Welch Allyn ProBP 2400 User Manual

Welch Allyn ProBP 2400 digital blood pressure device
EN
1
ProBP 2400 Digital Blood Pressure Device
Directions for use
AdvancingFrontline Care
CAUTION Federal law restricts this device to sale by or on the order of a physician or licensed healthcare provider
2
Introduction
This directions for use manual is a comprehensive guide designed to help you understand the capabilities and operation
of the ProBP 2400 digital blood pressure device. Read this manual thoroughly before attempting to setup, congure, use,
troubleshoot, or maintain the device.
Intended use
The ProBP 2400 is a non-invasive digital blood pressure device using oscillometric technique and an upper-arm blood pressure cuff to measure systolic and diastolic blood pressures, pulse rate and mean arterial pressure (MAP).
The device detects the appearance of irregular heartbeat during measurement, and gives a warning signal with the reading once the irregular heartbeat is detected.
The device can accurately measure blood pressure in pregnant patients including those with known or suspected preeclampsia.
The ProBP 2400 combines the advantages of an automatic blood pressure device and auscultatory sphygmomanometer. It
is designed to provide convenient, accurate and reliable ofce blood pressure measurements according to guidelines of the
European Society of Hypertension (ESH)
1
, American Heart Association (AHA)2, and World Health Organization (WHO)3 with
the only modication that the ProBP2400 performs 3 repeated measurements always, regardless of the result of the rst two
measurements.
EN
3
Welch Allyn ProBP 2400 digital blood pressure device
1
Pickering TG, Hall JE, Appel LJ, et al. Recommendations for blood pressure measurement in humans and experimental animals: part 1: blood pressure
measurement in humans: a statement for professionals from the subcommittee of professional and public education of the american heart association council on high blood pressure research. Circulation 2005;111:697-716.
2
Whitworth JA. 2003 World Health Organization (WHO)/International Society of Hypertension (ISH) statement on management of hypertension. J Hypertens
2003;21:1983-92.
3
O’Brien E, Asmar R, Beilin L, et al. Practice guidelines of the European Society of Hypertension for clinic, ambulatory and self blood pressure measurement. J Hypertens 2005;23:697-701. E, Asmar R, Beilin L, Imai Y, et al. European Society of Hypertension recommendations for conventional, ambulatory and home blood pressure measurement. European Society of Hypertension Working Group on Blood Pressure Monitoring. J Hypertens 2005;23:697-
701.2003,21:1983-1992.
Pregnancy
Approximately 20% of women develop hypertension during pregnancy (Pre-eclampsia or Toxemia) which may affect pregnancy. Pre-eclampsia, can usually be recognized by a clear increase in blood pressure and high urine protein levels after 20 weeks of gestation. Since many oscillometric devices appeared to be unsuitable for use in pregnancy and Pre-eclampsia
health care authorities require that blood pressure monitors used for this vulnerable patient group are specically tested.
The Welch Allyn ProBP 2400 digital blood pressure monitor has successfully passed this validation and therefore may be recommended for use during pregnancy and Pre-eclampsia.
4
EN
5
Welch Allyn ProBP 2400 digital blood pressure device
Symbols .................................................................................. 6
Warnings and cautions
General warnings and cautions .............................................. 7-8
NIBP (non-invasive blood pressure) warnings and cautions ...... 9
Product description
Name of parts and display .................................................. 10-11
Components and accessories ................................................. 12
Before using ProBP 2400
Attaching the power plug to the power adapter ........................ 13
Selecting Units of Measure ...................................................... 14
Selecting the correct cuff ......................................................... 15
Fitting the cuff properly ....................................................... 16-17
Taking measurements in Single Reading (1x), Manual and Three
Reading Average (
) Mode
«1x» Mode .............................................................................. 18
«Manual» Mode ................................................................. 19-20
«
» Mode .............................................................................. 21
Special Functions
MAP (Mean Arterial Pressure) ................................................. 22
Irregular heartbeat detector in «1x» Mode ............................... 22
Setting maximum ination pressure ........................................ 23
Taking fewer than three measurements .................................. 24
Skipping the countdown time ................................................... 24
Setting measurement interval times in «
» Mode ................... 25
Viewing the stored values .................................................... 26
Appendix
Rechargeable battery and power adapter ............................... 27
Troubleshooting ...................................................................... 28
Error messages.................................................................. 29-30
Safety, care, accuracy test and disposal ............................. 31-34
Technical specications .......................................................... 35
Table of Contents
6
Symbols
Documentation symbols
WARNING: The warning statements in this manual identify conditions or
practices that could lead to illness, injury, or death.
CAUTION: The caution statements in this manual identify conditions or
practices that could result in damage to the equipment or other property, or loss of data.
Consult operating instructions/directions for use (DFU). A copy of the
DFU is available on this web site. A printed copy of the DFU can be ordered from Welch Allyn for delivery within 7 calendar days.
F
Helpful notes.
Shipping, storing, and environment symbols
Fragile; handle with care
Recovery/Recyclable
Recycle the product separate from other disposables
Keep dry
Control symbols
Power on/Power off
Start/Stop
Memory
Adjust measurement interval times
Adjust maximum ination pressure
1x
Single Reading Mode
Manual Mode
Three Reading Average Mode
EN
7
Welch Allyn ProBP 2400 digital blood pressure device
WARNING The information in this directions for use is a comprehensive guide to the
operation of ProBP 2400. For best results, read this directions for use thoroughly before using the device.
WARNING The device is intended for use only in environments with clinician
supervision.
WARNING The device is designed for medical clinician use. Although this directions
for use may illustrate medical spot-check techniques, only a trained clinician should use this device.
WARNING The device is not intended for use during patient transport.
WARNING Fire and explosion hazard. Do not operate the device in the presence
of a ammable anesthetic mixture with air, oxygen, or nitrous oxide; in oxygen
enriched environments.
WARNING Every three months, inspect the blood pressure cuff and other
accessories for fraying or other damage. Replace as necessary.
WARNING Inaccurate measurement risk. Do not use the device on patients who
are connected to heart/lung machines.
WARNING Electric shock hazard. Do not open the device or attempt repairs. There
are no user-serviceable parts inside ProBP 2400 other than battery replacement.
Only perform routine cleaning and maintenance procedures specically
described in this directions for use. Inspection and servicing of internal parts shall
only be performed by qualied service personnel.
WARNING The device complies with applicable domestic and international
standards for electromagnetic interference and should not present problems to other equipment or be affected by other devices. As a precaution, avoid using the device in close proximity to other equipment.
General warnings and cautions
Miscellaneous symbols
Meets essential requirements of European Medical Device Directive
93/42/EEC
Regulatory Affairs Representative
Type BF applied parts
Manufacturer
8
WARNING Welch Allyn is not responsible for the integrity of any mounting
installation. Welch Allyn recommends that customers contact their Biomedical Engineering Department or maintenance service to ensure professional installation for safety and reliability of any mounting accessory.
WARNING The device may not function properly if dropped or damaged. Do not use
the device if you notice any signs of damage. Qualied service personnel must
check any device that is dropped or damaged for proper operation before putting the device back into use.
WARNING Defective batteries can damage the device. If the battery shows any
signs of damage, leakage, or cracking, it must be replaced immediately, and only with a battery recommended for or supplied with the device.
WARNING Improper disposal of batteries may create an explosion or contamination
hazard. Never dispose of batteries in refuse containers. Do not dispose of the
battery in re. Always recycle batteries according to local regulations.
WARNING Improper handling of the battery can lead to heat generation, smoke,
bursting, or re.
WARNING Do not disassemble, modify, or solder the battery.
WARNING For proper patient electrical isolation and battery charging, use only the
provided external power supply to charge the device.
WARNING Electric shock hazard. Before cleaning the device, disconnect the power
cord from the power source and the device.
WARNING Take care to prevent water or other uid from entering any connectors
on the device. Should this occur, dry the connectors with warm air. Check the accuracy of all operating functions.
Avoid simultaneously connecting patients to the device and high frequency surgical equipment.
CAUTION The device is not heat-resistant. Do not autoclave.
CAUTION Use the device within stated operating temperature ranges. The device
will not meet performance specications if used outside these temperatures
ranges.
CAUTION Always unplug the external power source from the outlet before moving
the device to a new location.
CAUTION Use only Welch Allyn approved accessories. Use of unapproved
accessories with the device can affect patient and operator safety, and can reduce product performance and accuracy.
CAUTION Federal law restricts this device to sale by or on the order of a physician or
licensed health care provider
EN
9
Welch Allyn ProBP 2400 digital blood pressure device
NIBP (non-invasive blood pressure) warnings and cautions
WARNING ProBP 2400 is not intended to measure blood pressure on children
younger than 12 years of age.
WARNING Do not compress the blood pressure hose or cuff. This may cause system
errors or patient safety risks to occur.
WARNING Inaccurate measurement risk. Do not use the device on patients who are
experiencing convulsions or tremors.
WARNING Use only blood pressure cuffs and hoses listed as approved accessories
to ensure safe and accurate of blood pressure measurements.
WARNING Patient injury risk. When several blood pressure measurements are
taken on the same patient, regularly check the cuff site and extremity for possible ischemia, purpura, and/or neuropathy.
WARNING Do not allow a blood pressure cuff to remain on the patient more than
5 minutes when inated above 15 mmHg. Excessive cuff tightness may cause
venous congestion, peripheral nerve injury, discoloration of the limb, and patient distress.
WARNING Patient injury risk. Never install Luer Lock connectors on Welch Allyn
blood pressure tubing. Using these connectors on blood pressure cuff tubing creates the risk of mistakenly connecting this tubing to a patient’s intravenous line and introducing air into the patient’s circulatory system.
WARNING NIBP measurements may be inaccurate in the presence of excessive
motion artefact. Minimize extremity and cuff motion during blood pressure readings.
WARNING The position and physiologic condition of the subject can affect a blood
pressure reading.
CAUTION If the blood pressure cuff is not at heart level, note the difference in reading
due to the hydrostatic effect.
CAUTION Proper blood pressure cuff size and placement is essential to the accuracy
of the blood pressure determination. See Blood pressure cuff selection for sizing information.
CAUTION The blood pressure cuff must be properly positioned to ensure blood
pressure accuracy and patient safety. Wrapping the cuff too loosely (preventing
proper ination) may result in inaccurate blood pressure readings.
CAUTION The patients should be comfortably seated, legs uncrossed, feet at on
the oor, back and arm supported. The middle of the cuff should be at the level of
the heart.
CAUTION Recommend the patients to sit down and relax for at least 5 minutes prior
to the measurement and not talk during the measurement procedure.
10
3 sec.1x 3x3 sec.
On/Off Switch
Power Adapter Socket
Battery Compartment
Cuff Socket
Recharge Indicator
Display
M Button (Memory)
Mode Switch
Start/Stop Button
Product description
Name of parts
EN
11
Welch Allyn ProBP 2400 digital blood pressure device
EN
Display
1x
3x
260 240 220 200 180 160 140 AUTO
34.7
32.0
29.3
26.7
24.0
21.3
18.7 AUTO
3x
M1 M2 M3
PAM MAP
/ MIN
SYS
DIA
mmHg kPa
M
Systolic Value
Diastolic Value
Memory
Mean Arterial Pressure
Pulse Rate
Pulse Detected
Mode Selection:
Single Reading,
Manual, Three Reading Average
Power
Low Battery
Unit of Measure
Auto Inflation
Maximum Inflation Pressure
Rest/Count Down
Irregular Heart Beat (IHB)
Power
Low Battery
Mode Selection:
Single Reading,
Manual, Three Reading Average
Unit of Measure
Maximum Ination Pressure
Rest/Count Down
Auto Ination
Memory
Systolic Value
Diastolic Value
Mean Arterial Pressure
Irregular Hear Beat (IHB)
Pulse Rate
Pulse Detected
12
Microlife Corporation 9F, 431, RuiGuang Road, NeiHu, Taipei, 11492, Taiwan, R.O.C.
Microlife AG Espenstrasse 139 9443 Widnau, Switzerland
Distributed by Welch Allyn Inc. 4341 State Street Road Skaneateles Falls, NY 13153
Welch Allyn
ProBP
TM
2400
digital blood pressure device
Directions for use
Version 1.01
English, , Deutsch, Español, Português, Français, Italiano,
Nederlands, Dansk, Norsk, Svenska, Suomi, Русский, Polski,
Türkçe, 한국어
Reusable ProBP 2400 Blood Pressure Cuff
22-32cm
ADULT
ProBP 2400 components and accessories
1 X ProBP 2400
1 X Adult (22cm~32cm)
1 X Large adult (32cm~42cm)
1 X CD Directions For Use
3 sec.1x 3x3 sec.
1 X Power Adapter
4 X Power Plugs (US, Europe, UK, Australia)
(Input: 100-240V~50/60Hz 0.48A Output: +7.5V 1.5A)