Quantum Data TPA-MHL-8R Quick Start Manual

Page 1
TPA-MHL-8R - Quick Start Guide 68-00224 - Rev. A4
Page 1
August 31, 2011
Quick Start Guide
Rev: A4
Page 2
TPA-MHL-8R - Quick Start Guide 68-00224 - Rev. A4
Page 2
August 31, 2011
Table of Contents
1 Overview of MHL 3 2 Quantum Data MHL Solutions 3
2.1 What is in the TPA-MHL-8R shipping box 6
2.2 Scope of this Quick Start Guide 6
2.3 Power Considerations 7
3 MHL Applications for Testing MHL Source Devices 8
3.1 Functional testing of MHL sources 8
3.2 Basic Protocol Testing of MHL Sources 10
3.3 Protocol Analysis of MHL Sources 12
3.4 MHL Source Compliance Testing 16
3.5 HDCP compliance testing for source devices 18
3.6 C-BUS Protocol Monitoring of MHL Source and MHL Sinks 20
4 MHL Applications for Testing MHL Sink Devices 22
4.1 MHL Sink Functional Testing 22
4.2 MHL Sink Compliance Testing 24
4.3 HDCP compliance testing for sink devices 26
Page 3
TPA-MHL-8R - Quick Start Guide 68-00224 - Rev. A4
Page 3
August 31, 2011
1 Overview of MHL
Mobile High-Definition Link (MHL) is a video interface derived from HDMI but targeted to provide an interface from mobile devices like smart phones to TVs or other displays. The MHL standard is designed to connect to smart phones by reusing the existing 5 pin micro-USB connector that many phones have rather than requiring a separate 19 pin HDMI connector. The MHL cable is micro-USB on the source mobile device and HDMI Type A on the other end.
2 Quantum Data MHL Solutions
Quantum Data offers a comprehensive suite of test solutions for testing MHL source and sink devices. The centerpiece of these solutions is a test point adapter (TPA-MHL-8R) that converts MHL streams to HDMI and HDMI streams to MHL. The TPA-MHL-8R facilitates basic sink emulation, functional testing, protocol analysis as well as HDMI compliance testing and HDCP compliance testing for MHL source and sink devices. Protocol analysis and HDMI Compliance testing requires the Quantum Data 980 Protocol Analyzer (297MHz module). HDCP compliance testing requires the Quantum Data 882EA.
Page 4
TPA-MHL-8R - Quick Start Guide 68-00224 - Rev. A4
Page 4
August 31, 2011
The rear of the TPA-MHL-8R is shown below. There are three application sections with associated connectors. The connections for each are described in the table below the diagram.
TPA-MHL-8R Section
Connector Label
Connection
980 CONNECTONS SOURCE DUT
OUT TO 980 PORT 1
Connect HDMI cable to the Rx port on 980 MHL Protocol Analyzer.
MHL IN FROM DUT
Connect MHL cable to MHL mobile device (MHL source).
882 CONNECTIONS SOURCE DUT
HDMI OUT
This is an HDMI output. The device it gets connected to depends on the source test application.
For basic functional testing of an MHL source -
An HDMI cable is connected to the HDMI IN port on an HDTV to monitor the video and audio.
For basic protocol testing of an MHL source – An
HDMI cable is connected to an HDMI input on the 882EA or 780 to check the timing and protocol data or to check the response to various EDIDs.
For HDCP compliance testing of an MHL source
– An HDMI cable is connected to a port on the Encryption Status Tester (EST).
MHL IN FROM DUT
Connect MHL cable from MHL source device under test, e.g. mobile phone.
Page 5
TPA-MHL-8R - Quick Start Guide 68-00224 - Rev. A4
Page 5
August 31, 2011
TPA-MHL-8R Section
Connector Label
Connection
882 CONNECTIONS SINK DUT
HDMI IN
This is an HDMI input used for HDCP compliance testing of MHL sink devices. An HDMI cable is connected to the Encryption Status Tester (EST) used with the 882EA HDMI output. For basic functional testing of an MHL sink device, an HDMI source devices or 780 Handheld Test Instrument is connected to this port.
MHL OUT TO DUT
Connect MHL cable to the MHL sink device under test, e.g. HDTV.
980 Connections CBUS ANALYSIS
MHL IN FROM SOURCE
Connect MHL cable from MHL source device under test, e.g. mobile phone.
MHL OUT TO SINK
Connect MHL cable to the MHL sink device under test, e.g. HDTV.
TO 980 TRIGGER IN CBUS
Connect coax cable to TRIG IN port on 980 Protocol Analyzer.
Page 6
TPA-MHL-8R - Quick Start Guide 68-00224 - Rev. A4
Page 6
August 31, 2011
2.1 What is in the TPA-MHL-8R shipping box
The TPA-MHL-8R instrument shipping container includes the items listed in the table below:
Table 1-1: TPA-MHL-8R Shipping Box Contents
Item
Part No.
TPA-MHL-8R
00-00230
Power Adapter assembly
12V DC power supply
25-00094
Line cord
30A00400A03
MHL Cable Kit
Contents
78-00061
RG590A BNC cable
30-00125
(2) MHL Cables
30-00189
HDMI Cable
30-00146
Quick Start Guide
68-00224
2.2 Scope of this Quick Start Guide
This Quick Start Guide provides application information for testing your MHL source or MHL sink device. Procedures for connecting up the TPA-MHL-8R are provided in this Guide. Procedures for basic functional testing and protocol testing are available in the 780 User Guide. Procedures for MHL protocol analysis and MHL compliance testing for MHL sources and sinks are provided with the 980 Protocol Analyzer User Guide. Detailed procedures for testing HDCP on MHL sources and sinks are provided in the 882EA User Guide.
Page 7
TPA-MHL-8R - Quick Start Guide 68-00224 - Rev. A4
Page 7
August 31, 2011
2.3 Power Considerations
The TPA-MHL-8R is powered through a 12VDC source jack. The TPA-MHL-8R is supplied with a12V DC power supply and charger and line cord.
Page 8
TPA-MHL-8R - Quick Start Guide 68-00224 - Rev. A4
Page 8
August 31, 2011
3 MHL Applications for Testing MHL Source Devices
This section describes the MHL test applications supported by Quantum Data test instruments and utilities when testing MHL source devices.
3.1 Functional testing of MHL sources
You can conduct basic functional testing to verify an MHL source device using the TPA­MHL-8R and a standard HDMI HDTV. Basic functional testing involves monitoring the video and audio on an HDMI sink device and verifying it against the expected video resolution and audio format based on the EDID of the HDTV.
MHL Cable
HDMI Cable
TPA-MHL-8R
HDMI Display
Source
MHL DUT
Page 9
TPA-MHL-8R - Quick Start Guide 68-00224 - Rev. A4
Page 9
August 31, 2011
Use the procedure below to make the connections to the TPA-MHL-8R for conducting basic functional tests of an MHL source device.
1. Apply power to the TPA-MHL-8R using the power adapter provided.
2. Make the connections as shown in the diagram below. For this application you will be using the section on the TPA-MHL-8R labeled 882 CONNECTIONS – SOURCE DUT.
a. Connect an MHL cable from the MHL source device under test to the MHL IN FROM
DUT connector on the TPA-MHL-8R.
b. Connect an HDMI cable from the HDMI OUT connector of the TPA-MHL-8R to the
HDMI IN connector on the HDMI HDTV.
3. Monitor the HDTV for proper operation.
Connect MHL
cable to MHL
mobile device
(MHL source)
Connect HDMI
cable to HDMI
IN port on
HDMI HDTV
Page 10
TPA-MHL-8R - Quick Start Guide 68-00224 - Rev. A4
Page 10
August 31, 2011
3.2 Basic Protocol Testing of MHL Sources
You can conduct basic protocol testing of MHL source devices using the TPA-MHL-8R and an HDMI analyzer such as the Quantum Data 882EA or 780 Handheld Test Instrument. Basic protocol analysis enables you to view snapshots of the MHL timing, video type and metadata (e.g. infoframes and data islands).
For more details on the analysis features and functions of the 882EA or 780, please refer to the Quantum Data products web page at:
www.quantumdata.com/products/index.asp. Refer to the diagram below for a depiction
of the test setup.
MHL Cable
HDMI Cable
780 HDMI Tester
Source
MHL DUT
TPA-MHL-8R
Page 11
TPA-MHL-8R - Quick Start Guide 68-00224 - Rev. A4
Page 11
August 31, 2011
Use the procedure below to make the connections to the TPA-MHL-8R for running basic protocol tests on an MHL source device.
1. Apply power to the TPA-MHL-8R using the power adapter provided.
2. Make the connections as shown in the diagram below. For this application you will be using the section on the TPA-MHL-8R labeled 882 CONNECTIONS – SOURCE DUT.
a. Connect an MHL cable from the MHL source device under test to the MHL IN FROM
DUT connector on the TPA-MHL-8R.
b. Connect an HDMI cable from the HDMI OUT connector of the TPA-MHL-8R to the
HDMI IN port of the 780 or 882EA.
3. Refer to the 780 or 882EA User Guide for details on how to conduct basic protocol and timing tests on an HDMI source device.
Connect MHL
cable to MHL
mobile device
(MHL source)
Connect HDMI
cable to HDMI
IN port on 780
or 882EA
Page 12
TPA-MHL-8R - Quick Start Guide 68-00224 - Rev. A4
Page 12
August 31, 2011
3.3 Protocol Analysis of MHL Sources
You can conduct protocol analysis of MHL source devices using the TPA-MHL-8R and the Quantum Data 980 Protocol Analyzer. Protocol analysis enables you to view the timing, video type, control data and all metadata (e.g. infoframes and data islands) with precise time stamps in real time or capture and store for analysis. You can also view the MHL C-Bus traffic.
For more details on the MHL protocol analysis functions of the 980, please refer to the 980 User Guides available on the Quantum Data 980 product web page at:
www.quantumdata.com/products/980.asp. Refer to the diagram below for a depiction of
the test setup.
MHL Cable
HDMI Cable
Source
MHL DUT
980 Protocol Analyzer
Real Time Monitoring
TPA-MHL-8R
980 Protocol Analyzer
Capture, Store &
Page 13
TPA-MHL-8R - Quick Start Guide 68-00224 - Rev. A4
Page 13
August 31, 2011
Use the procedures and diagram below to make the connections to the TPA-MHL-8R for running protocol analysis of an MHL source device.
1. Apply power to the TPA-MHL-8R using the power adapter provided.
2. Make the connections as shown in the diagram below. For this application you will be using the section on the TPA-MHL-8R labeled 980 CONNECTIONS – SOURCE DUT.
a. Connect an MHL cable from the MHL source device under test to the MHL IN FROM
DUT connector on the TPA-MHL-8R.
b. Connect an HDMI cable from the OUT TO 980 PORT 1 connector of the TPA-MHL-8R
to the Rx port on the 980 Protocol Analyzer.
3. Refer to the 980 User Guide for details on how to capture, store and analyze data from the MHL source.
Connect MHL cable to MHL mobile device (MHL source)
Connect HDMI cable to Rx Port on 980 Protocol
Analyzer
Page 14
TPA-MHL-8R - Quick Start Guide 68-00224 - Rev. A4
Page 14
August 31, 2011
4. Optionally, if you wish to monitor the C-Bus transactions while monitoring the MHL stream make the following connections and refer to the diagram that follows as a reference:
a. Connect an MHL cable from the MHL source device under test to the MHL IN FROM
SOURCE connector on the section of the TPA-MHL-8R labeled 980 CONNECTIONS – CBUS ANALYSIS.
b. Connect a Coax cable from the connector labeled TO 980 TRIGGER IN CBUS on the
section of the TPA-MHL-8R labeled to 980 CONNECTIONS – CBUS ANALYSIS to the BNC connector labeled TRIG IN on the 980 Protocol Analyzer.
c. Connect a second MHL cable from the MHL OUT TO SINK connector on the section of
the TPA-MHL-8R labeled 980 CONNECTIONS – CBUS ANALYSIS to the MHL IN FROM DUT connector on the section of the TPA-MHL-8R labeled 980 CONNECTIONS SOURCE DUT.
d. Connect an HDMI cable from the connector labeled OUT TO 980 PORT 1 on the
section of the TPA-MHL-8R labeled 980 CONNECTIONS – SOURCE DUT to the Rx port on the 980 Protocol Analyzer.
Page 15
TPA-MHL-8R - Quick Start Guide 68-00224 - Rev. A4
Page 15
August 31, 2011
Page 16
TPA-MHL-8R - Quick Start Guide 68-00224 - Rev. A4
Page 16
August 31, 2011
3.4 MHL Source Compliance Testing
You can run MHL compliance testing on source devices in accordance with Section 3 of the MHL Compliance Test Specification using the TPA-MHL-8R and the 980 HDMI Protocol and Compliance Tester.
Note: You must purchase the MHL source compliance test option as part of the 980 Protocol Analyzer.
For more details on the MHL Source Compliance Test functions of the 980, please refer to the 980 User Guides available on the Quantum Data 980 product web page at:
www.quantumdata.com/products/980.asp. Refer to the diagram below for a depiction of
the test setup.
980 Protocol Analyzer
& Compliance Tester
Source
MHL DUT
TPA-MHL-8R
MHL Cable
HDMI Cable
Page 17
TPA-MHL-8R - Quick Start Guide 68-00224 - Rev. A4
Page 17
August 31, 2011
Use the procedure below to make the connections the TPA-MHL-8R for running source compliance tests on an MHL source device.
1. Apply power to the TPA-MHL-8R using the power adapter provided.
2. Make the connections as shown in the diagram below. For this application you will be using the section on the TPA-MHL-8R labeled 980 CONNECTIONS – SOURCE DUT.
a. Connect an MHL cable from the MHL source device under test to the MHL IN FROM
DUT connector on the TPA-MHL-8R.
b. Connect an HDMI cable from the OUT TO 980 PORT 1 connector of the TPA-MHL-8R
to the Rx port on the 980 Protocol Analyzer.
3. Refer to the 980 MHL Compliance Test User Guide for details on how to run the MHL compliance test on an MHL source.
Connect MHL cable to MHL mobile device (MHL source)
Connect HDMI cable to Rx Port on 980 Protocol
Analyzer
Page 18
TPA-MHL-8R - Quick Start Guide 68-00224 - Rev. A4
Page 18
August 31, 2011
3.5 HDCP compliance testing for source devices
You can run HDCP compliance tests on MHL source devices in accordance with the HDCP 1.2 Compliance Test Specification. This test requires TPA-MHL-8R and 882EA with HDCP compliance test option.
For more details on the HDCP source compliance test feature of the User Guide for the 882EA Video Test Instrument available on the Quantum Data 882E product web page at: www.quantumdata.com/products/882E.asp.
EST 882EA
TPA-MHL-8R
Source
MHL DUT
HDMI Cable
MHL Cable
Page 19
TPA-MHL-8R - Quick Start Guide 68-00224 - Rev. A4
Page 19
August 31, 2011
Use the procedure below to make the connections to the TPA-MHL-8R for running the HDCP compliance tests on an MHL source device.
1. Apply power to the TPA-MHL-8R using the power adapter provided.
2. Make the connections as shown in the diagram below. For this application you will be using the section on the TPA-MHL-8R labeled 882 CONNECTIONS – SOURCE DUT.
a. Connect an MHL cable from the MHL source device under test to the MHL IN FROM
DUT connector on the TPA-MHL-8R.
b. Connect an HDMI cable from the HDMI OUT connector of the TPA-MHL-8R to the
appropriate connector on the Encryption Status Tester (EST) and then to the 882EA HDMI IN connector.
3. Refer to the 882EA User Guide for details on how to run HDCP compliance tests on a source device.
Connect MHL
cable to MHL
mobile device
(MHL source)
Connect HDMI
cable to the
EST used
with the 882E
Page 20
TPA-MHL-8R - Quick Start Guide 68-00224 - Rev. A4
Page 20
August 31, 2011
3.6 C-BUS Protocol Monitoring of MHL Source and MHL Sinks
You can passively monitor the C-BUS transactions between an MHL source device and an MHL sink device using the TPA-MHL-8R and the Quantum Data 980 Protocol Analyzer. Passively monitoring the C-BUS enables you to view the EDID exchange and HDCP authentication transactions in real time.
For more details on the analysis features and functions of the 980, please refer to the Quantum Data products web page at: www.quantumdata.com/products/980.asp. Refer to the diagram below for a depiction of the test setup.
Use the procedure below to make the connections to the TPA-MHL-8R for passively monitoring the C-Bus messages between an MHL source device and an MHL sink device.
1. Apply power to the TPA-MHL-8R using the power adapter provided.
2. Make the connections as shown in the diagram below. For this application you will be using the section on the TPA-MHL-8R labeled 980 CONNECTIONS – CBUS ANALYSIS.
TPA-MHL-8R
MHL Cable
BNC Coax cable
Source
MHL DUT
980 Protocol Analyzer
Real Time Mode
MHL HDTV
Page 21
TPA-MHL-8R - Quick Start Guide 68-00224 - Rev. A4
Page 21
August 31, 2011
a. Connect an MHL cable from the MHL source device under test to the MHL IN FROM
SOURCE connector on the TPA-MHL-8R.
b. Connect a second MHL cable from the MHL OUT TO SINK connector on the TPA-
MHL-8R to the MHL sink device.
c. Connect a Coax cable from the connector labeled TO 980 TRIGGER IN CBUS on the
TPA-MHL-8R to the connector labeled TRIG IN on the 980 Protocol Analyzer.
3. Refer to the Real Time monitoring section of the 980 User Guide for details on how to monitor the MHL C-Bus messages.
Connect MHL
cable to MHL
mobile device
(MHL source)
Connect MHL cable to MHL-capable
HDTV
Connect cable to TRIG IN port on 980 Protocol Analyzer
Page 22
TPA-MHL-8R - Quick Start Guide 68-00224 - Rev. A4
Page 22
August 31, 2011
4 MHL Applications for Testing MHL Sink Devices
This section describes the MHL test applications supported by Quantum Data test instruments and utilities for testing MHL sink devices.
4.1 MHL Sink Functional Testing
You can run a functional test on an MHL sink using the TPA-MHL-8R and the 780 HDMI Handheld Test Instrument.
For more details on the HDMI test functions of the 780, please refer to the 780 User Guide available on the Quantum Data 780 product web page at:
www.quantumdata.com/products/780.asp. Refer to the diagram below for a depiction of
the test setup.
Use the procedure below to make the connections for running the compliance tests on an MHL sink device.
MHL Cable
HDMI Cable
MHL Sink DUT
780 HDMI Source
TPA-MHL-8R
Page 23
TPA-MHL-8R - Quick Start Guide 68-00224 - Rev. A4
Page 23
August 31, 2011
1. Apply power to the TPA-MHL-8R using the power adapter provided.
2. Make the connections as shown in the diagram below. For this application you will be using the section on the TPA-MHL-8R labeled 882 CONNECTIONS – SINK DUT.
a. Connect an MHL cable from the MHL sink device under test to the MHL OUT TO DUT
connector on the TPA-MHL-8R.
b. Connect an HDMI cable from the HDMI IN connector of the TPA-MHL-8R to the HDMI
OUT port on the 780 Handheld Test Instrument.
3. Refer to the 780 User Guide for details on how to run functional tests on the MHL/HDMI sink devices.
Connect MHL cable to MHL HDTV (MHL sink)
Connect HDMI
cable to the
HDMI Out Port
of the 780
Page 24
TPA-MHL-8R - Quick Start Guide 68-00224 - Rev. A4
Page 24
August 31, 2011
4.2 MHL Sink Compliance Testing
You can run MHL compliance testing on MHL sink devices in accordance with Section 4 of the MHL Compliance Test Specification using the TPA-MHL-8R, 882EA and the 980 HDMI Protocol and Compliance Controller.
Note: You will have to purchase the MHL Sink Compliance Test option as part of the 980 Protocol Analyzer to run this test series.
For more details on the MHL Sink Compliance Test functions of the 980, please refer to the 980 User Guide for MHL compliance testing available on the Quantum Data 980 product web page at: www.quantumdata.com/products/980.asp. Refer to the diagram below for a depiction of the test setup.
Page 25
TPA-MHL-8R - Quick Start Guide 68-00224 - Rev. A4
Page 25
August 31, 2011
Use the procedure below to make the connections for running the compliance tests on an MHL sink device.
1. Apply power to the TPA-MHL-8R using the power adapter provided.
2. Make the connections as shown in the diagram below. For this application you will be using the section on the TPA-MHL-8R labeled 882 CONNECTIONS – SINK DUT.
a. Connect an MHL cable from the MHL sink device under test to the MHL OUT TO DUT
connector on the TPA-MHL-8R.
b. Connect an HDMI cable from the HDMI IN connector of the TPA-MHL-8R to the HDMI
OUT port on the 882E.
3. Refer to the 980 User Guide for details on how to use the MHL compliance test application to run compliance tests on an MHL sink.
Connect MHL cable to MHL HDTV (MHL sink)
Connect HDMI
cable to the
HDMI Out Port
of the 882EA
Page 26
TPA-MHL-8R - Quick Start Guide 68-00224 - Rev. A4
Page 26
August 31, 2011
4.3 HDCP compliance testing for sink devices
You can run HDCP compliance tests on MHL sink devices in accordance with the HDCP 1.2 Compliance Test Specification. This test requires TPA-MHL-8R and 882EA with HDCP compliance test option.
For more details on the HDCP source compliance test feature of the User Guide for the 882EA Video Test Instrument available on the Quantum Data 882E product web page at: www.quantumdata.com/products/882E.asp.
EST
882E
Sink MHL DUT TPA-MHL-8R
HDMI Cable
MHL Cable
Page 27
TPA-MHL-8R - Quick Start Guide 68-00224 - Rev. A4
Page 27
August 31, 2011
1. Apply power to the TPA-MHL-8R using the power adapter provided.
2. Make the connections as shown in the diagram below. For this application you will be using the section on the TPA-MHL-8R labeled 882 CONNECTIONS – SINK DUT.
a. Connect an MHL cable from the MHL sink device under test to the MHL OUT TO DUT
connector on the TPA-MHL-8R.
b. Connect an HDMI cable from the HDMI IN connector of the TPA-MHL-8R to the
appropriate port on the Encryption Status Tester (EST) and then to the 882EA HDMI OUT connector.
3. Refer to the 882EA User Guide for details on how to run HDCP compliance tests on a sink device.
Connect MHL cable to MHL HDTV (MHL sink)
Connect HDMI
cable to the
EST used with
the 882EA