3B SCIENTIFIC PHYSICS AM51 Instruction Sheet

3B SCIENTIFIC
®
PHYSICS
1
Analogue Multimeter AM51 1003074
Instruction sheet
1 Rotary switch to select measurement range 2 Rotary knob to set full-scale deflection for 0 3 Connection for voltage measurement 4 Connection for current and resistance measurement 5 Connection for ground 6 Adjustment screw for mechanical zero-point setting 7 Nub to latch the housing shut
1. Safety instructions
Before using the analogue multimeter, make sure you read the operating instructions carefully and that you comply with them completely.
The safety of the multimeter and the person operating it can only be guaranteed if it is used in accordance with the instructions. Do not operate or handle this unit incorrectly or inappropriately.
The device may only be used by persons, who are aware of the hazards of contact (for voltages over 30 V rms) and can undertake the appropriate safety precautions. This also includes the appearance of unforeseen volt­ages e.g. in defective units or charged ca­pacitors.
This multimeter is not a toy and must not fall into the hands of children.
Do not place it, keep it or use it anywhere within reach of children.
When the multimeter is used by teenagers, trainees etc., a suitable person should super­vise to ensure the equipment is used safely.
If measurements are made where there are any risks of coming into contact with elec­tricity, a second person is to be informed.
Be particularly careful when measuring volt­ages in excess of 33 V AC (RMS) or 70 V DC.
For voltages which are in excess of 33 V AC or 70 V DC, only use safety measurement leads conforming at least to CAT II.
In the case of voltage and current measure­ments the nominal voltage between the phase and neutral conductor may not exceed 300 V according to CAT II (in circuits that are directly connected to the mains) and CAT III (in building wiring installations) 300 V.
The analogue multimeter may not be used for measurements in circuits with corona discharge (high voltage).
In measurements involving RF circuits spe­cial care must be taken due to the existence of dangerous hybrid voltages.
The appropriate permissible measurement range may not be exceeded. Always change from a higher measurement range to a lower measurement range.
Before using the device, check the housing and the measurement cables for any damage.
The multimeter may only be used in a dry, dust-free environment with no risk of explo­sions occurring. Workplace, hands, shoes and floor must be dry.
2
Before opening the housing all measure­ment leads are disconnected from the de­vice.
Do not dispose of the battery in the regular household garbage. Follow the local regula­tions.
1.1 Symbol legend
Hazard, read instruction sheet
V Voltage A Current
Moving coil galvanometer
AC/DC quantities accuracy class 2,5
Use in horizontal position
DC quantities
AC quantities
CAT Measuring category as per IEC EN
61010-1
Double insulated casing
CE EU conformity mark
Earth symbol
Ground symbol
Battery symbol
2. Description
Handheld meter used to measure current, volt­age and resistance as well as gain levels or attenuation e.g. in chains of quadripoles.
Passive analog multimeter with a rotary switch for selecting the measurement ranges and a scale with mirrored background for parallaxfree readings. The device is extremely robust in terms of load capacity and is equipped with ex­cellent overload protection due to its two anti­parallel diodes as well as moving coil move­ment, which is not sensitive to external electro­magnetic fields. The safety connection sockets offer protection against accidental touch contact. The robust plastic housing and the spring­connected bearing jewels of the movement guarantee protection against damage caused by mechanical stress.
3. Technical data
Measurement ranges: Voltage measurement:
DC
Measurement range Internal resistance
100 mV 2 k
1 V 20 k 10 V 200 k 30 V 600 k
100 V 2 M 300 V 6 M 600 V 12 M
AC
Measurement range Internal resistance
10 V 66.7 k 30 V 200 k
100 V 667 k 300 V 2 M 600 V 4 M
Current measurement:
DC
Measurement range Voltage drop
50 µA 100 mV
1 mA 500 mV
10 mA 500 mV
100 mA 500 mV
1 A 590 mV
AC
Measurement range Voltage drop
3 mA 1.5 V
30 mA 1.6 V
300 mA 1.6 V
3 A 1.8 V
Widerstandsmessung:
Rotary
switch
Measurement range
and scale centre
Max. meas.
current
x 1 1 ...35 k...5 k 45 mA
x 10
10 ...350 k...50
k
4.5 mA
x 100
100 ...3.5 k...500
k
0.45 mA
Accuracy:: Class 2.5 Influencing variables and nominal operating ranges: Temperature 0 – 40° C: ± 1% / 10 K for DC
± 2.5% / 10 K for 100 mV/50 µA DC ± 1.5% / 10 K for AC Frequency (30 Hz...1 kHz): ± 2.5%