Zoll 9650-0209-01-C COVID-19 Technical specifications for portable ultrasound

RECTILINEAR BIPHASIC
RECTILINEAR BIPHASIC WAVEFORM
RECTILINEAR BIPHASICRECTILINEAR BIPHASIC DEFIBRILLATOR OPTION
DEFIBRILLATOR OPTION
DEFIBRILLATOR OPTIONDEFIBRILLATOR OPTION

General Information

General Information
General InformationGeneral Information
Introduction
M Series products are available with an advanced electrical design that provides a unique rectilinear biphasic waveform for
defibrillation and cardioversion. With this option, the electrical energy of the defibrillator is delivered in two successive current (voltage) phases of opposite polarity. This type of defibrillation waveform is commonly called biphasic, as opposed to the earlier monophasic damped sine wave common to most commercially available defibrillators.
The ZOLL M Series Rectilinear Biphasic Waveform Defibrillator Option produces a proprietary waveform designed for optimal clinical performance and tested extensively in multi-center clinical trials. These clinical trials have demonstrated that the waveform is clinically effective for both defibrillation and synchronized cardioversion.
This insert explains how the M Series Rectilinear Biphasic Waveform Defibrillator Option differs from the monophasic damped sine wave output of other M Series units. It is to be used in conjunction with the M Series Operators Guide. Important safety information relating to general use of the M Series is located in the Safety Considerations section of the
M Series Operators Guide.
M Series with Rectilinear Biphasic Waveform Defibrillator Option Indications for Use
WAVEFORM
WAVEFORM WAVEFORM
The ZOLL M Series Biphasic Option is to be used only by qualified medical personnel for converting Ventricular Fibrillation (VF), a cardiac rhythm incompatible with life, and/or Ventricular Tachycardias (VT) to sinus rhythm or other cardiac rhythms capable of producing hemodynamically stable heart beats.
In addition, this product is to be used in the synchronized mode only by qualified medical personnel to terminate Atrial Fibrillation (AF) at lower energies and currents than monophasic defibrillators. A qualified physician must decide when synchronized cardioversion is appropriate.
This product is also to be used in the synchronized mode only by qualified medical personnel to terminate Ventricular Tachycardias (VT). A qualified physician must decide when synchronized cardioversion is appropriate.
The Rectilinear Biphasic Waveform (RBW) has been successfully tested in multi-center, prospective, randomized, transthoracic defibrillator VT/VF and AF clinical trials, and proven to defibrillate and cardiovert adult patients at lower energies and currents than existing monophasic devices. The M Series Biphasic Option incorporates a broad range of user selectable energy settings some of which are lower than those used during those clinical trials.
There are currently no clinical studies related to the use of the Rectilinear Biphasic Waveform (RBW) in pediatric applications or for direct defibrillation of the heart during open chest surgical procedures.
The AED or advisory function should only be used to confirm ventricular fibrillation in patients meeting the following clinical criteria:
the patient should be unconscious and unresponsive.
the patient should be apneic (not breathing).
the patient should be pulseless.
WARNING
Do not use the units AED function on patients under 8 years of age. (Per AHA Guidelines for Adult Cardiopulmonary Resuscitation and AED, 3-5, 1998).
9650-0209-01 Biphasic - 1 Rev. C
OPTION INSERT

Defibrillator Function

The M Series Rectilinear Biphasic Waveform Defibrillator Option is a DC defibrillator capable of delivering up to 200 joules of energy. It can be used for defibrillation or in a synchronized mode for Cardioversion using the R-wave of the patients ECG as a timing reference. The unit operates with external paddles, or disposable pre-gelled, MFE Pads for defibrillation and cardioversion.
Energy Selection and Displays
Multiple energy levels, and the ability to program initial and subsequent shock energies, allow users to set the M Series to either a non-progressive or progressive sequence of shocks.
The M Series supports the use of progressive shock sequences to provide an energy reserve, allowing the delivery of a higher energy shock if a lower energy shock fails to terminate the arrhythmia. A sequence of 120J Biphasic, 150J Biphasic and 200J Biphasic would most closely approximate the current AHA recommended escalating energy sequence of 200J, 300J, and 360J for
defibrillator.
adult defibrillation using a monophasic
When an M Series device is equipped with the Biphasic Option, all Energy Displays shown in Section 3, Section 4, and Section 5 (for Manual Defibrillation, Advisory Defibrillation and Automated Defibrillation) of this M Series Operators Guide are modified to add the word BIPHASIC to the energy display as shown below.
The M Series is capable of delivering energy up to 200 Joules with external paddles or Multi-Function Pads. The initial default energy setting for the M Series Biphasic option is 120J with external paddles or Multi-Function Pads. The energy settings available with the M Series Bi-Phasic device are 1,2, 3, 4, 5, 6, 7, 8, 9, 10, 15, 20, 30, 50, 75, 100, 120, 150 and 200 Joules. Energy settings are selected by using the controls located on the Sternum Paddle or on the front panel of the unit.
Upon power-up M Series units equipped with the Rectilinear Biphasic Waveform Defibrillator Option automatically default to the 120 Joule setting. In units with automatic progressive energy selection enabled, Shock No. 1 is set at 120J; Shock No. 2 is set at 150J and Shock No. 3 is set at 200J. These are the default energy settings. They may be modified following the instructions in the M Series Configuration Guide.
Charge Time
Charge time: < 6 seconds with a new fully charged battery (first 15 charges to 200 Joules). Depleted batteries will result in a longer Defibrillator charge time.
All other operational aspects of the M Series Rectilinear Biphasic Waveform Defibrillator are identical to those described in the M Series Operators Guide.
Recorder Printer Annotations
In addition to the information normally printed on the M Series recorder and the M Series Summary Reports described in Section 2, units equipped with the Biphasic Option will also print Defibrillation Impedance (DEFIB
IMPEDANCE), and Delivered Current (PATIENT CURRENT). This information will be included in the Summary Report for each defibrillation shock.
NAME
BIPHASIC DEFIBRILLATION
ECG LEAD ECG SIZE
JOULES SELECTED JOULES DELIVERED PATIENT CURRE NT DEFIB IMPEDANCE
PADS
3.0
30 34 34 A 71 Ohms
PRE SHOCK 15:18:45 10 MAY 93
POST SHOCK 15:18:45
The maximum energy available from the M Series with Rectilinear Biphasic Waveform Option is 200 Joules. If the operator attempts to increase the energy level above 200 Joules the error message 200 Joules Max Biphasic will be displayed.
9650-0209-01 Biphasic- 2
RECTILINEAR BIPHASIC WAVEFORM
Energy Delivery Test
The energy delivery test for the Biphasic Option is performed at 30 J according to the instructions contained in Section 9 of the M Series Operators Guide.
When performing the energy delivery test at 30J the recorder strip should resemble that shown below.
12:46:30 02-SEP-99 PAD SIZE 1.0 HR = 30 30 JOULES TEST OK TEST_CUR = 11A DEFIB_IMPED = 1
Rectilinear Biphasic Waveform Defibrillator Option Information
The Biphasic Option has been designed to produce a Rectilinear Biphasic Waveform whose shape remains essentially constant from patient to patient by the device.
Current (I Patient)
BIPHASIC DEFIBRILLATION PULSE
120 JOULES INTO 50 OHM LOAD
Duration (msec)
The Rectilinear Biphasic Waveform consists of a 6 millisecond, essentially constant current first phase followed by a 4 millisecond, truncated exponential second phase. The first and second phases of the defibrillation waveform are of opposite polarity and their amplitudes vary based on the user selected therapeutic energy level. The initial amplitude of this waveform's second phase is approximately equal to the first phase's final amplitude. This waveform has an integrated patient impedance measurement sensing pulse at the beginning of the waveform. The positive and negative phases are separated by 100 µsec. The shape of the waveforms first phase is controlled by electronics and software which compensate for different transthoracic impedances to maintain an essentially constant current throughout the first phase.
When the highest energy setting is selected and patient impedance exceeds 85 ohms, the first phase of the waveform will droop. All other waveform parameters (phase duration, inter-phase delay and integrated impedance measurement sensing pulse) remain the same. The following Rectilinear Biphasic Waveform is produced when the M Series with Biphasic option is discharged into a 50 load at the default energy setting of 120 Joules. The vertical axis is in amperes; the horizontal axis is in milliseconds. (For more detailed information regarding the parameters of the Rectilinear Biphasic Waveform when discharged into 25, 50, and 100 loads at a maximum energy setting of 200 Joules refer to the Specification section of this Operators Insert.
9650-0209-01 Biphasic - 3
OPTION INSERT

Clinical Trials Results for the M Series Biphasic Waveform:

The Efficacy of ZOLLs Rectilinear Biphasic Waveform has been clinically verified during various studies for defibrillation of Ventricular Fibrillation (VF) / Ventricular Tachycardia (VT) and for synchronized cardioversion of Atrial Fibrillation (AF). Feasibility studies were performed initially for defibrillation of VF/VT (n=20) and synchronized cardioversion of AF (n=21) on two separate groups of patients to ensure waveform safety and energy selection. Subsequently two separate, multi-center, randomized clinical trials were performed to verify the waveforms efficacy. Descriptions of these studies are provided below. All studies were performed using ZOLL defibrillation systems consisting of ZOLL defibrillators, the ZOLL Rectilinear Biphasic Waveform and ZOLL Multi-Function Pads.
A) Randomized Multi-Center Clinical Trial for Defibrillation of Ventricular Fibrillation (VF) and Ventricular Tachycardia (VT):
Overview: The defibrillation efficacy of ZOLLs Rectilinear
Biphasic Waveform was compared to a monophasic damped sine waveform in a prospective, randomized, multi­center study of patients undergoing ventricular defibrillation for VF/VT during electro-physiological studies, ICD implants and tests. A total of 194 patients were enrolled in the study. Ten (10) patients who did not satisfy all protocol criteria were excluded from the analysis.
Objectives: The primary goal of this study was to compare the first shock efficacy of the 120J Rectilinear Biphasic Waveform with a 200J monophasic waveform. The secondary goal was to compare all shock (three consecutive 120, 150, 170J) efficacy of the Rectilinear Biphasic Waveform with that of a monophasic waveform (three consecutive 200, 300, 360J). A significance level of p=0.05 or less was considered statistically significant using Fischers Exact test. Also, differences between the two waveforms were considered statistically significant when the customary 95% or AHA recommended 90%* confidence interval between the two waveforms was greater than 0%.
Results: The study population of 184 patients had a mean
14 years. 143 patients were males. 98 patients
age of 63 were in the biphasic group (ventricular fibrillation/flutter, n=80, ventricular tachycardia, n=18) and 86 patients were in the monophasic group (ventricular fibrillation/flutter, n=76, ventricular tachycardia, n=10). There were no adverse events or injuries related to the study.
The first shock, first induction efficacy of biphasic shocks at 120J was 99% versus 93% for monophasic shocks at 200J (p=0.0517, 95% confidence interval of the difference of
---2.7% to 16.5% and 90% confidence interval of the difference of ---1.01% to 15.3%).
±
Monophasic Biphasic
1st Shock Efficacy 93% 99%
p-value 0.0517
95% Confidence. Interval -2.7% to 16.5%
90% Confidence Interval -1.01% to 15.3%
Successful defibrillation with rectilinear biphasic shocks was achieved with 58% less delivered current than with monophasic shocks (14 ±1 vs. 33 ±7 A, p=0.0001).
The difference in efficacy between the rectilinear biphasic and the monophasic shocks was greater in patients with high transthoracic impedance (greater than 90). The first shock, first induction efficacy of biphasic shocks was 100% versus 63% for monophasic shocks for patients with high impedance (p=0.02, 95% confidence interval of the difference of ---0.021% to 0.759% and 90% confidence interval of the difference of 0.037% to 0.706%).
Monophasic Biphasic
1st Shock Efficacy (High
Impedance Patients)
p-value 0.02
95% Confidence. Interval -0.021% to 0.759%
90% Confidence Interval 0.037% to 0.706%
A single patient required a second biphasic shock at 150J to achieve 100% efficacy versus six patients for whom shocks of up to 360J were required for 100% total defibrillation efficacy.
Conclusion: The data demonstrate the equivalent efficacy of low energy rectilinear biphasic shocks compared to standard high energy monophasic shocks for transthoracic defibrillation for all patients at the 95% confidence level. The data also demonstrate the superior efficacy of low energy rectilinear biphasic shocks compared to standard high energy monophasic shocks in patients with high transthoracic impedance at the 90% confidence level. There were no unsafe outcomes or adverse events due to the use of the rectilinear biphasic waveform
* Kerber, R., et. al., AHA Scientific Statement, Circulation, 1997; 95: 1677-1682:  the task force suggests that to demonstrate superiority of an alternative waveform over standard waveforms, the upper boundary of the 90% confidence interval of the difference between standard and alternative waveforms must be < 0% (i.e., alternative is greater than standard).
63% 100%
.
9650-0209-01 Biphasic- 4
RECTILINEAR BIPHASIC WAVEFORM
B). Randomized Multi-Center Clinical trial for Cardioversion of Atrial Fibrillation (AF).
Overview: The defibrillation efficacy of ZOLLs Rectilinear
Biphasic Waveform was compared to a monophasic damped sine waveform in a prospective randomized multi­center study of patients undergoing cardioversion of their atrial fibrillation. A total of 173 patients entered the study. Seven (7) patients who did not satisfy all protocol criteria were excluded from the analysis. ZOLL disposable gel electrodes with surface areas of 78 cm
2
(posterior) were used exclusively for the study.
cm
Objective: The primary goal of the study was to compare the total efficacy of four consecutive rectilinear biphasic shocks (70J, 120J, 150J, 170J) with four consecutive monophasic shocks (100J, 200J, 300J, 360J). The significance of the multiple shocks efficacy was tested statistically via two procedures, the Mantel-Haenszel statistic and the log-rank test, significance level of p=0.05 or less was considered statistically significant. The data are completely analogous to the comparison of two survival curves using a life-table approach where shock number plays the role of time.
The secondary goal was to compare the first shock success of rectilinear biphasic and monophasic waveforms. A
significance level of p=0.05 or
statistically significant using Fisher Exact tests. Also, differences between the two waveforms were considered statistically significant when the 95% confidence interval between the two waveforms was greater than 0%.
Results: The study population of 165 patients had a mean
12 years with 116 male patients.
age of 66
The total efficacy of consecutive rectilinear biphasic shocks was significantly greater than that of monophasic shocks. The following table displays the Kaplan-Meier (product-limit) survival curves for each of the two waveforms. As all patients begin in the failure mode, the estimated life-table probabilities refer to the chance of still being in failure after the k
±
th
shock (k=1,2,3,4):
2
(anterior) and 113
less was considered
Kaplan-Meier Estimate for the
Shock #
Probability of Shock Failure Biphasic Monophasic 0 1.000 1.000 1 0.318 0.792 2 0.147 0.558 3 0.091 0.324 4 0.057 0.208
As can be seen from the table, the Biphasic experience is superior over the entire course of shocks delivered. The one degree of freedom chi-square statistic for the Mantel­Haenszel test is 30.39 (p<0.0001). Similarly, the log-rank test, also a one degree of freedom chi-square statistic, is
30.38 (p<0.0001). The residual number of patients not successfully treated after four shocks is 5.7% for biphasic compared to 20.8% for monophasic.
There was a significant difference between the first shock efficacy of biphasic shocks at 70J of 68% and that of monophasic shocks at 100J of 21% (p=0.0001, 95% confidence interval of the difference of 34.1% to 60.7%).
Successful cardioversion with rectilinear biphasic shocks was achieved with 48% less delivered current than with
1 vs. 21±4 A, p<0.0001).
monophasic shocks (11
One half of the patients who failed cardioversion after four consecutive escalating monophasic shocks were subsequently successfully cardioverted using a biphasic shock at 170J. No patient was successfully cardioverted using a 360J monophasic shock after the patient had failed cardioversion with biphasic shocks.
Conclusion: The data demonstrate the superior efficacy of low energy rectilinear biphasic shocks compared to high energy monophasic shocks for transthoracic cardioversion of atrial fibrillation. There were no unsafe outcomes or adverse events due to the use of Rectilinear Biphasic Waveform.
±
Synchronized Cardioversion of Atrial Fibrillation
Cardioversion of Atrial Fibrillation (AF) and overall clinical effectiveness is enhanced by proper pad placement. Clinical studies (refer to above) of the M Series Biphasic Defibrillator Waveform Option demonstrated that high conversion rates are achieved when defibrillation pads are placed as shown in the diagram below.
Recommended Anterior/Posterior Placement
Back/
Posterior
Place the Front (Apex) pad on the third intercostal space, mid clavicular line on the right anterior chest. The Back/Posterior Pad should be placed in the standard posterior position as shown.
Front/
Apex
Defibrillation and Cardioversion Performance
Caution:
Defibrillator Waveform Option are based upon the use of ZOLL Multi-Function Pads. The combination of waveform, electrode properties and gel characteristics is essential to achieving efficacy results similar to those described above.
For synchronized cardioversion of Atrial Fibrillation, the combination of waveform, electrode properties, gel characteristics and pad placement is essential to achieving efficacy results similar to those above.
WARNING: Unnecessary skin damage can result from incorrect application or use of a defibrillation pad other than the type recommended.
The clinical results for the ZOLL Biphasic
9650-0209-01 Biphasic - 5
OPTION INSERT
Additional Messages and TroubleShooting
The following chart lists the messages that may appear on the M Series unit relating to the Biphasic Option, why the message appeared, and the action(s) to take.
The operator should become thoroughly familiar with this information before using the device.
Message Possible Cause(s) Recommended Action(s)
200 J MAX BIPHASIC Appears when trying to select an energy higher No higher energy is available. than 200 Joules. Use 200J setting.
BRIDGE TEST FAILED Biphasic module not operating properly while Try charging again. Attempt to charging. clear the message by turning the Selector Switch to OFF then back to the desired operating mode. If Fault persists contact ZOLL Technical Service Department.
BRIDGE SHORT Current higher than expected was detected. Ensure pads/paddles are used
properly. Attempt to clear the message by turning the Selector Switch to OFF then back to the desired operating mode. If Fault persists contact ZOLL Technical Service Department.
Additional Specifications and Changes

General

Refer to M Series Operators Guide for ALL Specifications except the following:
Waveform: Rectilinear Biphasic
Energy Settings: 1, 2, 3, 4, 5, 6, 7, 8, 9, 10, 15, 20, 30, 50, 75, 100, 120, 150, 200
Charge Time: Less than 6 seconds with a new fully charged battery
(first 15 charges at 200J)
Messages: 200 J MAX BIPHASIC, BRIDGE TEST FAILED, BRIDGE SHORT
Operating Time: For a new, fully charged battery pack at 20°C: 40 defibrillator
discharges at maximum energy (200J), or 2.75 hours minimum of continuous ECG monitoring, or 2.25 hours of continuous ECG monitoring/pacing at 60 mA, 80 beats/min.
9650-0209-01 Biphasic- 6
RECTILINEAR BIPHASIC WAVEFORM
The following table shows the Rectilinear Biphasic waveforms characteristics when discharged into 25Ω, 50Ω, and 100Ω loads at a maximum energy setting of 200 Joules.
I
01 = First Phase
MAX
Maximum Initial Current
I
01 = First Phase
AVG
Average Current
Discharged into 25ΩΩΩ Load Discharged into 50ΩΩΩΩ Load Discharged into 100ΩΩΩΩ Load
30 A 26 A 21 A
27 A 23 A 16 A
TD 01 = First Phase Duration
T
= Interphase
INTD
duration between first and second phases.
I
02 = Second Phase
MAX
Maximum Initial Current
I
02 = Second Phase
AVG
Average Current
TD 02 = Second Phase Duration
6 ms 6 ms 6 ms
100 µs 100 µs 100 µs
26 A 21 A 14 A
15 A 15 A 12 A
4 ms 4 ms 4 ms
9650-0209-01 Biphasic - 7
Loading...