QuikLink
INSTRUCTIONS FOR USE English
INTENDED USE 4
SUMMARY AND EXPLANATION 4
CONTENTS 5
QuikRead
®
QuikLink consists of two separete products 5
Devices and items required but not supplied 5
OPERATING INSTRUCTIONS 6 Preparations for the use of the QuikRead QuikLink 6 Measurement on QuikLink mode with bar code reader 6 Measurement on Temporary NON-LIS mode 10 Measurement on QuikLink mode without bar code reader 10
TROUBLESHOOTING 12
BRUKSANVISNING Svenska
AVSEDD ANVÄNDNING 14
ÖVERSIKT 14
INNEHÅLL 15
QuikRead
®
QuikLink består av två separata produkter 15
Tillbehör och artiklar som krävs, men som ej ingår 15
INSTRUKTIONER FÖR DRIFT 16 Förberedelser för användande av QuikRead QuikLink 16 Mätning i QuikLink läge med streckkodsläsare 16 Mätning i tillfälligt NON-LIS läge 20 Mätning i QuikLink läge utan streckkodsläsare 20
FELSÖKNING 22
3
English
QuikLink
INTENDED USE
Transportation of QuikRead assay results coupled with patient identication
to a remote Laboratory or Hospital Informa tion System (LIS/HIS) using an ASTM protocol.
Only for the QuikRead 101 Instrument.
SUMMARY AND EXPLANATION
QuikRead QuikLink is a supplementary product for the QuikRead Instrument. It allows transfer of QuikRead assay results to a remote LIS/HIS using the ASTM protocol. These instructions for use describe the use of the QuikRead QuikLink user interface and include supplementary instructions for the routine measurement protocol. Before starting any assay, read and follow the QuikRead Instrument and
QuikRead reagent kit instructions for use for the specic assay procedure
and sample requirements. QuikRead QuikLink has two optional protocols (Figure 1). A Quik Read assay result can be transferred to a remote LIS/HIS computer or a printer either with
patient identication (QuikLink mode with bar code reader) or with consecutive
sample number and time (QuikLink mode without bar code reader). The proto-
col is chosen and congured using the QuikRead QuikLink Setup Wizard.
QuikLink mode with bar code reader / Temporary NON-LIS mode
If the QuikLink mode with bar code reader is enabled, the QuikRead Instru­ment starts in the QuikLink mode during start-up. A bar code reader is used to
couple the assay results with patient and, optionally, operator identication (ID).
If the Accept result before sending option is enabled, the accepted results are sent to the remote LIS/HIS computer or printer after each measurement. If the Accept result before sending option is disabled, all results are automati­cally sent to the remote LIS/HIS computer or printer after each measurement without an acceptance procedure. If no patient ID is available, for example with emergency patients, the QuikRead Instrument can be switched to the Temporary NON-LIS mode. In the tempo­rary NON-LIS mode, patient ID is not requested for and the result is not sent to the LIS. After one successful temporary NON-LIS measurement, the instrument automatically returns to the QuikLink mode with bar code reader.
QuikLink mode without bar code reader
If the bar code reader is not in use, the QuikRead Instrument is set into QuikLink mode without bar code reader. In this mode, results are sent to the LIS computer automatically after each measurement without any patient
or operator identications or result acceptance. Data is sent coupled with
consecutive sample number and current time. The Temporary NON-LIS mode is not available in this mode.
4
Figure 1 QuikLink modes
Mode 1 QuikLink mode with bar code reader
Mode 2 QuikLink mode without bar code reader
CONTENTS
QuikRead QuikLink consists of two separate products:
®
QuikRead
QuikLink Cat. No. 06100
QuikRead QuikLink Setup Wizard CD-ROM. A PC program that installs the required software for the QuikRead 101 Instrument and congures the LIS/HIS properties.
QuikRead QuikLink Setup Wizard Instructions for use
QuikRead® QuikLink bar code reader Cat. No. 06109
(required only for the “QuikLink mode with bar code reader”)
Bar code reader
Interface cable.
Connects the bar code reader to the QuikRead Instrument.
QuikLink bar code tablet
Bar code reader user manual
QuikRead QuikLink Instructions for use
Devices and items required but not supplied
QuikRead 101 Instrument, Cat. No. 06040
Serial cable (null modem)
Remote LIS/HIS
QuikRead driver for the remote LIS/HIS (LIS/HIS dependent)
5
OPERATING INSTRUCTIONS
Preparations for the use of the QuikRead QuikLink
Use the QuikRead QuikLink Setup Wizard to congure the desired QuikLink
settings for your QuikRead Instrument. See the QuikRead QuikLink Setup
Wizard instructions for use.
Note that after servicing the QuikRead Instrument or the bar code reader, the
devices need to be recongured to the customised settings.
Measurement on QuikLink mode with bar code reader
These instructions for use describe the mode in which the assay results are sent to a remote LIS/HIS computer. The same procedure is also valid in the mode in which the results are sent to a printer.
START-UP
Switch on the QuikRead Instrument.
Check that the LCD displays:
QuikLink mode Operator ID enabled
QuikLink mode
or
Operator ID disabled
Use a magnetic card to set the analyte. The LCD will alternately display:
Ready for use <test> lot xxxx
Note that the magnetic card can only be read when the LCD displays
Read the card or Ready for use.
Ready for use
and
Read patient ID
USE OF THE BAR CODE READER
See the supplied bar code reader user manual for detailed instructions for use. Use the supplied interface cable to connect the bar code reader to the QuikRead Instrument. The cable is connected to the bar code reader con­nector located at the back of the QuikRead Instrument. See Figure 2 for the connections. To read a bar code, aim the reader at a bar code and pull the trigger in the
handle. Code scanning is performed along the center of the light bar emitted from the reading window. This bar must cover the entire code horizontally. Do
not aim the light bar direct at your eyes. The manufacturer’s warranty for the bar code reader covers defects in ma­terials or workmanship for 5 years from the date of shipment, provided that the product is operated under normal and proper conditions. In the event of a reader malfunction, contact your local supplier to receive information of the authorised service.
6
Figure 2
B
Cable connection
Serial cable
A
Bar code reader connector
B
Bar code reader interface cable
C
Power cord
D
A
D
C
PROCEDURE
Reading patient ID
1. Place the bar code reader above the bar code in the patient information
sheet or sample tube and press the trigger button in the handle. Wait for
a sound signal.
If the bar code reading fails several times, see Troubleshooting.
2. The LCD will display the patient ID. Make sure that it matches the patient
ID in the patient information sheet.
Ready for use <test> Patient <xxxxxxxx>
Reading operator ID
(Optional feature. If not in use go to step 6.)
3. If the operator ID is enabled, the LCD will display:
Ready for use <test> Read operator ID
4. Read the operator ID with the bar code reader. The LCD will display the
operator ID. Make sure that it matches the read operator ID.
Operator ID <oooooooo>
5. The LCD will then display:
Ready for use <test> Patient <xxxxxxxx>
Measurement
● with Accept result before sending enabled go to step 6
● with Accept result before sending disabled go to step 14
Measurement with Accept result before sending enabled
6. Make the measurement by following the instructions for use for the
QuikRead Instrument and QuikRead reagent kits and the texts displayed
by the LCD.
7. When the measurement is completed, the LCD will alternately display
the texts:
Patient <xxxxxxxx> Result <zz> <uuu>
where <xxxxxxxx> is the patient ID, <zz> is the result, and <uuu> is the
analyte.
and Accept or reject test
7
Accepting or rejecting a result
8. To accept a result, read the bar code OK/ACCEPT on the QuikLink bar
code tablet (bar codes also on page 2). To reject a measurement, read the bar code CANCEL/REJECT on the QuikLink bar code tablet.
9. When the result is accepted, the LCD will display:
Test accepted Remove cuvette
If the measurement was rejected, the LCD will display:
Test rejected Remove cuvette
10. Remove the cuvette and wait for the QuikRead Instrument to perform a
self-check.
Sending a result to the remote LIS/HIS computer
11. After a successful self-check, the accepted result is automatically sent to
the remote LIS/HIS computer. This can take a few seconds. During the
process, the LCD will display:
Sending result...
If the measurement was rejected, the result is not sent to the remote LIS/HIS computer. If the self-check is unsuccessful, the result is not sent
and an error message is displayed by the LCD. Delete the result.
12. When the result has been successfully transferred, the LCD will dis-
play:
Test result sent succesfully
13. If the QuikRead Instrument is unable to transfer the result to the remote
LIS/HIS computer, the LCD will alternately display the texts:
Sending failed! Print result?
Patient <xxxxxxxx>
and
Result <zz> <uuu>
Results can be recorded manually or sent to a printer. To send the results
to a printer, rst connect the printer to your QuikRead Instrument and
then read the OK/ACCEPT bar code on the QuikLink bar code tablet. See Troubleshooting.
When recording results manually, the error message must be acknowl-
edged by reading the CANCEL/REJECT bar code on the QuikLink bar code tablet.
In both cases, the result is removed from the internal memory of the
instrument and QuikRead is ready for the next measurement.
8
Measurement with Accept result before sending disabled
14. Make the measurement by following the instructions for use for the
QuikRead Instrument and QuikRead reagent kits and the text displayed
by the LCD.
15. When the measurement is completed, the LCD will alternately display
the texts:
Patient <xxxxxxxx> Result <zz> <uuu>
Remove cuvette to
and
start LIS/HIS sending
where <xxxxxxxx> is the patient ID, <zz> is the result, and <uuu> is the
analyte.
16. Remove the cuvette and wait for the QuikRead Instrument to perform a
self-check.
Sending a result to the remote LIS/HIS computer
17. After a succesful self-check, the QuikRead Instrument automatically sends
the result to the remote LIS/HIS computer. This can take a few seconds.
During the process, the LCD will display:
Sending result...
If the self-check is unsuccesful, the result is not sent to the remote LIS/
HIS computer and an error message is displayed by the LCD. Delete the
result.
18. When the result has been succesfully transferred, the LCD will alternately
display the texts:
Ready for use <test> lot xxx
Ready for use
and
Read patient ID
The instrument is ready for the next measurement. If the QuikRead Instrument is unable to transfer the result to the LIS/HIS
computer, the LCD will alternately display the texts:
Sending failed! Print result?
Patient <xxxxxxxx>
and
Result <zz> <uuu>
Results can be recorded manually or sent to a printer. To send the results
to a printer, rst connect the printer to your QuikRead Instrument and
then read the OK/ACCEPT bar code on the QuikLink bar code tablet. See Section 5 for Troubleshooting.
When recording results manually, the error message must be acknowl-
edged by reading the CANCEL/REJECT bar code on the QuikLink bar code tablet.
In both cases, the result is removed from the internal memory of the
instrument and QuikRead is ready for the next measurement.
Brief instructions for use of QuikLink mode with bar code reader
1.
Read patient ID.
2.
Read operator ID (optional).
3.
Make the measurement.
4.
Accept the result (optional).
9
Measurement on Temporary NON-LIS mode
Make sure that the QuikRead Instrument is in the idle state, i.e. the LCD alternately displays:
Ready for use <test> lot xxx
Ready for use
and
Read patient ID
SETTING INSTRUMENT TO TEMPORARY NON-LIS MODE
Read the Enter temporary NON-LIS mode bar code on the QuikLink bar
code tablet. The LCD will alternately display:
Ready for use <test> lot
Ready for use
and
NON-LIS mode
To exit Temporary NON-LIS mode without measuring any sample, read the Exit temporary NON-LIS mode bar code on the QuikLink bar code tablet.
PROCEDURE
1. Make the measurement by following the instrument and reagent kit in-
structions for use and the texts displayed by the LCD.
2. Record the result before the cuvette is removed from the measurement
well.
3. After a successful measurement, the QuikRead Instrument automatically returns to the QuikLink mode with bar code reader.
If a measurement fails (test cancelled), the QuikRead Instrument returns
to the start of the Temporary NON-LIS mode.
Measurement on QuikLink mode without bar code reader
START-UP
Switch on the QuikRead Instrument
Make sure that the LCD displays:
QuikLink mode Bar code reader disabled
Use a magnetic card to set the analyte. The LCD will display:
Ready for use <test> lot xxxx
Note that the magnetic card can only be read when the LCD dis plays
Read the card or Ready for use.
10
PROCEDURE
Measurement
1. Make the measurement by following the instructions for use for the
QuikRead Instrument and QuikRead reagent kits and the texts displayed
by the LCD.
2. When the measurement has been completed, the LCD will alternately display the texts:
Sample <xx> / <yyyyyy> Result <zz> <uuu>
Remove cuvette to
and
start LIS/HIS sending
where <xx> is the consecutive sample number, <yyyyyy> is the current
date, <zz> is the result and <uuu> is the analyte.
3. Remove the cuvette and wait for the QuikRead Instrument to perform a self-check.
Sending a result to the remote LIS/HIS computer
4. After a successful self-check, the QuikRead Instrument automatically
sends the result to the remote LIS/HIS computer. This can take a few
seconds. During this process, the LCD will display:
Sending result...
If the self-check is unsuccessful, the result is not sent and an error mes-
sage is displayed by the LCD. Delete the result.
5. When the result has been successfully transferred, the LCD will alternately display the texts:
Previous sending OK and
The instrument is ready for the next measurement. If the QuikRead Instrument is unable to transfer the result to the LIS
computer, the LCD will alternately display the following texts:
Ready for use <test> lot xxxx
Previous sending failed!
Sample <xx> / <yyyyyy> Result <zz> <uuu>
Read analyte card to proceed
and
and
The assay result shall be recorded manually. Use the analyte magnetic
card to proceed. The result is removed from the internal memory of the instrument and the QuikRead Instrument is ready for the next measure­ment. See Troubleshooting.
11
TROUBLESHOOTING
Error class Error type Error message
Bar code reading Bar code reader
sound signal absent after triggering.
A sound signal is heard but no message is
displayed by the LCD.
Error message on the LCD. Error while reading bar
Incorrect patient or operator ID shown
on the LCD.
Data transmission Error message on the
QuikRead Instrument display.
Data cannot be seen on the remote LIS/HIS computer.
Wrong data on the remote
LIS/HIS computer.
code. Please try again.
Sending failed! Print result?
12
Possible cause Corrective action
The interface cable connection is poor.
The bar code reader is wrongly
congured.
Unsuitable bar code reader Select right bar code reader model.
Defected bar code reader Contact your local supplier.
Poor bar code quality Use a new label.
Unsuitable bar code type Check that the QuikRead QuikLink
The QuikRead Instrument
is wrongly congured.
The interface cable connection is poor.
The interface cable is defective. Contact your local supplier.
Several causes Read again.
Unsuccesful bar code reading Aim the reader more carefully at the bar
Bar code reader does not sup­port the bar code type used for patient or operator ID and there­fore the code is read incorrectly.
The cable to the remote LIS/HIS computer is disconnected.
The cable to the remote LIS/HIS computer is defective.
Remote LIS/HIS computer is not responding.
Wrong driver version on the
remote LIS/HIS computer.
Other cause
Wrong driver version on the
remote LIS/HIS computer.
Other cause
Make sure that the cable is properly connected.
Use the QuikRead QuikLink Setup Wizard for reconguration.
Setup Wizard supports the bar code
families in use.
Use the QuikRead QuikLink Setup Wizard for reconguration.
Make sure the cable is properly connected.
If the message persists, contact your local supplier.
code. Ensure that the light bar covers the
entire code horizontally.
Check that the QuikRead QuikLink Setup Wizard supports the bar code families in
use.
Connect the cable.
Change the cable.
Check that the
remote computer is OK.
Contact your local administrator.
Contact your local administrator.
Also see the QuikRead 101 Instrument Instructions for use,
Appendix C: Troubleshooting / Error messages
13
Svenska
QuikLink
AVSEDD ANVÄNDNING
Överföring av analysresultat från QuikRead kopplade med patientidentika­tion till LIS/HIS (Laboratory or Hospital Information System, Laboratorie- eller Sjukhusdatasystem) som använder ASTM.
Endast för QuikRead 101 Instrument.
ÖVERSIKT
QuikRead QuikLink är en tilläggsprodukt till QuikRead Instrumentet. Den gör det möjligt att föra över analysresultat från QuikRead till LIS/HIS som använder ASTM.
Den här bruksanvisningen beskriver ödet vid användning av QuikRead
QuikLink och en del tilläggsinformation för det rutinmässiga analyserandet. Läs och följ instruktionerna för QuikRead instrumentet och kitinlagan för den analyt som ska analyseras innan någon analys utförs. QuikRead QuikLink har två valbara alternativ (Figur 1). Resultatet från en QuikRead analys förs över till en LIS/HIS dator eller printer antingen med patient-ID (QuikLink läge med streckkodsläsare) eller med löpnummer och tidsangivelse (QuikLink läge utan streckkodsläsare). Alternativen väljs och
kongureras i installationsprogrammet för QuikRead QuikLink.
QuikLink läge med streckkodsläsare / Tillfälligt NON-LIS läge
Om QuikLink-läge med streckkodsläsaren är inställt, startar QuikRead Instrumentet i QuikLink-läge. En streckkodsläsare används för att koppla ihop
analysresultatet med patienten och eventuell användaridentikation (ID). Om
funktionen Acceptera resultatet före överföring är innställd, sänds svaren till en LIS/HIS-dator eller skrivare efter varje mätning, först efter det att ett accepterande av svaret är gjort. Om funktionen Acceptera resultatet före överföring ej är inställd, sänds svaren automatiskt till en LIS/HIS-dator eller skrivare efter varje mätning utan att svaret accepteras. Om patient-ID inte är tillgängligt, exempelvis vid akuta fall, kan QuikRead Instrumentet ställas om till Tillfälligt NON-LIS läge. Vid Tillfälligt NON-LIS läge efterfrågas inte patient-ID och resultaten sänds inte till LIS/HIS datorn. Efter en lyckad analys i Tillfälligt NON-LIS läge återgår instrumentet automatiskt till
QuikLink läge med streckkodsläsare.
QuikLink läge utan streckkodsläsare
Om QuikLink läge utan streckkodsläsare är inställt, startar instrumentet i detta läge när det sätts igång. Resultaten sänds då automatiskt till LIS/ HIS datorn efter varje mätning, utan godkännande av användaren, och utan
patient- eller användareidentikation. Data sänds kopplade med löp nummer
och aktuell tid. Tillfälligt NON-LIS läge är inte tillgängligt i detta läge.
14
Figur 1 QuikLink lägen
Alternativ 1 QuikLink läge med streckkodsläsare
Alternativ 2 QuikLink läge utan streckkodsläsare
INNEHÅLL
QuikRead QuikLink består av två separata produkter:
QuikRead® QuikLink Artikelnr: 06100-OD
QuikRead QuikLink Setup Wizard CD-ROM.
Ett PC program som installerar den efterfrågade mjukvaran för QuikRead 101 Instrumentet och kongurerar egenskaperna för LIS/HIS.
QuikRead QuikLink installationsprogram bruksanvisning
QuikRead® QuikLink streckkodsläsare Artikelnr: 06109-OD
(krävs endast för “QuikLink läge med streckkodsläsare”)
Streckkodsläsare
Streckkodsläsarens kabel.
Ansluter streckkodsläsaren till QuikRead Instrumentet. “Streckkodstavla” för QuikLink
Streckkodsläsarens bruksanvisning
QuikRead QuikLink bruksanvisning
Tillbehör och artiklar som krävs, men som ej ingår
QuikRead 101 Instrument, artikelnummer 06040-OD
Seriell kabel (noll modem)
LIS/HIS
QuikRead drivrutin för LIS/HIS (Olika beroende på LIS/HIS)
15
INSTRUKTIONER FÖR DRIFT
Förberedelsen för användande av QuikRead QuikLink
Använd installationsprogrammet för QuikRead QuikLink för att kongurera de
inställningar som krävs för ditt QuikRead Instrument. Läs bruksanvisningen för installationsprogrammet till QuikRead QuikLink. Notera att efter service på QuikRead Instrumentet eller streck kodsläsaren
behöver en ny kongurering göras för att återfå tidigare inställningar.
Mätning i QuikLink läge med streckkodsläsare
Dessa instruktioner beskriver det läge i vilket mätresultat sänds till en LIS/HIS dator. Samma procedur gäller också i läget då resultat sänds till en skrivare.
UPPSTART
Sätt igång QuikRead Instrumentet.
Kontrollera att displayen visar:
Inställd för QuikLink användar-ID pa
Inställd för QuikLink
eller
användar-ID av
Använd ett magnetkort för att välja analyt. Displayen kommer nu omväx-
lande visa:
Klar att använda <test> lot xxxx
Notera att magnetkortet endast kan läsas när displayen visar Läs kortet
eller Klar att använda.
Klar att använda
och
Läs patient-ID
ANVÄNDNING AV STRECKKODSLÄSARE
Läs den medskickade bruksanvisningen för streckkodsläsaren för detalje­rade användar instruktioner. Använd streckkodsläsarens kabel för att ansluta streckkods läsaren till QuikRead Instrumentet. Kabeln ska anslutas till den
runda kontakten som nns på baksidan av QuikRead Instrumentet. Se Figur
2 för anslutningarna. För att läsa en streckkod, rikta läsaren över en streckkod och tryck in av-
tryckaren som nns i handtaget. Avläsningen utförs längs mitten av det ljusfält
som avges från avläsningsfönstret. Ljusfältet måste täcka hela streckkoden horisontellt. Rikta inte ljusfältet direkt mot ögonen. Tillverkargarantin för streckkodsläsaren täcker material och tillverkningsfel i 5 år från leveransdatum, förutsatt att produkten används på ett riktigt sätt under normala förhållanden. I händelse av att streckkodsläsaren inte skulle fungera kontakta din lokala återförsäljare för information om auktoriserad service.
16
Figur 2
B
Anslutning av streckkodsläsare
Seriell kabel
A
Kontakt till streckkodsläsaren
B
Streckkodsläsarens anslutnings kabel
C
Strömadapter
D
A
D
C
MÄTNINGSPROCEDUR
Att läsa patient-ID
1. Placera streckkodsläsaren över streckkoden på remissen eller provröret
och tryck in knappen i handtaget. Vänta på en ljud signal.
Om avläsningen av streckkoden misslyckas upprepade gånger, se Fel-
sökning.
2. Displayen visar patient-ID. Försäkra dig om att det överens stämmer med patient-ID på remissen.
Klar att använda <test> Patient <xxxxxxxx>
Läs användar-ID
(Valbar inställning. Om den inte är påslagen gå till steg 6.)
3. Om användar-ID är påslaget visar displayen:
Klar att använda <test> Läs användar-ID
4. Läs användar-ID med streckkodsläsaren. Displayen visar användar-ID. Försäkra dig om att det överensstämmer med läst användar-ID.
Användar-ID <oooooooo>
5. Displayen visar sen:
Klar att använda <test> Patient <xxxxxxxx>
Mätning
● Med inställningen Godkänn analyssvar innan sändning aktiverad gå vidare till steg 6
● Med inställningen Godkänn analyssvar innan sändning avaktiverad gå vidare till steg 14
Mätning med inställningen Godkänn analyssvar innan sänd­ning aktiverad
6. Utför mätningen genom att följa användarinstruktionerna för QuikRead Instrumentet, kitinlagan och texten i diplayen.
7. När mätningen är slutförd visar displayen texten:
Patient <xxxxxxxx> Resultat <zz> <uuu>
där <xxxxxxxx>är patient-ID, <zz> är resultatet, och <uuu> är analyten.
Acceptera eller
och
förkasta resultat
17
Acceptera eller förkasta resultat
8. För att acceptera ett resultat, läs streckkoden OK/ACCEPT på streck-
kodstavlan för QuikLink (streckkoderna nns också på sidan 2).
För att förkasta en mätning, läs streckkoden CANCEL/REJECT på
streckkodstavlan för QuikLink.
9. När resultatet accepteras visar displayen:
Analys godkänd Ta bort kyvett
Om resultatet förkastas visar displayen:
Analys ej godkänd Ta bort kyvett
10. Lyft upp kyvetten och invänta att QuikRead Instrumentet utför en själv­kontroll.
Sändning av resultat till LIS/HIS dator
11. Efter en lyckad självkontroll sänds automatiskt det accepterade resul-
tatet till LIS/HIS datorn. Detta kan ta några sekunder. Under tiden visar
displayen:
Sänder resultat...
Om resultatet förkastats sänds det inte till LIS/HIS datorn. Om självkon­trollen misslyckas sänds inte resultatet och ett felmeddelande visas i displayen. Radera resultatet.
12. När överföringen av resultatet lyckats visar displayen:
Lyckar sändning av analysresultat
13. Om QuikRead Instrumentet inte kan överföra resultatet till LIS/HIS datorn
visar displayen texten:
Misslyckar sändning! Skriv ut svaret?
Patient <xxxxxxxx>
och
Resultat <zz> <uuu>
Resultatet kan dokumenteras manuellt eller sändas till en skrivare. För att
sända resultatet till en skrivare ska skrivaren först anslutas till QuikRead Instrumentet och OK/ACCEPT läsas på QuikLink streckstreckkodstavlan. Se Felsökning.
Om resultatet dokumenteras manuellt, måste felmeddelandet erkännas
genom att CANCEL/REJECT läses på QuikLink streckkodstavla.
I båda fallen raderas resultatet från instrumentets interna minne och
QuikRead är klar för nästa mätning.
18
Mätning med inställningen Godkänn analyssvar innan sänd- ning avaktiverad
14. Utför mätningen genom att följa användarinstruktionerna för QuikRead
Instrumentet, kitinlagan och texten i diplayen.
15. När mätningen är avslutad visar displayen omväxlande:
Patient <xxxxxxxx> Resultat <zz> <uuu>
Ta bort kyvetten för att
och
starta LIS/HIS sändning
där <xx> är löpnummer, <yyyyyy> är aktuellt datum, <zz> är resultatet
och <uuu> är analyten.
16. Ta upp kyvetten och invänta att QuikRead Instrumentet utför en själv­kontroll.
Sändning av resultat till LIS/HIS dator
17. Efter en lyckad självkontroll sänds automatiskt det accepterade resul-
tatet till LIS/HIS datorn. Detta kan ta några sekunder. Under tiden visar
displayen:
Sänder resultat...
Om resultatet förkastats sänds det inte till LIS/HIS datorn. Om själv-
kontrollen misslyckas sänds inte resultatet och ett felmeddelande visas i displayen. Radera resultatet.
18. När överföringen av resultatet lyckats visar displayen omväxlande:
Klar att använda <test> lot xxx
Klar att använda
och
Läs patient-ID
Instrumentet är klart för nästä mätning. Om QuikRead Instrumentet inte kan överföra resultat till LIS/HIS datorn
visar displayen omväxlande texten:
Misslyckad sändning! Skriv ut svaret?
Patient <xxxxxxxx>
och
Resultat <zz> <uuu>
Resultatet kan dokumenteras manuellt eller sändas till en skrivare. För att
sända resultatet till en skrivare ska skrivaren först anslutas till QuikRead Instrumentet och OK/ACCEPT läsas på QuikLink streckstreckkodstavlan. Se Felsökning.
Om resultatet dokumenteras manuellt, måste felmeddelandet erkännas
genom att CANCEL/REJECT läses på QuikLink streckkodstavla.
I båda fallen raderas resultatet från instrumentets interna minne och
QuikRead är klar för nästa mätning.
Kort instruktion för användning av QuikLink läge med streckkodsläsare
1.
Läs patient-ID.
2.
Läs användar-ID (valbart).
3.
Utför mätningen.
4.
Acceptera resultatet (valbart).
19
Mätning i tillfälligt NON-LIS läge
Försäkra dig om att QuikRead Instrumentet är i inaktivt läge, dvs att displayen
omväxlande visar:
Klar att använda <test> lot xxx
Klar att använda
och
Läs patient-ID
INSTÄLLNING AV INSTRUMENTET FÖR TILLFÄLLIGT NON-LIS LÄGE
Läs Temporary NON-LIS mode streckkoden på QuikLink streck kodstavla.
Displayen visar omväxlande:
Klar att använda <test> lot
Klar att använda
och
Ej inställd för LIS
För att avsluta Tillfälligt NON-LIS läge utan att analysera något prov, läs Exit temporary NON-LIS mode streckkoden på QuikLink streckkodstavla.
MÄTNING
1. Utför mätningen genom att följa instruktionerna för instrumentet, kitinlagan
och texten i displayen.
2. Dokumentera resultatet innan kyvetten tas upp från mätbrunnen.
3. Efter en lyckad analys återvänder QuikRead Instrumentet automatiskt till QuikLink läge med streckkodsläsare. Om en analys misslyckas (test återkallat), återvänder QuikRead Instrumentet till början av Tillfälligt NON-LIS läge.
Mätning i QuikLink läge utan streckkodsläsare
UPPSTART
Sätt igång QuikRead Instrumentet.
Försäkra dig om att displayen visar:
Inställd för QuikLink streckkodsavläsare av
Använd ett magnetkort för att välja analyt. Displayen visar:
Klar att använda <test> lot xxxx
Notera att magnetkortet bara kan läsas när displayen visar Läs kortet
eller Klar att använda.
20
MÄTNINGSPROCEDUR
Mätning
1. Utför mätningen genom att följa instruktionerna för QuikRead Instrumentet,
kitinlagan och texten i displayen.
2. När mätningen är avslutad visar displayen omväxlande:
Prov <xx> / <yyyyyy> Resultat <zz> <uuu>
Ta bort kyvetten för att
och
starta LIS/HIS sändning
där <xx> är löpnummer, <yyyyyy> är aktuellt datum, <zz> är resultatet
och <uuu> är analyten.
3. Ta upp kyvetten och invänta att QuikRead Instrumentet utför en själv­kontroll.
Sändning av resultat till LIS/HIS dator
4. Efter en lyckad självkontroll sänder QuikRead Instrumentet automatiskt
resultatet till LIS/HIS datorn. Detta kan ta några sekunder. Under tiden
visar displayen:
Sänder resultat...
Om självkontrollen misslyckas sänds inte resultatet och ett felmeddelande visas i displayen Radera resultatet.
5. När överföringen av resultatet lyckats visar displayen omväxlande tex-
ten:
Föregaende sändning OK
Instrumentet är klart för nästa mätning. Om QuikRead Instrumentet inte kan överföra resultat till LIS/HIS datorn
visar displayen omväxlande texten:
Föregaende sändning misslyckades!
Prov <xx> / <yyyyyy> Resultat <zz> <uuu>
Läs analytkort för att ga vidare
Analysresultatet ska dokumenteras manuellt. Använd aktuell analyts
magnetkort för att fortsätta. Resultatet raderas från instrumentets interna minne och QuikRead Instrumentet är klart för nästa mätning. Se Felsök­ning.
Klar att använda
och
<test> lot xxxx
och
och
21
FELSÖKNING
Fel område Typ av fel Felmeddelande
Streckkodsläsning Streckkodsläsarens signal
uteblir efter intryckning.
Ett ljudsignal hörs men inget meddelande visas i displayen.
Felmeddelande i displayen.
Felaktig patient eller användar-ID visas i displayen.
Dataöverföring Felmeddelande på
QuikRead Instrumentets display.
Data kan inte överföras till LIS/HIS datorn.
Felaktiga data til LIS/HIS datorn.
Fel vid läsning av streckkod. Försök igen.
Misslyckad sänd ning! Skriv ut resultatet?
22
Trolig orsak Korrigerande åtgärd
Streckkodsläsarens anslutningskabel har dålig kontakt.
Streckkodsläsaren är felaktigt
kongurerad.
Olämplig streckkodsläsare Välj rätt modell av streckkods läsare.
Defekt streckkodsläsare Kontakta din lokala leverantör.
Dålig streckkodskvalitet Använd en ny etikett.
Oanvändbar streckkodstyp Kontrollera att installations programmet
QuikRead Instrumentet
är felaktigt kongurerat.
Streckkodsläsarens anslutnings kabel har dålig kontakt.
Strekkodsläsarens kabel är defekt.
Ett ertal orsaker Läs igen. Om meddelandet kvarstår,
Se till att kabeln är korrekt ansluten.
Använd installationsprogrammet för
QuikRead QuikLink för att kongurera på
nytt.
stöder streckkods typen som används.
Använd installationsprogrammet för
QuikRead QuikLink för att kongurera på
nytt.
Se till att kabeln är korrekt ansluten.
Kontakta din lokala leverantör.
kontakta din lokala leverantör.
Misslyckad streckkodsläsning Rikta läsaren mer noggrant över streck-
Streckkodsläsaren stöder inte den streckkodstyp som används för patient eller användar-ID och läser därför koden felaktigt.
Kabeln till LIS/HIS datorn är felaktigt ansluten.
Kabeln till LIS/HIS datorn är defekt.
LIS/HIS datorn svarar inte Kontrollera att datorn är OK.
Felaktig version av drivrutin hos LIS/HIS datorn
Annan orsak
Felaktig version av drivrutin hos LIS/HIS datorn
Annan orsak
Se också QuikRead 101 Instrument bruksanvisning, Appendix C:
Felsökningsguiden / Felmeddelanden.
koden. Se till att ljusfältet täcker hela koden horisontellt.
Kontrollera att installationsprogrammet stöder streckkods typen som används.
Anslut kabeln.
Byt kabel.
Kontakta din lokala administratör.
Kontakta din lokala administratör.
23