Siemens Mammomat 1000 Service Program

4.7 (7)

MAMMOMAT 1000/3000

Wiring Diagrams

 

 

 

 

 

 

 

 

 

 

 

 

© Siemens AB 1999

 

 

The reproduction, transmission or

 

 

use of this document or its contents

 

 

is not permitted without express

 

 

written authority. Offenders will be

 

 

liable for damages.

All rights,

 

 

including rights created by patent

 

 

grant or registration of a utility

 

 

model _or_ design,_are_ reserved.

 

Register 6

English

 

 

 

Print No.: SPB7-230.051.05.02.02

Doc. Gen. Date:

10.99

 

Replaces: SPB7-230.051.05.01.02

65 27 258

 

 

0 - 2

 

 

Revision

 

 

 

 

Chapter

Page

Revision

 

 

 

 

 

1

2

02

 

 

 

 

 

2

2, 4-9, 11-12

02

 

 

 

 

 

3

1-2, 11

02

 

 

 

 

 

5

2

02

 

 

 

 

 

MAMMOMAT 1000/3000

Register 6

Drucknr_kurz_V

Page 2 of 4

Siemens-Elema AB

Manual_V

 

Rev. _V 10.99

SPS-UD

Solna, Sweden

Contents

0 - 3

 

Page

This document is valid from serial No. 6500

 

1 _______System Overview ______________________________________________ 1 - 1

System Overview. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 - 1 MAMMOMAT 1000 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 - 1 MAMMOMAT 1000 WITH REMOTE CONTROL . . . . . . . . . . . . . . . . . 1 - 2 MAMMOMAT 3000 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 - 3

2 _______Stand Overview _______________________________________________

2 - 1

Placement of PC-boards and Components, Stand

. . . . . . . . . . . . . . . . . . 2 - 1

MAMMOMAT 1000 . . . . . . . . . . . . . .

. . . . . . . . . . . . . . . . . . 2 - 1

MAMMOMAT 3000 . . . . . . . . . . . . . . .

. . . . . . . . . . . . . . . . . . 2 - 2

List of Boards and Fuses, Stand . . . . . . . . .

. . . . . . . . . . . . . . . . . . 2 - 4

PCB Overview . . . . . . . . . . . . . . . . .

. . . . . . . . . . . . . . . . . . 2 - 4

Fuses Overview . . . . . . . . . . . . . . . .

. . . . . . . . . . . . . . . . . . 2 - 4

List of Switches and Components, Stand . . . . .

. . . . . . . . . . . . . . . . . . 2 - 5

Switches . . . . . . . . . . . . . . . . . . . .

. . . . . . . . . . . . . . . . .

2 - 5

Potentiometers . . . . . . . . . . . . . . . . .

. . . . . . . . . . . . . . . . . . 2

- 6

Motors . . . . . . . . . . . . . . . . . . . . .

. . . . . . . . . . . . . . . . . . 2

- 6

Miscellaneous . . . . . . . . . . . . . . . . .

. . . . . . . . . . . . . . . . . . 2

- 6

List of Signals and Test Points, Stand . . . . . . .

. . . . . . . . . . . . . . . . . . 2

- 8

List of Voltages and Test Points, Stand . . . . . .

. . . . . . . . . . . . . . . . . 2 - 12

3 _______Stand Functional Diagrams ______________________________________ 3 - 1

Block Diagram Cable Connections. . . . . . . . . . . . . . . . . . . . . . . . . . . 3 - 1 Grounding Connections . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3 - 2 Mains Connection, Power Supply CPU . . . . . . . . . . . . . . . . . . . . . . . 3 - 3 Grounding Connections . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3 - 3 Power Supply Motor Control . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3 - 4 Display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3 - 5

CPU Board Internal, Generator Interface . . . . . . . . . . . . . . . . . . . . . .

. 3

- 6

Tube, Temperature Monitoring . . . . . . . . . . . . . . . . . . . . . . . . . . .

. 3

- 7

Lift and Rotation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

 

3

- 8

Lift and Rotation Motor Drive . . . . . . . . . . . . . . . . . . . . . . . . . . . .

. 3

- 9

Lift and Rotation Switches . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3 - 10

Lift and Rotation Switches, M1000 with Remote Control . . . . . . . . . . . . . .

3

- 11

Compression Measurement Signals . . . . . . . . . . . . . . . . . . . . . . . . . 3 - 12

Compression Motor Drive . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

3 - 13

Tube Angle Measurement, Pot. Return,Tilt Signal . . . . . . . . . . . . . . . . .

3 - 14

Collimator 1 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

3

- 15

Collimator 2 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

3

- 16

Siemens-Elema AB

Register 6

SP B7-230.051.05

Page 3 of 4

Mammomat 1000/3000

Solna, Sweden

 

Rev. 02 10.99

SPS-UD

Wiring Diagrams

0 - 4

 

 

 

Contents

 

 

 

 

 

 

 

 

Page

Object Table 1 + 2 . . . . . . . . . . . . . . . . . .

. . . . .

. . . .

.

.

. .

.

.

.3 - 17

Automatic Exposure Control (AEC) . . . . . . . . . .

. . . . .

. . . .

.

.

. .

.

.

.3 - 18

4 ______ Generator Overview ____________________________________________ 4 - 1

Mains Voltage Connection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

4

- 1

List of PC-boards and Components, Generator . . . . . . . . . . . . . . . . . . .

4 - 2

Subassembly and PC board overview/Baugruppen und Platinenübersicht . . . .

4 - 2

Control Panel Functions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

4

- 3

Fuses . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

4

- 3

List of Switches and Components, Generator . . . . . . . . . . . . . . . . . . . .

4 - 4

Control Panel Functions. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 - 5

List of Switches and Components, Generator . . . . . . . . . . . . . . . . . . . .

4 - 6

Switches . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 - 6

Potentiometers . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

4

- 6

Component Overview (Generator)/Bauteileübersicht (Generator) . . . . . . . .

4 - 7

List of Signals and Test Points, Generator . . . . . . . . . . . . . . . . . . . . . .

4

- 8

Signal Diagram . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4 - 14

5 ______ General Functional Diagrams_____________________________________ 5 - 1

Block Diagram . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5 - 1

Ground, 0V and Protection Ground Distribution. . . . . . . . . . . . . . . . . . 5 - 2

Line Input. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5 - 3 Back Plane D700 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5 - 4 Principle Diagram, Master . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5 - 5

Principle Diagram, Tube Filament . . . . . . . . . . . . . . . . . . . . . . . . 5 - 6 Principle Diagram, Focus Selection Board . . . . . . . . . . . . . . . . . . . . 5 - 7 Principle Diagram, Inverter Control and kV Regulation . . . . . . . . . . . . . . 5 - 8

Principle Diagram, Inverter . . . . . . . . . . . . . . . . . . . . . . . . . . . 5 - 9

Principle Diagram, Control Panel . . . . . . . . . . . . . . . . . . . . . . . .5 - 10

Principle Diagram, Iontomat . . . . . . . . . . . . . . . . . . . . . . . . . . .5 - 11

6 ______ Generator PC Board Overview ____________________________________ 6 - 1

Contact Location . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6 - 1

Mammomat 1000/3000

Register 6

SP B7-230.051.05

Page 4 of 4

Siemens-Elema AB

Wiring Diagrams

 

Rev. 02 10.99

SPS-UD

Solna, Sweden

X041E

1-1

SYSTEM OVERVIEW

MAMMOMAT 1000

MAMMOMAT 1000

MAMMOMAT

Fig. 1 Mammomat 1000 with generator with integrated control panel and radiation shield.

Siemens Elema AB

Register 6

SP B7-230.051.05

Page 1 of 4

Mammomat 1000/3000

Solna, Sweden

 

Rev. 02 10.99

SPS-UD

Wiring Diagrams

1-2

X041E

SYSTEM OVERVIEW

MAMMOMAT 1000 WITH REMOTE CONTROL

MAMMOMAT 1000

MAMMOMAT 1000

MAMMOMAT

MAM00638

MAM00638

Fig. 2 Mammomat 1000 generator with integrated control panel and radiation shield.

Siemens Elema AB

Register 6

SP B7-230.051.05

Page 2 of 4

Mammomat 1000/3000

Solna, Sweden

 

Rev. 02 10.99

SPS-UD

Wiring Diagrams

X041E

1-3

SYSTEM OVERVIEW

MAMMOMAT 3000

MAM00421

Fig. 3

MAMMOMAT 3000 installation with pivoted object-table arm and generator with integrated control panel and radiation shield.

Alternatively, the control panel and the radiation shield can be separated from the generator, see overleaf.

Siemens Elema AB

Register 6

SP B7-230.051.05

Page 3 of 4

Mammomat 1000/3000

Solna, Sweden

 

Rev. 02 10.99

SPS-UD

Wiring Diagrams

1-4

X041E

SYSTEM OVERVIEW

MAMMOMAT 1000/3000

SIEMENS

MAMMOMAT 3000

MAM00423

MAM00424

 

Fig. 4 Free-standing radiation shield with separate

Fig. 5 Separate generator

controle console

 

SIEMENS

·

MAMMOMAT 3000

Counterweight

Fixed object-table arm

MAM00427

Fig. 6 Stand with fixed object-table arm

Pivoted object-table arm (“flying wing”) replaced by fixed object-table arm and counterweight.

Siemens Elema AB

Register 6

SP B7-230.051.05

Page 4 of 4

Mammomat 1000/3000

Solna, Sweden

 

Rev. 02 10.99

SPS-UD

Wiring Diagrams

X041E 2-1

PLACEMENT OF PC-BOARDS AND COMPONENTS, STAND

MAMMOMAT 1000

X877 M6

D813**** X878**

Tube

K7-K9

S867*

m5*

 

D807

Right side

 

D807

Left side

S865*

 

D813****

 

 

S866*

Lamp, X894

m4*

m7*

 

S897

S868*

R861

 

m8*

X874

 

 

 

 

R863

D811****

MAMMOMAT 1000

M2 (Compression)

Strain gauge

D804

 

 

 

Object table

 

 

D813****

S861

 

X858***

S1

S892

S862

X870

S2

X892

 

 

 

M

S883

S884

D803

X801

*not fitted with one filter collimator

**not used with one filter collimator

Z801

*** not used

D812**** D805

M1 (Lift), X887, R807

R871

S882

D801

D809

S880

S881

S869

M3 (Rotation)

X888

D802

D808 Right side

D808 Left side

R803

 

X898, R806

 

R805

 

X881

 

X885

 

X880, X803

 

T802

 

T801

 

X899

MAM00637

F801-805

 

Incoming mains

**** option, remote control

Fig. 7

Siemens Elema AB

Register 6

SP B7-230.051.05

Page 1 of 16 Mammomat 1000/3000

Solna, Sweden

 

Rev. 02 10.99

SPS-UD

Wiring Diagrams

Siemens Mammomat 1000 Service Program

2-2

X041E

PLACEMENT OF PC-BOARDS AND COMPONENTS, STAND

MAMMOMAT 3000

 

 

 

M6

 

 

X877

D810

Tube

D805

X878

 

 

 

K7-K9

S867, M5

D807 Right side

D807 Left side

S865

S866 Lamp, X894

M4

M7 S897

S868 R861

m8

X874

R863

M2 (Compression)

Strain gauge

D804

 

 

 

Object table

S890

 

 

S1

S896

 

S861

X858

S2

S862

S895

X870

M

 

S891

 

 

 

 

 

S883

X891

 

 

 

 

 

S884

 

 

 

 

D803

 

 

 

X801

 

M1 (Lift), X887, R807

R871

S882

D801

D809

S880

S896

S881

S892

X892

S869

M3 (Rotation)

X888

D802

S894

D808 Right side

D808 Left side

R803

S860

X898, R806

R805

X880, X803

X881

X885

T802

T801

X899

MAM00426

F801-805

 

Z801

Incoming mains

Fig. 8 Stand with pivoted object-table arm

Siemens Elema AB

Register 6

SP B7-230.051.05

Page 2 of 16 Mammomat 1000/3000

Solna, Sweden

 

Rev. 02 10.99

SPS-UD

Wiring Diagrams

X041E

2-3

 

PLACEMENT OF PC-BOARDS AND COMPONENTS, STAND

 

 

 

 

SIEMENS

·

Counterweight

Fixed object-table arm

MAM00427

Fig. 9 Stand with fixed object-table arm

Pivoted object-table aram (“flying wing”) replaced by fixed object-table arm and counterweight

Siemens Elema AB

Register 6

SP B7-230.051.05

Page 3 of 16 Mammomat 1000/3000

Solna, Sweden

 

Rev. 02 10.99

SPS-UD

Wiring Diagrams

2-4

 

 

 

 

 

X041E

LIST OF BOARDS AND FUSES, STAND

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

PCB Overview

 

 

 

 

 

 

D801

CPU BOARD

 

 

 

D802

MOTOR CONTROL BOARD

 

D803

DISPLAY BOARD

 

D804

FORCE AMPLIFIER BOARD

 

D805

WING BOARD

 

 

 

D807

BOARD FOR LIFT, ROTATION AND LAMP SWITCHES (NOT FIT-

 

 

TED IN M1000)

 

 

 

D808

BOARD FOR LIFT AND ROTATION SWITCHES

 

D809

THICKNESS ADAPTER BOARD

 

D810

TILT SWITCH BOARD (Not fitted in M1000)

 

Option, remote control (MAMMOMAT 1000 only)

 

D811

REMOTE CONTROL TRANSMITTER (2 INCLUDED)

 

D812

REMOTE CONTROL RECEIVER (1 INCLUDED)

 

D833

REMOTE CONTROL DETECTOR (4 INCLUDED)

 

Fuses Overview

 

 

 

 

 

 

PCB

FUSE

FUSED VOLTAGE

 

 

 

 

 

 

 

 

D801

F1

2

AT

+5V, +5V_REF, +5V_DSP, +7V

 

D801

F2

1 AT

+5V_MPS

 

D802

F1

5

AT

+5V, +5V_M, +14V, D805/+5V

 

D802

F2

3.15 AT

+24VF, +15V

 

D802

F3

6.3 AT

+24V

 

D802

F5

5AT

17VAC (LAMP)

 

D802

F4

1

AT

24VACF (FAN)

 

Chassis-mounted

 

 

 

 

 

 

fuses

 

 

 

 

 

 

 

F801

2 AT

11 VAC

 

 

F802

1 AT

11 V_MPS

 

 

F803

5 AT

14 VAC

 

 

F804

5 AT

17 VAC

 

 

F805

6.3 AT

24 VAC

NOTE! See spare parts list for replacement fuses.

Siemens Elema AB

Register 6

SP B7-230.051.05

Page 4 of 16 Mammomat 1000/3000

Solna, Sweden

 

Rev. 02 10.99

SPS-UD

Wiring Diagrams

X041E

 

2-5

 

 

LIST OF SWITCHES AND COMPONENTS, STAND

 

 

 

 

 

 

 

 

 

Switches

 

 

 

 

 

 

 

Switch

Page

Description

 

 

 

 

 

 

 

 

 

S1/D801

3-6/1D

Reset switch for stand CPU.

 

 

 

 

 

S1/Object table

3-16/2A

Grid out switch located in the object table.

 

 

 

 

 

S1-1-4/D802

3-9/3D

Current limit switches: 1=+3%, 2=+6%, 3=+12%, 4=+24%.

 

 

 

 

 

S1-5/D802

3-9/3D

Tacho On/Off switch for rotation motor drive.

 

 

 

 

 

S1-S6/D807*

3-10/2A

Control switches for lamp, liftand rotation movement. Left and right side.

 

 

 

 

 

S1-S6/D811****

 

Control swithces and transmitter for lamp, liftand rotation movement.

 

 

 

Left and right side. 2 provided.

 

 

 

 

 

S1-S4/D808

3-10/4B

Control switches for liftand rotation movement. Left and right side.

 

 

 

 

 

S2/Object table

3-16/2A

Cassette inserted switch.

 

 

 

 

 

S860*

3-16/3C

Stereo lever switch*.

 

 

 

 

 

S861

3-12/2G

Compression limit switch, compression unit.

 

 

 

 

 

S862

3-12/2G

Decompression limit switch, compression unit.

 

 

 

 

 

S865**

3-15/2E

18 x 24 Collimator optoswitch**.

 

 

 

 

 

S866**

3-15/3E

24 x 30 Collimator optoswitch**.

 

 

 

 

 

S867**

3-15/4E

Filter position, optoswitch**.

 

 

 

 

 

S868**

3-15/5E

Focus position optoswitch**.

 

 

 

 

 

S869

3-8/3G

Tacho optoswitch

 

 

 

 

 

S880

3-4/2B

Emergency stop, switches off 24VAC and 14VAC, machine blocked.

 

 

 

 

 

S881

3-9/2G

Lift up limit switch, rotating unit.

 

 

 

 

 

S882

3-4/2A

Limit switch, distance tube head - floor <50 mm, switches off 24VAC,

 

 

 

machine blocked.

 

 

 

 

 

S883

3-8/4B

Collision protection switch, risk of collision when tubehead is less than

 

 

 

120 mm from the floor.

 

 

 

 

 

S884

3-9/2G

Lift down limit switch, rotating unit.

 

 

 

 

 

S890*

3-16/3C

AEC in position switch (Wing 1)*.

 

 

 

 

 

S891*

3-16/3C

Object table locked in position (Wing1)*.

 

 

 

 

 

S892

3-16/2C

Object table locked in position (Wing 2)***.

 

 

 

 

 

S894*

3-16/3C

Select wing 1 switch*, ***.

 

 

 

 

 

S895*

3-16/3C

Select wing 2 switch*, ***.

 

 

 

 

 

S896*

3-13/5B

Tilt opto-switch. 0° or ±10 0 used with stereo option*.

 

S897

3-15/2G

Diaphragm switch.

 

 

 

 

 

S3

3-12/5F

Decompression button (only with separate control console).

 

 

 

 

*

**

***

****

Not fitted in M1000

Not fitted in M1000 with one filter Not fitted in M3000 one wing M1000 option, remote control

Siemens Elema AB

Register 6

SP B7-230.051.05

Page 5 of 16 Mammomat 1000/3000

Solna, Sweden

 

Rev. 02 10.99

SPS-UD

Wiring Diagrams

2-6

 

 

X041E

LIST OF SWITCHES AND COMPONENTS, STAND

 

 

 

 

 

 

 

 

 

 

 

Potentiometers

 

 

 

 

 

 

 

 

Potentiometer

Page

Description

 

 

 

 

 

 

 

 

 

 

 

 

R803

3-13/2A

Tube angle, 2k.

 

 

 

 

 

 

 

R861

3-11/3E

Preset force, 10k.

 

 

 

 

 

 

 

R863

3-11/3E

Thickness, 2k.

 

 

 

 

 

 

 

R871

3-13/2A

Preset angle, 10k.

 

 

 

 

 

 

Motors

Motor

Page

Description

 

 

 

M

3-16/2A

Grid motor, placed in object table with grid.

 

 

 

M1

3-9/2G

Lift motor.

 

 

 

M2

3-12/2G

Compression motor.

 

 

 

M3

3-9/3G

Rotation motor.

 

 

 

M4*

3-15/3D

Collimator stepping motor*.

 

 

 

M5*

3-15/4D

Filter stepping motor*.

 

 

 

M6

3-7/4E

Fan, tube cooling.

 

 

 

M7*

3-14/4G

Mirror magnet*.

 

 

 

M8*

3-15/5D

Focus stepping motor*.

 

 

 

M9

3-9/5G

Rotation brake.

 

 

 

M10

3-9/5G

Rotation brake.

 

 

 

* Not fitted in M1000 with one filter

 

Miscellaneous

 

 

 

 

 

Misc.

Page

Description

 

 

 

 

 

 

COMP. UNIT

3-11, 3-12

Compression unit.

 

 

 

CABLE DUCT

3-2/3A

Generator to stand cable protection.

 

 

 

GROUND RAIL

3-2/1E

System grounding terminal.

 

 

 

K7

3-7/4E

Fan thermostat.

 

 

 

K8

3-7/4E

Tube overtemperature thermostat.

 

 

 

K9

3-7/4E

Tube housing overpressure switch.

 

 

 

L804 - L809

3-9/SG

VHF choke (EMC)

 

3-12/2G

 

 

 

 

LAMP

3-14/5G

Field light lamp.

 

 

 

R805

3-9/5G

Brake resistor, 6.2Ω.

 

 

 

R806

3-9/5G

Brake varistor

 

 

 

R807

3-9/2G

Lift motor varistor.

 

 

 

R860

3-11/2A

Strain gauge (R=300-400 Ω)

 

 

 

T801

3-3/2C

Mains transformer.

 

 

 

T802

3-3/3C

Insolation transformer

 

 

 

Z801

3-3/4C

Stereo mains filter, stand

 

 

 

C801 - C803

3-12/5G

EMC capacitors

 

 

 

Siemens Elema AB

Register 6

SP B7-230.051.05

Page 6 of 16 Mammomat 1000/3000

Solna, Sweden

 

Rev. 02 10.99

SPS-UD

Wiring Diagrams

X041E 2-7

LIST OF SIGNALS AND TEST POINTS, STAND

This page intentionally left blank.

Siemens Elema AB

Register 6

SP B7-230.051.05

Page 7 of 16 Mammomat 1000/3000

Solna, Sweden

 

Rev. 02 10.99

SPS-UD

Wiring Diagrams

2-8

X041E

LIST OF SIGNALS AND TEST POINTS,STAND

Page

Signal Name

Origin

Via

Destination

Testpoint

LED

 

 

 

 

 

 

 

3-9/1C

ACT_SPEED

TACHO

 

D802

ACT_SPEED

 

 

 

 

 

 

 

 

3-16/3C

AEC_POS

S890

D801

D801

 

 

 

 

 

 

 

 

 

3-10/2A

ANGLE_MEM

A.M. -pot

D807

D801

 

 

 

 

 

 

 

 

 

3-6/2B

AR

VH

D801

GENERATOR

D801 AR

 

 

 

 

 

 

 

 

3-11/3B

BR_OPEN

D804

 

 

D804.X843 pin 8

V2

 

 

 

 

 

 

 

3-11/3B

BR_RET

D804

 

 

D804.X843 Pin 7

 

 

 

 

 

 

 

 

3-9/4B

*BRAKE

D801

 

D802

 

 

3-9/4E

*BRAKE_RET

D802

 

M9, M10

D802

 

3-9/4E

*BRAKE_RELAY

D802

 

not used

 

 

3-15/3C

C_LA

D805

 

M4

 

 

 

 

 

 

 

 

 

3-15/3C

C_LB

D805

 

M4

 

 

 

 

 

 

 

 

 

3-9/4E

C_LIM

D802

 

 

 

V11

 

 

 

 

 

 

 

3-15/3C

*C_PA1

D805

 

M4

 

 

3-15/3C

*C_PA2

D805

 

M4

 

 

3-15/3C

*C_PB1

D805

 

M4

 

 

3-15/3C

*C_PB2

D805

 

M4

 

 

3-12/2D

C_PWM

D802

 

M2

D802

 

 

 

 

 

 

 

 

3-16/1A

CASS_LOADED

OBJ. TABLE

D805

D801

 

 

 

 

 

 

 

 

 

3-16/2C

*COLL_DIR

D801

 

D805

D805 COLL_D

 

3-16/2C

*COLL_STEP

D801

 

D805

D801, D805 COLL_ST

 

3-16/3E

COLL_POS_18X24

18x24 switch

D805

D801

COLL_P1

 

 

 

 

 

 

 

 

3-16/4E

COLL_POS_24X30

24x30 switch

D805

D801

COLL_P2

 

 

 

 

 

 

 

 

3-12/1E

COMP_LIMIT

D802

D805

COMP. UNIT

 

 

 

 

 

 

 

 

 

3-12/1B

*COMP_DIR

D801

 

D802

D802 COMP_D

 

3-11/2F

COMP_FORCE

D804

 

D801

D804.X843 pin 5

 

 

 

 

 

 

 

 

3-12/3E

COMP_OK

D802

 

D801

 

D802 V2

 

 

 

 

 

 

COMP_ERR

 

 

 

 

 

 

 

3-12/1E

COMP_PROT

D802

 

D801

D802

 

 

 

 

 

 

 

 

3-12/1B

*COMP_SPEED

D801

 

D802

D801, D802

 

 

 

 

 

 

COMP_SP

 

3-12/4F

COMPRESS

PEDAL

D805

D801, D802

 

 

 

 

 

 

 

 

 

3-9/2D

D_PWM

D802

 

 

D802

 

 

 

 

 

 

 

 

3-12/1E

DECOMP_ LIMIT

D802

D805

COMP. UNIT

 

 

 

 

 

 

 

 

 

3-12/4F

DECOMPRESS

PEDAL / **

 

D801

 

 

 

 

 

 

 

 

 

3-9/3B

*DIR

D801

 

D802

D802

 

3-9/4F

DRIVE_OK

D802

 

D801

 

D802

 

 

 

 

 

 

V24DRIVE_ERR

 

 

 

 

 

 

 

3-9/1E

DRIVE_PROT

D802

 

D801

D802

 

 

 

 

 

 

 

 

3-15/3D

FI_LA

D805

 

M5

 

 

 

 

 

 

 

 

 

3-15/3D

FI_LB

D805

 

M5

 

 

 

 

 

 

 

 

 

3-15/3D

*FI_PA1

D805

 

M5

 

 

*SIGNAL” = “SIGNAL” active low

**plus decompression button on separate control console.

Siemens Elema AB

Register 6

SP B7-230.051.05

Page 8 of 16 Mammomat 1000/3000

Solna, Sweden

 

Rev. 02 10.99

SPS-UD

Wiring Diagrams

X041E

2-9

 

LIST OF SIGNALS AND TEST POINTS, STAND

 

 

 

 

Signal Name

Description

 

 

 

 

ACT_SPEED

Test point for frequency to voltage converted tacho signal.

 

 

AEC_POS

AEC-detector in position near breast.

 

 

ANGLE_MEM

Operator control signal for preset stop angle.

 

 

AR

Signal given to generator as exposure request. If grid used, grid has started.

 

 

BR_OPEN

Bridge open, compression force strain gauge not working correctly.

 

 

BR_RET

Compression force strain guage return line (appr. 0V)

 

 

*BRAKE

Control signal to brake reversing relay.

 

 

*BRAKE_RET

Output to the brakes for max or min braking when active.

 

 

*BRAKE RELAY

Output to brake reversing relay, max brake if *BRAKE_RET active.

C_LA

Output line A. Collimator stepping motor M4

 

 

C_LB

Output line B. Collimator stepping motor M4

 

 

C_LIM

Rotation and fast-acting current limit

 

 

*C_PA1

Output Phase A1. Collimator stepping motor M4

*C_PA2

Output Phase A2. Collimator stepping motor M4

*C_PB1

Output Phase B1. Collimator stepping motor M4

*C_PB2

Output Phase B2. Collimator stepping motor M4

C_PWM

Pulse Width Modulated power output to compression motor.

 

 

CASS_LOADED

Cassette inserted in object table.

 

 

*COLL_DIR

Direction Control signal . For the filter stepping motor.

*COLL_STEP

Step control signal for the collimator stepping motor.

COLL_POS_18X24

Signal from the 18 x 24 collimator optoswitch

 

 

COLL_POS_24X30

Signal from the 24 x 30 collimator optoswitch

 

 

COMP

Power output via compression limit switch to compression motor.

 

 

*COMP_DIR

Controls compression motor to compress (moving down)

COMP_FORCE

Analog signal indicating compression force.

 

 

COMP_OK

Compression motor working correctly, no over current.

 

 

COMP_PROT

Compression relay K4 is working correctly, checked 1s after pedal release.

 

 

*COMP_SPEED

Pulse Width Modulated signal to control speed of compression motor (20kHz).

COMPRESS

Operator control signal to move compression plate down.

 

 

D_PWM

Pulse Width Modulated power output to rotation or lift motor.

 

 

DECOMP

Power output via decompression limit switch to compression motor.

 

 

DECOMPRESS

Operator control signal to move compression plate up.

 

 

*DIR

Selecting direction of rotation or lift motor.

DRIVE_OK

Rotation and lift motor are working correctly, no over current, approx. 1.5 s delay.

 

 

DRIVE_PROT

DMG relay K6 is working correctly, checked 3s after rot/lift button is released.

 

 

FI_LA

Output line A. Filter stepping motor M5

 

 

FI_LB

Output line B. Filter stepping motor M5

 

 

*FI_PA1

Output phase A1. Filter stepping motor M5

Siemens Elema AB

Register 6

SP B7-230.051.05

Page 9 of 16 Mammomat 1000/3000

Solna, Sweden

 

Rev. 02 10.99

SPS-UD

Wiring Diagrams

2-10

X041E

LIST OF SIGNALS AND TEST POINTS, STAND

Page

Signal Name

Origin

Via

Destination

Testpoint

LED

 

 

 

 

 

 

 

 

 

 

 

 

 

 

3-15/4C

*FI_PA2

D805

 

M5

 

 

3-15/4C

*FI_PB1

D805

 

M5

 

 

3-15/4C

*FI_PB2

D805

 

M5

 

 

3-15/2C

*FILTER_DIR

D801

 

D805

D805 FILT_D

 

3-15/2B

FILTER_POS

Filter switch

D805

D801

FILT_P

 

 

 

 

 

 

 

 

3-15/2C

*FILTER_STEP

D801

 

D805

D805 FILT_ST

 

3-15/5C

FO_LA

D805

 

M8

 

 

 

 

 

 

 

 

 

3-15/5C

FO_LB

D805

 

M8

 

 

 

 

 

 

 

 

 

3-15/5C

*FO_PA1

D805

 

M8

 

 

3-15/5C

*FO_PA2

D805

 

M8

 

 

3-15/5C

*FO_PB1

D805

 

M8

 

 

3-15/5C

*FO_PB2

D805

 

M8

 

 

3-15/1C

*FOCUS_DIR

D801

 

D805

D805 FOCUS_D

 

3-15/1B

FOCUS_POS

Focus switch

D805

D801

FOCUS_P

 

 

 

 

 

 

 

 

3-15/1C

*FOCUS_STEP

D801

 

D805

D805 FOCUS_DT

 

3-16/3B

GRID_MOTOR_N

D802

D805

OBJ. TABLE

D802 GRID_M

 

 

 

 

 

 

 

 

3-16/2A

GRID_OUT

OBJ. TABLE

D805

D801

D801 GRID_O

 

 

 

 

 

 

 

 

3-16/2G

*GRID_SPEED

D801

 

D802

D801 GRID_SP

 

3-11/3D

I1_OUT

D804

 

 

D804.X843 pin 6

 

 

 

 

 

 

 

 

3-6/2B

*KVA

D801,D701

 

D700 (GENERATOR)

D801 (D705, X715)

 

 

 

D702, D705

 

 

 

 

 

 

 

 

 

 

 

3-14/5G

LAMP

D807, D812**

D805

D801

 

 

 

 

 

 

 

 

 

3-14/4C

*LAMP_CTRL

D801

 

D802

 

 

3-14/5F

LAMP_V

D802

 

LAMP

D802 LAMP

 

 

 

 

 

 

 

 

3-8/3D

LED

D802

 

TACHO switch

 

 

 

 

 

 

 

 

 

3-15/2E

LED_1

D805

 

18 x 24 switch

 

 

 

 

 

 

 

 

 

3-15/3E

LED_2

D805

 

24 x 30 switch

 

 

 

 

 

 

 

 

 

3-15/4E

LED_3

D805

 

Filter switch

 

 

 

 

 

 

 

 

 

3-15/5E

LED_4

D805

 

Focus switch

 

 

 

 

 

 

 

 

 

3-10/2A

LIFT_DOWN

D807, D808, D812**

D805

D801, D802

 

 

 

 

 

 

 

 

 

3-10/2A

LIFT_UP

D807, D808, D812**

D805

D801, D802

 

 

 

 

 

 

 

 

 

3-14/4F

*MIRROR_MAGNET

D802

 

M7

D802 MIRROR

 

3-6/3B

MPS

D801

 

GENERATOR

D801

 

 

 

 

 

 

 

 

3-7/4C

OKT1_

TUBE/K8

 

GENERATOR

 

 

 

 

 

 

 

 

 

3-7/4C

OKT2_

TUBE/K9

 

GENERATOR

 

 

 

 

 

 

 

 

 

3-9/2C

PGM_SPEED

D802

 

 

PGM_SPEED

 

 

 

 

 

 

 

 

3-13/2D

POT_RETURN

POT

 

D801

 

 

 

 

 

 

 

 

 

3-11/2E

PRES_FORCE

R861

 

D801

 

 

 

 

 

 

 

 

 

3-13/1A

PRESET_ANGLE

R871

D805

D801

 

 

 

 

 

 

 

 

 

*SIGNAL” = “SIGNAL” active low

** M1000 option, remote control, D807 not fitted

 

 

 

 

 

 

 

 

 

Siemens Elema AB

Register 6

SP B7-230.051.05

Page 10 of 16 Mammomat 1000/3000

Solna, Sweden

 

Rev. 02 10.99

SPS-UD

Wiring Diagrams

X041E

2-11

 

LIST OF SIGNALS AND TEST POINTS, STAND

 

 

 

 

Signal Name

Description

 

 

 

 

*FI_PA2

Output phase A2. Filter stepping motor M5

*FI_PB1

Output phase B1. Filter stepping motor M5

*FI_PB2

Output phase B2. Filter stepping motor M5

*FILTER_DIR

Control signal . For the filter stepping motor.

FILTER_POS

Signal from the filter pos collimator optoswitch

 

 

*FILTER_STEP

Step control. For the filter stepping motor.

FO_LA

Output line A. Focus stepping motor M8

 

 

FO_LB

Output line B. Focus stepping motor M8

 

 

*FO_PA1

Output Phase A1. Focus stepping motor M8

*FO_PA2

Output Phase A2. Focus stepping motor M8

*FO_PB1

Output Phase B1. Focus stepping motor M8

*FO_PB2

Output Phase B2. Focus stepping motor M8

*FOCUS_DIR

Control signal for the focus stepping motor.

FOCUS_POS

Signal from the focus pos collimator optoswitch.

 

 

*FOCUS_STEP

Step control signal for the focus stepping motor.

GRID_MOTOR_N

Grid motor return line.

 

 

GRID_OUT

Active when the grid is not in end position.

 

 

*GRID_SPEED

Pulse Width Modulated signal control speed of grid motor.

I1_OUT

Output signal from compression force preamplifier.

 

 

*KVA

Signal to terminate exposure. WIRED-OR line, see generator.

LAMP

Operator control signal for the field light lamp.

 

 

*LAMP_CTRL

Signal to switch on the field light lamp.

LAMP_V

Output to the field light lamp.

 

 

LED

Output to the LED in the tacho opto switch in the rotation motor.

 

 

LED_1

Output to the LED in the 18 x 24 collimator opto switch.

 

 

LED_2

Output to the LED in the 24 x 30 collimator opto switch.

 

 

LED_3

Output to the LED in the filter collimator opto switch.

 

 

LED_4

Output to the LED in the focus collimator opto switch.

 

 

LIFT_DOWN

Operator control signal to move compression plate down.

 

 

LIFT_UP

Operator control signal to move compression plate up.

 

 

*MIRROR_MAGNET

Output to mirror magnet. Enabled via *LAMP_CTRL.

MPS

Multi processor serial communication for information exchange Master - Slaves.

 

 

OKT1_

Signal from temperature switch K8 on the tube.

 

 

OKT2_

Signal from the over pressure switch K9 on the tube.

 

 

PGM_SPEED

Test point for duty cycle to voltage converted programmed speed signal.

 

 

POT_RETURN

Return line for potentiometers connected to 0VA via a 20Ω resistor.

 

 

PRES_FORCE

Analog signal as preset value for maximum compression force.

 

 

PRESET_ANGLE

Analog signal as preset value for requested rotation angle.

 

 

Siemens Elema AB

Register 6

SP B7-230.051.05

Page 11 of 16 Mammomat 1000/3000

Solna, Sweden

 

Rev. 02 10.99

SPS-UD

Wiring Diagrams

2-12

X041E

LIST OF VOLTAGES AND TEST POINTS, STAND

Page

Signal Name

Origin

Via

Destination

Testpoint

LED

 

 

 

 

 

 

 

 

 

 

 

 

 

 

3-12/2D

PU

D802

 

 

PU

 

 

 

 

 

 

 

 

3-9/4B

*RELEASE

D801

 

D802

D802

 

3-6/1E

*RESET

D801

 

 

 

 

3-6/1E

*RESET_HW

D801

 

 

 

 

 

 

 

 

 

 

 

3-6/1E

RESET_SW

D801

 

 

 

 

 

 

 

 

 

 

 

3-10/3A

ROT_CW

D807, D808,

D805

D801, D802

 

 

 

 

D812***

 

 

 

 

 

 

 

 

 

 

 

3-10/3A

ROT_CCW

D807, D808,

D805

D801, D802

 

 

 

 

D812***

 

 

 

 

 

 

 

 

 

 

 

3-13/4A

ROT_POS**

D806**

 

D801

 

 

 

 

 

 

 

 

 

3-6/3D

RXD

GEN.

 

D801

D801

 

 

 

 

 

 

 

 

3-9/3B

*SEL_ROT

D801

 

D802

D802 SEL_ROT

 

3-16/3C

SELECT_WING1

S891

 

D801

 

 

 

 

 

 

 

 

 

3-16/4C

SELECT_WING2

S892

 

D801

 

 

 

 

 

 

 

 

 

3-9/2B

*SPEED

D801

 

D802

D801, D802 SPEED

 

3-15/2C

*STEP_ENABLE

D801

 

D805

D805 STEP_EN

 

3-15/1B

STEREO_COLL

S897

 

D801

 

 

 

 

 

 

 

 

 

3-16

STEREO_LEVER

S860

D805

D801

 

 

 

 

 

 

 

 

 

3-14/1B

TABLE_CONFIG

OBJ.TABLE

D805

D801

 

 

 

 

 

 

 

 

 

3-8/3G

TACHO

Tacho switch

 

D802

D802 TACHO

 

 

 

 

 

 

 

 

3-11/3E

THICKNESS

R863

 

D801

 

 

 

 

 

 

 

 

 

3-11/5E

THICKNESS_B

THICKNESS

D809

 

 

 

 

 

 

 

 

 

 

3-11/5F

THICKNESS_STE-

D809

 

X885

 

 

 

REO

 

 

 

 

 

 

 

 

 

 

 

 

3-13/5B

TILT

S896/D810

D805

D801

 

D810 V8

 

 

 

 

 

 

 

3-13/2A

TUBE_ANGLE

R803

 

D801

 

 

 

 

 

 

 

 

 

3-6/4D

TXD

D801

 

GENERATOR

D801 TXD

 

 

 

 

 

 

 

 

3-6/3B

VH

GEN.

 

D801

 

 

 

 

 

 

 

 

 

3-11/3C

ZERO

D804

 

 

D804.X843 pin 4

 

 

 

 

 

 

 

 

“*SIGNAL” = “SIGNAL” active low.

**Not used with software v2.1 or higher.

***M1000 option, remote control, D807 not fitted

Siemens Elema AB

Register 6

SP B7-230.051.05

Page 12 of 16 Mammomat 1000/3000

Solna, Sweden

 

Rev. 02 10.99

SPS-UD

Wiring Diagrams

X041E

2-13

 

LIST OF VOLTAGES AND TEST POINTS, STAND

 

 

 

 

Signal Name

Description

 

 

 

 

PU

Pull up signal for error comparators = +15V, +5V, +24V voltages OK.

 

 

*RELEASE

Control signal to release the brakes from min or max to permanent magnet braking.

RESET

Reset signal to the CPU.

 

 

RESET_HW

Reset signal from the CPU to the reset cicuit, caused by slave restart and CPU watchdog

 

reset.

 

 

RESET_SW

Signal from the reset switch to the CPU via the reset circuit.

 

 

ROT_CW

Operator control signal to rotate the x-ray system clockwise.

 

 

ROT_CCW

Operator control signal to rotate the x-ray system counter clockwise.

 

 

ROT_POS

Low level signal pulse indicating +10 or -10 degrees of rotation.

 

 

RXD

Input for receiving data via MPS.

 

 

*SEL_ROT

Low level signal selects the rotation drive to be controlled.

SELECT_WING1

Signal from select wing 1 switch. Wing 1 selected.

 

 

SELECT_WING2

Signal from select wing 2 switch. Wing 2 selected.

 

 

*SPEED

Pulse Width Modulated signal to control speed of rotation or lift motor (20 kHz).

*STEP_ENABLE

Control signal to enable all stepping motors.

STEREO_COLL

External diaphragm switch, diaphragm in place. For stereo or spot diaphragm, 18 x 24

 

collimator in M1000 with one filter.

 

 

STEREO_LEVER

Signal from stereo lever switch, indicating lever not in stereo position.

 

 

TABLE_CONFIG

4 inputs from object table, see table configuration .

 

 

TACHO

Signal from rotation motor tachometer.

 

 

THICKNESS

Analog signal indicating the compression thickness.

 

 

THICKNESS_B

Buffered output to the thickness adapter board D809 (Stereo option)

 

 

THICKNESS_STE-REO

Isolated analog signal indicating compression thickness for stereo evaluation unit.

 

 

TILT

Indicates that the stereo angle is 0° or ±10°. Visible through vent opening behind tube.

 

 

TUBE ANGLE

Analog signal indicating the rotation angle.

 

 

TXD

Output for sending data via MPS.

 

 

VH

The stand is informed that the generator has completed its preparation.

 

 

ZERO

Used for adjusting offset of the compression force amplifiers.

 

 

Siemens Elema AB

Register 6

SP B7-230.051.05

Page 13 of 16 Mammomat 1000/3000

Solna, Sweden

 

Rev. 02 10.99

SPS-UD

Wiring Diagrams

Loading...
+ 49 hidden pages