "D" connector (TX)
RS-232 TX (pin 3 on the DB9 connector) ..... pin 15 of the 15 pin "D" connector (RX)
RS-232 GND (pin 5 on the DB9 connector) ..... pin 9 of the 15 pin "D" connector (GND)
4
5
Figure b-1 - XFM 15 PIN "D" CONNECTOR CONFIGURATION
IMPORTANT NOTES:
Generally, "D" Connector numbering patterns are standardized. There are, how­ever, some connectors with nonconforming patterns and the numbering sequence on your mating connector may or may not coincide with the numbering sequence shown in our pin configuration table above. It is imperative that you match the appropriate wires in accordance with the correct sequence regardless of the par­ticular numbers displayed on the mating connector.
Make sure power is OFF when connecting or disconnecting any cables in the system.
The (+) and (-) power inputs are each protected by a 300mA M (medium time-lag) resettable fuse. If a shorting condition or polarity reversal occurs, the fuse will cut power to the flow transducer circuit. Disconnect the power to the unit, remove the faulty condition, and reconnect the power. The fuse will reset once the faulty con­dition has been removed. DC Power cable length may not exceed 9.5 feet (3 meters). Use of the XFM flow transducer in a manner other than that specified in this manual or in writing from Aalborg, may impair the protection provided by the equipment.
PIN XFM FUNCTION
1 Common, Signal Ground For Pin 2
(4-20 mA return)
2 0-5 Vdc or 4-20mA Flow Signal Output 3 Relay No. 2 - Normally Open Contact 4 Relay No. 2 - Common Contact 5 Common, Power Supply
(- DC power for 11 to 26 Vdc)
6 Relay No. 1 - Common Contact 7 Plus Power Supply
(+ DC power for 11 to 26 Vdc)
8 RS485 (-) (Optional RS232 TX) 9 RS232 Signal GND (RS-485 GND Optional) 10 Do not connect (Test/Maintenance terminal) 11 Relay No. 2 - Normally Closed Contact 12 Relay No. 1 - Normally Open Contact 13 Relay No. 1 - Normally Closed Contact 14 Do not connect (Test/Maintenance terminal) 15 RS485 (+) (Optional RS232 RX) Shield Chassis Ground
3. PRINCIPLE OF OPERATION
The stream of gas entering the Mass Flow transducer is split by shunting a small portion of the flow through a capillary stainless steel sensor tube. The remainder of the gas flows through the primary flow conduit. The geometry of the primary con­duit and the sensor tube are designed to ensure laminar flow in each branch. According to principles of fluid dynamics the flow rates of a gas in the two laminar flow conduits are proportional to one another. Therefore, the flow rates measured in the sensor tube are directly proportional to the total flow through the transducer.
In order to sense the flow in the sensor tube, heat flux is introduced at two sec­tions of the sensor tube by means of precision wound heater-sensor coils. Heat is transferred through the thin wall of the sensor tube to the gas flowing inside. As gas flow takes place heat is carried by the gas stream from the upstream coil to the downstream coil windings. The resultant temperature dependent resistance differential is detected by the electronic control circuit. The measured temperature gradient at the sensor windings is linearly proportional to the instantaneous rate of flow taking place.
An output signal is generated that is a function of the amount of heat carried by the gases to indicate mass-molecular based flow rates.
Additionally, the XFM model Mass Flow Meter incorporates a Precision Analog Microcontroller (ARM7TDMI
®
MCU) and non-volatile memory that stores all hard­ware specific variables and up to 10 different calibration tables. The flow rate can be displayed in 23 different volumetric or mass flow engineering units. Flow meter parameters and functions can be programmed remotely via the RS-232/RS-485 interface or optional Profibus DP interface. XFM flow meters support various func­tions including: programmable flow totalizer, low, high or range flow alarm, auto­matic zero adjustment (activated via local button or communication interface), 2 pro­grammable SPDT relays output, 0-5 Vdc / 4-20 mA analog outputs (jumper selec­table), self diagnostic alarm, 36 internal and user defined K-factor. Optional local 2x16 LCD readout with adjustable back light provides flow rate and total volume reading in currently selected engineering units and diagnostic events indication.
6
7
4. SPECIFICATIONS
FLOW MEDIUM: Please note that XFM Mass Flow Meters are designed to work only with clean gases. Never try to measure flow rates of liquids with any XFM.
CALIBRATIONS: Performed at standard conditions [14.7 psia (101.4 kPa) and 70
F
F
(21.1
F
C)] unless otherwise requested or stated.
ENVIRONMENTAL (PER IEC 664): Installation Level II; Pollution Degree II.
FLOW ACCURACY (INCLUDING LINEARITY): ±1% of FS at calibration temperature and
pressure.
REPEATABILITY: ±0.15% of full scale.
FLOW TEMPERATURE COEFFICIENT: 0.15% of full scale/
F
C or better.
FLOW PRESSURE COEFFICIENT: 0.01% of full scale/psi (6.895 kPa) or better.
FLOW RESPONSE TIME: 1000ms time constant; approximately 2 seconds to
within ±2% of set flow rate for 25% to 100% of full scale flow.
MAXIMUM GAS PRESSURE: 500 psig (3447 kPa gauge).
MAXIMUM PRESSURE DROP: 0.18 PSID (at 10 L/min flow). 8 psi (at 100 L/min flow).
See Table IV for pressure drops associated with various models and flow rates.
GAS AND AMBIENT TEMPERATURE: 32
F
F to 122 FF (0 FC to 50 FC). 14 FF to 122 FF
(-10
F
C to 50 FC) - Dry gases only.
RELATIVE GAS HUMIDITY: Up to 70%.
LEAK INTEGRITY: 1 x 10
-9
sccs He maximum to the outside environment.
ATTITUDE SENSITIVITY: Incremental deviation of up to 1% from stated accuracy, after re­zeroing.
OUTPUT SIGNALS: Linear 0-5 Vdc (3000 ohms min load impedance);
Linear 4-20 mA (500 ohms maximum loop resistance). Maximum noise 20mV peak to peak (for 0-5 Vdc output).
TRANSDUCER INPUT POWER: 11 to 26 Vdc, 100 mV maximum peak to peak output noise.
Power consumption: +12Vdc (200 mA maximum);
+24Vdc (100 mA maximum); Circuit board have built-in polarity reversal protection, 300mA resettable fuse provide power input protection.
WETTED MATERIALS: Anodized aluminum, brass, 316 stainless steel, 416 stainless steel, VITON7 O-rings; BUNA-N7, NEOPRENE7 or KALREZ7 O-rings are optional.
CAUTION: Aalborg makes no expressed or implied guarantees of corrosion resistance of mass flow meters as pertains to different flow media reacting with components of meters. It is the customers' sole responsibility to select the model suitable for a particular gas based on the fluid contacting (wetted) materials offered in the different models.
INLET AND OUTLET CONNECTIONS:
Model XFM 17/37 standard 1/4” compression fittings, Model XFM 47/57 standard 3/8” compression fittings, Model XFM 67 standard 1/2” compression fittings, Model XFM 77 standard 3/4” FNPT fittings,
Optional 1/8” or 3/8” compression fittings and 1/4” VCR fittings are available.
DISPLAY: Optional local 2x16 characters LCD with adjustable backlight (2 lines of text).
CALIBRATION OPTIONS: Standard is one 10 points NIST calibration. Optional, up to 9
additional calibrations may be ordered at additional charge.
4.1 CE Compliance
EMC Compliance with 89/336/EEC as amended. Emission Standard: EN 55011:1991, Group 1, Class A, Immunity Standard: EN 55082-1:1992.
8
9
*Flow rates are stated for Nitrogen at STP conditions [i.e. 70 FF (21.1 FC) at 1 atm].
For other gases use the K factor as a multiplier from APPENDIX 2.
CODE
scc/min [N2]
CODE
std liters/min [N2]
00 0 to 5 07 0 to 1 01 0 to 10 08 0 to 2
02 0 to 20 09 0 to 5 03 0 to 50 10 0 to 10 04 0 to 100 05 0 to 200
06 0 to 500
TABLE I XFM 17 LOW FLOW MASS FLOW METER*
TABLE II XFM 37 MEDIUM FLOW MASS FLOW METER*
FLOW RANGES
CODE
std liters/min [N2]
30 20 31 30 32 40 33 50
TABLE III XFM XFM 47/57/67/77 HIGH FLOW MASS FLOW METERS*
CODE
std liters/min [N2]
40 60 41 80 42 100 50 200 60 500 70 1000
10
TABLE IV PRESSURE DROPS
TABLE V APPROXIMATE WEIGHTS
MODEL
FLOW RATE
[std liters/min]
MAXIMUM PRESSURE DROP
[mm H2O] [psid] [kPa]
XFM 17 up to 10 130 0.18 1.275
XFM 37
20
TBD TBD TBD
30
TBD TBD TBD
40
TBD TBD TBD
50
TBD TBD TBD
XFM 47
60
TBD TBD TBD
100
TBD TBD TBD
XFM 57
200
TBD TBD TBD
XFM 67
500
TBD TBD TBD
XFM 77
1000
TBD TBD TBD
MODEL
WEIGHT
SHIPPING WEIGHT
XFM 17 transmitter 2.20 lbs. (1.00 kg) 3.70 lbs. (1.68 kg)
XFM 37/47 transmitter 2.95 lbs. (1.33 kg) 4.34 lbs. (1.97 kg)
XFM 57 transmitter TBD TBD XFM 67 transmitter TBD TBD XFM 77 transmitter TBD TBD
11
5. OPERATING INSTRUCTIONS
5.1 Preparation and Warm Up
It is assumed that the Mass Flow Meter has been correctly installed and thor­oughly leak tested as described in section 2. Make sure the flow source is OFF. When applying power to a flow meter within the first two seconds, you will see on the LCD display: the product name, the software version, and revision of the EEP­ROM table (applicable for LCD option only).
Figure b-2: XFM first Banner Screen
Within the next two seconds, the RS-485 network address, the analog output set­tings, and currently selected gas calibration table will be displayed (applicable for LCD option only).
Figure b-3: XFM second Banner Screen
Note: Actual content of the LCD screen may vary depending on the
model and device configuration.
After two seconds, the LSD display switches to the main screen with the following information:
- Mass Flow reading in current engineering units (upper line).
- Totalizer Volume reading in current volume or mass based engineering units (lower line).
Figure b-4: XFM Main Screen
AALBORG XFM 485
S: Ver1.4 Rev.A0
Ad: 11 Out: 0-5Vdc
Gas# 1 AIR
F: 50.0 L/min
T: 75660.5 Ltr
12
During initial powering of the XFM transducer, the flow output signal will be indi­cating a higher than usual output. This is an indication that the XFM transducer has not yet attained its minimum operating temperature. This condition will auto­matically cancel within a few minutes and the transducer should eventually indi­cate zero.
For the XFM transducer with LCD option: If the LCD diagnostic is activated, the second line of the LCD will display the time remaining until the end of the warm up period (Minutes:Seconds format) and will alternatively switch to Totalizer read­ing indication every 2 seconds.
Figure b-5: XFM Main Screen during Sensor Warm up period.
5.2 Swamping Condition
If a flow of more than 10% above the maximum flow rate of the Mass Flow Meter is taking place, a condition known as "swamping" may occur. Readings of a "swamped" meter cannot be assumed to be either accurate or linear. Flow must be restored to below 110% of maximum meter range. Once flow rates are lowered to within calibrated range, the swamping condition will end. Operation of the meter above 110% of maximum calibrated flow may increase recovery time.
Note: Allow the Digital Mass Flow Meter to warm-up for a MINIMUM of 6 minutes.
Note: During the first 6 minutes of the initial powering of the XFM transducer, the status LED will emit CONSTANT UMBER light.
Note: After 6 minutes of the initial powering of the XFM transducer, the status LED will emit a constant GREEN light (normal operation, ready to measure). For XFM with LCD option, the screen will reflect flow and totalizer reading. (see Figure b-4).
F: 50.0 L/min
** WarmUp 2:39 **
13
5.3 XFM Parameters Settings
5.3.1 Engineering Units Settings
The XFM Mass Flow Meter is capable of displaying flow rate with 23 different Engineering Units. Digital interface commands (see paragraph 8.3 ASCII Command Set “XFM SOFTWARE INTERFACE COMMANDS”) are provided to:
- get currently active Engineering Units
- set desired Engineering Units.
The following Engineering Units are available:
TABLE VI UNITS OF MEASUREMENT
NUMBER INDEX
FLOW RATE
ENGINEERING
UNITS
TOTALIZER
ENGINEERING
UNITS
DESCRIPTION
1 0 % %s Percent of full scale
2 1 mL/sec mL Milliliter per second
3 2 mL/min mL Milliliter per minute
4 3 mL/hr mL Milliliter per hour
5 4 L/sec Ltr Liter per second
6 5 L/ min Ltr Liter per minute
7 6 L/hr Ltr Liter per hour
87
m
3
/sec m
3
Cubic meter per second
98
m
3
/ min m
3
Cubic meter per minute
10 9
m
3
/hr m
3
Cubic meter per hour
11 10
f
3
/sec f
3
Cubic feet per second
12 11
f
3
/min f
3
Cubic feet per minute
13 12
f
3
/hr f
3
Cubic feet per hour
14 13 g/sec g Grams per second
15 14 g/min g Grams per minute
16 15 g/hr g Grams per hour
17 16 kg/sec kg Kilograms per second
18 17 kg/min kg Kilograms per minute
19 18 kg/hr kg Kilograms per hour
20 19 Lb/sec Lb Pounds per second
21 20 Lb/min Lb Pounds per minute
22 21 Lb/hr Lb Pounds per hour
23 22 User UD User defined
5.3.2 Gas Table Settings.
The XFM Mass Flow Meter is capable of storing calibration data for up to 10 dif­ferent gases. Digital interface commands are provided to:
- get currently active Gas Table number and Gas name
- set desired Gas Table.
5.3.3 Totalizer Settings.
The total volume of the gas is calculated by integrating the actual gas flow rate with respect to the time. Digital interface commands are provided to:
- reset the totalizer to ZERO
- start the totalizer at a preset flow
- assign action at a preset total volume
- start/stop (enable/disable) totalizing the flow
- read totalizer via digital interface
The Totalizer has several attributes which may be configured by the user. These attributes control the conditions which cause the Totalizer to start integrat­ing the gas flow and also to specify actions to be taken when the Total Volume is outside the specified limit.
Totalizer action conditions become true when the totalizer reading and preset "Stop at Total" volumes are equal.
14
Note: Once Flow Unit of Measure is changed, the Totalizer’s Volume/Mass based Unit of Measure will be changed automatically.
Note: By default the XFM is shipped with at least one valid calibration table (unless optional additional calibrations were ordered). If instead of the valid Gas name (for example NITROGEN), the LCD screen or digital interface displays Gas designator as “Uncalibrated”, then the user has chosen the Gas Table which was not calibrated. Using an “Uncalibrated” Gas Table will result in erroneous reading.
Note: Before enabling the Totalizer, ensure that all totalizer settings are configured properly. Totalizer Start values have to be entered in %F.S. engineering unit. The Totalizer will not totalize until the flow rate becomes equal to or more than the Totalizer Start value. Totalizer Stop values must be entered in currently active volume / mass based engineering units. If the Totalizer Stop at preset total volume feature is not required, then set Totalizer Stop value to zero.
Loading...
+ 39 hidden pages