Aalborg TIO Totalizer I,O Flow Monitor,Controller User Manual

OPERATING MANUAL
TIO
Totalizer-Input/Output
Flow Monitor/Controller
Technical Data Sheet No.: TD-TIO-0911 Rev C Date of Issue: January 2013
Aalborg7is a registered trademark of Aalborg7 Instruments & Controls.
discretion at any time without notice. For certified dimensions please contact Aalborg7.
TABLE OF CONTENTS
1. UNPACKING THE TIO TOTALIZER......................................................
1.1 Inspecting Package for External Damage..................................................................
1.2 Unpacking the TIO Totalizer.....................................................................................
1.3 Returning Merchandise for Repair.............................................................................
2. SAFETY INSTRUCTIONS / INTRODUCTION...............................................
3. SPECIFICATIONS.............................................................................
4. ELECTRICAL CONNECTIONS................................................................
4.1 Power Supply Connections........................................................................................
4.2 Process Variable (PV) Input Signal Connections..................................................
4.3 Set Point (SP) Output Signal Connections............................................................
4.4 Digital Communication Interface Connections......................................................
4.5 Digital and Pulse Optically-Isolated Output Connections....................................
5. LCD KEY-PAD OPERATION: DATA ENTRY AND CONFIGURATION.....................
5.1 Display Indications....................................................................................................
5.1.1 Set Point Control (only for devices set as controller)..........................................
5.2 Menu Structure....................................................................................................
5.3 Parameter Entry.........................................................................................................
5.3.1 Submenu “Change PP Password”.............................................................................
5.3.2 Submenu “Device Information”...........................................................................
5.3.3 Submenu “Measuring Units”................................................................................
5.3.4 Submenu “User-Defined Units”..................................................................................
5.3.5 Submenu “K-Factors Settings”.............................................................................
5.3.6 Submenu “Alarm Settings”........................................................................................
5.3.7 Submenu “Totalizer #1”............................................................................................
5.3.8 Submenu “Totalizer #2”.......................................................................................
5.3.9 Submenu “Pulse Output”...........................................................................................
5.3.10 Submenu “Opt. Outputs Settings”............................................................................
5.3.11 Submenu “Display Settings”.................................................................................
5.3.12 Submenu “Device Function”......................................................................................
5.3.13 Submenu “Communication Settings”....................................................................
5.3.14 Submenu “Device Calibration”..............................................................................
5.3.15 Submenu “Signal Conditioner”............................................................................
5.3.16 Submenu “Program Set Point”..................................................................................
5.3.17 Submenu “Event Register Menu”.........................................................................
5.3.18 Submenu “Diagnostic Menu” “Event Register Menu”...........................................
6. INSTALLATION............................................................................
6.1 General Directions.....................................................................................................
6.2 Hardware Installation............................................................................................
6.2.1 Connecting TIO to GFM series flow meter............................................................
6.2.2 Connecting TIO to GFC series flow controller...........................................................
6.2.3 Connecting TIO to flow meters / controllers from other manufacturers
(stand alone)............................................................................................................
1
1 1 1
2
3
5
6 6 7 8 11
13 13 15 17 18 19 21 20 22 25 25 27 29 32 33 34 35 36 37 40 42 43 47
49
49 49 50 57
63
7. TROUBLESHOOTING....................................................................
7.1 Common Conditions...........................................................................................
7.2 Troubleshooting Guide........................................................................................
APPENDIX A TIO Totalizer Input/Output Flow Monitor/Controller EEPROM Variables APPENDIX B Internal K-Factors Table APPENDIX C Totalizer-IO ASCII Commands Set APPENDIX D Mechanical Drawings APPENDIX E Circuit Layout Diagrams APPENDIX F Warranty
68
68 69
71 76 77 92 93 95
1. UNPACKING THE TIO TOTALIZER
1.1 Inspect Package for External Damage
Your TIO Totalizer-Input/Output Flow Monitor/Controller was carefully packed in a sturdy cardboard carton with anti-static cushioning materials to withstand ship­ping shock. Upon receipt, inspect the package for possible external damage. In case of external damage to the package, contact the shipping company immedi­ately.
1.2 Unpack the TIO Totalizer
Open the carton carefully from the top and inspect for any sign of concealed ship­ping damage. In addition to contacting the shipping carrier, please forward a copy of any damage report to your distributor or to Aalborg7 directly. When unpacking the instrument make sure that you have all the items indicated on the Packing List. Please promptly report any shortages.
1.3 Returning Merchandise for Repair
Please contact your distributor’s customer service representative, or Aalborg7 if you purchased your TIO instrument directly. Request a Return Authorization
Number (RAN). Equipment returned without an RAN will not be accepted.
Aalborg7 reserves the right to charge the customer a fee for equipment returned under warranty claims if the instruments are tested and found to be free from war­rantied defects. Shipping charges are borne by the customer. Instruments returned “collect” will not be accepted! It is mandatory that any equipment
returned for servicing be purged and neutralized of any dangerous contents including but not limited to toxic, bacterially infectious, corrosive or radioac­tive substances. No work shall be performed on a returned product unless the customer submits a fully executed, signed SAFETY CERTIFICATE. Please request form from the Service Manager.
1
The TIO Totalizer-Input/Output Flow Monitor/Controller is a microcontroller-driven device designed to linearize the flow meter/controller flow curve and to display instantaneous Flow Rate, Total and Accumulated Total. This product is designed to be used primarily with Aalborg7 series GFM/GFC analog flow meters/controllers but can be also used with any commercial flow meters/controllers with analog 0­5 (0-10) Vdc or 4-20 mA interface.
The following functions and features are supported:
• Built-in Flow Linearizer (10 point linearization of the flow curve).
• Up to 47 different volumetric and mass flow engineering units (including user-
defined).
• Graphic LCD with large 13mm (0.51") digits for Flow Rate and 5.5mm (0.21")
for Total.
• User-adjustable LCD back light and contrast level.
• Digital RS-232 or RS-485 interface (multidrop capability for up to 64 devices).
• Compact design for unit mount, panel mount, wall mount or field mount
applications.
• Two independent programmable Totalizers.
• User-programmable, optically-isolated pulse output.
• Two programmable optically-isolated digital outputs for different events.
• Low and High Flow Alarms with programmable Action Delay.
• Flow controllers Set Point command control via local LCD or digital interface.
• Programmable Set Point table with ramping up/down capability for up to 16
steps.
• Free Configuration and Monitoring Utility Software.
2
AALBORG7 warranties and all other responsibilities by direct or implied are voided if users fail to follow all instructions and procedures described in this manual.
LIFE SUPPORT APPLICATIONS: The TIO is not designed for use in life support applications where malfunctioning of the device may cause personal injury. Customers using or selling this device for use in such applications do so at their own risk and agree to be fully responsible for any damages resulting from improper use or sale.
Some of the IC devices used in the TIO are static-sensitive and may be damaged by improper handling. When adjusting or servicing the device, use of a grounded wrist strap is recommended to prevent inadvertent damage to the integral solid-state circuitry.
2. SAFETY INSTRUCTIONS / INTRODUCTION
3
3. SPECIFICATIONS
ADC/DAC RESOLUTION: 12 bit.
ACCURACY: ±0.1% F.S.
ANALOG INPUTS: 0-5 Vdc, 4-20 mA, 5-10 Vdc (jumper-selectable),
0-10 Vdc (special order).
ANALOG OUTPUTS: 0-5 Vdc, 4-20 mA (jumper-selectable), 0-10Vdc
(special order).
LCD: 128x64 graphic LCD with instantaneous Flow reading
and Total volume indication. Adjustable LCD contrast and back light. Refresh rate 10 times/sec.
KEY-PAD: Local 6 tactical push buttons.
DATA PROTECTION: EEPROM backup of all settings, backup Totalizer #1
every one second. Data retention at least 10 years. Password protection for configuration data.
PULSE OUTPUT: User-programmable, optically-isolated, with preset
active low time interval (10 – 6550 ms), UCE 40Vdc, ICE 150 mA (Voltage Isolation: 250 Vrms).
DIGITAL OUTPUT: Two programmable, optically-isolated. UCE ≤ 40Vdc,
ICE 150 mA (Voltage Isolation: 250 Vrms).
DIGITAL INTERFACE: RS-232 or RS-485 (multidrop capability up to 64
devices).
POWER REQUIREMENTS: 12 – 26 Vdc, 100 mV maximum peak-to-peak output
noise (up to 60 mA maximum load).
Protocol: Proprietary ASCII software interface command set. Speed: 1200 - 2400 - 4800 - 9600 -19200 – 38400 – 57600 – 115200
baud (user-selectable). Default 9600 baud.
Configuration: Stop bit: 1
Data bits: 8 Parity: None
Flow Control: None Addressing: Maximum 255 addresses, default
address 11 hex. (for RS-485 option only).
Type: RS232 or RS485 2-wire.
4
INTERFACE CONNECTORS: Process I/O signals and digital RS-232/RS-485
interface: miniature 9 pin female D-SUB connector. Digital optically-isolated outputs: TERMINAL BLOCK HEADER 4POS 3.5MM male pins, Shrouded (Mated connector: Tyco Electronics P/N: 284510-4).
ENVIRONMENT: Installation Level II; Pollution Degree II.
ELECTROMAGNETIC COMPATIBILITY:
Compliant ref. 89/336/EEC as amended. Emission. Standard: EN 55011:1991, Group 1, Class A. Immunity Standard: EN 55082-1:1992.
OPERATING TEMPERATURE:-10 °C to +70 °C (14 °F to +158 °F).
DIMENSIONS: 86.4 x 76.2 x 19.1 mm (3.4" x 3.0" x 0.75") - W x H x D.
WEIGHT: Approximately 125g / 0.3 lbs.
5
4. ELECTRICAL CONNECTION
PIN FUNCTION NOTE
1
Power supply, common Power input
2
Power supply, positive Power input 12 – 26 Vdc
3
RS232 RX, Optional RS485 (+)
Communication (RS-232 – input, RS-485 – input/output)
4
Analog Input (+), PV input Input
5
Analog Output (+), PV set point Output
6
RS232 Signal GND (RS-485 GND Optional)
Communication reference
7
RS232 TX, Optional RS485 (-)
Communication (RS-232 – output, RS-485 – input/output)
8
Analog Input/Output reference (common for pins 4 and 5)
9
+5Vdc reference input (for 5-10 Vdc interface only)
Figure 4.1 - TIO 9 PIN "D" CONNECTOR CONFIGURATION
9
5
4
8
7
6
3
2
1
6
4.2 Process Variable (PV) Input Signal Connections
Depending on the jumper J2 configuration Input signal can be set to 0-5, 5-10, 0-10 Vdc or 4-20 mA.
CAUTION: Do not apply power voltage above 28Vdc.
Doing so will cause device damage or faulty operation.
CAUTION: When connecting the external signals to the input terminals always check actual jumper J2 configuration. Do not exceed the rated
values shown in the specifications (see Table 4.1). Failure to do so
might cause damage to this device. Be sure to check if the wiring and the polarity of the power supply and PV signals are correct before turning the power ON. Wiring error may cause damage or faulty operation.
Make sure power is OFF when connecting or disconnecting any cables or wires in the system.
The power supply (PS), process variable (PV) input, set point (SP) control out­put, and digital communication interface signals are connected to the TIO via miniature 9 pin female D-SUB connector.
4.1 Power Supply Connections
The power supply requirements for TIO are: 12 to 26 Vdc, (unipolar power sup­ply).
DC Power (+) --------------- pin 2 of the 9 pin "D" connector DC Power (-) --------------- pin 1 of the 9 pin "D" connector
7
4.3 Set Point (SP) Output Signal Connections
Set Point (SP) output signal connection is required only if TIO is mated to the flow controller and will be used as a source for a Set Point control signal. Depending on the jumper J2 configuration, SP output signal can be set to 0-5, 0-10 Vdc or 4-20 mA.
Table 4.1 Maximum rated values for PV input signals
PV
INPUT
TYPE
J2 JUMPER
CONFIGURATION
MAXIMUM
SIGNAL
LEVEL
NOTE
J2D J2E J2F
0-5 Vdc 10 -11 14 -15 17-18
6 Vdc
5-10 Vdc 11 -12 14 -15 17-18
11 Vdc
+5Vdc reference signal must be used (GFM/GFC option)
0-10 Vdc 11 -12 14 -15 17-18
11 Vdc
Special Order option! (PCB hardware must be changed)
4-20 mA 10 -11 13 -14 16-17
25 mA
(249 Ohm passive, not isolated current input)
3
2
1
6
5
4
9
8
7
12
11
10
15
14
13
18
17
16
21
20
19
A
B
C
D
E
F
G
Figure 4.2 - TIO Input/Output Configuration Jumpers
PV input (+) --------------- pin 4 of the 9 pin “D” connector. PV input (-) --------------- pin 8 of the 9 pin “D” connector.
8
Table 4.2 Maximum rated load impedance for SP output signals
SP
OUTPUT
TYPE
J2 JUMPER
CONFIGURATION
MAXIMUM
LOAD
IMPEDANCE
NOTE
J2A J2B J2C
0 - 5 Vdc 2 - 3 5 - 6 8 - 9
1000 Ohm
0 - 10 Vdc 2 - 3 5 - 6 8 - 9
5000 Ohm
Special Order option! (PCB hardware must be changed).
4 – 20 mA 1 - 2 4 - 5 7 - 8
900 Ohm
(24 Vdc PS)
Self-powered (non-isolated) current loop. For 12 Vdc PS the load impedance should not exceed 400 Ohm.
CAUTION: When connecting the load to the output terminals always check actual jumper J2 configuration. Do not exceed the rated values shown in the specifications (see Table 4.2). Failure to do so might cause damage to this device. Be sure to check if the wiring and the polarity of the power supply and SP signals are correct before turning the power ON. Wiring error may cause damage or faulty operation. Do not connect external voltage source to the SP output terminals.
PV output (+) --------------- pin 5 of the 9 pin "D" connector PV output (-) --------------- pin 8 of the 9 pin "D" connector
WARNING: The 4-20 mA current loop output is self-powered (non­isolated). Do not connect an external voltage source to the output signals.
4.4 Digital Communication Interface Connections
The digital interface operates via RS-232 (optional RS-485) and provides access to all applicable internal configuration parameters and data.
Communication Settings for RS-232/RS-485 communication interface:
Baud rate: default 9600 baud (user-selectable. See specification section).
Stop bit: .................... 1
Data bits: .................... 8
Parity: .................... None
Flow Control: .................... None
9
RS-232 Communication Interface Connection:
Crossover connection must be established:
RS-232 RX (pin 2 on the host PC DB9 connector)------pin 7 of the 9 pin "D" connector (TX-)
RS-232 TX (pin 3 on the host PC DB9 connector)------pin 3 of the 9 pin "D" connector (RX+)
RS-232 SIGNAL GND (pin 5 on the host PC DB9 connector)------pin 6 of the 9 pin "D" connector
RS-485 Communication Interface Connection:
The RS485 converter/adapter must be configured for: multidrop, 2 wire, half duplex mode (See Figure 4.3). The transmitter circuit must be enabled by TD or RTS (depending on which is available on the converter/adapter). Settings for the receiver circuit should follow the selection made for the transmitter circuit in order to eliminate echo.
RS-485 T(-) or R(-) ----------------- pin 7 of the 9 pin "D" connector (TX-) RS-485 T(+) or R(+) ----------------- pin 3 of the 9 pin "D" connector (RX+) RS-485 GND (if available) ----------------- pin 6 of the 9 pin "D" connector
10
Figure 4.3 RS-485 Multidrop Half Duplex Two Wire System
11
When the TIO device is set as the last device on the RS-485 bus segment and 220 Ohm bus termination is required, set the jumper J2G to position 19-20. This will result in connection 220 Ohm resistor between RS-485 (+) and (-) terminals.
4.5 Digital and Pulse Optically-Isolated Outputs
Connections
TIO is equipped with two programmable digital optically-isolated outputs. Each output can be assigned to any one of many different system events or config­ured as a pulse output (see Paragraph 5.3.9).
Digital optically-isolated outputs use dedicated 4 position 3.5mm male terminal block header J1 located on the top side of the TIO enclosure (see Figure 6.1). (Mated interface connector: Tyco Electronics P/N: 284510-4).
Optocoupler #1 - Terminal J1 (pins 1 and 2):
Plus (+) (passive) --------------- Terminal J1 pin 1 Minus (-) (passive) --------------- Terminal J1 pin 2
Optocoupler #2 - Terminal J1 (pins 3 and 4):
Plus (+) (passive) --------------- Terminal J1 pin 3 Minus (-) (passive) --------------- Terminal J1 pin 4
WARNING: Optically-isolated outputs require application of external DC voltage across terminals. Do not exceed maximum allowed limits for voltage and current provided below:
2 V < UCE < 40 V
0.2 mA < ICE < 150 mA
12
WARNING: Optically-isolated outputs have maximum absolute voltage rating 250 Vdc RMS. Do not exceed maximum allowed limits for voltage. Doing so may cause personal injury or damage to this device.
13
5. LCD KEY-PAD OPERATION: DATA ENTRY AND
CONFIGURATION
5.1 Display Indications
Initially, after the power is first turned on, the Banner Screen is shown for 2 sec­onds, then device firmware and EEPROM data base table revisions on the first line, communication interface type on the second line, baud rate and RS-485 hexadec­imal address value on third and fourth lines are shown for another 2 seconds. Subsequently, the actual process information (PI) is displayed.
Figure 5.1: TIO Firmware and Communication Interface Info Screen
Based on device configuration (Device Function as flow meter or flow controller), different parameters may be displayed in the Process Information (PI) screen by pressing the UP or DN pushbuttons.
Process Information screens can be configured to be static or dynamic (see Paragraph 5.3.11 “Display Menu”). Using Screen Mask settings user can enable (unmask) or disable (mask) up to 4 different process information combinations (see Figure 5.4). In the Static Mode the UP button pages through the PI screens in the forward direction, the DN button pages through the PI screens in the reverse direction. When the last PI screen is reached, the firmware “wraps around” and scrolls to the initial PI screen once again.
In the Dynamic Display Mode, firmware initiates automatic screen sequencing with user- adjustable screen Cycle Time (see Paragraph 5.3.11 “Display Menu”). When the last PI screen is reached, the firmware “wraps around” and scrolls to the ini­tial PI screen once again.
Fw: A002 Tbl: A001
Interface: RS-232
Baud Rate: 9600
RS485 address: 11
NOTE: Actual content of the LCD screen may vary depending on the model and device configuration.
,
NOTE: Actual content of the LCD screen may vary depending on the model and device configuration.
14
,
99.97
A:D litr/min
T1: 1589324.5 litr
99.97
S: 100.0 litr/min
T1: 1589324.5 litr
Figure 5.2: TIO Initial PI Screen (Device Function: Flow Meter)
Figure 5.3: TIO Initial PI Screen (Device Function: Flow Controller
15
Device Function: Flow Controller
Figure 5.4: TIO PI Screen (based on device function)
Device Function: Flow Meter
5.1.1 Set Point Control
(only for devices set as controller)
When TIO is configured as controller it can be used to control set point value for mated flow controller using analog output interface.
NOTE: Your TIO device input / output jumpers were factory configured according to your order. Make sure the mated flow controller has an analog input interface compatible with the TIO analog output configuration. Before applying the power and process signals make sure the input /output jumpers are installed in the correct position (See Table 6.5).
,
51.01
S: 51.0 litr/min
T1: 254898.0
litr
51.01
KF:D litr/min
T2: 436.5
51.01 litr/min
A:D KF:D
T1: 254898.0 litr
T2: 436.5
51.01 litr/min
S 51.0 0.0s
CS: 1 0.0 0
M:D L:O S: Off
litr
litr
51.01
A:D litr/min
T1: 254898.0
51.01
KF:D litr/min
T2: 436.5
51.01 litr/min
A:D KF:D
T1: 254898.0 litr
T2: 436.5
51.01 litr/min A:D KF:D
Events Reg: No Events
PO:E U/P: 1.000 litr DO# 1: Pulse Output DO# 2: Disabled
litr
litr
litr
16
The Set Point value can be adjusted locally using LCD/keypad, remotely via RS­232/RS-485 digital interface or can be programmed in advance using user-preset programs of up to sixteen steps (Program Set Point Mode).
a) Adjusting Set Point value using local LCD/keypad
Current Set Point value is displayed on the second line of the main PI screen, next to the ‘S’ character.
Pressing the ENT button while in the main PI screen will activate Set Point adjust­ment mode. The first character of the Set Point value will start to flash. Use UP or DN button to increment / decrement digit value from 0-9. Use RIGHT or LEFT but­ton to move cursor to another digit position. When desired Set Point value is entered use the ENT button to accept (save in the EEPROM) new Set Point value.
If at the end of the Set Point value entry the ESC button is pressed instead ENT, the original Set Point value will be restored and Set Point adjustment mode will be deactivated. To exit from the Set Point adjustment mode before Set Point value is accepted, press the ESC button.
b) Controlling Set Point value using Program Set Point mode
To activate Program Set Point mode the following must to be done:
1. Program Set Point mode must be Enabled (see paragraph 5.3.16 a).
2. Program Loop parameter must be set to desired value (On/Off).
3. Program Run parameter must be set to “On” (default settings is Off).
NOTE: Since the Set Point value entered via local LCD/keypad is stored in the non-volatile memory (EEPROM), it will be executed on the next device power up event.
,
NOTE: If the Program Set Point mode is enabled and the program is running, the Set Point value can be changed at any moment by the execution of the next active step. The Set Point entered via local LCD/keypad can be also changed via digital RS-232/RS-485 interface.
,
99.97
S: 100.0 litr/min
T1: 1589324.5 litr
17
NOTE: Before executing, the program should be entered in the program table (see Paragraph 5.3.16)
,
NOTE: While Program Set Point mode is running, the current Set Point value also can be changed from local LCD/keypad and digital RS-232/RS-485 communication interface. In this case, new Set Point value will be kept only until the next successive program step is executed.
,
As shown in the above drawing the Program Run parameter can be toggled “On” or “Off” by pressing the RIGHT and LEFT keypad buttons, while PI screen #4 is active. If Program Run status parameter is set to “Off”, the program execution will pause and current SP value will freeze until Program Run status parameter is set to “On”.
5.2 Menu Structure
The diagram on the Figure 5.7 gives a general overview of the standard top-level display menu structure when running firmware version A001. The ESC pushbutton is used to toggle between the Process Mode (PI screens) and the Setup menus.
UP and DN buttons must be used to move through the menu items. When the last item in the menu is reached, the menu “wraps around” and scrolls back to the beginning of the menu items list. Similarly, when the first menu item is highlight­ed and the UP button is pressed, the menu “wraps around” and scrolls down to the end of the menu item’s list.
All process configuration parameter’s settings are password-protected. In order to access or change them, Program Protection should be disabled. Each time the device is powered up, the Program Protection is enabled automatically. By default, the device is shipped from the factory with the Program Protection (PP) password set to Zero (PP Disabled). If PP password is set to Zero (Disabled), entering a PP password is not required. A subsequent screen will appear and the Program Protection menu item will be selected:
Current flow rate value
51.01 litr/min
Current SP value
Current program step
Program Mode: E/D
Program Loop Mode: On/Off
Program Run Status: On/Off
S 51.0 0.0s
CS: 1 0.0 0
M:D L:Off S: Off
Current step time
elapsed
18
Figure 5.5
Pressing the UP or DN button to select the Disabled option and then the ENT but- ton to save settings will disable program protection.
If PP password is set to any value more than Zero, the firmware will prompt with “Enter PP Password” (see Figure 5.6). User must enter up to 3 digits program protection code, in order to be able to access password protected menus. Once the correct password is entered, Program Protection is turned off until the unit is powered up again.
Figure 5.6
5.3 Parameter Entry
There are two methods of data entry: • Direct numerical number entry.
• Tabular Input from a table menu.
If the menu with direct numerical entry is selected use the UP or DN button to increment / decrement digit value from 0-9. Use the RIGHT or LEFT button to move the cursor to another digit position. When the desired value is entered, use ENT button to accept (save in the EEPROM) the new value.
19
If the menu with tabular entry is selected, the available menu options can be set with the UP and DN buttons and are accepted by pressing the ENT button.
5.3.1 Submenu “Change PP Password”
In order to get access to “Change PP Password” menu, Program Protection must be disabled. If PP password is set to Zero (Disabled), entering PP Password is not required and PP can be disabled from “Program Protection” menu (see Figure
5.5). If PP Password is set to any value more than Zero, the firmware will prompt with “Enter PP Password” (see Figure 5.6). User must enter program protection code (up to 3 digits). If PP password is lost or forgotten, contact the factory or your distributor.
NOTE: During data entry the input vales are checked for acceptability. If data is not acceptable, it is rejected and a message is generated indicating that the new data has not been accepted.
,
Program Protaction (PP)
Change PP Password
Device Information
EngUnits & K-Factor
Totalizer Settings
Opt Output Settings
General Settings
Device Diagnostic
Alarm Settings
***** Main Menu *****
Up/Dn
Event Register Menu
Analog Input counts
Analog Output Value
LCD Back Light Set.
Pulse Output Queue
CPU Temperature
Device Diagnostic
Opt Output #1 Set
Opt Output #2 Set
Totalizer #1
Totalizer #2
Pulse Output
Flow Alarm Mode
Low Flow Alarm
High Flow Alarm
Flow Alarm Delay
Flow Alarm Latch
Measuring Units
UserDefined Units
K-Factor Settings
Device ID & FS Flow
Analog/Com Interface
Firmware/EE Version
Alarm Settings
Totalizer#1 Settings
Totalizer#2 Settings
Pulse & Dig. Outputs
Mated Device Info
General Settings
Display Mode
Screen Cycle Time
Screen Mask
Display Back Light
Display Contrast
Flow Meter
Flow Controller
Baud Rate Settings
RS-485 Bus Address
Full Scale Range
Low Flow Cut-off
Flow PowerUp Delay
Fluid Std. Density
Analog Output Cal.
Analog Input Cal.
Pilot Cal. Timer
Signal Condit. Mode
NRF Num. of Samples
NRF Time Interval
Aver.Filter Damping
FlowLinearizer Mode
Program SP Mode
PSP Loop Mode
PSP Steps Mask
PSP Steps Settings
Totalizer #1 Mode
Tot#1 Flow Start
Tot#1 Action Vol.
Tot#1 PowerOn Delay
Tot#1 Auto Reset
Tot#1 AutoRes Delay
Reset Totalizer #1
Totalizer #2 Mode
Tot#2 Configuration
Tot#2 Flow Start
Tot#2 Action Vol.
Tot#2 PowerOn Delay
Tot#2 Auto Reload
Tot#2 AutoRel Delay
Reset Totalizer #2
PulseOutput Mode
Pulse Flow Start
[Unit]/Pulse
Pulse Active Time
Disabled
Low F. Alarm
High F. Alarm
F. Range H-L
Total#1 Event
Total#2 Event
Pulse Output
Diagnostic
Manual On
UD Unit K-Factor
UD Unit Time Base
UD Unit Use Density
K-Factor Mode
Int. K -Factor Index
User Def'd K-Factor
%FS
ml/sec
ml/min
ml/hr
ml/day
litr/sec
litr/min
litr/hr
litr/day
m^3/sec
m^3/min
m^3/hr
m^3/day
f^3/sec
f^3/min
f^3/hr
f^3/day
gal/sec
gal/min
gal/hr
gal/day
gram/sec
gram/min
gram/hr
gram/day
kg/sec
kg/min
kg/hr
kg/day
lb/sec
lb/min
lb/hr
lb/day
Mton/min
Mton/hr
Igal/sec
Igal/min
Igal/hr
Igal/day
MilL/min
MilL/hr
MilL/day
bbl/sec
bbl/min
bbl/hr
bbl/day
User
Ent
Event Reg. Status
Event Latch Mask
Event Reg. Mask
Reset Event Reg.
Display Settings
Device Function
Communication Sett.
Device Calibration
Signal Conditioner
Program Set Point
Ent
Ent
Ent
Ent
Ent
Ent
Ent
20
Figure 5.7
TIO Upper Levels
Menu Structure
21
Once “Change PP Password” menu is selected, the following screen will appear:
In order to protect device configuration parameters when changing the PP pass­word, the old PP password must be entered.
Once old and new passwords are entered the firmware will prompt with a confir­mation message (see Figure 5.9) that the new password has been saved.
5.3.2 Submenu “Device Information”
This submenu contains information about the device’s main configuration param­eters. These items are informational only, not password-protected, and can’t be changed (read only).
Figure 5.8
Figure 5.9
NOTE: By default the device is shipped from the factory with Program Protection (PP) password set to Zero (PP Disabled).
,
22
Figure 5.10
5.3.3 Submenu “Measuring Units”
Use the “Engineering Units and K-Factor Menu” to navigate to “Measuring Units” menu option. This option allows configuration of the flow meter/controller with the desired units of measurement. These are global settings and determine what appears on all process information screens and data log records. Units should be selected to meet your particular metering needs. A total of 47 different volumetric and mass-based engineering units are supported (See Table 5.1).
NOTE: Program the Measuring Units first because subsequent menus may be based on the units selected. Once Flow Unit of Measure is changed the Totalizer’s Volume based Unit of Measure will be automatically changed.
,
5.3.4 “Submenu User-Defined Units”
In addition to conventional flow units user-defined flow engineering units may be selected. Use the “Engineering Units and K-Factor Menu” to navigate to the “User-defined Units” menu option. This option enables user-defined configura­tion of any engineering unit required for process measurement.
The following three parameters are available for this function:
a) UD Unit volume K-Factor (defined in Liters), b) UD Unit time base (defined in Seconds), c) UD Unit use density (units with or without density support).
Before using the User-defined Unit, be sure the proper conversion factor of the new unit with respect to one liter is set (the default entry is 1.00 Liter). Also, prop­er time base values for User-Defined Units must be set.
23
TABLE 5.1 SUPPORTED ENGINEERING UNITS LIST
NUMBER
FLOW RATE
ENGINEERING
UNITS
TOTALIZER
ENGINEERING
UNITS
DESCRIPTION
1 %FS %s Percent of full scale
2 ml/sec ml Milliliter per second
3 ml/min ml Milliliter per minute
4 ml/hr ml Milliliter per hour
5 ml/day ml Milliliter per day
6 litr/sec litr Liter per second
7 litr/ min litr Liter per minute
8 litr/hr litr Liter per hour
9 litr/day litr Liter per day
10 m^3/sec m^3 Cubic meter per second
11 m^3/ min m^3 Cubic meter per minute
12 m^3/hr m^3 Cubic meter per hour
13 m^3/day m^3 Cubic meter per day
14 f^3/sec f^3 Cubic feet per second
15 f^3/min f^3 Cubic feet per minute
16 f^3/hr f^3 Cubic feet per hour
17 f^3/day f^3 Cubic feet per day
18 gal/sec gal Gal per second
19 gal/min gal Gal per minute
20 gal/hr gal Gal per hour
21 gal/day gal Gal per day
22 gram/sec gram Grams per second
23 gram/min gram Grams per minute
24 gram/hr gram Grams per hour
25 gram/day gram Grams per day
26 kg/sec kg Kilograms per second
27 kg/min kg Kilograms per minute
The following selections are available: 1 second, 60 seconds (1 minute), 3600 sec­onds (1 Hour), 86400 seconds (1 Day). The default entry is 60 seconds. If a mass­based User-defined Unit is desired, the “UD Unit Use Density” parameter must be set to “YES”. The default entry is “NO” so the Fluid STD Density parameter is not used for flow rate calculation.
24
NUMBER
FLOW RATE
ENGINEERING
UNITS
TOTALIZER
ENGINEERING
UNITS
DESCRIPTION
28 kg/hr kg Kilograms per hour
29 kg/day kg Kilograms per day
30 lb/sec lb Pounds per second
31 lb/min lb Pounds per minute
32 lb/hr lb Pounds per hour
33 lb/day lb Pounds per day
34 Mton/min Mton Metric Ton per minute
35 Mton/hr Mton Metric Ton per hour
36 Igal/sec Igal Imperial Gal per second
37 Igal/min Igal Imperial Gal per minute
38 Igal/hr Igal Imperial Gal per hour
39 Igal/day Igal Imperial Gal per day
40 MilL/min MilL Million Litr per minute
41 MilL/hr MilL Million Litr per hour
42 MilL/day MilL Million Litr per day
43 bbl/sec bbl Barrel per second
44 bbl /min bbl Barrel per minute
45 bbl /hr bbl Barrel per hour
46 bbl /day bbl Barrel per day
47 User User User-defined
25
5.3.5 Submenu “K-Factors Settings”
Conversion factors relative to Nitrogen are convenient to use when the flow meter/controller mated to the TIO is calibrated for Nitrogen and another gas is required to be measured/controlled.
Conversion factors relative to Nitrogen for up to 22 common gases are stored in the TIO (see APPENDIX II). In addition, provision is made for a user-defined con- version factor. Conversion factors may be applied to all units of measure (except %FS unit) via LCD/Keypad or digital communication interface.
The following three parameters are available for this function:
a) K-Factor Mode: Disable, Internal Index, user-defined (default Disabled) b) Internal K Factor Index: 1 – 22 (from internal K-Factor table, see APPENDIX II) c) User-defined K-Factor: 0.001 – 999.9 (default value is 1.000).
5.3.6 Submenu “Alarm Settings”
TIO provides the user with a flexible Alarm/warning system that monitors the Fluid Flow for conditions that fall outside configurable limits as well as visual feedback for the user via the LCD or via an optically-isolated output. The Flow Alarm has several attributes which may be configured by the user via LCD/Keypad or digital communication interface. These attributes control the conditions which cause the Alarm to occur and to specify actions to be taken when the flow rate is outside the specified conditions.
Depending on the TIO function configuration (flow meter or controller) there are two Alarm algorithms. If the TIO is configured as a flow meter, Flow Alarm condi­tions become true when the current flow reading is equal to or Higher/Lower than corresponding values of High and Low Flow Alarm levels. If TIO is configured as flow controller, Flow Alarm conditions become true when difference between Set Point value and current flow reading is equal or Higher/Lower than corresponding values of High and Low Flow Alarm levels.
Alarm action can be assigned with preset Delay Interval (0-3600 seconds) to acti­vate the optically-isolated output (separate for High and Low Alarm). Latch Mode control feature allows each optical output to be latched on or follow the corre­sponding Alarm status.
Following settings are available for Flow Alarm (see Figure 5.7):
Note: The conversion factors will not be applied for the % F.S. engineering unit.
,
26
a) Flow Alarm Mode (Tabular entry)
This function determines whether the Flow Alarm is Enabled or Disabled. The following selections are available: Enabled or Disabled. The default entry is Disabled. Alarm Mode selections can be set with the UP and DN buttons and are accepted by pressing ENT button.
b) Low Flow Alarm (Numerical entry)
The limit of required Low Flow Alarm value can be entered in increments of 0.1% from 0 - 100%F.S.
If a Low Alarm occurs, and one of the two optical outputs is assigned to the Low Flow Alarm Event (see paragraph 5.3.10) the optically-isolated output will be activated:
• For Flow Meter function: when the flow is less than the Low Flow Alarm value.
• For Flow Controller function: when the absolute difference between Set Point value and actual flow reading is equal to or higher than the Low Flow Alarm value and Actual Flow value is less than Set Point value.
The Low Flow Alarm condition is also indicated on the corresponding Process Information Screen by displaying L character.
c) High Flow Alarm (Numerical entry)
The limit of required High Flow Alarm value can be entered in increments of 0.1% from 0 - 100% F.S. If a High Alarm occurs, and one of the two optical outputs is assigned to the High Flow Alarm Event (see paragraph
5.3.10) the optically-isolated output will be activated for:
• Flow Meter function: when the flow is more than the High Flow Alarm value.
• Flow Controller function: when absolute difference between Set Point value and Actual Flow reading is equal to or higher than the High Flow Alarm value and actual flow value is more than set point value.
The High Flow Alarm condition is also indicated on the corresponding Process Information Screen by displaying the H character.
NOTE: For Flow Meter function the value of the Low Flow Alarm must be less than the value of the High Flow Alarm.
,
Loading...
+ 68 hidden pages