Young 06206 User Manual

4.5 (2)

METEOROLOGICAL INSTRUMENTS

INSTRUCTIONS

WIND TRACKER

MODEL 06206

R.M. YOUNG COMPANY 2801 AERO PARK DRIVE, TRAVERSE CITY, MICHIGAN 49686, USA TEL: (231) 946-3980 FAX: (231) 946-4772 WEB: www.youngusa.com

PN: 06206-90

REV: G043014

Young 06206 User Manual

MODEL 06206

WIND TRACKER

INTRODUCTION

The YOUNG Model 06206 Marine Wind Tracker is a compact wind speed and direction display with advanced features relative wind angle and NMEA serial I/O making it suitable for shipboard use.

FEATURES

3-digit wind speed display

3-digit maximum wind speed or wind direction display

Multi-color wind direction display with variability

Wind speed and direction alarms with delay

RS-485/NMEA serial connections

Calibrated 0-5 VDC outputs

Display brightness control

Compatible with YOUNG wind sensors & 4-20 mA inputs

PRECAUTIONS

INDOOR USE ONLYunless placed in approved enclosure

Operatingtemperaturerange0-50°C(32-122°F),0-95%RH

Useonlyrecommendedpowersources.12-30VDC,3.5W

Disconnect power when connecting or servicing

Alarm contact rating 24 VAC/30 VDC, 5 A resistive, 2 A inductive maximum

FACTORY DEFAULT SETTINGS

INPUT:

-Sensor Input Setting: "05" Wind Monitor

-Input selector switch position: "Down" (Sensor)

OPTIONS:

-Averaging: "No"

-Right hand display window: "Max Wind Speed"

-Alarms: "No"

OUTPUT:

-Voltage Outputs: 0-5Vdc=0-100m/s / 0-5Vdc = 0-360°

-NMEA Serial Output Rate: "FST"

IMPORTANT NOTE!:

Before connecting any external devices, verify the above settings are compatible with your application. If different, refer to the wiring diagrams in the back of the manual for the correct settings.

FRONT PANEL

1.Wind speed display

2.Wind speed units indicator

3.Alarm status indicators

4.Data averaging indicator

5.Maximum wind speed or direction display

6.Wind direction and variability display

7.Brightness control (operate) or Enter key (setup)

8.Maximum Reset (operate) or Select (setup)

BACK PANEL

1.Power input coaxial jack (12-30 VDC)

2.Power input terminals (12-30 VDC)

3.Sensor or 4-20 mA inputs

4.Earth ground connection

5.RS-485 serial input/output

6.0-5 VDC calibrated outputs

7.Alarm relay connections (Normally Open)

8.Input Selector Switch (Sensor or 4-20mA)

Page 1

06206-90(G)

MOUNTING AND START-UP

1.For best visibility, place the Wind Tracker in a location free of direct sunlight. Mount it using the attached bracket or remove bracketforflushmountingtoabulkheadorpanelcutout.Panel cutout dimensions are given in the specifications.An optional rack mounting panel (Model 06280) and protective enclosure (Model 06260) are available from your YOUNG supplier.

2.Connect cables to terminals according to wiring diagram.

IMPORTANT NOTE! Please observe correct position of back-panel Input Selector Switch according to wiring diagram.

3.Connect GND terminal to suitable earth ground.

4.Insert power supply plug into power jack, plug power supply into a suitable AC wall outlet or connect to suitable 12 to 30 VDC power source to terminals.

IMPORTANT NOTE! Do not connect two different power sources to the Marine Wind Tracker at the same time.

5.Whenpowerisapplied,theWindTrackerwilldisplayfirmware version number then begin to display wind information as follows:

Wind speed

Wind speed units

Maximum wind speed or direction degrees

Wind direction (single orange indicator)

Direction variability (green indicators)

Alarm status indicators (if selected)

Data averaging indicator (if selected)

6.Observe display to confirm proper operation.

CHANGING SETTINGS

Marine Wind Tracker parameters may be inspected or changed in SETUP mode. Press and hold both ENTER and SELECT keys for about 4 seconds. When SETUP mode is active, abbreviations identify each function and available options as listed below. The SELECT key changes options or values. The ENTER key saves and moves to the next parameter.

Appearance of options depends on parameter settings. Some options may be hidden.

DISPLAY

SETUP FUNCTION

InP

03

Input / Sensor Type

Wind Sentry

 

04

Wind Monitor-Jr

 

05

Wind Monitor

 

SEr

NMEA serial input

 

Ld2

Line Driver 4-20mA input (0-100 m/s)

 

Hd

Wind Monitor HD

 

Ldi

Line Driver 4-20mA input (0-50 m/s)

SPd

unt

Wind Speed Units (annunciator blinks)

SELECT key changes units. ENTER to save

Out

FSt

NMEA Serial Output Rate

16 times per second

 

SLO

Once per second

dSP

no

Display Averaging (annunciator blinks)

Instantaneous data displayed

 

YES

Average data displayed

PEr

030

Set averaging period in seconds (0-999).

 

 

Display will update at this interval.

dSP

SPd

Right Display Window Selection

Maximum wind speed

 

dir

Wind Direction degrees

ALr

no

Wind Direction Alarm (annunciator blinks)

Direction alarm not armed

 

YES

Direction alarm armed

ALr

dir

SELECT key sets direction alarm sector start.

ALr

SPn

ENTER key saves.

SELECT key sets direction alarm sector span.

 

 

ENTER key saves.

ALr

no

Wind Speed Alarm (annunciator blinks)

Speed alarm not armed

 

YES

Speed alarm armed

ALr

000

Alarm set-point. SELECT key increments value.

 

 

ENTER key saves.

dLY

030

Alarm Delay Time

Alarm delay time in seconds (0-999). SELECT key

 

 

increments value. ENTER key saves.

Snd

no

Sound

No sound

 

YES

Audible beep with alarm activations or average

 

 

update.

dir

360

Wind Direction Voltage Output Scale

0-360 degrees

 

540

0-540 degrees

tSt

no

Test Functions

No test

 

YES

Test

tSt

ALr

SELECT key closes alarm relays.

CAL

0.00

SELECT key alternates between 0.00 and

tSt

dsP

5.00 VDC output to calibrate external devices.

SELECT key tests display sections.

OPERATION

ALARMS

Wind speed and direction alarms each have their own set-point, LEDstatusindicator,andrelaycontacts.AnAlarmDelayparameter establishesthetimedurationinoroutoftheset-pointrangeneeded for the alarm to change state. During operation, front panel LEDs indicate alarm status.

LED Off = Alarm not armed and OFF. Relay open LED Steady = Alarm armed and OFF. Relay open

LED Blinking = Alarm ON. Relay closed. Audible beep if Sound parameter has been set.

AVERAGING

When averaging is enabled, the front-panel AVG annunciator is illuminated, and average wind speed and direction values are displayed at intervals set by the Period (PEr) parameter. When disabled, instantaneous wind values are displayed.

06206-90(G)

Page 2