MANUAL, REF LB24M INTERNATIONAL
English . . . . . . . . 1 Français . . . . . . 31 Español . . . . . . 61 Português . . . . 91 Deutsch . . . . . 121 Italiano . . . . . . 153. . . . . . .183. . . . . . . 211
CaviCide® is a registered trademark of Metrex Research Corporation (800) 841-1428.
Lysol® is a registered trademark of Reckitt Benckiser (800) 677-9218.
Solarc® is a registered trademark of Welch Allyn, Inc. (800) 535-6663.
Table of Contents
Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
Indications. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
Contraindications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
User Profile . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
Symbol Description. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
Warnings. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
Cautions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
System Description and Operation . . . . . . . . . . . . . . . .11
Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .13
Lamp Replacement. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
Cleaning . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
Ordering Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
Troubleshooting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18
Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .19
Electrical . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19
Dimensions. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19
Illumination System . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19
Environment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20
Equipment Classification . . . . . . . . . . . . . . . . . . . . . . . . . . 20
Patents . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20
Approvals . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21
Emissions and Immunity Information . . . . . . . . . . . . . . . . . 22
Warranty and Service . . . . . . . . . . . . . . . . . . . . . . . . . 27
Warranty . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 27
Service . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 28
2 |
Welch Allyn Light Source |
3
This manual describes the setup, installation, start-up, and use of the Welch Allyn Light Source.
Use of the Welch Allyn Light Source is indicated for medical lighting applications requiring a remotely located, high-quality white light. This product must only be used by qualified medical personnel.
The use of the Welch Allyn Light Source is contraindicated for applications where high intensity light might burn or damage tissues; such as for neonate transillumination and ophthalmic procedures.
For further contraindications, the user must also consult the operating manual of the instrument being used with the Light Source in the medical procedure.
This product must only be used by qualified medical personnel.
4 Chapter 1 Introduction
|
Caution: Consult accompanying |
|
documentation |
|
Light; Lighting; Illumination |
|
Intensity |
IPXØ |
Not Protected Against the |
Ingress of Water |
|
|
Type BF Equipment |
Filter In / Filter Out
Lamp On
Storage Humidity
MAX
|
|
ER |
TE |
|
|
|
T |
|
|||
|
IN |
|
|
K |
|
|
|
|
|
CM |
|
C |
LI |
S TE |
D |
US |
|
|
|
3045357
ITS, Nationally Recognized
Testing Laboratory (NRTL) Mark
c
Direct Current
Welch Allyn Light Source
Not AP/APG Equipment
Caution, Hot Surface
Bright Light
Alternating Current
Indoor/dry location use only
Lamp Off
Transport Temperature
UL Recognized Component Mark
US
Double Insulation
Directions for Use |
Chapter 1 Introduction |
5 |
European CE Mark |
EMC Framework of Australia |
|
N344 |
Caution |
Hazard Warning |
Recycling SymbolDo not dispose of this product as unsorted municipal waste. Prepare this product for reuse or separate collection as specified by Directive 2002/96/EC of the European Parliament and the Council of the European Union on Waste Electronic and Electrical Equipment (WEEE). If this product is contaminated, this directive does not apply. See www.welchallyn.com/weee or contact Welch Allyn Customer Service.
6 |
Chapter 1 Introduction |
Welch Allyn Light Source |
A warning statement in this manual identifies a condition or practice which, if not corrected or discontinued immediately, could lead to patient injury, illness or death.
WARNING RISK OF FIRE. Do not operate on, near, or over combustible material. Operate only on smooth, hard, flat surface. This device is not suitable for use in
the presence of a flammable anesthetic mixture with air or oxygen or nitrous oxide. An explosion may result.
WARNING The high intensity light at the front of the light source and at the tip of the fiber optic bundle will create high temperatures and bright light. To minimize the risk of injury, avoid direct viewing or contact with sensitive tissues.
WARNING Do not use for neonate transillumination.
Erythema may result.
WARNING To prevent temporary blinding, always plug the fiber optic bundle into the Welch Allyn Light Source and set the light intensity control to minimum before turning the power ON.
WARNING If this light source is used during invasive surgery, the failure of a component within the system, causing loss of illumination, may indirectly cause risk to the patient. Therefore, it is recommended to have alternative means of illumination available.
WARNING When the unit is on, never disconnect the fiber optic bundle from the light source or the accessory from the end of the fiber bundle. High intensity light from the end of the fiber optic bundle can ignite combustible material.
Directions for Use |
Chapter 1 Introduction |
7 |
WARNING PROVIDE ADEQUATE VENTILATION TO PREVENT OVERHEATING. Do not cover or drape this light source. Provide a 3 inch (7.62 cm) distance between the Welch Allyn Light Source and any solid objects.
WARNING Use only non-electrically conductive fiber bundles. Use of conductive fiber bundles will
compromise the safety and effectiveness of this device.
WARNING RISK OF SHOCK. Do not operate the Light Source without a lamp installed. Disconnect power before attempting to change the lamp.
WARNING This product comes with a power cord that is intended for use only with this product. The power cord has not been tested and approved for use with other products that may have the same power connectors. If you cannot locate the original power cord, please contact Welch Allyn to obtain replacement parts.
8 |
Chapter 1 Introduction |
Welch Allyn Light Source |
A caution statement in this manual identifies a condition or practice which, if not corrected or discontinued immediately, could lead to equipment failure, equipment damage, or data loss.
Caution United States federal law restricts this device to sale by or on the order of a health care practitioner.
Caution Allow the lamp to cool before replacing.
Caution Since the Solarc Light Source uses a custom lamp, always have a spare lamp available for replacement. USE CAUTION BECAUSE LAMP MAY BE HOT. Allow lamp to cool before replacing.
Caution Do not sterilize the Welch Allyn Light Source. Refer to Cleaning section for cleaning instructions.
Caution Do not switch the Welch Allyn Light Source ON and OFF in rapid succession. Doing so will dramatically shorten lamp life. Allow lamp to cool down 10 to 15 seconds between uses.
Caution Do not use or store liquids above or on the
Welch Allyn Light Source.
Caution Do not operate device without the cover in place. Operate in the horizontal orientation only.
Caution Only qualified medical personnel should operate the Light Source.
Caution System integrator is responsible for all safety testing, including optical and thermal testing.
Directions for Use |
Chapter 1 Introduction |
9 |
Caution It is recommended to test dental composites with and without the optional filter prior to use with a patient.
Caution Any changes or modifications made to this device that are not expressly approved by Welch Allyn, Inc. will void the warranty and may compromise product safety.
10 |
Chapter 1 Introduction |
Welch Allyn Light Source |
11
|
2 |
System Description and |
|
|
Operation |
|
|
|
The Welch Allyn Solarc Light Source is used to supply high quality, white light for video and optical illumination systems.
1.Verify the power switch is in the OFF position ( ). Place the light source on a smooth, hard, level surface with at least 3” air clearance between the light source and any solid objects. Plug the power supply outlet connector to the light source power input (see Figure 1) and then plug the power supply into an appropriately rated power outlet (refer to Section 4, Electrical Specifications).
2.Turn the light intensity control knob to the minimum position (see Figure 1). Plug the fiber optic bundle into the light port and connect the distal end to the instrument.
Note For proper lamp performance, verify the fiber optic bundle is plugged in at all times.
3.Turn the power switch to the ON position ().The lamp will take 10 - 15 seconds to complete the warm-up cycle.
4.Set the light intensity knob to the desired level (see Figure 1).
5.Turn the unit OFF when not in use.
12 |
Chapter 2 System Description and Operation |
Welch Allyn Light Source |
.
Caution Do not restart the Solarc Light Source immediately after shutting it off - it may fail to light. If this occurs, turn the power off and wait 10 to 15 seconds before turning the unit back on.
In dental applications, the optional filter blocks light wavelengths that cure some dental composites. For white light, switch the filter out of the light path. When working with dental composites, switch the filter into the light path (Figure 1).
Note It is recommended to test dental composites with and without the optional filter prior to use with a patient.
Figure 1. Front and Rear View
Filter In
(Optional)
Maximum
Filter Out
Light (Optional)
Control
Minimum
Power Switch |
Power Input |
Fan |
13
3 |
Maintenance |
This section provides information to replace a lamp and to clean the Light Source. There are no other serviceable parts within the Light Source.
WARNING RISK OF SHOCK. Do not operate the Light Source without a lamp installed. Disconnect power before attempting to change lamp.
Caution LAMP MAY BE HOT. Allow lamp to cool before handling.
Recommended lamp cooling time is a minimum of 10 minutes.
Caution Do not operate device without the cover in place. Operate in the horizontal position only.
To replace the lamp:
1.Turn the Solarc Light Source off and unplug the power cord from both the wall outlet and the rear of the unit.
14 |
Chapter 3 Maintenance |
Welch Allyn Light Source |
2.Remove the five screws from the Solarc Light Source using a #2 philips head screwdriver. Pivot the cover to expose the lamp.
Figure 2. Moving the Top Cover
Pivot |
Removable Screws (two on opposite side) |
3.Flip up the retaining latch. Pull the lamp base back and up against the retaining spring to remove the lamp.
4.Remove the lamp, with harness attached and then disconnect the lamp connector.
Figure 3. Inside the Light Source
Retaining Latch
Retaining Spring
Lamp Base
Lamp Connector
32
5.Replace with Welch Allyn replacement lamp 09500-U only. DO NOT touch the lamp or the interior reflective surface of the lamp. Skin oils on the lamp will cause premature lamp failure. Only hold the outside of the reflector or the base of
the lamp. Remove any grease or fingerprints with a clean
Directions for Use |
Chapter 3 Maintenance |
15 |
Q-tip moistened with alcohol. Do not leave any lint on the lamp.
6.Connect the new lamp to the connector and insert lamp to properly seat the alignment pin located on the bottom of the lamp into the slot located on the lamp bracket.
Figure 4. Bottom View of the Lamp Slot
Lamp Slot.
7.Close the retaining latch and replace the cover. Secure the five screws on the Solarc Light Source. Reconnect the power supply.
Do not operate the device without the cover in place. Dispose of the lamp according to local, state, and federal laws.
Caution Do not submerge the Solarc Light Source in water, cold steriliants, or in an ultrasonic cleaner.
Caution Do not use an autoclave, chemiclave, glutaraldehyde, iodophor or any other sterilization method, or damage may result.
Clean the Solarc Light Source monthly, or as required, as follows:
1.Prior to cleaning, turn power off and disconnect the power cord from both the wall outlet and the Solarc Light Source.
16 |
Chapter 3 Maintenance |
Welch Allyn Light Source |
2.Clean the external surfaces with a cloth dampened with mild soap and water or use the following cleaning/disinfecting solutions. Always follow the manufacturer’s recommendations.
•A quaternary ammonium disinfectant such as Cavicide®
•A phenol, alcohol-based spray such as Lysol® brand disinfectant
•Gluteraldehyde
•70% isopropyl alcohol
•10% mild bleach solution
•Metracide
•10% Wescodyne
•Banicide
•Wavecide-01
If there are any questions, please contact Customer Service.
3.The power cord may be wiped clean using the same cleaning/ disinfecting solutions listed above. Wipe dry with a clean cloth. Care must be taken not to get the plug prongs wet. DO NOT IMMERSE.
4.Do not plug the unit into a wall outlet until it is thoroughly dry.
Directions for Use Chapter 3 Maintenance 17
LB24MS00-004-00 |
Welch Allyn Light Source with ACMI Adapter |
|
and North American Power Supply |
LB24MS00-004-01 |
Welch Allyn Light Source with ACMI Adapter |
|
and European Power Supply |
Power Supply: |
|
North America: |
LBM-PS-00 |
International: |
LBM-PS-01 |
Replacement Lamp: |
09500-U |
Power Cords: |
|
Europe: |
704904 |
Australia: |
704905 |
UK: |
704906 |
18 |
Chapter 3 Maintenance |
Welch Allyn Light Source |
If lamp is not illuminating and...
Fan Is Not
Operating,
|
|
|
Light Source is not plugged into |
Plug the power cord into the wall |
|
either the wall outlet or the |
outlet and power cord socket at back |
|
power cord socket. |
of unit. |
|
|
|
|
Power switch is not turned on. |
Depress the “ |
“side of the power |
|
switch. |
|
|
|
|
Power cord is damaged. |
Replace the power supply with a new |
|
This may cause intermittent |
Welch Allyn power supply. |
|
operation. |
|
|
|
|
|
No power is supplied to the wall |
Verify that the wall outlet has power |
|
outlet. |
by checking the circuit breaker status |
|
|
and by operating other equipment at |
|
|
that wall outlet. |
|
Fan Is |
The lamp was not able to “cool |
Wait at least 15 seconds after turning |
Operating, |
down” before restarting. |
the Light Source off before turning it |
|
|
back on. |
|
|
|
|
The lamp has burned out. |
Replace with a new Welch Allyn |
|
|
lamp. Refer to the lamp replacement |
|
|
section. |
Low Light |
Lamp is not properly seated in |
Output |
recess. |
Remove the lamp and re-install per the lamp replacement instructions. Verify that the lamp is properly seated.
Light intensity control knob |
Verify that the light intensity control |
closed. |
knob is fully open. |
|
|
19
4 |
Specifications |
Caution Use Welch Allyn power supplies:
•LBM-PS-00
•LBM-PS-01
Power Supply Input: Voltage: 100–240 V~, 50–60 Hz
Power Supply Output: 12V, 3.4A
Current: 1A (@ 100 V~)
Light Box Input: 12V
5.8" (14.73 cm) (L) x 3.5" (8.89 cm) (W) x 2.8" (7.11 cm) (H)
Welch Allyn Solarc Lamp
Lamp: 60 Volts, 24 watts
20 |
Chapter 4 Specifications |
Welch Allyn Light Source |
Environment |
|
|
|
Operating: |
Transport/Storage: |
Temperature: 50° F (10° C) to 86° F (30° C) Humidity: 15 to 95% RH non-condensing
Temperature: -4° F (-20° C) to 120° F (49° C) Humidity: 15 to 95% RH non-condensing Atmospheric pressure: 700 hpa to 1060 hpa
Mode of Operation: Continuous Operation
System Classification: Safety Class II
Type BF Equipment
Ingress of Water: IPXØ
Not AP/APG Equipment
This product is covered by the following patents:
5,144,201; 5,083,059; 5,138,228; 5,117,154; 5,420,477; 5,539,273
Directions for Use |
Chapter 4 Specifications |
21 |
|
|
ERT |
E |
|
|
|
T |
|
|||
|
I |
|
|
K |
|
|
N |
|
|
|
|
|
|
|
|
CM |
|
C |
I |
|
|
D |
|
L |
S TE |
|
US |
||
|
|
3045357
EN/IEC 60601-1
UL60601-1
CAN/CSA C22.2 No. 601.1-M90
IEC60601-1-2
EMC Framework of Australia
The CE mark on this product indicates it has been tested to and conforms with the provisions noted within the 93/42/EEC Medical Device Directive.
European Regulatory Manager
Welch Allyn Ltd.,
Navan Business Park
Dublin Road
Navan, County Meath, Republic of Ireland.
Tel.: +353 46 90 67700
Fax: +353 46 90 67755
22 |
Chapter 4 Specifications |
Welch Allyn Light Source |
Electromagnetic Emissions
The Welch Allyn Light Source is intended for use in the electromagnetic environment specified below. The customer or user of the Welch Allyn Light Source should assure that it is used in such an environment.
Emissions test |
Compliance |
Electromagnetic environment - guidance |
|
|
|
|
|
RF emissions |
Group 1 |
The Welch Allyn Light Source uses RF energy only for its |
|
CISPR 11 |
|
internal function. Therefore, its RF emissions are very low |
|
|
|
and are not likely to cause any interference in nearby |
|
|
|
electronic equipment. |
|
|
|
|
|
RF emissions |
Class B |
The Welch Allyn Light Source is suitable for use in all |
|
CISPR 11 |
|
establishments including domestic and those directly |
|
|
|
connected to the public low-voltage power supply network |
|
Harmonic emissions |
Class A |
||
that supplies buildings used for domestic purposes. |
|||
IEC 61000-3-2 |
|
||
|
|
||
|
|
|
|
Voltage fluctuations/ |
Complies |
|
|
flicker emissions |
|
|
|
IEC 61000-3-3 |
|
|
|
|
|
|
Electromagnetic Immunity
The Welch Allyn Light Source is intended for use in the electromagnetic environment specified below. The customer or user of the Welch Allyn Light Source should assure that it is used in such an environment.
Immunity test |
IEC 60601 |
Compliance |
Electromagnetic environment - |
|
test level |
level |
guidance |
|
|
|
|
Electrostatic |
± 6 kV contact |
± 6 kV contact |
Floors should be wood, concrete, or |
discharge (ESD) |
± 8 kV air |
± 8 kV air |
ceramic tile. If floors are covered with |
IEC 61000-4-2 |
|
|
synthetic material, the relative |
|
|
|
humidity should be at least 30%. |
|
|
|
|
Electrical fast |
±2 kV mains |
±2 kV mains |
Mains power quality should be that of |
transient (EFT)/ |
±1 kV I/O |
Not applicable |
a typical commercial or hospital |
burst |
|
|
environment. |
IEC 61000-4-4 |
|
|
|
|
|
|
|
Directions for Use |
|
|
Chapter 4 Specifications |
23 |
|
|
|
|
|
Surge |
±1 kV differential |
±1 kV differential |
Mains power quality should be that of |
|
IEC 61000-4-5 |
±2 kV common |
±2 kV common |
a typical commercial or hospital |
|
|
|
|
environment. |
|
|
|
|
|
|
Voltage dips, short |
>95% dip in |
>95% dip in |
Mains power quality should be that of |
|
interruptions, and |
0.5 cycle |
0.5 cycle |
a typical commercial or hospital |
|
voltage variations |
|
|
environment. If the user of the Welch |
|
on power supply |
60% dip in |
60% dip in |
Allyn Light Source requires continued |
|
input lines. |
5 cycles |
5 cycles |
operation during power mains |
|
IEC 61000-4-11 |
|
|
interruptions, it is recommended that |
|
|
30% dip for |
30% dip for |
the Welch Allyn Light Source be |
|
|
25 cycles |
25 cycles |
powered from an uninterruptible |
|
|
>95% dip in |
>95% dip in |
power supply or battery. |
|
|
|
|
||
|
5 seconds |
5 seconds |
|
|
|
|
|
|
|
Power frequency |
3 A/m |
3 A/m |
Power frequency magnetic fields |
|
(50/60Hz) |
|
|
should be at levels characteristic of a |
|
magnetic field |
|
|
typical location in a typical |
|
IEC 61000-4-8 |
|
|
commercial or hospital environment. |
|
|
|
|
|
|
Electromagnetic Immunity
The Welch Allyn Light Source is intended for use in the electromagnetic environment specified below. The customer or user of the Welch Allyn Light Source should assure that it is used in such an environment.
Immunity |
IEC 60601 |
Compliance Electromagnetic environment - guidance |
test |
test level |
level |
|
|
|
|
|
Portable and mobile RF communications equipment |
|
|
should be used no closer to any part of the Welch Allyn |
|
|
Light Source, including cables, than the recommended |
|
|
separation distance calculated from the equation |
|
|
applicable to the frequency of the transmitter. |
Recommended separation distance
Conducted |
3 Vrms |
3 Vrms |
d = (1.17) P |
RF, IEC |
150 kHz to |
|
|
61000-4-6 |
80 MHz |
|
|
24 |
Chapter 4 |
Specifications |
|
|
Welch Allyn Light Source |
|||
Radiated |
|
3 V/m |
|
|
3 V/m |
|
d = (1.17) |
P 80 MHz to 800 MHz |
|
|
|
||||||
RF, IEC |
|
80 MHz to |
|
|
|
|
P 800 MHz to 2.5 GHz |
|
61000-4-3 |
2.5 GHz |
|
|
|
|
d = (2.33) |
||
|
|
|
|
|
|
|
where P is the maximum output power rating of the |
|
|
|
|
|
|
|
|
transmitter in watts according to the transmitter |
|
|
|
|
|
|
|
|
manufacturer and d is the recommended separation |
|
|
|
|
|
|
|
|
distance in meters. |
|
|
|
|
|
|
|
|
Field strengths from fixed RF transmitters, as |
|
|
|
|
|
|
|
|
determined by an electromagnetic site survey,a should |
|
|
|
|
|
|
|
|
be less than the compliance level in each frequency |
|
|
|
|
|
|
|
|
range.b |
|
|
|
|
|
|
|
|
Interference may occur in the vicinity of equipment |
|
|
|
|
|
|
|
|
marked with the following symbol: |
|
|
|
|
|
|
|
|
|
|
Note 1: At 80 MHz and 800 MHz, the higher frequency range applies.
Note 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects, and people.
aField strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy. To assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey should be considered. If the measured field strength in the location in which the Welch Allyn Light Source is used exceeds the applicable RF compliance level above, the Welch Allyn Light Source should be observed to verify normal operation. If abnormal performance is observed, additional measures may be necessary, such as reorienting or relocating the Welch Allyn Light Source.
bOver the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m.
Directions for Use |
Chapter 4 Specifications |
25 |
Recommended Separation Distances Between Portable and Mobile RF Communications
Equipment and the Welch Allyn Light Source
The Welch Allyn Light Source is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled. The customer or user of the Welch Allyn Light Source can help prevent electromagnetic interference by maintaining a minimum distance between portable and mobile RF communications equipment (transmitters) and the Welch Allyn Light Source as recommended below, according to the maximum output power of the communications equipment.
Rated max. |
Separation distance according to frequency of transmitter (m) |
|||
output power |
|
|
|
|
of transmitter |
150 kHz to 80 MHz |
80 MHz to 800 MHz |
800 MHz to 2.5 GHz |
|
d = (1.17) P |
d = (1.17) P |
d = (2.33) P |
||
(W) |
||||
|
|
|
|
|
0.01 |
0.117 |
0.117 |
0.233 |
|
|
|
|
|
|
0.1 |
0.37 |
0.37 |
0.74 |
|
|
|
|
|
|
1 |
1.17 |
1.17 |
2.33 |
|
|
|
|
|
|
10 |
3.70 |
3.70 |
7.37 |
|
|
|
|
|
|
100 |
11.70 |
11.70 |
23.30 |
|
|
|
|
|
For transmitters rated at a maximum output power not listed above, the recommended separation distance d in meters (m) can be estimated using the equation applicable to the frequency of the transmitter, where P is the maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer.
Note 1: At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
Note 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects, and people.
26 |
Chapter 4 Specifications |
Welch Allyn Light Source |
27
5 |
Warranty and Service |
Welch Allyn warrants the Welch Allyn Light Source when new, to be free of defects in material and workmanship and to perform in accordance with manufacturer’s specifications when subject to normal use and service for a period of one year from the date of purchase from Welch Allyn or an authorized agent. Welch Allyn will either repair or replace any components found to be defective or at variance from manufacturer’s specifications within this time at no cost to the customer. It shall be the purchaser’s responsibility to return the instrument directly to the authorized distributor, agent, or service representative.
This warranty does not cover the Solarc Lamp or breakage or failure due to tampering, misuse, neglect, accidents, improper installation, modification, shipping, or to improper maintenance, service, and cleaning procedures. This warranty is also void if the instrument is not used in accordance with manufacturer’s recommendations or if required or serviced by other than Welch Allyn or an authorized agent. Purchase date determines warranty requirements. No other express or implied warranty
is given.
28 |
Chapter 5 Warranty and Service |
Welch Allyn Light Source |
If you have an equipment problem that you cannot resolve, you may call the Welch Allyn Service Center nearest you for assistance. Technical service support is available by telephone on normal business days at the phone numbers listed below.
If you are advised to return a product to Welch Allyn for repair or routine maintenance, schedule the repair with the service center nearest you.
Before returning a product for repair you must obtain authorization from Welch Allyn. An RMA (Return MerChandise Authorization) number will be given to you by our service personnel. Be sure to note this number on the outside of your shipping box. Returns without an RMA number will not be accepted.
USA Customers |
Canada Customers |
Welch Allyn, Inc., U.S.A. |
Welch Allyn, Ltd., Canada |
Tel: 1-315-685-4347 |
Tel: 905-890-0004 |
Fax: 315-685-2854 |
Fax: 905-890-0008 |
Welch Allyn, GmbH & Co KG |
Welch Allyn, Ltd., Singapore |
Tel: 00-49-7477-927173 |
Tel: 011-656-291-0882 |
Fax: 00-49-7477-927293 |
Fax: 011-656-291-5780 |
Welch Allyn France SARL |
Welch Allyn, China |
Tel: (011) 33 1.6009.3366 |
Tel: 011-86-21-63279631 |
Fax: (011) 33 1.6009.6797 |
Fax: 011-86-21-63279632 |
Welch Allyn, Ltd., Australia |
Welch Allyn, Japan |
Tel: 011-61-294-183-155 |
Tel: 011-813-5212-7391 |
Fax: 011-61-294-183-650 |
Fax: 011-813-3261-7372 |
Welch Allyn UK Ltd |
Latin American Customers |
Tel: 0207-365-6780 |
MD International |
Fax: 0207-365-9694 |
Tel: 1-305-669-9003 |
|
Fax: 1-305-669-1971 |