ZOLL M Series User Manual

5 (1)
ZOLL M Series User Manual

M Series® Service Manual

9650-0450-01 Rev R

The issue date for the M Series Service Manual (REF 9650-0450-01 Rev. R) is March, 2013.

If more than 3 years have elapsed since the issue date, contact ZOLL Medical Corporation to determine if additional product information updates are available.

Copyright © 2013 ZOLL Medical Corporation. All rights reserved. Rectilinear Biphasic, M Series, and ZOLL are trademarks or registered trademarks of ZOLL Medical Corporation in the United States and/or other countries.

Masimo is a trademark or registered trademark of Masimo Corporation in the United States and/or other countries.

ZOLL Medical Corporation

269 Mill Road Chelmsford, MA USA 01824-4105

ZOLL International Holding B.V.

Newtonweg 18

6662 PV ELST

The Netherlands

0123

9650-0450-01 Rev R

M Series Service Manual

T A B L E O F C O N T E N T S

PREFACE .......................................................................................................................................................................................

 

V

Overview ..............................................................................................................................................................................................

v

Safety Considerations ...........................................................................................................................................................................

v

Additional Reference Material ...........................................................................................................................................................

vi

Conventions .......................................................................................................................................................................................

vii

Service Policy Warranty ....................................................................................................................................................................

vii

Technical Service ..............................................................................................................................................................................

vii

Technical Service for International Customers ................................................................................................................................

viii

CHAPTER 1 MAINTENANCE TESTS ..............................................................................................................................................

1

Overview ..............................................................................................................................................................................................

1

Before You Begin the Maintenance Tests ............................................................................................................................................

2

Equipment that You Need to Perform the M Series Maintenance Tests ..............................................................................................

3

Equipment You Need for the M Series Options Maintenance Tests ...................................................................................................

4

1.0

Physical Inspection of the Unit......................................................................................................................................................

5

2.0

Front Panel Button Test.................................................................................................................................................................

7

3.0

3, 5, and 12 Leads Test................................................................................................................................................................

10

4.0

Power Supply Test (Optional) .....................................................................................................................................................

11

5.0

Leakage Current Test...................................................................................................................................................................

15

6.0

Paddles Test (If applicable) .........................................................................................................................................................

16

7.0

Heart Rate Display Test...............................................................................................................................................................

18

8.0

Calibrating Pulses on Strip Chart Test ........................................................................................................................................

19

9.0

Notch Filter Test..........................................................................................................................................................................

20

10.0 Heart Rate Alarm Test...............................................................................................................................................................

21

11.0

Defibrillator Self Test................................................................................................................................................................

23

12.0

Synchronized Cardioversion Test..............................................................................................................................................

25

13.0 Shock Test .................................................................................................................................................................................

26

14.0 Summary Report Test................................................................................................................................................................

29

15.0 Advisory Message Test .............................................................................................................................................................

30

16.0

Pacer Test ..................................................................................................................................................................................

31

i

M Series Service Manual

17.0

SpO2 Monitor Test for SpO2 Option ........................................................................................................................................

33

18.0

EtCO2 Monitor Test (for EtCO2 Option)..................................................................................................................................

36

19.0

Temperature Test........................................................................................................................................................................

38

20.0

IBP Monitoring Test.................................................................................................................................................................

39

21.0

NIBP Transducer Calibration Test .............................................................................................................................................

44

22.0

NIBP Monitor Test .....................................................................................................................................................................

47

23.0

NIBP Volume Leak Test with Fluke Biomedical NIBP Analyzer.............................................................................................

49

24.0

Bluetooth Test............................................................................................................................................................................

51

CHAPTER 2 TROUBLESHOOTING ................................................................................................................................................

53

Overview ............................................................................................................................................................................................

53

Troubleshooting ..................................................................................................................................................................................

54

Zoll M Series Error Messages ............................................................................................................................................................

58

CHAPTER 3 DISASSEMBLY PROCEDURES ..................................................................................................................................

75

Overview ............................................................................................................................................................................................

75

Required Equipment ...........................................................................................................................................................................

76

Parts That May Need Replacing After Disassembly ..........................................................................................................................

76

Safety Precautions ..............................................................................................................................................................................

77

Overview of Modules .........................................................................................................................................................................

78

1.

Removing the ZIF Keeper ............................................................................................................................................

81

2.

Removing the Front Panel ............................................................................................................................................

82

2A.Removing the Display .................................................................................................................................................

83

2B.Removing the Control Board ......................................................................................................................................

84

3.

Removing the Upper Housing Assembly .....................................................................................................................

86

4.

Removing the System Board Assembly .......................................................................................................................

87

5.

Removing the Battery Interconnect Board Assembly ..................................................................................................

90

6.

Removing the High Voltage/Charger Assembly .........................................................................................................

91

7.

Removing the High Voltage Module Assembly ..........................................................................................................

94

8.

Removing the High Voltage Capacitor Assembly .......................................................................................................

96

9.

Removing the System Interconnect Board ...................................................................................................................

98

10.Removing the Printer/Recorder Motor ......................................................................................................................

100

11.Removing the Lower Housing Assembly ..................................................................................................................

102

12.Removing the Print Head Assembly ..........................................................................................................................

103

13.Removing the PCMCIA Card Slot Assembly ...........................................................................................................

104

ii

M Series Service Manual

 

14.Removing the Paddle Release Latch .........................................................................................................................

105

CHAPTER 4 REPLACEMENT PARTS .........................................................................................................................................

107

Overview ..........................................................................................................................................................................................

107

Replacement Parts ............................................................................................................................................................................

108

Field Replacement Parts ...................................................................................................................................................................

112

CHAPTER 5 FUNCTIONAL DESCRIPTION .................................................................................................................................

115

Overview ..........................................................................................................................................................................................

115

Main System Board ..........................................................................................................................................................................

116

Main System Board Functions .........................................................................................................................................................

118

Power Supply ....................................................................................................................................................................................

120

ECG Front End .................................................................................................................................................................................

121

Multifunction Electrode (MFE)/PADS (System Board and High Voltage Module) .......................................................................

121

CPU and EPU ...................................................................................................................................................................................

121

High Voltage Module .......................................................................................................................................................................

122

Defibrillator Charging and Discharging ...........................................................................................................................................

123

High Voltage Capacitor Monitor ......................................................................................................................................................

124

Pacer/Defibrillator Control Signals ..................................................................................................................................................

125

Internal Discharge Resistor Module .................................................................................................................................................

127

AC/DC Charger Module ...................................................................................................................................................................

127

System Interconnect Module ............................................................................................................................................................

127

Stripchart Recorder ...........................................................................................................................................................................

128

PCMCIA Slots ..................................................................................................................................................................................

128

Front Panel and Controls PWBA ......................................................................................................................................................

128

M Series Options ..............................................................................................................................................................................

128

Isolated Power Supply Module .........................................................................................................................................................

129

12 Lead Option .................................................................................................................................................................................

129

Pulse Oximetry (SpO2) ....................................................................................................................................................................

130

End Tidal Carbon Dioxide (EtCO2) .................................................................................................................................................

130

Biphasic Waveform ..........................................................................................................................................................................

131

APPENDIX ...................................................................................................................................................................................

135

Overview ..........................................................................................................................................................................................

135

Interconnect Diagram for the M Series Monophasic Unit ...............................................................................................................

136

Interconnect Diagram for the M Series Biphasic Unit .....................................................................................................................

137

Interconnect Diagram for the M Series CCT Biphasic Unit ............................................................................................................

138

iii

 

M Series Service Manual

INDEX ..........................................................................................................................................................................................

139

iv

M Series Service Manual

Preface

Overview

ZOLL Medical Corporation’s M Series Service Manual is intended for the service technician whose responsibility is to identify malfunctions and/or make repairs at the subassembly level. The Zoll M Series Service Manual has five main sections and one appendix.

Preface—Contains safety warnings and an overview of the manual’s contents. Be sure to review this section thoroughly before attempting to use or service the M Series unit.

Chapter 1—Maintenance Tests explains how to check the defibrillator’s performance using a series of recommended checkout procedures to be conducted every six months.

Chapter 2—Troubleshooting provides a listing of the procedures and error messages to help the service technician detect faults and repair them.

Chapter 3—Disassembly Procedures describes step-by-step procedures for removing subassemblies from the M Series unit.

Chapter 4—Replacement Parts List displays a complete list of ZOLL part numbers for field replaceable parts available for the M Series unit, allowing the service person to identify and order replacement parts from ZOLL.

Chapter 5—Functional Description provides technical descriptions for the M Series major subassembly modules. Appendix A—M Series Operator’s Manual.

Safety Considerations

The following section describes general warnings and safety considerations for operators and patients. Service technicians should review the safety considerations prior to servicing any equipment and read the manual carefully before attempting to disassemble the unit. Only qualified personnel should service the M Series unit.

Federal (U.S.A.) law restricts this unit for use by or on the order of a physician.

Safety and effectiveness data submitted by ZOLL Medical Corporation to the Food and Drug Administration (FDA) under section 510(K) of the Medical Device Act to obtain approval to market is based upon the use of ZOLL accessories such as disposable electrodes, patient cables and batteries. The use of external pacing/defibrillation

v

M Series Service Manual

 

electrodes and adapter units from sources other than ZOLL is not recommended. ZOLL makes no representations or

 

warranties regarding the performance or effectiveness of its products when used in conjunction with pacing/

 

defibrillation electrodes and adapter units from other sources. If unit failure is attributable to pacing/defibrillation

 

electrodes or adapter units not manufactured by ZOLL, this may void ZOLL's warranty.

 

Only qualified personnel should disassemble the M Series unit.

 

 

 

WARNING!

This unit can generate up to 4500 volts with sufficient current to cause lethal shocks.

 

All persons near the equipment must be warned to “STAND CLEAR” prior to discharging the defibrillator.

 

 

Do not discharge the unit’s internal energy more than three times in one minute or damage to the unit may result.

 

Do not discharge a battery pack except in a Base PowerCharger4x4 or compatible ZOLL Battery Charging/Testing unit.

 

Do not use the M Series in the presence of flammable agents (such as gasoline), oxygen-rich atmospheres, or

 

flammable anesthetics. Using the unit near the site of a gasoline spill may cause an explosion.

 

Do not use the unit near or within puddles of water.

NOTE

The M Series is protected against interference from radio frequency emissions typical of two-way radios and cellular

 

phones (digital and analog) used in emergency service/public safety activities. Users of the M Series should assess the

 

unit’s performance in their typical environment of use for the possibility of radio frequency interference from

 

high-power sources. Radio Frequency Interference (RFI) may be observed as shifts in monitor baseline, trace

 

compression, or transient spikes on the display.

Additional Reference Material

In addition to this guide, there are several other components to the Zoll M Series documentation. They include:

Operator’s Guide - A comprehensive reference work that describes all the user tasks needed to operate the M Series.

Configuration Guide - Describes the M Series features and functions whose operation can be customized by authorized users.

vi

M Series Service Manual

 

Conventions

 

 

 

WARNING!

Warning statements describe conditions or actions that can result in personal injury or death.

 

 

 

 

CAUTION

Caution statements describe conditions or actions that can result in damage to the unit.

 

NOTE

Notes contain additional information on using the defibrillator.

 

Service Policy Warranty

 

In North America: Consult your purchasing agreement for terms and conditions associated with your warranty. Outside

 

of North America, consult ZOLL authorized representative.

 

In order to maintain this warranty, the instructions and procedures contained in this manual must be strictly followed.

 

For additional information, please call the ZOLL Technical Service Department 1-800-348-9011 in North America.

 

Technical Service

 

If the ZOLL M Series unit requires service, contact the ZOLL Technical Service Department:

 

Telephone: 1-978-421-9655; 1-800-348-9011

 

Fax 1-978-421-0010

 

Have the following information available for the Technical Service representative:

 

Unit serial number.

 

Description of the problem.

 

Department where equipment is used.

 

Sample chart recorder strips documenting the problem, if applicable.

 

Purchase Order to allow tracking of loan equipment.

 

Purchase Order for a unit with an expired warranty.

vii

M Series Service Manual

If the unit needs to be sent to ZOLL Medical Corporation, obtain a service order request number from the Technical Service representative. Return the unit in its original container to:

ZOLL Medical Corporation 269 Mill Road

Chelmsford, Massachusetts 01824-4105 Attn: Technical Service Department

Telephone: 1-800-348-9011; 1-978-421-9655 FAX: 978-421-0010

Technical Service for International Customers

International customers should return the unit in its original container to the nearest authorized ZOLL Medical Corporation Service Center. To locate an authorized service center, contact the International Sales Department at ZOLL Medical at the above address.

Units are available on loan while your unit is being repaired.

viii

M Series Service Manual

Chapter 1

Maintenance Tests

Overview

The M Series has two checkout procedures: the operator’s shift checklist and the extensive six-month maintenance tests checkout procedures.

Because the M Series units must be maintained ready for immediate use, it is important for users to conduct the Operator’s Shift Checklist procedure at the beginning of every shift. This procedure can be completed in a few minutes and requires no additional test equipment. (See the ZOLL M Series Operator’s Guide for the Operator’s Shift Checklist.)

A qualified biomedical technician must perform a more thorough maintenance test checkout every six months to ensure that the functions of the M Series unit work properly. This chapter describes the step by step procedures for performing the six month maintenance test checkout. Use the checklist at the back of this document (ZOLL M Series Maintenance Tests Checklist) to record your results of the maintenance tests.

This chapter describes the following M Series maintenance tests:

1.Physical Inspection of the Unit

2.Front Panel Button

3.3, 5, and 12 Leads

4.Power Supply

5.Leakage Current

6.Paddles

7.Heart Rate Display

8.Calibrating Pulses on Strip Chart

9.Notch Filter

10.Heart Rate Alarm

1

M Series Service Manual

11.Defibrillator Self Test

12.Synchronized Cardioversion

13.Shock

14.Summary Report

15.Advisory Message

16.Pacer

17.SpO2 Monitor

18.EtCO2 Monitor

19.Temperature Test

20.IBP Monitoring Test

21.NIBP Transducer Calibration Test

22.NIBP Monitor Test

23.NIBP Volume Leak Test with Fluke Biomedical NIBP Analyzer

24.Bluetooth Test

Before You Begin the Maintenance Tests

Assemble the tools or specialized testing equipment listed in the “Equipment You Need to Perform the Maintenance Tests” section shown below.

Keep an extra fully charged ZOLL M Series compatible battery available.

Schedule an hour to conduct the entire maintenance test.

Photocopy the checklist at the back of this document and use the copy to record your results. As you conduct each step of a procedure, mark the

Pass/Fail/NA check boxes on your checklist and then save it for your maintenance file.

Perform the tests in the order presented.

Perform all the steps of each test procedure.

Complete all the steps of the procedure before evaluating the test results.

2

M Series Service Manual

Equipment that You Need to Perform the M Series Maintenance Tests

This section lists equipment that we use to perform the maintenance tests that we describe in this chapter. You can substitute an equivalent device for a listed device; however, not all simulators and analyzers will produce the same results. Be sure to follow the manufacturer’s recommendations for conducting the maintenance tests.

We recommend the use of the following equipment when performing M Series Maintenance Tests:

ZOLL Medical Electrode Adapter from Fluke Biomedical (ZOLL Part Number 3010-0378).

QED 6 Defibrillator Analyzer

MFC Test Port Connector, 1004-0053-99

Fluke Biomedical 601 Pro Series International Safety Analyzer.

ECG Simulator; 12 Lead Simulator for 12 Lead test (e.g., Symbio CS1201).

Stop watch.

Standard series II PC flash memory cards.

1 red miniature alligator to miniature alligator test lead.

1 black miniature alligator to miniature alligator test lead.

DC power supply (15 amp minimum).

0.1 ohm 1% resistor (1/4 watt or greater).

1000 ohm 1% resistor (1/4 watt or greater).

Fluke 75 Multimeter or equivalent.

Printer Paper.

Battery.

AC line cord.

3 lead, 5 lead and 12 lead ECG cables. (12 lead cable needed if 12 lead option is installed.

3

M Series Service Manual

Equipment You Need for the M Series Options Maintenance Tests

Fluke Biomedical Index 2 SpO2 Simulator or equivalent. (For SpO2 units only.)

SpO2 cable and sensor (if option is installed).

EtCO2 cable, and CAPNOSTAT 3 Mainstream cable with airway adapter, or CAPNOSTAT 3 Sidestream cable with cannula (if option is installed).

Novametrix Capnostat Simulator Yb 1265/7100

Paddles (if used).

)

Fluke Biomedical BP Pump 2 NIBP Monitor Analyzer (For NIBP units only) with NIBP cable and cuff (if NIBP option is installed), or

Fluke Biomedical Cufflink Analyzer (if NIBP option is installed)

IBP Temperature Simulator 9100-0402-TF

DNI Nevada 214B Patient SImulator, with M Series interface cable

NOTE The Fluke Biomedical BP Pump NIBP Monitor Analyzer and the Fluke Biomedical BP Pump NIBP Analyzer use different technologies for testing NIBP monitors and therefore, the manual provides two different procedures for performing the NIBP Volume Leak test with each of these types of test equipment.

4

M Series Service Manual

1.0 Physical Inspection of the Unit

Tools Needed

Battery.

 

 

 

 

Test Setup

None.

 

 

 

 

 

 

 

 

 

 

 

Observe this...

Pass/Fail/NA

 

 

 

 

 

 

 

 

 

1.1

Housing

o

o

 

 

 

 

Is the unit clean and undamaged?

 

 

 

 

 

 

 

 

 

 

 

 

1.2

Does the unit show signs of excessive wear?

o

o

 

 

 

 

 

 

 

 

 

 

1.3

Does the handle work properly?

o

o

 

 

 

 

 

 

 

 

 

 

1.4

Does the recorder door open and close properly?

o

o

 

 

 

 

 

 

 

 

 

 

1.5

Are input connectors clean and undamaged?

o

o

 

 

 

 

 

 

 

 

 

 

1.6

Are there any cracks in the housing?

o

o

 

 

 

 

 

 

 

 

 

 

1.7

Do the front panel or selector switches have any damage or cracks?

o

o

 

 

 

 

 

 

 

 

 

 

1.8

Are there any loose housing parts?

o

o

 

 

 

 

 

 

 

 

 

 

1.9

Cables

o

o

 

 

 

 

Are all cables free of cracks, cuts, exposed or broken wires?

 

 

 

 

 

 

 

 

 

 

 

 

1.10

Are all cable bend/strain reliefs undamaged and free of excessive cable

o

o

 

 

 

 

wear?

 

 

 

 

 

 

 

 

 

 

 

 

1.11

Paddles

o

o

o

 

 

 

Do the adult and pedi plates have major scratches or show signs of damage?

 

 

 

 

 

 

 

 

 

 

 

 

1.12

Do the adult shoes slide on and off easily to expose the covered pedi plates?

o

o

o

 

 

 

 

 

 

 

 

 

1.13

Are the paddles clean (e.g., free of gel) and undamaged? (if applicable)

o

o

o

 

 

 

 

 

 

 

5

M Series Service Manual

Observe this...

Pass/Fail/NA

 

 

 

 

1.14

Battery

o

o

 

Place battery in battery well.

 

 

 

 

 

 

1.15

Is the battery seated in the battery well correctly?

o

o

 

 

 

 

Record your results on the Maintenance Test Checklist.

6

M Series Service Manual

 

2.0 Front Panel Button Test

Tools Needed

QED 6 Defibrillator Analyzer

Test Setup

Install strip chart paper into the recorder tray.

Install the battery in the unit or connect the A/C power cord to the unit and then plug the cord into an electrical outlet.

Connect the universal cable and ECG cable (3 lead, 5 lead, or 12 lead) to the QED 6 Analyzer (or equivalent).

 

 

Do this...

Verify that...

Pass/Fail/

 

 

 

 

 

NA

 

 

 

 

 

 

 

 

 

 

 

 

2.1

Turn the selector switch to

The unit sounds 4 beep tones. PADS and MONITOR

o

o

 

 

 

 

MONITOR.

display on the monitor.

 

 

 

 

 

 

(For AED units, turn the selector

Note: PADS is a factory default setting.

 

 

 

 

 

 

switch to ON and select Manual

 

 

 

 

 

 

 

mode.)

 

 

 

 

 

 

 

 

 

 

 

 

 

 

2.2

Set the QED 6 analyzer to NSR

As you press the SIZE button five times (0.5, 1.0, 1.5,

o

o

 

 

 

 

of 120 BPM. To check the size of

2.0, 3.0), the size of the ECG waveform appropriately

 

 

 

 

 

 

the ECG waveform, press the

changes on the display.

 

 

 

 

 

 

SIZE button.

 

 

 

 

 

 

 

 

 

 

 

 

 

 

2.3

Press the ALARM SUSPEND

Alarm symbol changes from disabled to enabled. If the

o

o

o

 

 

 

button.

alarm sounds, press the ALARM SUSPEND button to

 

 

 

 

 

 

 

turn it off. The alarm will only be suspended for 90

 

 

 

 

 

 

 

seconds at this point. Press and hold the ALARM

 

 

 

 

 

 

 

SUSPEND button for 3 seconds to disable alarms.

 

 

 

 

 

 

 

 

 

 

 

 

 

2.4

Press the RECORDER button.

The strip chart paper moves out of the unit from the

o

o

o

 

 

 

 

paper tray. Check that the correct time, date, ECG

 

 

 

 

 

 

 

lead annotation and waveform are recorded on the

 

 

 

 

 

 

 

paper. (Set Time and Date, if necessary.)

 

 

 

 

 

 

 

 

 

 

 

 

 

2.5

Open the paper compartment

The CHECK RECORDER message appears on the

o

o

o

 

 

 

door, then press the

monitor.

 

 

 

 

 

 

RECORDER button.

 

 

 

 

 

 

 

 

 

 

 

 

 

 

2.6

Close the paper compartment

The strip chart paper flows out of paper tray and the

o

o

o

 

 

 

door, then press the

monitor no longer displays the CHECK RECORDER

 

 

 

 

 

 

RECORDER button.

message.

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

7

 

 

 

 

M Series Service Manual

 

Do this...

Verify that...

Pass/Fail/

 

 

 

NA

 

 

 

 

 

 

 

 

2.7

Press the RECORDER button.

The strip chart paper stops flowing out of the paper

o

o

o

 

 

tray.

 

 

 

 

 

 

 

 

 

2.8

Press the VOLUME softkey.

The volume bar graph displays.

o

o

 

 

 

Note: The QRS tone is on or off. There is no gradual

 

 

 

 

 

change in volume. If equipped, voice prompts

 

 

 

 

 

are gradual. Note: The voice volume has 5

 

 

 

 

 

settings. Setting 3 is in the mid-range.

 

 

 

 

 

 

 

 

 

2.9

Press the INC softkey.

The bar graph appears on the display, indicating an

o

o

 

 

 

increase in volume.

 

 

 

 

 

Note: This action does not increase the volume of the

 

 

 

 

 

unit’s audio prompts.

 

 

 

 

 

 

 

 

 

2.10

Press the DEC softkey.

The bar graph appears on the display, indicating a

o

o

 

 

 

decrease in volume.

 

 

 

 

 

Verify that pressing the DEC softkey until only the last

 

 

 

 

 

bar appears silences the unit’s audio prompts,

 

 

 

 

 

otherwise the volume of the audio prompts does not

 

 

 

 

 

decrease.

 

 

 

 

 

 

 

 

 

2.11

Press the CONTRAST button.

The monitor screen displays the Contrast Menu.

o

o

 

 

 

 

 

 

 

2.12

Press the INC softkey.

The screen contrast and bar graph increases on the

o

o

o

 

 

display.

 

 

 

 

 

 

 

 

 

2.13

Press the DEC softkey.

The screen contrast and bar graph decreases on the

o

o

 

 

 

display.

 

 

 

 

 

 

 

 

 

2.14

Press the SUMMARY button.

The monitor screen displays the Summary menu,

o

o

o

 

 

showing the summary report options.

 

 

 

 

 

 

 

 

 

2.15

Press the CODEMARKER

The monitor screen displays the Code marker menu.

o

o

o

 

button.

 

 

 

 

 

 

 

 

 

 

2.16

Connect to AC or DC current

The CHARGER ON indicator lights (either the Green

o

o

o

 

and install the battery. Turn the

or Amber indicator lights).

 

 

 

 

unit off.

 

 

 

 

 

 

 

 

 

 

8

M Series Service Manual

 

Do this...

Verify that...

Pass/Fail/

 

 

 

NA

 

 

 

 

 

 

 

 

2.17

Remove the battery.

The charger light alternately flashes green and amber.

o

o

 

 

 

 

 

 

 

2.18

Replace the battery and the turn

The amber charger light illuminates.

o

o

 

 

unit on.

 

 

 

 

 

 

 

 

 

 

2.19

Press the ANALYZE button.

The SELECT DEFIB MODE message appears on the

o

o

o

 

 

monitor (for manual devices).

 

 

 

 

 

 

 

 

 

2.20

Connect the universal cable to

The display shows that the unit is charging. The

o

o

o

 

the QED 6. Move the selector

SHOCK button lights when the unit is charged, and

 

 

 

 

switch to DEFIB. Select 2J.

the Ready tone for DEFIB sounds.

 

 

 

 

Press the CHARGE button.

 

 

 

 

 

 

 

 

 

 

2.21

Press and hold the ENERGY

The unit discharges internally and selected energy

o

o

o

 

SELECT down arrow.

decrements to 1J.

 

 

 

 

 

 

 

 

 

2.22

Press and release the ENERGY

The following selections appear: Verify the following

o

o

o

 

SELECT up arrow 19 times.

energy amounts display incrementally:

 

 

 

 

 

Biphasic:2-10, 15, 20,30, 50, 75, 100,120, 150, 200J

 

 

 

 

 

Monophasic(DSW): 2-10, 15,20,30,50,75, 100, 150,

 

 

 

 

 

200, 300, 360J

 

 

 

 

 

 

 

 

 

2.23

Press the CHARGE button.

The display shows the unit charged up to 200J (360J

o

o

o

 

 

monophasic) and the SHOCK button lights.

 

 

 

 

 

 

 

 

 

2.24

Press the SHOCK button.

The unit discharges and the SHOCK button is no

o

o

o

 

 

longer illuminated.

 

 

 

 

 

A 15 second strip chart automatically prints, displaying

 

 

 

 

 

the number of joules delivered (if equipped with

 

 

 

 

 

recorder and configured to print event).

 

 

 

 

 

 

 

 

 

Record your results on the Maintenance Test Checklist.

9

M Series Service Manual

3.0 3, 5, and 12 Leads Test

Tools Needed:

3 lead, 5 lead, and 12 lead cables. Test each cable separately. QED 6 Defibrillator Analyzer.

 

 

Test Setup:

The M

Series unit must be configured to display ECG LEAD OFF message.

 

 

 

 

Connect the lead wires appropriate to each test to the QED 6 Defibrillator Analyzer.

 

 

 

 

 

 

 

 

 

 

 

 

Do this...

Observe this...

Pass/Fail/

 

 

 

 

 

NA

 

 

 

 

3.1

Turn the selector switch to

The monitor displays the NO ECG LEADS OFF

o

o

o

 

 

 

MONITOR. Select leads.

message.

 

 

 

 

 

 

 

 

 

 

 

 

 

3.2

Disconnect one lead from the

The monitor displays the ECG LEAD OFF message

o

o

o

 

 

 

simulator.

within 3 seconds (if configured).

 

 

 

 

 

 

 

 

 

 

 

 

 

3.3

Reconnect the lead. Repeat

The ECG LEAD OFF message appears when the

o

o

o

 

 

 

step 3.2 with the remaining

lead is disconnected and clears the lead is

 

 

 

 

 

 

leads.

reconnected.

 

 

 

 

 

 

 

 

 

 

 

 

 

3.4

Repeat 3.2 and 3.3 for 5 lead

Note: If heart rate alarm sounds, press and hold the

o

o

o

 

 

 

and 12 lead cables.

ALARM SUSPEND button for 4 seconds to

 

 

 

 

 

 

 

disable the alarms.

 

 

 

 

 

 

 

Note: When testing the 12 lead cable, the ECG

 

 

 

 

 

 

 

LEAD OFF message displays when you pull

 

 

 

 

 

 

 

off a limb lead. When you pull off a V lead, the

 

 

 

 

 

 

 

ECG VX LEAD OFF message displays, where

 

 

 

 

 

 

 

“X” is the number between 1 and 6.

 

 

 

 

 

 

 

 

 

 

 

Record your results on the Maintenance Tests Checklist.

10

M Series Service Manual

4.0 Power Supply Test (Optional)

Tools Needed:

1 red miniature alligator to miniature alligator test lead.

 

1 black miniature alligator to miniature alligator test lead.

 

• DC power supply (15 Amp minimum).

 

0.1 ohm resistor (¼W or greater).

 

1000 ohm 1% resistor (¼W or greater).

 

Fluke 75 multimeter or equivalent.

Test Setup:

Make sure the unit and power supply are turned off.

 

Connect one end of the black lead to the “-” terminal in the battery well.

 

Connect the other end of the black lead to the “-” terminal of the power supply.

 

Connect the red lead to “+” terminal socket of the battery well. Use the middle pin with the plastic guard around it.

 

Connect the other end of the red lead to the “+” terminal of the power supply.

 

Set the power supply voltage to 7V.

 

 

 

CAUTION

Be sure to connect the power supply properly to the M Series battery well terminals or damage to the unit may result. Do NOT raise

 

the power supply voltage above 12V.

 

11

M Series Service Manual

Red

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Battery Well

 

 

15 Amp

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

?

Supply

 

 

 

 

 

 

 

 

?

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Black

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Do this...

Verify that...

 

 

Pass/

 

 

 

 

 

 

 

 

 

 

 

 

Fall

 

4.1

Turn the selector switch to

The unit does not turn on.

 

 

o

o

 

MONITOR.

 

 

 

 

 

 

 

 

 

(For AED units, turn the selector

 

 

 

 

 

 

 

 

 

switch to ON and select Manual

 

 

 

 

 

 

 

 

 

mode.)

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

4.2

Turn the selector switch to the

 

 

 

 

 

 

 

 

 

OFF position.

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

4.3

Adjust the power supply voltage to

The unit turns on.

 

 

o

o

 

10.3V and turn the selector switch

 

 

 

 

 

 

 

 

 

to MONITOR (for AED units, turn

 

 

 

 

 

 

 

 

 

the selector switch to the ON

 

 

 

 

 

 

 

 

 

position).

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

4.4

Low Battery Test

The unit does not display the LOW BATTERY message.

o

o

 

Set voltage to 9.8V.

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

4.5

Set voltage to 9.4V.

The unit displays the LOW BATTERY message within 30

o

o

 

 

 

 

 

 

seconds.

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

4.6

Shut Down Voltage Test

The unit shuts off within 30 seconds.

 

 

o

o

 

Set voltage to 8.5V.

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Record your results on the Maintenance Tests Checklist.

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

12

M Series Service Manual

Test Setup

Remove red lead from power supply and connect to 0.1ohm resistor.

Connect other end of resistor to “+” terminal of power supply using a second red label. Connect multimeter across the resistor.

Set voltage scale (if DVM is not autoranging) to 220mV.

 

_

Red

DMM

 

+

+

 

 

 

 

Battery Well

 

 

 

_

 

 

 

 

 

 

 

 

 

 

 

 

 

+

 

 

 

 

 

 

 

 

 

15 Amp

Black

 

 

 

 

 

Supply

 

 

 

 

 

 

 

_

 

 

 

 

 

 

 

 

 

 

 

Do this....

Observe this...

Pass/

 

 

 

Fail/

 

 

 

 

4.7

System Current Test

 

 

 

Set power supply to 10.3V.

 

 

 

 

 

 

13

M Series Service Manual

 

Do this....

Observe this...

Pass/

 

 

 

Fail/

 

 

 

 

 

 

4.8

Turn the selector switch to

Voltage across resistor should be 80 mV or less (<800

o

o

 

MONITOR. (For AED units, turn

mA of ON current). Note: Without optional parameters.

 

 

 

the selector switch to ON and

a) With green screen or LCD and no options <80mV.

 

 

 

select Manual mode.

 

 

 

b) With yellow screen and no options <81mV.

 

 

 

 

 

 

 

 

c) With yellow screen and SpO2 <104mV.

 

 

 

 

d) With yellow screen and voice recording <91mV.

 

 

 

 

e) With yellow screen, voice recording, and SpO2

 

 

 

 

<114mV.

 

 

 

 

f) All devices with EtCO2 <121mV.

 

 

 

 

 

 

 

4.9

Turn unit off.

 

o

o

 

 

 

 

 

Record your results on the Maintenance Test Checklist

 

 

 

 

 

 

 

 

Off Current Test

Test Setup

Remove 0.1 ohm resistor and replace with 1K

 

Connect DMM across resistor.

 

Set voltage scale to DCV.

 

Measure voltage across resistor.

4.10

Measure across resistor with unit

Voltage should be less than 450mV (<450uA of current).

o o

 

turned off.

 

 

 

 

 

 

Record your results on the Maintenance Test Checklist

 

 

 

 

 

14

M Series Service Manual

 

 

5.0 Leakage Current Test

 

 

Tools Needed:

See the manufacturer’s instructions or supplied specifications for the leakage tester you use.

Test Setup:

See the manufacturer’s instructions or supplied specifications for the leakage tester you use. Repeat leakage test with

 

accessories: MFC, external paddles, and anterior/posterior paddles.

 

 

 

 

Table 1:

 

 

 

 

 

 

 

 

 

 

 

 

Maximum Leakage Acceptance Limits

 

 

 

 

 

 

 

 

 

 

 

 

Normal Condition

Single Fault Condition*

 

 

 

 

 

 

 

 

 

 

ECG

 

10

50

 

 

 

 

 

 

 

 

 

 

Universal

 

100

100

 

 

 

 

 

 

 

 

 

 

Earth

 

500

1000

 

 

 

 

 

 

 

 

 

*Single fault considered AC mains on applied part.

 

 

 

 

 

 

 

 

 

15

M Series Service Manual

6.0 Paddles Test (If applicable)

Tools Needed:

QED 6 Defibrillator Analyzer.

 

 

 

Test Setup:

If applicable, connect the universal cable to the paddles and place the paddles in paddle wells.

 

 

 

 

 

 

 

 

 

 

 

Do this...

Verify that...

Pass/Fail/

 

 

 

 

 

NA

 

 

 

6.1

Turn the selector switch to

The energy selection decreases to 1J.

o

o

 

 

 

DEFIB (For AED units,turn the

 

 

 

 

 

 

selector switch to ON and

 

 

 

 

 

 

select Manual mode.) Press

 

 

 

 

 

 

and hold the ENERGY DOWN

 

 

 

 

 

 

button on the sternum paddle.

 

 

 

 

 

 

 

 

 

 

 

 

6.2

Press and release the ENERGY

The joules selection increases as follows: Biphasic 1-

o

o

 

 

 

UP button on the sternum paddle

10, 15, 20, 30, 50, 75, 100, 120, 150, 200J.

 

 

 

 

 

for each setting.

Monophasic (DSW): 1-10,15, 20, 30, 50, 75, 100, 150,

 

 

 

 

 

 

200, 300, 360.

 

 

 

 

 

 

 

 

 

 

 

6.3

Press and release the

The recorder starts printing.

o

o o

 

 

 

RECORDER button on the

 

 

 

 

 

 

sternum paddle.

 

 

 

 

 

 

 

 

 

 

 

 

6.4

Press and release the

The recorder stops printing.

o

o

 

 

 

RECORDER button again.

 

 

 

 

 

 

 

 

 

 

 

 

6.5

Select 30J using the paddle

The unit charges to 30J, then the red LED charge

o

o

 

 

 

ENERGY button. Press the

indicator illuminates and the charge tone sounds.

 

 

 

 

 

CHARGE button on the Apex

Note: The front panel shock button does not

 

 

 

 

 

paddle.

illuminate.

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

6.6

Press and release the APEX

The unit does not discharge.

o

o

 

 

 

SHOCK button.

 

 

 

 

 

 

 

 

 

 

 

 

6.7

Press and release the

The unit does not discharge.

o

o

 

 

 

STERNUM SHOCK button.

 

 

 

 

 

 

 

 

 

 

 

 

6.8

Press and hold both paddles

The unit discharges. The TEST OK message displays

o

o

 

 

 

SHOCK buttons.

and the red LED turns off. If configured, the recorder

 

 

 

 

 

 

prints a strip chart.

 

 

 

 

 

 

 

 

 

16

M Series Service Manual

 

...Do this

...Verify that

Pass/Fail/

 

 

 

NA

Record your results on the Maintenance Tests Checklist.

17

M Series Service Manual

7.0 Heart Rate Display Test

Tools Needed:

 

ECG Cable (3, 5 or 12 leads).

 

 

 

 

 

QED 6 Defibrillator Analyzer.

 

 

 

Test Setup:

Turn the selector switch to MONITOR (For AED units, turn the selector switch to ON and select Manual override.)

 

Press LEAD button until “I” displays.

 

 

 

 

Connect the ECG leads to the QED 6 Defibrillator Analyzer.

 

 

 

Connect the ECG cable to the unit.

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Do this...

Verify that...

Pass/Fail/

 

 

 

 

 

 

 

NA

 

 

 

 

7.1

Set the QED 6 Defibrillator

The Heart Rate displays as 120 +/- 2 bpm.

o o o

 

 

 

 

 

Analyzer to 120BPM.

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Record your results on the Maintenance Tests Checklist.

 

 

 

 

 

 

 

 

 

 

18

M Series Service Manual

8.0 Calibrating Pulses on Strip Chart Test

Tools Needed:

None

 

 

 

Test Setup:

None.

 

 

 

 

 

 

 

 

 

 

 

 

Do this...

Verify that...

Pass/Fail/

 

 

 

 

 

NA

 

 

8.1

Press the RECORDER button.

 

 

 

 

 

 

 

 

8.2

Press and hold SIZE button to

The strip chart displays a signal of 300 ppm with an

o o o

 

 

 

activate the calibration signal.

amplitude of 10 mm +/- 1 mm. The signal also appears

 

 

 

 

 

on the video display.

 

 

 

 

 

 

 

Record your results on the Maintenance Tests Checklist.

19

M Series Service Manual

9.0 Notch Filter Test

Tools Needed:

QED 6 Defibrillator Analyzer.

 

 

Test Setup:

Connect the ECG cable to the QED 6 Defibrillator Analyzer.

 

 

Connect the ECG cable to the unit.

 

 

 

 

 

 

 

 

 

 

 

Do this...

Verify that...

Pass/Fail/

 

 

 

 

 

NA

 

 

9.1

Turn the selector switch to

 

 

 

 

 

MONITOR mode.

 

 

 

 

 

(For AED units, turn the selector

 

 

 

 

 

switch to ON and select Manual

 

 

 

 

 

mode.)

 

 

 

 

 

 

 

 

 

9.2

Select lead I, size 3x.

 

 

 

 

 

Select 60Hz (or 50 Hz for a 50Hz

 

 

 

 

 

unit) on

 

 

 

 

 

the QED 6 Defibrillator Analyzer.

 

 

 

 

 

 

 

 

9.3

Press RECORDER button.

Verify that the waveform amplitude on the strip chart is

o o o

 

 

 

 

less than 1.5 mm.

 

 

 

 

 

 

 

 

9.4

Press RECORDER button to

 

 

 

 

 

stop printing.

 

 

 

 

 

 

 

 

Record your results on the Maintenance Tests Checklist.

20

Loading...
+ 124 hidden pages