3M ATTEST User Manual

3M™ ATTEST™ STEAM INCUBATOR (56ºC) a
Product Description
The 3M™ Attest™ Steam Incubator (56ºC) is a compact, round incubator with 14 incubation wells. The 3M™ Attest™ Steam Incubator is pre-set at the appropriate temperature to promote growth of the test organism,
stearothermophilus
automatically maintained. If a temperature measurement device is used to monitor the incubator, the sensor should be in contact with the metal heating block wall.
IntendedUse
The 3M™ Attest™ Steam Incubator is designed and has been tested to be used with 3M™ Attest™ Biological Indicators (BI) 1261 and 1261P or 1262 and 1262P and allows rapid, reliable monitoring of steam sterilization processes. As a quick reference, the steam monitoring system is color-coded blue and brown, respectively. Use with other biological indicators may lead to incorrectresults.
Contraindications
None.
Warnings
• To reduce the risk of electricshock: – do not pour any liquids into the metal heating blockholes – do not immerse the incubator in anyliquid – use indoorsonly – disconnect power cord beforecleaning – use only a disinfectant or mild detergent and wipe with a dampcloth – there are no user serviceableparts – make sure a properly rated power cord is installed on theincubator.
• To reduce the risk of using a load from a failed sterilizationcycle: – match the color on the exterior label of the incubator with the same color cap on theBI – verify the incubation conditions are correct by using a positive control BI each day a test BI isincubated.
• There is a glass ampoule inside the plastic vial of the biologicalindicator: – crushing or excessive handling of the BI before cooling may cause the glass ampoule toburst – wear safety glasses and heat-resistant gloves when removing the BI from thesterilizer – wear safety glasses when crushing the biologicalindicator – handle the BI by the cap whencrushing – do not use your fingers to crush the glassampoule.
Precautions
• To avoid possible environmental contamination, dispose of the incubator according to applicable
governmentalregulations.
• Avoid contact with the hot surface of the metal heatingblock.
• If this product is used in a manner not specified, the protection provided by the product may beimpaired.
. The incubator’s internal temperatures are rigidly controlled to 56ºC +/-2ºC (133ºF +/-3ºF) and
Explanation of Product and Packaging Label Symbols (Note: not all symbols appear on allmodels)
Attention – Refer to the instructions foruse
Caution – Risk of electricshock
Warning – Do notimmerse
Warning – Do not pourliquids
Caution – HotSurface
Waste Electrical and Electronic Equipment(WEEE)
Mark of Conformity to EuropeanDirectives
UL Listed to US and Canadian SafetyStandards
Manufacturer
Authorized representative for the EuropeanCommunity
Environmental OperatingConditions
Environmental Condition
Altitude 2500 (max) meters
Operating Temperature 15 – 35 ºC
Relative Humidity 20 – 80 (non-condensing) %
Voltage Range
Frequency 50/60 Hertz
Current
Installation/Over Voltage Category II
Pollution Degree 2
Operating Condition Units
Model 116:120 Model 118: 220-240
Model 116:0.25 Model 118: 0.13
VoltsAC Volts AC
Amps Amps
Geobacillus
Attest™ Incubator 120 volt
Model 116 (14vials) 56 +/- 2ºC (133 +/- 3ºF)
See above See above
Directions forUse
1. Plug incubator in and allow to warm-up for 30 minutes before use. It is
recommended that the unit be left on to eliminate warm-upperiods.
2. After completion of the cycle and while wearing safety glasses and heat-resistant
gloves, remove the Attest™ biological indicator from the sterilizer.
3. When the biological indicator is not contained in a test pack or other heat
absorbing packaging material, allow it to cool for an additional 10 minutes prior tocrushing.
4. When the biological indicator is contained in a test pack or other heat absorbing
packaging material, the test pack or any other heat absorbing packaging material should be removed from the sterilizer and opened up for 5 minutes to dissipate heat prior to removing the biological indicator. Then allow the indicator to cool outside the test pack for an additional 10 minutes prior tocrushing.
5. Position the biological indicator in the incubator’s metal heating block (See
Figure 1). Place the bottom of the indicator into the metal heating block so that the indicator is at an angle of approximately 45° and the cap is above the metalblock.
6. Push the indicator straight back (See Figure 2). This crushes the ampoule and
activates theindicator.
7. Push the activated indicator down to seat it in the metal heating block (See Figure
3). Be sure that the cap remains above the metalblock.
Refer to the Attest™ biological indicator Instructions for Use for further directions for use, including the interpretation of results and use ofcontrols.
Customer Support
In the USA, call the 3M Health Care Help Line at 1-800-228-3957. Outside of the USA, contact your local 3M Subsidiary.
Disposal
Dispose of instrument in accordance with federal, state, and localrequirements. For further information, please contact your local 3M representative or contact us at
3M.com and select your country.
Attest™ Incubator 220‑240volt
Model 118 (14vials) 56 +/- 2ºC (133 +/- 3ºF)
Sterilization Process Monitored
Steam (except 270ºF/132ºC gravity displacement)
Steam gravity displacement 270ºF/132ºC
Attest™ Biological Indicators
1262 & 1262P Brown
1261 &1261P Blue
Color Coding
Made in USA with Globally Sourced Materials for
3M Company
2510 Conway Ave. St. Paul, MN 55144 USA 1-800-228-3957 (USA Only) 3M.com
3M and Attest are trademarks of3M. Used under license in Canada. © 2020, 3M. All rights reserved.
3M et Attest sont des marques commerciales de3M utilisées sous licence au Canada. © 2020, 3M. Tous droits réservés.
Date of Issue: 2019-09, rev. 3 34-8725-0690-1