AMX−4+ Ratings and Specifications
(Model 2169360, 2236420 & 2275938 Series)
Copyrighte 1996, 1997, 1999, 2000, 2003 and 2004 By General Electric Co.
All Rights Reserved
AMX−4+ RATINGS & SPECIFICATIONS
GE HEALTHCARE(MODEL 2169360, 2236420 & 2275938 SERIES)
REV 8DIRECTION 2173221−100
THIS PAGE INTENTIONALLY LEFT BLANK.
ii
AMX−4+ RATINGS & SPECIFICATIONS
GE HEALTHCARE(MODEL 2169360, 2236420 & 2275938 SERIES)
REV 8DIRECTION 2173221−100
iii
AMX−4+ RATINGS & SPECIFICATIONS
GE HEALTHCARE(MODEL 2169360, 2236420 & 2275938 SERIES)
REV 8DIRECTION 2173221−100
iv
AMX−4+ RATINGS & SPECIFICATIONS
GE HEALTHCARE(MODEL 2169360, 2236420 & 2275938 SERIES)
REV 8DIRECTION 2173221−100
v
AMX−4+ RATINGS & SPECIFICATIONS
GE HEALTHCARE(MODEL 2169360, 2236420 & 2275938 SERIES)
REV 8DIRECTION 2173221−100
vi
AMX−4+ RATINGS & SPECIFICATIONS
GE HEALTHCARE(MODEL 2169360, 2236420 & 2275938 SERIES)
REV 8DIRECTION 2173221−100
vii
AMX−4+ RATINGS & SPECIFICATIONS
GE HEALTHCARE(MODEL 2169360, 2236420 & 2275938 SERIES)
REV 8DIRECTION 2173221−100
viii
AMX−4+ RATINGS & SPECIFICATIONS
GE HEALTHCARE(MODEL 2169360, 2236420 & 2275938 SERIES)
REV 8DIRECTION 2173221−100
ix
AMX−4+ RATINGS & SPECIFICATIONS
GE HEALTHCARE(MODEL 2169360, 2236420 & 2275938 SERIES)
REV 8DIRECTION 2173221−100
x
AMX−4+ RATINGS & SPECIFICATIONS
GE HEALTHCARE(MODEL 2169360, 2236420 & 2275938 SERIES)
REV 8DIRECTION 2173221−100
xi
AMX−4+ RATINGS & SPECIFICATIONS
GE HEALTHCARE(MODEL 2169360, 2236420 & 2275938 SERIES)
REV 8DIRECTION 2173221−100
THIS PAGE INTENTIONALLY LEFT BLANK.
xii
AMX−4+ RATINGS & SPECIFICATIONS
GE HEALTHCARE(MODEL 2169360, 2236420 & 2275938 SERIES)
REV 8DIRECTION 2173221−100
IMPORTANT! . . . X-RAY PROTECTION
X-ray equipment if not properly used may cause injury. Accordingly, the instructions herein contained
should be thoroughly read and understood by everyone who will use the equipment before you attempt to
place this equipment in operation. The General Electric Company, Medical Systems Group, will be glad to
assist and cooperate in placing this equipment in use.
Although this apparatus incorporates a high degree of protection against x-radiation other than the useful beam, no
practical design of equipment can provide complete protection. Nor can any practical design compel the operator to
take adequate precautions to prevent the possibility of any persons carelessly exposing themselves or others to
radiation.
It is important that everyone having anything to do with x-radiation be properly trained and fully acquainted with the
recommendations of the National Council on Radiation Protection and Measurements as published in NCRP Reports
available from NCRP Publications, 7910 Woodmont Avenue, Room 1016, Bethesda, Maryland 20814, and of the
International Commission on Radiation Protection, and take adequate steps to protect against injury.
The equipment is sold with the understanding that the General Electric Company, Medical Systems Group, its agents,
and representatives have no responsibility for injury or damage which may result from improper use of the equipment.
Various protective material and devices are available. It is urged that such materials or devices be used.
CAUTION: United States Federal law restricts this device to use by or on the order of a physician.
CERTIFIED ELECTRICAL CONTRACTOR STATEMENT
All electrical installations that are preliminary to positioning of the equipment at the site prepared for the equipment shall be
performed by licensed electrical contractors. In addition, electrical feeds into the Power Distribution Unit shall be performed
by licensed electrical contractors. Other connections between pieces of electrical equipment, calibrations, and testing shall
be performed by qualified GE Medical personnel. The products involved (and the accompanying electrical installations) are
highly sophisticated, and special engineering competence is required. In performing all electrical work on these products,
GE will use its own specially trained field engineers. All of GE’s electrical work on these products will comply with the
requirements of the applicable electrical codes.
The purchaser of GE equipment shall only utilize qualified personnel (i.e., GE’s field engineers, personnel of third-party
service companies with equivalent training, or licensed electricians) to perform electrical servicing on the equipment.
DAMAGE IN TRANSPORTATION
All packages should be closely examined at time of delivery. If damage is apparent, have notation “Damage in
Shipment” written on all copies of the freight or express bill before delivery is accepted or “signed for” by a
General Electric representative or a hospital receiving agent. Whether noted or concealed, damage MUST be
reported to the carrier immediately upon discovery, or in any event, within 14 days after receipt, and the contents
and containers held for inspection by the carrier. A transportation company will not pay a claim for damage if an
inspection is not requested within this 14−day period.
To file a report:
−Call 1−800−548−3366 and use option 8.
−Fill out a report on http://egems.med.ge.com/edq/home.jsp
−Contact your local service coordinator for more information on this process
Rev. June 13, 2006
xiii
AMX−4+ RATINGS & SPECIFICATIONS
GE HEALTHCARE(MODEL 2169360, 2236420 & 2275938 SERIES)
REV 8DIRECTION 2173221−100
THIS PAGE INTENTIONALLY LEFT BLANK.
xiv
AMX−4+ RATINGS & SPECIFICATIONS
GE HEALTHCARE(MODEL 2169360, 2236420 & 2275938 SERIES)
(Section 2−9−1).
4Apr. 12, 1999Added AMX−4+ Model 2236420.
5Nov. 8, 2000Added AMX−4+ Model 2275938.
6Nov. 20, 2003Revised moving efforts in section 2−5.
7Aug. 20, 2006Added new collimator 46−270615P3 (Medys) reference in Section 2−13.
8May 09, 2008Updated Section 2−12 for CFR 1020.30/31/32 regulatory compliance.
Updated for international compliance.
Updated model numbers in Table 1−2.
Added model numbers to note in Table 1−2.
Added model numbers to Section 2−2−1, step 1.
Added model numbers to Section 2−3−1, step 2, 2nd bullet.
Added model numbers to Section 2−4, step 1a and 1c.
xvii
AMX−4+ RATINGS & SPECIFICATIONS
GE HEALTHCARE(MODEL 2169360, 2236420 & 2275938 SERIES)
REV 8DIRECTION 2173221−100
THIS PAGE INTENTIONALLY LEFT BLANK.
xviii
AMX−4+ RATINGS & SPECIFICATIONS
GE HEALTHCARE(MODEL 2169360, 2236420 & 2275938 SERIES)
REV 8DIRECTION 2173221−100
SECTION 1
INTRODUCTION
1-1Identification and Compatibility
See Illustration 1−1. The AMX−4+ is identified by Model Number on the rating plate located on the top
cover. Model part and catalog numbers are identified in Table 1−1.
Japan Short Column2169360−6A0659D2236420−6 & 2275938−6A0659JE
240V Short Column2275938−12A0659JM
240V Short Column, AEC2275938−13A0659JN
240V Short Column, Tech Switch2275938−14A0659JP
240V Short Column, AEC, Tech Switch2275938−15A0659JR
IEC, Tech Switch − Chinese2275938−16A0659JS
IEC, AEC, Tech Switch − Chinese2275938−17A0659JW
Catalog
Number
Part Number
Catalog
Number
1
AMX−4+ RATINGS & SPECIFICATIONS
GE HEALTHCARE(MODEL 2169360, 2236420 & 2275938 SERIES)
REV 8DIRECTION 2173221−100
ILLUSTRATION 1−1
AMX−4+ IDENTIFICATION
RATING
PLATE
The applicable components making up the AMX−4+ are identified with the nameplate statement, “this
product conforms to all applicable standards under 21 CFR part 1020,” or “complies with radiation
performance standards, 21 CFR sub- chapter J.”
TABLE 1−2
AMX−4+ COMPONENTS
ComponentModel No.Nameplate Location
HV Transformer46−270954G2Duplicate plate on front of top cover above circuit
breaker
HRT X−ray Tube Housing46−155750G8Rear trim cover
Maxiray 75 TH 111 X−ray
Tube Housing
Collimator46−270615P3Right side of Collimator
Mobil−AID D1300 Automatic
Exposure Control*
*Note:Mobil−AID is present only on AMX−4+ Models 2169360−2, −4, −8, −10, 2236420−2, −4, −8, −10 and 2275938−2,
−4, −8, −10, −13, −15, −17.
2222709Rear trim cover
46−279732P4 (A8413BA)Duplicate plate on front of top cover above AEC
paddle
1-2General
The AMX−4+ contains operating safeguards to provide maximum safety. Before calling for service, be
certain proper operating procedures are being used. Refer to Direction 2166913−1EN, AMX−4+Operation for proper operating procedures.
Satisfactory equipment performance requires the use of service personnel specially trained on X−ray
apparatus. GE Healthcare is responsible for the effects on safety, reliability, and performance only if the
following conditions are met:
−The electrical wiring of the relevant rooms complies with all national and local codes.
2
AMX−4+ RATINGS & SPECIFICATIONS
GE HEALTHCARE(MODEL 2169360, 2236420 & 2275938 SERIES)
REV 8DIRECTION 2173221−100
−All assembly operations, extensions, re-adjustments, modifications, or repairs are carried out by GE
Healthcare, authorized service representatives.
−The equipment is used in accordance with the instructions for use. Refer to Direction
2166913−1EN, AMX−4+ Operating Manual for proper operating procedures.
AMX−4+ contains high voltage circuits for generating X−rays. Only trained
and qualified personnel should be permitted access to the internal parts
of this equipment.
1-3Inspection
In order to assure continued performance of this X−ray equipment, a Periodic Maintenance program must
be established in accordance with 2173227−100 AMX−4+ Periodic Maintenance. It is the owner’s
responsibility to supply or arrange for this service.
1-4HHS Testing
The United States Department of Health and Human Services (HHS) has established performance
requirements for diagnostic X−ray equipment. These requirements are defined in Title 21 of the Code of
Federal Regulations, and apply only to certain specified components identified as “certified equipment.”
The manufacturer of specified diagnostic X−ray components must certify that the components:
1. Perform as required by the HHS standard when installed, adjusted, and tested as specified in the
manufacturer’s instructions to the assembler.
Refer to Direction 2173222−100 AMX−4+ Installation for HHS compliance testing as it pertains to the
AMX−4+ during installation.
2. Will continue to comply when maintained in accordance with the manufacturer’s instructions.
Refer to Direction 2173227−100 AMX−4+ Periodic Maintenance for periodic HHS compliance testing
for the AMX−4+.
3
AMX−4+ RATINGS & SPECIFICATIONS
GE HEALTHCARE(MODEL 2169360, 2236420 & 2275938 SERIES)
REV 8DIRECTION 2173221−100
SECTION 2
SPECIFICATIONS
2-1Accuracy Criteria
Specified accuracies are subject to additional errors for accuracy of instrumentation used during
calibration and measurement. To achieve generator accuracies, the following criteria must be met:
−The WAIT message must not appear on the Message Display when taking an exposure.
−The READY FOR EXPOSURE message must appear on the Message Display.
General Electric Healthcare reserves the right to change specifications at any time without incurring any
obligation to incorporate new features in products previously sold.
2-2Physical Characteristics
2-2-1Dimensions (All measures are nominal)
1. Height: 70 inches (1778 mm) for Models 2169360−6, 2236420−6, and 2275938−6, −12, −13, −14,
−15; all others 76 inches (1930 mm).
2. Width: 25−3/16 inches (640 mm)
3. Length: 45−3/8 inches (1153 mm)
4. Weight: 1080 pounds (490 kg)
2-2-2Environmental Limits
1. Operating temperature range: 59 to 100 Degrees Fahrenheit (15 to 38 Degrees Celsius) at 80%
non−condensing humidity.
2. Storage temperature range −40 to +140 Degrees Fahrenheit (−40 to +60 Degrees Celsius)
3. Maximum operating altitude: 8,000 feet (2440 meters).
2-3Battery
2-3-1Specifications
1. Nine 12.9 volt batteries connected in series provide approximately 116 volts at full charge.
Voltage must be maintained to +/−10% of the nominal value during charging.
4
AMX−4+ RATINGS & SPECIFICATIONS
GE HEALTHCARE(MODEL 2169360, 2236420 & 2275938 SERIES)
REV 8DIRECTION 2173221−100
2-3-2Battery Capacity
The AMX−4+ battery capacity can be measured by one of the following five methods. All capacities are
measured after the AMX−4+ has been charged to the “CHARGE COMPLETE” state. Available capacity
as stated applies only to new battery sets free of defective cells. Capacity may decrease as the battery
nears the end of its useful life.
method 1: exams
The AMX−4+ batteries will provide capacity for more than 20 typical EXAMs. An “EXAM” is defined as:
−Two 70 kVp, 10 mAs X−ray exposures including:
>7 seconds of prep (rotor and filament drive)
>25 seconds of field light
−5 minutes of drive time
−9 minutes of idle time
The formula in the “VARIED USAGE” section can be used to determine the number of total EXAMs
available for usage regimes different from this typical case.
method 2: X−ray Exposures
The AMX−4+ batteries will provide enough capacity for 165 or more 100 kVp, 100 mAs X−ray exposures.
Each exposure includes 4 seconds of prep (rotor and filament drive) time and 30 seconds of idle time for
battery recovery. This number may be reduced by additional idle time required for X−ray tube cooling.
method 3: drive time
The AMX−4+ batteries will provide enough capacity for 140 minutes of continuous drive time. This time
is typically independent of driving conditions, however, it may be reduced if a significant portion of the
drive is on carpeting or up ramps.
method 4: idle time
The AMX−4+ batteries will provide capacity for 23.3 hours of continuous idle time. “Idle” is the time when
the AMX−4+ is ON but not being used.
method 5: Varied usage
For varied usage, the AMX−4+ batteries will provide capacity according to the following formula:
{ (idle time in minutes ) 3} +
{ (drive time in minutes ) 30} +
{ ( field light time in minutes ) 25 } +
{ ( prep time in minutes ) 30} +
{ ( exposure energy* ) 2.17} = 4200
*exposure energy = cumulative { ( kVp mAs ) 1000 }
EXAMPLE: Assume one desires to estimate the number exams available from an AMX−4+ used in a
particular pediatric ward. It is determined that a typical exam for this case is comprised of:
−Two 70 kVp, 0.8 mAs X−ray exposures including:
>3 seconds of prep
>15 seconds of field light
5
AMX−4+ RATINGS & SPECIFICATIONS
GE HEALTHCARE(MODEL 2169360, 2236420 & 2275938 SERIES)
REV 8DIRECTION 2173221−100
−1 minute of drive
−5 minutes of idle
Using the above formula we can estimate the number of exams as follows:
c.Tube Column rotation: 25 to 40 pound−feet (34 to 54 Newton−meters).
2-6Drive Speed
There are two speeds: Drive Speed with the horizontal Tube Arm secured for transport, and Maneuvering
Speed with the horizontal Tube Arm removed from the Transport Latch.
Drive speed is measured on a smooth, hard and level surface. Speed will be reduced by inclines,
carpeted or soft surfaces.
1. Drive Speed is 264 feet (6705 mm) per minute +/− 25%;
2. Maneuvering Speed is 30% to 60% of drive speed.
2-7Tube Unit Ratings
Tube Housing and Insert Specifications are given in Direction 46−017226, Tube Ratings, HRT 50 and 60
Hz., or in 2236721−100, Product Data Sheet Maxiray 75 TH 11 X−ray Tube. Also, refer to note on
maximum allowable kVp and mAs ratings in 2166913−1EN AMX−4+ Operating Manual.
2-8Generator Operator Indicators
Check for proper operation of tones or buzzers and labels as required by regulations. Reference
“Generator Operator Indicators” in Tab 3 of Direction 46−013894, System Field Test for HHS.
1. X−ray on indicator lights during an exposure.
2. Audible tone occurs during exposure.
3. Audible tone occurs with safety timed termination of AEC exposures.
4. Safety timed termination of AEC exposures requires resetting before taking another exposure.
5. X−ray warning label is legible.
7
AMX−4+ RATINGS & SPECIFICATIONS
GE HEALTHCARE(MODEL 2169360, 2236420 & 2275938 SERIES)
REV 8DIRECTION 2173221−100
2-9kVp Accuracy
1. Rise time of the kVp wave form from 10% to 90% of the maximum kVp is 1.2 millisecond or less.
2. Fall time of the kVp wave form from 90% of the maximum kV to 20 kV is 2.5 milliseconds or less.
3. Accuracy of the kVp wave form to selected kVp is +/− 8% of the value displayed on the operator
panel for the first 20 ms and +/− 5% after 20 ms. Accuracy applies within the range of the bar graph
battery charge indicator.
Note:These specifications do not apply to switching transients which occur during the first millisecond and
the last millisecond of an exposure.
2-9-1Test Method
Preferred
Use a Keithley Non−Invasive kVp Divider (Model 35080A with Deviation 535 or Model 35080B, both using
Mobile Filter Pack Plus 37946C and optional Low Range Filter Pack 38237C). A Triad 35050A Dosimeter
can also be used to provide digital readout of corrected kVp values. No other substitutions for
non−invasive kVp Dividers are approved!
The set−up procedure and linear correction curves (non−Triad systems) for using the Keithley divider are
covered in Section 3 of Direction 46−017561 HHS Control and Tube Assembly Tests and Keithley’s
Operation and Maintenance manuals.
Alternate
Use the GE C1515A Invasive Bleeder. If this method is used, the unit should be calibrated and verified
with this meter.
Note:If an attempt is made to verify a unit calibrated with a C1515A bleeder with either of the above Keithley
Non−Invasive dividers, kVp readings will read 5−7 kVp higher than when read with the C1515A. This
is due to impedance changes in the high voltage circuit with the bleeder removed from the circuit and
due to frequency compensation errors present using the C1515A divider with the AMX−4+ waveform.
The procedure for connecting the C1515A High Voltage divider is covered in Direction 46−013288Bleeder, High−Voltage Dual Type T8005G and C1515A Connection ... Applications, and Direction
2196272−100, High Voltage Cable Installation and Troubleshooting Procedures.
Make exposures and calculations as described in “Technique Accuracy − kV/mA” in Tab 3 of Direction
46−013894 System Field Test for HHS.
2-10mAs Metering Accuracy
Accuracy of mAs is the integral of X−ray tube emission current between the time at which the kV wave
form reaches 75% of the indicated peak value at the beginning of an exposure and the time at which it
falls to 75% of the indicated peak value when the exposure is terminated.
Actual mAs shall match selected mAs within +/−10%.
8
AMX−4+ RATINGS & SPECIFICATIONS
GE HEALTHCARE(MODEL 2169360, 2236420 & 2275938 SERIES)
REV 8DIRECTION 2173221−100
2-10-1 Test Method
This is an indirect procedure which verifies accuracy of the mAs metering circuitry and mAs calibration.
Measuring mAs Metering Accuracy is done by injecting 100 mA into the mAs integrating circuit and
comparing the response with a meter installed in the circuit. Reference Direction 2173223−100, AMX−4+Calibration− familiarity with this direction is assumed. Also reference “Technique Accuracy − mAs” in Tab
3 of Direction 46−013894, System Field Test for HHS.
1. Enter mAs calibration and install meter.
2. At the prompt ENTER VALUE compare meter reading with displayed reading.
3. Correct reading to include meter accuracy before comparing with requirements.
Note:Accuracy of mA reading at the approximate 100 mA test condition is 0.1 mA. Meter accuracy must
be added to the mA reading before making judgment on the final reading.
2-11Reproducibility
The coefficient of variation for radiation output is less than 0.045 for successive exposures having
constant technique factors.
Coefficient of variation is measured and calculated as described in “Reproducibility of Exposure” in Tab 3
of Direction 46−013894, System Field Test for HHS. This applies to all units for non−AEC mode −
reference the procedure “for exposures made without the use of A.E.C.” For units equipped with
Mobil−AID, also reference the procedure “for exposures made with an A.E.C.”
2-12Beam Quality
The half−value layer of the useful beam at 80 kVp shall not be less than 2.5 millimeters of aluminum. This
requirement differs slightly from and supersedes NCRP 33. The specific test point is at 80 kVp requiring a
half value layer of 2.5 millimeters of aluminum. This applies to systems manufactured before June 10th,
2006.
For systems manufactured on or after June 10th, 2006 the following new regulation applies; The
half−value layer of the useful beam at 80 kVp shall not be less than 2.9 millimeters of aluminum. The
specific test point is at 80 kVp requiring a half value layer of 2.9 millimeters of aluminum.
Reference the System Rating Plate to determine manufactured date. Note: The rating plate shows month
and year only. For systems manufactured during June, 2006 you must verify the manufactured date in the
GE GIB database. Contact your local Service Representative for assistance.
2-12-1 Test Method
The procedure for measuring Beam Quality is presented in “Beam Quality (Half Value Layer)” in Tab 4 of
Direction 46−013894 System Field Test for HHS. The following noted exceptions to that procedure apply
for the AMX−4+:
1.Select initial technique factors of 80 kVp and 20 to 48 mAs.
2.When making an exposure without the absorber, adjust mAs so that the radiation meter reading
contains three significant digits.
3.Use an AMX−4 absorber 46−173632G2 in the collimator rails.
9
AMX−4+ RATINGS & SPECIFICATIONS
GE HEALTHCARE(MODEL 2169360, 2236420 & 2275938 SERIES)
REV 8DIRECTION 2173221−100
>For systems manufactured On or After June 10th, 2006 add additional aluminum filtration from
the HVL Attenuator set part # 46−194427P274 found in the HHS Kit part # 46−303879G1 or
46−303879G2. Total test filtration = 3.1 mm of Aluminum. Affix these plates to the AMX−4
absorber with tape. Ensure the tape is not in the usable beam path.
>For tube changes there is an additional 1.2 mm Aluminum cup added to the X−Ray Tube. This
cup snaps onto the collimator mounting plate and is easily identified since the tube glass port is
not visible. This cup must be present and retained for systems manufactured On or After June
10th, 2006. Likewise this cup if provided with the replacement tube it must be removed for
systems manufactured before June 10th, 2006. Reference the system rating plate affixed to the
AMX unit.
2-13Collimator Function
A manual collimator, model 46−270615P2 (Micro Medical Systems) or model 46−270615P3 (Medys),
bearing an HHS certification label is installed on the AMX−4+. Reference “Functional Test − Manual Rad
Collimator Version” in Tab 6 of Direction 46−013894, System Field Test for HHS.
1. Minimum source to skin distance is limited to more than 30 centimeters by the skin spacers at the
beam exit of the collimator.
2. Full 17 by 17 inch (43 by 43 centimeters) radiographic coverage at 40 inch (1.02 meter) Source to
Image Distance.
3. Minimum inherent filtration of 2.0 mm aluminum equivalent at 100 kVp.
2-14Collimator Alignment and SID
Collimator Alignment and SID tests are performed in accordance with Tab 6 of Direction 46−013894,System Field Test for HHS, reference:
SID Test
Light to X−ray Field Test
Center to Center Test
Field Indicator (Pointers to Actual Size)
1. The X−ray field size must agree with the indicated field size within 1.8% of SID.
2. Total misalignment of parallel edges of the light field with the edges of its X−ray field must not
exceed 1.7% of SID.
3. The difference between the indicated SID and the actual SID must not exceed 1.8% of the indicated
SID.
2-15Collimator Light Field Intensity
The average illumination at a distance of 100 cm (39.37 inch) from the focal spot shall be 16 foot candles
(170 lux) or more. Reference “Collimator Light Field Intensity” in Tab 6 of Direction 46−013894, System
Field Test for HHS.
10
AMX−4+ RATINGS & SPECIFICATIONS
GE HEALTHCARE(MODEL 2169360, 2236420 & 2275938 SERIES)
REV 8DIRECTION 2173221−100
2-16AEC Minimum Exposure Time
(Applies to units equipped with Mobil−AID)
When the X−ray kVp is equal to or greater than 50 kVp, the minimum exposure time will be equal to or
greater than the time interval required to deliver 5 mAs. Reference “AEC Minimum Exposure Time” in Tab
3 of Direction 46−013894, System Field Test for HHS.
11
AMX−4+ RATINGS & SPECIFICATIONS
GE HEALTHCARE(MODEL 2169360, 2236420 & 2275938 SERIES)
REV 8DIRECTION 2173221−100
SECTION 3
HHS TEST CHECKLISTS
The importance of HHS compliance testing cannot be over emphasized. Before starting these
procedures, review Direction 46−013894 System Field Test for HHS. Pay particular attention to
“Introduction” (Tab 1 of Direction 46−013894).
With the exception of Collimator Function and Beam Quality, all field tests for HHS compliance are to be
performed at specified intervals.
3-1HHS Tests (Excluding AEC)
Requirements are pass or fail. Service is required if any Requirement is failed. After each Requirement is
verified, place a check in its box. Record notes and comments next to the Requirement.
InspectRequirementInspector’s Notes
3-1-1Generator Operator
Indicators
Required during Installation,
preventative maintenance
calls, or when replacing
major components.
3-1-2Technique
Accuracy kVp/mA
Required during Installation,
preventative maintenance calls,
and repair.
3-1-3Indirect Linearity
Required during Installation,
preventative maintenance
calls, and repair.
3-1-4Technique
Accuracy − mAs
Required during Installation,
preventative maintenance
calls, and repair.
[ ] Check for proper operation of tones and
buzzers as required by regulations. Refer
to “Generator Operator Indicators” in
Section 2 of Direction 2173221−100,
AMX−4+ Ratings And Specifications, and
perform defined inspection.
[ ] Actual kVp matches select kVp. Refer to
“kV Accuracy” in Section 2 of Direction
2173221−100, AMX−4+ Ratings &Specifications, and perform procedure.
[ ] Refer to “Indirect Linearity” in Tab 3 of
Direction 46−013894, System Field Test for HHS.
[ ] Refer to “mAs Metering Accuracy” in
Section 2 of Direction 2173221−100,
AMX−4+ Ratings & Specifications, and
perform defined procedure.
3-1-5Reproducibility Of
Exposures −
Non−AEC
Required during Installation,
preventative maintenance
calls, and repair.
[ ] Refer to “Reproducibility” in Section 2 of
Direction 2173221−100, AMX−4+ Ratings &Specifications, and perform defined
procedure for non−AEC mode exposures.
12
AMX−4+ RATINGS & SPECIFICATIONS
GE HEALTHCARE(MODEL 2169360, 2236420 & 2275938 SERIES)
REV 8DIRECTION 2173221−100
InspectInspector’s NotesRequirement
3-1-6Beam Quality
(Half Value Layer)
Required during Installation,
or after replacement of tube
unit, collimator, or other
absorption between source
and patient.
3-1-7Collimator
Function
Required during Installation,
or after replacement of
collimator.
3-1-8Collimator
Alignment and SID
Required during Installation,
preventative maintenance
calls, and repair.
3-1-9Collimator Light
Field Intensity
Required during Installation,
preventative maintenance
calls, and repair.
[ ] Refer to “Beam Quality” in Section 2 of
Direction 2173221−100, AMX−4+ Ratings &Specifications, and perform defined
procedure.
[ ] Refer to “Collimator Function” in Section 2
of Direction 2173221−100, AMX−4+Ratings & Specifications, and perform
defined checks.
[ ] Refer to “Collimator Alignment and SID” in
Section 2 of Direction 2173221−100,
AMX−4+ Ratings & Specifications, and
perform defined procedures.
[ ] Refer to “Collimator Light Field Intensity” in
Section 2 of Direction 2173221−100,
AMX−4+ Ratings & Specifications, and
perform defined procedure.
3-2AEC HHS Tests
(Units with Mobil−AID)
Requirements are pass or fail. Service is required if any Requirement is failed. After each Requirement is
verified, place a check in its box. Record notes and comments next to the Requirement.
InspectRequirementInspector’s Notes
3-2-1Reproducibility Of
Exposures − AEC
Required during Installation,
preventative maintenance
calls, and repair.
3-2-2AEC Minimum
Exposure Time
Required during Installation,
preventative maintenance
calls, and repair.
[ ] Refer to “Reproducibility” in Section 2 of
Direction 2173221−100, AMX−4+ Ratings &Specifications, and perform defined
procedure for AEC mode exposures.
[ ] Refer to “AEC Minimum Exposure Time” in
Section 2 of Direction 2173221−100,
AMX−4+ Ratings & Specifications, and
perform defined procedure.
13
AMX−4+ RATINGS & SPECIFICATIONS
GE HEALTHCARE(MODEL 2169360, 2236420 & 2275938 SERIES)
REV 8DIRECTION 2173221−100
THIS PAGE INTENTIONALLY LEFT BLANK.
14
GE Healthcare
GE Healthcare: Telex 3797371
P.O. Box 414, Milwaukee, Wisconsin 53201 U.S.A.
(Asia, Pacific, Latin America, North America)
GE Healthcare — Europe: Telex 698626
283, rue de la Miniére, B.P. 34, 78533 Buc Cedex
France
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.