3D Instruments DPG-6700 User guide

Page 1
KT-077
8/04
Page 1 of 20
DPG-6700 Instruction Sheet
The“Direct Drive Difference” in Digital
Form DPG-6700
The 3D Digital Pressure Gauge DPG-6700 is a loop pow­ered electronic instrument designed for applications needing high accuracy, excellent long-term stability, trouble-free operation, high performance and long life. No separate power supply is needed, the DPG-6700 takes power directly from the cur­rent loop. It incorporates a large 0.75" four digit, 25 bar LCD display with a four key sealed membrane switch.
The integral pressure transducer is a high quality, all stainless steel media-isolated sen­sor with no internal “O” rings, no silicone oil and no welds. The gauge is intended for use in the measurement of gases and liquids compati­ble with stainless steel.
The accuracy of the gauge is +/- 0.25% of full scale over a temperature of 0°C to 55°C (32°F to 131°F). Full operating temperature range is -40°C to 85°C (-40°F to 185°F).
The gauge is ideally suited to process applications with features such as an adjustable software filter to stabilize readings where the pressure is fluctuating. The pressure units are selectable and are displayed on the LCD. When a new unit is selected, the gauge automatically converts the reading to match the units. A configurable bargraph gives the user a quick visual indication as to where the pressure is relative to low and high values and whether it is increasing or decreasing. A high and low pressure reading detection shows the opera­tor the pressure extremes that the process was exposed to. In addition, the input pres­sure is continuously compared against two adjustable setpoints that can be adjusted using the front keypad on the gauge. The 4-20 mA loop can be configured anywhere within the range of the gauge, as well as reverse acting. The gauge uses the loop current as its pri­mary power. With the absence of loop current it automatically switches to the internal “C” size 3.6 volt replaceable lithium cell. It continues to operate as a battery-powered gauge until loop power returns. The gauge is designed with the latest microcontroller technology for extremely long battery life. To extend battery life even longer, a programmable power­on-timer can be set from 0 to 720 minutes.
The enclosure is the standard 3D 4-1/2" ‘Dyna-Mount’ ABS case with a special swivel fit­ting and moveable flange. The gauge can be converted to front or back flange and bot­tom or back mount in the field with no more than a screwdriver . The digital pressure gauge accepts all of the optional accessories offered in the mechanical gauge line includ­ing the metallic GP-140 chemical seal, elastomeric isolator GP-400 and the 316 SS sin­tered filter RS-380-1 for particulate laden process medias.
And finally, the DPG-6700 is manufactured and serviced by the company that makes THE Premier Pressure Gauge on the market today. Get The Direct Drive Difference in
Digital Toda y!
INTRODUCTION
Page 2
The DPG-6700 digital pressure gauge is shipped with the internal battery installed so that it is ready to use. Press any key and the gauge will power-up. A complete LCD segment test is performed followed by the full scale reading of the gauge in psi units. The gauge may be adjusted to local atmospheric condition using the ZERO/TARE function - see page 4. The full scale reading of the gauge in psi units can be displayed at any time by simultane-
ously pressing both arrow keys (and ).
The DPG-6700 measures “gauge” pres­sure. “Gauge” pressure measures the input pressure referenced to ambient atmospheric pressure. The gauge will also read vacuum down to absolute zero. The vacuum readings will be displayed as minus in the chosen unit of measure.
Mount the gauge to a suitable flat surface using the flange and hardware supplied. Avoid mounting the gauge with the LCD display facing direct sunlight. The display temperature range is -40°F to +185°F. (-40°C to +85°C)
NOTE: Although flange mounting is sug­gested for high vibration applications, the flange does not have to be used to mount the gauge. Mounting the gauge using its process connection to a pipe is acceptable as long as excessive vibration is not expe­rienced.
Attach the gauge to the pressure fitting using teflon tape or equivalent. Tighten the process connection by hand and con­tinue to tighten 2 to 3 turns using a 9/16" wrench on the stem of the gauge. Do not use the gauge case to tighten pressure fit­ting.
WARNINGS AND CAUTIONS:
To ensure safe and accurate operation please comply with the following warnings:
W
ARNINGS:
HAZARDOUS AREA OPERATION - The
DPG-6700 series gauge is CSA-certified for CLASS I, DIVISION 1, GROUPS C and D areas intrinsically safe, when used with a CSA approved intrinsic safety barrier.
SUBSTITUTION OF COMPONENTS MAY IMPAIR INTRINSIC SAFETY. To prevent
ignition of a hazardous atmosphere, batter­ies must only be changed in an area known to be nonhazardous.
SEVERE INJURY OR DAMAGE CAN OCCUR THROUGH IMPROPER USE OF PRESSURE INSTRUMENTS. Do not
exceed recommended pressure limits of transducer, tubing and fittings. Be certain all pressure connections are secured.
Deviation from these installation instruc­tions may lead to improper operation of the monitored machine which could cause per­sonal injury to operators or other nearby personnel.
CA
UTIONS:
THIS GAUGE CAN DISPLAY ZERO PRESSURE EVEN WITH UP TO 20% OF FULL SCALE PRESSURE APPLIED. Do
not rely on the display indication when dis­connecting pressure - it may not be indi­cating true pressure.
The DPG-6700 digital pressure gauge is suitable for use in Class I, Division 2, Groups C and D hazardous area locations, when installed in accordance with these instructions.
OPERATING INSTRUCTIONS
KT-077 8/04 Page 2 of 20
Page 3
KT-077
8/04
Page 3 of 20
LOOP WIRING
The DPG-6700 is designed to transmit analog 4 to 20 mA signals over an industry standard 2-wire current loop for the control of valves, actuators, and other devices commonly used in process control. The gauge also is powered directly from the two wires of the current loop. Connect the loop wires to terminals 1(+) and 2(-); power requirement is 12 to 32 VDC. See wiring diagram page 15. The loop is protected against open and short circuits and reverse wiring. The maximum load resist­ance that can be tolerated in the loop is determined by the supply voltage, but should not exceed 1000 ohms. Refer to page 15 for the maximum load resistor verses the supply voltage.
- A 250 ohm loop resistor can be used over the supply voltage range from 14 to 32 Vdc.
- At 12 VDC supply voltage, the maxi­mum loop resistor for 20 mA loop output current is 150 ohms.
- If using a 500 ohm loop resistor and a desired full scale loop output of 20 mA, the supply voltage must be between 17 and 32 VDC.
HAZARDOUS AREA OPERA
TION
The DPG-6700 gauge is CSA-certified for CLASS I, DIVISION 2, GROUPS C and D areas when mounted in a suitable enclo­sure. The device may be operated as CLASS I, DIVISION I, GROUPS C and D intrinsically safe, if the following conditions are met:
A. The gauge is powered from a CSA-certi-
fied zener barrier rated 30 volts max., 120 ohms min. A suitable barrier is a Stahl part no. 9002/13-280-110-00; fol­low the installation instructions supplied with the barrier.
B. The switch outputs, if used, are connect-
ed to the sensor inputs of an Altronic DA or DD annunciator system with the 690 series power supply. Contact 3D Instruments for additional information.
In addition, the following requirements must be met (see NFPA standard no. 493):
1. The intrinsically-safe gauge wires within the panel enclosure must be kept at least two (2) inches away from other wiring.
2. Wiring to valves, actuators and other control devices must have a grade of insulation capable of withstanding an AC voltage of 500 volts RMS.
3. Control wires must be run in separate conduits and junction boxes from high voltage wires such as ignition, fuel valve, and other high voltage wiring. Refer to page 14 for wiring instructions.
W
ARNING:
SUBSTITUTION OF COMPONENTS MAY IMPAIR INTRINSIC SAFETY AND/OR SUITABILITY FOR CLASS I, DIV. 2.
DO NOT DISCONNECT EQUIPMENT IN DIV. 2 ENVIRONMENT UNLESS POWER IS SWITCHED OFF OR THE AREA IS KNOWN TO BE NON-HAZARDOUS.
OUTPUT SWITCH
WIRING
There are two user-programmable output switches that can be set anywhere within the range of the gauge. A fault condition will cause one or both of the output switch­es to turn ON/OFF to their common. These switches are solid state, open drain FET’s and are isolated from the loop sup­ply. The common is also isolated from the loop supply thus allowing for low-side or high-side switching. The switches are rated at 100 VDC, 1 Amp continuous. Please refer to the wiring diagram on page
16.
Page 4
ZERO /
TARE
The zero function is used to “zero” the gauge to compensate for any ambient effects that cause an offset at zero pres­sure. The zero offset is limited to +/-2 psi to +/-10 psi depending on the range of the gauge. To “zero” the gauge, vent the pres­sure to the gauge to atmosphere (0 psig); press the SETUP key, the gauge will read “ZErO”, press ENTER and the gauge will be zeroed at the current pressure value.
The tare function can be used to offset the zero point of the gauge to any vacuum or positive gauge pressure from 0 to 20% of full range of the gauge. Both the under and over arrows will blink to indicate when tare is used. To access the tare function, press the SETUP key until the display reads “tare”, press ENTER and the gauge will be zeroed at the current pressure value. The tare offset can be canceled at any time by pressing SETUP and then the up or down arrow key then ENTER. The zero offset will return to the original zero reading. Note that offsetting the zero using the tare function can limit the range of the gauge. Using a 0-100 psi gauge as an example, if the tare function is exer­cised at 20 psi reading, the maximum pressure that can be displayed will be 110% of actual full scale (110 psi minus 20 psi offset = 90 psi displayed).
CAUTION: Before disconnecting the gauge from its pressure source make sure that the pressure is at zero psi by cancel­ing the tare function as described above.
O
VER-PRESSURE CONDITIONS
The DPG-6700 series gauge has a dis­played range of -14.7 psig (or equivalent) to 110% of gauge range. If these limits are exceeded, the low or high out of range arrows will be displayed and the reading will remain at the limit indicated above.
NOTE: If tare was used to offset zero, the max reading will be offset by the zero off­set amount. For an explanation, see ZERO / TARE section above.
IMPORTANT: The internal pressure trans­ducer in the gauge will withstand a maxi­mum of 2 times the rated pressure without damage. Burst pressure is 5 times or 25,000 psi, whichever is less. If the over­load rating is exceeded, failure may occur. Pressure fluctuations may not be dis­played, especially if the filter value is set for high dampening. Pressure fluctuations occur in most systems; select the gauge range with a rating high enough to prevent overload by peak pressures of pulsations. It is recommended that a pressure snubber be used which will reduce the peak pres­sure applied to the gauge. The life of the gauge will be extended with the use of a snubber or pulsation dampener.
SELECTING UNITS
A number of pressure units are available with the DPG-6700 series gauge; the most common ones are displayed on the LCD. To select a different unit press the SETUP key until the display reads “Unit”, use the
or (up or down) arrow key to scroll
through the displayed available units, select a unit and press ENTER. To config­ure the gauge for pressure units that are not on the display, select “SPEC” (special) and press ENTER. Using the flowchart on page 10 as a guide, configure the decimal point location and the low and high values. The low value will be the pressure value in the chosen units equivalent to -14.7 psig, and the high value will be the pressure value in the chosen units at 110% of gauge range.
NOTE: The zero and tare offsets are not adjustable, when using the “special” con­figuration.
KT-077 8/04 Page 4 of 20
Page 5
KT-077
8/04
Page 5 of 20
MEASURING V
ACUUM
All pressure ranges of the DPG-6700 can measure vacuum down to -14.7 psig or equivalent in another unit of measure. When measuring a pressure less than ambient barometric conditions, a minus (-) sign will be displayed. If the input goes below -14.7 psig, the display will read -
14.7 (or equivalent) and the lower out of range arrow will be displayed.
NOTE: The accuracy for readings below 0 psi is +/-2% of highest vacuum reading.
ADJUSTING
THE DISPLAY FILTER
The display filter can be used to stabilize the display reading of a pulsating pressure input. The filter value is read-out from 1 to 255; 1 being minimum filter value, 255 being maximum filter value. Below are some typical filter values and their effect on the display reading. Settling values are approximate times in seconds to reach 90% of new reading. To set the filter value press the SETUP key until the display reads "FILt". Press the ENTER key and the display will read the previously set filter
value. Use the ▲ or ▼ arrow key to
increase or decrease the filter value and press ENTER to save the new filter value.
NOTE: The filter values listed are when the gauge is powered from loop power, when operating from internal battery power the filter times will be approximately 7 times longer.
FILTER VALUE 1 64 128 200 SETTLING, SEC. .5 1 1.3 1.5
FILTER VALUE 210 220 230 240 SETTLING, SEC. 1.8 2.3 3 5
FILTER VALUE 245 247 253 255 SETTLING, SEC. 8.5 11 30 90
LO / HI DETECTION
The DPG-6700 continuously records the lowest and highest filtered pressure applied to the gauge. To view the low and high values, press the SETUP key until the display reads “LO.HI” and press ENTER. The display will read the low value record­ed. Press ENTER to read the high value recorded. The lower half of the bargraph will display to indicate the reading is the low value and the upper half of the bar­graph will display to indicate the reading is the high value. These readings will remain stored if the gauge is powered down.
NOTE: Values will be lost when battery is replaced.
RESETTING RECORDED LO / HI VALUES
To clear the current values, display either the low or high value and press either the
or arrow key; the display will show
“Clrd” and the values will begin new low and high values.
Page 6
KT-077 8/04 Page 6 of 20
CONFIGURING THE B
ARGRAPH
The bargraph shows the gauge pressure in a graph form between two points. The bargraph can be configured to display pressure in different modes. To reach the bargraph menu, press the SETUP key until the display reads “bAr”, the previous con­figuration will be displayed. To change the
type of bargraph, press either the ▲ or ▼
arrow key to scroll between the different configurations. Upon selecting the desired bargraph type, press ENTER to save.
"On IIIIII" Bar mode between the
two max points. When this mode is selected the low point will be 0 psig or equivalent and the bars will increase to 100% of range.
"On I I I" Single bar between
the two max points. When this mode is selected the low point will be 0 psig or equivalent and a single bar will increase to 100% of range.
"On.HL IIIIII" Bar mode between
two selected points. When this mode is selected press ENTER and adjust the low point to the desired value. Press ENTER and adjust the high point. The bars will increase from the low point to the high point.
"On.HL I I I" Single bar between
two selected points. When this mode is selected press ENTER and adjust the low point to the desired value. Press ENTER and adjust the high point. A sin­gle bar will increase from the low point to the high point.
"OFF" No bargraph displayed.
When selected the bargraph will be off.
CURRENT LOOP
The 4-20 mA current loop allows the user to output a signal proportional to the pres­sure being measured and displayed. The nominal range is 4 mA to 20 mA; the maxi­mum range is 3 mA to 25 mA. To config­ure the current loop, press the SETUP key until the display reads "LOOP" and press ENTER. The display will read the value for the previously set 4 mA pressure point in
the previously selected units. Use the or arrow key to increase or decrease
the numeric value for the 4 mA point. Press ENTER and the numeric value for the 20 mA point will be displayed. Again
use the ▲ or ▼ arrow key to adjust the
desired 20 mA numeric value. Press ENTER to save the new 4-20 mA configu­ration and return to the normal reading. The 4-20 mA current loop can be config­ured for reverse action. Simply configure the 4 mA point with a higher pressure than the 20 mA value and visa-versa.
NOTE: Output signal goes to zero when using battery power.
ADJUSTING AND CONFIGURING SET
-
POINTS
There are two individually adjustable and configurable setpoints which can be set anywhere within the range of the gauge. NOTE: The display and setpoints will remain active with the loss of loop power (they will be kept active by the internal bat­tery). However if both setpoints are not used (set to “OFF”), and the power-on­timer is set to something other than zero and it expires, the gauge will go into power down mode to extend battery life until a key is pressed or loop power returns. To view or configure the setpoints, press the SETUP key until the display reads “SP.1" or “SP.2" and press the ENTER key. The configuration menus are as follows:
LOW, HIGH, LOOP, OFF - Each switch can individually be set to trip on a high
Page 7
pressure, a low pressure, on loss of loop or off. The display will read either “LO”,
“HI”, LOOP or OFF; use the ▲ or ▼ arrow
key to make a selection and press ENTER. If set for “HI”, the switch will change states on increasing pressure at the setpoint value. If set for “LO”, the switch will change states on decreasing pressure at the setpoint value. If set for “LOOP”, the switch will change states on the loss of the 4-20 mA loop voltage/current. If set for “OFF” the switch will be set to off or open and will not change states. When set to off, no further setpoint configurations will be displayed; the display will show “SAVE/donE” and return to the normal dis­play screen.
SETPOINT INDICATION - When the set­point indication is set to “On” and a set­point is tripped, the display reading will alternate between the pressure reading and a setpoint label. The setpoint labels are LO.1, LO.2, HI.1, HI.2, LP.1, and LP.2. If the setpoint indication is set to “OFF” and a setpoint is tripped, the display will continuously read the input pressure. Each setpoint can be individually config­ured to either show an indication or not.
Use the ▲ or ▼ arrow key to select “On”
or “OFF” and press ENTER to save.
NORMALLY OPEN OR NORMALLY CLOSED - Each switch can be set to nor­mally open “nO” or normally closed “nC” for the normal or safe state. For example, if set for normally open and a high set­point, the switch will be open below the pressure setpoint and close upon increas­ing pressure at the setpoint value. Use the
or arrow key to select normally open
“nO” or normally closed “nC” and press ENTER to save.
ADJUSTING SETPOINTS - With the set-
point value displayed, press the ▲ or ▼
arrow key to increase or decrease the value until the desired trip point for that switch is reached. Press ENTER to accept the new value. The new setpoint value will be saved to nonvolatile memory at the end of the setpoint menu when the ENTER key
is pressed and the display shows “SAVE/donE”. Pressing SETUP while in this menu will cause the display to return to the normal screen with no changes tak­ing effect.
NOTE: When adjusting the setpoints, the previous setpoint values are monitored, and the new value is monitored only at the end of the setpoint menu when the ENTER key is pressed and the display reads SAVE/donE. If no key is pressed for 15 seconds, the display will return to the nor­mal mode and the configuration will revert back to the previous parameters.
HYSTERESIS - Hysteresis sometimes is referred to as a deadband value. It is the numeric value above the low setpoint value that the switch will clear, and below the high setpoint value that the switch will clear. The hysteresis value is displayed in absolute pressure units that the gauge is configured for. For example, if a 90 psi high setpoint with a 10 psi hysteresis is desired (80 psi release point), set the "HySt" reset value at 10 psi. The hystere­sis can be individually set for each switch. Press ENTER and the display will show "HySt" for one second and then display the previously set hysteresis value. Use the
or arrow key to increase or decrease
the hysteresis value and then press ENTER to save the new value. The hys­teresis value range for a low setpoint is from the setpoint value up to 110% of full scale, and for a high setpoint, from the set­point value to absolute zero pressure units.
NOTE: If the hysteresis value is set so that it cannot be cleared easily by gauge pressure change, the setpoint can be cleared manually by going to the setpoint menu and turning the setpoint off. Make sure the gauge pressure is below the high setpoint or above the low setpoint and go back into the setpoint menu and turn the setpoint on again. Setting a large hystere­sis value will in effect act like latching out­puts.
KT-077
8/04
Page 7 of 20
Page 8
KT-077 8/04 Page 8 of 20
POWER ON
TIMER
The power-on-timer is used when the loop is not powered and the unit is operating of f of battery power. When received, the power-on-timer is set to 60 minutes. The gauge will turn on with any key press and will remain on for 60 minutes without a key press. The power-on-timer can be set anywhere between 1 and 720 minutes (12 hours). If “0" is entered, the gauge will remain on continuously. To change the timer, press the SETUP key until the dis­play reads “POt”, press the ENTER key and the current timer value will be dis-
played. Press the ▲ or ▼ arrow key to
increase or decrease the on time; press ENTER to save.
NOTE: Both setpoints must be set to “OFF” for the power-on-timer to be active.
B
ATTERY - LOW INDICATION AND
REPLA
CEMENT
The gauge is designed to give four (4) years of service life with the supplied, high energy density, extra long life lithium bat­tery. When the battery needs replacement, the display will alternate between the pres­sure reading and “L.BAt”. At this time the gauge will continue to work properly. When the “L.BAt” indication displays con­tinuously, the battery is exhausted and must be replaced. Use 1 “C” size 3.6 volt high energy lithium battery, Tadiran models TL-5920 or TL-2200 or Saft model LS26500 or equivalent. See page 17 for battery replacement instructions.
NOTE: DO NOT use a 1.5 volt “C” size battery. Such use will void the warranty.
If adjustment is required, we recommend returning the gauge to the factory. Factory service offers many benefits. We have the facilities to provide calibration reports that include test data at a variety of tempera­tures utilizing NIST traceable standards. In addition, upgrades may be available to add or enhance operating features. We designed the product to last, and we sup­port it so that you can get the most from your investment. Although we prefer that you return the DPG-6700 to 3D instruments for calibra­tion, ordinary adjustments may be per­formed by a qualified person with appropri­ate equipment.
The calibration standard used should have an accuracy of 0.05% of full scale or bet­ter.
NOTE: Be sure that the units of the cali­brator match the units of the gauge before performing a calibration.
Calibration may be performed many times over the life of the gauge. Calibration can be performed from the front keypad without disassembling the gauge. To calibrate the gauge, use a dead weight tester or the 3D Star Calibrator as the calibration standard.
NOTE: During calibration, the gauge allows 2 minutes between keystrokes to change or save a new calibration. If 2 minutes lapse without a keystroke, the gauge will automatically return to the nor­mal mode without making any changes. The new calibration information is saved only if the ENTER key is pressed and the gauge reads "SAVE/donE".
CALIBRATION PROCEDURE - To cali-
brate the gauge, press and hold the ▲ key
and press the SETUP key until the display reads "CAL", press ENTER; the display will read "CAL.L" for the low or zero calibration value. Adjust the standard for a reading at zero and press ENTER. After the “bar”
timer expires, use the ▲ or ▼ arrow key to
CALIBRATION
Page 9
KT-077
8/04
Page 9 of 20
increase or decrease the display reading to match the reading of the standard and press ENTER; the display will read "CAL.H" for the high or span calibration value. Adjust the standard for a reading at or near full scale and press ENTER. After
the bar timer expires, use the ▲ or ▼
arrow key to increase or decrease the dis­play reading to match the standard and press ENTER. The display will read "SAVE/donE" and return to the normal reading with the new calibration values stored in permanent memory.
The DPG-6700 gauge has a feature that allows a slight adjustment of either the zero or span values individually. This type of calibration can be used to "tweak" the readout to match that of a known value without actually performing a formal cali­bration procedure.
NOTE: This type of adjustment will invali­date calibration settings resulting from the procedures in the above section.
ZERO ADJUSTMENT ONLY - To make a small adjustment on the zero calibration value of the gauge, enter the calibration mode, the display reads "CAL", press ENTER; the display will read "CAL.L" for the low or zero calibration value. Adjust the standard for a reading at zero and press ENTER. After the “bar” timer
expires, use the ▲ or ▼ arrow key to
increase or decrease the display reading to match the reading of the standard and press ENTER. The display will read "CAL.H"; press the SETUP key and the display will read "SAVE/donE" and will return to the normal reading with the new zero calibration value stored in permanent memory.
SPAN ADJUSTMENT ONLY - To make a small adjustment on the span point of the gauge, enter the calibration mode, the dis­play reads "CAL", press ENTER; the dis­play will read "CAL.L". Press the SETUP key and the display will read "CAL.H" for the high or span calibration value. With the standard at the desired span value
press ENTER and after the “bar” timer
expires use the ▲ or ▼ arrow key to
increase or decrease the display reading to match the standard and press ENTER. The display will read "SAVE/donE" and will return to the normal reading with the new span calibration value stored in permanent memory.
SELECTING THE DEFAULT CALIBRA­TION - The user can at any time during the life of the gauge return to the factory cali-
bration values. Press and hold the
arrow key and press the SETUP key until the display reads "CAL", now press the SETUP key until the gauge reads “dFLt”. Press ENTER and the factory calibration values will be recalled and used for the calibration of the gauge.
For the 4
1
2" black ABS plastic case config- uration, the product part number for the various pressure ranges are as follows:
67544-xxB71
Pressure Range Code (xx) 50 psig = 18 100 psig = 23 200 psig = 25 500 psig = 27 1000 psig = 29 2500 psig = 47 5000 psig = 35 10,000 psig = 38
300 Series SS case available. Consult 3D Instruments for details.
DPG-6700 PART NUMBERING SYSTEM
Page 10
KT-077 8/04 Page 10 of 20
End
End
SETUP SETUP
SETUP SETUP
UnIt FILt
LO.HI
1-255
dPnt
XXXX
XXXX
UnIt
SETUP SETUP
SETUP SETUP
SETUP
ENTER ENTER
ENTER
ENTER
SPEC
ENTER
ENTER
ENTER
ENTER
ENTER
ENTER
ENTER
ENTER
CLEARS BOTH
ADJUST LOW POINT
ADJUST HI POINT
XXXX
CLrD
XXXX
psi
psi
CAL
CAL.L
CAL.H
CLP
XXXX
XXXX
LP.4
XXXX
ENTER
ENTER
ENTER
ENTER
ENTER
ENTER
LP.20
XXXX
ENTER
ENTER
ENTER
ENTER
NO CHANGE
NORMAL ZErO tArE
0.0
OR
CLrd
SAVE/ donE
Err
L < H
L > H
TEST
CAL
OR
OR
HOLD THEN SETUP
DISP FOR 1 SEC
psl H2O Hg Kg/cm2 mmHg mmH2O bar mbar SPEC
m
m
FLOWCHART
LOOP
DIGITAL
Page 11
KT-077
8/04
Page 11 of 20
bAr
LO, HI LOOP,
OR OFF
LO, HI LOOP,
OR OFF
ADJUST SETPOINT ONE
ADJUST HYSTERESIS VALUE
DISP FOR 1 SEC
N/O OR N/C
LO/HI/
LOOP/OFF
SETPOINT INDICATOR
SETUP
POWER ON TIMER
POt
0-720
min
XXXX
FLOWCHART KEY
DOUBLE BARS­USE UP AND DOWN ARROW KEYS TO SCROLL
DASHED LINES­MAKE SELECTION
AND
NOTES:
1. PRESS BOTH
2. PRESS SETUP TO ESCAPE OUT OF 2ND LEVEL MENUS WITH NO CHANGE.
TOGETHER TO DISPLAY THE FULL SCALE READING OF THE GAUGE.
SETUPSETUP
ENTER
ENTER
ENTER
ENTER
ADJUST LOW POINT
ADJUST HI POINT
ADJUST 4 mA POINT
ADJUST 20 mA POINT
XXXX
LOOP
XXXX
psi
psi
psi
psi psi
psi
psi
psi
psi
ENTER
ENTER
ENTER
ENTER
ENTER
ENTER
ENTER
ENTER
ENTER
ENTER
ENTER
ENTER
ENTER
dFLt
ENTERSETUP
SETUP
SP.1
S.Ind
SP.2
On.HL
psi H2O Hg Kg/cm2 mmHg mmH2O bar mbar SPEC
On
On
On.HL
On.HL
On.HL
OFF
XXXX
XXXX
LO/HI/
LOOP/OFF
On/OFF
ENTER
ENTER
nO/nC
XXXX
0-xxxx
HySt
ADJUST SETPOINT TWO
ADJUST HYSTERESIS VALUE
DISP FOR 1 SEC
N/O OR N/C
SETPOINT INDICATOR
ENTER
ENTER
ENTER
S.Ind
On/OFF
ENTER
ENTER
nO/nC
XXXX
0-xxxx
HySt
DPG-6700 POWERED GAUGE
Page 12
KT-077 8/04 Page 12 of 20
A
CCURACY
0 to Full Scale +/- 0.25% of Full Scale 0 to -14.7 psig +/- 2% of highest (or equivalent) vacuum reading
Loop 4-20 mA +/- 0.25% of Full Scale
(positive pressure) +/- 2% of highest reading (vacuum)
NOTE: Accuracy specs are for one year and include all effects of linearity, hystere­sis, repeatability and temperature, within the specified operating temperature range.
Exposure to environmental extremes of temperature, shock or vibration may war­rant a more frequent re-certification period.
MEDIA COMP
ATIBILITY
Liquids and gases compatible with 316 Stainless Steel.
PO
WER
Loop Power 10 to 32 VDC nominal
reverse polarity protected
Battery 1 each 3.6 volt lithium “C”
cell, Tadiran model TL-5920 (8.5 Ah), Tadiran model TL­2200 (7.2 Ah) or Saft model LS26500 (7.3 Ah) or equiva­lent. NOTE: Do NOT use a 1.5 volt “C” size battery.
Battery Life 4 years of continuous opera-
tion.
Low Battery Two indicators ­Indicator 1st: alternating reading of
pressure and L.BAt 2nd: a continuous reading of L.BAt.
CURRENT LOOP OUTPUT
Nominal range: 4 mA to 20 mA
Max range: 3 mA to 25 mA
OUTPUT SWITCHES
2 each independent, solid state, isolated open drain FET rated 100 VDC, 1.0 amp continuous.
DISPLA
Y AND CURRENT LOOP
UPDATE RA
TE
Display - 4 updates per second with loop power; 1 update per second when battery powered Loop - approx. 12 per second
PRESSURE FITTING
1/4" male NPT, – 316 SS
ENCLOSURE
Case ABS or optional
300 Series SS Hardware Stainless Steel Dimensions Refer to Page 18 Weight 514 g (18oz)
including battery
TEMPERA
TURE
Operating & Compensated Range: 0°C to 55°C (32°F to 131°F) Overall Operating Range:
-40°C to 85°C (-40°F to 185°F) Storage Range:
-40°C to 85°C -40°F to 185°F)
HUMIDITY
Temperature Range Humidity 0 to 10°C Uncontrolled 10 to 30°C 0 to 95° Relative 30 to 40°C 0 to 75° Relative 40 to 50°C 0 to 45° Relative
SPECIFICATIONS
Page 13
KT-077
8/04
Page 13 of 20
REPLACEMENT PARTS
User replaceable battery - see page 17.
PRESSURE CONVERSIONS
1 psi = 27.6806 inches of water column
(water at 4°C [39.2°F])
1 psi = 2.03602 inches of mercury
(mercury at 0°C [32°F]) 1 psi = 6.8948 kilopascals 1 psi = 51.7149 millimeters of mercury
(mercury at 0°C [32°F]) 1 psi = 703.087 millimeters of water
column (water at 4°C [39.2°F]) 1 psi = 0.068948 bar 1 psi = 68.948 millibar 1 psi = 0.070307 kilograms per square
centimeter
OPTIONS & A
CCESSORIES
NIST Certificate of Conformance (stan­dard) NIST Certificate w/ applied pressure and reading at 5 points P/N 8110-300 (Pneumatic Handpump: 0 to 300 psi) P/N 8111-300 (Duplex Handpump: -13.7 to 300 psi) P/N 8112-3000 (Hydraulic Handpump: 0 to 3000 psi) P/N 5150-xxxxx (Star Pressure Calibrator) P/N RS380-1 (Stainless Steel Filter) P/N GP140KT-B/X (Metallic Diaphragm Seal) GP1007 Sanitary 1 1/2" Ladish Tri-clamp GP1002 Sanitary 2" Ladish Tri-clamp P/N 2001-0090 AC Adapter (110V
24VDC) P/N 2230-0003 Replacement 3.6 VDC Lithium Cell
PRESSURE RANGE/DISPLAY RESOLUTION CHART
Available Pressure Ranges & Display Resolution
Pressure Max Over Display Resolution
Range Pressure Available Units
psig psig psig In In mm mm bar mbar Kg/cm2
H2O Hg H2O Hg
-14.7 – 50 100 0.1 1 0.1 1 1 0.001 1 0.001
-14.7 – 100 200 0.1 1 0.1 1 1 0.01 1 0.01
-14.7 – 200 400 0.1 1 1 1 1 0.01 1 0.01
-15 – 500 1,000 1 1 1 1 0.01 1 0.01
-15 – 1,000 2,000 1 1 1 1 0.1 1 0.1
-15 – 2,500 5,000 1 1 0.1 0.1
-15 – 5,000 10,000 1 1 0.1 0.1
-15 – 10,000 20,000 1 1 1 1
Notes:
1. Measured pressure range is -14.7 psig to 1 10% of Full Scale for pressure ranges 200 psi and below and -15 to 110% of Full Scale for pressure ranges above 200 psi.
2. Burst pressure is 5X or 25,000 psi, whichever is less.
3. Display maximum is 9999 regardless of engineering unit.
Page 14
KT-077 8/04 Page 14 of 20
WIRING DIAGRAM - INTRINSICALLY SAFE
NOTES:
1. DC power must be from a CSA certified zener barrier rated 30 volts max. 120
min. A suitable barrier is a Stahl part no. 9002/13-280-110-00. Follow the instal­lation instructions supplied with the barrier.
2. With Stahl 9002/13-280-110-00 the sensing resistor can be either on the positive
or negative side of the supply.
3. Loop voltage must be 24 VDC nominal for use with barrier .
4. The switch outputs, if used must be connected to the sensor inputs of an Altronic
DA or DD annunciator system with 690 series power supply. Contact 3D Instruments for additional information.
5. Maximum non-hazardous area voltage must not exceed 250V.
6. Install in accordance with C.E.C. Part I.
SAFE LOCATION HAZARDOUS LOCATION
CLASS 1, DIV. 1, GROUP D
SENSING RESISTOR
TYPICAL
250
ANNUNCIATOR
SAFE LOCATION
LOOP VOLTAGE
24 VDC TYPICAL
-
4-20 mA
DA OR DD
SERIES
TO ANNUNCIATOR
POINTS #10-57
LOOP
HAZARDOUS LOCATION
CLASS I, DIV.1, GROUP D
+
BARRIER
3
42
1
690 SERIES POWER
SUPPLY
DPG-6700 PRESSURE
GAUGE
-
+
+-+
123456
SW1
-
SW2
Page 15
KT-077
8/04
Page 15 of 20
WIRING DIAGRAM - CURRENT LOOP
Page 16
KT-077 8/04 Page 16 of 20
WIRING DIAGRAM - OUTPUT SWITCHES
Page 17
KT-077
8/04
Page # of 20
BATTERY REPLACEMENT
Use 1 each “C” size 3.6 volt high energy lithium battery , Tadiran models TL-5920 or TL-2200 or Saft model LS26500 or equivalent.
NOTE: DO NOT use a 1.5 volt “C” size battery.
1. Remove the four access screws on the back of the gauge as shown.
2. Remove snap ring from the front of the gauge.
3. Carefully separate the front of the gauge from the back housing.
4. Install battery as indicated on battery holder.
Page 18
KT-077 8/04 Page 18 of 20
DIMENSIONAL DRAWING
Page 19
KT-077
8/04
Page 19 of 20
WARRANTY
vice,
we can be reached at:
3D Instruments, LP warrants the DPG-6700 to be free from defects in material and work­manship under normal use and service for one (1) year from date of purchase to the origi­nal purchaser. It does not apply to batteries or when the product has been misused, altered or damaged by accident or abnormal conditions of operation.
Within one (1) year from date of purchase, 3D Instruments will, at our option, repair or replace a defective device free of charge and the device will be returned, transportation prepaid. However, if we determine the failure was caused by misuse, alteration, accident or abnormal condition of operation, you will be billed for the repair.
3D INSTRUMENTS, LP MAKES NO WARRANTY OTHER THAN THE LIMITED WAR­RANTY STATED ABOVE. ALL WARRANTIES, INCLUDING IMPLIED WARRANTIES OF MERCHANTABILITY OR FITNESS FOR ANY PARTICULAR PURPOSE, ARE LIMITED TO A PERIOD OF ONE (1) YEAR FROM THE DATE OF PURCHASE. 3D INSTRU­MENTS, LLC SHALL NOT BE LIABLE FOR ANY SPECIAL, INCIDENTAL OR CONSE­QUENTIAL DAMAGES,WHETHER IN CONTRACT, TORT OR OTHERWISE.
For warranty or non-warranty ser
Phone........................714•399•9200
Email .........................info@3dinstruments.com
Address .....................3D Instruments, LP
Web...........................www.3dinstruments.com
Return Authorization numbers are not required for servicing. Please return, freight pre­paid, to the address above and include a Contact Name, Address, Phone and Fax
Number. If you wish to be notified of the charges before any service is done, 3D Instruments will contact you after evaluating the unit. Units evaluated but not serviced are subject to an evaluation charge. Defective units need to be returned to 3D Instruments, LP within
90 days of identification of a problem.
Attn: Accu-Cal Plus Service Department 2900 E. White Star Avenue Anaheim, CA 92806 U.S.A.
Page 20
© 2013 3D Instruments, LP
http://www.3dinstruments.com
The“Direct Drive Difference” in Digital
2900 E. White Star Avenue
Anaheim, CA 92806
Ph: 714•399•9200
E-mail: info@3dinstruments.com
Web Site
Loading...