ZOLL X Series User Manual

0 (0)
ZOLL X Series User Manual

SurePowerTM II Battery Pack Guide

"

70

70

9650-000840-01 Rev. A

The issue date for the SurePower™ II Battery Pack Guide (REF 9650-000840-01 Rev. A) is January, 2012.

If more than three years have elapsed since the issue date, contact ZOLL Medical Corporation to determine if additional product information updates are available.

SurePower, X Series, and ZOLL are trademarks or registered trademarks of ZOLL Medical Corporation in the United States and/or other countries.

All other trademarks are the property of their respective owners. Copyright © 2012 ZOLL Medical Corporation. All rights reserved.

ZOLL Medical Corporation

269 Mill Road

Chelmsford, MA, 01824

USA

Tel.: +1 978-421-9655

Fax: +1 978-421-0001

ZOLL International Holding B.V.

Newtonweg 18

6662 PV ELST

Nederland

Tel.: +31 (0) 481 366410

Fax: +31 (0) 481 366411

Introduction

The SurePowerTM II Battery Pack is a rechargable, sealed lithium ion battery for use in compatible ZOLL ® defibrillator/monitors and chargers.

It is important that you read and understand the information in this manual. The proper use of the SurePower II Battery Pack is critical to the reliable operation of your ZOLL defibrillator/ monitor.

Conventions

This manual uses the following conventions:

 

Warning statements describe conditions or actions that can result in personal injury or

 

death.

 

 

 

 

Caution

Caution statements describe conditions or actions that can result in damage to the SurePower II

 

Battery Pack.

 

Note: Notes contain additional information on using and maintaining the SurePower II

 

Battery Pack.

 

The manual displays button and LED labels in bold face type.

Safety

 

 

The following sections provide important warnings, cautions, and notes regarding the ZOLL

 

SurePower II Battery Pack.

Warnings

Locate ZOLL defibrillator/monitor units where they cannot harm the patient or medical personnel if they fall -- accidental sliding or drop of the SurePower II Battery Pack may cause physical injury. Damaged batteries may expose users to a safety hazard. Should the SurePower II Battery Pack be dropped, inspect and test the unit prior to further use following the procedures described in the Maintenance section of this Guide.

Avoid prolonged exposure of the SurePower II Battery Pack to direct sunlight to minimize heating and damage that may occur from exposure to ultraviolet light.

SurePower II Battery Packs have been tested against interference from radio frequency emissions typical of two-way radios used in emergency service/public safety activities. Refer to the appropriate ZOLL defibrillator/monitor operator’s guide to determine recommended operating distances from RF transmission equipment.

Do not use SurePower II Battery Packs unless they have been visually inspected, tested, and recalibrated by appropriate personnel and you have been adequately trained in the proper use of these batteries. See the Maintenance section of this guide for more information.

Do not disassemble the SurePower II Battery Pack.

Charge the SurePower II Battery Pack in a well-ventilated area.

You must perform periodic inspection, capacity testing, and functional testing of rechargable defibrillator batteries to ensure the availability of safe and reliable batteries. See the Maintenance section of this guide for more information.