Error Codes

When responding to an error code, always check the patient first and then the device.

Blood Pressure Error Codes

C02 Autozero failure: Check for air obstruction; limit patient movement. C03 Inflation too rapid: Check for kinked hose or air obstruction.

C04 Excessive inflation time: Check for air leaks.

C05 Excessive noise: Check patient condition, cuff placement; limit patient movement. C06 Measurement outside of device’s range: Check patient condition.

E10 Cuff overpressure: Check patient condition.

Temperature Error Codes

C20 Broken/missing probe: Replace probe.

P Loss of tissue contact: Ensure proper probe positioning.

E0.2-E0.3 Ambient temperature out of range: Adjust room temperature or device location. C22 10-minute diagnostic limit exceeded: Check patient. Read the Directions for Use.

Verify Spot Vital Signs is not being used for monitoring purposes.

SpO2 Error Codes

C6

SpO2 heart rate out of range.

C8

Bad sensor: Replace sensor.

E7

Internal malfunction: Contact Technical Service.

C9

10-minute diagnostic limit exceeded: Check patient. Read the Directions for Use.

 

Verify Spot Vital Signs is not being used for monitoring purposes.

General Error Codes

E11 Internal safety violation: Check patient; contact Technical Service.

C12 Ambient temperature out of range: Adjust room temperature or device location. C13 Battery Depletion: Use wall transformer.

E0.0-E9.9 Temperature module malfunction: Contact Technical Service.

E20-E50 General internal malfunction: Contact Technical Service.

Error Codes

When responding to an error code, always check the patient first and then the device.

Blood Pressure Error Codes

C02 Autozero failure: Check for air obstruction; limit patient movement. C03 Inflation too rapid: Check for kinked hose or air obstruction.

C04 Excessive inflation time: Check for air leaks.

C05 Excessive noise: Check patient condition, cuff placement; limit patient movement. C06 Measurement outside of device’s range: Check patient condition.

E10 Cuff overpressure: Check patient condition.

Temperature Error Codes

C20 Broken/missing probe: Replace probe.

P Loss of tissue contact: Ensure proper probe positioning.

E0.2-E0.3 Ambient temperature out of range: Adjust room temperature or device location. C22 10-minute diagnostic limit exceeded: Check patient. Read the Directions for Use.

Verify Spot Vital Signs is not being used for monitoring purposes.

SpO2 Error Codes

C6

SpO2 heart rate out of range.

C8

Bad sensor: Replace sensor.

E7

Internal malfunction: Contact Technical Service.

C9

10-minute diagnostic limit exceeded: Check patient. Read the Directions for Use.

 

Verify Spot Vital Signs is not being used for monitoring purposes.

General Error Codes

E11 Internal safety violation: Check patient; contact Technical Service.

C12 Ambient temperature out of range: Adjust room temperature or device location. C13 Battery Depletion: Use wall transformer.

E0.0-E9.9 Temperature module malfunction: Contact Technical Service.

E20-E50 General internal malfunction: Contact Technical Service.

Error Codes

When responding to an error code, always check the patient first and then the device.

Blood Pressure Error Codes

C02 Autozero failure: Check for air obstruction; limit patient movement. C03 Inflation too rapid: Check for kinked hose or air obstruction.

C04 Excessive inflation time: Check for air leaks.

C05 Excessive noise: Check patient condition, cuff placement; limit patient movement. C06 Measurement outside of device’s range: Check patient condition.

E10 Cuff overpressure: Check patient condition.

Temperature Error Codes

C20 Broken/missing probe: Replace probe.

P Loss of tissue contact: Ensure proper probe positioning.

E0.2-E0.3 Ambient temperature out of range: Adjust room temperature or device location. C22 10-minute diagnostic limit exceeded: Check patient. Read the Directions for Use.

Verify Spot Vital Signs is not being used for monitoring purposes.

SpO2 Error Codes

C6

SpO2 heart rate out of range.

C8

Bad sensor: Replace sensor.

E7

Internal malfunction: Contact Technical Service.

C9

10-minute diagnostic limit exceeded: Check patient. Read the Directions for Use.

 

Verify Spot Vital Signs is not being used for monitoring purposes.

General Error Codes

E11 Internal safety violation: Check patient; contact Technical Service.

C12 Ambient temperature out of range: Adjust room temperature or device location. C13 Battery Depletion: Use wall transformer.

E0.0-E9.9 Temperature module malfunction: Contact Technical Service.

E20-E50 General internal malfunction: Contact Technical Service.

Error Codes

When responding to an error code, always check the patient first and then the device.

Blood Pressure Error Codes

C02 Autozero failure: Check for air obstruction; limit patient movement. C03 Inflation too rapid: Check for kinked hose or air obstruction.

C04 Excessive inflation time: Check for air leaks.

C05 Excessive noise: Check patient condition, cuff placement; limit patient movement. C06 Measurement outside of device’s range: Check patient condition.

E10 Cuff overpressure: Check patient condition.

Temperature Error Codes

C20 Broken/missing probe: Replace probe.

P Loss of tissue contact: Ensure proper probe positioning.

E0.2-E0.3 Ambient temperature out of range: Adjust room temperature or device location. C22 10-minute diagnostic limit exceeded: Check patient. Read the Directions for Use.

Verify Spot Vital Signs is not being used for monitoring purposes.

SpO2 Error Codes

C6

SpO2 heart rate out of range.

C8

Bad sensor: Replace sensor.

E7

Internal malfunction: Contact Technical Service.

C9

10-minute diagnostic limit exceeded: Check patient. Read the Directions for Use.

 

Verify Spot Vital Signs is not being used for monitoring purposes.

General Error Codes

E11 Internal safety violation: Check patient; contact Technical Service.

C12 Ambient temperature out of range: Adjust room temperature or device location. C13 Battery Depletion: Use wall transformer.

E0.0-E9.9 Temperature module malfunction: Contact Technical Service.

E20-E50 General internal malfunction: Contact Technical Service.

For technical questions inside the U.S., please call Welch Allyn at 800-535-6663. Outside the U.S., please contact your local distributor.

Reorder 4200-88E

706287 Ver. B

For technical questions inside the U.S., please call Welch Allyn at 800-535-6663. Outside the U.S., please contact your local distributor.

Reorder 4200-88E

706287 Ver. B

For technical questions inside the U.S., please call Welch Allyn at 800-535-6663. Outside the U.S., please contact your local distributor.

Reorder 4200-88E

706287 Ver. B

For technical questions inside the U.S., please call Welch Allyn at 800-535-6663. Outside the U.S., please contact your local distributor.

Reorder 4200-88E

706287 Ver. B

Error Codes

When responding to an error code, always check the patient first and then the device.

Blood Pressure Error Codes

C02 Autozero failure: Check for air obstruction; limit patient movement. C03 Inflation too rapid: Check for kinked hose or air obstruction.

C04 Excessive inflation time: Check for air leaks.

C05 Excessive noise: Check patient condition, cuff placement; limit patient movement. C06 Measurement outside of device’s range: Check patient condition.

E10 Cuff overpressure: Check patient condition.

Temperature Error Codes

C20 Broken/missing probe: Replace probe.

P Loss of tissue contact: Ensure proper probe positioning.

E0.2-E0.3 Ambient temperature out of range: Adjust room temperature or device location. C22 10-minute diagnostic limit exceeded: Check patient. Read the Directions for Use.

Verify Spot Vital Signs is not being used for monitoring purposes.

SpO2 Error Codes

C6

SpO2 heart rate out of range.

C8

Bad sensor: Replace sensor.

E7

Internal malfunction: Contact Technical Service.

C9

10-minute diagnostic limit exceeded: Check patient. Read the Directions for Use.

 

Verify Spot Vital Signs is not being used for monitoring purposes.

General Error Codes

E11 Internal safety violation: Check patient; contact Technical Service.

C12 Ambient temperature out of range: Adjust room temperature or device location. C13 Battery Depletion: Use wall transformer.

E0.0-E9.9 Temperature module malfunction: Contact Technical Service.

E20-E50 General internal malfunction: Contact Technical Service.

Error Codes

When responding to an error code, always check the patient first and then the device.

Blood Pressure Error Codes

C02 Autozero failure: Check for air obstruction; limit patient movement. C03 Inflation too rapid: Check for kinked hose or air obstruction.

C04 Excessive inflation time: Check for air leaks.

C05 Excessive noise: Check patient condition, cuff placement; limit patient movement. C06 Measurement outside of device’s range: Check patient condition.

E10 Cuff overpressure: Check patient condition.

Temperature Error Codes

C20 Broken/missing probe: Replace probe.

P Loss of tissue contact: Ensure proper probe positioning.

E0.2-E0.3 Ambient temperature out of range: Adjust room temperature or device location. C22 10-minute diagnostic limit exceeded: Check patient. Read the Directions for Use.

Verify Spot Vital Signs is not being used for monitoring purposes.

SpO2 Error Codes

C6

SpO2 heart rate out of range.

C8

Bad sensor: Replace sensor.

E7

Internal malfunction: Contact Technical Service.

C9

10-minute diagnostic limit exceeded: Check patient. Read the Directions for Use.

 

Verify Spot Vital Signs is not being used for monitoring purposes.

General Error Codes

E11 Internal safety violation: Check patient; contact Technical Service.

C12 Ambient temperature out of range: Adjust room temperature or device location. C13 Battery Depletion: Use wall transformer.

E0.0-E9.9 Temperature module malfunction: Contact Technical Service.

E20-E50 General internal malfunction: Contact Technical Service.

Error Codes

When responding to an error code, always check the patient first and then the device.

Blood Pressure Error Codes

C02 Autozero failure: Check for air obstruction; limit patient movement. C03 Inflation too rapid: Check for kinked hose or air obstruction.

C04 Excessive inflation time: Check for air leaks.

C05 Excessive noise: Check patient condition, cuff placement; limit patient movement. C06 Measurement outside of device’s range: Check patient condition.

E10 Cuff overpressure: Check patient condition.

Temperature Error Codes

C20 Broken/missing probe: Replace probe.

P Loss of tissue contact: Ensure proper probe positioning.

E0.2-E0.3 Ambient temperature out of range: Adjust room temperature or device location. C22 10-minute diagnostic limit exceeded: Check patient. Read the Directions for Use.

Verify Spot Vital Signs is not being used for monitoring purposes.

SpO2 Error Codes

C6

SpO2 heart rate out of range.

C8

Bad sensor: Replace sensor.

E7

Internal malfunction: Contact Technical Service.

C9

10-minute diagnostic limit exceeded: Check patient. Read the Directions for Use.

 

Verify Spot Vital Signs is not being used for monitoring purposes.

General Error Codes

E11 Internal safety violation: Check patient; contact Technical Service.

C12 Ambient temperature out of range: Adjust room temperature or device location. C13 Battery Depletion: Use wall transformer.

E0.0-E9.9 Temperature module malfunction: Contact Technical Service.

E20-E50 General internal malfunction: Contact Technical Service.

Error Codes

When responding to an error code, always check the patient first and then the device.

Blood Pressure Error Codes

C02 Autozero failure: Check for air obstruction; limit patient movement. C03 Inflation too rapid: Check for kinked hose or air obstruction.

C04 Excessive inflation time: Check for air leaks.

C05 Excessive noise: Check patient condition, cuff placement; limit patient movement. C06 Measurement outside of device’s range: Check patient condition.

E10 Cuff overpressure: Check patient condition.

Temperature Error Codes

C20 Broken/missing probe: Replace probe.

P Loss of tissue contact: Ensure proper probe positioning.

E0.2-E0.3 Ambient temperature out of range: Adjust room temperature or device location. C22 10-minute diagnostic limit exceeded: Check patient. Read the Directions for Use.

Verify Spot Vital Signs is not being used for monitoring purposes.

SpO2 Error Codes

C6

SpO2 heart rate out of range.

C8

Bad sensor: Replace sensor.

E7

Internal malfunction: Contact Technical Service.

C9

10-minute diagnostic limit exceeded: Check patient. Read the Directions for Use.

 

Verify Spot Vital Signs is not being used for monitoring purposes.

General Error Codes

E11 Internal safety violation: Check patient; contact Technical Service.

C12 Ambient temperature out of range: Adjust room temperature or device location. C13 Battery Depletion: Use wall transformer.

E0.0-E9.9 Temperature module malfunction: Contact Technical Service.

E20-E50 General internal malfunction: Contact Technical Service.

For technical questions inside the U.S., please call Welch Allyn at 800-535-6663. Outside the U.S., please contact your local distributor.

Reorder 4200-88E

For technical questions inside the U.S., please call Welch Allyn at 800-535-6663. Outside the U.S., please contact your local distributor.

Reorder 4200-88E

For technical questions inside the U.S., please call Welch Allyn at 800-535-6663. Outside the U.S., please contact your local distributor.

Reorder 4200-88E

For technical questions inside the U.S., please call Welch Allyn at 800-535-6663. Outside the U.S., please contact your local distributor.

Reorder 4200-88E

706287 Ver. B

706287Ver. B

706287Ver. B

706287 Ver. B

Loading...
+ 1 hidden pages