3M™ ATTEST™ STEAM INCUBATOR (56ºC) a
Product Description
The 3M™ Attest™ Steam Incubator (56ºC) is a compact, round incubator with 14 incubation wells. The 3M™ Attest™
Steam Incubator is pre-set at the appropriate temperature to promote growth of the test organism,
stearothermophilus
automatically maintained. If a temperature measurement device is used to monitor the incubator, the sensor should be in
contact with the metal heating block wall.
IntendedUse
The 3M™ Attest™ Steam Incubator is designed and has been tested to be used with 3M™ Attest™ Biological Indicators
(BI) 1261 and 1261P or 1262 and 1262P and allows rapid, reliable monitoring of steam sterilization processes. As a quick
reference, the steam monitoring system is color-coded blue and brown, respectively. Use with other biological indicators may
lead to incorrectresults.
Contraindications
None.
Warnings
• To reduce the risk of electricshock:
– do not pour any liquids into the metal heating blockholes
– do not immerse the incubator in anyliquid
– use indoorsonly
– disconnect power cord beforecleaning
– use only a disinfectant or mild detergent and wipe with a dampcloth
– there are no user serviceableparts
– make sure a properly rated power cord is installed on theincubator.
• To reduce the risk of using a load from a failed sterilizationcycle:
– match the color on the exterior label of the incubator with the same color cap on theBI
– verify the incubation conditions are correct by using a positive control BI each day a test BI isincubated.
• There is a glass ampoule inside the plastic vial of the biologicalindicator:
– crushing or excessive handling of the BI before cooling may cause the glass ampoule toburst
– wear safety glasses and heat-resistant gloves when removing the BI from thesterilizer
– wear safety glasses when crushing the biologicalindicator
– handle the BI by the cap whencrushing
– do not use your fingers to crush the glassampoule.
Precautions
• To avoid possible environmental contamination, dispose of the incubator according to applicable
governmentalregulations.
• Avoid contact with the hot surface of the metal heatingblock.
• If this product is used in a manner not specified, the protection provided by the product may beimpaired.
. The incubator’s internal temperatures are rigidly controlled to 56ºC +/-2ºC (133ºF +/-3ºF) and
Explanation of Product and Packaging Label Symbols (Note: not all symbols appear on allmodels)
Attention – Refer to the instructions foruse
Caution – Risk of electricshock
Warning – Do notimmerse
Warning – Do not pourliquids
Caution – HotSurface
Waste Electrical and Electronic Equipment(WEEE)
Mark of Conformity to EuropeanDirectives
UL Listed to US and Canadian SafetyStandards
Manufacturer
Authorized representative for the EuropeanCommunity
Environmental OperatingConditions
Environmental
Condition
Altitude 2500 (max) meters
Operating Temperature 15 – 35 ºC
Relative Humidity 20 – 80 (non-condensing) %
Voltage Range
Frequency 50/60 Hertz
Current
Installation/Over Voltage Category II
Pollution Degree 2
Operating Condition Units
Model 116:120
Model 118: 220-240
Model 116:0.25
Model 118: 0.13
VoltsAC
Volts AC
Amps
Amps
Geobacillus
Attest™ Incubator
120 volt
Model 116 (14vials)
56 +/- 2ºC (133 +/- 3ºF)
See above See above
Directions forUse
1. Plug incubator in and allow to warm-up for 30 minutes before use. It is
recommended that the unit be left on to eliminate warm-upperiods.
2. After completion of the cycle and while wearing safety glasses and heat-resistant
gloves, remove the Attest™ biological indicator from the sterilizer.
3. When the biological indicator is not contained in a test pack or other heat
absorbing packaging material, allow it to cool for an additional 10 minutes prior
tocrushing.
4. When the biological indicator is contained in a test pack or other heat absorbing
packaging material, the test pack or any other heat absorbing packaging material
should be removed from the sterilizer and opened up for 5 minutes to dissipate
heat prior to removing the biological indicator. Then allow the indicator to cool
outside the test pack for an additional 10 minutes prior tocrushing.
5. Position the biological indicator in the incubator’s metal heating block (See
Figure 1). Place the bottom of the indicator into the metal heating block so
that the indicator is at an angle of approximately 45° and the cap is above the
metalblock.
6. Push the indicator straight back (See Figure 2). This crushes the ampoule and
activates theindicator.
7. Push the activated indicator down to seat it in the metal heating block (See Figure
3). Be sure that the cap remains above the metalblock.
Refer to the Attest™ biological indicator Instructions for Use for further directions for
use, including the interpretation of results and use ofcontrols.
Customer Support
In the USA, call the 3M Health Care Help Line at 1-800-228-3957.
Outside of the USA, contact your local 3M Subsidiary.
Disposal
Dispose of instrument in accordance with federal, state, and localrequirements.
For further information, please contact your local 3M representative or contact us at
3M.com and select your country.
Attest™ Incubator
220‑240volt
Model 118 (14vials)
56 +/- 2ºC (133 +/- 3ºF)
Sterilization
Process Monitored
Steam (except
270ºF/132ºC gravity
displacement)
Steam gravity
displacement
270ºF/132ºC
Attest™
Biological
Indicators
1262 & 1262P Brown
1261 &1261P Blue
Color
Coding
Made in USA with
Globally Sourced Materials for
3M Company
2510 Conway Ave.
St. Paul, MN 55144 USA
1-800-228-3957 (USA Only)
3M.com
3M and Attest are trademarks of3M.
Used under license in Canada.
© 2020, 3M. All rights reserved.
3M et Attest sont des marques commerciales de3M
utilisées sous licence au Canada.
© 2020, 3M. Tous droits réservés.
Date of Issue: 2019-09, rev. 3
34-8725-0690-1
3M Deutschland GmbH
Health Care Business
Carl-Schurz-Str. 1
41453 Neuss, Germany