Yamato Scientific BB600 Low Constant Temperature Water Bath User Manual

Loading...

Low Constant Temperature

Water Bath

Model

BB300/400/600

Instruction Manual

- Fourth Edition -

zThank you for purchasing "Low Constant Temperature Water Bath, BB Series" of Yamato Scientific Co., Ltd.

zTo use this unit properly, read this "Instruction Manual" thoroughly before using this unit.

Keep this instruction manual around this unit for referring at anytime.

WARNING!:

Carefully read and thoroughly understand the important warning items described in this

Yamato Scientific Co. LTD.

 

 

Contents

Cautions in Using with Safety................................................................

1

Explanation....................................................................................................................

1

Table of Illustrated Symbols ..........................................................................................

2

Fundamental Matters of "WARNING!" and "CAUTION!"...............................................

3

Before Using This Unit ...........................................................................

4

Requirements for Installation.........................................................................................

4

Installation Procedure....................................................................................................

7

Description and Function of Each Part .................................................

9

Main Unit .......................................................................................................................

9

Control Panel...............................................................................................................

10

Characters of the Controller ........................................................................................

11

Operation Method .................................................................................

12

Operation Mode and Function List ..............................................................................

12

Operation Mode, Function Setting Key, and Characters .............................................

14

Setting of Overheating Prevention Device ..................................................................

15

Fixed Temperature Operation......................................................................................

16

Quick Auto Stop Operation..........................................................................................

18

Auto Stop Operation....................................................................................................

19

Auto Start Operation....................................................................................................

21

Calibration Offset Function..........................................................................................

23

Lock Function..............................................................................................................

24

Temperature Output Terminal......................................................................................

25

RS485 Communication Function.................................................................................

27

Cooling curve, cooling capacity curve (reference data)...............................................

39

Flow Rate and Head (reference data) .........................................................................

41

Nybrine Freezing Temperature and Viscosity (reference data) ...................................

42

Device to Install (reference data) ................................................................................

43

Handling Precautions ...........................................................................

44

Maintenance Method.............................................................................

47

Daily Inspection and Maintenance ..............................................................................

47

Long storage and disposal...................................................................

49

When not using this unit for long term / When disposing ............................................

49

In the Event of Failure… .......................................................................

50

Safety Device and Error Code.....................................................................................

50

Trouble Shooting .........................................................................................................

51

After Service and Warranty ..................................................................

52

Specification..........................................................................................

53

 

Wiring Diagram......................................................................................

54

 

BB300..........................................................................................................................

54

 

BB400..........................................................................................................................

55

 

BB600..........................................................................................................................

56

 

Piping Diagram......................................................................................

57

 

Replacement Parts Table......................................................................

58

 

Reference...............................................................................................

59

 

List of Dangerous Substances ....................................................................................

59

 

Installation Standard Manual................................................................

60

Cautions in Using with Safety

Explanation

MEANING OF ILLUSTRATED SYMBOLS

Illustrated Symbols

Various symbols are used in this safety manual in order to use the unit without danger of injury and damage of the unit. A list of problems caused by ignoring the warnings and improper handling is divided as shown below.Be sure that you understand the warnings and cautions in this manual before operating the unit.

WARNING!

CAUTION!

If the warning is ignored, there is the danger of a problem that may cause a serious accident or even fatality.

If the caution is ignored, there is the danger of a problem that may cause injury/damage to property or the unit itself.

Meaning of Symbols

This symbol indicates items that urge the warning (including the caution). A detailed warning message is shown adjacent to the symbol.

This symbol indicates items that are strictly prohibited.

A detailed message is shown adjacent to the symbol with specific actions not to perform.

This symbol indicates items that should be always performed.

A detailed message with instructions is shown adjacent to the symbol.

1

Cautions in Using with Safety

Table of Illustrated Symbols

Warning

Warning,

Warning,

Warning,

Warning,

Warning,

generally

high voltage

high temperature

drive train

explosive

Caution

Caution,

Caution,

Caution,

Caution,

Caution,

generally

electrical shock

scald

no road heating

not to drench

Caution,

Caution,

water only

deadly poison

Prohibit

Prohibit,

Prohibit,

Prohibit,

Prohibit,

generally

inflammable

to disassemble

to touch

Compulsion

Compulsion,

Compulsion,

Compulsion,

Compulsion,

Compulsion,

generally

connect to the

install on a flat

disconnect the

periodical

 

grounding

surface

power plug

inspection

 

terminal

 

 

 

2

Cautions in Using with Safety

Fundamental Matters of "WARNING!" and "CAUTION!"

WARNING!

Do not use this unit in an area where there is flammable or explosive gas

Never use this unit in an area where there is flammable or explosive gas. This unit is not explosion-proof. An arc may be generated when the power switch is turned on or off, and fire/explosion may result. (Refer to page 59 "List of Dangerous Substances".)

Always ground this unit

Always ground this unit on the power equipment side in order to avoid electrical shock due to a power surge.

If a problem occurs

If smoke or strange odor should come out of this unit for some reason, turn off the circuit breaker right away, and then disconnect the power plug. Immediately contact a service technician for inspection. If this procedure is not followed, fire or electrical shock may result. Never perform repair work yourself, since it is dangerous and not recommended.

Do not use the power cord if it is bundled or tangled

Do not use the power cord if it is bundled or tangled. If it is used in this manner, it can overheat and fire may be caused.

Do not process, bend, wring, or stretch the power cord forcibly

Do not process, bend, wring, or stretch the power cord forcibly. Fire or electrical shock may result.

Substances that can not be used

Never use explosive substances, flammable substances and substances that include explosive or flammable ingredients in this unit. Explosion or fire may occur. (Refer to page 59 "List of Dangerous Substances".)

Do not disassemble or modify this unit

Do not disassemble or modify this unit. Fire or electrical shock or failure may be caused.

Do not touch high-temperature parts

The inside of the body or the door may become hot during and just after operation. It may cause burns.

CAUTION!

During a thunder storm

During a thunderstorm, turn off the power key immediately, then turn off the circuit breaker and the main power. If this procedure is not followed, fire or electrical shock may be caused.

3

Before Using This Unit

Requirements for Installation

WARNING!

1.Choose a proper place for installation

Do not install this unit in a place where:

Rough or dirty surface.

Flammable gas or corrosive gas is generated.

Ambient temperature above 30°C.

Ambient temperature fluctuates violently.

There is direct sunlight.

There is excessive humidity and dust.

There is a constant vibration.

Winds from the air conditioner, etc. hit the sample container directly.

Install this unit on a stable place with the space as shown below.

More than 20cm

More than

Main Unit

More than

20cm

20cm

 

More than 20cm

2.Installation on horizontal surface

Place this unit as flat a place as possible. If the casters are not in uniform contact with the floor surface, noise or vibration may result. Additionally, the unit may cause a problem or malfunction.

Weight of unit is BB300 type: 46 kg BB400 type: 53kg BB600 type: 70kg. Conveyance and installation should be done carefully by two or more persons.

3.Before/after installing

It may cause injure to a person if this unit falls down or moves by the earthquake and the impact. etc..To prevent, take measures that the unit cannot fall down, and not install to busy place.

4

Before Using This Unit

Requirements for Installation

4.Secure the ventilation of device

Do not operate while the device side/back is obstructed. The temperature inside the device rises and may become the cause of an accident, a failure and a fire.

5.Do not use in the place that the device is exposed to liquid

Do not operate in the place that the device is exposed to liquid. If liquid go inside of the device, it will become the cause of an accident, failure, an electric shock, and a fire.

6.Do not use this unit in an area where there is flammable or explosive gas

Never use this unit in an area where

there is flammable or explosive gas.

 

This unit is not explosion-proof. An arc

 

may be generated when the power

 

switch is turned ON or OFF, and

 

fire/explosion may result.

 

To know about flammable or explosive gas, refer to page 59 “List of Dangerous Substances”.

5

Before Using This Unit

Requirements for Installation

CAUTION!

7.Choose a correct power distribution board or receptacle

Choose a correct power distribution board or receptacle that meets the unit’s rated electric capacity.

Electric capacity: BB300: 100V AC, 12A BB400: 100V AC, 13A BB600: 100V AC, 20A

NOTE)

There could be the case that the unit does not run even after turning ON the power. Inspect whether the voltage of the main power is lowered than the specified value, or whether other device(s) uses the same power line of this unit. If the phenomena might be found, change the power line of this unit to the other power line.

Starburst connection with a branching receptacle or extended wiring with a cord reel lowers electrical power voltage, which may cause the degradation of refrigeration capability.

Connect the unit to only the power supply. If it is connected to a gas pipe, water pipe or telephone line, an accident or malfunction may result.

8.Handling of power code

Do not entangle the power cord. This will cause overheating and possibly a fire.

Do not bend or twist the power cord, or apply excessive tension to it. This may cause a fire and electrical shock.

Do not lay the power cord under a desk or chair, and do not allow it to be pinched in order to prevent it from being damaged and to avoid a fire or electrical shock.

Keep the power cord away from any heating equipment such as a room heater. The cord's insulation may melt and cause a fire or electrical shock.

If the power cord becomes damaged (wiring exposed, breakage, etc.), immediately turn off the power at the rear of this unit and shut off the main supply power. Then contact your nearest dealer for replacement of the power cord. Leaving it may cause a fire or electrical shock.

Connect the power plug to the receptacle which is supplied appropriate power and voltage.

9.Always ground this unit

Be sure to connect the earth wire (the green cable of power cord) to the grounding conductor or ground terminal to prevent accidents caused by electric leakage.

Do not connect the earth wire to gas or water pipes. If not, fire disaster may be caused.

Do not connect the earth wire to the ground for telephone wire or lightning conductor. If not, fire disaster or electric shock may be caused.

Please consult your local electrical contractor for power connecting work.

BB600 type

 

Green (to ground)

Rounded terminal 3. 5-4

Black (to power supply)

Rounded terminal 3. 5-5

White (to power supply)

 

6

Before Using This Unit

Installation Procedure

 

 

1 Release the stopper lock of the casters.

Button position: high

Locked

Push down the stopper button of the casters as

(during installation)

shown in the right figure. It will be unlocked.

 

(Only the two casters on the front side of the unit

 

are equipped with a caster.)

 

 

Button position: low

 

 

 

Push to unlock the

 

caster (movable)

 

Caster

 

 

2Move the unit to the place of installation.

If there is a bump on the floor, the casters may receive excessive load and get damaged. In this case, lift and move carefully by two or more persons.

3After the unit is placed in the desired position, lock the stopper button of the casters.

4 Check the drain cock.

Close

Position during operation

Confirm that the drain cock on the backside of the

 

 

 

unit is in the "Close" position (perpendicular to the

 

 

cock).

Open

 

Confirmation of overflow.

 

Position during draining

Connect the attached hose for overflow and

 

prepare a container etc. separately.

 

 

5For external sealed unit connection

(The package does not include any circulation hose. Please prepare it yourself.)

Connect the connection port of the main part to the circulation route of the external sealing system so that there may be no leakage. Connect the hose to the return port (IN) and discharge port (OUT) of the main part. The diameter of a hose nipple isφ10.5 . Refer to the following figure.

External unit

(sealed system)

BB unit

Insulation hose

Return

 

 

 

Discharge

(IN)

 

(OUT)

(IN

 

OUT

Insulation hose

7

Before Using This Unit

Installation Procedure

6Precautions about the circulating path

Carefully check the direction of circulation, and connect the hoses properly. Improper connection results in an accident or malfunction of the unit and the circulating path.

Minimize the length of the circulating path. If resistance inside the piping increases, the quantity of circulating fluid decreases, resulting in lower cooling efficiency. For the capacity of the circulating pump, see "Cooling curve, cooling capacity curve (reference data)" on page 39.

Be sure to connect for making a circulating route. Do not connect with water service or gas pipe, etc. It may cause an accident or a failure.

Check the circulation capacity and withstand pressure of the circulating path. Excessive circulation or pressure may result in an accident or malfunction.

Do not connect any powered unit or a unit with a motor to the circulating path. It may cause an accident or failure.

When changing flux, execute the operation slowly. A rapid change of flux may damage the durability of the pump.

If the unit is to be connected to a circulation unit installed in a higher place than it, beware of the backflow of the circulating fluid. If the fluid flows back, it may overflow the water bath of the unit. Add a valve to the circulating path or take other proper measures to prevent a backflow.

7Connecting the power.

Confirm that the earth leakage breaker is turned off, then connect the power plug (BB300/400 type) and the circle terminal (BB600 type) to switchboard or receptacle.

8Pour circulating fluid into the bath.

Check

overflow.

Connect

the

attached

35mm (BB300/400 type)

 

hose for overflow to the overflow port, and

30mm (BB600 type)

 

place

the hose point in a

container etc.

Bath top

 

Prepare a container separately.

 

 

 

 

 

 

Confirm that the drain cock is closed. Pour

Bath

 

the circulating fluid over the level of the

 

 

 

cooling coil.

Set

the circulating

fluid level

Proper water level

 

as follows.

 

 

 

 

 

 

 

 

 

 

Cooling coil fixing plate

Bath capacity

BB300 type: 6,

BB400

 

 

type: 13, BB600 type: 26

outside

if the

Cooling coil

 

Be careful that it

may leak

 

quantity is beyond the capacity.

Go on to the next operation at this state.

Turn on the earth leakage breaker and RUN/STOP key, and circulate the fluid.

In external sealing system connection, open the discharge valve for circulation and circulate to the device of external sealed system to be cooled. After it is stabilized, supply circulating fluid to the proper level shown as below.

After the resupply of the circulating fluid, turn "off" the leakage breaker.

Caution) Slowly pour the circulating oil.

Exercise care not to allow the circulating fluid to get on the unit. If it gets on any electric part, leakage or electric shock may result. If it splashes on the operation panel, wipe it out.

8

Description and Function of Each Part

Main Unit

Front view

Cover

Control panel

Refrigerator error lamp (Red) (Lights when the refrigerator is overloaded.)

Condenser filter

Rear view

Temperature output terminal (RS485 communication function)

Return port

Discharge port

Drain cock

Power cord

Rating notice sticker

Earth leakage breaker

Caster

(Two front casters have stopper)

Refrigerator error lamp (Green) (Lights when the refrigerator is in operation.)

Overflow

BB300/400 BB600

9

Description and Function of Each Part

Control Panel

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

No. Name

RUN/STOP Key

▲▼ Key

SUB MENU Key

ENTER Key

FIXED TEMP Key

TIMER Key

HEATER Lamp

ALARM Lamp

AUTO STOP Lamp

AUTO START Lamp

FIXED TEMP Lamp

Measurement Temperature

Display

Setting Temperature

Display

Overheating Prevention Temperature Display

Function

Starts/stops the operation.

Uses for rising UP/lowering DOWN the setting value.

Uses for setting the overheating prevention temperature, calibration offset temperature, or key lock function.

Settles the inputted value.

Chooses the fixed temperature operation.

Chooses the timer operation (Quick Auto Stop/Auto Stop/Auto Start). Lights while the heater works.

Lights up when an error occurs. (Buzzer sounds simultaneously.)

Blinks while setting quick auto stop timer or auto stop timer. Lights while quick auto stop timer or auto stop timer is running.

Blinks while setting auto start timer. Lights while auto start timer is running.

Blinks while setting fixed temperature operation. Lights while fixed temperature operation is running.

Displays the measured temperature, setting character, alarm information.

Displays the setting temperature, setting value for timer mode, remaining time.

Displays the setting temperature for overheating prevention device.

10

Description and Function of Each Part

Characters of the Controller

The characters controller shows are as follows:

 

Character

Identifier

Name

Purpose

 

 

FiX

Fixed Temperature

Used for setting the fixed temperature

 

 

 

 

Setting Mode

operation.

 

 

 

 

 

 

 

 

 

 

Sv

Temperature Setting

Used for setting the temperature.

 

 

 

 

 

 

 

 

 

 

 

 

 

 

AStP

Auto Stop Setting

Used for setting the auto stop operation.

 

 

 

 

 

 

 

 

 

 

 

 

 

 

AStr

Auto Start Setting

Used for setting the auto start operation.

 

 

 

 

 

 

 

 

 

 

 

 

 

 

tim

Time Setting

Used for setting the time.

 

 

 

 

 

 

 

 

 

 

 

 

 

 

End

Time-up

Displayed when timer operation is ended.

 

 

 

 

 

 

Used for inputting the calibration offset

 

 

 

 

 

 

cAL

Calibration Offset

 

 

 

 

temperature. (Refer to Page 23 "Calibration

 

 

Setting

 

 

 

 

Offset Function".)

 

 

 

 

 

 

oH

Overheating

Used for setting temperature for overheating

 

 

 

 

prevention device. (Refer to Page 15 "Setting of

 

 

Prevention Setting

 

 

 

 

Overheating Prevention Device ".)

 

 

 

 

 

 

LocK

Key Lock

Locks the keys on control panel to protect from

 

 

 

 

unnecessary operation. (Refer to Page 24

 

 

 

 

"Lock Function".)

 

 

 

 

* Also refer to Page 14 "Operation Mode, Function Setting Key, and Characters".

11

Operation Method

Operation Mode and Function List

The operation modes of this unit are as follows;

Name

Description

Page

 

 

 

 

Pressing the FIXED TEMP key enters into the fixed temperature

 

Fixed Temperature

operation setting mode.

 

Pressing it again enters into the temperature setting mode. The "

16

Operation

▲▼" are used to set temperature.

 

 

 

 

Pressing the RUN/STOP key starts or stops operation.

 

 

 

 

 

This operation is used to specify the period up to automatic stop

 

 

during operation.

 

 

The period up to operation stop can be set by pressing the TIMER

 

Quick Auto Stop

key during fixed temperature operation.

18

Operation

The "▲▼" are used to set the time.

 

 

Pressing the START key starts the quick auto stop operation,

 

 

activates the timer function and stops the operation automatically

 

 

after specified period.

 

 

This operation is used to specify the automatic stop time in the fixed

 

 

temperature operation.

 

Auto Stop Operation

Pressing the TIMER key displays "AStP".

 

The setting temperature "Sv" can be set by pressing the ENTER

19

 

key.

 

 

The operation time "tim" can be set by pressing it again.

 

 

Pressing the RUN/STOP key starts the auto stop operation.

 

 

This operation is used to specify the period up to automatic start

 

 

after power on.

 

Auto Start Operation

Pressing the TIMER key displays "AS r".

21

The setting temperature "Sv" can be set by pressing the ENTER

 

key.

 

 

The operation time "tim" can be set by pressing it again.

 

 

Pressing the RUN/STOP key starts the auto start operation.

 

NOTE) This unit is impossible to be changed the mode during the operation. If the mode requires to be changed, stop the operation.

12

Operation Method

Operation Mode and Function List

The operation functions of this unit are as follows;

 

Name

 

 

 

Description

 

 

 

Page

 

Auto overheating

This function is set to be automatically activated (auto

 

 

reset) when

the

temperature

exceeds

the

setting

 

 

prevention function

 

 

temperature by 6 .

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Overheating

 

Though the device shares power source, display, and

 

 

key input with

the

controller,

it

has

independent

15

prevention

 

 

temperature

measurement circuit,

CPU,

sensor and

function

Overheating

 

output circuit.

Overheating prevention temperature can

 

 

 

 

prevention device

be set using the operation panel.

 

 

 

 

 

 

 

The unit stops operation when the device is activated.

 

 

 

The unit starts operation again when the POWER switch

 

 

 

is pressed again (manual reset).

 

 

 

 

 

 

 

This calibration offset function is for calibrating the

 

 

 

difference occurred between the required inbath

 

Calibration offset function

temperature

and

control

temperature

(sensor

23

temperature) of the controller. This unit can be calibrated

 

 

 

 

 

toward either plus side or minus side of the whole

 

 

 

temperature range.

 

 

 

 

 

 

Setting value locking

This function locks the established operation status.

24

It can be set and cancelled with the SUB MENU key.

 

 

 

Temperature Output Terminal

Transmits and outputs the measured temperature of the

25

controller at 4 to 20 mA.

 

 

 

 

 

 

 

 

 

 

 

 

 

The function to allow communication between the VS3

 

 

 

controller and a personal computer or another unit. An

 

RS485 Communication Function

optional RS485-RS232C conversion adapter is required

27

for external communication.

 

 

 

 

 

 

 

 

 

 

 

 

 

A sample program is uploaded on our website.

 

 

 

 

http://www.yamato-net.co.jp/support/program/index.htm

 

13

Operation Method

Operation Mode, Function Setting Key, and Characters

The operation mode setting and function setting use the key operation and characters show in the following figure.

14

Operation Method

Setting of Overheating Prevention Device

The unit has the overheating prevention device (manual reset) that consists of independent temperature measurement circuit, CPU, sensor and output circuit (it shares power source, display, and key input with the controller) in addition to the automatic overheating prevention function (auto reset) in the controller.

Setting range/function

The unit has failsafe functions against overheating. One of them is built in the controller and previously set at factory shipment so to be automatically activated when the temperature exceeds the setting temperature of temperature controller by 6 , where the heater repeats on and off.

The other is united with the controller, which can be set by operating the keys on the controller. The setting range of latter is from 0 to 50 .

In case the temperature in bath exceeds the setting temperature of controller to reach to that of overheating prevention device, the circuit is shut off and "Er19" is displayed with blinking on the screen of controller with buzzer sound.

If the device is once activated,"Er19"continues to be displayed until the power is newly turned on.

Temperature setting procedure

Notes:

1.Turn on the power (turn on the breaker in front)

The default value is displayed for about four seconds after turning on the power. The screen then displays the initial setting. The current temperature in bath, operation mode character and setting temperature of overheating prevention device are displayed on respective screens.

2.Set the temperature for overheating prevention

Press the SUB MENU key.

Press the " ▼▲" several times to select the setting character of overheating prevention temperature "OH".

Press the ENTER key. The current setting temperature is displayed with blinking on the setting temperature screen.

Note: To prevent improper operation, set the value 10 or more over the setting temperature of controller.

Select the value using the " ▼▲"and then press the ENTER key. This completes the setting.

The standard setting temperature of device is "the maximum setting temperature of unit plus 10 " or "setting temperature plus 10 ". If the unit performs improper operation, increase it 5 more.

The setting range of overheating prevention device is from 0 to 50 . Improper setting of temperature may cause inoperative of unit, malfunction of device, e.g. it is activated during increasing in temperature in bath, or unexpected accidents such as fire disaster. To prevent such matters, set a proper value.

The temperature is set to 90 at factory shipment.

The purpose of overheating prevention device is to protect the unit from overheating. It does not intend to protect the samples, or to protect them from the accident caused by the use of explosive or inflammability.

15

Operation Method

Fixed Temperature Operation

In this mode, the unit starts to operate by pressing RUN/STOP key and continues operating at the set temperature until RUN/STOP key is re-pressed, as shown in the figure below.

TEMP

SV::Set Temp.

SV

 

 

TIME

RUN/STOP key:ON

RUN/STOP key:OFF

t

 

 

 

 

Fixed temperature operation procedure

1.Turn on the power (turn on the breaker in front)

Current version of the software is displayed for about four seconds after turning on the power. The screen then displays the initial setting. The current temperature in bath, operation mode character and setting temperature of overheating prevention device are displayed on respective screens.

Measurement temperature screen:

Displays the current temperature in bath.

Setting temperature screen:

Displays the operation mode character. (Refer to Page 13)

Overheating prevention screen:

Displays the setting temperature of overheating prevention device

2.Select the operation mode

Press the FIXED TEMP key to display "FIX", which indicates the fixed temperature operation, on the center display screen.

3.Set the temperature

Press the FIXED TEMP key again.

The setting temperature screen displays the character "Sv" which indicates the temperature setting. Also it displays the current setting temperature with blinking. The FIXED TEMP lamp blinks, too.

Set the temperature by pressing the "▼▲".

The temperature can be set to the first decimal place.

16

Operation Method

Fixed Temperature Operation

4. Start operation

Press the orange RUN/STOP key for about one second. The unit starts operation and the blinking FIXED TEMP lamp lights on.

5. Stop operation

Press the orange RUN/STOP key for about one second. The unit stops operation and the FIXED TEMP lamp lights off. The screen returns to the initial setting screen.

To correct or check setting…

Press the FIXED TEMP key again to correct or check the setting.

Changing the setting temperature during operation is also possible by pressing the FIXED TEMP key. Press the ENTER key after changing the setting.

17