Relay PCA
P/N 04523
Addendum
to M100E, M200E and M400E Operator Manuals
© TELEDYNE INSTRUMENTS ADVANCED POLLUTION INSTRUMENTATION
(T-API)
9480 CARROLL PARK DRIVE
SAN DIEGO, CA 92121-5201
USA
Toll-free Phone: 800-324-5190
Phone: 858-657-9800
Fax: 858-657-9816
Email: api-sales@teledyne.com
Website: http://www.teledyne-api.com/
|
05118 Rev B3 |
|
DCN 5362 |
Copyright 2007 T-API Inc. |
20 March 2009 |
THIS PAGE INTENTIONALLY BLANK
Relay PCA 04523 |
Addendum to E-Series Operator Manuals |
SAFETY MESSAGES
For general safety information please refer to your M100E/M200E/M400E operator’s manual.
CAUTION
Never disconnect electronic circuit boards, wiring harnesses or electronic subassemblies while the unit is under power.
Printed circuit assemblies (PCAs) are sensitive to electro-static discharges too small to be felt by the human nervous system. Failure to use ESD protection when working with electronic assemblies will void the instrument warranty.
See Chapter 12 of your M100E/M200E/M400E operator’s manual for more information on preventing ESD damage.
NOTE
Technical Assistance regarding the use and maintenance of the any Teledyne Instruments product can be obtained by:
Contacting Teledyne Instruments’ Customer Service Department at 800-324-5190
or
Via the internet at http://www.teledyne-api.com/inquiries.asp
05118 Rev B3 |
i |
Addendum to E-Series Operator Manuals |
Relay PCA 04523 |
THIS PAGE INTENTIONALLY BLANK
ii |
05118 Rev B3 |
Relay PCA 04523 |
Addendum to E-Series Operator Manuals |
|
TABLE OF CONTENTS |
|
|
1. INTRODUCTION ........................................................................................................................... |
|
1 |
1.1. What’s the Same..................................................................................................................... |
|
1 |
1.2. What’s Different ...................................................................................................................... |
|
2 |
1.3. Relay PCA Location and Layout.................................................................................................. |
|
2 |
1.3.1. Version 045230200 ........................................................................................................... |
|
3 |
1.3.2. Version 045230100 ........................................................................................................... |
|
3 |
2. OPERATION AND CONFIGURATION.............................................................................................. |
|
5 |
2.1. Power Distribution ................................................................................................................... |
|
5 |
2.1.1. AC Power Configuration ..................................................................................................... |
|
5 |
2.1.1.1. AC configuration – Internal Pump (JP7).......................................................................... |
|
7 |
2.1.1.2. AC Configuration – Standard Heaters (JP2)..................................................................... |
|
8 |
2.1.1.3. AC Configuration –Heaters for Option Packages (JP6)....................................................... |
9 |
|
2.2. Valve Control.......................................................................................................................... |
|
9 |
2.3. Status LEDs & Watch Dog Circuitry .......................................................................................... |
|
10 |
2.3.1. Watchdog Indicator (D1).................................................................................................. |
|
10 |
2.4. Heater Control Loop............................................................................................................... |
|
11 |
2.4.1. Thermocouple Inputs and Configuration Jumper (JP5) .......................................................... |
12 |
|
2.5. DC Power Supply Test Points................................................................................................... |
|
14 |
3. TROUBLE SHOOTING AND REPAIR.............................................................................................. |
|
15 |
3.1. Removing / Replacing the Relay PCA from the Instrument ........................................................... |
15 |
|
4. SCHEMATICS AND SPARE PARTS FOR RELAY PCA P/N 04523 ..................................................... |
16 |
|
LIST OF FIGURES |
|
|
Figure 1-1: Relay PCA Layout (P/N04523).......................................................................... |
|
2 |
Figure 1-2: Relay PCA P/N 045230200 with AC Relay Retainer In Place .................................. |
3 |
|
Figure 1-3: Relay PCA P/N 045230100 with Safety Shield In Place......................................... |
3 |
|
Figure 2-1: Power Distribution Block Diagram..................................................................... |
|
5 |
Figure 2-2: Location of AC power Configuration Jumpers...................................................... |
6 |
|
Figure 2-3: Pump AC Power Jumpers (JP7) ........................................................................ |
|
7 |
Figure 2-4: Typical Set Up of AC Heater Jumper Set (JP2).................................................... |
8 |
|
Figure 2-5: Typical Set Up of AC Heater Jumper Set (JP2).................................................... |
9 |
|
Figure 2-6: Status LED Locations – Relay PCA 04523......................................................... |
|
10 |
Figure 2-7: Heater Control Loop Block Diagram. ............................................................... |
|
11 |
Figure 2-8: Thermocouple Configuration Jumper (JP5) Pin-Outs .......................................... |
13 |
|
Figure 3-1: Relay PCA 04523 Mounting Screw Locations .................................................... |
15 |
|
LIST OF TABLES |
|
|
Table 1-1: List of Analyzers Using New Relay PCA P/N04523 ................................................ |
1 |
|
Table 2-1: AC Power Configuration for Internal Pumps (JP7)................................................. |
7 |
|
Table 2-2: Power Configuration for Standard AC Heaters (JP2) ............................................. |
8 |
|
Table 2-3: Power Configuration for Optional AC Heaters (JP6)............................................... |
9 |
|
Table 2-4: Thermocouple Configuration Jumper (JP5) Pin-Outs ........................................... |
12 |
|
Table 2-5: Typical Thermocouple Settings for M100E/M200E series analyzers ....................... |
13 |
|
Table 2-5: DC Power Test Point and Wiring Color Code ...................................................... |
|
14 |
Table 2-6: DC Power Supply Acceptable Levels................................................................. |
|
14 |
Table 4-1: List of Spare Parts......................................................................................... |
|
16 |
Table 4-2: List of Included Electronic Schematics.............................................................. |
|
16 |
05118 Rev B3 |
i |
Addendum to E-Series Operator Manuals |
Relay PCA 04523 |
THIS PAGE INTENTIONALLY BLANK
ii |
05118 Rev B3 |
Relay PCA 04523 |
Addendum to E-Series Operator Manuals |
This addendum documents a recent upgrade to the relay card installed in some of the M100E, M200E and M400E series analyzers. Specifically, a new relay PCA design, P/N 04523 replaces P/N 03955 and all of its versions in the following instruments.
It also corrects several errors related to the function of the relay PCA.
Table 1-1: List of Analyzers Using New Relay PCA P/N04523
PRODUCT FAMILY |
MODEL |
RELAY PCA VERSION1 |
||
045230100 |
045230200 |
|||
|
|
|||
|
M100E |
|
|
|
|
|
|
|
|
UV Fluorescence |
M100EH |
|
|
|
|
|
|
||
M101E |
|
|
||
|
|
|||
|
M102E |
|
|
|
|
M200E |
|
|
|
|
|
|
|
|
Chemiluminescence |
M200EH |
|
|
|
|
|
|
||
M200EM |
|
|
||
|
|
|||
|
M201E |
|
|
|
UV Absorption |
M400E |
|
|
|
Analyzers |
|
|||
|
|
|
||
1 See Sections 1.3.1 and |
1.3.2 |
|
|
For those instruments, the information included in this addendum supersedes the information related to the relay PCA that appears in the chapters on Theory of Operation and Troubleshooting & Repair of your M100E/M200E/M400E operator’s manual’s.
Because the operator’s manual for each model family of the E-Series instruments is slightly different there is no direct correspondence between the sections and chapters of this addendum and those of the manual.
In most respects the electronic operation of the Relay PCA 04523 is the same as Relay PCA 03955.
Both are used by the CPU to operate various AC and DC heaters that control the temperature of key components of the E-Series analyzers to such as sensor modules, reaction cells and converters as well as certain optional components such as IZS permeation tubes.
In both cases a set of configuration jumpers located on the relay card is used to configure the heaters for 110 VAC, 60 Hz vs. 230 VAC, 50 Hz operation.
Both Relay cards, based on commands from the CPU received over the I2C bus, energize/de-energize the various solenoid valves used to control the sample gas path of the analyzers (e.g. reference/measure valves; zero-span-cal valve options; IZS options; etc.)
Both distribute the AC and DC power to the various electrically powered components and assemblies for the instrument.
Both include a special watchdog circuit watches that will automatically shut of all valves as
well as turn off the UV Source(s) and all heaters should either the CPU cease issuing commands or the I2C bus fail. In both cases the sample pump will continue to run.
05118 Rev B3 |